LF Conductivity Measuring Probe
|
|
- Nancy Dennis
- 5 years ago
- Views:
Transcription
1 LF Conductivity Measuring Probe Contents 1 General 1.1 Introduction 1.2 Application range 2 Safety 2.1 Warnings and symbols 3 Transport and storage 4 Application 5 Measuring principle 6 Assembly 7 Construction 8 Technical data 8.1 Explosion protection 9 Calibration 10 Startup 10.1 Electrical connection 11 Spare parts 12 Related documents 13 Glass Testing 14 Avoiding damages to peripheral devices Annex 1 EC-Type-Examination-Certificate Annex 2 Declaration of conformity Operating Instructions e
2 LF Conductivity Measuring Probe 1 General 1.1 Introduction The probe may only be installed, started up and serviced by authorized personnel with special skills in measuring technology and in strict compliance with the present instructions as well as the valid provisions. Non-compliance with these provisions no matter whether by intention or by negligence shall release the manufacturer from all responsibility and reliability. The present operating instructions are designed to familiarize users with the equipment of the probe and its use. The operating instructions should be available to the operating and maintenance personnel and should be studied before performing any installation and/or maintenance work. By knowing these operating instructions, you can avoid damages to the measuring equipment and ensure trouble-free operation. The representations and data given in these Operating Instructions are subject to modification and further improvement. The latest edition hereof will always supersede all previous ones. 1.2 Application range LF probes are used to measure the conductivity of changes in conductivity. Never operate this measuring device outside its permissible operating conditions. 2 Safety The safety instructions refer exclusively to the scope of delivery. Please also note the valid safety regulations concerning electrical installations and facilities. 2.1 Warnings and symbols In the operating instructions, the danger symbol is used to draw your attention to especially important information. m Compliance with these mandatory instructions helps to avoid personal injury and damage to property. m Please note the following instructions: Do not practice any working methods which may endanger safety! 3 Transport and storage The probe should only be transported and stored in its closed original packaging. In order to guarantee an as-new condition of the probe, maintain the following storage conditions: dry and dust-free, steady temperature and ventilation When used probes are sent to the manufacturer or third parties for repair or other purposes, all parts must be cleaned and decontaminated. Appropriate safety instructions for further handling must be attached to the outside of the packaging. The probes do not need any preservatives, they are resistant to normal environmental influences. Figure 1 LF measuring probes 4 Application The fully glasslined LF measuring probe is used to measure the electrolytic conductivity and can be installed directly inside reactors or pipes. For the resistance of the glass lining used, please refer to our publication no The LF probe is preferably used for measurement in concentrated acids at high temperatures and high pressures up to 40 bar (special option up to 100 bar). 2 Pfaudler GmbH
3 LF Conductivity Measuring Probe 5 Measuring principle The 4-electrode method is used. A constant AC current flows through the external electrodes. The measuring-circuit voltage is picked up in the form of a resistancedependent voltage drop across the internal electrodes without current load. The voltage value is then supplied to a transmitter with a high-resistance input. This measuring method is largely insensitive to contamination of the electrodes. Polarization errors are impossible because polarization only occurs at the current electrodes. A built-in Pt 100 resistance thermometer compensates for the temperature dependence of the conductivity. By combining steel, glass and rhodium (for the electrodes) as materials for the probe, the LF probe was given a number of outstanding characteristics: high resistance to corrosion large temperature application range between C resistance to pressure: bar (option 100 bar) resistance to thermal shock Figure 2 LF measuring probe in pipes Rhodium electrodes equipotential bonding conductor, if necessary U R i = i Due to these properties, it is even suitable for extreme applications. 6 Assembly When measuring in pipes made of plastic materials or steel with an insulating lining, the use of conductive gaskets offers advantages. These gaskets are connected to the equipotential bonding system of the probe. DN50 part number DN80 part number Glass l k U Pt 100 Metal body Process conductometer ms/cm % by weight ma LF When installing the probe in steel pipes, only normal, non-conductive gaskets may be used due to the possible formation of electrolytic element couples between components. DN50 part number DN80 part number Figure 3 Structure of the LF measuring probes For the tightening torques of the split flange connection, please refer to Service Instructions 318. After the first temperature cycle (but by no means later than after 24 hours) the screws must be retightened with the prescribed torque. Pfaudler GmbH 3
4 LF Conductivity Measuring Probe 7 Construction The four metal electrodes with a diameter of 2 mm have been fused into the glass layer of a measuring probe carrier so as to ensure insulation between them. Measuring probe carriers may include baffles, thermometer wells or spacing rings. The supply conductors for the electrodes are also fused into glass and lead to a terminal box at the flange of the probe carrier. The Pt 100 has been integrated into the steel body of the probe carrier. Baffles and thermometer wells are available in any desired size, spacing rings may be manufactured from DN ms/cm 1 g/l 0,1 1 10,05 100,6 0,1 0,01 0, % by weight Figure 4 Concentration table H 2 SO 4 20 C 8 Technical data Conductivity range: Temperature range: Pressure: Electrode material: Cell constant: 1 to ms/cm depends on the temperature compensation of the transmitter used glass limits between 25 and 200 C 40 bar as a standard; up to 100 bar in special design Rhodium 2 mm in diameter Glass lining: Glass WWG Explosion protection When using the measuring probe type LF (type of protection: intrinsic safety Ex ia IIC) in potentially explosive areas of zone 0, it is approved for connection to intrinsically safe circuits category ia with electrical isolation and the maximum electrical values specified in the certificate of conformity PTB No. Ex-88.B.2127X. approx. 0.2 cm 1, depending on design 9 Calibration The glasslined conductivity measuring probe is an open measuring cell, therefore, it has to be calibrated in the reactor/in the pipe in order to allow the stray field to be fully generated. For this reason, a transmitter is to be used which can be calibrated for the conductivity data of the product and which automatically calculates the cell constant after calibration. For calibration, the measuring probe is installed inside the reactor or the pipe and filled with product. Measure the conductivity of a product sample using an external conductivity meter. Then enter the measured conductivity manually into the transmitter. Please note that the product temperature must be identical to the sample temperature. 4 Pfaudler GmbH
5 LF Conductivity Measuring Probe # Conductor color Assignment 1 white voltage electrode coaxial cable 2 brown voltage electrode coaxial cable 3 brown current electrode + Pt green Pt blue current electrode 6 black Pt 100 Cable shield (total shield on connector to frame) 8 violet coaxial cable shield 1 (insulated at connector) 9 orange coaxial cable shield 2 (insulated at connector) Startup Connect the measuring probe and the transmitter according to the applicable wiring diagram. Startup comprises the parameterization and calibration of the transmitter. The wiring diagrams of suitable conductivity meters are available on request. During parameterization, enter a cell constant of approx. 0.2 cm 1. Measuring range of 1 ms and higher, for a conductivity 1 ms transmitter messages may be output (sensor not stable), depending on the design and place of installation of the transmitter. m Caution! When the probe has been installed in another measuring location, it has to be re-calibrated. Once a cell constant has been determined, it can always be used if the probe and place of installation are identical Electrical connection Figure 5 Connection cable assignment Figure 6 Terminal box Pfaudler GmbH 5
6 LF Conductivity Measuring Probe Figure 7 Wiring diagram female connector male connector female connector protective cap screwed connection conductivity processor (conductive) mains current output 14 Avoiding damages to peripheral devices When performing a spark test, please note the following information; otherwise, the components and/or the electronic transmitter may get damaged. The measuring transducers for temperature, glass monitoring, capacitive sensors and other electronic/ electric components that have been attached to the valve or the baffle must be disconnected prior to the spark test. Suitable equipment must be employed for the test (impulse voltage). We recommend using the GlaSparker, our high-voltage tester For glass lined measuring probes, the max. test voltage must not exceed 7 kv. The contact window around the measuring probe site (e.g. P) must not be tested. 11 Spare parts Designation Part no. 12 Related documents Service Instruction 318 Tightening Torques 400 mm O-ring made of Viton for protective cap, 25 x 1.5 mm D Connection cable, blue, 2 m with connector Connection cable, blue, 5 m with connector Connection cable, blue, 10 m with connector Glass testing During a spark test, inflammable sparks may occur at the pores in the form of an electric arc. Therefore, spark testing may only be carried out outside of potentially explosive atmospheres mm Figure 8 Contacs Measuring Points In general, however, we recommend calling our technical service for performing the test. 6 Pfaudler GmbH
7 Annex 1 PTB 03 ATEX 2132 X Pfaudler GmbH 7
8 Annex 1 PTB 03 ATEX 2132 X 8 Pfaudler GmbH
9 Annex 1 PTB 03 ATEX 2132 X Pfaudler GmbH 9
10 Annex 1 PTB 03 ATEX 2132 X 10 Pfaudler GmbH
11 Annex 1 PTB 03 ATEX 2132 X Pfaudler GmbH 11
12 Annex 1 PTB 03 ATEX 2132 X 12 Pfaudler GmbH
13 Annex 1 PTB 03 ATEX 2132 X Pfaudler GmbH 13
14 Annex 2 Declaration of conformity 14 Pfaudler GmbH
15 Notes Pfaudler GmbH 15
16 The information provided in this documentation corresponds to the state of the art at the time of printing. It is published in good faith. However, we will accept no warranty claims based on the information provided in this documentation. We reserve the right to include improvements, amendments and new findings in this documentation without prior notice. The actual design of products may deviate from the information contained in the calatoge if technical alterations and product improvements so require. The proposal made by Pfaudler for a concrete application will be binding in such cases. The present documentation is made available free of charge to our customers and other interested parties. The right to print or copy this documentation, or any parts there of, or to convert the same into electronic form shall be subject to our written permission. All rights reserved by us. Pfaudler Werke GmbH P.O. Box 1780 D Schwetzingen Pfaudlerstraße D Schwetzingen Phone Telefax info@pfaudler-instrumentation.com
Temperature probes type T/TW
Temperature probes type T/TW Contents 1 General Preface 1.1 2 Introduction General instructions 1.2 2.1 Scope Operating range 1.3 2.2 Continuous Warnings contained glass monitoring in the operating 2 Safety
More informationWATERFLUX 3000 Quick Start
WATERFLUX 3000 Quick Start Electromagnetic flowmeter The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE CONTENTS WATERFLUX 3000
More informationB Operating Instructions V1.00/EN/
Conductive conductivity/ultra-pure water sensors with a 2-electrode system, types 202922, 202923, 202924, 202925 Glass conductivity sensors, type 202922/30 Diaphragm tubes, type 201083 Compensation thermometer,
More informationWATERFLUX 3000 Quick Start
WATERFLUX 3000 Quick Start Electromagnetic flow sensor The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE CONTENTS WATERFLUX 3000
More informationIntrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751
Data sheet Intrinsically safe pressure transmitter MBS 420, MBS 425, MBS 470 and MBS 475 The intrinsically safe pressure transmitter program is designed for use in hazardous environments and offers a reliable
More informationTechnical Information Condumax CLS21D/CLS21
TI00085C/07/EN/14.14 71268310 Products Solutions Services Technical Information Condumax CLS21D/CLS21 Conductivity sensors, analog or digital with Memosens technology Cell constant k = 1.0 cm 1 Application
More information... for hydraulic and general applications in hazardous areas
ATEX Approved Inductive... for hydraulic and general applications in hazardous areas Inductive Namur sensors Explosion protected Ignition class "intrinsically safe" when used together a suitable certified
More informationED701 General Industry Pressure Transmitter
ED701 General Industry Pressure Transmitter Standard industrial process connections Complete range of electrical connections 4... 20 ma and Voltage outputs Accuracy: 0.1%, 0.2% and 0.4% FS Quick response
More informationTransducer transmitter BILT 4 II 2(1) G. Technical Manual
GB Transducer transmitter BILT 4 II 2(1) G Technical Manual Transducer transmitter BILT 4 Contents Introduction General... 3 Ex.safety description... 4 Technical data... 5 Installation General... 6 Mechanical
More informationConductivity sensor for hygienic applications
Conductivity sensor for hygienic applications 2- or 4-electrode technology Extended measuring range 0.05 S/cm... 500 ms/cm Process connections, materials, surface finish designed for use in hygienic applications
More informationTechnical Information Condumax CLS15D/CLS15
TI0009C/07/EN/4.4 7268306 Products Solutions Services Technical Information Condumax CLS5D/CLS5 Conductivity sensors, analog or digital with Memosens technology Cell constant k = 0.0 cm or k = 0. cm Application
More informationInstallation & Operating Manual. iwap202
Installation & Operating Manual iwap202 This page is intentionally left blank. Document Number 409345 (based on 407655) (See Last Page for Revision Details) For warranty information, refer to Terms and
More informationExpert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters
Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters ll 2G EEx ia llc T6 As our products are continuously improved, we reserve the right to make any change in the stated specifications
More informationOPTIWAVE 7300 C Supplementary instructions
OPTIWAVE 7300 C Supplementary instructions Radar (FMCW) Level Transmitter for agitated liquids in process applications Supplementary Instructions for ATEX applications KROHNE CONTENTS OPTIWAVE 7300 C General
More informationOPTIFLUX 5000 Technical Datasheet
OPTIFLUX 5000 Technical Datasheet Electromagnetic flowmeter in flanged version Exceptional long-term stability and accuracy For highly aggressive and abrasive fluids Fully vacuum-resistant with high-tech
More informationOperating Instructions
Level and Pressure Operating Instructions VEGATOR 620, 621, 622 max. 0 10 min. 0 10 on VEGATOR 622! 6 7 8 9 10 11 12 13 14 in out Contents Contents Safety information... 2 Note Ex area... 2 1 Product description
More informationIntrinsically Safe Compact Controller CTR 210i. Intrinsically Safe Bargraph Indicator. BGI 210i
Intrinsically Safe Compact Controller CTR 210i Intrinsically Safe Bargraph Indicator BGI 210i DMT 02 ATEX E 148 2. Supplement also grahics display Revision 4 IBS BatchControl GmbH Im Sträßchen 2 4 Tel.:
More informationConductivity Transmitter
Conductivity Transmitter Programmable outputs: two transistor and single or dual analog 4-20 ma Removable backlighted display Universal process connection Type can be combined with... Three cell constants
More informationEX SSB 70 EX SSB 150 Protective cabinet system heating SSB
ERICHR I C H O T T EX SSB 70 EX SSB 150 Protective cabinet system heating SSB Identification Planned certification from an order quantity of 25pc,. EC- type examination certificate Ambient temperature
More informationKFS. Flap-type flow meter. Design and range of application
Design and range of application The measuring device operates largely independent of viscosity and is suitable for indicating the flow rate of water, acids, alkaline solutions and gases. Every device is
More informationPharma Temperature Sensor TFP-EX
SENSORS FOR AND BIO. Product Information TFP-EX Pharma Temperature Sensor TFP-EX Application/Specified Usage Temperature measurement: of liquid media in vessels and pipes suitable for applications in potentially
More informationPHV RO High Voltage Passive Probe. Instruction Manual
PHV 1000-3-RO High Voltage Passive Probe Instruction Manual Copyright 2012 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications
More informationTrimod Besta Level Switch type I 8 / IE9 8
Besta AG, CH8610 Uster / Switzerland Issue : November 2008 Page 1/8 Trimod Besta Level Switch type I 8 / IE9 8 for use in intrinsically safe installation acc. to directive 94/9/EC (ATEX) Contents 1. Safety
More informationPressure transmitters MBS 4201, MBS 4251, MBS 4701 and MBS 4751
060R9345 Instructions Pressure transmitters MBS 4201, MBS 4251, MBS 4701 and MBS 4751 Content Page 1 Description/application 2 Identification 3 Specifications 4 Safety instructions 5 Installation/dimensions
More informationOperator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique
Operator Manual Transmitters Series KAT-... Analog transmitters in 2-wire technique Important Safety Notes! In the case of non-observance of the installation and safety notes or improper use of the transmitter
More informationMiniature resistance thermometer Explosion-protected version Model TR34, thread-mounted
Electrical temperature measurement Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted WIKA data sheet TE 60.34 Applications Machine building, plant and vessel construction
More informationPHV 1000-RO High Voltage Passive Probe. Instruction Manual
PHV 1000-RO High Voltage Passive Probe Instruction Manual Copyright 2014 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications
More informationOPTIFLUX 5000 Technical Datasheet
OPTIFLUX 5000 Technical Datasheet Electromagnetic flowmeter in flanged version Exceptional long-term stability and accuracy For highly aggressive and abrasive fluids Fully vacuum-resistant with high-tech
More informationIntrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments Model IS-20-H
Replacement product: Model IS-3 Electronic Pressure Measurement Intrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments WIKA Data Sheet PE 81.51 Applications
More informationIntrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F
Replacement product: Model IS-3 Electronic pressure measurement Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F WIKA data sheet PE
More informationDMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure
DMP 4 Industrial Pressure Transmitter Without Media Isolation accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 0 mbar up to 0... 000 mbar Product characteristics excellent linearity
More informationIntrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F, IS-21-F
Replacement product: Model IS-3 Electronic Pressure Measurement Intrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F,
More informationLEO RECORD Ei Intrinsically Safe Digital Manometer with Record Function for use in Hazardous Applications. Accuracy: 0,1 %FS
LEO RECORD Ei Actual Value Status Indication SELECT ENTER G1/4 II 2 G Ex ia IIC T4 PTB 05 ATEX 2012 X Intrinsically Safe Digital Manometer with Record Function for use in Hazardous Applications. Accuracy:
More informationPHV 1000-RO. High impedance passive probe. Features: CeramCore TM hybrid probe. Modular construction. Coaxial design
High impedance passive probe Features: CeramCore TM hybrid probe Modular construction Coaxial design Interchangeable spring contact tip Certificate of calibration available on request Read-out BNC connector
More informationBAR-ANT-N-N-EX. Antenna barrier for the hazardous area. Data sheet. 1 Description
BAR--N-N-EX Antenna barrier for the hazardous area Data sheet 060_en_00 PHOENIX CONTACT 05-06-0 Description Using an antenna barrier, you can execute HF outputs of wireless modules with intrinsic safety.
More informationMJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400
Data Sheet MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400 Submersible Hydrostatic s Rugged and versatile We reserve the right to continuously improve our products and make any change in
More informationOperating Manual. isolate501
Operating Manual isolate501 Document Number 413569 (See Last Page for Revision Details). Copyright Extronics Ltd, 2015 Page 2 of 21 Contents 1 Introduction... 4 2 Safety Information and Notes... 5 2.1
More informationIntroduction. Additional Information. Temperature transmitter for HART protocol. Suitable for all standard requirements. Measurement made easy
ABB MEASUREMENT & ANALYTICS COMMISSIONING INSTRUCTION TTF200 Field-mount temperature transmitter Temperature transmitter for HART protocol. Suitable for all standard requirements Measurement made easy
More informationph or ORP Transmitter
ph or ORP Transmitter Programmable outputs: two transistor and single or dual analog 4-20 ma (Process + Temp) Removable backlighted display Universal process connection Compatible with 120 mm ph/ ORP probes
More informationDMP 343. Industrial Pressure Transmitter. Without Media Isolation. appmeas.co.uk +44 (0)
DMP 4 Industrial Pressure Transmitter Without Media Isolation accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 0 mbar up to 0... 000 mbar Product characteristics excellent linearity
More informationResistance Thermometers Model Series TR7X0, Sheathed Design
Electrical Temperature Measurement Resistance Thermometers Model Series TR7X0, Sheathed Design WIKA Data Sheet TE 60.40 Applications T Suitable for all industrial and laboratory applications Special Features
More informationDesign and applications
Design and applications The measuring device operates largely independent of viscosity and is suitable for indicating the flow rate of water, acids, alkaline solutions and gases. Every device is calibrated
More informationPV 102. IECEx. Piezoelectric velocity sensor FEATURES. From the Vibro-Meter product line
FEATURES From the Vibro-Meter product line Designed for medium temperature industrial applications: 50 to +120 C ( 58 to +248 F) Frequency response: 6.0 to 2500 Hz: ±10 % to 5000 Hz: ±3 db Ultra low-noise
More informationIntrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F
Replacement product: Model IS-3 Electronic pressure measurement Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F WIKA data sheet PE
More informationOperator s Manual. PP016 Passive Probe
Operator s Manual PP016 Passive Probe 2017 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials is strictly prohibited. Customers are permitted
More informationB Operating Instructions /
Conductivity/ultra-pure water measuring cells with a twin-electrode system Types 202922, 202923, 202924 and 202925 Glass conductivity cells, type 201080 Diaphragm tubes, type 201083 Compensation thermometer,
More informationRAD-IN/OUT-8D. Digital extension modules for the bidirectional wireless system. INTERFACE Data sheet _en_05. 1 Description
RAD-IN/OUT-D Digital extension modules for the bidirectional wireless system INTERFACE Data sheet 102122_en_0 PHOENIX CONTACT 2010-02-2 1 Description The RAD-ISM-...-SET-BD-BUS-ANT bidirectional wireless
More informationInductive Conductivity Meter ILM
1 Product Information ILM-2 ILM-3 Inductive Conductivity Meter ILM f o o d Application / Specified Usage Authorizations Inductive measurement of the specific conductivity of liquid media in the range of
More informationPHV RO. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design
High impedance passive probe Features: CeramCore TM Hybrid Probe Modular Construction Coaxial Design Interchangeable Spring Contact Tip Certificate of Calibration available on request Read-out BNC Connector
More information2-WIRE UNIVERSAL TEMPERATURE TRANSMITTER
SAFE INSTALLATION MANUAL (FM APPROVAL) -WIRE UNIVERSAL TEMPERATURE TRANSMITTER (HART communication, intrinsically safe/explosion-proof) MODEL B6U/B6U-B BEFORE USE... SAFETY PRECAUTIONS This manual describes
More informationMINI MCR-SL-UI-I-LP-NC
2-way isolation amplifier Data sheet 105263_en_02 PHOENIX CONTACT 2013-12-13 1 Description The configurable 2-way isolation amplifiers are used to electrically isolate, convert and filter standard signals.
More informationSupplied in carry case with additional accessories The PHV 1000-RO is a 400 MHz, standard sized, 100:1 passive probe designed for instruments
High impedance passive probe Features: CeramCore TM hybrid probe Modular construction Coaxial design Interchangeable spring contact tip Certificate of calibration available on request Read-out BNC connector
More informationE8EB-N0C2B E8EB-N0B2B
Slim Sensor Ideal for Workpiece Position and Original Checking The to 1 kpa model can be used for workpiece position checking. The to 1MPa model is ideal for original pressure checking. Degree of protection
More informationSignal converter for electromagnetic flowmeters
Quick Start Signal converter for electromagnetic flowmeters Electronic revision: ER 3.0.xx The documentation is only complete when used in combination with the relevant documentation for the flow sensor.
More informationType EXT with cable, probe sold
standard signal FF-GLT-Series Description Type INT for duct mounting, probe installed Type EXT with cable, probe sold Type AF for wall mounting, probe sold Characteristic features Standard signal 0...
More informationDigital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version
Electrical temperature measurement Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version WIKA data sheet TE 15.01 Applications Process industry Machine building
More informationOEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted
Electrical temperature measurement OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted WIKA data sheet TE 60.31 further approvals see page 10 Applications Machine building, plant
More informationTechnical Information Orbipore CPS91D and CPS91
TI00375C/07/EN/14.15 71299929 Products Solutions Services Technical Information Orbipore CPS91D and CPS91 ph electrodes, analog or with digital Memosens technology With open aperture for contaminated media,
More informationCapacitive level meters CLM 40 INSTRUCTIONS MANUAL. průmyslová elektronika
průmyslová elektronika INSTRUCTIONS MANUAL Capacitive level meters CLM 40 Before level meters usage please read instructions in this service manual and keep it for further use. Service manual is for all
More information09746E00. ATEX / IECEx NEC 505 NEC 506 NEC 500 Class I Class I Class II Class III Zone Zone Division Ex i
Temperature Transmitter Series 9182 www.stahl.de > One unit for nearly all temperature sensors indivdually configurable > Intrinsically safe input [Ex ia] IIC > Signal duplication possible > Galvanic isolation
More informationInstallation guide 971 SmartRadar LTi
Installation guide 971 SmartRadar LTi March 2009 Part no. 4416.715 Revision 3 Enraf B.V. P.O. Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100 Fax : +31 15 2701 111 E-mail : enraf-nl@honeywellenraf.nl
More informationMANUAL. Electrostatic Field Meter EFM 235 / 255. System Prof. Dr.-Ing. Hans Kleinwächter
MANUAL Electrostatic Field Meter EFM 235 / 255 System Prof. Dr.-Ing. Hans Kleinwächter 2 Manual EFM 120 Contents General information about electrostatics... 3 Origin of electrostatic charge... 3 Product
More informationGETTING STARTED GUIDE NI AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous
GETTING STARTED GUIDE NI 9228 8 AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous This document explains how to connect to the NI 9228. Note Before you begin, complete the software and hardware installation procedures
More informationTechnical Information Memosens CLS82D
TI01188C/07/EN/03.16 71318348 Products Solutions Services Technical Information Memosens CLS82D Hygienic conductivity sensors, digital with Memosens technology, cell constant k = 0.57 cm -1 Application
More informationJUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: 49 661 60030 Fax: 49 661 6003607 Email: mail@jumo.net Internet: www.jumo.net JUMO
More informationTAG5000 WIRELESS PHASER. Instruction Manual HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS U.S.A.
TAG5000 WIRELESS PHASER Instruction Manual TM HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS 60085 U.S.A. PHONE 847.473.4980 FAX 847.473.4981 website: www.hdelectriccompany.com DESCRIPTION
More informationOperating Instructions for Pressure Sensor with Ceramic Sensor Element. Model: SEN-9601
Operating Instructions for Pressure Sensor with Ceramic Sensor Element Model: SEN-9601 We don t accept warranty and liability claims neither upon this publication nor in case of improper treatment of the
More information3 Ex i Isolators. Temperature Transmitter without Output 0/ ma, with Limit Value Contact (Field Circuit Ex i) Type 9182/.
Ex i Isolators Temperature Transmitter without Output 0/4... 20 ma, with Limit Value Contact Field Circuit Ex i) One unit for nearly all temperature sensors indivdually configurable Intrinsically safe
More informationCondumax CLS21 and CLS21D
Technical Information Condumax CLS21 and CLS21D Conductivity sensors, analog or digital with Memosens technology Cell constant k = 1 cm -1 Application Measurements in media of medium and high conductivities:
More informationSafety Manual. Omni-ID UHF Tags
Safety Manual Omni-ID UHF Tags This page is intentionally left blank. Document Number 330556 (See Last Page for Revision Details) 2012 Extronics Limited. This document is Copyright Extronics limited. Extronics
More informationAnalogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting
Electrical Temperature Measurement Analogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting WIKA Data Sheet TE 24.01 Applications T Machinery, plant construction T Process industry
More informationJUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: 49 661 60030 Fax: 49 661 6003607 Email: mail@jumo.net Internet: www.jumo.net JUMO
More informationResistance thermometer Model TR30, compact design
Electrical temperature measurement Resistance thermometer Model TR30, compact design WIKA data sheet TE 60.30 Applications Machine building, plant and vessel construction Propulsion technology, hydraulics
More informationKSR-Level sensor KSR-Schwimmer-Magnetschalter KSR Contact Protection Relays
KSR-Level sensor KSR-Schwimmer-Magnetschalter 1001-5 KSR Contact Protection s 1 KSR Contact Protection Type KR230-Ex and KR24-Ex General Description The contact protection relays KR230-Ex and KR24-Ex transmit
More informationOPTIFLUX 1000 Handbook
OPTIFLUX 1000 Handbook Electromagnetic flow sensor in sandwich design The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE : IMPRINT
More information3 Ex i Isolators. Temperature Transmitter with Output 0/ ma, without Limit Value Contact (Field Circuit Ex i) Type 9182/.0-5.
Ex i Isolators Temperature Transmitter with Output 0/4... 20 ma, without Limit Value Contact Field Circuit Ex i) One unit for nearly all temperature sensors indivdually configurable Intrinsically safe
More informationMT MOTORI ELETTRICI. Installation, operation, maintenance and safety manual for motors used in hazardous areas 1-II-2G 21-II-2D
MT MOTORI ELETTRICI Installation, operation, maintenance and safety manual for motors used in hazardous areas 1-II-2G 21-II-2D TABLE OF CONTENTS 1. Introduction 2. Scope of application 3. Installation
More informationMeasuring and Controlling Unit Multronic OC
Measuring and Controlling Unit Multronic OC One channel measuring and controlling unit for ph / conductivity / Chlorine measuring Automatic temperature compensation Measuring and control behaviour simultaneously
More informationRosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna
Rosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna Product Data Sheet November 2016 DS 2000-10-008 Ed 03 Rosemount TGU 65 Tank Radar Gauge, Still Pipe Antenna is a marine pipe guided level gauge that
More informationDescription of equipment
Applicant: Electrical Apparatus: KROHNE Messtechnik GmbH & Co. KG Level Radar BM 70M-Ex Description of equipment The Types BM 70M-Ex Level Radar gauges serve to continuously measure the product level in
More informationOPTITEMP TT 10 C/R Technical Datasheet
OPTITEMP TT 10 C/R Technical Datasheet Analogue 2-wire temperature transmitter Temperature linear 4...20 ma output Rangeable with solder pads and potentiometers Easy wiring through large center hole The
More informationSubmersible pressure sensor For water and wastewater Model LW-1
Replacement product: Model LF-1 Level measurement Submersible pressure sensor For water and wastewater Model LW-1 WIKA data sheet LM 40.03 Applications Level measurement in rivers and lakes Deep well and
More informationJUMO BlackLine CR-GT/-EC/-GS Conductive 2-Electrode Conductivity Sensor
sales@jumo.co.uk info.us@jumo.net Vorläufig V Page 1/8 JUMO BlackLine CR-GT/-EC/-GS Conductive 2-Electrode Conductivity Sensor Cell constants K = 0.01 1.0 202922 Series for universal application, e.g.
More informationIsolating Spark Gaps. Isolating spark gaps at an additional earth ring conductor. Isolating spark gaps for Ex zones
Isolating Spark Gaps Isolating spark gaps at an additional earth ring conductor Isolating spark gaps for Ex zones 187 Protective Spark Gaps For use at roof supports for overhead lines For indirect connection
More informationMANUAL HYDROBAR. Manufactured by: 7991 CZ DWINGELOO 7990 AA DWINGELOO The Netherlands
MANUAL HYDROBAR WARNING Before installing the HYDROBAR, read the warnings and advisements on page 9 for personal and system safety, and for optimum performance, make sure you thoroughly understand the
More informationRefer to manual. Failure to read message could result in personal injury or serious damage to the equipment or both.
CERAVAC Transmitter CTR 100 N Short form manual 300544675_002_C0 Part numbers: 230 300V02 230 310V02 230 301V02 230 311V02 230 302V02 230 312V02 230 303V02 230 313V02 230 304V02 230 314V02 230 305V02 230
More informationWind Transmitter - with frequency output xx.xxx
THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for Use 021531/11/09 Wind Transmitter - with frequency output 4.3303.xx.xxx ADOLF THIES GmbH & Co. KG Hauptstraße
More information3 Ex i Isolators. Switching Repeater with Power Relay (250 V / 4 A) Type 9170/ Type 9170/
3 Ex i Isolators Switching Repeater with Power Relay Intrinsically safe input [Ex ia] IIC Galvanic isolation between input, output and power supply Open-circuit / short-circuit monitoring and messaging
More informationCE620. Piezoelectric accelerometer with integrated electronics. IECEx
Piezoelectric accelerometer with integrated electronics FEATURES» From the Vibro-Meter product line» Voltage output signal: 100 or 500 mv/ g» Frequency response: 2 Hz to 10 khz» Temperature range: 55 to
More informationDawson DDM190. Digital Multimeter User s Manual
Dawson DDM190 Digital Multimeter User s Manual TABLE OF CONTENTS LIMITED WARRANTY AND LIMITATION OF LIABILITY... 3 Out of the Box... 3 Accessories.. Error! Bookmark not defined. Safety Information... 7
More informationMANUAL HYDROBAR. Manufactured by: Fax: H/US/Hydrobar/ /00 Klay Instruments Page 1
MANUAL HYDROBAR WARNING Before installing the Hydrobar, read the warnings and advisements on the last page for personal and system safety, and for optimum performance, make sure you thoroughly understand
More informationOperator s Manual. PP022 Passive Probe
Operator s Manual PP022 Passive Probe 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com PP022 Passive Probe Instruction Manual 2017 Teledyne
More informationDawson DDM181. Pocket-Size Autorange Digital Meter User s Manual
Dawson DDM181 Pocket-Size Autorange Digital Meter User s Manual 1 Table of Contents LIMITED WARRANTY AND LIMITATION OF LIABILITY... 3 Out of the Box... 3 Accessories... 4 Important Safety Information...
More informationUltrasonic level measuring device, non-contact
Ultrasonic level measuring device, non-contact For level measurement up to 8 m 4... 20 ma/hart - 2 wires Suitable for solids ATEX approvals Type 8177 can be combined with... Type 8611 Type 8793 Type 8802-GB
More informationVBS TBS KTS BSS LFS UFS
VBS TBS KTS BSS LFS UFS TBS. Surge protection systems Order information and technical data Protection and isolating spark gaps Protection spark gap or isolating spark gap? Isolating spark gaps Protection
More informationInstallation guide 877 FDI Field Display & Interface
Installation guide 877 FDI Field Display & Interface November 2008 Part no. 4416.264 Rev. 5 Enraf BV PO Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100 Fax : +31 15 2701 111 E-mail : enraf-nl@honeywell.com
More informationUser manual. Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD
User manual Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD Contents Precautions Intended use General 1 Specification 3
More informationSMARTPAT COND 1200 Technical Datasheet
SMARTPAT COND 1200 Technical Datasheet Digital conductive conductivity sensor for the water and wastewater industry 2-wire loop powered sensor with integrated transmitter technology Wide measuring range
More informationTIDALFLUX 2300 F Technical Datasheet
Technical Datasheet Electromagnetic flow sensor for partially filled pipes Measurement in partially filled pipes up to DN1600 / 64" Patented, non-contact level measurement Measurement possible down to
More informationThermocouple - Transducer
Thermocouple - Transducer TM3 Characteristics: Thermocouple type J, K, R, S, T, E, B, N Linearity error < 1% Measurement range configurable -210...+1820 C Built-in cold junction compensation Galvanic 3-way
More information