The Speaker Study. By: Jay Bliefnick. Acoustical Testing 1. Attn: Dr. Dominique Chéenne, Dr. Lauren Ronsse. Group Members:

Size: px
Start display at page:

Download "The Speaker Study. By: Jay Bliefnick. Acoustical Testing 1. Attn: Dr. Dominique Chéenne, Dr. Lauren Ronsse. Group Members:"

Transcription

1 The Speaker Study By: Jay Bliefnick Acoustical Testing 1 Attn: Dr. Dominique Chéenne, Dr. Lauren Ronsse Group Members: Hannah Knorr, Michael Hanson, Matt Johnson, Miles Possing, & Ming Yu 11/27/13

2 Table of Contents Abstract... 3 Introduction. 3 Anechoic Testing Setup....4 Non-Anechoic Testing Setup... 9 Frequency Response Analysis Crossover Frequency Analysis..12 Time Response Analysis. 14 Polar Response Analysis. 15 Non-Anechoic Response Comparison...16 Earthworks Microphone Comparison.. 19 Conclusion

3 Abstract In the speaker study, students were required to thoroughly analyze a speaker in an anechoic environment. An Event 8XL reference monitor was selected as the test speaker, and all tests were conducted using the Goldline TEF-20 system and an Electrovoice RE-55 microphone. The frequency response of the speaker was measured in a variety of orientations and frequency resolutions to determine an accurate reading across the entire spectrum. The crossover point between the two components was identified, and the time response was tested at the highest time resolution TEF allowed. The full-frequency polar response was also found using an automated turntable. The speaker s frequency and time responses were also taken outside of an anechoic environment to compare the two sound fields. Finally, the anechoic frequency and time response tests were completed using a second microphone: an Earthworks M-50 condenser. Through these tests, the performance of this speaker was verified across a wide variety of conditions. Introduction The goal of the speaker study was to determine as much about a selected test speaker as possible, whether this be parameters or possible test conditions. The first round of data was collected in the anechoic chamber (Figure 1) at Columbia College, which provided a reflectionfree environment in which to test. This allowed for direct analysis of the test speaker with as little room interaction as possible. Figure 1: Anechoic Chamber at Columbia College. The speaker was placed in the approximate center of the chamber with the microphone placed on-axis, 1 m away. This was the first speaker-microphone orientation utilized. 3

4 The anechoic chamber was actually recently remodeled to repair flood damage caused several years ago. Over the summer, the temporary ceiling absorption was removed, and in its place three layers of Roxul sound-rated insulation were installed. Two layers of RHT-80 (the densest panels) and one layer of RHT-60 were fastened to the ceiling, with all side areas being filled as well. The final layer of RHT-40 (least dense) was suspended ~12 lower on a wire support system. The finished ceiling not only looked better, but also provided a much more even frequency response in the room, without the extreme notch at 1 khz seen before the new construction (caused by the previous wedge-style absorption panels). The one issue that still remained with the chamber was the relatively small dimensions of x 10 2 x 8 1 (Volume = 1137 ft 3 ), which created some modal influence. With the anechoic chamber providing an accurate method of isolating the speaker s properties, it was important to research the comparative responses in a non-anechoic environment. The open area in the LL01 classroom provided the necessary conditions for these tests. The frequency response was taken and compared to show how the influence of the room (reverberation) affected the results. The time response was also tested, utilizing multiple reflection panels. These created peaks which could be measured and quantified into distances, then compared with the anechoic results. Finally, the anechoic tests were run once again using a second microphone to compare the two and confirm the response of the speaker. Anechoic Testing Setup The first round of testing was conducted in the aforementioned anechoic chamber. The Event 8XL studio reference monitor was selected for study due to its consistent response, which could be expected to produce accurate results. The Electrovoice RE-55 was selected for use as the primary microphone. This dynamic, omni-directional microphone allowed for precise measurements to be taken without the need for phantom power. This combination of speaker 4

5 and microphone would be used for all tests, except for two additional tests taken at the end of the study with an Earthworks M-50 condenser microphone. In the anechoic chamber, four main features of the speaker were investigated: the frequency response, the crossover frequency, the time response, and the polar directivity pattern. These were completed under a variety of setup orientations and testing parameters to thoroughly analyze all potential issues. These tests were completed using the Goldline TEF-20 analysis system installed on workstation computer 2, adjacent the anechoic chamber. This was accomplished by using the multiple patch bays connecting the facility (Figure 2). Computer Running Station 2 Central Anechoic Event 8XL TEF 20 Software Patch Bay Patch Bay Snake EV RE-55 Figure 2: Signal chain used to connect the speaker and microphone in the anechoic chamber to the computer using the patching system. The process was simplified by using the same number for each step: input 1 for all microphone connections and output 3 for speaker connections. The first speaker attribute tested in the anechoic chamber was the frequency response. This was completed using three setup positions, two independent methods, and a variety of parameter settings. The microphone-speaker relationship remained constant for the tests at 1 m, on-axis with the speaker s center. This orientation was chosen due to it being the industry-wide standard. For the initial placement, the speaker was set in the center of the chamber facing the microphone parallel to the sidewalls. The second setup rotated the system clockwise approximately 30 degrees, whereas the final position placed the system about 30 degrees counterclockwise from the original arrangement (Figure 3). These tests provided a look into the influence the room had on the measurements and how much modal interaction was present. In the initial placement, the frequency response was gathered two separate ways: as a single test across the entire spectrum, and by breaking the range into multiple tests, compiling the graph piecemeal. To begin, a single sweep was run from 20 Hz 20 khz using a sweep time 5

6 of 180 seconds. This provided a frequency resolution of 18.6, which was good enough for an initial set of data, but was not accurate enough in the low end. To generate a frequency response with an adequate resolution for all ranges, the test was completed once more, only this time in octave intervals. By using this method, a high resolution could be maintained while limiting the amount of time required to run the tests. Overall, seven tests with approximately 2-minute sweep times were conducted in octave intervals (with the low frequency test accounting for several octaves). This method produced similar data to that gathered using the single sweep, but with a higher degree of accuracy. Finally, a long 5000-second single-sweep was attempted, which pushed the TEF system to its maximum frequency resolution of 2.0 Hz. Figure 3: Three speaker orientations used in the anechoic chamber. The multiple positions allowed the investigation of modal properties within the space. Using the second and third speaker-microphone positions, the long and short singlesweep tests were completed. Given the uniformity between the piecemeal response and the long sweep test conducted at position one, it was decided that the single-sweep tests would suffice. The responses were largely the same, especially above 500 Hz, but as expected, the chamber s modal properties and absorption created variances in the mid and low frequencies. The second property of the speaker investigated was the crossover frequency between the woofer and the tweeter. As the Event 8XL was a bi-amped powered speaker with a mono input, each component had its own frequency response. These would blend into a unified field after a 6

7 sufficient distance, but at close ranges, the two pieces could be isolated. The crossover was tested in two ways, generating somewhat contradictory results. The first method for testing the crossover frequency involved blocking the sound generated from one component and then testing the generating source. Each test was conducted 1 from each source, on-axis. To test the woofer, the tweeter was blocked with a layer of Auralex Acoustics SonoFiber absorber panels (Figure 4). To test the tweeter, the same process was repeated for covering the woofer using the same absorbers (Figure 5). The first test using this method demonstrated the under whelming performance of the absorbers under this application, so an additional layer was added underneath to improve the results. Figure 4: To test the woofer, the tweeter was blocked using an Auralex Acoustics Sonofiber absorber panel. Figure 5: The same process was used to test the tweeter. Additional layers of absorption were added beneath the visible Auralex panel. The techniques of testing the two components worked as intended, and accurate fullfrequency responses were taken of both. It was found that the crossover range was 1 khz 5 khz (generous approximation), so those frequencies were retested at a higher resolution (5 Hz) for the woofer and tweeter. Their graphs were overlaid to see where the inflection point was. The one issue with this process was the fact that the two components had different sensitivities, creating level variances at the 1 testing distance. Due to these factors, the crossover frequency could only be narrowed to between 2 khz and 3 khz. 7

8 The second method of testing the crossover frequency did not involve any form of absorption or any isolating materials. Instead, the microphone was place 1 from the face of the speaker, on-axis, looking directly between the woofer and the tweeter. The same, limited frequency range test was conducted and demonstrated interesting results. The crossover frequency appeared to be lower, at approximately 1.8 khz utilizing this process. With the frequency response and crossover frequency tests complete, the next parameter to study was the time response of the speaker. Speaker orientation one was utilized for time response tests. The most important factor in obtaining accurate time response data was to maximize the time resolution within the TEF system. To achieve this level of precision in the time domain, resolution in the frequency domain needed to be sacrificed. Therefore, the time response tests were conducted encompassing the full 20 Hz 20 khz range. The sweep time was also limited to.77 seconds to improve results. The tests were run with two sweep time/sample amount parameter settings. The first test had a sweep time of 10.9 s using 512 samples, and the second had a sweep time of 175 ms with 8192 samples. These produced identical results, which was expected considering: Time Resolution = Sweep Time Total Samples For both examples, the time resolution of the system was 21.3 μs, the maximum allowable resolution in the TEF system. The effect of the speaker and microphone stand were also investigated utilizing absorption coverings. They were seen to have negligible influence. The final speaker parameter tested within the anechoic chamber was the polar directivity response. This was accomplished using the automated turntable (seen clearly in Figure 4), which was operated through TEF. The test consisted of multiple frequency response tests (as performed before), conducted at a selected rotation degree step, to obtain a full 360-degree speaker response. The time length and sample size of each sweep could be chosen, which would 8

9 determine the frequency resolution for each test, and thus the accuracy of the end result. The rotation step size could also be selected: from 22.5 degrees, down to a tiny 2.5 degrees. Both of these parameter selections could improve the resolution of the eventual polar response, but the sacrifice was the time required to complete the full test. As a compromise to achieve valid data with a relatively short test cycle, a sweep length of 300 seconds per test with a 15-degree step size was chosen. This resulted in a total test time of approximately 2.5 hours. The process went through without a hitch, but considering concurrent tests were being conducted in the adjacent Real World Transmission Loss Chamber, accuracy needed to be verified. An identical test was run after hours, while the building was empty, and showed nearly indistinguishable results. Once reliable data was gathered from the two polar tests, a third was conducted utilizing a much higher frequency resolution and smaller step size. The sweep time for each test was raised to 600 seconds, while the rotation step size was decreased to 10 degrees. These changes inflated the overall test time to more than 6.5 hours. The analysis was again run while no one was utilizing LL01, ensuring isolation. Eventually the process was complete and an even more accurate polar response was obtained. Non-Anechoic Testing Setup In addition to testing the speaker within the anechoic chamber, the response from a nonanechoic environment was needed to provide comparison. The open area in the LL01 classroom served as the testing venue for this analysis. This was made possible due to the incredibly high signal to noise ratio within TEF, allowing accurate measurements even in noisy environments. The speaker was set upon the same stand and the Electrovoice RE-55 was set up as before. First, the frequency response of the speaker was tested using the single-sweep method. Utilizing a sweep length of 300 seconds, a frequency resolution of 11 Hz was achieved: accurate 9

10 enough for this test. The response graph was similar in general shape, but with the addition of the room interaction, it was obvious the speaker was no longer isolated. The time response was the other parameter tested in the non-anechoic environment. The same parameters were used as before to ensure the highest time resolution possible. The first test was conducted in the exact setup as the frequency test. The time response resembled the anechoic results, save for a second peak seen after the initial decay. It was suspected that this peak was caused by a reflection off of the floor (the only close surface). To test this theory, absorption was applied to the floor between the speaker and microphone (Figure 7). Using the same parameters, the second peak disappeared, confirming the reflection from the floor. To push these results further, another reflecting surface was added into the equation, this time in the form of a moveable solid panel (Figure 6). This time three tests were completed: no absorption, absorption on the floor, and absorption on the reflector panel. With no absorption, two additional peaks were seen after the direct sound. When the absorption was applied to both the floor and the panel, their respective peaks on the time response graph disappeared. This confirmed the hypothesis that the time response peaks were due to close reflecting surfaces. The distances of these components were then measured and compared against the computed values. Figure 6: Non-anechoic speaker tests with reflector panel. Frequency response and initial time response conducted without the panel. Figure 7: To confirm the reflection source, absorption was added to the floor (and eventually the reflector panel), which eliminated the peaks. 10

11 Frequency Response Analysis The frequency response was computed in a variety of ways, and in three separate orientations. In general, all graphs were quite similar, especially above 500 Hz, but some variances could be seen in the lower registers. The first position was tested in two different manners: as a single sweep and piecemeal using multiple individual sweeps (Figures 8 & 9). Figure 8: Frequency response of the speaker in position 1 using the piecemeal method. The different colors represent the individual tests conducted in octave bands. The frequency resolution was much higher than had a single sweep been run, given the time spent on the tests. Figure 9: Frequency response of the speaker in position 1 using the single sweep method. The maximum TEF sweep time was used to acquire the highest frequency resolution. The two responses appear identical. 11

12 Once accurate data was confirmed for speaker position one, the setup was rotated and retested in positions two and three. By placing the system in different orientations within the anechoic chamber, the influence from the room itself could be ascertained using the same second single sweep method. Figures 10 and 11 demonstrate the variance between the initial positioning. Test setup two showed general similarities to the first graph, but the peaks/troughs from 300 Hz to 500 Hz were not congruent. Test setup three, however, displayed extreme differences between the placements within the chamber. In this position, the microphone was the closest to the sidewall, which could have contributed to the altered response. Figure 10: Anechoic frequency test position 2. Figure 11: Anechoic frequency test position 3. Crossover Frequency Analysis The crossover frequency of the speaker was determined between the tweeter and the woofer in two separate manners: by physically isolating each component and by testing directly between the two. These methods produced two different results for the possible crossover frequency, which needed explanation. By isolating the components of the speaker, the individual frequency response graphs from the tweeter and woofer could be found (Figures 12 & 13). These displayed a region of 1 khz 5 khz where the woofer began to decrease in power and the tweeter began to increase. 12

13 This region was further investigated by testing this specific frequency range at a higher resolution, and then plotted on the same graph. This resulted in a potential crossover range of 2 khz 3 khz. The exact frequency could not be narrowed further due to the relative levels of the two components: it was impossible to plot the two responses equally (Figure 14). Woofer Response Tweeter Response Figure 12: Isolated woofer response. Figure 13: Isolated tweeter response. Figure 14: Combined tweeter Lower graph was with and woofer response. additional layers of absorption. The second method for testing the crossover frequency placed the microphone directly between the tweeter and the woofer, 1 from the speaker plane. By selecting this condition, the relative levels of the components could be seen simultaneously (Figure 15). They showed an obvious dip in the frequency response of ~20 db at 1.8 khz. This was well below the values the previous tests generated. The phase relationship was also plotted over the range, and it displayed a shift at the exact same frequency. It would seem that this solved the crossover frequency question. However, by conducting research online, the manual for the Event 8XL was discovered, revealing the published crossover value of 2.6 khz. This value confirms the first crossover tests, but did not answer the issues raised by the second test. One possible explanation could be that 1.8 khz was where the two components cancelled each other out due to phase 13

14 interference. On either side of this frequency, either the tweeter or woofer would dominate, limiting the phase problems, but at that frequency, a depression was formed. Considering the fact that the tweeter had a higher sensitivity, by this reasoning it would make sense that the crossover frequency was raised above 1.8 khz. Frequency Response Phase Response Figure 15: Crossover frequency test with microphone placed 1 from speaker plane between the tweeter and the woofer. The white line was the frequency response; the red line was the phase response. The graph shows the extreme dip at 1.8 khz as well as the phase shift at the same frequency. Time Response Analysis The time response of the speaker was tested in the anechoic chamber to determine the maximum time resolution within the TEF-20 system, and to provide an adequate baseline to compare the non-anechoic tests against. A time resolution of 21.3 μs was achieved by using a full-frequency range, a sweep time of 175 ms and 8192 total samples (Figure 16). More tests were conducted, but the time response graphs were identical. The tests showed an even decay pattern, after the reception of the direct sound at 3.1 ms. This equated to 1.06 meters, based on a 343 m/s speed of sound. Beyond the initial decay, the time response remained consistently low, as expected. This demonstrated that there were no sources of reflections within the anechoic chamber that influenced time response results. 14

15 Figure 16: Anechoic chamber time response. The smooth decay curve was due to the reflection-free environment. Polar Response Analysis The polar response of the speaker was collected over the course of three separate tests. The first two were conducted using identical testing parameters: 300 s sweep times with a rotation step size of 15 degrees. For the third test, the resolution was increased for both the sweeps and the steps: 600 s sweep times and 10-degree step sizes were used. Given the higher accuracy, the third test was primarily used to evaluate the polar response (Figure 17). Below 250 Hz, the speaker was very omni-directional: showing less than a 10 db difference between 0 and 180 degree orientations. At 56 Hz, the level of the speaker response began diminishing, as the frequency graph predicted. Between 500 Hz and 1 khz, the rearward projection remained strong, but lobes began to form, indicating discrepancies in the response. Above 2 khz, the speaker became much more directional, as the difference between 0 and 90 degrees was greater than 10 db. The rear response was also substantially decreased with even more pattern deviations. As the frequency increased, the directivity of the speaker became even more pronounced, with the response at 16 khz being 20 db greater on-axis than off. This narrowing of the polar response was attributed to the tweeter becoming the dominant source component, as they are generally more directional than woofers. 15

16 Figure 17: Polar response of Event 8XL in an anechoic environment. Polar test 3 was plotted (highest resolutions) in octave bands across the entire spectrum. Non-Anechoic Response Comparison With the speaker thoroughly studied under anechoic conditions, the next step was to test the frequency and time responses under free-field conditions. Because the crossover frequency was determined using close microphone techniques, a re-test was unnecessary, and the polar response was wired to only be completed in the anechoic chamber. The frequency response was the first test to be run, setup with no modifications to the test environment. There was no absorption placed near the test, and the additional reflector panel was set to the side. Using a sweep time of 300 seconds, a frequency resolution of 11 Hz was achieved. The graph displayed the same basic response pattern as before, but with significantly more deviations (Figure 18). 16

17 This could be attributed to the influence of the room and the pattern of reflections that occurred under non-anechoic conditions. The reverberation characteristics of the space were mixed with the speaker s response resulting in a very disruptive graph, as opposed to the much smoother response gathered under more suitable test conditions (Figure 9). Figure 18: Frequency response taken in LL01 under non-anechoic conditions. The much more aggressive pattern could be attributed to the reverberation characteristics of the space. The time response under these test parameters was also gathered to determine if the source of these reflections. The same sweep time, time span, and sample size values were utilized as before, ensuring the highest time resolution possible. The first test was conducted as the frequency response was: with no absorption or reflector panel. The second test included absorption placed on the floor directly between the speaker and the microphone. Both graphs displayed the direct sound at 3.1 ms and initial decay, but only the test without the absorption displayed the second peak at 6.91 ms (Figure 19). By extrapolating these two numbers out, the distance for the direct sound was 1.06 m and the first reflection was 2.37 m. Using a tape measure, the distance from the front of the speaker to the front of the microphone was 1.01 m, and the path reflecting off of the floor was approximately 2.34 m. These values confirmed the source of the first reflection as the floor between the speaker and microphone. 17

18 No Absorption With Absorption Figure 19: Time response taken in LL01 under non-anechoic conditions. The blue line was with no absorption present, whereas the red line was with absorption between the speaker and microphone. To further validate the source of the reflections as seen on the time response graph, a reflector panel was set up adjacent to the speaker and microphone. It was placed parallel to the system, slightly more than a meter away from the direct sound path, which was expected to create a second peak slightly after the first reflection. Three tests were completed under these conditions: no absorption, absorption on the floor, and absorption on the panel (Figure 20). As before, with the absorption in place, the expected reflection peaks were eliminated. No Absorption Absorption on Floor Absorption on Panel Figure 20: Time response taken in LL01 under non-anechoic conditions. The black line was with no absorption present, the blue line was with absorption on the floor, and the orange line had absorption on the panel. 18

19 Earthworks Microphone Comparison To compare the response of more than just the speaker, a second microphone was employed to see if differences could be found. An Earthworks M-50 condenser microphone was chosen due to its exceptionally flat frequency response. Because it was a condenser, though, it needed phantom power to operate, so the Aphex preamplifier, located in the equipment rack, was hooked up. The one issue with the Earthworks was the small capsule size, because there was no proper way to properly calibrate the microphone. A standard 94 db at 1 khz calibrator was put to use, but since much of the sound was escaping due to the larger capsule slot, the microphone was not correctly set, resulting in lower than expected values. Regardless, the frequency and time responses were both taken for position one in the anechoic chamber. A 90 second sweep time was selected for the frequency response, resulting in a 37.5 Hz resolution (Figure 21). The graph showed similar features to the Electrovoice tests, but with an even smoother curve. The regions above 10 khz and below 70 Hz were especially different: the Earthworks displayed much stronger results. Figure 21: Frequency response of the speaker in position 1 in the anechoic chamber using the Earthworks M-50 microphone. This graphs displays many similarities, but is much smoother in the lows and highs. The time response was once again conducted with the highest time resolution allowed in TEF. This time, the response of the speaker was significantly quicker than ones previously taken 19

20 (Figure 22). It only took about 1.5 ms for the direct sound decay into the nominal level, as opposed to more than 2.5 ms for the Electrovoice microphone. This was likely due to the quicker transient response of the Earthworks, due the fact that it was a condenser. Figure 22: Time response of the speaker in position 1 in the anechoic chamber using the Earthworks M-50 microphone. The condenser microphone displayed a much quicker response than the dynamic. Conclusion After thoroughly testing the Event 8XL studio monitor, many parameters were ascertained: the frequency response, the crossover frequency, the time response, and the polar directivity pattern. All of these tests were conducted under anechoic conditions, but the frequency and time responses were also re-tested under non-anechoic conditions. The result was a full, accurate view of the performance of the speaker under a variety of testing situations. Although much was learned, further tests could reveal even more about the details of the speaker. The polar and crossover tests could be completed using the Earthworks microphone, which could also be used in the non-anechoic environment. For the crossover, the microphone could be placed once again between the tweeter and woofer, but tested at several distances to determine if that factored into where the frequency dip and phase switch occurred. The function of the port could also be investigated by testing the response of it at varying distances. 20

Non-Anechoic Testing Environment

Non-Anechoic Testing Environment Non-Anechoic Testing Environment Acoustical Testing 1 Dr. Lauren Ronsse, Dr. Dominique Chéenne November 15 th, 2014 By: IRINEO JAIMES TEAM: CHRIS KEZON, HAYDEN JUBERA, NATHAN NEGRU 1 Table of Contents

More information

Acoustical Testing I Loudspeaker Study

Acoustical Testing I Loudspeaker Study Acoustical Testing I Loudspeaker Study Submitted to: Dr. Dominique Chéenne & Dr. Lauren Ronsse Columbia College Chicago November 19 th, 2014 By: Andrew Hulva Cody Elston, Daniel O Brien, Erich Pfister

More information

Acoustical Testing 1

Acoustical Testing 1 Material Study By: IRINEO JAIMES TEAM ANDREW MILLER SAM SHROYER NATHAN NEGRU ERICH PFISTER Acoustical Testing 1 Dr. Lauren Ronsse, Dr. Dominique Chéenne 11/05/2014 Table of Contents Abstract. 3 Introduction....3

More information

Processor Setting Fundamentals -or- What Is the Crossover Point?

Processor Setting Fundamentals -or- What Is the Crossover Point? The Law of Physics / The Art of Listening Processor Setting Fundamentals -or- What Is the Crossover Point? Nathan Butler Design Engineer, EAW There are many misconceptions about what a crossover is, and

More information

BIG 3 WAY SPEAKER: INTEGRATION OF BASS AND MIDRANGER DRIVERS. 3D Acoustics Research, January

BIG 3 WAY SPEAKER: INTEGRATION OF BASS AND MIDRANGER DRIVERS. 3D Acoustics Research, January BIG 3 WAY SPEAKER: INTEGRATION OF BASS AND MIDRANGER DRIVERS 1. Introduction 3D Acoustics Research, January 2010 www.3dar.ru In this article we show how 3D Response simulator can be used in low mid frequency

More information

The Naim Balanced Mode Radiator The Naim Ovator Bass Driver

The Naim Balanced Mode Radiator The Naim Ovator Bass Driver 1 The Naim Balanced Mode Radiator The Naim Ovator Bass Driver Lampos Ferekidis & Karl-Heinz Fink Fink Audio Consulting on behalf of Naim Audio Southampton Road, Salisbury SP1 2LN, ENGLAND The Balanced

More information

ONLINE TUTORIALS. Log on using your username & password. (same as your ) Choose a category from menu. (ie: audio)

ONLINE TUTORIALS. Log on using your username & password. (same as your  ) Choose a category from menu. (ie: audio) ONLINE TUTORIALS Go to http://uacbt.arizona.edu Log on using your username & password. (same as your email) Choose a category from menu. (ie: audio) Choose what application. Choose which tutorial movie.

More information

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024 Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 1 Suwanee, GA 324 ABSTRACT Conventional antenna measurement systems use a multiplexer or

More information

Airborne Sound Insulation

Airborne Sound Insulation Airborne Sound Insulation with XL2-TA Sound Level Meter This application note describes the verification of the airborne sound insulation in buildings with the XL2-TA Sound Level Meter. All measurements

More information

SIA Software Company, Inc.

SIA Software Company, Inc. SIA Software Company, Inc. One Main Street Whitinsville, MA 01588 USA SIA-Smaart Pro Real Time and Analysis Module Case Study #2: Critical Listening Room Home Theater by Sam Berkow, SIA Acoustics / SIA

More information

How To... Commission an Installed Sound Environment

How To... Commission an Installed Sound Environment How To... Commission an Installed Sound Environment This document provides a practical guide on how to use NTi Audio instruments for commissioning and servicing Installed Sound environments and Evacuation

More information

DESIGN OF ROOMS FOR MULTICHANNEL AUDIO MONITORING

DESIGN OF ROOMS FOR MULTICHANNEL AUDIO MONITORING DESIGN OF ROOMS FOR MULTICHANNEL AUDIO MONITORING A.VARLA, A. MÄKIVIRTA, I. MARTIKAINEN, M. PILCHNER 1, R. SCHOUSTAL 1, C. ANET Genelec OY, Finland genelec@genelec.com 1 Pilchner Schoustal Inc, Canada

More information

RD75, RD50, RD40, RD28.1 Planar magnetic transducers with true line source characteristics

RD75, RD50, RD40, RD28.1 Planar magnetic transducers with true line source characteristics RD75, RD50, RD40, RD28.1 Planar magnetic transducers true line source characteristics The RD line of planar-magnetic ribbon drivers represents the ultimate thin film diaphragm technology. The RD drivers

More information

Technique for the Derivation of Wide Band Room Impulse Response

Technique for the Derivation of Wide Band Room Impulse Response Technique for the Derivation of Wide Band Room Impulse Response PACS Reference: 43.55 Behler, Gottfried K.; Müller, Swen Institute on Technical Acoustics, RWTH, Technical University of Aachen Templergraben

More information

Multi-channel Active Control of Axial Cooling Fan Noise

Multi-channel Active Control of Axial Cooling Fan Noise The 2002 International Congress and Exposition on Noise Control Engineering Dearborn, MI, USA. August 19-21, 2002 Multi-channel Active Control of Axial Cooling Fan Noise Kent L. Gee and Scott D. Sommerfeldt

More information

HOW TO CREATE EASE LOUDSPEAKER MODELS USING CLIO

HOW TO CREATE EASE LOUDSPEAKER MODELS USING CLIO Daniele Ponteggia A procedure to measure loudspeaker polar patterns using CLIOwin 7 software and thus create a model for EASE 3.0 and EASE 4.1 for Windows software is described. Magnitude

More information

1. LOW PASS Frequency: Sets the - 3dB point of the low-pass filters from 110 to 440 Hz in 1/2-octave steps.

1. LOW PASS Frequency: Sets the - 3dB point of the low-pass filters from 110 to 440 Hz in 1/2-octave steps. The Electronic Crossover/Equalizer Unit The active Crossover/Equalizer control unit,which is part of the Reference Standard 4.5 system and optional with the RS 2.5, makes it possible for the user to compensate

More information

A White Paper on Danley Sound Labs Tapped Horn and Synergy Horn Technologies

A White Paper on Danley Sound Labs Tapped Horn and Synergy Horn Technologies Tapped Horn (patent pending) Horns have been used for decades in sound reinforcement to increase the loading on the loudspeaker driver. This is done to increase the power transfer from the driver to the

More information

Electro-Voice S40. Full Range Compact Speaker System 160 Watts Power Handling Available is Black or White

Electro-Voice S40. Full Range Compact Speaker System 160 Watts Power Handling Available is Black or White Electro-Voice S40 Full Range Compact Speaker System 160 Watts Power Handling Available is Black or White NOTE: This data sheet refers to several graphs. In order to keep the size of this document reasonable

More information

From time to time it is useful even for an expert to give a thought to the basics of sound reproduction. For instance, what the stereo is all about?

From time to time it is useful even for an expert to give a thought to the basics of sound reproduction. For instance, what the stereo is all about? HIFI FUNDAMENTALS, WHAT THE STEREO IS ALL ABOUT Gradient ltd.1984-2000 From the beginning of Gradient Ltd. some fundamental aspects of loudspeaker design has frequently been questioned by our R&D Director

More information

Holographic Measurement of the Acoustical 3D Output by Near Field Scanning by Dave Logan, Wolfgang Klippel, Christian Bellmann, Daniel Knobloch

Holographic Measurement of the Acoustical 3D Output by Near Field Scanning by Dave Logan, Wolfgang Klippel, Christian Bellmann, Daniel Knobloch Holographic Measurement of the Acoustical 3D Output by Near Field Scanning 2015 by Dave Logan, Wolfgang Klippel, Christian Bellmann, Daniel Knobloch LOGAN,NEAR FIELD SCANNING, 1 Introductions LOGAN,NEAR

More information

What you Need: Exel Acoustic Set with XL2 Analyzer M4260 Measurement Microphone Minirator MR-PRO

What you Need: Exel Acoustic Set with XL2 Analyzer M4260 Measurement Microphone Minirator MR-PRO How To... Handheld Solution for Installed Sound This document provides a practical guide on how to use NTi Audio instruments for commissioning and servicing Installed Sound environments and Evacuation

More information

LOW FREQUENCY SOUND IN ROOMS

LOW FREQUENCY SOUND IN ROOMS Room boundaries reflect sound waves. LOW FREQUENCY SOUND IN ROOMS For low frequencies (typically where the room dimensions are comparable with half wavelengths of the reproduced frequency) waves reflected

More information

Residual Phase Noise Measurement Extracts DUT Noise from External Noise Sources By David Brandon and John Cavey

Residual Phase Noise Measurement Extracts DUT Noise from External Noise Sources By David Brandon and John Cavey Residual Phase Noise easurement xtracts DUT Noise from xternal Noise Sources By David Brandon [david.brandon@analog.com and John Cavey [john.cavey@analog.com Residual phase noise measurement cancels the

More information

Application Note 4. Analog Audio Passive Crossover

Application Note 4. Analog Audio Passive Crossover Application Note 4 App Note Application Note 4 Highlights Importing Transducer Response Data Importing Transducer Impedance Data Conjugate Impedance Compensation Circuit Optimization n Design Objective

More information

AXIHORN CP5TB: HF module for the high definition active loudspeaker system "NIDA Mk1"

AXIHORN CP5TB: HF module for the high definition active loudspeaker system NIDA Mk1 CP AUDIO PROJECTS Technical paper #4 AXIHORN CP5TB: HF module for the high definition active loudspeaker system "NIDA Mk1" Ceslovas Paplauskas CP AUDIO PROJECTS 2012 г. More closely examine the work of

More information

Audiofrog UMI-1 Tuning CD Liner Notes

Audiofrog UMI-1 Tuning CD Liner Notes Audiofrog UMI-1 Tuning CD Liner Notes We have chosen and arranged the tracks on this CD to help make tuning your system using common tools and a real time analyzer as straightforward as possible. This

More information

Holographic Measurement of the 3D Sound Field using Near-Field Scanning by Dave Logan, Wolfgang Klippel, Christian Bellmann, Daniel Knobloch

Holographic Measurement of the 3D Sound Field using Near-Field Scanning by Dave Logan, Wolfgang Klippel, Christian Bellmann, Daniel Knobloch Holographic Measurement of the 3D Sound Field using Near-Field Scanning 2015 by Dave Logan, Wolfgang Klippel, Christian Bellmann, Daniel Knobloch KLIPPEL, WARKWYN: Near field scanning, 1 AGENDA 1. Pros

More information

PRELIMINARY. group S. M i cro W e dge M W 15 Speci f i cations. 2-Way full-range stage monitor See NOTES TABULAR DATA for details FEATURES DESCRIPTION

PRELIMINARY. group S. M i cro W e dge M W 15 Speci f i cations. 2-Way full-range stage monitor See NOTES TABULAR DATA for details FEATURES DESCRIPTION FEATURES Highly coherent coaxial design (15 inch LF) Enhanced fidelity and power handling Upgraded transducers and crossover design Switchable single-amp/bi-amp modes Convenient input connector location

More information

JBL Professional Application Note. Loudspeaker Array Low-Frequency Pattern Control using Filtered Array Technology

JBL Professional Application Note. Loudspeaker Array Low-Frequency Pattern Control using Filtered Array Technology JBL Professional Application Note Loudspeaker Array Low-Frequency Pattern Control using Filtered Array Technology 1: Overview Array directivity control theory is not new. Olson s Acoustical Engineering

More information

CHAPTER 3 THE DESIGN OF TRANSMISSION LOSS SUITE AND EXPERIMENTAL DETAILS

CHAPTER 3 THE DESIGN OF TRANSMISSION LOSS SUITE AND EXPERIMENTAL DETAILS 35 CHAPTER 3 THE DESIGN OF TRANSMISSION LOSS SUITE AND EXPERIMENTAL DETAILS 3.1 INTRODUCTION This chapter deals with the details of the design and construction of transmission loss suite, measurement details

More information

PRELIMINARY. group S. MicroWedge MW12 Specifications 2-WAY FULL-RANGE STAGE MONITOR FEATURES DESCRIPTION. CONFIGURATION Subsystem

PRELIMINARY. group S. MicroWedge MW12 Specifications 2-WAY FULL-RANGE STAGE MONITOR FEATURES DESCRIPTION. CONFIGURATION Subsystem FEATURES Highly coherent coaxial design (12-inch LF) Enhanced fidelity and power handling Upgraded transducers and crossover design Switchable single-amp/bi-amp modes Convenient input connector location

More information

ECMA-108. Measurement of Highfrequency. emitted by Information Technology and Telecommunications Equipment. 4 th Edition / December 2008

ECMA-108. Measurement of Highfrequency. emitted by Information Technology and Telecommunications Equipment. 4 th Edition / December 2008 ECMA-108 4 th Edition / December 2008 Measurement of Highfrequency Noise emitted by Information Technology and Telecommunications Equipment COPYRIGHT PROTECTED DOCUMENT Ecma International 2008 Standard

More information

App Note Highlights Importing Transducer Response Data Generic Transfer Function Modeling Circuit Optimization Digital IIR Transform IIR Z Root Editor

App Note Highlights Importing Transducer Response Data Generic Transfer Function Modeling Circuit Optimization Digital IIR Transform IIR Z Root Editor Application Note 6 App Note Application Note 6 Highlights Importing Transducer Response Data Generic Transfer Function Modeling Circuit Optimization Digital IIR Transform IIR Z Root Editor n Design Objective

More information

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types Exercise 1-3 Radar Antennas EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the role of the antenna in a radar system. You will also be familiar with the intrinsic characteristics

More information

Application Note 7. Digital Audio FIR Crossover. Highlights Importing Transducer Response Data FIR Window Functions FIR Approximation Methods

Application Note 7. Digital Audio FIR Crossover. Highlights Importing Transducer Response Data FIR Window Functions FIR Approximation Methods Application Note 7 App Note Application Note 7 Highlights Importing Transducer Response Data FIR Window Functions FIR Approximation Methods n Design Objective 3-Way Active Crossover 200Hz/2kHz Crossover

More information

Monitor Setup Guide The right monitors. The correct setup. Proper sound.

Monitor Setup Guide The right monitors. The correct setup. Proper sound. Monitor Setup Guide 2017 The right monitors. The correct setup. Proper sound. Table of contents Genelec Key Technologies 3 What is a monitor? 4 What is a reference monitor? 4 Selecting the correct monitors

More information

Transfer Function (TRF)

Transfer Function (TRF) (TRF) Module of the KLIPPEL R&D SYSTEM S7 FEATURES Combines linear and nonlinear measurements Provides impulse response and energy-time curve (ETC) Measures linear transfer function and harmonic distortions

More information

Operating Mode: PERFORMANCE Operating Range: 50 Hz to 19 khz Nominal Beamwidth: Horz 90 Vert 12 Axial Sensitivity (whole space SPL):

Operating Mode: PERFORMANCE Operating Range: 50 Hz to 19 khz Nominal Beamwidth: Horz 90 Vert 12 Axial Sensitivity (whole space SPL): FEATURES Extremely high output to size & weight ratio. Size and output capabilities allow it to be used in the widest variety of venues Exceptional pattern control due to symmetrical design, large horn(s)

More information

A mobile reverberation cabin for acoustic measurements in an existing anechoic room

A mobile reverberation cabin for acoustic measurements in an existing anechoic room A mobile reverberation cabin for acoustic measurements in an existing anechoic room Elsa PIOLLET 1 ; Simon LAROCHE 2 ; Marc-Antoine BIANKI 3 ; Annie ROSS 4 1,2,3,4 Ecole Polytechnique de Montreal, Canada

More information

[Q] DEFINE AUDIO AMPLIFIER. STATE ITS TYPE. DRAW ITS FREQUENCY RESPONSE CURVE.

[Q] DEFINE AUDIO AMPLIFIER. STATE ITS TYPE. DRAW ITS FREQUENCY RESPONSE CURVE. TOPIC : HI FI AUDIO AMPLIFIER/ AUDIO SYSTEMS INTRODUCTION TO AMPLIFIERS: MONO, STEREO DIFFERENCE BETWEEN STEREO AMPLIFIER AND MONO AMPLIFIER. [Q] DEFINE AUDIO AMPLIFIER. STATE ITS TYPE. DRAW ITS FREQUENCY

More information

ODEON APPLICATION NOTE ISO Open plan offices Part 2 Measurements

ODEON APPLICATION NOTE ISO Open plan offices Part 2 Measurements ODEON APPLICATION NOTE ISO 3382-3 Open plan offices Part 2 Measurements JHR, May 2014 Scope This is a guide how to measure the room acoustical parameters specially developed for open plan offices according

More information

Description. 175 watts Test Voltages: 34.2 volts rms 68.3 volts peak R SR (1.15 R E. ): 6.67 ohms. Suspending the Xi-1082

Description. 175 watts Test Voltages: 34.2 volts rms 68.3 volts peak R SR (1.15 R E. ): 6.67 ohms. Suspending the Xi-1082 Xi-1082 X-Array Install Two-Way, Full-Range, Sound-Reinforcement System 8-inch two-way 90 x 40 horn coverage 175-watts long term power capacity Ring-Mode Decoupling (RMD ) Full-range with internal passive

More information

Large Diaphragm Studio Condenser Microphone PROFESSIONAL MICROPHONES

Large Diaphragm Studio Condenser Microphone PROFESSIONAL MICROPHONES Large Diaphragm Studio Condenser Microphone PROFESSIONAL MICROPHONES Table Of Contents Introduction 3 CL7 Features 3 Operating the CL7 4 Powering the CL7 4 Setting up the Signal Level 4 Using the PAD switch

More information

DESIGN OF VOICE ALARM SYSTEMS FOR TRAFFIC TUNNELS: OPTIMISATION OF SPEECH INTELLIGIBILITY

DESIGN OF VOICE ALARM SYSTEMS FOR TRAFFIC TUNNELS: OPTIMISATION OF SPEECH INTELLIGIBILITY DESIGN OF VOICE ALARM SYSTEMS FOR TRAFFIC TUNNELS: OPTIMISATION OF SPEECH INTELLIGIBILITY Dr.ir. Evert Start Duran Audio BV, Zaltbommel, The Netherlands The design and optimisation of voice alarm (VA)

More information

EQ s & Frequency Processing

EQ s & Frequency Processing LESSON 9 EQ s & Frequency Processing Assignment: Read in your MRT textbook pages 403-441 This reading will cover the next few lessons Complete the Quiz at the end of this chapter Equalization We will now

More information

Loudspeaker Array Case Study

Loudspeaker Array Case Study Loudspeaker Array Case Study The need for intelligibility Churches, theatres and schools are the most demanding applications for speech intelligibility. The whole point of being in these facilities is

More information

Technical Notes Volume 1, Number 25. Using HLA 4895 modules in arrays: system controller guidelines

Technical Notes Volume 1, Number 25. Using HLA 4895 modules in arrays: system controller guidelines Technical Notes Volume 1, Number 25 Using HLA 4895 modules in arrays: system controller guidelines Introduction: The HLA 4895 3-way module has been designed for use in conjunction with the HLA 4897 bass

More information

Operating Mode: PERFORMANCE Operating Range: Nominal Beamwidth: (rotatable) Horz 90 Vert 45 Axial Sensitivity (whole space SPL):

Operating Mode: PERFORMANCE Operating Range: Nominal Beamwidth: (rotatable) Horz 90 Vert 45 Axial Sensitivity (whole space SPL): 2-WAY FULL-RANGE LOUDSPEAKER 90 45 See NOTES TABULAR DATA for details CONFIGURATION Subsystem: Operating Mode: Transducer LF 1 15 in cone HF 1 1.4 in exit, 3 in voice coil compression driver Single-amp

More information

Operating Mode: PERFORMANCE 1 Operating Range: 57 Hz to 15 khz Nominal Beamwidth: (rotatable) Horz 120 Vert 60 Axial Sensitivity (whole space SPL):

Operating Mode: PERFORMANCE 1 Operating Range: 57 Hz to 15 khz Nominal Beamwidth: (rotatable) Horz 120 Vert 60 Axial Sensitivity (whole space SPL): DESCRIPTION Premium Performance Application Flexibility The MK5300 Series of 2-way, high output, trapezoidal loudspeaker systems are designed as main PA elements for smaller venues, including small houses

More information

Excelsior Audio Design & Services, llc

Excelsior Audio Design & Services, llc Charlie Hughes March 05, 2007 Subwoofer Alignment with Full-Range System I have heard the question How do I align a subwoofer with a full-range loudspeaker system? asked many times. I thought it might

More information

INTERNATIONAL BACCALAUREATE PHYSICS EXTENDED ESSAY

INTERNATIONAL BACCALAUREATE PHYSICS EXTENDED ESSAY INTERNATIONAL BACCALAUREATE PHYSICS EXTENDED ESSAY Investigation of sounds produced by stringed instruments Word count: 2922 Abstract This extended essay is about sound produced by stringed instruments,

More information

The Mimir. Enclosure and stuffing. Drive units

The Mimir. Enclosure and stuffing. Drive units The Mimir Named after Mimir, a primal god of Norse mythology who was renowned for his knowledge and wisdom, we present a new high-end two-way speaker kit. The Mimir consist of an 18 cm long throw woofer

More information

SSE 26 SPECIFICATIONS

SSE 26 SPECIFICATIONS SPECIFICATIONS SSE 26 Description The new SSE enclosure series features high power ratings with custom components, a sleek look and a unique, snap-in metal grille. The SSE 26 is a two-way system featuring

More information

Product Description. Features. Applications

Product Description. Features. Applications Product Description The Tannoy i7 is a tall and slim loudspeaker cabinet equipped with four mm (4. ) low frequency direct radiating drive units and one 25mm (. ) titanium dome tweeter arranged as a line

More information

C30/C & C30/HC Hanging High Definition Microphone for Choir User s Guide

C30/C & C30/HC Hanging High Definition Microphone for Choir User s Guide Made in U.S.A. Earthworks, Inc. 37 Wilton Rd. Milford, NH 03055 603-654-6427 www.earthworksaudio.com email: sales@earthworksaudio.com Printed in U.S.A. C30/C & C30/HC Hanging High Definition Microphone

More information

EFFECT OF ARTIFICIAL MOUTH SIZE ON SPEECH TRANSMISSION INDEX. Ken Stewart and Densil Cabrera

EFFECT OF ARTIFICIAL MOUTH SIZE ON SPEECH TRANSMISSION INDEX. Ken Stewart and Densil Cabrera ICSV14 Cairns Australia 9-12 July, 27 EFFECT OF ARTIFICIAL MOUTH SIZE ON SPEECH TRANSMISSION INDEX Ken Stewart and Densil Cabrera Faculty of Architecture, Design and Planning, University of Sydney Sydney,

More information

REVERBERATION CHAMBER FOR EMI TESTING

REVERBERATION CHAMBER FOR EMI TESTING 1 REVERBERATION CHAMBER FOR EMI TESTING INTRODUCTION EMI Testing 1. Whether a product is intended for military, industrial, commercial or residential use, while it must perform its intended function in

More information

CHRISTIAN S. LÖTBÄCK PATANÉ. Master of Science Thesis

CHRISTIAN S. LÖTBÄCK PATANÉ. Master of Science Thesis Reverberation Chamber Performance and Methods for Estimating the Rician K-factor Evaluation of Reverberation Chamber Measurements at the National Institute of Standards and Technology in Boulder, Colorado,

More information

Design and Measurement of a Dipole Microphone (online section)

Design and Measurement of a Dipole Microphone (online section) Design and Measurement of a Dipole Microphone (online section) Mark Williamsen 5/17/09 submitted to Audio Amateur, Inc. for audioxpress Magazine This is the online section of the article Design and Measurement

More information

FISCHER CUSTOM COMMUNICATIONS, INC.

FISCHER CUSTOM COMMUNICATIONS, INC. FISCHER CUSTOM COMMUNICATIONS, INC. Current Probe Catalog FISCHER CUSTOM COMMUNICATIONS, INC. Fischer Custom Communications, Inc., is a manufacturer of custom electric and magnetic field sensors for military

More information

Application Note 5. Analog Audio Active Crossover

Application Note 5. Analog Audio Active Crossover App Note Highlights Importing Transducer Response Data Generic Transfer Function Modeling Circuit Optimization Cascade Circuit Synthesis n Design Objective 3-Way Active Crossover 4th Order Crossover 200Hz/2kHz

More information

Small Room and Loudspeaker Interaction

Small Room and Loudspeaker Interaction The common questions Several common questions are often asked related to loudspeaker s sound reproduction, such as: 1. Why does a loudspeaker sound different when moved to another room? 2. Why does my

More information

Design of a Line Array Point Source Loudspeaker System

Design of a Line Array Point Source Loudspeaker System Design of a Line Array Point Source Loudspeaker System -by Charlie Hughes 6430 Business Park Loop Road Park City, UT 84098-6121 USA // www.soundtube.com // 435.647.9555 22 May 2013 Charlie Hughes The Design

More information

OCV 6. Product Description. Features. Applications

OCV 6. Product Description. Features. Applications Product Description Tannoy OCV Series pendant loudspeakers deliver the renowned acoustic performance and exceptional value of Tannoy s CVS Series in-ceiling products in an architecturally sensitive, self-contained,

More information

Active Control of Energy Density in a Mock Cabin

Active Control of Energy Density in a Mock Cabin Cleveland, Ohio NOISE-CON 2003 2003 June 23-25 Active Control of Energy Density in a Mock Cabin Benjamin M. Faber and Scott D. Sommerfeldt Department of Physics and Astronomy Brigham Young University N283

More information

NTT DOCOMO Technical Journal. Method for Measuring Base Station Antenna Radiation Characteristics in Anechoic Chamber. 1.

NTT DOCOMO Technical Journal. Method for Measuring Base Station Antenna Radiation Characteristics in Anechoic Chamber. 1. Base Station Antenna Directivity Gain Method for Measuring Base Station Antenna Radiation Characteristics in Anechoic Chamber Base station antennas tend to be long compared to the wavelengths at which

More information

not overpower the audience just below and in front of the array.

not overpower the audience just below and in front of the array. SPECIFICATIONS SSE LA Description Designed for use in permanent professional installations in churches, theaters, auditoriums, gyms and theme parks, the SSE LA is a dual-radius dius curved line array that

More information

MK2399i Specifications

MK2399i Specifications DESCRIPTION Premium Performance Application Flexibility The MK2300i Series of 2-way, high output, trapezoidal loudspeaker systems are designed as main PA elements for smaller venues, including small houses

More information

Sound Design and Technology. ROP Stagehand Technician

Sound Design and Technology. ROP Stagehand Technician Sound Design and Technology ROP Stagehand Technician Functions of Sound in Theatre Music Effects Reinforcement Music Create aural atmosphere to put the audience in the proper mood for the play Preshow,

More information

Normalized Site Attenuation Test Report

Normalized Site Attenuation Test Report NVLAP LAB CODE 200974-0 Normalized Site Attenuation Test Report Test Specification NORMALIZED SITE ATTENUATION (NSA) Range 30 MHz 1GHz using the methods of ANSI C63.4-2009; EN 50147-2 (1997); CISPR 16-1-4

More information

INTRODUCTION. Please read this manual carefully for a through explanation of the Decimator ProRackG and its functions.

INTRODUCTION. Please read this manual carefully for a through explanation of the Decimator ProRackG and its functions. INTRODUCTION The Decimator ProRackG guitar noise reduction system defines a new standard for excellence in real time noise reduction performance. The Decimator ProRackG was designed to provide the maximum

More information

Measuring Instrument Combinations

Measuring Instrument Combinations PISTONPHONE START/STOP GRP NUM SLM RTA CAL LIGHT STORE MENU 1 FREQ WEIGHT TIME PUSH ON OFF ENT MODE 2 PAUSE/CONT 4 LEVEL 3 POWER PISTONPHONE Measuring Instrument Combinations 1 2 Acoustic Measurement Calibration

More information

How Accurate is Your Directivity Data?

How Accurate is Your Directivity Data? How Accurate is Your Directivity Data? A white paper detailing an idea from Ron Sauro: A new method and measurement facility for high speed, complex data acquisition of full directivity balloons By Charles

More information

Validation of lateral fraction results in room acoustic measurements

Validation of lateral fraction results in room acoustic measurements Validation of lateral fraction results in room acoustic measurements Daniel PROTHEROE 1 ; Christopher DAY 2 1, 2 Marshall Day Acoustics, New Zealand ABSTRACT The early lateral energy fraction (LF) is one

More information

Mei Wu Acoustics. By Mei Wu and James Black

Mei Wu Acoustics. By Mei Wu and James Black Experts in acoustics, noise and vibration Effects of Physical Environment on Speech Intelligibility in Teleconferencing (This article was published at Sound and Video Contractors website www.svconline.com

More information

SSE S5 SPECIFICATIONS

SSE S5 SPECIFICATIONS SPECIFICATIONS SSE S5 Description Designed for use in professional permanent installation in churches, theaters, auditoriums, gyms, and theme parks, the SSE S5 is a two-way speaker system, which provides

More information

PACX81. 1 OF of 6 PAGES PACX81 SPEAKER. Vision Series

PACX81. 1 OF of 6 PAGES PACX81 SPEAKER. Vision Series Vision Series The model (black) and W (white) are professional full-range, ultra-compact coaxial loudspeaker systems providing wide bandwidth and light weight in a low profile enclosure. The coaxial configuration

More information

ECMA-108. Measurement of Highfrequency. emitted by Information Technology and Telecommunications Equipment. 5 th Edition / December 2010

ECMA-108. Measurement of Highfrequency. emitted by Information Technology and Telecommunications Equipment. 5 th Edition / December 2010 ECMA-108 5 th Edition / December 2010 Measurement of Highfrequency Noise emitted by Information Technology and Telecommunications Equipment Reference number ECMA-123:2009 Ecma International 2009 COPYRIGHT

More information

Assessing the accuracy of directional real-time noise monitoring systems

Assessing the accuracy of directional real-time noise monitoring systems Proceedings of ACOUSTICS 2016 9-11 November 2016, Brisbane, Australia Assessing the accuracy of directional real-time noise monitoring systems Jesse Tribby 1 1 Global Acoustics Pty Ltd, Thornton, NSW,

More information

DC CIRCUITS AND OHM'S LAW

DC CIRCUITS AND OHM'S LAW July 15, 2008 DC Circuits and Ohm s Law 1 Name Date Partners DC CIRCUITS AND OHM'S LAW AMPS - VOLTS OBJECTIVES OVERVIEW To learn to apply the concept of potential difference (voltage) to explain the action

More information

Low frequency section: 500 Watts continuous 1,000 Watts program 2,000 Watts peak

Low frequency section: 500 Watts continuous 1,000 Watts program 2,000 Watts peak SPECIFICATIONS QW 3 Frequency response, 1 meter on-axis, swept-sine in an anechoic environment: 50 Hz 16 khz (±3 db) Usable low frequency limit (-10 db point): 33 Hz Power handling: Full range: 1,000 Watts

More information

Radiated Spurious Emission Testing. Jari Vikstedt

Radiated Spurious Emission Testing. Jari Vikstedt Radiated Spurious Emission Testing Jari Vikstedt jari.vikstedt@ets-lindgren.com What is RSE? RSE = radiated spurious emission Radiated chamber Emission EMI Spurious intentional radiator 2 Spurious Spurious,

More information

SSE 10 SPECIFICATIONS

SSE 10 SPECIFICATIONS SPECIFICATIONS SSE 10 Description The SSE is a new enclosure series featuring high power ratings with custom components, a sleek look and a unique, snap-in metal grille. The SSE 10 is a twoway system featuring

More information

THE USE OF VOLUME VELOCITY SOURCE IN TRANSFER MEASUREMENTS

THE USE OF VOLUME VELOCITY SOURCE IN TRANSFER MEASUREMENTS THE USE OF VOLUME VELOITY SOURE IN TRANSFER MEASUREMENTS N. Møller, S. Gade and J. Hald Brüel & Kjær Sound and Vibration Measurements A/S DK850 Nærum, Denmark nbmoller@bksv.com Abstract In the automotive

More information

Exercise 4. Angle Tracking Techniques EXERCISE OBJECTIVE

Exercise 4. Angle Tracking Techniques EXERCISE OBJECTIVE Exercise 4 Angle Tracking Techniques EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the principles of the following angle tracking techniques: lobe switching, conical

More information

Quadra 15 Available in Black and White

Quadra 15 Available in Black and White S P E C I F I C A T I O N S Quadra 15 Available in Black and White Frequency response, 1 meter onaxis, swept-sine in anechoic environment: 64 Hz to 18 khz (±3 db) Usable low frequency limit (-10 db point):

More information

SSE 12 SPECIFICATIONS

SSE 12 SPECIFICATIONS SPECIFICATIONS SSE 12 Description The SSE is a new enclosure series featuring high power ratings with custom components, a sleek look and a unique, snap-in metal grille. The SSE 12 is a twoway system featuring

More information

APPLICATION NOTE MAKING GOOD MEASUREMENTS LEARNING TO RECOGNIZE AND AVOID DISTORTION SOUNDSCAPES. by Langston Holland -

APPLICATION NOTE MAKING GOOD MEASUREMENTS LEARNING TO RECOGNIZE AND AVOID DISTORTION SOUNDSCAPES. by Langston Holland - SOUNDSCAPES AN-2 APPLICATION NOTE MAKING GOOD MEASUREMENTS LEARNING TO RECOGNIZE AND AVOID DISTORTION by Langston Holland - info@audiomatica.us INTRODUCTION The purpose of our measurements is to acquire

More information

Practical Impedance Measurement Using SoundCheck

Practical Impedance Measurement Using SoundCheck Practical Impedance Measurement Using SoundCheck Steve Temme and Steve Tatarunis, Listen, Inc. Introduction Loudspeaker impedance measurements are made for many reasons. In the R&D lab, these range from

More information

Loudspeakers. Juan P Bello

Loudspeakers. Juan P Bello Loudspeakers Juan P Bello Outline 1. Loudspeaker Types 2. Loudspeaker Enclosures 3. Multiple drivers, Crossover Networks 4. Performance Measurements Loudspeakers Microphone: acoustical sound energy electrical

More information

Audio Engineering Society. Convention Paper. Presented at the 141st Convention 2016 September 29 October 2 Los Angeles, USA

Audio Engineering Society. Convention Paper. Presented at the 141st Convention 2016 September 29 October 2 Los Angeles, USA Audio Engineering Society Convention Paper Presented at the 141st Convention 2016 September 29 October 2 Los Angeles, USA This paper is peer-reviewed as a complete manuscript for presentation at this Convention.

More information

Report Of. Shielding Effectiveness Test For. DefenderShield. Test Date(s): September 1 October 2, 2012

Report Of. Shielding Effectiveness Test For. DefenderShield. Test Date(s): September 1 October 2, 2012 Report Of Test For Test Date(s): September 1 October 2, 2012 UST Project No: Total Number of Pages Contained Within This Report: 15 3505 Francis Circle Alpharetta, GA 30004 PH: 770-740-0717 Fax: 770-740-1508

More information

New ON/OFF Absorption Technology That Includes Low Frequencies

New ON/OFF Absorption Technology That Includes Low Frequencies Toronto, Canada International Symposium on Room Acoustics 2013 June 9-11 ISRA 2013 New ON/OFF Absorption Technology That Includes Low Frequencies Niels W. Adelman-Larsen (nwl@flexac.com) Flex Acoustics

More information

Quadra 12 Available in Black and White

Quadra 12 Available in Black and White S P E C I F I C A T I O N S Quadra 12 Available in Black and White Frequency response, 1 meter onaxis, swept-sine in anechoic environment: 76 Hz to 18 khz (±3 db) Usable low frequency limit (-10 db point):

More information

Definition Install iw63 DC

Definition Install iw63 DC Defintion_iW63_Data_Sheet_DJ_amends:Layout 1 16/1/28 11:39 Page 1 Product Description The in wall speaker has been designed around the unique Tannoy Dual Concentric drive unit to bring audiophile performance

More information

ON MEASUREING LOUDSPEAKERS

ON MEASUREING LOUDSPEAKERS ON MEASUREING LOUDSPEAKERS BACKGROUND For the most part loudspeakers in the audio marketplace are only evaluated with subjective tests. Historically the dominance of these tests made sense because acoustic

More information

Radar Signatures and Relations to Radar Cross Section. Mr P E R Galloway. Roke Manor Research Ltd, Romsey, Hampshire, United Kingdom

Radar Signatures and Relations to Radar Cross Section. Mr P E R Galloway. Roke Manor Research Ltd, Romsey, Hampshire, United Kingdom Radar Signatures and Relations to Radar Cross Section Mr P E R Galloway Roke Manor Research Ltd, Romsey, Hampshire, United Kingdom Philip.Galloway@roke.co.uk Abstract This paper addresses a number of effects

More information

4GHz / 6GHz Radiation Measurement System

4GHz / 6GHz Radiation Measurement System 4GHz / 6GHz Radiation Measurement System The MegiQ Radiation Measurement System (RMS) is a compact test system that performs 3-axis radiation pattern measurement in non-anechoic spaces. With a frequency

More information