LF Conductivity Measuring Probe

Size: px
Start display at page:

Download "LF Conductivity Measuring Probe"

Transcription

1 LF Conductivity Measuring Probe Contents 1 General 1.1 Introduction 1.2 Application range 2 Safety 2.1 Warnings and symbols 3 Transport and storage 4 Application 5 Measuring principle 6 Assembly 7 Construction 8 Technical data 8.1 Explosion protection 9 Calibration 10 Startup 10.1 Electrical connection 11 Spare parts 12 Related documents 13 Glass Testing 14 Avoiding damages to peripheral devices Annex 1 EC-Type-Examination-Certificate Annex 2 Declaration of conformity Operating Instructions e

2 LF Conductivity Measuring Probe 1 General 1.1 Introduction The probe may only be installed, started up and serviced by authorized personnel with special skills in measuring technology and in strict compliance with the present instructions as well as the valid provisions. Non-compliance with these provisions no matter whether by intention or by negligence shall release the manufacturer from all responsibility and reliability. The present operating instructions are designed to familiarize users with the equipment of the probe and its use. The operating instructions should be available to the operating and maintenance personnel and should be studied before performing any installation and/or maintenance work. By knowing these operating instructions, you can avoid damages to the measuring equipment and ensure trouble-free operation. The representations and data given in these Operating Instructions are subject to modification and further improvement. The latest edition hereof will always supersede all previous ones. 1.2 Application range LF probes are used to measure the conductivity of changes in conductivity. Never operate this measuring device outside its permissible operating conditions. 2 Safety The safety instructions refer exclusively to the scope of delivery. Please also note the valid safety regulations concerning electrical installations and facilities. 2.1 Warnings and symbols In the operating instructions, the danger symbol is used to draw your attention to especially important information. m Compliance with these mandatory instructions helps to avoid personal injury and damage to property. m Please note the following instructions: Do not practice any working methods which may endanger safety! 3 Transport and storage The probe should only be transported and stored in its closed original packaging. In order to guarantee an as-new condition of the probe, maintain the following storage conditions: dry and dust-free, steady temperature and ventilation When used probes are sent to the manufacturer or third parties for repair or other purposes, all parts must be cleaned and decontaminated. Appropriate safety instructions for further handling must be attached to the outside of the packaging. The probes do not need any preservatives, they are resistant to normal environmental influences. Figure 1 LF measuring probes 4 Application The fully glasslined LF measuring probe is used to measure the electrolytic conductivity and can be installed directly inside reactors or pipes. For the resistance of the glass lining used, please refer to our publication no The LF probe is preferably used for measurement in concentrated acids at high temperatures and high pressures up to 40 bar (special option up to 100 bar). 2 Pfaudler GmbH

3 LF Conductivity Measuring Probe 5 Measuring principle The 4-electrode method is used. A constant AC current flows through the external electrodes. The measuring-circuit voltage is picked up in the form of a resistancedependent voltage drop across the internal electrodes without current load. The voltage value is then supplied to a transmitter with a high-resistance input. This measuring method is largely insensitive to contamination of the electrodes. Polarization errors are impossible because polarization only occurs at the current electrodes. A built-in Pt 100 resistance thermometer compensates for the temperature dependence of the conductivity. By combining steel, glass and rhodium (for the electrodes) as materials for the probe, the LF probe was given a number of outstanding characteristics: high resistance to corrosion large temperature application range between C resistance to pressure: bar (option 100 bar) resistance to thermal shock Figure 2 LF measuring probe in pipes Rhodium electrodes equipotential bonding conductor, if necessary U R i = i Due to these properties, it is even suitable for extreme applications. 6 Assembly When measuring in pipes made of plastic materials or steel with an insulating lining, the use of conductive gaskets offers advantages. These gaskets are connected to the equipotential bonding system of the probe. DN50 part number DN80 part number Glass l k U Pt 100 Metal body Process conductometer ms/cm % by weight ma LF When installing the probe in steel pipes, only normal, non-conductive gaskets may be used due to the possible formation of electrolytic element couples between components. DN50 part number DN80 part number Figure 3 Structure of the LF measuring probes For the tightening torques of the split flange connection, please refer to Service Instructions 318. After the first temperature cycle (but by no means later than after 24 hours) the screws must be retightened with the prescribed torque. Pfaudler GmbH 3

4 LF Conductivity Measuring Probe 7 Construction The four metal electrodes with a diameter of 2 mm have been fused into the glass layer of a measuring probe carrier so as to ensure insulation between them. Measuring probe carriers may include baffles, thermometer wells or spacing rings. The supply conductors for the electrodes are also fused into glass and lead to a terminal box at the flange of the probe carrier. The Pt 100 has been integrated into the steel body of the probe carrier. Baffles and thermometer wells are available in any desired size, spacing rings may be manufactured from DN ms/cm 1 g/l 0,1 1 10,05 100,6 0,1 0,01 0, % by weight Figure 4 Concentration table H 2 SO 4 20 C 8 Technical data Conductivity range: Temperature range: Pressure: Electrode material: Cell constant: 1 to ms/cm depends on the temperature compensation of the transmitter used glass limits between 25 and 200 C 40 bar as a standard; up to 100 bar in special design Rhodium 2 mm in diameter Glass lining: Glass WWG Explosion protection When using the measuring probe type LF (type of protection: intrinsic safety Ex ia IIC) in potentially explosive areas of zone 0, it is approved for connection to intrinsically safe circuits category ia with electrical isolation and the maximum electrical values specified in the certificate of conformity PTB No. Ex-88.B.2127X. approx. 0.2 cm 1, depending on design 9 Calibration The glasslined conductivity measuring probe is an open measuring cell, therefore, it has to be calibrated in the reactor/in the pipe in order to allow the stray field to be fully generated. For this reason, a transmitter is to be used which can be calibrated for the conductivity data of the product and which automatically calculates the cell constant after calibration. For calibration, the measuring probe is installed inside the reactor or the pipe and filled with product. Measure the conductivity of a product sample using an external conductivity meter. Then enter the measured conductivity manually into the transmitter. Please note that the product temperature must be identical to the sample temperature. 4 Pfaudler GmbH

5 LF Conductivity Measuring Probe # Conductor color Assignment 1 white voltage electrode coaxial cable 2 brown voltage electrode coaxial cable 3 brown current electrode + Pt green Pt blue current electrode 6 black Pt 100 Cable shield (total shield on connector to frame) 8 violet coaxial cable shield 1 (insulated at connector) 9 orange coaxial cable shield 2 (insulated at connector) Startup Connect the measuring probe and the transmitter according to the applicable wiring diagram. Startup comprises the parameterization and calibration of the transmitter. The wiring diagrams of suitable conductivity meters are available on request. During parameterization, enter a cell constant of approx. 0.2 cm 1. Measuring range of 1 ms and higher, for a conductivity 1 ms transmitter messages may be output (sensor not stable), depending on the design and place of installation of the transmitter. m Caution! When the probe has been installed in another measuring location, it has to be re-calibrated. Once a cell constant has been determined, it can always be used if the probe and place of installation are identical Electrical connection Figure 5 Connection cable assignment Figure 6 Terminal box Pfaudler GmbH 5

6 LF Conductivity Measuring Probe Figure 7 Wiring diagram female connector male connector female connector protective cap screwed connection conductivity processor (conductive) mains current output 14 Avoiding damages to peripheral devices When performing a spark test, please note the following information; otherwise, the components and/or the electronic transmitter may get damaged. The measuring transducers for temperature, glass monitoring, capacitive sensors and other electronic/ electric components that have been attached to the valve or the baffle must be disconnected prior to the spark test. Suitable equipment must be employed for the test (impulse voltage). We recommend using the GlaSparker, our high-voltage tester For glass lined measuring probes, the max. test voltage must not exceed 7 kv. The contact window around the measuring probe site (e.g. P) must not be tested. 11 Spare parts Designation Part no. 12 Related documents Service Instruction 318 Tightening Torques 400 mm O-ring made of Viton for protective cap, 25 x 1.5 mm D Connection cable, blue, 2 m with connector Connection cable, blue, 5 m with connector Connection cable, blue, 10 m with connector Glass testing During a spark test, inflammable sparks may occur at the pores in the form of an electric arc. Therefore, spark testing may only be carried out outside of potentially explosive atmospheres mm Figure 8 Contacs Measuring Points In general, however, we recommend calling our technical service for performing the test. 6 Pfaudler GmbH

7 Annex 1 PTB 03 ATEX 2132 X Pfaudler GmbH 7

8 Annex 1 PTB 03 ATEX 2132 X 8 Pfaudler GmbH

9 Annex 1 PTB 03 ATEX 2132 X Pfaudler GmbH 9

10 Annex 1 PTB 03 ATEX 2132 X 10 Pfaudler GmbH

11 Annex 1 PTB 03 ATEX 2132 X Pfaudler GmbH 11

12 Annex 1 PTB 03 ATEX 2132 X 12 Pfaudler GmbH

13 Annex 1 PTB 03 ATEX 2132 X Pfaudler GmbH 13

14 Annex 2 Declaration of conformity 14 Pfaudler GmbH

15 Notes Pfaudler GmbH 15

16 The information provided in this documentation corresponds to the state of the art at the time of printing. It is published in good faith. However, we will accept no warranty claims based on the information provided in this documentation. We reserve the right to include improvements, amendments and new findings in this documentation without prior notice. The actual design of products may deviate from the information contained in the calatoge if technical alterations and product improvements so require. The proposal made by Pfaudler for a concrete application will be binding in such cases. The present documentation is made available free of charge to our customers and other interested parties. The right to print or copy this documentation, or any parts there of, or to convert the same into electronic form shall be subject to our written permission. All rights reserved by us. Pfaudler Werke GmbH P.O. Box 1780 D Schwetzingen Pfaudlerstraße D Schwetzingen Phone Telefax info@pfaudler-instrumentation.com

Temperature probes type T/TW

Temperature probes type T/TW Temperature probes type T/TW Contents 1 General Preface 1.1 2 Introduction General instructions 1.2 2.1 Scope Operating range 1.3 2.2 Continuous Warnings contained glass monitoring in the operating 2 Safety

More information

WATERFLUX 3000 Quick Start

WATERFLUX 3000 Quick Start WATERFLUX 3000 Quick Start Electromagnetic flowmeter The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE CONTENTS WATERFLUX 3000

More information

B Operating Instructions V1.00/EN/

B Operating Instructions V1.00/EN/ Conductive conductivity/ultra-pure water sensors with a 2-electrode system, types 202922, 202923, 202924, 202925 Glass conductivity sensors, type 202922/30 Diaphragm tubes, type 201083 Compensation thermometer,

More information

WATERFLUX 3000 Quick Start

WATERFLUX 3000 Quick Start WATERFLUX 3000 Quick Start Electromagnetic flow sensor The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE CONTENTS WATERFLUX 3000

More information

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751 Data sheet Intrinsically safe pressure transmitter MBS 420, MBS 425, MBS 470 and MBS 475 The intrinsically safe pressure transmitter program is designed for use in hazardous environments and offers a reliable

More information

Technical Information Condumax CLS21D/CLS21

Technical Information Condumax CLS21D/CLS21 TI00085C/07/EN/14.14 71268310 Products Solutions Services Technical Information Condumax CLS21D/CLS21 Conductivity sensors, analog or digital with Memosens technology Cell constant k = 1.0 cm 1 Application

More information

... for hydraulic and general applications in hazardous areas

... for hydraulic and general applications in hazardous areas ATEX Approved Inductive... for hydraulic and general applications in hazardous areas Inductive Namur sensors Explosion protected Ignition class "intrinsically safe" when used together a suitable certified

More information

ED701 General Industry Pressure Transmitter

ED701 General Industry Pressure Transmitter ED701 General Industry Pressure Transmitter Standard industrial process connections Complete range of electrical connections 4... 20 ma and Voltage outputs Accuracy: 0.1%, 0.2% and 0.4% FS Quick response

More information

Transducer transmitter BILT 4 II 2(1) G. Technical Manual

Transducer transmitter BILT 4 II 2(1) G. Technical Manual GB Transducer transmitter BILT 4 II 2(1) G Technical Manual Transducer transmitter BILT 4 Contents Introduction General... 3 Ex.safety description... 4 Technical data... 5 Installation General... 6 Mechanical

More information

Conductivity sensor for hygienic applications

Conductivity sensor for hygienic applications Conductivity sensor for hygienic applications 2- or 4-electrode technology Extended measuring range 0.05 S/cm... 500 ms/cm Process connections, materials, surface finish designed for use in hygienic applications

More information

Technical Information Condumax CLS15D/CLS15

Technical Information Condumax CLS15D/CLS15 TI0009C/07/EN/4.4 7268306 Products Solutions Services Technical Information Condumax CLS5D/CLS5 Conductivity sensors, analog or digital with Memosens technology Cell constant k = 0.0 cm or k = 0. cm Application

More information

Installation & Operating Manual. iwap202

Installation & Operating Manual. iwap202 Installation & Operating Manual iwap202 This page is intentionally left blank. Document Number 409345 (based on 407655) (See Last Page for Revision Details) For warranty information, refer to Terms and

More information

Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters

Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters ll 2G EEx ia llc T6 As our products are continuously improved, we reserve the right to make any change in the stated specifications

More information

OPTIWAVE 7300 C Supplementary instructions

OPTIWAVE 7300 C Supplementary instructions OPTIWAVE 7300 C Supplementary instructions Radar (FMCW) Level Transmitter for agitated liquids in process applications Supplementary Instructions for ATEX applications KROHNE CONTENTS OPTIWAVE 7300 C General

More information

OPTIFLUX 5000 Technical Datasheet

OPTIFLUX 5000 Technical Datasheet OPTIFLUX 5000 Technical Datasheet Electromagnetic flowmeter in flanged version Exceptional long-term stability and accuracy For highly aggressive and abrasive fluids Fully vacuum-resistant with high-tech

More information

Operating Instructions

Operating Instructions Level and Pressure Operating Instructions VEGATOR 620, 621, 622 max. 0 10 min. 0 10 on VEGATOR 622! 6 7 8 9 10 11 12 13 14 in out Contents Contents Safety information... 2 Note Ex area... 2 1 Product description

More information

Intrinsically Safe Compact Controller CTR 210i. Intrinsically Safe Bargraph Indicator. BGI 210i

Intrinsically Safe Compact Controller CTR 210i. Intrinsically Safe Bargraph Indicator. BGI 210i Intrinsically Safe Compact Controller CTR 210i Intrinsically Safe Bargraph Indicator BGI 210i DMT 02 ATEX E 148 2. Supplement also grahics display Revision 4 IBS BatchControl GmbH Im Sträßchen 2 4 Tel.:

More information

Conductivity Transmitter

Conductivity Transmitter Conductivity Transmitter Programmable outputs: two transistor and single or dual analog 4-20 ma Removable backlighted display Universal process connection Type can be combined with... Three cell constants

More information

EX SSB 70 EX SSB 150 Protective cabinet system heating SSB

EX SSB 70 EX SSB 150 Protective cabinet system heating SSB ERICHR I C H O T T EX SSB 70 EX SSB 150 Protective cabinet system heating SSB Identification Planned certification from an order quantity of 25pc,. EC- type examination certificate Ambient temperature

More information

KFS. Flap-type flow meter. Design and range of application

KFS. Flap-type flow meter. Design and range of application Design and range of application The measuring device operates largely independent of viscosity and is suitable for indicating the flow rate of water, acids, alkaline solutions and gases. Every device is

More information

Pharma Temperature Sensor TFP-EX

Pharma Temperature Sensor TFP-EX SENSORS FOR AND BIO. Product Information TFP-EX Pharma Temperature Sensor TFP-EX Application/Specified Usage Temperature measurement: of liquid media in vessels and pipes suitable for applications in potentially

More information

PHV RO High Voltage Passive Probe. Instruction Manual

PHV RO High Voltage Passive Probe. Instruction Manual PHV 1000-3-RO High Voltage Passive Probe Instruction Manual Copyright 2012 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications

More information

Trimod Besta Level Switch type I 8 / IE9 8

Trimod Besta Level Switch type I 8 / IE9 8 Besta AG, CH8610 Uster / Switzerland Issue : November 2008 Page 1/8 Trimod Besta Level Switch type I 8 / IE9 8 for use in intrinsically safe installation acc. to directive 94/9/EC (ATEX) Contents 1. Safety

More information

Pressure transmitters MBS 4201, MBS 4251, MBS 4701 and MBS 4751

Pressure transmitters MBS 4201, MBS 4251, MBS 4701 and MBS 4751 060R9345 Instructions Pressure transmitters MBS 4201, MBS 4251, MBS 4701 and MBS 4751 Content Page 1 Description/application 2 Identification 3 Specifications 4 Safety instructions 5 Installation/dimensions

More information

Operator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique

Operator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique Operator Manual Transmitters Series KAT-... Analog transmitters in 2-wire technique Important Safety Notes! In the case of non-observance of the installation and safety notes or improper use of the transmitter

More information

Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted

Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted Electrical temperature measurement Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted WIKA data sheet TE 60.34 Applications Machine building, plant and vessel construction

More information

PHV 1000-RO High Voltage Passive Probe. Instruction Manual

PHV 1000-RO High Voltage Passive Probe. Instruction Manual PHV 1000-RO High Voltage Passive Probe Instruction Manual Copyright 2014 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications

More information

OPTIFLUX 5000 Technical Datasheet

OPTIFLUX 5000 Technical Datasheet OPTIFLUX 5000 Technical Datasheet Electromagnetic flowmeter in flanged version Exceptional long-term stability and accuracy For highly aggressive and abrasive fluids Fully vacuum-resistant with high-tech

More information

Intrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments Model IS-20-H

Intrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments Model IS-20-H Replacement product: Model IS-3 Electronic Pressure Measurement Intrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments WIKA Data Sheet PE 81.51 Applications

More information

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F Replacement product: Model IS-3 Electronic pressure measurement Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F WIKA data sheet PE

More information

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure DMP 4 Industrial Pressure Transmitter Without Media Isolation accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 0 mbar up to 0... 000 mbar Product characteristics excellent linearity

More information

Intrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F, IS-21-F

Intrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F, IS-21-F Replacement product: Model IS-3 Electronic Pressure Measurement Intrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F,

More information

LEO RECORD Ei Intrinsically Safe Digital Manometer with Record Function for use in Hazardous Applications. Accuracy: 0,1 %FS

LEO RECORD Ei Intrinsically Safe Digital Manometer with Record Function for use in Hazardous Applications. Accuracy: 0,1 %FS LEO RECORD Ei Actual Value Status Indication SELECT ENTER G1/4 II 2 G Ex ia IIC T4 PTB 05 ATEX 2012 X Intrinsically Safe Digital Manometer with Record Function for use in Hazardous Applications. Accuracy:

More information

PHV 1000-RO. High impedance passive probe. Features: CeramCore TM hybrid probe. Modular construction. Coaxial design

PHV 1000-RO. High impedance passive probe. Features: CeramCore TM hybrid probe. Modular construction. Coaxial design High impedance passive probe Features: CeramCore TM hybrid probe Modular construction Coaxial design Interchangeable spring contact tip Certificate of calibration available on request Read-out BNC connector

More information

BAR-ANT-N-N-EX. Antenna barrier for the hazardous area. Data sheet. 1 Description

BAR-ANT-N-N-EX. Antenna barrier for the hazardous area. Data sheet. 1 Description BAR--N-N-EX Antenna barrier for the hazardous area Data sheet 060_en_00 PHOENIX CONTACT 05-06-0 Description Using an antenna barrier, you can execute HF outputs of wireless modules with intrinsic safety.

More information

MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400

MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400 Data Sheet MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400 Submersible Hydrostatic s Rugged and versatile We reserve the right to continuously improve our products and make any change in

More information

Operating Manual. isolate501

Operating Manual. isolate501 Operating Manual isolate501 Document Number 413569 (See Last Page for Revision Details). Copyright Extronics Ltd, 2015 Page 2 of 21 Contents 1 Introduction... 4 2 Safety Information and Notes... 5 2.1

More information

Introduction. Additional Information. Temperature transmitter for HART protocol. Suitable for all standard requirements. Measurement made easy

Introduction. Additional Information. Temperature transmitter for HART protocol. Suitable for all standard requirements. Measurement made easy ABB MEASUREMENT & ANALYTICS COMMISSIONING INSTRUCTION TTF200 Field-mount temperature transmitter Temperature transmitter for HART protocol. Suitable for all standard requirements Measurement made easy

More information

ph or ORP Transmitter

ph or ORP Transmitter ph or ORP Transmitter Programmable outputs: two transistor and single or dual analog 4-20 ma (Process + Temp) Removable backlighted display Universal process connection Compatible with 120 mm ph/ ORP probes

More information

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. appmeas.co.uk +44 (0)

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. appmeas.co.uk +44 (0) DMP 4 Industrial Pressure Transmitter Without Media Isolation accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 0 mbar up to 0... 000 mbar Product characteristics excellent linearity

More information

Resistance Thermometers Model Series TR7X0, Sheathed Design

Resistance Thermometers Model Series TR7X0, Sheathed Design Electrical Temperature Measurement Resistance Thermometers Model Series TR7X0, Sheathed Design WIKA Data Sheet TE 60.40 Applications T Suitable for all industrial and laboratory applications Special Features

More information

Design and applications

Design and applications Design and applications The measuring device operates largely independent of viscosity and is suitable for indicating the flow rate of water, acids, alkaline solutions and gases. Every device is calibrated

More information

PV 102. IECEx. Piezoelectric velocity sensor FEATURES. From the Vibro-Meter product line

PV 102. IECEx. Piezoelectric velocity sensor FEATURES. From the Vibro-Meter product line FEATURES From the Vibro-Meter product line Designed for medium temperature industrial applications: 50 to +120 C ( 58 to +248 F) Frequency response: 6.0 to 2500 Hz: ±10 % to 5000 Hz: ±3 db Ultra low-noise

More information

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F Replacement product: Model IS-3 Electronic pressure measurement Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F WIKA data sheet PE

More information

Operator s Manual. PP016 Passive Probe

Operator s Manual. PP016 Passive Probe Operator s Manual PP016 Passive Probe 2017 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials is strictly prohibited. Customers are permitted

More information

B Operating Instructions /

B Operating Instructions / Conductivity/ultra-pure water measuring cells with a twin-electrode system Types 202922, 202923, 202924 and 202925 Glass conductivity cells, type 201080 Diaphragm tubes, type 201083 Compensation thermometer,

More information

RAD-IN/OUT-8D. Digital extension modules for the bidirectional wireless system. INTERFACE Data sheet _en_05. 1 Description

RAD-IN/OUT-8D. Digital extension modules for the bidirectional wireless system. INTERFACE Data sheet _en_05. 1 Description RAD-IN/OUT-D Digital extension modules for the bidirectional wireless system INTERFACE Data sheet 102122_en_0 PHOENIX CONTACT 2010-02-2 1 Description The RAD-ISM-...-SET-BD-BUS-ANT bidirectional wireless

More information

Inductive Conductivity Meter ILM

Inductive Conductivity Meter ILM 1 Product Information ILM-2 ILM-3 Inductive Conductivity Meter ILM f o o d Application / Specified Usage Authorizations Inductive measurement of the specific conductivity of liquid media in the range of

More information

PHV RO. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design

PHV RO. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design High impedance passive probe Features: CeramCore TM Hybrid Probe Modular Construction Coaxial Design Interchangeable Spring Contact Tip Certificate of Calibration available on request Read-out BNC Connector

More information

2-WIRE UNIVERSAL TEMPERATURE TRANSMITTER

2-WIRE UNIVERSAL TEMPERATURE TRANSMITTER SAFE INSTALLATION MANUAL (FM APPROVAL) -WIRE UNIVERSAL TEMPERATURE TRANSMITTER (HART communication, intrinsically safe/explosion-proof) MODEL B6U/B6U-B BEFORE USE... SAFETY PRECAUTIONS This manual describes

More information

MINI MCR-SL-UI-I-LP-NC

MINI MCR-SL-UI-I-LP-NC 2-way isolation amplifier Data sheet 105263_en_02 PHOENIX CONTACT 2013-12-13 1 Description The configurable 2-way isolation amplifiers are used to electrically isolate, convert and filter standard signals.

More information

Supplied in carry case with additional accessories The PHV 1000-RO is a 400 MHz, standard sized, 100:1 passive probe designed for instruments

Supplied in carry case with additional accessories The PHV 1000-RO is a 400 MHz, standard sized, 100:1 passive probe designed for instruments High impedance passive probe Features: CeramCore TM hybrid probe Modular construction Coaxial design Interchangeable spring contact tip Certificate of calibration available on request Read-out BNC connector

More information

E8EB-N0C2B E8EB-N0B2B

E8EB-N0C2B E8EB-N0B2B Slim Sensor Ideal for Workpiece Position and Original Checking The to 1 kpa model can be used for workpiece position checking. The to 1MPa model is ideal for original pressure checking. Degree of protection

More information

Signal converter for electromagnetic flowmeters

Signal converter for electromagnetic flowmeters Quick Start Signal converter for electromagnetic flowmeters Electronic revision: ER 3.0.xx The documentation is only complete when used in combination with the relevant documentation for the flow sensor.

More information

Type EXT with cable, probe sold

Type EXT with cable, probe sold standard signal FF-GLT-Series Description Type INT for duct mounting, probe installed Type EXT with cable, probe sold Type AF for wall mounting, probe sold Characteristic features Standard signal 0...

More information

Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version

Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version Electrical temperature measurement Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version WIKA data sheet TE 15.01 Applications Process industry Machine building

More information

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted Electrical temperature measurement OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted WIKA data sheet TE 60.31 further approvals see page 10 Applications Machine building, plant

More information

Technical Information Orbipore CPS91D and CPS91

Technical Information Orbipore CPS91D and CPS91 TI00375C/07/EN/14.15 71299929 Products Solutions Services Technical Information Orbipore CPS91D and CPS91 ph electrodes, analog or with digital Memosens technology With open aperture for contaminated media,

More information

Capacitive level meters CLM 40 INSTRUCTIONS MANUAL. průmyslová elektronika

Capacitive level meters CLM 40 INSTRUCTIONS MANUAL. průmyslová elektronika průmyslová elektronika INSTRUCTIONS MANUAL Capacitive level meters CLM 40 Before level meters usage please read instructions in this service manual and keep it for further use. Service manual is for all

More information

09746E00. ATEX / IECEx NEC 505 NEC 506 NEC 500 Class I Class I Class II Class III Zone Zone Division Ex i

09746E00. ATEX / IECEx NEC 505 NEC 506 NEC 500 Class I Class I Class II Class III Zone Zone Division Ex i Temperature Transmitter Series 9182 www.stahl.de > One unit for nearly all temperature sensors indivdually configurable > Intrinsically safe input [Ex ia] IIC > Signal duplication possible > Galvanic isolation

More information

Installation guide 971 SmartRadar LTi

Installation guide 971 SmartRadar LTi Installation guide 971 SmartRadar LTi March 2009 Part no. 4416.715 Revision 3 Enraf B.V. P.O. Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100 Fax : +31 15 2701 111 E-mail : enraf-nl@honeywellenraf.nl

More information

MANUAL. Electrostatic Field Meter EFM 235 / 255. System Prof. Dr.-Ing. Hans Kleinwächter

MANUAL. Electrostatic Field Meter EFM 235 / 255. System Prof. Dr.-Ing. Hans Kleinwächter MANUAL Electrostatic Field Meter EFM 235 / 255 System Prof. Dr.-Ing. Hans Kleinwächter 2 Manual EFM 120 Contents General information about electrostatics... 3 Origin of electrostatic charge... 3 Product

More information

GETTING STARTED GUIDE NI AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous

GETTING STARTED GUIDE NI AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous GETTING STARTED GUIDE NI 9228 8 AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous This document explains how to connect to the NI 9228. Note Before you begin, complete the software and hardware installation procedures

More information

Technical Information Memosens CLS82D

Technical Information Memosens CLS82D TI01188C/07/EN/03.16 71318348 Products Solutions Services Technical Information Memosens CLS82D Hygienic conductivity sensors, digital with Memosens technology, cell constant k = 0.57 cm -1 Application

More information

JUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter

JUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: 49 661 60030 Fax: 49 661 6003607 Email: mail@jumo.net Internet: www.jumo.net JUMO

More information

TAG5000 WIRELESS PHASER. Instruction Manual HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS U.S.A.

TAG5000 WIRELESS PHASER. Instruction Manual HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS U.S.A. TAG5000 WIRELESS PHASER Instruction Manual TM HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS 60085 U.S.A. PHONE 847.473.4980 FAX 847.473.4981 website: www.hdelectriccompany.com DESCRIPTION

More information

Operating Instructions for Pressure Sensor with Ceramic Sensor Element. Model: SEN-9601

Operating Instructions for Pressure Sensor with Ceramic Sensor Element. Model: SEN-9601 Operating Instructions for Pressure Sensor with Ceramic Sensor Element Model: SEN-9601 We don t accept warranty and liability claims neither upon this publication nor in case of improper treatment of the

More information

3 Ex i Isolators. Temperature Transmitter without Output 0/ ma, with Limit Value Contact (Field Circuit Ex i) Type 9182/.

3 Ex i Isolators. Temperature Transmitter without Output 0/ ma, with Limit Value Contact (Field Circuit Ex i) Type 9182/. Ex i Isolators Temperature Transmitter without Output 0/4... 20 ma, with Limit Value Contact Field Circuit Ex i) One unit for nearly all temperature sensors indivdually configurable Intrinsically safe

More information

Condumax CLS21 and CLS21D

Condumax CLS21 and CLS21D Technical Information Condumax CLS21 and CLS21D Conductivity sensors, analog or digital with Memosens technology Cell constant k = 1 cm -1 Application Measurements in media of medium and high conductivities:

More information

Safety Manual. Omni-ID UHF Tags

Safety Manual. Omni-ID UHF Tags Safety Manual Omni-ID UHF Tags This page is intentionally left blank. Document Number 330556 (See Last Page for Revision Details) 2012 Extronics Limited. This document is Copyright Extronics limited. Extronics

More information

Analogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting

Analogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting Electrical Temperature Measurement Analogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting WIKA Data Sheet TE 24.01 Applications T Machinery, plant construction T Process industry

More information

JUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter

JUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: 49 661 60030 Fax: 49 661 6003607 Email: mail@jumo.net Internet: www.jumo.net JUMO

More information

Resistance thermometer Model TR30, compact design

Resistance thermometer Model TR30, compact design Electrical temperature measurement Resistance thermometer Model TR30, compact design WIKA data sheet TE 60.30 Applications Machine building, plant and vessel construction Propulsion technology, hydraulics

More information

KSR-Level sensor KSR-Schwimmer-Magnetschalter KSR Contact Protection Relays

KSR-Level sensor KSR-Schwimmer-Magnetschalter KSR Contact Protection Relays KSR-Level sensor KSR-Schwimmer-Magnetschalter 1001-5 KSR Contact Protection s 1 KSR Contact Protection Type KR230-Ex and KR24-Ex General Description The contact protection relays KR230-Ex and KR24-Ex transmit

More information

OPTIFLUX 1000 Handbook

OPTIFLUX 1000 Handbook OPTIFLUX 1000 Handbook Electromagnetic flow sensor in sandwich design The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE : IMPRINT

More information

3 Ex i Isolators. Temperature Transmitter with Output 0/ ma, without Limit Value Contact (Field Circuit Ex i) Type 9182/.0-5.

3 Ex i Isolators. Temperature Transmitter with Output 0/ ma, without Limit Value Contact (Field Circuit Ex i) Type 9182/.0-5. Ex i Isolators Temperature Transmitter with Output 0/4... 20 ma, without Limit Value Contact Field Circuit Ex i) One unit for nearly all temperature sensors indivdually configurable Intrinsically safe

More information

MT MOTORI ELETTRICI. Installation, operation, maintenance and safety manual for motors used in hazardous areas 1-II-2G 21-II-2D

MT MOTORI ELETTRICI. Installation, operation, maintenance and safety manual for motors used in hazardous areas 1-II-2G 21-II-2D MT MOTORI ELETTRICI Installation, operation, maintenance and safety manual for motors used in hazardous areas 1-II-2G 21-II-2D TABLE OF CONTENTS 1. Introduction 2. Scope of application 3. Installation

More information

Measuring and Controlling Unit Multronic OC

Measuring and Controlling Unit Multronic OC Measuring and Controlling Unit Multronic OC One channel measuring and controlling unit for ph / conductivity / Chlorine measuring Automatic temperature compensation Measuring and control behaviour simultaneously

More information

Rosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna

Rosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna Rosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna Product Data Sheet November 2016 DS 2000-10-008 Ed 03 Rosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna is a marine pipe guided level gauge that

More information

Description of equipment

Description of equipment Applicant: Electrical Apparatus: KROHNE Messtechnik GmbH & Co. KG Level Radar BM 70M-Ex Description of equipment The Types BM 70M-Ex Level Radar gauges serve to continuously measure the product level in

More information

OPTITEMP TT 10 C/R Technical Datasheet

OPTITEMP TT 10 C/R Technical Datasheet OPTITEMP TT 10 C/R Technical Datasheet Analogue 2-wire temperature transmitter Temperature linear 4...20 ma output Rangeable with solder pads and potentiometers Easy wiring through large center hole The

More information

Submersible pressure sensor For water and wastewater Model LW-1

Submersible pressure sensor For water and wastewater Model LW-1 Replacement product: Model LF-1 Level measurement Submersible pressure sensor For water and wastewater Model LW-1 WIKA data sheet LM 40.03 Applications Level measurement in rivers and lakes Deep well and

More information

JUMO BlackLine CR-GT/-EC/-GS Conductive 2-Electrode Conductivity Sensor

JUMO BlackLine CR-GT/-EC/-GS Conductive 2-Electrode Conductivity Sensor sales@jumo.co.uk info.us@jumo.net Vorläufig V Page 1/8 JUMO BlackLine CR-GT/-EC/-GS Conductive 2-Electrode Conductivity Sensor Cell constants K = 0.01 1.0 202922 Series for universal application, e.g.

More information

Isolating Spark Gaps. Isolating spark gaps at an additional earth ring conductor. Isolating spark gaps for Ex zones

Isolating Spark Gaps. Isolating spark gaps at an additional earth ring conductor. Isolating spark gaps for Ex zones Isolating Spark Gaps Isolating spark gaps at an additional earth ring conductor Isolating spark gaps for Ex zones 187 Protective Spark Gaps For use at roof supports for overhead lines For indirect connection

More information

MANUAL HYDROBAR. Manufactured by: 7991 CZ DWINGELOO 7990 AA DWINGELOO The Netherlands

MANUAL HYDROBAR.   Manufactured by: 7991 CZ DWINGELOO 7990 AA DWINGELOO The Netherlands MANUAL HYDROBAR WARNING Before installing the HYDROBAR, read the warnings and advisements on page 9 for personal and system safety, and for optimum performance, make sure you thoroughly understand the

More information

Refer to manual. Failure to read message could result in personal injury or serious damage to the equipment or both.

Refer to manual. Failure to read message could result in personal injury or serious damage to the equipment or both. CERAVAC Transmitter CTR 100 N Short form manual 300544675_002_C0 Part numbers: 230 300V02 230 310V02 230 301V02 230 311V02 230 302V02 230 312V02 230 303V02 230 313V02 230 304V02 230 314V02 230 305V02 230

More information

Wind Transmitter - with frequency output xx.xxx

Wind Transmitter - with frequency output xx.xxx THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for Use 021531/11/09 Wind Transmitter - with frequency output 4.3303.xx.xxx ADOLF THIES GmbH & Co. KG Hauptstraße

More information

3 Ex i Isolators. Switching Repeater with Power Relay (250 V / 4 A) Type 9170/ Type 9170/

3 Ex i Isolators. Switching Repeater with Power Relay (250 V / 4 A) Type 9170/ Type 9170/ 3 Ex i Isolators Switching Repeater with Power Relay Intrinsically safe input [Ex ia] IIC Galvanic isolation between input, output and power supply Open-circuit / short-circuit monitoring and messaging

More information

CE620. Piezoelectric accelerometer with integrated electronics. IECEx

CE620. Piezoelectric accelerometer with integrated electronics. IECEx Piezoelectric accelerometer with integrated electronics FEATURES» From the Vibro-Meter product line» Voltage output signal: 100 or 500 mv/ g» Frequency response: 2 Hz to 10 khz» Temperature range: 55 to

More information

Dawson DDM190. Digital Multimeter User s Manual

Dawson DDM190. Digital Multimeter User s Manual Dawson DDM190 Digital Multimeter User s Manual TABLE OF CONTENTS LIMITED WARRANTY AND LIMITATION OF LIABILITY... 3 Out of the Box... 3 Accessories.. Error! Bookmark not defined. Safety Information... 7

More information

MANUAL HYDROBAR. Manufactured by: Fax: H/US/Hydrobar/ /00 Klay Instruments Page 1

MANUAL HYDROBAR.   Manufactured by: Fax: H/US/Hydrobar/ /00 Klay Instruments Page 1 MANUAL HYDROBAR WARNING Before installing the Hydrobar, read the warnings and advisements on the last page for personal and system safety, and for optimum performance, make sure you thoroughly understand

More information

Operator s Manual. PP022 Passive Probe

Operator s Manual. PP022 Passive Probe Operator s Manual PP022 Passive Probe 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com PP022 Passive Probe Instruction Manual 2017 Teledyne

More information

Dawson DDM181. Pocket-Size Autorange Digital Meter User s Manual

Dawson DDM181. Pocket-Size Autorange Digital Meter User s Manual Dawson DDM181 Pocket-Size Autorange Digital Meter User s Manual 1 Table of Contents LIMITED WARRANTY AND LIMITATION OF LIABILITY... 3 Out of the Box... 3 Accessories... 4 Important Safety Information...

More information

Ultrasonic level measuring device, non-contact

Ultrasonic level measuring device, non-contact Ultrasonic level measuring device, non-contact For level measurement up to 8 m 4... 20 ma/hart - 2 wires Suitable for solids ATEX approvals Type 8177 can be combined with... Type 8611 Type 8793 Type 8802-GB

More information

VBS TBS KTS BSS LFS UFS

VBS TBS KTS BSS LFS UFS VBS TBS KTS BSS LFS UFS TBS. Surge protection systems Order information and technical data Protection and isolating spark gaps Protection spark gap or isolating spark gap? Isolating spark gaps Protection

More information

Installation guide 877 FDI Field Display & Interface

Installation guide 877 FDI Field Display & Interface Installation guide 877 FDI Field Display & Interface November 2008 Part no. 4416.264 Rev. 5 Enraf BV PO Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100 Fax : +31 15 2701 111 E-mail : enraf-nl@honeywell.com

More information

User manual. Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD

User manual. Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD User manual Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD Contents Precautions Intended use General 1 Specification 3

More information

SMARTPAT COND 1200 Technical Datasheet

SMARTPAT COND 1200 Technical Datasheet SMARTPAT COND 1200 Technical Datasheet Digital conductive conductivity sensor for the water and wastewater industry 2-wire loop powered sensor with integrated transmitter technology Wide measuring range

More information

TIDALFLUX 2300 F Technical Datasheet

TIDALFLUX 2300 F Technical Datasheet Technical Datasheet Electromagnetic flow sensor for partially filled pipes Measurement in partially filled pipes up to DN1600 / 64" Patented, non-contact level measurement Measurement possible down to

More information

Thermocouple - Transducer

Thermocouple - Transducer Thermocouple - Transducer TM3 Characteristics: Thermocouple type J, K, R, S, T, E, B, N Linearity error < 1% Measurement range configurable -210...+1820 C Built-in cold junction compensation Galvanic 3-way

More information