Investigation of the film Kodak MX 125 in Agfa NDT S eco processing

Size: px
Start display at page:

Download "Investigation of the film Kodak MX 125 in Agfa NDT S eco processing"

Transcription

1 Investigation of the film Kodak MX 125 in Agfa NDT S eco processing Test Report-No.: BAM VIII.3 / 6568b Date: Name and address Kodak GmbH, of the customer: Hedelfinger Straße 60, Stuttgart, Germany Investigation: Measurement of the film system parameters according to DIN EN 584-1, pren 584-1:2004, ISO and ASTM E (2001) of the film Kodak MX125 Emulsion No. REF , LOT packed in evacuated Contactpak with lead screens (0.027 mm) and film processing according to the test specification Test specification: Investigation of the film Kodak MX125 in Agfa Structurix Eco chemistry for automatic processing in Agfa NDT S eco machine with 5 min cycle time, immersion time 63 seconds temperature 30 C Date of testing: Place of testing: BAM Berlin This report consists of 6 pages of text including one figure. The publication of the test report - also in parts - in word, writing and picture, sound, film and television requires the written consent of the Federal Institute. The same applies to each use of the test report in the same manner for the purposes of advertising. The consent can be revoked by the Federal Institute at any time.

2 BAM Test Report page 2 of 6 pages BAM Ref.:VIII.3/6568b Aim of the investigation: Processing of the emulsion LOT of the mentioned non-destructive testing film Kodak MX 125 in the processing chemistry Agfa Structurix Eco (processor AGFA NDT S eco, cycle time 5 min, immersion time 63 s, temperature 30 C) and measurement of the film system parameters gradient and granularity as well as determination of the gradient/granularity-ratio according to the standards DIN EN 584-1, pren 584-1:2004, ISO and ASTM E (2001). Test Program: 1. Exposure of the films in steps of optical density with an automatic film transport unit at conditions according to the above standards. 2. Development of the films. 3. Measurement of the optical density of the steps to determine the characteristic film curve and the film gradients at D-D 0 = 2.00 and D-D 0 = Measurement of the granularity at D-D 0 = 2.00 ± Instruments for processing and measurements: The exposures of the films (packed in Contactpak with mm lead screens) were performed with the filtered (8 mm Cu) 220 kv radiation of a highly stabilized X-ray facility (320 Isovolt D for dosimetry of the Seifert company) and an automated computer controlled film transport for exposure. For the measurement of the diffuse optical density a calibrated densitometer (Macbeth TD 502) was used, while the granularity was measured with a computer controlled microdensitometer of the type Joyce Loebl 3CS with a quadratic aperture of 100 µm x 100 µm. Film processing: For the automatic processing in the developer Agfa Structurix ecodev (part A: G , , part B: G , ) the following parameters were observed: temperature of the developer 30 C, immersion time 63s, cycle time of the process 5 minutes. Procedure for determination of the film system parameters according to EN 584-1, pren 584-1:2004, ISO and ASTM E (2001): The system parameters gradient G(2) and G(4) at the optical densities 2.00 and 4.00 above fog and base (D 0 ), the granularity σ D and the gradient/granularity-ratio G(2)/σ D are determined for a film system, which comprises a film type and the processing chemistry inclusive the processing. In detail the following processing steps were carried out: - procurement of films and processing chemicals,

3 BAM Test Report page 3 of 6 pages BAM Ref.:VIII.3/6568b - preparation of the developer and fixer according to the prescriptions of the manufacturer, - activation of the developer with a starter and approximately 1 m 2 film material - test exposure for the determination of the suitable exposure times for the steps of optical density, - exposure of at least 6 film samples for each system with filtered 220 kv radiation on at least 12 steps of optical density between D=1 and D= 4.7 in each case (see procedure in DIN EN 584-1, pren 584-1:2004, ISO , ASTM E (2001)), - processing of the films according to the prescriptions of the manufacturer and controlling by a film test strip according to DIN EN and ISO , - measurement of the processed films with the Macbeth densitometer and storage of the exposure data and the density values as the average of 3 measured values per density step in a file as well as the density of fog and film base, - evaluation of the data by approximation of the film curve by a third order polynomial and determination of the gradient G(2) for the density D - D 0 = 2.00 and the gradient G(4) at D - D 0 = 4.00 in accordance with DIN EN 584-1, pren 584-1:2004, ISO and ASTM E (2001), - measurement of the local density fluctuations of the films at density D - D 0 = 2.00 ± 0.05 with a microdensitometer (scan length 140 mm), - calibration of the microdensitometer into diffuse optical density with the aid of a film sample of the investigated film system, - calculation of the root mean square σ D of the optical density after conversion of the specular density into diffuse density, a digital high pass filtering to cut off low frequency noise below 0.1 mm -1 and the determination of the median according to ISO/DIS and pren 584-1:2004, - conversion of the granularity value into the corresponding value for measurements with a circular aperture with a diameter of 100 µm, - correction of the granularity value on the basis of measured density above fog and base to D - D 0 = 2.00, - calculation of the average of the relevant data from at least 6 film samples. Results: The diffuse density of fog and film base results in: D 0 = 0.23 The characteristic curve of the film shows up to the density 4.5 above fog and base the usual shape. Under the given conditions the exposed dose for density 2 above fog and base amounts to: K S = (8.5 ± 0.2) mgy

4 BAM Test Report page 4 of 6 pages BAM Ref.:VIII.3/6568b According to table 2 of DIN EN this corresponds to a CEN speed of: S = 125 The values of the gradients determined under the given conditions amount to: G (2) = 4.48 ± 0.02 G (4) = 7.71 ± 0.03 The granularity σ D at density D - D 0 = 2.00 results in: σ D = ± The gradient/granularity-ratio is calculated to: G(2) / σ D = 215 According to table 1 of DIN EN the limiting values for the film system class C3 amount to: G (2) 4.1, G (4) 6.8, σ D 0.023, G(2)/σ D 180 The film system consisting of emulsion LOT of the Kodak film type MX125, developed in the developer Agfa Structurix Eco (automatic processing in AGFA NDT S eco, cycle time 5 min), immersion time 63 seconds, temperature 30 C, meets the requirements of the film system class C3 of the standard DIN EN and pren 584-1:2004, system class I of ASTM E (2001) and system class T2 of ISO The values and its standard deviations mentioned above were determined from the respective values of the gradients and the granularity at the given densities (D - D 0 = 2.00 and D - D 0 = 4.00) from 6 film samples. The plots in the added graph (page 6, optical density (above fog and base) and gradient) result from the average of 6 film curves measured on 8 different film samples of the emulsion. Certified Agfa Structurix PMC strips with serial number E029, exp. date , were used to proof the state of the developing chemistry. The following indices were determined corresponding to EN and ISO : S x = 2.22 (+8%) compared to S r =2.05 and C x = 1.47 (-2%) comp. to C r = The certified Kodak Industrex preexposed control strips batch no. 074, exp. date , provide the following indices (as reference values the values for Kodak Industrex single part developer at 100s and 26 C have been used): S x = 2.11 (+11%) compared to S r =1.9 and C x = 1.10 (+0%) compared to C r = 1.1. The Kodak certified process monitor strips indicate deviations larger than 10% from the certified values, but the reference value provided in the certificate is for Kodak Industrex single part developer at 26 C. The AGFA certified process monitor strips (reference values for the

5 BAM Test Report page 5 of 6 pages BAM Ref.:VIII.3/6568b chemistry used, but 29 C developer temperature) indicates that the chemistry is within the specification (see EN 584-2, paragraph 6). Finally, the Agfa Structurix Thio-Test was performed to estimate the archivability of the processed films. The evaluation of the results according to ISO results in: L.E. = 100 (colour of the light step indicating a Life Expectancy of 100 years) Remark: This report summarizes only the results of the measurements at a single emulsion and is valid only for this special emulsion, especially it states no general classification of a film system. For a system classification corresponding to EN and pr EN 584-1:2004 a continuous measurement of all films of all manufactured emulsions by an independent third party is necessary, or, the continuous own product control in the house of the manufacturer in connection with the participation in a proficiency test according to pren 584-1:2004 for the harmonisation of the measuring procedures and algorithms and sample test measurements by an independent third party. Federal Institute for Materials Research and Testing (BAM) Bundesanstalt für Materialforschung und prüfung (BAM) D Berlin, Division VIII.3 Nondestructive Testing and Characterization; Radiation Techniques Working group VIII.35 Digital Radiology and Image Analysis By order By order Prof. Dr. U. Ewert Dr. U. Zscherpel

6 BAM Test Report page 6 of 6 pages BAM Ref.:VIII.3/6568b

Investigation of the film FUJIFILM IX 80 in TETENAL machine processing

Investigation of the film FUJIFILM IX 80 in TETENAL machine processing Investigation of the film FUJIFILM IX 80 in TETENAL machine processing Test Report No.: BAM 8.3 / 7315 d Date: 03. 02. 2012 Name and address: TETENAL AG & CO. KG Schützenwall 31-35 22844 Norderstedt Investigation:

More information

GB/T Translated English of Chinese Standard: GB/T

GB/T Translated English of Chinese Standard: GB/T Translated English of Chinese Standard: GB/T19348.1-2014 www.chinesestandard.net Sales@ChineseStandard.net GB NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA ICS 37.040.25; 19.100 J 04 GB/T 19348.1-2014

More information

Computed Radiography

Computed Radiography BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel

More information

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr Industrial Computed Radiography with storage phosphor imaging plates results of a classification of the system Dürr HD-CR 35 NDT Plus scanner using HD-IP Plus imaging plates (HD-IP + ) BAM reference BAM

More information

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr lndustrial Computed Radiography with storage phosphor imaging plates results of a classification of the system Dürr HD-CR 35 NDT scanner with blue HD- IP Plus imaging plates (HD-IP + ) BAM reference BAM

More information

Experiences of users in Digital Radiography

Experiences of users in Digital Radiography Computed Radiography Products & Applications Experiences of users in Digital Radiography Jimmy Opdekamp May Jimmy 2006Opdekamp Global Product Manager CR Int l Workshop Imaging NDT Chennai, 25-28 April

More information

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations uwe.ewert@bam.de Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations Uwe Ewert and Uwe Zscherpel BAM Federal Institute for

More information

Pre Conference Workshop on Advanced Digital imaging

Pre Conference Workshop on Advanced Digital imaging Pre Conference Workshop on Advanced Digital imaging Uwe Zscherpel, uwez@bam.de BAM Federal Institute for Materials Research and Testing, Division 8.3 Radiological Methods, Berlin, Germany 1 Content: 1.

More information

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Film Replacement in Radiographic Weld Inspection The New ISO Standard BAM Berlin Film Replacement in Radiographic Weld Inspection The New ISO Standard 17636-2 Uwe Ewert, Uwe Zscherpel, Mirko Jechow Requests and information to: uwez@bam.de 1 Outline - The 3 essential parameters

More information

Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography

Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography 7 th European-American Workshop on Reliability of NDE Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography Uwe EWERT, Uwe ZSCHERPEL, Justus VOGEL, Fangzhou ZHANG

More information

Strategies for Film Replacement in Radiography - a comparative study -

Strategies for Film Replacement in Radiography - a comparative study - IV Conferencia Panamericana de END Buenos Aires Octubre 2007 Strategies for Film Replacement in Radiography - a comparative study - 1 Uwe Ewert, 1 Uwe Zscherpel, 2 Klaus Bavendiek 1 BAM-Berlin, Federal

More information

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices Moving from film to digital: A study of digital x-ray benefits, challenges and best practices H.U. Pöhler 1 and N. D Ademo 2 DÜRR NDT GmbH & Co. KG, Höpfigheimer Straße 22, Bietigheim-Bissingen, 74321,

More information

Essential Parameters and Conditions for Optimum Image Quality in Digital Radiology

Essential Parameters and Conditions for Optimum Image Quality in Digital Radiology 18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa Essential Parameters and Conditions for Optimum Image Quality in Digital Radiology Uwe EWERT, Uwe ZSCHERPEL, Mirko

More information

5th Pan American Conference for NDT 2-6 October 2011, Cancun, Mexico

5th Pan American Conference for NDT 2-6 October 2011, Cancun, Mexico QUALIFICATION PROCESS AND INSPECTION VALIDATION OF COMPUTED RADIOGRAPHY TECHNIQUE FOR DWDI WELD INSPECTION Carla Alves MARINHO 1, José Maurício RABELLO, Marcos AIUB, Eduardo IGUCHI, Ricardo LOPES 2, Davi

More information

KODAK Medical X-ray Film / General Purpose / Blue / 3528 / MXB

KODAK Medical X-ray Film / General Purpose / Blue / 3528 / MXB TECHNICAL INFORMATION DATA SHEET Copyright, Eastman Kodak Company, 1998 KODAK Medical X-ray Film / General Purpose / Blue / 3528 / MXB 1) Description KODAK Medical X-ray Film / General Purpose / Blue /

More information

KODAK X-OMAT BT Film / 4530 / XBT

KODAK X-OMAT BT Film / 4530 / XBT TECHNICAL INFORMATION DATA SHEET Copyright, Eastman Kodak Company, 2003 KODAK X-OMAT BT Film / 4530 / XBT 1) Description KODAK X-OMAT BT Film / 4530 is a full-speed, blue-sensitive medical x-ray screen

More information

A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS

A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic More Info at Open Access Database www.ndt.net/?id=16394 A COMPARATIVE STUDY ON THE PERFORMANCE

More information

FOMA NDT SYSTEM X-Ray films for industrial radiography INDUX and FOMADUX Processing chemicals Представляем новинки shrinkage mark CD4

FOMA NDT SYSTEM X-Ray films for industrial radiography INDUX and FOMADUX Processing chemicals Представляем новинки shrinkage mark CD4 www.foma.eu Представляем новинки elongated/worm hole porosity shrinkage mark CD4 FOMA NDT SYSTEM X-Ray films for industrial radiography INDUX and FOMADUX Processing chemicals FILMS FOR INDUSTRIAL RADIOGRAPHY

More information

A Modified Detectability Criterion for Conventional Radiography Simulation

A Modified Detectability Criterion for Conventional Radiography Simulation 19 th World Conference on Non-Destructive Testing 2016 A Modified Detectability Criterion for Conventional Radiography Simulation David TISSEUR 1, Caroline VIENNE 1, Pierre GUÉRIN 2, Angela PETERZOL PARMENTIER

More information

KODAK INDUSTREX AA400 Film Versatility is the key.

KODAK INDUSTREX AA400 Film Versatility is the key. AIRCRAFT INSPECTION AERO-ENGINES ARCHEOLOGICAL ARTIFACTS ASSEMBLIES CASTINGS CONC ELECTRICAL COMPONENTS FORENSICS FORESTRY MUNITIONS BOMB DISPOSAL NUCLEAR APPLICAT PAINTINGS SCULPTURES PIPELINES TIRES

More information

KODAK Simulation Film / 7157

KODAK Simulation Film / 7157 TECHNICAL INFORMATION DATA SHEET Copyright, Eastman Kodak Company, 1998 1) Description KODAK Simulation Film / 7157 KODAK Simulation Film / 7157 is a high-speed, wide latitude, orthochromatic medical x-ray

More information

ISO INTERNATIONAL STANDARD. Photography Root mean square granularity of photographic films Method of measurement

ISO INTERNATIONAL STANDARD. Photography Root mean square granularity of photographic films Method of measurement INTERNATIONAL STANDARD ISO 10505 First edition 2009-05-15 Photography Root mean square granularity of photographic films Method of measurement Photographie Moyenne quadratique granulaire de films photographiques

More information

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations More Info at Open Access Database www.ndt.net/?id=16708 Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations Uwe EWERT, Uwe

More information

APPLICATION OF THE DIGITAL RADIOGRAPHY IN WELD INSPECTION OF GAS AND OIL PIPELINES

APPLICATION OF THE DIGITAL RADIOGRAPHY IN WELD INSPECTION OF GAS AND OIL PIPELINES APPLICATION OF THE DIGITAL RADIOGRAPHY IN WELD INSPECTION OF GAS AND OIL PIPELINES Davi F. OLIVEIRA, Edson V. MOREIRA, Aline S. S. SILVA, José M. B. RABELLO, Ricardo T. LOPES, Marcelo S. PEREIRA, Uwe ZSCHERPEL

More information

GE Sensing & Inspection Technologies. X-Ray Radiographic Film Systems

GE Sensing & Inspection Technologies. X-Ray Radiographic Film Systems GE Sensing & Inspection Technologies X-Ray Radiographic Film Systems GE Sensing & Inspection Technologies A global leader in technology-driven inspection solutions that deliver productivity, quality and

More information

ScienceDirect. Evaluation of Pipe Wall Thickness Based on Contrast Measurement using Computed Radiography (CR)

ScienceDirect. Evaluation of Pipe Wall Thickness Based on Contrast Measurement using Computed Radiography (CR) Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 69 ( 2014 ) 1216 1224 24th DAAAM International Symposium on Intelligent Manufacturing and Automation, 2013 Evaluation of Pipe

More information

ISO INTERNATIONAL STANDARD. Photography Root mean square granularity of photographic films Method of measurement

ISO INTERNATIONAL STANDARD. Photography Root mean square granularity of photographic films Method of measurement INTERNATIONAL STANDARD ISO 10505 First edition 2009-05-15 Photography Root mean square granularity of photographic films Method of measurement Photographie Moyenne quadratique granulaire de films photographiques

More information

Process Monitoring: Sensitometric Parameters

Process Monitoring: Sensitometric Parameters Process Monitoring: Sensitometric Parameters Revised 7/04 In addition to monitoring the key chemical and mechanical process parameters and processing KODAK Audit Strips, Process E-6, you should continue

More information

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers More info about this article: http://www.ndt.net/?id=22782 Abstract

More information

COMPUTED RADIOGRAPHY THE STATE OF THE ART

COMPUTED RADIOGRAPHY THE STATE OF THE ART 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 COMPUTED RADIOGRAPHY

More information

KODAK Panchromatic Separation Film 2238

KODAK Panchromatic Separation Film 2238 TECHNICAL INFORMATION DATA SHEET Copyright, Eastman Kodak Company, 2015 KODAK Panchromatic Separation Film 2238 1) Description KODAK Panchromatic Separation Film 2238 is a black-and-white film intended

More information

M. Senoner 1), Th. Wirth 1), W. E. S. Unger 1), M. Escher 2), N. Weber 2), D. Funnemann 3) and B. Krömker 3) INTRODUCTION

M. Senoner 1), Th. Wirth 1), W. E. S. Unger 1), M. Escher 2), N. Weber 2), D. Funnemann 3) and B. Krömker 3) INTRODUCTION Testing of Lateral Resolution in the Nanometre Range Using the BAM-L002 - Certified Reference Material: Application to ToF-SIMS IV and NanoESCA Instruments M. Senoner 1), Th. Wirth 1), W. E. S. Unger 1),

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16371-1 First edition 2011-10-01 Non-destructive testing Industrial computed radiography with storage phosphor imaging plates Part 1: Classification of systems Essais non destructifs

More information

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes

More information

Certified Reference Material

Certified Reference Material Bundesanstalt für Materialforschung und -prüfung (BAM) in co-operation with the Committee of Chemists of the GDMB Gesellschaft der Metallurgen und Bergleute e.v. Element Certified Reference Material BAM-M376a

More information

NEW POSSIBILITIES OF RADIATION CONTROL OF QUALITY OF WELDED JOINTS

NEW POSSIBILITIES OF RADIATION CONTROL OF QUALITY OF WELDED JOINTS NDT of Welded Joints NEW POSSIBILITIES OF RADIATION CONTROL OF QUALITY OF WELDED JOINTS V.A. TROITSKY E.O. Paton Electric Welding Institute, NASU 11 Bozhenko Str., 03680, Kiev, Ukraine. E-mail: office@paton.kiev.ua

More information

AVITONE PD 1 P-OS AND PD 3 P-OS

AVITONE PD 1 P-OS AND PD 3 P-OS AVITONE PD 1 P-OS AND PD 3 P-OS Orthochromatic continuous tone, direct positive copying film Avitone PD 1 p-os and PD 3 p-os are orthochromatic, continuous tone, direct positive copying films, coated on

More information

Using KODAK Control Strips for Processes AR-5 and AN-6

Using KODAK Control Strips for Processes AR-5 and AN-6 Using KODAK Control Strips for Processes AR-5 and AN-6 TECHNICAL DATA / CHEMICAL March 2010 TI-2220 INTRODUCTION This information is excerpted from a section on control strips in the process manual entitled

More information

GE Measurement & Control. X-Ray

GE Measurement & Control. X-Ray GE Measurement & Control X-Ray Radiographic Film Systems s GE Measurement & Control A global leader in technology-driven inspection solutions that deliver productivity, quality and safety to our customers.

More information

State of the Art Film Dosimetry

State of the Art Film Dosimetry State of the Art Film Dosimetry Micke A., Lewis D. Advanced Materials Ashland proprietary technology, patents pending Film Dosimetry Radiochromic Film EBT2/EBT3 One-Scan Protocol Multi-channel Film Dosimetry

More information

Gamex Lite 3.0 Program description and operating manual

Gamex Lite 3.0 Program description and operating manual Gamex Lite 3.0 Program description and operating manual Issue No. : 3.0 Date of Issue : 12-2013 Z.U.T. NDT SOFT http://www.ndtsoft.eu Copyright (c) 2013 by Z.U.T. NDT SOFT All Rights Reserved Disclaimer

More information

Application of Random Dot Model-to-Fog Granularity Caused by High-Energy Radiation of Silver Halide Emulsions in Color Systems

Application of Random Dot Model-to-Fog Granularity Caused by High-Energy Radiation of Silver Halide Emulsions in Color Systems Application of Random Dot Model-to-Fog Granularity Caused by High-Energy Radiation of Silver Halide Emulsions in Color Systems David E. Fenton Eastman Kodak Company Rochester, New York/USA Abstract The

More information

A panchromatic negative film for technical, industrial and aerial photography

A panchromatic negative film for technical, industrial and aerial photography :AVIPHOT PAN 200 A panchromatic negative film for technical, industrial and aerial photography :Aviphot Pan 200 PE1 is a panchromatic negative film, coated onto a transparent polyester base providing excellent

More information

:AVIPHOT COLOR N400 PE1

:AVIPHOT COLOR N400 PE1 :AVIPHOT COLOR N400 PE1 Panchromatic Negative Colour Film :Aviphot Color N400 PE1 is a panchromatic negative colour film, designed for aerial photography from low to medium altitudes. This film gives excellent

More information

Introduction of a Single Chip TLD System for Patient Dosimetry

Introduction of a Single Chip TLD System for Patient Dosimetry Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute

More information

KODAK MIN-R S Film / 4906

KODAK MIN-R S Film / 4906 TECHNICAL INFORMATION DATA SHEET Copyright, Eastman Kodak Company, 2002 KODAK MIN-R S Film / 4906 1) Description KODAK MIN-R S Film / 4906 is a medium speed, dual coated, ortho-sensitive medical x-ray

More information

Camera Negative Microfilm Data Sheet

Camera Negative Microfilm Data Sheet Camera Negative Microfilm Data Sheet KODAK IMAGELINK HQ, CS, CP and FS Microfilms Description Kodak Imagelink Microfilms have been specially designed to produce high-quality images from a wide variety

More information

AF3-024E COLOR REVERSAL FILMS 1. FEATURES AND USES 2. SPEED

AF3-024E COLOR REVERSAL FILMS 1. FEATURES AND USES 2. SPEED AF3-024E COLOR REVERSAL FILMS FUJICHROME 64T TYPE II Professional [RTPII] 1. FEATURES AND USES FUJICHROME 64T TYPE II Professional [RTP II] is an ISO 64 speed tungsten-type high image quality color reversal

More information

EASTMAN TRI-X Reversal Film 7278

EASTMAN TRI-X Reversal Film 7278 MPTVI Data Sheet XXXXXXXXXXX XX KODAK XX XX TInet XX XXXXXXXXXXX Technical Information Copyright, Eastman Kodak Company, 1994 1) Description EASTMAN TRI-X Reversal Film 7278 EASTMAN TRI-X Reversal Film

More information

A panchromatic negative film for aerial photography.

A panchromatic negative film for aerial photography. AVIPHOT PAN 80 A panchromatic negative film for aerial photography. Aviphot Pan 80 PE0 is a panchromatic aerial negative high resolution film, coated onto a transparent polyester base providing excellent

More information

A panchromatic negative film for aerial photography.

A panchromatic negative film for aerial photography. AVIPHOT PAN 80 A panchromatic negative film for aerial photography. Aviphot Pan 80 PE0 is a panchromatic aerial negative high resolution film, coated onto a transparent polyester base providing excellent

More information

OPTIMIZATION OF SCANNING PARAMETERS OF COMPUTED RADIOGRAPHY FOR CHARACTERIZATION OF EXPANDED POLYSTYRENE

OPTIMIZATION OF SCANNING PARAMETERS OF COMPUTED RADIOGRAPHY FOR CHARACTERIZATION OF EXPANDED POLYSTYRENE Danijela Hrman Marko Rakvin Damir Markučič ISSN 1333-1124 eissn 1849-1391 OPTIMIZATION OF SCANNING PARAMETERS OF COMPUTED RADIOGRAPHY FOR CHARACTERIZATION OF EXPANDED POLYSTYRENE UDC 620.179.152 Summary

More information

The advantages of variable contrast

The advantages of variable contrast Contrast Control with Color Enlargers Calibration of dichroic heads to ISO paper grades -an- by Ralph W. Lambrecht The advantages of variable contrast paper over graded paper have made it the prime choice

More information

Engineering Policy & Procedure

Engineering Policy & Procedure FPD > Engineering > Global Standards Engineering Policy & Procedure Revision History Number: G2-4 Section: G Subject: Radiographic Examination Procedure 1.0 SCOPE This procedure specifies the requirements

More information

The Influence of Film Processing Temperature and Time on Mammographic film Characteristics

The Influence of Film Processing Temperature and Time on Mammographic film Characteristics ffi;j*hjh[;~, 8: -9, 997 Bull Seh Health Sci, Okayama Univ (Original) The Influence of Film Processing Temperature and Time on Mammographic film Characteristics Toshinori MARUYAMA, Sachiko GOTO, Yoshiharu

More information

NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE

NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE 80 Hour Course Length 1.0 NDT Qualification and Introduction (3 Hours) 1.1 NDT Introduction 1.2 NDT Qualification and Certification 1.2.1 NAS 410 1.2.2 SNT-TC-1A

More information

FUJICHROME PROVIA 1600 Professional [RSP]

FUJICHROME PROVIA 1600 Professional [RSP] AF3-798E COLOR REVERSAL FILMS FUJICHROME PROVIA 1600 Professional [RSP] 1 FEATURES AND USES FUJICHROME PROVIA 1600 Professional [RSP] is an ultra-high speed daylight-type color reversal film designed for

More information

Technical Data Range of AgfaPhoto standard films

Technical Data Range of AgfaPhoto standard films Technical Data Range of AgfaPhoto standard films -1 Vista 100 Vista 200 Vista 400 APS star 200 CTprecisa 100 400 500 This brochure contains information on the quality and features of AgfaPhoto amateur

More information

THE CR SPECIALISTS HD-CR CR 35 NDT

THE CR SPECIALISTS HD-CR CR 35 NDT THE CR SPECIALISTS HD-CR CR 35 NDT COMPUTED RADIOGRAPHY SYSTEMS Digital Intelligence Ready to Change. www.duerr-ndt.com VERSATILE AND DEPENDABLE HIGH-TECH MADE IN GERMANY Highest resolution DÜRR NDT is

More information

RADIOGRAPHIC EXAMINATION PROCEDURE

RADIOGRAPHIC EXAMINATION PROCEDURE Page 1 of 19 Doc. No. KNS/RT/01.REV 0 RADIOGRAPHIC EXAMINATION PROCEDURE Page: 19 of 19 Date: 15.09.2012 RADIOGRAPHIC EXAMINATION PROCEDURE PREPARED & APPROVED BY NDE- Level III CERTIFIED BY REVIEWED BY

More information

This document is a preview generated by EVS

This document is a preview generated by EVS INTERNATIONAL STANDARD ISO 14096-1 First edition 2005-06-15 Non-destructive testing Qualification of radiographic film digitisation systems Part 1: Definitions, quantitative measurements of image quality

More information

X-Ray Radiographic Film Systems

X-Ray Radiographic Film Systems GE Sensing & Inspection Technologies X-Ray Radiographic Film Systems GEIT-40007EN_Filmbroch.indd 1 14-10-2008 11:42:03 GE Sensing & Inspection Technologies A global leader in technology-driven inspection

More information

AVIPHOT ASP 400 S PE1/PE0

AVIPHOT ASP 400 S PE1/PE0 AVIPHOT ASP 400 S PE1/PE0 Panchromatic negative B/W film for traffic infringement camera ASP 400 S is a panchromatic halftone aerial negative film with near IR sensitivity. The emulsion is coated onto

More information

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography ECNDT 26 - Poster 17 Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography Khairul Anuar MOHD SALLEH, Ab. Razak HAMZAH and Mohd Ashhar

More information

Standard Test Method for Determining the L/D Ratio of Neutron Radiography Beams 1

Standard Test Method for Determining the L/D Ratio of Neutron Radiography Beams 1 Designation: 91 (Reapproved 1996) An American National Standard Standard Test Method for Determining the L/D Ratio of Neutron Radiography Beams 1 This standard is issued under the fixed designation ; the

More information

Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition)

Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition) Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition) New Jersey Department of Environmental Protection Bureau of X-ray Compliance PO Box 420 Mail Code 25-01 Trenton, NJ 08625-0420 FAX 609-984-5811

More information

A COMPARISON BETWEEN ASTM E588 AND SEP 1927 RELATING RESOLUTION LIMITS AT DETERMINATION OF THE PURITY GRADE

A COMPARISON BETWEEN ASTM E588 AND SEP 1927 RELATING RESOLUTION LIMITS AT DETERMINATION OF THE PURITY GRADE 19 th World Conference on Non-Destructive Testing 2016 A COMPARISON BETWEEN ASTM E588 AND SEP 1927 RELATING RESOLUTION LIMITS AT DETERMINATION OF THE PURITY GRADE Daniel KOTSCHATE 1, Dirk GOHLKE 1, Rainer

More information

Monitoring Process RA-4 for Silver-Halide Digital Printers

Monitoring Process RA-4 for Silver-Halide Digital Printers CURRENT INFORMATION SUMMARY July 2005 CIS-272 Monitoring Process RA-4 for Silver-Halide Digital Printers Photographic Systems Management involves using a comprehensive quality control program to ensure

More information

RA Duplicating Film X-OMAT 5000 RA Processor; RP X-OMAT Chemicals; 90 Second Cycle; Diffuse Visual Densitometry. Density

RA Duplicating Film X-OMAT 5000 RA Processor; RP X-OMAT Chemicals; 90 Second Cycle; Diffuse Visual Densitometry. Density Technical Data Sheet Issued 2014-02 RA Duplicating Film / PicturePerfect Duplicating Film RA Duplicating Film provides excellent image quality and accurate information due to its sensitometric curve shape.

More information

Heiland electronic GmbH TD / TD1 / TD2. B&W-Densitometers. USERS MANUAL Version 5

Heiland electronic GmbH TD / TD1 / TD2. B&W-Densitometers. USERS MANUAL Version 5 Heiland electronic GmbH TD / TD1 / TD2 B&W-Densitometers USERS MANUAL Version 5 2 Table of Contents 1. GENERAL INFORMATION...4 2. SAFETY REGULATIONS...5 3. AREA OF APPLICATIONS...5 4. INSTRUMENT DESCRIPTION...6

More information

Using KODAK Negative Film Processing Chemicals, Type L

Using KODAK Negative Film Processing Chemicals, Type L Using KODAK Negative Film Processing Chemicals, Type L CURRENT INFORMATION SUMMARY December 2017 CIS-253 KODAK Negative Film Processing Chemicals, Type L, and KODAK Rinse Tablets are specially made for

More information

KODAK EKTASCAN IR Laser Imaging Film / 1356 / EIR

KODAK EKTASCAN IR Laser Imaging Film / 1356 / EIR TECHNICAL INFORMATION DATA SHEET TI5 Issued 29-9 Carestream Health, Inc., 29 1) Description KODAK EKTASCAN IR Laser Imaging Film / 1356 is a very fine grain, high resolution, black-and-white film suitable

More information

EASTMAN EXR 200T Film 5287, 7287

EASTMAN EXR 200T Film 5287, 7287 TECHNICAL INFORMATION DATA SHEET TI2124 Issued 6-94 Copyright, Eastman Kodak Company, 1994 EASTMAN EXR 200T Film 5287, 7287 1) Description EASTMAN EXR 200T Film 5287 (35 mm) and 7287 (16 mm) is a medium-high

More information

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications 19 th World Conference on Non-Destructive Testing 2016 Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications David WALTER 1, Uwe ZSCHERPEL 1, Uwe EWERT 1 1 BAM Bundesanstalt

More information

FUJICOLOR NPC 160 PROFESSIONAL [NPC]

FUJICOLOR NPC 160 PROFESSIONAL [NPC] AF3-065E COLOR NEGATIVE FILMS FUJICOLOR NPC 160 PROFESSIONAL [NPC] 1. FEATURES AND USES FUJICOLOR NPC 160 PROFESSIONAL [NPC] is an ISO 160 daylight color negative film that is suitable for highcontrast

More information

EASTMAN EXR 500T Film 5298

EASTMAN EXR 500T Film 5298 TECHNICAL INFORMATION DATA SHEET TI2082 Revised 12-98 Copyright, Eastman Kodak Company, 1993 1) Description EASTMAN EXR 500T Films 5298 (35 mm) is a high-speed tungsten-balanced color negative camera film

More information

Nuclear tracks analysis in passive Radon detector using scanner

Nuclear tracks analysis in passive Radon detector using scanner WORKSHOP: The second radon-in-field international intercomparison for passive measurement devices: dwellings and workplaces. Milano 21-22 sept 2017 Nuclear tracks analysis in passive Radon detector using

More information

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data Description GAFCHROMIC HD-810 dosimetry film is designed for the measurement of absorbed dose of high-energy

More information

NDT Supply.com 7952 Nieman Road Lenexa, KS USA

NDT Supply.com 7952 Nieman Road Lenexa, KS USA Durr NDT ScanX Computed Radiography System The Workhorse of Portable Digital Radiography in NDT There are 2 models available: ScanX Discover HR ScanX Discover HC ScanX Discover HR ScanX Discover HC The

More information

DRAWINGS ON 35mm MICROFILM INSPECTION PROCEDURES SILVER MICROFILM

DRAWINGS ON 35mm MICROFILM INSPECTION PROCEDURES SILVER MICROFILM AT& T PRACTICE AT& T 006-110-500 Standard Issue 6, March 1987 DRAWINGS ON 35mm MICROFILM INSPECTION PROCEDURES SILVER MICROFILM 1. GENERAL 1.01 This practice covers inspection procedures for 35mm silver

More information

KODAK VISION Expression 500T Color Negative Film / 5284, 7284

KODAK VISION Expression 500T Color Negative Film / 5284, 7284 TECHNICAL INFORMATION DATA SHEET TI2556 Issued 01-01 Copyright, Eastman Kodak Company, 2000 1) Description is a high-speed tungsten-balanced color negative camera film with color saturation and low contrast

More information

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University

More information

KODAK AEROCHROME III MS Film 2427

KODAK AEROCHROME III MS Film 2427 AS-2569 KODAK AEROCHROME III MS Film 2427 KODAK AEROCHROME III MS Film 2427 is a very fine-grain, medium-speed color-reversal aerial camera film. This film has excellent color rendition and good image

More information

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China 17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China Real-time Radiographic Non-destructive Inspection for Aircraft Maintenance Xin Wang 1, B. Stephen Wong 1, Chen Guan Tui

More information

1 MPTVI DATA SHEET XXXXXXXXXXX

1 MPTVI DATA SHEET XXXXXXXXXXX 1 MPTVI DATA SHEET XXXXXXXXXXX TI1664 XX KODAK XX Reissued 6-92 XX TInet XX XXXXXXXXXXX ================================================================== TECHNICAL INFORMATION Copyright, Eastman Kodak

More information

NDT-PRO Services expands service offering

NDT-PRO Services expands service offering NDT-PRO Services expands service offering NDT-PRO Services announced the formal release of two advanced NDT methods, Phased Array (including TOFD) and Eddy Current. What are they and where are the used?

More information

DOSE REDUCTION BY USE OF DIGITAL X-RAY DETECTORS. Bernhard REDMER

DOSE REDUCTION BY USE OF DIGITAL X-RAY DETECTORS. Bernhard REDMER 14.03.2016 DOSE REDUCTION BY USE OF DIGITAL X-RAY DETECTORS Bernhard REDMER Digital Technology and Networking Imaging Plates Scanner for Imaging Plates Data Transmission Reporting Computed Radiography

More information

Preliminary Investigation of a New X-ray Film

Preliminary Investigation of a New X-ray Film 256 J. Appl. Cryst. (1981). 14, 256-260 Preliminary Investigation of a New X-ray Film BY S. AB~HAMSSON Department of Structural Chemistry, Faculty of Medicine, University of G6teborg, S-400 33 Gb'teborg

More information

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF

More information

EASTMAN PLUS-X Reversal Film / 7276

EASTMAN PLUS-X Reversal Film / 7276 MPTVI Data Sheet XXXXXXXXXXX XX KODAK XX XX TInet XX XXXXXXXXXXX Technical Information Copyright, Eastman Kodak Company, 1995 1) Description EASTMAN PLUS-X Reversal Film / 7276 EASTMAN PLUS-X Reversal

More information

Determination of Optimum X-Ray Tube Output Parameters kv and ma for Digital Radiography Testing of Welded Tubes

Determination of Optimum X-Ray Tube Output Parameters kv and ma for Digital Radiography Testing of Welded Tubes More info about this article: http://www.ndt.net/?id=21129 Determination of Optimum X-Ray Tube Output Parameters kv and for Digital Radiography Testing of Welded Tubes AyazJhanorwala, SatishTilva, L &

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés

More information

Digital Quality Control Phantom for X-ray Mammography Image

Digital Quality Control Phantom for X-ray Mammography Image Digital Quality Control Phantom for X-ray Mammography Image Yuichi Nagai, Atsuko Takada National: Cancer Hospital East Department of Radiology Collaborated by Nobuyuki Niwa : Kyoto Kagaku Co., Ltd. Introduction

More information

Using the KODAK Negative Film Processing Cartridges FC1 and FC2

Using the KODAK Negative Film Processing Cartridges FC1 and FC2 Using the KODAK Negative Film Processing Cartridges FC1 and FC2 CURRENT INFORMATION SUMMARY December 2017 CIS-254 The KODAK Negative Film Processing Cartridges FC1 and FC2 and KODAK Rinse Tablets are designed

More information

KODAK PRIMETIME 640T Teleproduction Film / 5620,7620

KODAK PRIMETIME 640T Teleproduction Film / 5620,7620 TECHNICAL INFORMATION DATA SHEET TI2299 Issued 0-96 Copyright, Eastman Kodak Company, 996 KODAK PRIMETIME 640T Teleproduction Film / 5620,7620 ) Description KODAK PRIMETIME 640T Teleproduction Film / 5620,7620

More information

Results of Proof-of-Concept 50keV electron multi-beam Mask Exposure Tool (emet POC)

Results of Proof-of-Concept 50keV electron multi-beam Mask Exposure Tool (emet POC) Results of Proof-of-Concept 50keV electron multi-beam Mask Exposure Tool (emet POC) Elmar Platzgummer *, Christof Klein, and Hans Loeschner IMS Nanofabrication AG Schreygasse 3, A-1020 Vienna, Austria

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Using the KODAK Negative Film Processing Cartridges FC1 and FC2

Using the KODAK Negative Film Processing Cartridges FC1 and FC2 CURRENT INFORMATION SUMMARY May 2005 CIS-254 Using the Processing Cartridges FC1 and FC2 The Processing Cartridges FC1 and FC2 and KODAK Rinse Tablets are designed for use in FUJI FP363SC and FP563SC Film

More information

HPX1-Plus. For Non-Destructive Testing. THE BENCHMARK IN COMPUTED RADIOGRAPHY.

HPX1-Plus. For Non-Destructive Testing.  THE BENCHMARK IN COMPUTED RADIOGRAPHY. HPX1-Plus For Non-Destructive Testing THE BENCHMARK IN COMPUTED RADIOGRAPHY. Introducing the The HPX family of products has earned many of the NDT industry s most prestigious awards. It was no surprise

More information

Certified Reference Material

Certified Reference Material Certified Reference Material BAM-F012 (brief excerpt) Glass-based Multi-emitter Fluorescence Standard as day-to-day intensity and wavelength standard for instrument performance validation (IPV) of fluorescence

More information