Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P)

Size: px
Start display at page:

Download "Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P)"

Transcription

1 Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P) The Cressington sputter system is designed for only one purpose which is the deposition of gold onto a sample to reduce charging when analyzed in the SEM. This system is only available for reservation through a NanoFab Staff member. NanoFab staff will make the reservation and operate the system for the intended user. Here is a brief summary about film thickness from the Cressington Operations Manual. The thickness of film required to prevent charging is very dependent on the sample and the operating voltage in the SEM. A thickness of between 3nm 5nm will usually prevent charging. The film needs to be thin enough not to obscure fine detail, but thick enough to prevent charging. Operating the system: 1

2 1) Verify that the system is powered OFF (Fig. 1). (Fig. 1) 2) Open the UPPER LID (Fig. 2a) of the chamber, if the upper lid will not lift up easily then the chamber may not be fully vented; in this case you must press and hold the RED VENT BUTTON (Fig. 2b) located on the left side of the table until the chamber lid will lift up. (Fig. 2a) (Fig. 2b) 3) Remove the UPPER GLASS CHAMBER and the ADAPTER RING (Fig. 3) that is located underneath it and sit them off to the side. This allows for easy access to the sample table. 2

3 (Fig. 3) 4) Sit the sample to be coated onto the center of the SAMPLE TABLE (Fig. 3a) facing upwards. The table can be tilted if better side coverage is needed on the sample. This is done by rotating the LARGE BLACK KNOB (Fig. 3b) at the rear of the chamber. (Fig. 3a) (Fig. 3b) 5) Ensure that the CRYSTAL HOLDER (Fig. 4) is facing upwards towards the target. 3

4 (Fig. 4) 6) Reinstall the GLASS UPPER CHAMBER and ADAPTER RING and lower the UPPER LID (See Fig. 2a & 3). 7) Turn on the MAIN POWER SWITCH (Fig. 5). The mechanical pump will start and the chamber will begin pumping down. Within a minute the front panel gauge should begin to show a drop in vacuum pressure. (Fig. 5) 4

5 8) Let the system pump down to <.04 mbar as measured by the lower range of the FRONT PANEL GAUGE (Fig. 6). This should typically occur within 10 minutes of turning the system on. (Fig. 6) 9) Press the AUTO - MANUAL (FIG. 7) button to illuminate the MANUAL LED located directly above the button. (Fig. 7) 5

6 10) The DEPOSITION TIME is displayed in the TIMER section of the front panel. The deposition rate of this system is approximately.33 Å per second. To set the timer for the deposition thickness desired press and hold the PAUSE/TEST button (Fig. 8) and use the UP/DOWN buttons to the right of the TIME DISPLAY to adjust the deposition time. (Fig. 8) 11) Turn ON the Thickness monitor power, wait a few seconds for it to initialize and then momentarily press the ZERO button (Fig. 9). (Fig. 9) 6

7 12) Press the LEAK button (Fig. 10) and give the system a minute for the pressure to stabilize. Ensure that the SPUTTERING PRESSURE is set for.07 mbar -.08 mbar on the front panel gauge (See Fig. 6). If the pressure is incorrect then it will have to be adjusted, this is performed by turning the needle valve located on the back of the system (Contact NanoFab Staff for this). LEAK button (Fig. 10) 13) Now press the AUTO/MANUAL button again (See Fig. 7), the AUTO LED should now be illuminated. 14) If you need a highly uniform film on the surface of the sample then turn ON the sample table rotation by turning the SPEED SELECTION KNOB (Fig. 11) clockwise, there are 5 different rotation speeds that can be selected. (Fig. 11) 7

8 15) Open the shutter by turning the SHUTTER ACTUATION KNOB (Fig. 12) clockwise until it stops rotating. (Fig. 12) 16) Press the CYCLE button (Fig. 13) to start the automatic deposition cycle. Once this has been started the system will run a number of purge cycles before beginning deposition. A glow will be evident from the plasma and the timer will begin counting down once the deposition has started. The timer will reset to your preset amount when the deposition has completed. (Fig. 13) 17) Once the deposition has completed you may turn OFF the rotation (See Fig. 11). Then turn OFF both the system power (See Fig. 1) and the deposition monitor power (See Fig. 9). 18) Rotate the SHUTTER ACTUATION KNOB (See Fig. 12) counterclockwise approximately ¼ turn, back to the point it covers the target. 8

9 19) To vent the system you must press and hold the RED VENT BUTTON (See Fig. 2b) located to the left of the system. Once the lid lifts easily by hand you can release the RED VENT BUTTON. 20) Remove the UPPER GLASS CHAMBER and the ADAPTER RING (See Fig. 3) that is located underneath it and sit them off to the side. 21) Remove your sample and then reinstall the UPPER GLASS CHAMBER and the ADAPTER RING, then lower the UPPER LID. Leave the system in its powered down state when you are finished. 9

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016

CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016 CHA ELECTRON BEAM EVAPORATOR SYSTEM OPERATING MANUAL Version: 2.0 August 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1.0 DESCRIPTION.... 3 2.0 SAFETY....... 3

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual NanoFactor NVG-200A Silicon Grinder Version 1.1 Page 1 of 18 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Possible Grinding Materials

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

EXPERIMENT # 3: Oxidation and Etching Tuesday 2/3/98 and 2/5/98 Thursday 2/10/98 and 2/12/98

EXPERIMENT # 3: Oxidation and Etching Tuesday 2/3/98 and 2/5/98 Thursday 2/10/98 and 2/12/98 EXPERIMENT # 3: Oxidation and Etching Tuesday 2/3/98 and 2/5/98 Thursday 2/10/98 and 2/12/98 Experiment # 3: Oxidation of silicon - Oxide etching and Resist stripping Measurement of oxide thickness using

More information

isel=vakuum UV exposure device

isel=vakuum UV exposure device isel=vakuum UV exposure device Manual I isel-vakuum UV exposure device About these instructions In this instructions, you find different symbols marking important information for easy and fast reference.

More information

Thin Film Deposition

Thin Film Deposition Thin Film Deposition Section Eleven 11 11.1 General Information 2 11.2 Deposition Monitors 3 11.3 Crystal Feedthroughs 4 11.4 s 5 11.5 Cables, s, Crystals & Accessories 6 Nor-Cal Products, Inc. 1967 South

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent

More information

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an

More information

EXPERIMENT # 3: Oxidation and Etching Week of 1/31/05 and 2/7/05

EXPERIMENT # 3: Oxidation and Etching Week of 1/31/05 and 2/7/05 EXPERIMENT # 3: Oxidation and Etching Week of 1/31/05 and 2/7/05 Experiment # 3: Oxidation of silicon - Oxide etching and Resist stripping Measurement of oxide thickness using different methods The purpose

More information

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users are expected to have read and understood this document.

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

First published : May 1997 Fourth edition : January No

First published : May 1997 Fourth edition : January No First published : May 1997 Fourth edition : January 2006 No. 050153 INTRODUCTION Thank you for your purchasing Kansai Special's FX Series. Read and study this instruction manual carefully before beginning

More information

Strata DB235 FESEM FIB

Strata DB235 FESEM FIB Strata DB235 FESEM FIB Standard Operating Procedure Revision: 5.0 Last Updated: August 16/2016, revised by Li Yang Overview This document will provide a detailed operation procedure of the Focused Ion

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

Thin Film Deposition

Thin Film Deposition Thin Film Deposition Section Eleven 11 11.1 General Information 2 11.2 Deposition Monitors 3 11.3 Crystal Feedthroughs 4 11.4 Crystal s 5 11.5 Cables, s, Crystals & Accessories 6 Nor-Cal Products, Inc.

More information

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014 1. Utility Requirements a. System power is supplied by two 120 VAC/20 A circuits. When doing maintenance

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name:

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Equipment Name: Coral Name: Nanoimprinter Revision Number: 1.1 Model: NX-B200 Revisionist: M. Fisher Location: Bay 4 Date: 2/12/2010 1 Description Nanonex NX-B200 nanoimprinter is another method of transfer

More information

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check

More information

7. Michelson Interferometer

7. Michelson Interferometer 7. Michelson Interferometer In this lab we are going to observe the interference patterns produced by two spherical waves as well as by two plane waves. We will study the operation of a Michelson interferometer,

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

Setting up and Using Digital Micrometer Controlled Lapping Fixtures

Setting up and Using Digital Micrometer Controlled Lapping Fixtures Setting up and Using Digital Micrometer Controlled Lapping Fixtures Purpose polishing fixtures are commonly used in materials preparation labs around the world. Lapping fixtures provide stability, precision,

More information

Gencoa 3G Circular Magnetron

Gencoa 3G Circular Magnetron Gencoa 3G Circular Magnetron Presenting Gencoa s 3 rd generation range of circular magnetrons November 2017 www.gencoa.com 1 Contents Product Overview Introduction 3G Design Features 3G Mechanical Options

More information

Niobium Coating of Copper Cavities by UHV Cathodic Arc: progress report

Niobium Coating of Copper Cavities by UHV Cathodic Arc: progress report Niobium Coating of Copper Cavities by UHV Cathodic Arc: progress report L. Catani, A. Cianchi, D. Digiovenale, J. Lorkiewicz, Prof. S. Tazzari, INFN-Roma "Tor Vergata", Italy Roberto Russo, Istituto di

More information

HEX02 EMBOSSING SYSTEM

HEX02 EMBOSSING SYSTEM HEX02 EMBOSSING SYSTEM LOCATION: Hot Embossing Area PRIMARY TRAINER: 1. Scott Munro (2-4826, smunro@ualberta.ca) OVERVIEW The hot embosser is available to users who require polymer mold fabrication. This

More information

FE-SEM SU-8020 Operating manual (Preliminary version)

FE-SEM SU-8020 Operating manual (Preliminary version) FE-SEM SU-8020 Operating manual (Preliminary version) 2016/04/11 Seimitsu Bunseki sitsu lab. Starting up 1.Turn on the Display switch. Windows OS is starting up 2. Select the user SU-8000. 3. Click the

More information

Tinker Tuesday Project - Stainless Steel Tumblers

Tinker Tuesday Project - Stainless Steel Tumblers Tinker Tuesday Project - Stainless Steel Tumblers 1. Open CorelDRAW and create a new document. Near the top left corner of the screen, click File, and then click Import on the resulting menu. Select an

More information

DELUXE AIR PRESSURE CASTING MACHINE MODEL # S

DELUXE AIR PRESSURE CASTING MACHINE MODEL # S DELUXE AIR PRESSURE CASTING MACHINE MODEL # S INSTALLATION 99-002-D4AP 99-002-D4AP-A MAINTENANCE The points to look for in order to obtain a long and trouble free operation are: a. See that the oil fog

More information

SEM OPERATION IN LOW VACUUM MODE

SEM OPERATION IN LOW VACUUM MODE SEM OPERATION IN LOW VACUUM MODE Instructions for JEOL 5800 LV The EVAC light of the SEM specimen chamber should be already lit when you approach the SEM & the SEM will have been left in the high vacuum

More information

Model SU3500 Scanning Electron Microscope

Model SU3500 Scanning Electron Microscope Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction

More information

Hydraulic Clamp Carrier. Installation & Operation Manual

Hydraulic Clamp Carrier. Installation & Operation Manual Hydraulic Clamp Carrier Installation & Operation Manual Hydraulic Clamp Carrier Installation & Operation Manual Quick Machinery Company 8272 Peninsula Drive Kelseyville, CA 95451 phone: (707) 272-6719

More information

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

1.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

Maintenance Information

Maintenance Information 16601023 Edition 2 January 2014 Air Impact Wrench 2705P1 Maintenance Information Save These Instructions Product Safety Information WARNING Failure to observe the following warnings, and to avoid these

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

EMO. Service Instruction. created by Frank Weithöner. Table of contents: Special Tools Assembling Mixing Chamber

EMO. Service Instruction. created by Frank Weithöner. Table of contents: Special Tools Assembling Mixing Chamber EMO Service Instruction created by Frank Weithöner Table of contents: - Special Tools Disassembling Mixing Chamber Assembling Mixing Chamber Adjustment Rotor / Level Indicator Unit Temperature Compensating

More information

2. OPERATING PROCEDURES 3. WARNING AND SAFETY HAZARDS 4. QUALITY CONTROL 5. MAINTENANCE 6. INSTALLATION 7. WARRANTY AND LIMITATIONS

2. OPERATING PROCEDURES 3. WARNING AND SAFETY HAZARDS 4. QUALITY CONTROL 5. MAINTENANCE 6. INSTALLATION 7. WARRANTY AND LIMITATIONS SUSS. MJB 3 MASK ALIGNER Operator's Reference Manual Model No. Serial No. Date of Manufacture. 1. GENERAL DESCRIPTION AND PRINCIPLES OF OPERATION ( 2. OPERATING PROCEDURES 3. WARNING AND SAFETY HAZARDS

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

UK10 UK11. First published: June No.KX03023

UK10 UK11. First published: June No.KX03023 UK10 UK11 First published: June 2003 No.KX03023 INTRODUCTION Thank you for purchasing Kansai Special s UK series machine. Please study this instruction manual carefully before operating the machine. 1.

More information

DISCO DICING SAW SOP. April 2014 INTRODUCTION

DISCO DICING SAW SOP. April 2014 INTRODUCTION DISCO DICING SAW SOP April 2014 INTRODUCTION The DISCO Dicing saw is an essential piece of equipment that allows cleanroom users to divide up their processed wafers into individual chips. The dicing saw

More information

CLUSTERLINE RAD VERSATILE DYNAMIC SPUTTER SYSTEM OPTOELECTRONICS, MEMS, PHOTONICS, WIRELESS

CLUSTERLINE RAD VERSATILE DYNAMIC SPUTTER SYSTEM OPTOELECTRONICS, MEMS, PHOTONICS, WIRELESS CLUSTERLINE RAD VERSATILE DYNAMIC SPUTTER SYSTEM OPTOELECTRONICS, MEMS, PHOTONICS, WIRELESS CLUSTERLINE RAD Enabling your roadmap in thin film deposition The combination of Evatec s process know-how and

More information

This product is designed for a water supply pressure range of 20 psi (137kPa) minimum to 80psi (550 kpa) maximum.

This product is designed for a water supply pressure range of 20 psi (137kPa) minimum to 80psi (550 kpa) maximum. We appreciate your commitment to Kohler quality. Please take a few minutes to review this manual before you start installation. If you encounter any installation or performance problems, please don t hesitate

More information

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control

More information

1. Specimen Holder Removal, Loading, and Insertion

1. Specimen Holder Removal, Loading, and Insertion OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope

More information

Syncore Analyst Technical data sheet

Syncore Analyst Technical data sheet STOP PROGRAMM SET DOWN UP DOWN UP Syncore Analyst Technical data sheet With the Syncore Analyst, BUCHI is offering a solution for a fast, gentle and environmentfriendly multiple sample concentration to

More information

Adjustable Center Crossheads Operations Manual

Adjustable Center Crossheads Operations Manual Plastic Extrusion Tooling Specialists ph: 800.272.8878 Adjustable Center Crossheads Operations Manual BH25, BH30, BH75, BH225, BH250 & BH750 B&H TOOL COMPANY FIRST IN EXTRUSION TOOLING Dear Customer, Thank

More information

FBX1104P FBX1104 FBX1106P FBX1106

FBX1104P FBX1104 FBX1106P FBX1106 FBX1104P FBX1104 FBX1106P FBX1106 Second edition : September 2004 No. 040037 INTRODUCTION Thank you for your purchasing Kansai Special's FBX Series. Read and study this instruction manual carefully before

More information

Tinker Tuesday Project - Drinking Glasses

Tinker Tuesday Project - Drinking Glasses Tinker Tuesday Project - Drinking Glasses 1. Open CorelDRAW and create a new document. Near the top left corner of the screen, click File, and then click Import on the resulting menu. Select an image from

More information

2. Raise HT to 200kVby following the procedure explained in 1.6.

2. Raise HT to 200kVby following the procedure explained in 1.6. JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,

More information

Band-Master ATS Nano Pneumatic Banding Tool Operating Instructions

Band-Master ATS Nano Pneumatic Banding Tool Operating Instructions Band-Master ATS 601-118 Nano Pneumatic Banding Tool CONTENTS 601-118 Overview... 3 Safety.... 5 Initial Tool Set-up... 5 Regulator assembly mounting... 5 Attach tool head to regulator.... 6 Operating instructions...

More information

Bodyspray INSTALLATION INSTRUCTIONS

Bodyspray INSTALLATION INSTRUCTIONS Bodyspray P22077 P22079 2014 KALLISTA 1 of 8 THANK YOU FOR CHOOSING KALLISTA Bodyspray We appreciate your commitment to KALLISTA quality products. Please take a moment to review this manual before you

More information

SCREEN PRINTING V0.3 DRAFT. Not for Release STANDARD OPERATING PROCEDURE

SCREEN PRINTING V0.3 DRAFT. Not for Release STANDARD OPERATING PROCEDURE V0.3 DRAFT SCREEN PRINTING STANDARD OPERATING PROCEDURE Not for Release Professor Chris Bondy, School of Media Science, RIT Akshat Pardiwala, Graduate Student, MS, Print Media Mitchell Morgan, Undergrad

More information

1 Placing particles on the slide

1 Placing particles on the slide Aerosols Transport Particle Removal Experiment E.S. Kenney, J.A. Taylor, and G. Ahmadi 1 Placing particles on the slide 1.1 Swing the light beneath the test section down and to the left. Figure 1: Light

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

plasmonic nanoblock pair

plasmonic nanoblock pair Nanostructured potential of optical trapping using a plasmonic nanoblock pair Yoshito Tanaka, Shogo Kaneda and Keiji Sasaki* Research Institute for Electronic Science, Hokkaido University, Sapporo 1-2,

More information

INSTRUCTION BX1425P,PSM,PTV BX1433P,PSM,PTV BX1025P,PSM BX1033P,PSM. No First published : November 1997

INSTRUCTION BX1425P,PSM,PTV BX1433P,PSM,PTV BX1025P,PSM BX1033P,PSM. No First published : November 1997 INSTRUCTION Industrial Sewing Machines BX1425P,PSM,PTV BX1433P,PSM,PTV BX1025P,PSM BX1033P,PSM First published : November 1997 No. 970112 INTRODUCTION Thank you for your purchasing Kansai Special's BX

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

QUICK PROGRAMMING FOR THERMO SUNIS INDOOR WIREFREE SENSOR

QUICK PROGRAMMING FOR THERMO SUNIS INDOOR WIREFREE SENSOR Adding a Thermo Sunis Indoor Sensor During initial programming, it is recommended that power is provided only to motorized window covering being programmed. STEP 1: Carefully remove rear cover to expose

More information

RAISING THE BAR ON REACTIVE DEPOSITION SPUTTER RATES Douglas R. Pelleymounter

RAISING THE BAR ON REACTIVE DEPOSITION SPUTTER RATES Douglas R. Pelleymounter RAISING THE BAR ON REACTIVE DEPOSITION SPUTTER RATES Douglas R. Pelleymounter ABSTRACT Sputtering non-conductive oxides at a high deposition rate for a long period of time has always been a goal for process

More information

Week IX: INTERFEROMETER EXPERIMENTS

Week IX: INTERFEROMETER EXPERIMENTS Week IX: INTERFEROMETER EXPERIMENTS Notes on Adjusting the Michelson Interference Caution: Do not touch the mirrors or beam splitters they are front surface and difficult to clean without damaging them.

More information

AZ 1512 RESIST PHOTOLITHOGRAPHY

AZ 1512 RESIST PHOTOLITHOGRAPHY AZ 1512 RESIST PHOTOLITHOGRAPHY STANDARD OPERATIONAL PROCEDURE Faculty Supervisor: Prof. R. Bruce Darling Students: Katherine Lugo Danling Wang Department of Electrical Engineering Spring, 2009 TABLE OF

More information

RG REPAIR GUIDELINE 10/23/2015. Repair Description: Tools Required: Step List: Approximate time: minutes

RG REPAIR GUIDELINE 10/23/2015. Repair Description: Tools Required: Step List: Approximate time: minutes Repair Description: Replace BP54000/BP55000 with PA54000 for repairs Tools Required: ) # Philips head screwdriver or Drill ) Small, flat head screwdriver ) Dremel Tool or ¼ round file 4) Caulking Gun 5)

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Buehler EcoMet TM 300 Polisher Version 1.0 Page 1 of 19 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Possible Polishing Materials 2.3

More information

IL550 & IL560 Series Optical Monitors for The ULTIMATE in Thin Film Coating Precision, Accuracy & Control

IL550 & IL560 Series Optical Monitors for The ULTIMATE in Thin Film Coating Precision, Accuracy & Control IL550 & IL560 Series Optical Monitors for The ULTIMATE in Thin Film Coating Precision, Accuracy & Control Slide 0 Why Use Optical Monitoring? Quartz crystal measures the deposited mass Typical accuracy

More information

Z 406 Bench work Manual

Z 406 Bench work Manual 08686 Z 406 Bench work Manual Rev A Custodian: Brian Simeone Table of Contents Section Gantry Removal & Replacements Page 1.. Remove & Replace Printer Module.......3 2.. Remove & Replace Head Card Assembly.....

More information

System Options. Magnetic Property Measurement System. AC Susceptibility. AC Susceptibility Specifications

System Options. Magnetic Property Measurement System. AC Susceptibility. AC Susceptibility Specifications System Options AC Susceptibility Magnetic Property Measurement System Many materials display dissipative mechanisms when exposed to an oscillating magnetic field, and their susceptibility is described

More information

Owner s Manual. PhotoSharp E. Southern Avenue, Phoenix, AZ USA Workhorseproducts.com 1

Owner s Manual. PhotoSharp E. Southern Avenue, Phoenix, AZ USA Workhorseproducts.com 1 Owner s Manual PhotoSharp 3730 E. Southern Avenue, Phoenix, AZ 85040 USA 800-778-8779 Workhorseproducts.com 1 Table of Contents I. Introduction........... 3 II. Specifications................ 4 III. Safety

More information

Automatic Film Applicator BEVS 1811/2 User Manual

Automatic Film Applicator BEVS 1811/2 User Manual Automatic Film Applicator BEVS 1811/2 User Manual This manual shall be read carefully before starting. Directions included in this operation manual shall be strictly followed. PAGE 1 Content 1 Company

More information

MDX DC 1 kw & 1.5 kw Series. Tight regulation Superior arc control Low stored output energy

MDX DC 1 kw & 1.5 kw Series. Tight regulation Superior arc control Low stored output energy MDX DC 1 kw & 1.5 kw Series Tight regulation Superior arc control Low stored output energy Benefits Tight regulation Improved yield Reduced target burn-in time High reliability Tight regulation, superior

More information

High Rise Sit-Stand Desk Converter

High Rise Sit-Stand Desk Converter High Rise Sit-Stand Desk Converter Assembly Instructions for Model DC300 Patent Pending PRE-ASSEMBLY Please read all instructions before beginning assembly. We strongly recommend you watch the video at

More information

Advanced Plasma Technology. High precision film thickness trimming for the TFH industry. Roth & Rau AG September 2009

Advanced Plasma Technology. High precision film thickness trimming for the TFH industry. Roth & Rau AG September 2009 Advanced Plasma Technology High precision film thickness trimming for the TFH industry Roth & Rau AG September 2009 Product Overview IonScan Equipment for ultra-precise Surface Processing IonScan 800 Wafer

More information

AUTOMATION OF THIN FILM DEPOSITION PROCESS BASED ON MAGNETRON SPUTTERING 1. INTRODUCTION

AUTOMATION OF THIN FILM DEPOSITION PROCESS BASED ON MAGNETRON SPUTTERING 1. INTRODUCTION Journal of Machine Engineering, Vol. 12, No. 2, 2012 production automation, magnetron flow production line Piotr DOMANOWSKI 1 Andrzej WAWRZAK 2 AUTOMATION OF THIN FILM DEPOSITION PROCESS BASED ON MAGNETRON

More information

Voltage Control for Reactive Sputtering: Improving Typical Sputter Rate while Dramatically Reducing Input Power Requirements

Voltage Control for Reactive Sputtering: Improving Typical Sputter Rate while Dramatically Reducing Input Power Requirements Voltage Control for Reactive Sputtering: Improving Typical Sputter Rate while Dramatically Reducing Input Power Requirements C. Gruber, J. German, M. Wade, T. Valinski, J. Valek, and B. Bednar, Cardinal

More information

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL Version 5.1 EM Facility CMSE-SEF Massachusetts Institution of Technology TABLE OF CONTENTS 1. Specifications...2 1.1 Performance...2

More information

King Cloud III. Warning. Obstruction Detection. Instructions for Reclining Sofas

King Cloud III. Warning. Obstruction Detection. Instructions for Reclining Sofas Instructions for Reclining Sofas Warning Do not sit on footrest when in open reclined/extended position. MAXIMUM LOAD 15 KG. Always RETRACT FOOTREST prior to standing. Obstruction Detection IF THE RECLINER

More information

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE by Doug Bray Department of Biological Sciences University of Lethbridge Revised September, 2000 Note: The terms in bold in this document represent

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

MOTOR REPLACEMENT : STEP BY STEP.

MOTOR REPLACEMENT : STEP BY STEP. (D) REPLACEMENT : STEP BY STEP. (V) Adjustment screw (Allen key 3/16 ) (E) (U) (I) Tightening screw (Allen key 5/32 ) Blocking screw (H) (Allen key 5/32 ) (G) (Y) (T) (ZZ) (Z) (K) (X) (L) (S) (W) (R) (Q)

More information

UFNF YES Image Reversal & HMDS Oven Revision 6.0 1/22/2014 Page 1 of 5. YES Image Reversal and HMDS Oven SOP

UFNF YES Image Reversal & HMDS Oven Revision 6.0 1/22/2014 Page 1 of 5. YES Image Reversal and HMDS Oven SOP 1/22/2014 Page 1 of 5 YES Image Reversal and HMDS Oven SOP Table of Contents 1.0 Safety 2.0 Quality Control and Calibrations 3.0 Processes Description 4.0 Process Information for Lift Off 5.0 Operation

More information

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T:

CNC Using the FlexiCam CNC and HMI Software. Guldbergsgade 29N, P0 E: T: CNC Using the FlexiCam CNC and HMI Software Guldbergsgade 29N, P0 E: makerlab@kea.dk T: +46 46 03 90 This grey box is the NC controller. Let s start by turning the red switch to the ON position, then press

More information

INSTRUCTION. Industrial Sewing Machines V7100/D,DE,F,ML W8100/D,DE,F,C W8042 W V7002-1S W8103-1S. No Third edition : March 2001

INSTRUCTION. Industrial Sewing Machines V7100/D,DE,F,ML W8100/D,DE,F,C W8042 W V7002-1S W8103-1S. No Third edition : March 2001 INSTRUCTION Industrial Sewing Machines V7100/D,DE,F,ML W8100/D,DE,F,C W8042 W8042-1 V7002-1S W8103-1S Third edition : March 2001 No. 010012 INTRODUCTION Thank you for your purchasing Kansai Special's V.W

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

Optical spray painting practice and training system

Optical spray painting practice and training system University of Northern Iowa UNI ScholarWorks Patents (University of Northern Iowa) 9-14-1999 Optical spray painting practice and training system Richard J. Klein II Follow this and additional works at:

More information

Standard Operating Procedure for Cutting Stainless Steel on the Rofin Laser Machining Station

Standard Operating Procedure for Cutting Stainless Steel on the Rofin Laser Machining Station Standard Operating Procedure for Cutting Stainless Steel on the Rofin Laser Machining Station Version 1.1, August 15, 2007, by Steven Sergenese Reviewed and up-dated 02-02-2010, by D.Turnbull. Introduction

More information

Introduction of New Products

Introduction of New Products Field Emission Electron Microscope JEM-3100F For evaluation of materials in the fields of nanoscience and nanomaterials science, TEM is required to provide resolution and analytical capabilities that can

More information

VS7550 VUV/UV Mini Spectrograph Operating Manual

VS7550 VUV/UV Mini Spectrograph Operating Manual Document RD 15 11 No: VS7550 VUV/UV Mini Spectrograph Operating Manual VS7550 Operating Manual 1 Table of Contents Table of Contents Overview Specifications Vacuum Interface Software and Drivers Packing

More information

OPERATING INSTRUCTIONS FOR MODEL SPR-25 Manual Stencil Printer

OPERATING INSTRUCTIONS FOR MODEL SPR-25 Manual Stencil Printer OPERATING INSTRUCTIONS FOR MODEL SPR-25 Manual Stencil Printer TABLE OF CONTENTS I. INSTALLATION...3 II. SET-UP...4 III. Z AXIS ADJUSTMENT (Screen Height)...6 IV. X, Y AND Ø ADJUSTMENTS...6 V. SQUEEGEE

More information

Fabrication of narrow bandpass filters for wavelength division multiplexing applications A feasibility study

Fabrication of narrow bandpass filters for wavelength division multiplexing applications A feasibility study Indian Journal of Engineering & Materials Sciences Vol. 14, April 2007, pp. 125-132 Fabrication of narrow bandpass filters for wavelength division multiplexing applications A feasibility study A Basu*,

More information

* * APPLICABLE MODELS: 2014 > MAZDA 3

* * APPLICABLE MODELS: 2014 > MAZDA 3 PART NUMBER: 0000 8C L46 GENUINE ACCESSORIES INSTALLATION INSTRUCTIONS Rev. AAA *550-0604-000* APPLICABLE MODELS: 204 > MAZDA 3 REQUIRED COMPONENTS: ITEM QTY DESCRIPTION Usage Chart MIRROR ASSEMBLY: Mirror

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

1 Connector X14270, 47-pin 2 Plug connector to distribution box X4010, 23-pin. 3 Connector X14271, 54-pin 4 Connector X14272, 54-pin

1 Connector X14270, 47-pin 2 Plug connector to distribution box X4010, 23-pin. 3 Connector X14271, 54-pin 4 Connector X14272, 54-pin Junction box electronics: E87, E90, E91, E92 Construction Item Description Item Description 1 Connector X14270, 47-pin 2 Plug connector to distribution box X4010, 23-pin 3 Connector X14271, 54-pin 4 Connector

More information

Installation and Operating Instructions Desk top Laser

Installation and Operating Instructions Desk top Laser www.katanatrading.com Installation and Operating Instructions Desk top Laser This manual covers the following: A. Unpacking and setup of the engraver B. Installation of Newly Seal software C. Basic operation

More information

OPERATION AND MAINTENANCE INSTRUCTIONS AVK UNDERGROUND FIRE HYDRANT

OPERATION AND MAINTENANCE INSTRUCTIONS AVK UNDERGROUND FIRE HYDRANT 1. INTRODUCTION Operation and maintenance manual for the series underground CLEARWAY fire hydrant 1 of 7 The designs, materials and specifications shown are subject to change without notice due to our

More information

Micro Automation Programmable Dicing Saw, Model 1006

Micro Automation Programmable Dicing Saw, Model 1006 Micro Automation Programmable Dicing Saw, Model 1006 Purpose: Operational and procedural instructions used to separate individual die on a silicon wafer or glass substrate. Initial Precautions (1) Edges

More information