REPORT. EUROMET project no EUROMET.EM-S18 supplementary comparison of 1Ω and 10 kω resistance standards. P.O.Hetland, T. Sørsdal (JV, pilot 1 Ω)

Size: px
Start display at page:

Download "REPORT. EUROMET project no EUROMET.EM-S18 supplementary comparison of 1Ω and 10 kω resistance standards. P.O.Hetland, T. Sørsdal (JV, pilot 1 Ω)"

Transcription

1 REPORT EUROMET project no. 710 EUROMET.EM-S18 supplementary comparson of 1Ω and 10 kω resstance standards P.O.Hetland, T. Sørsdal (JV, plot 1 Ω) G. Eklund, O. Gunnarsson (SP, plot 10 kω), H. D. Jensen (DFM), A. Satrapnsk (MIKES) May 005 Abstract: Ths report presents the result of the supplementary comparson of measurements on 1 Ω and 10 kω resstance standards performed by the natonal nsttutes of Sweden, Norway, Denmark and Fnland. 1

2 Content 1. Introducton Partcpants and schedule Comparson schedule Organsaton of the comparson The travellng standards and measurement nstructons Descrpton of the travellng standards Measurement nstructons Method of measurements Behavour of the travellng standards Measurement results Results of the partcpatng nsttutes Calculaton of the reference value and ts uncertanty Degree of equvalence (DoE) of the partcpatng nsttutes wth respect to the reference value Summary and concluson... 1 References... 1 Appendx A: Method of measurements Appendx B: Uncertanty budgets... 15

3 1. Introducton Many natonal laboratores use the quantum Hall effect as a means to establsh a reference standard of dc resstance. When the lowest possble uncertanty s desrable for the scalng from the Quantum Hall Resstance (QHR) to conventonal resstance standards a cryogenc current comparator (CCC) s often used. The scalng from QHR to a 100 Ω resstance standard s often the frst step n the scalng process. A EUROMET key comparson of Resstance at 100 Ω, EUROMET.EM- K10 began n the sprng of 003 and wll soon be completed. Comparson of the scalng from QHR or 100 Ω to 1 Ω and 10 kω s also very mportant. Such a BIPM comparson was carred out n In May to November 003 the natonal nsttutes of Denmark, Fnland, Norway and Sweden performed a supplementary comparson on 1 Ω and 10 kω resstance standards. Ths comparson allows for a clear and unequvocal comparson of the measurement results and wll show the equvalence of measurng results obtaned wth varous quantzed Hall systems or other measurement systems for resstance n the partcpatng natonal nsttutes.. Partcpants and schedule There were four laboratores partcpatng n ths comparson. The laboratores are lsted n Table 1. Laboratory Country JV Norwegan Metrology Servce Norway DFM Dansh Insttute for Fundamental Metrology Denmark MIKES Centre for Metrology and Accredtaton Fnland SP Swedsh Natonal Testng and Research Insttute Sweden Table 1: Lst of partcpants.1 Comparson schedule Crculaton Tme Schedule Supplementary Comparson EUROMET project 710 ( ) Set A1 & A Insttuton Country Start date Tme for Comment measurement and transportaton SP (start 10 kω) Sweden May weeks 10 kω JV (start 1 Ω) Norway June weeks 10 kω +1 Ω DFM Denmark July weeks 10 kω +1 Ω MIKES Fnland August weeks 10 kω +1 Ω SP (plot 10 kω) Sweden September weeks 10 kω +1 Ω JV (plot 1 Ω) Norway November weeks 1 Ω 3

4 . Organsaton of the comparson For ths supplementary comparson JV was plot laboratory for the 1Ω travellng standards and SP for the 10 kω travellng standard. The comparson was performed n parallel wth the ongong 100 Ω EUROMET key comparson, EUROMET.EM-K10. The 1 Ω and 10 kω travellng standards were crculated together wth two 100 Ω resstors. All travellng standards were transported by car and kept n temperature regulated enclosures. After crculaton the travellng standards were returned to the plot laboratores. Each laboratory had 4 weeks to perform the measurements. 3. The travellng standards and measurement nstructons 3.1 Descrpton of the travellng standards The travellng resstance standards were one 10 kω ESI standard and two CSIRO 1Ω standards. The 10 kω standard had a temperature stablsed enclosure wth nternal temperature measurement usng a Pt 100 thermometer. The 1Ω standards where kept n olbaths at each laboratory but transported n a temperature regulated box. DFM used the temperature regulated box nstead of an olbath. Set A1, 10 kω (SP): Standard Resstor 10 kω ESI, QHR transfer standard wth temperature stablsed enclosure, S/N SP91010 Pt 100 measurement cable Power supply 1V Instructon Manual. Set A, CSIRO 1 Ω (JV): Standard Resstor 1 Ω CSIRO, S/N 64179, Standard Resstor 1 Ω CSIRO, S/N Temperature regulated transport box 3. Measurement nstructons The measurand s the value of the resstance at DC, based on the conventonal value of the von Kltzng constant R K-90 = 5 81,807 Ω. In practce, DC means that the watng tme between the end of a current reversal and the start of data acquston should not be shorter than 8 s for the 10 kω travellng standard and 4 s for the 1 Ω travellng standards. It s up to the partcpants to ether make a guarded measurement where the resstor case s guarded, leave the resstor floatng wth respect to the case, or connect one pont of the resstor to ts case. The soluton adopted should be reported. A short descrpton of the methods used should be ncluded n the fnal report together wth the measurement results. After nstallaton of the travellng standards n ther respectve bath and thermostat n the laboratory, a mnmum settlng tme of two days s requred for the 1Ω travellng 4

5 standards and three days for the 10 kω travellng standard. The measurements should be carred out wth the followng preferred condtons: drect or ndrect (usng a 100 Ω resstor) comparson wth the QHR usng a CCC brdge, amed uncertanty less than (95% confdence level), 50 ma through the 1 Ω resstor, appled measurement voltage 10 kω standard 0.5V 1V (max appled 3 V) ol bath temperature for the 1Ω standards (3,00 ± 0,01) C ambent temperature for the10 kω enclosure (0,0 4,0) C Laboratores not usng the QHR as ther prmary standard of resstance shall measure the resstors wth ther respectve best measurement capablty. For these measurements the source of traceablty has to be ncluded n the measurement report. The travellng standard temperature and ambent pressure should be recorded and reported as well as the heght of ol above the top plate of the resstors n the ol bath. If known, the densty of the ol n the ol bath should be reported. Resstors may have a long thermal tme constant (several hours)! The measurements should be made at dfferent dates durng the perod n the laboratory. The temperature and pressure coeffcents of the travellng standard have been determned to allow for correctons. They are not gven n the protocol. 4. Method of measurements Dfferent measurement set-ups were used n the comparson. One of the laboratores used a conventonal DCC brdge and has traceablty to BIPM through 10 kω resstor standards. The scalng from 10 kω to 1 Ω s acheved prmarly va two Hamon transfer devces. The three remanng laboratores have traceablty to ther own QHR standards. The travellng standards were compared drectly aganst 100 Ω standards usng CCC resstance brdges. The 100 Ω standards were calbrated aganst QHR. The 10 kω travellng standard plot laboratory (SP) measurements was drectly aganst the QHR usng the dc-ccc brdge. Two of these three laboratores use the same type of commercal dc-ccc brdge. These brdges are based upon a Natonal Physcal Laboratory (NPL) desgn [1]. The thrd laboratory uses a self desgned ac-ccc brdge (0.1 Hz to 0.3 Hz) []. Further detals are gven n appendx A. 5. Behavour of the travellng standards The plot laboratores measured the respectve travellng standards before and after the crculaton. From these measurements the drft and stablty of the travellng standards have been evaluated. The estmated drft rate has also been compared wth the hstorcal drft rate of the travellng standards. All three travellng standards had a drft rate as expected from the hstorcal data durng the comparson. The temperature regulaton of the 10 kω standard also showed a very good behavour 5

6 wth small temperature varatons between the measurements at the partcpatng laboratores. 6. Measurement results 6.1 Results of the partcpatng nsttutes Each laboratory has reported a mean value, R lab, the temperature, the mean date, ambent condtons and the measurement uncertanty for each travellng standard. The sngle values from whch the mean values are estmated and complete uncertanty budgets are also reported. The travellng standards have small temperature dependences. The two 1 Ω travellng standards also have a small pressure dependence. These two effects are corrected for n tables -4. The corrected values for all three resstors are shown n fgures 1-3. Due to the small varatons of the temperature of the 10 kω travellng standard enclosure the temperature correctons were for all laboratores below 3 nω/ω for the 10 kω travellng standard. Laboratory Date 1 Ω sn (Ω) Exp. uncertanty (nω) JV , DFM , MIKES , SP , JV , Table : Corrected measurement results for 1Ω travellng standard sn Laboratory Date 1 Ω sn (Ω) Exp. uncertanty (nω) JV , DFM , MIKES , SP , JV , Table 3: Corrected measurement results for 1Ω travellng standard sn Laboratory Date 10 kω sn. SP91010 (Ω) Exp. uncertanty (mω) SP , ,6 JV , ,09 DFM , ,78 MIKES , ,50 SP , ,6 Table 4: Corrected measurement results for 10 kω travellng standard sn.sp

7 Nordc 1 Ω (sn ) EUROMET.EM-S18 supplementary comparson 100 Devaton from nomnal value µω/ω JV (before/after) DFM MIKES SP Ref.value Measurement date Fgure 1: Results for 1Ω travellng standard, sn , wth expanded uncertanty Nordc 1 Ω (sn ) EUROMET.EM-S18 supplementary comparson Devaton from nomnal value µω/ω JV (before/after) DFM MIKES SP Ref.value Measurement date Fgure : Results for 1Ω travellng standard, sn , wth expanded uncertanty 7

8 Nordc 10 kω EUROMET.EM-S18 supplementary comparson 80 Devaton from nomnal value -6.9 µω/ω SP JV DFM MIKES Ref. value Measurement date Fgure 3: Results for 10 kω travellng standard, sn. SP91010, wth expanded uncertanty 6. Calculaton of the reference value and ts uncertanty For a stable reference standard t s possble to estmate the reference value usng the method of least squares. Stablty of the travellng standards was provded by organsaton of careful transportaton and we assume that change of the resstors was small. The reference value can be determned by the weghted mean of the laboratores measurements n that case. The measurements results from the plot laboratores and old drft data ndcate that some drft n the standards has to be ncluded n the evaluaton of the reference values. Our proposal for the determnaton of the reference value s to determne the weghted ft to a straght lne, Y=a+bX, where only parameter a s ftted. The parameter b s calculated from the plot laboratory results. a = w y b w w x where the weght w s c w = u x ) ( wth c=0,5 for the plot laboratory and c=1 for the other laboratores. 8

9 In addton to the weght based on the measurement uncertanty, the plot laboratores are gven a weght of 0,5 because they contrbute wth two measurement results. The calculated reference values are shown as a dashed lne n fgure 1-3. The standard uncertanty of the reference value s determned from u 1 ( x = ) u 1 ( x ) ( ref 1 x N u ) 6.3 Degree of equvalence (DoE) of the partcpatng nsttutes wth respect to the reference value The laboratores degree of equvalence wth respect to the reference value are calculated accordng to the CCEM gudelnes [3]. For nsttute =1,., N the degree of equvalence s gven by d = x x ref wth the expanded uncertanty U ( d ) = u( d ) when u(d ) s gven by u ( d ) = u ( x ) u ( x ref ) Laboratory DoE ± U(DoE) 1Ω sn Ω sn kω sn SP91010 JV 19 ±7 17 ±5 0 ± 3 DFM 35 ±540 4 ± ± 178 MIKES -6 ±9-14 ±7 0 ± 49 SP -19 ±33-3 ±68 5 ± 5 Table 5: Degrees of equvalence (DoE) wth expanded uncertanty for the ndvdual travellng standards. Snce two travellng standards have been used n the 1 Ω measurements, t s normal to report only one combned DoE. The two 1 Ω results for each laboratory can be defned to be R 1, and R,. The measurants we want to compare are the average results from the two 1 Ω measurements. 9

10 R avg, = R 1, + R, wth the uncertanty ( u ( R ) + u ( R )) 1 u( Ravg ) = u ( R0, ) + 1,, 4, R 0, are the type B components wth an uncertanty u(r 0, ) whch represent the system uncertanty common for both 1 Ω measurements. δr 1, and δr, are the type A components for each 1 Ω measurements wth the uncertanty u(δr 1, ) and u(δr, ). Then wth R avg, and u(r avg, ) for each laboratory a reference value and DoE among the laboratores are calculated n the same way as for the ndvdual 1 Ω standards shown above. Laboratory DoE ± U(DoE) 1Ω 10 kω sn SP91010 JV 19 ± 5 0 ± 3 DFM 40 ± ± 178 MIKES -9 ± 8 0 ± 49 SP -0 ± 38 5 ± 5 Table 6: Degrees of equvalence (DoE) wth expanded uncertanty for the 1Ω and 10 kω measurements. 10

11 Degrees of equvalence - 1 Ω Degrees of equvalence (n Ω ) Date JV DFM MIKES SP Degrees of equvalence - 10 kω sn. SP91010 Degrees of equvalence (m Ω ) Date JV DFM MIKES SP Fgure 4: DoE wth respect to the reference value for the 1Ω and 10 kω measurements. 11

12 7. Summary and concluson The comparson ndcates that t s possble to compare resstance standards and estmate systematc effects n measurement equpment wth uncertantes of parts n 10-8 for 1Ω and 10 kω usng conventonal resstance standards and careful transport. Personally transported standards, mantaned at stable temperature, showed very stable behavour. Ths s mportant for the possblty to check for systematc uncertantes n resstance scalng n the range 1Ω - 10 kω wth CCC-brdges of dfferent desgn. The results of the 1 Ω comparson show good agreement among the partcpants. The agreement of the 10 kω comparson s excellent. References [1] J.M. Wllams and A. Hartland, An Automated Cryogenc Current Comparator Resstance Rato Brdge, IEEE Trans. Instrum. Meas., Vol. 40, pp 67-70, [] H. Seppä and A. Satrapnsk, AC Resstance Brdge Based on the Cryogenc Current Comparator, IEEE Trans. Instrum. Meas., Vol. 46, No., pp , 1997 [3] M.G. Cox, The evaluaton of key comparson data, Metrologa, 00, 39,

13 Appendx A: Method of measurements The measurement method of SP The measurements are based on the quantum Hall effect. The QHE cryostat has a varable temperature nsert and QHE samples can be cooled to 1,6 K. The magnet can produce operatng felds to 1 Tesla at 4, K. The Hall sample s mounted on a header of the TO-8 type and the room temperature connectons are gold plated bnd post termnals. The used QHE sample s charactersed wth a constant current source connected to the current termnals. The measurements have been performed followng the techncal gudelnes n Delahaye F 1989 Metrologa The realsed QHE resstance, the QHR, s compared wth room temperature resstance standards wth nomnal values 100 Ω and 10 kω wth a cryogenc current comparator (CCC) or a Josephson potentometer (10 kω measurements). All results reported here are from CCC measurements. The CCC system s manufactured by Oxford Instruments under a technology transfer agreement wth Natonal Physcal Laboratory UK [1]. The CCC measurements are performed n a postve-negatve-postve sequence wth two current reversals. The total number of measurements s the same n postve and negatve polarty. For measurements below the 100 Ω level the CCC measurements have to be made n two steps. In the case of the measurements on the 1 Ω standards reported here, a temperature controlled 100 Ω resstance standard s used as a transfer standard. The measurement method of MIKES The 1 Ω travellng standards were measured wth the MIKES ac Cryogenc Current Comparator (CCC) Resstance Brdge. The 1 Ω resstors were compared drectly wth the two 100 Ω transfer standards, Tr1 and Tr. These 100 Ω resstors were measured n July 003 aganst the MIKES QHR standard. The 1 Ω standards were measured at the frequences 0,1 Hz 0,3 Hz, and wth the currents (rms value) n the 100 Ω: 0,35 ma, 0,5 ma and 0,7 ma, (or equvalent dc currents n 1 Ω 35 ma, 50 ma, and 70 ma). The temperature of the 1 Ohm resstors was measured by temperature meter Hart Scentfc (model 159R), wth the calbrated Pt100 sensors placed nsde the wells of the resstors. The 100 Ω references are n ther own enclosures wth the mantaned temperatures: T Tr1 = 30,73 C, and T Tr = 30,5 C. These temperatures are mantaned permanently wth the nstabltes of 0,003 C and, at these temperatures the temperature coeffcents are close to zero. 13

14 The 10 kω was compared also drectly wth the two 100 Ω transfer standards, Tr1 and Tr wth the MIKES ac Cryogenc Current Comparator (CCC) Resstance Brdge. The applcaton of low frequency currents allows comparng the magnary components of the measured standards. Durng measurements wth the 10 kω a bgger magnary part and a hgher frequency dependence n the real part was notced, compared to other types of 10 kω. Ths ndcates probably addtonal dsspaton n capactve components of that standard. The ncreased frequency dependence and the uncertanty of nterpolaton were the reasons of an addtonal contrbuton to the uncertanty budget of that resstor. The measurement method of JV JV s system s an automatc QHR-system from Oxford Instruments. It conssts of a 16T magnet wth a top loadng 3 He nsert and a Cryogenc Current Comparator (CCC). The CCC-brdge s produced by Oxford Instruments under a technology transfer agreement wth NPL [1]. The CCC brdge s used n the calbraton of the 1 Ω and 10 kω resstors usng a 100 Ω resstor as reference standard. Ths reference standard s calbrated aganst the QHR standard. The measurement method of DFM DFM acheves at present ts traceablty for resstance from BIPM. A set of three ESI SR kω resstors serve as DFM reference standard, and each of these are n turn calbrated by BIPM over a three year perod. Scalng over the range 1 Ω to 10 kω s acheved prmarly va two Hamon transfer devces, Guldlne 9350/1 kω and Guldlne 9350/10 kω. The scalng s ndependently checked va a set of thermalsed resstors n a 1::5 sequence. Workng standards at 10 kω, 100 Ω, etc. are used as check standards. Resstance rato measurements are performed usng a Guldlne 6675A Drect Current Comparator Brdge. The resstors are connected to a Measurement Internatonal 40A Four-Termnal Matrx scanner. The case of the ESI and Hamon transfer resstors s connected to the reference potental va the screen of the potental carryng nterconnectng cable. The screen of the current carryng cable s connected to the reference pont at the scanner sde. A program connect the resstors, set up the measurement parameters of the brdge, start the measurement, and read the resstance rato measured by the brdge. Temperature measurements, Pt 100 sensor readngs, as well as recordng of the envronmental condtons are performed smultaneously. 14

15 Appendx B: Uncertanty budgets The uncertanty budget of MIKES for 1Ω Source Table II Error budget n determnaton of 1 Ω from the 100 Ω by CCC brdge. Input Quantty X Estmate x I Standard Uncertanty of nput u(x ) Probablty Dstrbuton / method of evaluaton (A,B) Senstvty Coeffcent c I (ΔR x / ΔX ) Uncertanty Contrbuton u (R x ) = u(x )* (ΔR x / ΔX ) Degree of Freedom ν I 100 Ohm Reference resstance, Calbraton from QHR, low frequency fluctuaton. Reference resstance, Current dependence, 100 Ω Square / B Infnte Power coeffcent µω/ω / mw Square / B Infnte Rato, compensaton current error I comp / I Square / B Infnte Rato, CCC wndng error, dc -1 Hz N 1 / N, 16: Square / B Infnte Rato, dynamc error, FB gan I 1 / I Square / B Infnte CCC brdge Nose rectfcaton, squd voltage Voltage V Square / B 10-7 / V Infnte bas changes, flux jumps, nonlnearly of null-detector Error due to uncompensated ac Voltage Square / B Infnte currents and leakage currents n voltage lnk Zero offset n brdge Voltage Square / B Infnte Extrapolaton to dc Rato Square / B Infnte Contact resstance n rotary Resstance Normal / A swtch of I comp µω/ω Temperature* 3.0 C C Square / B 0.5 µω/ω / C* Infnte 1 Ohm Resstor Temperature coeffcents (α, β)* 0.5* µω/ω / C Current dependence* Power coeffcent <0.5 µω/ω / C < µω/ω / mw 1 µω/ω/ < Infnte Square / B (µω/ω/ C) Square / B 1 < Infnte Pressure* 1013 hpa 0.1 hpa Square / B < / hpa < Infnte Low frequency fluctuaton** Resstance Square / B Infnte Measurement*** Voltage Normal / A R1 Ω ν eff = 8984 The expanded uncertanty s 36 nω/ω for k=. 15

16 Source The uncertanty budget of MIKES for 10 kω Table II Error budget n determnaton of 10 kω from the 100 Ω by ac CCC brdge. Input Quantty X Estmate x I Standard Uncertanty of nput u(x ) Probablty Dstrbuton / method of evaluaton (A,B) Senstvty Coeffcent c I (ΔR x / ΔX ) Uncertanty Contrbuton u (R x ) = u(x )* (ΔR x / ΔX ) Degree of Freedom ν I 100 Ohm Reference resstance, Calbraton from QHR, low frequency fluctuaton. 100 Ω Square / B Infnte Rato, compensaton current error I comp / I Square / B Infnte Rato, CCC wndng error, dc -1 Hz N 1 / N, 16: Square / B Infnte Rato, dynamc error, FB gan I 1 / I Square / B Infnte CCC brdge Nose rectfcaton, squd Voltage 10 - V Square / B 10-7 / V Infnte voltage bas changes, flux jumps Error due to Voltage Square / B Infnte uncompensated ac currents and leakage currents n voltage lnk Zero offset n brdge Voltage Square / B Infnte Extrapolaton to dc Rato Square / B Infnte 10 kohm Resstor Contact resstance n Resstance Normal / A rotary swtch of I comp µω/ω Temperature coeffcents 30 C C Square / B 0.1 µω/ω / C* Infnte Temperature coeffcents µω/ω 1 µω/ω/ Infnte (α, β)* µω/ω / C / C square / B (µω/ω/ C) Pressure* 1013 hpa 0.1 hpa Square / B < / hpa < Infnte Low frequency Resstance Square / B Infnte fluctuaton** Measurement*** Voltage Normal / A R10 kω ν eff = 8984 The expanded uncertanty s 50 nω/ω for k=. * temperature and pressure coeffcents are not known and the uncertanty are estmated approxmately. ** estmate of possble combned maxmum nfluence. *** Estmate for typcal parameters n one run of measurements: current (rms) n 100 Ω 5 ma, frequency 0.1 Hz, tme of measurement - 0 mn (degree of freedom 180). 16

17 The uncertanty budget of SP for 1 Ω The model for the measurements s: Rx = Rs (1+ δtsd) r ( 1+ δcwr + δclr + δcsr + δrxl ) Where Rx s the unknown 1 Ω resstor. Rs s the 100 Ω QHR transfer resstor. δtsd s the relatve error due to the QHR 100 Ω transfer standard drft. r s the rato Rx/Rs measured by the CCC brdge. δcwr s the relatve error due to the CCC wndng rato devaton from nomnal. δclr s the relatve error due the CCC leakage resstance. δcsr s the relatve error due to the error of the shunt resstor value. δrxl s the relatve error due to the error of the nternal lead resstance of the shunted resstor. The relatve standard uncertanty s gven by : u( Rx) = Rx u( R S) u( r) + RS r + u( δ ) Where u ( r) r s the relatve standard devaton of the mean for the measurement results. Uncertanty budget for R X = 1 Ω travellng standard s/n Quan- Estmate Relatve Probablty Senstvty Relatve Degree tty standard dstrbuton coeffcent uncertanty of X x uncertanty u(x ), (10-9 ) / method of evaluaton (A,B) c contrbuton u (R x ), freedom ν Rs Ω 13.0 Normal/A+B δtsd Normal/B r Normal/A δqpl Rectangular/B δcwr Normal/A δclr Rectangular/B δcsr Normal/B δrxl Normal/B R X Ω The expanded uncertanty s 39 nω/ω for k=.1 17

18 Uncertanty budget for R X = 1 Ω travellng standard s/n Quan- Estmate Relatve Probablty Senstvty Relatve Degree tty standard dstrbuton coeffcent uncertanty of X x uncertanty u(x ), (10-9 ) / method of evaluaton (A,B) c contrbuton u (R x ), freedom ν Rs Ω 13.0 Normal/A+B δtsd Normal/B r Normal/A δqpl Rectangular/B δcwr Normal/A δclr Rectangular/B δcsr Normal/B δrxl Normal/B R X Ω The expanded uncertanty s 7 nω/ω for k=.87 The uncertanty budget of SP for 10 kω The model for the measurements s: Rx = QHR (1+ δqhr + δqpl) r ( 1+ δcwr + δclr + δcsr + δrxl ) Where Rx s the unknown 10 kω resstor. QHR s the realsed quantum Hall resstance at plateau = wth the exact numercal value 1906,4035 Ω. δqhr s the relatve error of the realsed Hall resstance due to mperfect quantzaton and effects of mperfect contacts on the Hall sample. δqpl s the relatve error due to the QHR probe leakage resstance. r s the rato Rx/QHR measured by the CCC brdge. δcwr s the relatve error due to the CCC wndng rato devaton from nomnal. δclr s the relatve error due the CCC leakage resstance. δcsr s the relatve error due to the error of the shunt resstor value. δrxl s the relatve error due to the error of the nternal lead resstance of the shunted resstor. 18

19 The relatve standard uncertanty s gven by : u( Rx) = Rx u( r r) + u( δ ) Where u ( r) r s the relatve standard devaton of the mean for the measurement results. Uncertanty budget for R X = 10 kω travellng standard s/n SP91010 Quantty X Estmate x Relatve standard uncertanty u(x ), (10-9 ) Probablty dstrbuton / method of evaluaton (A,B) Senstvty coeffcent c Relatve uncertanty contrbuton u (R x ), Degree of freedom ν QHR 1906,4035 Ω r Normal/A δqhr Normal/B δqpl Rectangular/B δcwr Normal/A δclr Rectangular/B δcsr 0.0 Normal/B 1.0 δrxl 0 0. Normal/B 1 0. R X Ω The expanded uncertanty s 5 nω/ω for k= 19

20 Uncertanty budget for R X = 10 kω travellng standard s/n SP91010 Quantty X Estmate x Relatve standard uncertanty u(x ), (10-9 ) Probablty dstrbuton / method of evaluaton (A,B) Senstvty coeffcent c Relatve uncertanty contrbuton u (R x ), Degree of freedom ν QHR 1906,4035 Ω r Normal/A δqhr Normal/B δqpl Rectangular/B δcwr Normal/A δclr Rectangular/B δcsr 0.0 Normal/B 1.0 δrxl 0 0. Normal/B 1 0. R X Ω The expanded uncertanty s 6 nω/ω for k= 0

21 Budget of uncertanty of JV for 1 Ω The model for the measurement s: R x = R r s ( wnd leak bal shunt rect p 1+ δ + δ + δ + δ + δ + δ ) The components are: R x : the unknown resstor R s : the 100 Ω reference standard: 100, Ω r : the rato measured by the CCC-brdge δ wnd : the relatve wndng rato error δ leak : the relatve error due to leakage resstance δ bal : the relatve error due to brdge balancng δ shunt : the relatve error due to the stablty and calbraton of the shunt resstor δ rect : the relatve error due to nose rectfcaton δ p : the relatve error due to change n power dsspaton n the 100 reference standard The relatve standard uncertanty s then gven by : u( R R x x ) = u( R Rs s ) u( r + r ) + u ( δ ) Whch gve the followng uncertanty budget for R x : Uncertanty budget for R X = 1 Ω travellng standard s/n Quantty Estmate Relatve Probablty Senstvty Relatve Degree standard dstrbuton coeffcent uncertanty of X x uncertanty / method of u(x ), evaluaton(a,b) c contrbuton u (R x ), freedom R s Ω 3.6 normal r normal δ wnd 0 1 normal 1 1 δ leak 0 1 rectangular 1 1 δ bal 0.0 normal δ shunt normal δ rect 0 1 rectangular 1 1 δ p 0 15 rectangular 1 15 R X Ω The expanded uncertanty s 34 nω/ω for k=. ν 1

22 Budget of uncertanty of JV for 10 kω The model for the measurement s: R x = Rs r ( 1 + δ wnd + δ leak + δ bal + δ shunt + δ rect ) The components are: R x : the unknown resstor R s : the 100 Ω reference standard: Ω r : the rato measured by the CCC-brdge δ wnd : the relatve wndng rato error δ leak : the relatve error due to leakage resstance δ bal : the relatve error due to brdge balancng δ shunt : the relatve error due to the stablty and calbraton of the shunt resstor δ rect : the relatve error due to nose rectfcaton The relatve standard uncertanty s then gven by : u( R R x x ) = u( R Rs s ) u( r + r ) + u ( δ ) Whch gve the followng uncertanty budget for R x : Uncertanty budget for R X = 10 kω travellng standard s/n SP91010 Quantty Estmate Relatve Probablty Senstvty Relatve Degree X x standard dstrbuton uncertanty / method of u(x ), evaluaton(a,b) coeffcent uncertanty of c contrbuton u (R x ), freedom R s Ω 3.6 normal r normal δ wnd 0 1 normal 1 1 δ leak 0 1 rectangular 1 1 δ bal 0.0 normal δ shunt normal δ rect 0 1 rectangular 1 1 R X Ω The expanded uncertanty s 9.0 nω/ω for k=. ν

23 The uncertanty budget of DFM for 1 Ω The model for the measurement s: ε X = ε + δε S SP + δ ( + δrs + δrc ) T X T, S + r δ, where ε X s the relatve devaton from nomnal (RDN) of the unknown resstor, ε S s the RDN of the reference resstor. The δε SP s the seres-parallel transfer error of the Hamon transfer devce. The term δ T,S s the temperature correcton of the reference, r s the rato as measured, δr S s the specfcaton of the current comparator brdge, δr C s the correcton of the brdge error, and δ T,X s the temperature correcton of the unknown resstor. The uncertanty budget becomes: Quantty (unt) Dstrbuton x U(x ) n c u (y) 1 ε s, rel.dev. from nomnal of standard (10-6 ) Normal , ,150 δε SP, Seres-parallel transfer error (10-6 ) Rectangular 0,000 0,058 nfnty 1 0,058 3 δ T,S, Temp devaton standard ±0,1 C (10-6 ) Rectangular 0,000 0,080 nfnty 1 0,080 4 r, Rato devaton as determned (10-6 ) Normal -4,547 0, ,005 5 δr S, Specfcaton for Guldlne brdge (10-6 ) Rectangular 0,000 0,058 nfnty 1 0,058 δr C, Correcton of brdge error (10-6 ) Normal 0,000 0, ,001 y ε X, Rel. dev. From nom. of unknown (10-6 ) Normal -0,9757 0,73 37 The expanded uncertanty s 540 nω/ω for k=. The uncertanty budget of DFM for 10 kω The model for the measurement s: ε X = ε ( r + δrs + δrc ) T, X ref, A + ε ref, B + δε + δt, S + δ ε S where ε X s the relatve devaton from nomnal (RDN) of the unknown resstor, ε S s the RDN of the reference resstor, contanng the value obtaned from BIPM (type-a and type-b components) and the drft snce calbraton. The term δ T,S s the temperature correcton of the reference, r s the rato as measured, δr S s the specfcaton of the current comparator brdge, δr C s the correcton of the brdge error, and δ T,X s the temperature correcton of the unknown resstor. Quantty (unt) Dstrbuton x u(x ) n c u (y) 1 ε s, rel.dev. from nomnal of standard (10-6 ) Normal 0,668 0, ,0151 δ T,S, Temp devaton standard ±0,1 C (10-6 ) Rectangular 0 0,017 nfnty 1 0, r, Rato devaton as determned (10-6 ) Normal -7,778 0, , δr S, Specfcaton for Guldlne brdge (10-6 ) Rectangular 0 0,058 nfnty 1 0, δr C, Correcton of brdge error (10-6 ) Normal 0,141 0, ,010 y ε X, Rel. dev. From nom. of unknown (10-6 ) Normal -6,969 0,089 nfnty The expanded uncertanty s 180 nω/ω for k=. 3

24 It s further estmated that there s a correlaton of the DFM result for resstor SP91010 to the BIPM workng standards of r = 0,17 (estmated from the quoted type-b relatve uncertanty of BIPM of 1, ) 4

BUREAU INTERNATIONAL DES POIDS ET MESURES. Final report

BUREAU INTERNATIONAL DES POIDS ET MESURES. Final report BUREAU INTERNATIONAL DES POIDS ET MESURES Blateral Comparson of Ω standards (ongong BIPM key comparson BIPM.EM-K.a) between the NIST (USA) and the BIPM. R. Goebel *, R. Elmqst **, N. Fletcher * and M.

More information

Uncertainty in measurements of power and energy on power networks

Uncertainty in measurements of power and energy on power networks Uncertanty n measurements of power and energy on power networks E. Manov, N. Kolev Department of Measurement and Instrumentaton, Techncal Unversty Sofa, bul. Klment Ohrdsk No8, bl., 000 Sofa, Bulgara Tel./fax:

More information

CCEM KEY COMPARISON CCEM.RF-K1d.W (GT-RF/97-3) Measurement Techniques and Results of an Intercomparison. for RF Power in R 900

CCEM KEY COMPARISON CCEM.RF-K1d.W (GT-RF/97-3) Measurement Techniques and Results of an Intercomparison. for RF Power in R 900 CCEM KEY COMPARISON CCEM.RF-K1d.W (GT-RF/97-3) Measurement Technques and Results of an Intercomparson for RF Power n R 900 at frequences of 75 and 94 GHz. Fnal Report of the Plot Laboratory Joanna Beaumont

More information

MTBF PREDICTION REPORT

MTBF PREDICTION REPORT MTBF PREDICTION REPORT PRODUCT NAME: BLE112-A-V2 Issued date: 01-23-2015 Rev:1.0 Copyrght@2015 Bluegga Technologes. All rghts reserved. 1 MTBF PREDICTION REPORT... 1 PRODUCT NAME: BLE112-A-V2... 1 1.0

More information

Guidelines for CCPR and RMO Bilateral Key Comparisons CCPR Working Group on Key Comparison CCPR-G5 October 10 th, 2014

Guidelines for CCPR and RMO Bilateral Key Comparisons CCPR Working Group on Key Comparison CCPR-G5 October 10 th, 2014 Gudelnes for CCPR and RMO Blateral Key Comparsons CCPR Workng Group on Key Comparson CCPR-G5 October 10 th, 2014 These gudelnes are prepared by CCPR WG-KC and RMO P&R representatves, and approved by CCPR,

More information

Passive Filters. References: Barbow (pp ), Hayes & Horowitz (pp 32-60), Rizzoni (Chap. 6)

Passive Filters. References: Barbow (pp ), Hayes & Horowitz (pp 32-60), Rizzoni (Chap. 6) Passve Flters eferences: Barbow (pp 6575), Hayes & Horowtz (pp 360), zzon (Chap. 6) Frequencyselectve or flter crcuts pass to the output only those nput sgnals that are n a desred range of frequences (called

More information

NATIONAL RADIO ASTRONOMY OBSERVATORY Green Bank, West Virginia SPECTRAL PROCESSOR MEMO NO. 25. MEMORANDUM February 13, 1985

NATIONAL RADIO ASTRONOMY OBSERVATORY Green Bank, West Virginia SPECTRAL PROCESSOR MEMO NO. 25. MEMORANDUM February 13, 1985 NATONAL RADO ASTRONOMY OBSERVATORY Green Bank, West Vrgna SPECTRAL PROCESSOR MEMO NO. 25 MEMORANDUM February 13, 1985 To: Spectral Processor Group From: R. Fsher Subj: Some Experments wth an nteger FFT

More information

Sensors for Motion and Position Measurement

Sensors for Motion and Position Measurement Sensors for Moton and Poston Measurement Introducton An ntegrated manufacturng envronment conssts of 5 elements:- - Machne tools - Inspecton devces - Materal handlng devces - Packagng machnes - Area where

More information

POLYTECHNIC UNIVERSITY Electrical Engineering Department. EE SOPHOMORE LABORATORY Experiment 1 Laboratory Energy Sources

POLYTECHNIC UNIVERSITY Electrical Engineering Department. EE SOPHOMORE LABORATORY Experiment 1 Laboratory Energy Sources POLYTECHNIC UNIERSITY Electrcal Engneerng Department EE SOPHOMORE LABORATORY Experment 1 Laboratory Energy Sources Modfed for Physcs 18, Brooklyn College I. Oerew of the Experment Ths experment has three

More information

Calculation of the received voltage due to the radiation from multiple co-frequency sources

Calculation of the received voltage due to the radiation from multiple co-frequency sources Rec. ITU-R SM.1271-0 1 RECOMMENDATION ITU-R SM.1271-0 * EFFICIENT SPECTRUM UTILIZATION USING PROBABILISTIC METHODS Rec. ITU-R SM.1271 (1997) The ITU Radocommuncaton Assembly, consderng a) that communcatons

More information

To: Professor Avitabile Date: February 4, 2003 From: Mechanical Student Subject: Experiment #1 Numerical Methods Using Excel

To: Professor Avitabile Date: February 4, 2003 From: Mechanical Student Subject: Experiment #1 Numerical Methods Using Excel To: Professor Avtable Date: February 4, 3 From: Mechancal Student Subject:.3 Experment # Numercal Methods Usng Excel Introducton Mcrosoft Excel s a spreadsheet program that can be used for data analyss,

More information

Figure 1. DC-DC Boost Converter

Figure 1. DC-DC Boost Converter EE46, Power Electroncs, DC-DC Boost Converter Verson Oct. 3, 11 Overvew Boost converters make t possble to effcently convert a DC voltage from a lower level to a hgher level. Theory of Operaton Relaton

More information

DETERMINATION OF WIND SPEED PROFILE PARAMETERS IN THE SURFACE LAYER USING A MINI-SODAR

DETERMINATION OF WIND SPEED PROFILE PARAMETERS IN THE SURFACE LAYER USING A MINI-SODAR DETERMINATION OF WIND SPEED PROFILE PARAMETERS IN THE SURFACE LAYER USING A MINI-SODAR A. Coppalle, M. Talbaut and F. Corbn UMR 6614 CORIA, Sant Etenne du Rouvray, France INTRODUCTION Recent mprovements

More information

antenna antenna (4.139)

antenna antenna (4.139) .6.6 The Lmts of Usable Input Levels for LNAs The sgnal voltage level delvered to the nput of an LNA from the antenna may vary n a very wde nterval, from very weak sgnals comparable to the nose level,

More information

Walsh Function Based Synthesis Method of PWM Pattern for Full-Bridge Inverter

Walsh Function Based Synthesis Method of PWM Pattern for Full-Bridge Inverter Walsh Functon Based Synthess Method of PWM Pattern for Full-Brdge Inverter Sej Kondo and Krt Choesa Nagaoka Unversty of Technology 63-, Kamtomoka-cho, Nagaoka 9-, JAPAN Fax: +8-58-7-95, Phone: +8-58-7-957

More information

Strain Gauge Measuring Amplifier BA 660

Strain Gauge Measuring Amplifier BA 660 Stran Gauge Measurng Amplfer BA 660 Orgnal of the Manual BA660 / IP20 BA660 / IP66 Table of Contents 1. Safety precautons...2 1.1. Feld of applcaton...2 1.2. Installaton...2 1.3. Mantenance...2 2. Functon...2

More information

Application of Intelligent Voltage Control System to Korean Power Systems

Application of Intelligent Voltage Control System to Korean Power Systems Applcaton of Intellgent Voltage Control System to Korean Power Systems WonKun Yu a,1 and HeungJae Lee b, *,2 a Department of Power System, Seol Unversty, South Korea. b Department of Power System, Kwangwoon

More information

High Speed ADC Sampling Transients

High Speed ADC Sampling Transients Hgh Speed ADC Samplng Transents Doug Stuetzle Hgh speed analog to dgtal converters (ADCs) are, at the analog sgnal nterface, track and hold devces. As such, they nclude samplng capactors and samplng swtches.

More information

Figure 1. DC-DC Boost Converter

Figure 1. DC-DC Boost Converter EE36L, Power Electroncs, DC-DC Boost Converter Verson Feb. 8, 9 Overvew Boost converters make t possble to effcently convert a DC voltage from a lower level to a hgher level. Theory of Operaton Relaton

More information

Shunt Active Filters (SAF)

Shunt Active Filters (SAF) EN-TH05-/004 Martt Tuomanen (9) Shunt Actve Flters (SAF) Operaton prncple of a Shunt Actve Flter. Non-lnear loads lke Varable Speed Drves, Unnterrupted Power Supples and all knd of rectfers draw a non-snusodal

More information

Comparison of Two Measurement Devices I. Fundamental Ideas.

Comparison of Two Measurement Devices I. Fundamental Ideas. Comparson of Two Measurement Devces I. Fundamental Ideas. ASQ-RS Qualty Conference March 16, 005 Joseph G. Voelkel, COE, RIT Bruce Sskowsk Rechert, Inc. Topcs The Problem, Eample, Mathematcal Model One

More information

Learning Ensembles of Convolutional Neural Networks

Learning Ensembles of Convolutional Neural Networks Learnng Ensembles of Convolutonal Neural Networks Lran Chen The Unversty of Chcago Faculty Mentor: Greg Shakhnarovch Toyota Technologcal Insttute at Chcago 1 Introducton Convolutonal Neural Networks (CNN)

More information

AFV-P 2U/4U. AC + DC Power Solutions. series. Transient Generation for Disturbance Tests. only. High Performance Programmable AC Power Source

AFV-P 2U/4U. AC + DC Power Solutions. series. Transient Generation for Disturbance Tests. only. High Performance Programmable AC Power Source AFV-P seres Hgh Performance Programmable AC Power Source only 2U/4U Intutve Touch Screen HMI Output Frequency up to 15-1000Hz Power Lne Smulatons: Step & Ramp Features Fast Response Tme: 300μs AC Source

More information

Boise State University Department of Electrical and Computer Engineering ECE 212L Circuit Analysis and Design Lab

Boise State University Department of Electrical and Computer Engineering ECE 212L Circuit Analysis and Design Lab Bose State Unersty Department of Electrcal and Computer Engneerng ECE Crcut Analyss and Desgn ab Experment #6: Transformers, Mutual Inductance, and Coupled Cols Objectes The objectes of ths laboratory

More information

Impact of Interference Model on Capacity in CDMA Cellular Networks. Robert Akl, D.Sc. Asad Parvez University of North Texas

Impact of Interference Model on Capacity in CDMA Cellular Networks. Robert Akl, D.Sc. Asad Parvez University of North Texas Impact of Interference Model on Capacty n CDMA Cellular Networks Robert Akl, D.Sc. Asad Parvez Unversty of North Texas Outlne Introducton to CDMA networks Average nterference model Actual nterference model

More information

Customer witness testing guide

Customer witness testing guide Customer wtness testng gude Ths gude s amed at explanng why we need to wtness test equpment whch s beng connected to our network, what we actually do when we complete ths testng, and what you can do to

More information

Figure.1. Basic model of an impedance source converter JCHPS Special Issue 12: August Page 13

Figure.1. Basic model of an impedance source converter JCHPS Special Issue 12: August Page 13 A Hgh Gan DC - DC Converter wth Soft Swtchng and Power actor Correcton for Renewable Energy Applcaton T. Selvakumaran* and. Svachdambaranathan Department of EEE, Sathyabama Unversty, Chenna, Inda. *Correspondng

More information

Control Chart. Control Chart - history. Process in control. Developed in 1920 s. By Dr. Walter A. Shewhart

Control Chart. Control Chart - history. Process in control. Developed in 1920 s. By Dr. Walter A. Shewhart Control Chart - hstory Control Chart Developed n 920 s By Dr. Walter A. Shewhart 2 Process n control A phenomenon s sad to be controlled when, through the use of past experence, we can predct, at least

More information

Triferential Subtraction in Strain Gage Signal Conditioning. Introduction

Triferential Subtraction in Strain Gage Signal Conditioning. Introduction Trferental Subtracton n Stran Gage Sgnal Condtonng Karl F. Anderson Vald Measurements 3751 W. Ave. J-14 Lancaster, CA 93536 (661) 722-8255 http://www.vm-usa.com Introducton The general form of NASA's Anderson

More information

ECE315 / ECE515 Lecture 5 Date:

ECE315 / ECE515 Lecture 5 Date: Lecture 5 Date: 18.08.2016 Common Source Amplfer MOSFET Amplfer Dstorton Example 1 One Realstc CS Amplfer Crcut: C c1 : Couplng Capactor serves as perfect short crcut at all sgnal frequences whle blockng

More information

AN ALTERNATE CUT-OFF FREQUENCY FOR THE RESPONSE SPECTRUM METHOD OF SEISMIC ANALYSIS

AN ALTERNATE CUT-OFF FREQUENCY FOR THE RESPONSE SPECTRUM METHOD OF SEISMIC ANALYSIS ASIAN JOURNAL OF CIVIL ENGINEERING (BUILDING AND HOUSING) VOL. 11, NO. 3 (010) PAGES 31-334 AN ALTERNATE CUT-OFF FREQUENCY FOR THE RESPONSE SPECTRUM METHOD OF SEISMIC ANALYSIS M. Dhleep a*, N.P. Shahul

More information

Mesh Equations. Introduction

Mesh Equations. Introduction Mesh Equatons Introducton The crcuts n ths problem set consst of resstors and ndependent sources. We wll analyze these crcuts by wrtng and solvng a set of mesh equatons. To wrte mesh equatons, we. Express

More information

A Comparison of Two Equivalent Real Formulations for Complex-Valued Linear Systems Part 2: Results

A Comparison of Two Equivalent Real Formulations for Complex-Valued Linear Systems Part 2: Results AMERICAN JOURNAL OF UNDERGRADUATE RESEARCH VOL. 1 NO. () A Comparson of Two Equvalent Real Formulatons for Complex-Valued Lnear Systems Part : Results Abnta Munankarmy and Mchael A. Heroux Department of

More information

29. Network Functions for Circuits Containing Op Amps

29. Network Functions for Circuits Containing Op Amps 9. Network Functons for Crcuts Contanng Op Amps Introducton Each of the crcuts n ths problem set contans at least one op amp. Also each crcut s represented by a gven network functon. These problems can

More information

Unit 1. Current and Voltage U 1 VOLTAGE AND CURRENT. Circuit Basics KVL, KCL, Ohm's Law LED Outputs Buttons/Switch Inputs. Current / Voltage Analogy

Unit 1. Current and Voltage U 1 VOLTAGE AND CURRENT. Circuit Basics KVL, KCL, Ohm's Law LED Outputs Buttons/Switch Inputs. Current / Voltage Analogy ..2 nt Crcut Bascs KVL, KCL, Ohm's Law LED Outputs Buttons/Swtch Inputs VOLTAGE AND CRRENT..4 Current and Voltage Current / Voltage Analogy Charge s measured n unts of Coulombs Current Amount of charge

More information

PRACTICAL, COMPUTATION EFFICIENT HIGH-ORDER NEURAL NETWORK FOR ROTATION AND SHIFT INVARIANT PATTERN RECOGNITION. Evgeny Artyomov and Orly Yadid-Pecht

PRACTICAL, COMPUTATION EFFICIENT HIGH-ORDER NEURAL NETWORK FOR ROTATION AND SHIFT INVARIANT PATTERN RECOGNITION. Evgeny Artyomov and Orly Yadid-Pecht 68 Internatonal Journal "Informaton Theores & Applcatons" Vol.11 PRACTICAL, COMPUTATION EFFICIENT HIGH-ORDER NEURAL NETWORK FOR ROTATION AND SHIFT INVARIANT PATTERN RECOGNITION Evgeny Artyomov and Orly

More information

Dynamic Optimization. Assignment 1. Sasanka Nagavalli January 29, 2013 Robotics Institute Carnegie Mellon University

Dynamic Optimization. Assignment 1. Sasanka Nagavalli January 29, 2013 Robotics Institute Carnegie Mellon University Dynamc Optmzaton Assgnment 1 Sasanka Nagavall snagaval@andrew.cmu.edu 16-745 January 29, 213 Robotcs Insttute Carnege Mellon Unversty Table of Contents 1. Problem and Approach... 1 2. Optmzaton wthout

More information

Multichannel Frequency Comparator VCH-315. User Guide

Multichannel Frequency Comparator VCH-315. User Guide Multchannel Frequency Comparator VCH-315 User Gude Table of contents 1 Introducton... 3 2 The workng prncple of the Comparator... 6 3 The computed functons... 8 3.1 Basc ratos... 8 3.2 Statstcal functons...

More information

Performance Testing of the Rockwell PLGR+ 96 P/Y Code GPS receiver

Performance Testing of the Rockwell PLGR+ 96 P/Y Code GPS receiver Performance Testng of the Rockwell PLGR+ 96 P/Y Code GPS recever By Santago Mancebo and Ken Chamberlan Introducton: The Rockwell PLGR (Precson Lghtweght GPS Recever) + 96 s a Precse Postonng Servce P/Y

More information

Fiber length of pulp and paper by automated optical analyzer using polarized light (Five-year review of T 271 om-12) (no changes since Draft 1)

Fiber length of pulp and paper by automated optical analyzer using polarized light (Five-year review of T 271 om-12) (no changes since Draft 1) OTICE: Ths s a DRAFT of a TAPPI Standard n ballot. Although avalable for publc vewng, t s stll under TAPPI s copyrght and may not be reproduced or dstrbuted wthout permsson of TAPPI. Ths draft s OT a currently

More information

@IJMTER-2015, All rights Reserved 383

@IJMTER-2015, All rights Reserved 383 SIL of a Safety Fuzzy Logc Controller 1oo usng Fault Tree Analyss (FAT and realablty Block agram (RB r.-ing Mohammed Bsss 1, Fatma Ezzahra Nadr, Prof. Amam Benassa 3 1,,3 Faculty of Scence and Technology,

More information

ECE 2133 Electronic Circuits. Dept. of Electrical and Computer Engineering International Islamic University Malaysia

ECE 2133 Electronic Circuits. Dept. of Electrical and Computer Engineering International Islamic University Malaysia ECE 2133 Electronc Crcuts Dept. of Electrcal and Computer Engneerng Internatonal Islamc Unversty Malaysa Chapter 12 Feedback and Stablty Introducton to Feedback Introducton to Feedback 1-4 Harold Black,

More information

100 Watt DC-DC Converters Q Series

100 Watt DC-DC Converters Q Series Bengn Envronment DC-DC Converters 100 Watt Q Seres 100 Watt DC-DC Converters Q Seres Input to output solaton Sngle output: Seres 24Q...48Q1000 Double output: Seres 24Q...48Q2000 Very hgh effcency of up

More information

A Current Differential Line Protection Using a Synchronous Reference Frame Approach

A Current Differential Line Protection Using a Synchronous Reference Frame Approach A Current Dfferental Lne rotecton Usng a Synchronous Reference Frame Approach L. Sousa Martns *, Carlos Fortunato *, and V.Fernão res * * Escola Sup. Tecnologa Setúbal / Inst. oltécnco Setúbal, Setúbal,

More information

MASTER TIMING AND TOF MODULE-

MASTER TIMING AND TOF MODULE- MASTER TMNG AND TOF MODULE- G. Mazaher Stanford Lnear Accelerator Center, Stanford Unversty, Stanford, CA 9409 USA SLAC-PUB-66 November 99 (/E) Abstract n conjuncton wth the development of a Beam Sze Montor

More information

Air Exchange and Ventilation in an Underground Train Station

Air Exchange and Ventilation in an Underground Train Station Ar Echange and Ventlaton n an Underground Tran Staton Mkael Björlng 1* 1 Unversty of Gävle, Faculty of Technology and Envronment, Department of Buldngs, Energy, and Envronment, 1 76 Gävle * Correspondng

More information

Research of Dispatching Method in Elevator Group Control System Based on Fuzzy Neural Network. Yufeng Dai a, Yun Du b

Research of Dispatching Method in Elevator Group Control System Based on Fuzzy Neural Network. Yufeng Dai a, Yun Du b 2nd Internatonal Conference on Computer Engneerng, Informaton Scence & Applcaton Technology (ICCIA 207) Research of Dspatchng Method n Elevator Group Control System Based on Fuzzy Neural Network Yufeng

More information

Digital Transmission

Digital Transmission Dgtal Transmsson Most modern communcaton systems are dgtal, meanng that the transmtted normaton sgnal carres bts and symbols rather than an analog sgnal. The eect o C/N rato ncrease or decrease on dgtal

More information

Simulation of Distributed Power-Flow Controller (Dpfc)

Simulation of Distributed Power-Flow Controller (Dpfc) RESEARCH INVENTY: Internatonal Journal of Engneerng and Scence ISBN: 2319-6483, ISSN: 2278-4721, Vol. 2, Issue 1 (January 2013), PP 25-32 www.researchnventy.com Smulaton of Dstrbuted Power-Flow Controller

More information

Network Theory. EC / EE / IN. for

Network Theory.   EC / EE / IN. for Network Theory for / / IN By www.thegateacademy.com Syllabus Syllabus for Networks Network Graphs: Matrces Assocated Wth Graphs: Incdence, Fundamental ut Set and Fundamental rcut Matrces. Soluton Methods:

More information

SMD Common Mode Filter for Signal Line. PRODUCT IDENTIFICATION 2012 F P - E under development. (1) (2) (3) (4) (5) (6) APPLICATIONS

SMD Common Mode Filter for Signal Line. PRODUCT IDENTIFICATION 2012 F P - E under development. (1) (2) (3) (4) (5) (6) APPLICATIONS Motocraft SMD Common Mode Flter for Sgnal Lne CMF2012F Seres (Sheldng Type) FEATURES Ths common mode flter s characterzed by ts small szed Hghly effectve n nose suppresson,hght common mode mpedance at

More information

RC Filters TEP Related Topics Principle Equipment

RC Filters TEP Related Topics Principle Equipment RC Flters TEP Related Topcs Hgh-pass, low-pass, Wen-Robnson brdge, parallel-t flters, dfferentatng network, ntegratng network, step response, square wave, transfer functon. Prncple Resstor-Capactor (RC)

More information

Optimization Frequency Design of Eddy Current Testing

Optimization Frequency Design of Eddy Current Testing Optmzaton Frequency Desgn of Eddy Current Testng NAONG MUNGKUNG 1, KOMKIT CHOMSUWAN 1, NAONG PIMPU 2 AND TOSHIFUMI YUJI 3 1 Department of Electrcal Technology Educaton Kng Mongkut s Unversty of Technology

More information

SMALL POWER GENERATOR FOR WIND POWER PLANTS

SMALL POWER GENERATOR FOR WIND POWER PLANTS Zeszyty Problemowe Maszyny Elektryczne Nr 8/29 9 Karel hmelík, Stanslav Mšák Techncal Unversty of Ostrava, Ostrava SMALL POWER GENERATOR FOR WND POWER PLANTS GENERATORY MAŁEJ MOY DO ELEKTROWN WATROWYH

More information

Webinar Series TMIP VISION

Webinar Series TMIP VISION Webnar Seres TMIP VISION TMIP provdes techncal support and promotes knowledge and nformaton exchange n the transportaton plannng and modelng communty. DISCLAIMER The vews and opnons expressed durng ths

More information

THE ARCHITECTURE OF THE BROADBAND AMPLIFIERS WITHOUT CLASSICAL STAGES WITH A COMMON BASE AND A COMMON EMITTER

THE ARCHITECTURE OF THE BROADBAND AMPLIFIERS WITHOUT CLASSICAL STAGES WITH A COMMON BASE AND A COMMON EMITTER VOL. 0, NO. 8, OCTOBE 205 ISSN 89-6608 2006-205 Asan esearch Publshng Network (APN. All rghts reserved. THE ACHITECTUE OF THE BOADBAND AMPLIFIES WITHOUT CLASSICAL STAGES WITH A COMMON BASE AND A COMMON

More information

N- and P-Channel 2.5-V (G-S) MOSFET

N- and P-Channel 2.5-V (G-S) MOSFET S456DY N- and P-Channel.5-V (G-S) MOSFET PRODUCT SUMMARY V DS (V) R DS(on) (Ω) (A).5 at 7. N-Channel.35 at V GS =.5 V 6. FEATURES Halogen-free Accordng to IEC 649-- Defnton TrenchFET Power MOSFET:.5 Rated

More information

Graph Method for Solving Switched Capacitors Circuits

Graph Method for Solving Switched Capacitors Circuits Recent Advances n rcuts, ystems, gnal and Telecommuncatons Graph Method for olvng wtched apactors rcuts BHUMIL BRTNÍ Department of lectroncs and Informatcs ollege of Polytechncs Jhlava Tolstého 6, 586

More information

TECHNICAL NOTE TERMINATION FOR POINT- TO-POINT SYSTEMS TN TERMINATON FOR POINT-TO-POINT SYSTEMS. Zo = L C. ω - angular frequency = 2πf

TECHNICAL NOTE TERMINATION FOR POINT- TO-POINT SYSTEMS TN TERMINATON FOR POINT-TO-POINT SYSTEMS. Zo = L C. ω - angular frequency = 2πf TECHNICAL NOTE TERMINATION FOR POINT- TO-POINT SYSTEMS INTRODUCTION Because dgtal sgnal rates n computng systems are ncreasng at an astonshng rate, sgnal ntegrty ssues have become far more mportant to

More information

PRO- CRIMPER III Hand Crimping Tool Assembly DESCRIPTION (Figures 1 and 2)

PRO- CRIMPER III Hand Crimping Tool Assembly DESCRIPTION (Figures 1 and 2) PRO- CRIMPER* III Hand Crmpng Instructon Sheet Tool Assembly 58495-1 408-9819 Wth De Assembly 58495-2 22 JUL 09 PROPER USE GUIDELINES Cumulatve Trauma Dsorders can result from the prolonged use of manually

More information

onlinecomponents.com

onlinecomponents.com PRO- CRIMPER* III Hand Crmpng Instructon Sheet Tool Assembly 58535-1 wth 408-4021 De Assembly 58535-2 29 JUL 09 PROPER USE GUIDELINES Cumulatve Trauma Dsorders can result from the prolonged use of manually

More information

1. REVIEW 2. DELIVERY SET

1. REVIEW 2. DELIVERY SET power status ON steps DIP test reset ON DIP off modes lght sensor / led sensor speed on sensor 2 speed off on/off delay pause to off REVIEW LED lghtng devce s desgned for automatc lghtng of starway steps

More information

VRT014 User s guide V0.8. Address: Saltoniškių g. 10c, Vilnius LT-08105, Phone: (370-5) , Fax: (370-5) ,

VRT014 User s guide V0.8. Address: Saltoniškių g. 10c, Vilnius LT-08105, Phone: (370-5) , Fax: (370-5) , VRT014 User s gude V0.8 Thank you for purchasng our product. We hope ths user-frendly devce wll be helpful n realsng your deas and brngng comfort to your lfe. Please take few mnutes to read ths manual

More information

A High-Sensitivity Oversampling Digital Signal Detection Technique for CMOS Image Sensors Using Non-destructive Intermediate High-Speed Readout Mode

A High-Sensitivity Oversampling Digital Signal Detection Technique for CMOS Image Sensors Using Non-destructive Intermediate High-Speed Readout Mode A Hgh-Senstvty Oversamplng Dgtal Sgnal Detecton Technque for CMOS Image Sensors Usng Non-destructve Intermedate Hgh-Speed Readout Mode Shoj Kawahto*, Nobuhro Kawa** and Yoshak Tadokoro** *Research Insttute

More information

EE301 AC Source Transformation and Nodal Analysis

EE301 AC Source Transformation and Nodal Analysis EE0 AC Source Transformaton and Nodal Analyss Learnng Ojectves. Construct equvalent crcuts y convertng an AC voltage source and a resstor to an AC current source and a resstor. Apply Nodal Analyss to an

More information

Control of Venturini Method Based Matrix Converter in Input Voltage Variations

Control of Venturini Method Based Matrix Converter in Input Voltage Variations IMECS 9, March 8 -, 9, Hong Kong Control of Venturn Method Based Matrx Converter n Input Voltage Varatons Hulus Karaca, Ramazan Akkaya Abstract Matrx converter s a sngle-stage converter whch drectly connects

More information

THE GENERATION OF 400 MW RF PULSES AT X-BAND USING RESONANT DELAY LINES *

THE GENERATION OF 400 MW RF PULSES AT X-BAND USING RESONANT DELAY LINES * SLAC PUB 874 3/1999 THE GENERATION OF 4 MW RF PULSES AT X-BAND USING RESONANT DELAY LINES * Sam G. Tantaw, Arnold E. Vleks, and Rod J. Loewen Stanford Lnear Accelerator Center, Stanford Unversty P.O. Box

More information

A Facts Device: Distributed Power-Flow Controller (DPFC)

A Facts Device: Distributed Power-Flow Controller (DPFC) A Facts Devce: Dstrbuted Power-Flow Controller (DPFC) Guda Pryanka 1, K. Jaghannath 2, D. Kumara Swamy 3 1, 2, 3 Dept. of EEE, SVS Insttute of Technology, Hanamkonda, T.S, Inda Emal address: ujjwalak32@gmal.com

More information

Evaluate the Effective of Annular Aperture on the OTF for Fractal Optical Modulator

Evaluate the Effective of Annular Aperture on the OTF for Fractal Optical Modulator Global Advanced Research Journal of Management and Busness Studes (ISSN: 2315-5086) Vol. 4(3) pp. 082-086, March, 2015 Avalable onlne http://garj.org/garjmbs/ndex.htm Copyrght 2015 Global Advanced Research

More information

Voltage Quality Enhancement and Fault Current Limiting with Z-Source based Series Active Filter

Voltage Quality Enhancement and Fault Current Limiting with Z-Source based Series Active Filter Research Journal of Appled Scences, Engneerng and echnology 3(): 246-252, 20 ISSN: 2040-7467 Maxwell Scentfc Organzaton, 20 Submtted: July 26, 20 Accepted: September 09, 20 Publshed: November 25, 20 oltage

More information

The Dynamic Utilization of Substation Measurements to Maintain Power System Observability

The Dynamic Utilization of Substation Measurements to Maintain Power System Observability 1 The Dynamc Utlzaton of Substaton Measurements to Mantan Power System Observablty Y. Wu, Student Member, IEEE, M. Kezunovc, Fellow, IEEE and T. Kostc, Member, IEEE Abstract-- In a power system State Estmator

More information

HUAWEI TECHNOLOGIES CO., LTD. Huawei Proprietary Page 1

HUAWEI TECHNOLOGIES CO., LTD. Huawei Proprietary Page 1 Project Ttle Date Submtted IEEE 802.16 Broadband Wreless Access Workng Group Double-Stage DL MU-MIMO Scheme 2008-05-05 Source(s) Yang Tang, Young Hoon Kwon, Yajun Kou, Shahab Sanaye,

More information

Rejection of PSK Interference in DS-SS/PSK System Using Adaptive Transversal Filter with Conditional Response Recalculation

Rejection of PSK Interference in DS-SS/PSK System Using Adaptive Transversal Filter with Conditional Response Recalculation SERBIAN JOURNAL OF ELECTRICAL ENGINEERING Vol., No., November 23, 3-9 Rejecton of PSK Interference n DS-SS/PSK System Usng Adaptve Transversal Flter wth Condtonal Response Recalculaton Zorca Nkolć, Bojan

More information

PRO- CRIMPER* III Hand Crimping

PRO- CRIMPER* III Hand Crimping PRO- CRIMPER* III Hand Crmpng Instructon Sheet Tool Assembly 91338-1 408-8377 wth De Assembly 91338-2 22 JUL 09 PROPER USE GUIDELINES Cumulatve Trauma Dsorders can result from the prolonged use of manually

More information

PRO- CRIMPER* III Hand

PRO- CRIMPER* III Hand PRO- CRIMPER* III Hand Instructon Sheet Crmpng Tool Assembly 58529-1 408-9999 wth De Assembly 58529-2 11 AUG 14 PROPER USE GUIDELINES Cumulatve Trauma Dsorders can result from the prolonged use of manually

More information

THEORY OF YARN STRUCTURE by Prof. Bohuslav Neckář, Textile Department, IIT Delhi, New Delhi. Compression of fibrous assemblies

THEORY OF YARN STRUCTURE by Prof. Bohuslav Neckář, Textile Department, IIT Delhi, New Delhi. Compression of fibrous assemblies THEORY OF YARN STRUCTURE by Prof. Bohuslav Neckář, Textle Department, IIT Delh, New Delh. Compresson of fbrous assembles Q1) What was the dea of fbre-to-fbre contact accordng to van Wyk? A1) Accordng to

More information

High Gain Soft-switching Bidirectional DC-DC Converters for Eco-friendly Vehicles

High Gain Soft-switching Bidirectional DC-DC Converters for Eco-friendly Vehicles Hgh Gan Soft-swtchng Bdrectonal DC-DC Converters for Eco-frendly ehcles Mnho Kwon, Junsung Park and Sewan Cho, EEE Senor Member Department of Electrcal and nformaton Engneerng Seoul Natonal Unversty of

More information

PRO- CRIMPER* III Hand Crimping

PRO- CRIMPER* III Hand Crimping PRO- CRIMPER* III Hand Crmpng Instructon Sheet Tool Assembly 58641-1 wth 408-4379 De Assembly 58641-2 18 JUN 09 PROPER USE GUIDELINES Cumulatve Trauma Dsorders can result from the prolonged use of manually

More information

ABSTRACT 1. INTRODUCTION 2. I-V CURVE MEASUREMENTS. Rodrigo Moretón, Eduardo Lorenzo and Javier Muñoz

ABSTRACT 1. INTRODUCTION 2. I-V CURVE MEASUREMENTS. Rodrigo Moretón, Eduardo Lorenzo and Javier Muñoz A 500kW PV GENERATOR Rodrgo Moretón, Eduardo Lorenzo and Javer Muñoz IVCURVE ABSTRACT Ths paper presents the measurement of the IV curve of a 500kW PV generator by means of an ownmade capactve load. It

More information

Design of Shunt Active Filter for Harmonic Compensation in a 3 Phase 3 Wire Distribution Network

Design of Shunt Active Filter for Harmonic Compensation in a 3 Phase 3 Wire Distribution Network Internatonal Journal of Research n Electrcal & Electroncs Engneerng olume 1, Issue 1, July-September, 2013, pp. 85-92, IASTER 2013 www.aster.com, Onlne: 2347-5439, Prnt: 2348-0025 Desgn of Shunt Actve

More information

A Novel Soft-Switching Two-Switch Flyback Converter with a Wide Operating Range and Regenerative Clamping

A Novel Soft-Switching Two-Switch Flyback Converter with a Wide Operating Range and Regenerative Clamping 77 Journal of ower Electroncs, ol 9, No 5, September 009 JE 9-5- A Novel Soft-Swtchng Two-Swtch Flybac Converter wth a Wde Operatng Range and Regeneratve Clampng Marn-Go Km and Young-Seo Jung * Dvson of

More information

Development of a High Bandwidth, High Power Linear Amplifier for a Precision Fast Tool Servo System

Development of a High Bandwidth, High Power Linear Amplifier for a Precision Fast Tool Servo System Development of a Hgh Bandwdth, Hgh Power near Amplfer for a Precson Fast Tool Servo System S. Rakuff 1, J. Cuttno 1, D. Schnstock 2 1 Dept. of Mechancal Engneerng, The Unversty of North Carolna at Charlotte,

More information

Beam quality measurements with Shack-Hartmann wavefront sensor and M2-sensor: comparison of two methods

Beam quality measurements with Shack-Hartmann wavefront sensor and M2-sensor: comparison of two methods Beam qualty measurements wth Shack-Hartmann wavefront sensor and M-sensor: comparson of two methods J.V.Sheldakova, A.V.Kudryashov, V.Y.Zavalova, T.Y.Cherezova* Moscow State Open Unversty, Adaptve Optcs

More information

Q Series Watt DC-DC Converters

Q Series Watt DC-DC Converters 66 - Watt DC-DC Converters Features RoHS lead-free-solder and lead-solder-exempted products are avalable 5 year warranty for RoHS complant products wth an extended temperature range Class I equpment Complant

More information

aperture David Makovoz, 30/01/2006 Version 1.0 Table of Contents

aperture David Makovoz, 30/01/2006 Version 1.0 Table of Contents aperture 1 aperture Davd Makovoz, 30/01/2006 Verson 1.0 Table of Contents aperture... 1 1 Overvew... 2 1.1 Input Image Requrements... 2 2 aperture... 2 2.1 Input... 2 2.2 Processng... 4 2.3 Output Table...

More information

PRO- CRIMPER III Hand Crimping Tool Assembly INSTALLATION AND REMOVAL OF DIE SET AND LOCATOR ASSEMBLY (Figure 2)

PRO- CRIMPER III Hand Crimping Tool Assembly INSTALLATION AND REMOVAL OF DIE SET AND LOCATOR ASSEMBLY (Figure 2) PRO- CRIMPER* III Hand Crmpng Instructon Sheet Tool Assembly 90547-1 wth 408-9884 De Assembly 90547-2 02 NOV 09 PROPER USE GUIDELINES Cumulatve Trauma Dsorders can result from the prolonged use of manually

More information

1 GSW Multipath Channel Models

1 GSW Multipath Channel Models In the general case, the moble rado channel s pretty unpleasant: there are a lot of echoes dstortng the receved sgnal, and the mpulse response keeps changng. Fortunately, there are some smplfyng assumptons

More information

A GBAS Testbed to Support New Monitoring Algorithms Development for CAT III Precision Approach

A GBAS Testbed to Support New Monitoring Algorithms Development for CAT III Precision Approach A GBAS Testbed to Support New Montorng Algorthms Development for CAT III Precson Approach B. Belabbas, T. Dautermann, M. Felux, M. Rppl, S. Schlüter, V. Wlken, A. Hornbostel, M. Meurer German Aerospace

More information

Time-frequency Analysis Based State Diagnosis of Transformers Windings under the Short-Circuit Shock

Time-frequency Analysis Based State Diagnosis of Transformers Windings under the Short-Circuit Shock Tme-frequency Analyss Based State Dagnoss of Transformers Wndngs under the Short-Crcut Shock YUYING SHAO, ZHUSHI RAO School of Mechancal Engneerng ZHIJIAN JIN Hgh Voltage Lab Shangha Jao Tong Unversty

More information

Locator Pin Indexing Pin. Wire Size Marking CAUTION NOTE TOOLING ASSISTANCE CENTER PRODUCT INFORMATION

Locator Pin Indexing Pin. Wire Size Marking CAUTION NOTE TOOLING ASSISTANCE CENTER PRODUCT INFORMATION ROTA-CRIMP* Crmpng Tool Instructon Sheet 408-2681 68321-1 13 APR 12 Statonary De (Nest) Anvl De Hold-Down Devce and Contact Locator Front of Tool Locator Pn Indexng Pn Wre Sze Markng The hold-down devce

More information

ANNUAL OF NAVIGATION 11/2006

ANNUAL OF NAVIGATION 11/2006 ANNUAL OF NAVIGATION 11/2006 TOMASZ PRACZYK Naval Unversty of Gdyna A FEEDFORWARD LINEAR NEURAL NETWORK WITH HEBBA SELFORGANIZATION IN RADAR IMAGE COMPRESSION ABSTRACT The artcle presents the applcaton

More information

A study of turbo codes for multilevel modulations in Gaussian and mobile channels

A study of turbo codes for multilevel modulations in Gaussian and mobile channels A study of turbo codes for multlevel modulatons n Gaussan and moble channels Lamne Sylla and Paul Forter (sylla, forter)@gel.ulaval.ca Department of Electrcal and Computer Engneerng Laval Unversty, Ste-Foy,

More information

Instruction Sheet AMPMODU* MTE CONNECTORS Mar 11 Rev A

Instruction Sheet AMPMODU* MTE CONNECTORS Mar 11 Rev A Instructon Sheet AMPMODU* MTE CONNECTORS 408-6919 10 Mar 11 PROPER USE GUIDELINES Cumulatve Trauma Dsorders can result from the prolonged use of manually powered hand tools. Hand tools are ntended for

More information

PRO- CRIMPER* III Hand

PRO- CRIMPER* III Hand PRO- CRIMPER* III Hand Instructon Sheet Crmpng Tool Assembly 90759-1 408-9962 wth De Assembly 90759-2 03 MAY 11 PROPER USE GUIDELINES Cumulatve Trauma Dsorders can result from the prolonged use of manually

More information

UNIT 11 TWO-PERSON ZERO-SUM GAMES WITH SADDLE POINT

UNIT 11 TWO-PERSON ZERO-SUM GAMES WITH SADDLE POINT UNIT TWO-PERSON ZERO-SUM GAMES WITH SADDLE POINT Structure. Introducton Obectves. Key Terms Used n Game Theory.3 The Maxmn-Mnmax Prncple.4 Summary.5 Solutons/Answers. INTRODUCTION In Game Theory, the word

More information

Research on Peak-detection Algorithm for High-precision Demodulation System of Fiber Bragg Grating

Research on Peak-detection Algorithm for High-precision Demodulation System of Fiber Bragg Grating , pp. 337-344 http://dx.do.org/10.1457/jht.014.7.6.9 Research on Peak-detecton Algorthm for Hgh-precson Demodulaton System of Fber ragg Gratng Peng Wang 1, *, Xu Han 1, Smn Guan 1, Hong Zhao and Mngle

More information

Particle Filters. Ioannis Rekleitis

Particle Filters. Ioannis Rekleitis Partcle Flters Ioanns Reklets Bayesan Flter Estmate state x from data Z What s the probablty of the robot beng at x? x could be robot locaton, map nformaton, locatons of targets, etc Z could be sensor

More information

Available online at ScienceDirect. Procedia Engineering 202 (2017)

Available online at   ScienceDirect. Procedia Engineering 202 (2017) Avalable onlne at www.scencedrect.com ScenceDrect Proceda Engneerng (17) 159 167 4th Internatonal Colloquum "Transformer Research and Asset Management Frequency response and harmonc dstorton testng of

More information

PRO-CRIMPER* III Hand Crimping Tool Assembly with Die AssemblY

PRO-CRIMPER* III Hand Crimping Tool Assembly with Die AssemblY PRO-CRIMPER* III Hand Crmpng Tool Assembly 90758-1 wth De AssemblY 90758-2 Instructon Sheet 408-9938 01 NOV 11 PROPER USE GUIDELINES Cumulatve Trauma Dsorders can result from the prolonged use of manually

More information