8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

Size: px
Start display at page:

Download "8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey"

Transcription

1 Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities 1

2 Radiologist Responsibilities Radiologist Responsibilities Radiologist Responsibilities 2

3 Radiologist Responsibilities Radiologist Responsibilities Radiologist Responsibilities 3

4

5

6

7

8 22 S/N S/N 23 Serial Number 8

9

10

11

12 1042 8/2/ D = Kgcs D = Mean Glandular Dose K = Entrance surface air kerma g = glandularity of 50% c = corrects for difference in composition (age dependent) s = X-ray spectrum correction (Target/Filter) Note: g and c depend on thickness, glandularity, and HVL Primary Ref: DR Dance, et al Additional for the Estimation of Mean Glandular Breast Dose Using the UK Mammography Dosimetry Protocol Physics in Medicine and Biology 45, ,

13 38 39 CR Cassette 40 13

14 SN 8/2/

15

16 Film Printer Setup Print image acquired in "ACR DM Phantom Image Quality" test Print from workstation/computer/pacs used for printing most clinical films Do not adjust window and level settings prior to printing Evaluate the phantom for artifacts & score Note: If possible, print phantom image from same x-ray unit each time for this test Note: Use film size most commonly used for mammography Note: For each printer, only print a single image from a single x-ray unit or workstation Date: Film Size (8 x 10 or 10 x 12): ID of x-ray unit or workstation for printed phantom: Dmax Note: Dmax to be measured on perimeter of film If not available, print a clinical image and measure in the non-breast area SN /2/ Background OD Cavity OD Dmax Radiologic Technologist's Section - DRAFT Page 20 of 25 4 TECH TESTS - Forms-FDA Rev_

17

18

19 56 57 Phantom Scoring: Artifact Evaluation There are 2 important distinctions here: Artifacts that fail vs Artifacts that don t fail The MP & Tech must fail if artifact evaluation meets the above criteria If artifact doesn t fail per above criteria, the MP & Tech should still comment, troubleshoot, and/or recommend for repair Just because the artifact does not Fail doesn t mean it should be ignored 59 19

20 Phantom Scoring: Artifact Evaluation Phantom image failure Will depend on the number of artifacts that appear and where they are located: ---- If only 1 to 3 speck-like artifacts appear in a location that could impact clinical interpretation, the image should not fail However, the artifacts should be evaluated and eliminated if possible ---- If speck-like artifacts are more widespread and appear in a location that could impact clinical interpretation, these are probably clinically significant and the image should fail These artifacts must be eliminated in order for the phantom image to pass Artifact Example Series Serial Number 63 Artifacts Fail Artifact Example Series Serial Number 64 Artifacts do not Fail but troubleshooting and comments are important 20

21 Artifact Example Series Serial Number 65 Artifacts may not Fail but troubleshooting and comments are important (2 outer specks may not be in location that could impact clinical interpretation*) Description: Massive amounts of dust & debris, texture or grid can be seen Pass or Fail for Artifacts?: --Detector blotchiness & visible texture: Fail --White speck-like: Fail Phantom image failure Will depend on the number of artifacts that appear and where they are located: ---- If only 1 to 3 speck-like artifacts appear in a location that could impact clinical interpretation, the image should not fail However, the artifacts should be evaluated and eliminated if possible ---- If speck-like artifacts are more widespread and appear in a location that could impact clinical interpretation, these are probably clinically significant and the image should fail These artifacts must be eliminated in order for the phantom image to pass Description: Blotchy background and linear artifacts Pass or Fail for Artifacts?: --Detector blotchiness & horizontal lines: Fail --White speck-like: Fail Phantom image failure Will depend on the number of artifacts that appear and where they are located: ---- If only 1 to 3 speck-like artifacts appear in a location that could impact clinical interpretation, the image should not fail However, the artifacts should be evaluated and eliminated if possible ---- If speck-like artifacts are more widespread and appear in a location that could impact clinical interpretation, these are probably clinically significant and the image should fail These artifacts must be eliminated in order for the phantom image to pass 21

22 Description: Non-uniformities, white speck like artifacts Pass or Fail for Artifacts?: --White speck-like: Fail Phantom image failure Will depend on the number of artifacts that appear and where they are located: ---- If only 1 to 3 speck-like artifacts appear in a location that could impact clinical interpretation, the image should not fail However, the artifacts should be evaluated and eliminated if possible ---- If speck-like artifacts are more widespread and appear in a location that could impact clinical interpretation, these are probably clinically significant and the image should fail These artifacts must be eliminated in order for the phantom image to pass Description: White dots Pass or Fail for Artifacts?: --White speck-like: Fail Phantom image failure Will depend on the number of artifacts that appear and where they are located: ---- If only 1 to 3 speck-like artifacts appear in a location that could impact clinical interpretation, the image should not fail However, the artifacts should be evaluated and eliminated if possible ---- If speck-like artifacts are more widespread and appear in a location that could impact clinical interpretation, these are probably clinically significant and the image should fail These artifacts must be eliminated in order for the phantom image to pass Description: Non-uniformities, white speck like artifacts Pass or Fail for Artifacts?: --Linear artifacts: Fail --White speck-like: Fail Phantom image failure Will depend on the number of artifacts that appear and where they are located: ---- If only 1 to 3 speck-like artifacts appear in a location that could impact clinical interpretation, the image should not fail However, the artifacts should be evaluated and eliminated if possible ---- If speck-like artifacts are more widespread and appear in a location that could impact clinical interpretation, these are probably clinically significant and the image should fail These artifacts must be eliminated in order for the phantom image to pass 72 22

23 Description: White speck like artifacts Pass or Fail for Artifacts?: Fail Phantom image failure Will depend on the number of artifacts that appear and where they are located: ---- If only 1 to 3 speck-like artifacts appear in a location that could impact clinical interpretation, the image should not fail However, the artifacts should be evaluated and eliminated if possible ---- If speck-like artifacts are more widespread and appear in a location that could impact clinical interpretation, these are probably clinically significant and the image should fail These artifacts must be eliminated in order for the phantom image to pass Thank you, And now for questions 23

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

Facility, Unit and Test Equipment Data

Facility, Unit and Test Equipment Data Facility, Unit and Test Equipment Data Medical hysicist's Tests - SAMLE FORMS Facility Information Facility Name Happy Valley Mammography Address Suite 1 Address 1 Oak Street City, State, Zip Anywhere

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and

More information

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May 2010 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics, Inc. Physicist's Role in the

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys Outline SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys Robert J. Pizzutiello, MS, FACR, FAAPM, FAC Residency Program Director, Upstate Medical Physics, PC Senior Vice President,

More information

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle

More information

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING Maynard High New York Medical College SS2001-M.High 1 Objectives: Review MQSA and ACR annual QC tests as opportunities for troubleshooting before a significant

More information

Half value layer and AEC receptor dose compliance survey in Estonia

Half value layer and AEC receptor dose compliance survey in Estonia Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:

More information

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests Slide 1 Digital Mammography QA: Comparing the Manufacturers Recommendations Eric A. Berns, Ph.D. Slide 2 Introduction What is QC and why is it important? Review & compare QC tests Key take home points

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Aspire HD. Program Manual. 2nd Edition - October 2012

Aspire HD. Program Manual. 2nd Edition - October 2012 Quality Control 1 Aspire HD Quality Control Program Manual 2nd Edition - October 2012 Overview Installation of FDR Mammography QC Program Weekly Test 2 3 4 Quarterly Test 5 Semi-annual Test 6 Annual Test

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL Implementation date: 31 March 2009 Contents

More information

Ansur TNT Users Manual. Plug-In

Ansur TNT Users Manual. Plug-In Ansur TNT 12000 Plug-In Users Manual August 2009, Rev. 2, 12/09 2009 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

FFDM -FCRm QC Requirements- What You REALLY Need to Know

FFDM -FCRm QC Requirements- What You REALLY Need to Know FFDM -FCRm QC Requirements- What You REALLY Need to Know Melissa C. Martin, M.S., FACR, FAAPM, FACMP AAPM Annual Meeting July 28, 2008 FDA Approval for Mammography Fuji FCRm System approved for Mammography

More information

Imaging Technique Optimization of Tungsten Anode FFDM System

Imaging Technique Optimization of Tungsten Anode FFDM System Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,

More information

Patient-Assisted Compression Impact on Image Quality and Workflow

Patient-Assisted Compression Impact on Image Quality and Workflow Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled

More information

4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences

4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences Quality Control Activities for the RadiologicTechnologist Quality Control Tests 2D QC Tomosynthesis QC DICOM Printer Quality Control Weekly Detector Flat Field Calibration Weekl Artifact Evaluation Weekly

More information

of sufficient quality and quantity

of sufficient quality and quantity of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam

More information

Image Display and Perception

Image Display and Perception Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration

More information

ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0

ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0 Heggie et al ACPSEM Position Paper: Digital Mammography V4.0 ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0 JCP Heggie 1, P Barnes 2, L Cartwright 3, J Diffey

More information

Digital Mammography Quality Control for the Mammographic Technologist

Digital Mammography Quality Control for the Mammographic Technologist Ontario Breast Screening Program Digital Mammography Quality Control for the Mammographic Technologist Authors: G.E. Mawdsley, A.K. Bloomquist, M.J. Yaffe October 2011 Revision 3.1 Mammographic Physics

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis Rosangela Requi Jakubiak* a, Humberto Remigio Gamba a, Maria Manuela Ramos a, Gislene Gabrielle Faversani

More information

1.1.Clinical Artefacts

1.1.Clinical Artefacts 1.1.Clinical Artefacts While the incidence of artefact on digital mammographic images 1 is typically less than with film based mammography, artefacts can be produced on digital systems. This section provides

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies

More information

Digital Image Management: the Basics

Digital Image Management: the Basics Digital Image Management: the Basics Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University Outline From screen/film to digital radiography PACS/Tele

More information

Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial

Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial Martin J. Yaffe, Aili K. Bloomquist, Gordon E. Mawdsley, Etta D. Pisano, R. Edward Hendrick, Laurie L. Fajardo,

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations Shirley Weddle, RT(R)(M), CIIP, BBA OBJECTIVES Define Computed Radiography (CR) Discuss CR vendor

More information

Tomosynthesis and Motion

Tomosynthesis and Motion Tomosynthesis (3D) Motion Unsharpness Occurs at about the same frequency as conventional mammography (2D) Presents the same issues as 2D motion, EXCEPT that motion may go undetected Most common patient-related

More information

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a

More information

Published text: Institute of Cancer Research Repository Please direct all s to:

Published text: Institute of Cancer Research Repository   Please direct all  s to: This is an author produced version of an article that appears in: MEDICAL PHYSICS The internet address for this paper is: https://publications.icr.ac.uk/1316/ Copyright information: http://www.aip.org/pubservs/web_posting_guidelines.html

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

Open. the Digitized world. Fuji Computed Radiography

Open. the Digitized world. Fuji Computed Radiography Open the Digitized world Fuji Computed Radiography If just one of these applies to you... Managing developing fluid is hard and darkroom work is a hassle... Images are not stable... Isn t digitalization

More information

Breast Imaging Basics: Module 10 Digital Mammography

Breast Imaging Basics: Module 10 Digital Mammography Module 10 Transcript For educational and institutional use. This test bank is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook DIAGNOSTIC ACCREDITATION PROGRAM Radiology and CT Quality Control Procedures Workbook Quality Control Procedures Radiography/CR/DR Safety Code 35 Summary For more detail about each quality control (QC)

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals DOI 10.1007/s00247-011-2290-1 ORIGINAL ARTICLE Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals Mervyn D. Cohen & Richard Markowitz

More information

Digital Breast Tomosynthesis

Digital Breast Tomosynthesis Digital Breast Tomosynthesis OLIVE PEART MS, RT(R) (M) HTTP://WWW.OPEART.COM 2D Mammography Not 100% effective Limited by tissue superimposition Overlapping tissue can mask tumors False negative Overlapping

More information

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

NJDEP Medical Physicist s Radiographic QC Survey Registration Number: Facility Name NJDEP ID # NJDEP Medical Physicist s Radiographic QC Survey PLEASE PRINT Facility Information Unit Information Manufacturer Model Console Model # Console serial # Tube serial # Location (room)

More information

GE Healthcare. Senographe 2000D Full-field digital mammography system

GE Healthcare. Senographe 2000D Full-field digital mammography system GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 7 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 7 Quality 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging:

More information

X-RAY MEDICAL EQUIPMENT

X-RAY MEDICAL EQUIPMENT X-RAY MEDICAL EQUIPMENT CHEST RADIOGRAPHY GENERAL RADIOGRAPHY & FLUOROSCOPY RADIOTHERAPY MOBILE HEALTHCARE MAMMOGRAPHY MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Biopsy Attachment џ MAMMOSCAN an ADANI

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Dose Management in Digital Radiography Raija Seuri (HUS Medical Imaging Center, FI) Cristina Almeida (Centro Hospitalar de Lisboa Central,

More information

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S. Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,

More information

Learning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition

Learning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition Converting the Radiology Department from Film-Screen to Digital: Making the Transition S. Jeff Shepard, MS, DABR University of Texas M. D. Anderson Cancer Center Houston, Texas jshepard@di.mdacc.tmc.edu

More information

White Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results...

White Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results... White Paper The Advantages of Hybrid Gamma PXL CONTENTS 1 Introduction... 2 2 Outline of Hybrid Gamma PXL... 4 3 Study... 5 4 Results... 7 5 Conclusions... 10 No.17-001 Revision A November 2017 Product

More information

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. Tingberg, Anders Published in: Radiation Protection Dosimetry DOI: 10.1093/rpd/ncs302 Published: 2013-01-01 Link to publication Citation for published

More information

ACR Update in Nuclear Medicine Accreditation

ACR Update in Nuclear Medicine Accreditation Disclaimer ACR Update in Nuclear Medicine Accreditation Beth A. Harkness, MS, DABR, FACR Henry Ford Health System Detroit, MI ACR physics subcommittee for nuclear medicine accreditation. My facility is

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

2012 :15th SESSION of ESMP

2012 :15th SESSION of ESMP 2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities

More information

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN

More information

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology

More information

DR _ solutions. We understand that customers don t need just products, they want. solutions

DR _ solutions. We understand that customers don t need just products, they want. solutions DR _ solutions We understand that customers don t need just products, they want solutions index company profile 1974-2005 2006-2007 - 2008 ITALRAY Srl was founded in 1974 as the production branch of Marzocchi

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

MILADY. Product Data. Page 1 of 8

MILADY. Product Data. Page 1 of 8 Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,

More information

KODAK DIRECTVIEW CR Mammography Feature User s Guide

KODAK DIRECTVIEW CR Mammography Feature User s Guide KODAK DIRECTVIEW CR Mammography Feature User s Guide 17 September 2010 9G3741 Version 1.0 Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 CARESTREAM, DIRECTVIEW, and DRYVIEW are trademarks

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

NEMA. Cardiology Phantom Nuclear Associates Model Diagnostic Imaging. Introduction. Features. Applications

NEMA. Cardiology Phantom Nuclear Associates Model Diagnostic Imaging. Introduction. Features. Applications NEMA Cardiology Phantom Nuclear Associates Model 07-680 Diagnostic Imaging! Phantom and test procedures simulate a range of fluoroscopically-guided invasive and interventional procedures! Provides simultaneous

More information

2 nd generation TOMOSYNTHESIS

2 nd generation TOMOSYNTHESIS 2 nd generation TOMOSYNTHESIS 2 nd generation DBT true innovation in breast imaging synthesis graphy Combo mode Stereotactic Biopsy Works in progress: Advanced Technology, simplicity and ergonomics Raffaello

More information

DIGITAL RADIOGRAPHY ARTIFACTS

DIGITAL RADIOGRAPHY ARTIFACTS IMAGING LAB MPHY 487 DIGITAL RADIOGRAPHY ARTIFACTS Mohammad Esmael Alsulimane B.Sc, M.Sc Medical Physics Lecturer - Physics Department All Rights Reserved: Some information and figures in this presentation

More information

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) SUMMARY OF 2003 FLUOROSCOPY SURVEY

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) SUMMARY OF 2003 FLUOROSCOPY SURVEY CRCPD Publication E-09-5 Available Online at No Charge NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) SUMMARY OF 2003 FLUOROSCOPY SURVEY September 2009 Published by Conference of Radiation Control Program

More information

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Draft version 0.10 February 2013 European Reference Organisation for Quality Assured Breast

More information

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology EVALUATING X-RAY TUBE AND GENERATOR PERFORMANCE : DEMO for PRACTICAL QUALITY CONTROL (QC) Dr Slavik Tabakov Dept. Medical Eng. & Physics, King's College London slavik.tabakov@kcl.ac.uk QC in Diagnostic

More information

RaySafe X2. Effortless measurements of X-ray

RaySafe X2. Effortless measurements of X-ray RaySafe X2 Effortless measurements of X-ray At your fingertips We ve grown accustomed to intuitive interactions with our devices. After all, it s not the device that s most important, but what you can

More information

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Draft version 0.15 January 2014 European Reference Organisation for Quality Assured Breast

More information

Collimation Assessment Using GAFCHROMIC XR-M2

Collimation Assessment Using GAFCHROMIC XR-M2 Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing

More information

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria International Journal of Innovative Scientific & Engineering Technologies Research 5(4):-5, Oct.-Dec., 07 SEAHI PUBLICATIONS, 07 www.seahipaj.org ISSN: 60-896X Assessment of Beam Alignment, Collimation

More information

History of digital imaging

History of digital imaging CR/QA RADCHEX History of digital imaging Early, crude digital detectors were developed in the 1970 s Image quality was problematic Processing time of digital images was untenable Viewing, transfer and

More information

Things you may want to know

Things you may want to know Things you may want to know 1 / 9 Last updated 18-Nov-05 Why medical monitors are special? Unlike commercial displays, Medical displays are used in the hospital and they are displaying contain life critical

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

Tomophan TSP004 Manual

Tomophan TSP004 Manual T h e P h a n t o m L a b o r a t o r y 1 Tomophan TSP004 Manual Copyright 2016 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order and

More information

Image Quality Artifacts in Digital Imaging

Image Quality Artifacts in Digital Imaging MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,

More information

DRAFT Technical evaluation of Philips Microdose SI digital mammography system

DRAFT Technical evaluation of Philips Microdose SI digital mammography system DRAFT Technical evaluation of Philips Microdose SI digital mammography system NHSBSP Equipment Report 1310 August 2013 About the NHS Cancer Screening Programmes The national office of the NHS Cancer Screening

More information

Threshold contrast visibility of micro calcifications in digital mammography

Threshold contrast visibility of micro calcifications in digital mammography Threshold contrast visibility of micro calcifications in digital mammography Ann-Katherine Carton*, Hilde Bosmans*, Dirk Vandenbroucke, Chantal Van Ongeval*, Geert Souverijns*, Frank Rogge*, Guy Marchal*

More information

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Gham Hur, M.D., Yoon Joon Hwang, M.D., Soon Joo Cha, M.D., Su Young Kim, M.D., Yong Hoon Kim, M.D.

More information

Measurement of table feed speed in modern CT

Measurement of table feed speed in modern CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department

More information

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Estimation of signal transfer property for wireless digital detector in different measurement schemes Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information