Nivotester FTR325. Technical Information. 1-channel switch amplifier for the microwave barrier Soliwave FDR50/FQR50
|
|
- Laura Glenn
- 6 years ago
- Views:
Transcription
1 Technical Information Nivotester 1-channel switch amplifier for the microwave barrier Soliwave FDR50/FQR50 Area of application The Nivotester is suitable as a 1-channel switch amplifier for the Soliwave FDR50/FQR50 microwave barrier. You can use the for level detection for any type of solids or for control and counting tasks for packaged goods. Associated apparatus for the microwave barrier with intrinsically safe circuits. Advantages at a glance High level of operational safety through line monitoring up to the sensor LED display of operating and switching state Selectable pick up and/or drop out delay of the switching output Compact housing for easy inline mounting on standard rack rails Easy wiring via pluggable terminal block connectors TI00377F/97/EN/
2 Table of contents Function and system design... 3 Measuring principle... 3 Measuring system... 3 Limit switch signal function... 3 Function monitoring system... 3 Device combinations... 4 Input... 4 Measured variable... 4 Input signal... 4 Output... 4 Output signal... 4 Switching frequency... 4 Power supply... 5 Electrical connection... 5 Ring wiring... 5 Star wiring... 5 Supply voltage... 6 Cable specification... 6 Power consumption... 6 Installation... 7 Installation instructions... 7 Clearances... 7 Human interface Control concept Display elements Operating elements Certificates and approvals CE approval Ex approval Other standards and guidelines Ordering information Ordering information Nivotester Notes on product structure Safety instructions Zone classification Safety notes for electrical equipment for potentially explosive atmospheres Accessories Protective housing Documentation Operating instructions (KA) Tecnical informations Safety instructions Environment... 8 Ambient temperature range... 8 Storage temperature... 8 Degree of protection... 8 Electromagnetic compatibility (EMC)... 8 Mechanical construction... 9 Design, dimensions... 9 Weight... 9 Material... 9 Housing... 9 Terminals
3 Function and system design Measuring principle The Nivotester monitors the Soliwave FDR50/FQR50 microwave barrier s open collector output and uses a relay (change-over gate) to provide this output for many user applications. An LED displays the switching state of the relay. Problems such as a cable breakage are also displayed. The input on the Nivotester is galvanically isolated from main power and output. It is connected to the microwave barrier via a three-wire cable (ring or star wiring), whereupon the Nivotester provides distribution (intrinsically safe circuit, DC) voltage to the microwave barrier. Measuring system Emitter Receiver Nivotester The complete measuring system for limit detection consists of: an emitter FQR50, a receiver FDR50 and an evaluator Nivotester Optical or acoustic signallers, contactors, relays, solenoids etc. may be connected to the Nivotester. Limit switch signal function The appropriate setting for the limit switch signal function (see page 9) ensures that the relay for signal output and fault is always activated (fail-safe mode) in idle state. In combination with the Soliwave FDR50/FQR50 microwave barrier the following secure modes are possible: The relay drops out when the switch point is reached (microwave barrier interrupted), a fault occurs or the distribution voltage fails. The relay drops out when the switch point is underrun (microwave barrier uninterrupted), a fault occurs or the distribution voltage fails. Function monitoring system The Nivotester is equipped with a function monitoring system for improved fail-safe operation. A fault is indicated by an LED, the fault relay falls off. In combination with the microwave barrier Soliwave FQR50/FDR50-C* and -D* resp. the Nivotester -B* and -D* resp. can recognise the following faults and use an LED to indicate them: Breakage in the power supply cable to the emitter FQR50 Microwave barrier power input (emitter FQR50 and/or receiver FDR50) too high Short-circuit (emitter FQR50, receiver FDR50 or wiring) Manual function monitoring can be carried out by pushing the test button. 3
4 Device combinations The following equipment combinations are possible: Nivotester -B* (ATEX II (1)G [Ex ia Ga] IIC and ATEX II (1)D [Ex ia Da] IIIC resp.) with FQR50/ FDR50-C* (ATEX II 1/2G Ex ia IIC T4 Ga/Gb and ATEX II 1/2D Ex ia IIIC T98 C Da/Db resp.) Nivotester -D* (IECEx [Ex ia Ga] IIC and IECEx [Ex ia Da] IIIC resp.) with FQR50/FDR50-D* (IECEx Ex ia IIC T4 Ga/Gb and IECEx Ex ia IIIC T98 C Da/Db resp.) The following equipment combinations are impossible: Nivotester -A* (non hazardous area) with FQR50/FDR50-C* (ATEX II 1/2G and ATEX II 1/2D resp.) Nivotester -A* (non hazardous area) with FQR50/FDR50-D* (IECEx) Nivotester -B* (ATEX II (1)G and ATEX II (1)D resp.) with FQR50/FDR50-A* (non hazardous area) Nivotester -B* (ATEX II (1)G and ATEX II (1)D resp.) with FQR50/FDR50-B*(ATEX II 1/2D) Nivotester -B* (ATEX II (1)G and ATEX II (1)D resp.) with FQR50/FDR50-D* (IECEx) Nivotester -D* (IECEx) with FQR50/FDR50-A*(non hazardous area) Nivotester -D* (IECEx) with FQR50/FDR50-B* (ATEX II 1/2D) Nivotester -D* (IECEx) with FQR50/FDR50-C* (ATEX II 1/2G and ATEX II 1/2D resp.) Input Measured variable Input signal A limit signal is generated when the microwave barrier is either interrupted or uninterrupted, depending on the type of limit signal selected. input: - Soliwave FDR50/FQR50 microwave barrier - input galvanically isolated from power supply - intrinsically safe circuit Output Output signal Relay output: - galvanically isolated from power supply - a potential-free switching contact for the limit signal - a potential-free switching contact for fault notification Breaking capacity for relay contacts: U~ max. 253 VAC I~ max. 2 A (AC) P~ max. 500 VA (cos φ 0.7) U max. 40 VDC I max. 2 A (DC) P max. 80 W Service life: min switching operations at max. contact load Function indicators: LEDs for - On (green) - Limit signal (yellow) - Fault (red) Switching frequency max. 2 Hz 4
5 Power supply Electrical connection Ring wiring When connected to the Nivotester a microwave barrier can be wired via a choice of ring or star connection. The required backup power is provided by the. Emitter FQR50 Receiver FDR Contact circuit (terminal 3, 4, 5) and alarm output (terminal 15, 16) resp. Power supply (terminal 1, 2) Star wiring Emitter FQR50 Receiver FDR Contact circuit (terminal 3, 4, 5) and alarm output (terminal 15, 16) resp. Power supply (terminal 1, 2) 5
6 Supply voltage AC voltage model - Voltage range: VAC, 50/60 Hz DC voltage model - Voltage range: VAC / VDC - Reverse voltage protection Internally there is a micro-fuse built into the supply voltage circuit so that only one fuse for short circuits in the feed line to the must be upstream. Cable specification Power consumption -B* with FQR50/FDR50-C* and -D* with FQR50/FDR50-D* resp.: Off-the-shelf installation cable, 3-conductor, non-shielded Resistance: 15 Ω/km R' 150 Ω/km Inductance: 0.4 mh/km L' 1 mh/km Capacitance: 45 nf/km C' 200 nf/km Length of spurs max m (IIC) and 5000 m (IIB) respectively AC voltage model: 6,9 VA DC voltage model: 1,85 W 6
7 Installation Installation instructions The Nivotester is mounted upright on a DIN rail (TS 35 as per EN 50022). A protective case (IP 66, see accessories) is available for external installation. min. IP mm = 3.94 in Din rail EN x7.5 / EN x15 Clearances -20 C C / -4 F F -20 C C / -4 F F -20 C C / -4 F F 7
8 Environment Ambient temperature range - 20 C C (if installed individually) - 20 C C (if installed in-line without clearance spaces) - 20 C C (if installed in protective case) Storage temperature Degree of protection IP 20 Electromagnetic compatibility (EMC) Note! If possible mount devices so that they are not subjected to direct sunlight. This applies especially to use in warmer climates C C (preferably at + 20 C) Interference Emission to EN 61326, Electrical Equipment Class B Interference Immunity to EN 61326, Appendix A (Industrial) 8
9 Mechanical construction Design, dimensions DIN rail EN x7.5 / EN x15 Weight 0.3 kg Material Housing: - Polycarbonat, Polyamid PA6 - Color (main part): Light grey, RAL 7035 Housing Housing: In-line case (Minipac shape) made of plastic Terminals X2 Refer to the nameplate! X1 N L1 L- N L+ L1 F X X2 max. 1.5 mm 2 (max. AWG 16) The four plug terminals are a part of the device. 9
10 Human interface Control concept Display elements Operating elements Onsite adjustments using switches and potentiometers behind the hinged face plate. Green LED: operational state Yellow LED: switching state Red LED: fault Limit switch signal function Test button Fault notification LED row YE - yellow (switch state) RD - red (fault) GN - green (operation) Limit switch signal function: - Switches the limit switch relay on or off when the microwave barrier is either interrupted or uninterrupted, depending on the type of limit signal selected. - Optical indication of switching state via yellow LED Fault notification: - When activated, switches on the fault relay when a fault occurs - Optical indication of a fault via red LED Test button: - Simulation of a fault when fault notification switched on - Simulation of a limit signal when fault notification is switched off (for uninterrupted microwave barrier) 10
11 Switch delay (pick up delay) t(t) = s m T T t(s) = s n S S Switch delay (drop out delay) Switch delay (pick up delay): - When switch delay activated limit switch relay switches on pick up delay after time t(t) - Adjustable time: t(t) = s Switch delay (drop out delay): - When switch delay activated limit switch relay switches off drop out delay after time t(s) - Adjustable time: t(s) = s Note: The switching frequency (maximum 2 Hz) is given by the microwave barrier Soliwave FDR50/FQR50. You can find detailed information on the different setting options in the KA00205F/97/A6 operating instructions. 11
12 Certificates and approvals CE approval Ex approval Other standards and guidelines The Nivotester fulfils the legal requirements of the EC directives. The manufacturer confirms the successful examination of the equipment by using the CE mark. See "Ordering information" EN Degrees of protection through housing (IP code) EN Safety requirements for electrical equipment for measurement, control and laboratory use EN X EMC product family standard for electrical equipment for measurement, control and laboratory use Ordering information Ordering information Nivotester 10 Approval: B ATEX II (1)G [Ex ia Ga] IIC ATEX II (1)D [Ex ia Da] IIIC D IECEx [Ex ia Ga] IIC IECEx [Ex ia Da] IIIC Y Special version, to be specified 20 Model: 1 DIN rail mounting, b = 45 mm 9 Special version, to be specified 30 Power supply: A VAC, 50/60 Hz E VDC VAC, 50/60 Hz Y Special version, to be specified 40 Output: 1 1x level SPDT + 1x alarm/level SPST 9 Special version, to be specified - Notes on product structure The following equipment combinations are possible: Nivotester -B* (ATEX II (1)G [Ex ia Ga] IIC and ATEX II (1)D [Ex ia Da] IIIC resp.) with FQR50/ FDR50-C* (ATEX II 1/2G Ex ia IIC T4 Ga/Gb and ATEX II 1/2D Ex ia IIIC T98 C Da/Db resp.) Nivotester -D* (IECEx [Ex ia Ga] IIC and IECEx [Ex ia Da] IIIC resp.) with FQR50/FDR50-D* (IECEx Ex ia IIC T4 Ga/Gb and IECEx Ex ia IIIC T98 C Da/Db resp.) 12
13 Safety instructions Zone classification Emitter FQR50-C* FQR50-D* Zone 1 / Zone 21 Zone 0 / Zone 20 Receiver FDR50-C* FDR50-D* Nivotester -B* -D* Safety notes for electrical equipment for potentially explosive atmospheres The following equipment combinations are allowed to operate in explosive area: - -B* with FQR50/FDR50-C* (see Safety Instructions XA00219F/97/A3) - -D* with FQR50/FDR50-D* (see Safety Instructions XA00484F/97/EN) Install the device according to the manufacturer's instructions and any other valid standards and regulations (e.g. EN/IEC ). Incorrect use of the device poses a danger. The device may be installed, hooked up and put into operation only by qualified and authorized persons with special attention paid to - the operating manual, - the relevant standards and - legal rules and regulations. The Nivotester can and may be serviced by the manufacturer only. 13
14 Accessories Protective housing A protective housing for field installation of top-hat rail devices e. g. Nivotester is available. In this housing there is space for several top-hat rail devices (to a maximum installation width of 145 mm). Technical data - Order code: Protection type (according to EN 60529): IP 66 - Lower housing section: fibre-glass reinforced polycarbonate, grey - Upper housing section: polycarbonate, transparent - Seal: PU seal - Top-hat rail (EN 50022): galvanized - Cable entries: 5 pieces M20x1.5 14
15 Documentation Operating instructions (KA) Nivotester KA00205F/97/A6 Soliwave FQR50/FDR50 KA00206F/97/A6 Tecnical informations Safety instructions Soliwave FQR50/FDR50 TI00378F/97/EN Soliwave FQR50/FDR50-C* with the Nivotester -B* XA00219F/97/A3 Soliwave FQR50/FDR50-D* with the Nivotester -D* XA00484F/97/EN 15
16 Subject to modification Instruments International Instruments International AG Kaegenstrasse Reinach Switzerland Tel Fax info@ii.endress.com TI00377F/97/EN/ InDesign 6.0.6
Nivotester FTR325. Technical Information. 1-channel switch amplifier for the Soliwave M microwave barrier
Technical Information Nivotester FTR325 1-channel switch amplifier for the Soliwave M microwave barrier Area of application The Nivotester FTR325 is suitable as a 1-channel switch amplifier for the Soliwave
More informationLevel Limit Switch nivotester FTL 325 P
Technical Information TI 350F/24/ae Level Limit Switch nivotester FTL 325 P With intrinsically safe signal circuit for connection to the Liquiphant and Soliphant measuring sensor Application Level limit
More informationSoliwave M FQR50 / FDR50
Technical Information Soliwave M FQR50 / FDR50 Microwave barrier Area of application The microwave barrier uses a contact free procedure for detection. It can be installed in containers, conduits, shafts
More informationTechnical Information Nivector FTC968, FTC968Z
TI0007F/00/EN/7.6 7494 Products Solutions Services Technical Information Nivector FTC968, FTC968Z Capacitance Point level switch for powder and fine-grained bulk solids Application The Nivector is a small-sized
More informationMiniature resistance thermometer Explosion-protected version Model TR34, thread-mounted
Electrical temperature measurement Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted WIKA data sheet TE 60.34 Applications Machine building, plant and vessel construction
More informationSoliwave FQR56/FDR56. Technical Information. Microwave Barrier. FQR56 emitter. FDR56 receiver
Technical Information Soliwave FQR56/FDR56 Microwave Barrier FQR56 emitter FDR56 receiver Application The Soliwave microwave barrier uses a contact-free procedure for detection. It can be installed in
More informationRMA42. Technical Information
Technical Information RMA42 Process transmitter with control unit Digital process transmitter for mounting on a top-hat rail for monitoring and visualizing analog measured values Application Plant and
More informationLevel Limit Switch nivotester FTL 370, FTL 372
Technical Information TI 98F/00/en Limit Switch nivotester FTL 370, FTL 37 With intrinsically-safe signal circuits for connection to sensors Liquiphant, Nivopuls, Soliphant. Features at a Glance High operational
More informationSET Installation and Operating Instructions. Level switch for two sensors
Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel: + 358 29 006 260 Fax: + 358 29 006 1260 20.2.2013 Internet: www.labkotec.fi 1/14 SET-2000 Level switch for two sensors Copyright 2013 Labkotec
More informationDigital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version
Electrical temperature measurement Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version WIKA data sheet TE 15.01 Applications Process industry Machine building
More informationOEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted
Electrical temperature measurement OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted WIKA data sheet TE 60.31 further approvals see page 10 Applications Machine building, plant
More informationDigital Input Output Module 24 V for Ex n Zone 2 Series 9472/35
www.stahl.de > 16 channels can be adjusted in pairs as digital inputs or outputs > Suitable for NAMUR proximity switches, 3-wire PNP proximity switches, contacts and solenoid valves (24 V / 0.5 A). > Up
More informationInductive sensor slot-type SI2-K08-AP7
SI2-K08-AP7 Slot sensor, height 8 mm Plastic, polypropylene Mechanical end stop, removable, for analog pointer instruments 3-wire DC, 10 30 VDC NO contact, PNP output Cable connection Wiring diagram Type
More informationRemote Switching Unit nivotester FTL 320
Technical Information TI 203F/24/ae Remote Switching Unit nivotester For connection to intrinsically safe Liquiphant level limit switches Applications Limit detection of fluids and bulk solids in tanks
More informationIsolators A3/1. Ex i Power Supply Field Circuit Ex i Series
> Intrinsically safe output [Ex ib] IIC/IIB > Stable output voltage > Galvanic isolation between output and power supply > Power supply or 85... 230 V AC > Compact design www.stahl.de 07673E00 Basic function:
More informationTemperature Input Module for Zone 1 Series 9482/32
www.stahl.de > 8 channels for temperature sensors > Intrinsically safe inputs Ex ia > For Pt-, Ni- and Cu-resistance temperature detectors according to DIN, IEC and GOST in 2-, 3- and 4-wire circuits >
More informationDigital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version
Temperature Digital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version WIKA data sheet TE 15.01 for further approvals see page 10 Applications
More informationDMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure
DMP 4 Industrial Pressure Transmitter Without Media Isolation accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 0 mbar up to 0... 000 mbar Product characteristics excellent linearity
More information06528E00. ATEX / IECEx NEC 505 NEC 506 NEC 500
Switching Repeater Series 9170 > Intrinsically safe input [Ex ia] IIC > Galvanic isolation between input, output and power supply > Open-circuit / short-circuit monitoring and messaging (can be switched
More informationDMP 343. Industrial Pressure Transmitter. Without Media Isolation. appmeas.co.uk +44 (0)
DMP 4 Industrial Pressure Transmitter Without Media Isolation accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 0 mbar up to 0... 000 mbar Product characteristics excellent linearity
More informationRAD-IN/OUT-8D. Digital extension modules for the bidirectional wireless system. INTERFACE Data sheet _en_05. 1 Description
RAD-IN/OUT-D Digital extension modules for the bidirectional wireless system INTERFACE Data sheet 102122_en_0 PHOENIX CONTACT 2010-02-2 1 Description The RAD-ISM-...-SET-BD-BUS-ANT bidirectional wireless
More informationAssembly. Front view. LED yellow: Relay output. LED red: LB/SC
Switch Amplifier Features Assembly -channel isolated barrier 30 V AC supply Dry contact or NAMUR inputs Relay contact output Line fault detection (LFD) Reversible mode of operation Up to SIL acc. to IEC
More informationIsolators Isolating Repeater Field Circuit Ex i ISpac 9165/ s Art. No
9165/16-11-10s 207909 Compact one- and two-channel Ex i output isolating repeater wire-breakage and short-circuit monitoring system, which can be disconnected and features a signalling contact can be used
More informationTemperature Transmitter itemp RTD DIN rail TMT 127
Technical Information TI 095R/09/en No. 51004790 Temperature Transmitter itemp RTD DIN rail TMT 127 Temperature Transmitter for Pt100 Resistance Thermometer Application Temperature transmitter with fixed
More informationGlass-fibre-reinforced polycarbonate. Dead zone Adaptive or fixed from % 1.5 s adjustable. 20 ms. 60 ms. 60 ms. Approx. 1.
General data Mounting On linear actuators On quarter turn actuators ARCA-integrated or to VDI/VDE 3847-1 or IEC 534-6 (NAMUR) Range of stroke: 3... 130 mm Integrated to VDI/VDE 3847-2 or VDI/VDE 3845 Angle
More informationJUMO dtrans T02 Programmable 4-wire Transmitter (Smart Transmitter)
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net
More informationSALES SERVICE +358 (0)
SALES SERVICE +358 (0)17 262 3555 info@econosto.fi ADCATROL ELECTRO-PNEUMATIC POSITIONERS PE 986 (ATEX) DESCRIPTION The ADCATROL PE986 positioner requires an input signal of 4 20 ma for proportional control
More informationAnalog Universal Module HART for Zone 2 Series 9468/33
> 8 channels can be adjusted individually as analog inputs or outputs > Intrinsically safe inputs/outputs Ex ia > For 0/4 20 ma + HART signals > Line fault monitoring per channel > Diagnostics based on
More information0 2 1 â Ó. itemp HART TMT182. Technical Information
Technical Information itemp HART TMT182 Temperature head transmitter Head transmitter for resistance thermometers (RTD), thermocouples (TC), resistance and voltage transmitters (mv), HART -protocol, for
More informationIsolators A3/1. Vibration Transducer Supply Unit Series
> For vibration, acceleration and speed sensors in 2- and 3-wire design > Space-saving dual-channel version > Signal frequencies up to 50 khz > Easy setting by means of front-side rotary switch > Galvanic
More informationAnalog Input Module HART Ex i / I.S. Inputs, Channels Type 9461/
> 4 channels for 2-wire HART transmitters and 4 channels for 4-wire HART transmitters > Intrinsically safe inputs Ex ia IIC > Galvanic separation between inputs and system > Open-circuit and short-circuit
More informationTechnical Information RMA42
TI00150R/09/EN/05.16 71315973 Products Solutions Services Technical Information RMA42 Process transmitter with control unit Digital process transmitter for monitoring and visualizing analog measured values
More informationTemperature transmitter itemp PCP DIN rail TMT 121
Technical information TI 087R/09/en 510 03671 Temperature transmitter itemp PCP DIN rail TMT 121 Universal temperature transmitter for resistance thermometers (RTD), thermocouples, resistance and voltage
More informationBinary Output Series 9175
Binary Output > Intrinsically safe output [Ex ia] IIC / [Ex ib] IIC > Galvanic isolation between input, output and power supply > Open-circuit and short-circuit monitoring (can be switched off) > For use
More informationLevel Measurement silometer FMX 570
Technical Information TI 01F/00/en Level Measurement silometer FMX 570 Measurement of liquids and bulk solids Suitable for explosion hazardous areas For use with capacitance and hydrostatic probes The
More informationâ Ó itemp TC TMT128 Technical information DIN rail mounted Temperature Transmitter for Thermocouples (TC)
Technical information itemp TC TMT128 DIN rail mounted Temperature Transmitter for Thermocouples (TC) Application Temperature transmitter with fixed measuring ranges for converting TC input signals into
More informationDigital Input Output Module for Zone 2 Series 9470/33
www.stahl.de > 16 channels can be adjusted in pairs as digital input or output > Intrinsically safe inputs/outputs Ex ia > For contacts, NAMUR proximity switches and low-power solenoid valves > Up to 8
More informationOEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted
Temperature OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted WIKA data sheet TE 60.31 further approvals see page 11 Applications Machine building, plant and vessel construction
More informationDigital Input Module NAMUR Ex i / I.S. Inputs, 16 Channels for Zone 1 / Div. 1 Series 9470/22
www.stahl.de > 16 channels for contacts and NAMUR proximity switches (EN 60947-5-6) > Intrinsically safe inputs Ex ia IIC > Galvanic separation between inputs and system > Open-circuit and short-circuit
More informationUser manual. Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD
User manual Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD Contents Precautions Intended use General 1 Specification 3
More informationitemp HART DIN rail TMT112
Temperature transmitter itemp HART DIN rail TMT112 Universal temperature transmitter for resistance thermometers (RTD), thermocouples, resistance and voltage transmitters, incorporating HART protocol Application
More informationOperating Instructions
Level and Pressure Operating Instructions VEGATOR 620, 621, 622 max. 0 10 min. 0 10 on VEGATOR 622! 6 7 8 9 10 11 12 13 14 in out Contents Contents Safety information... 2 Note Ex area... 2 1 Product description
More information09746E00. ATEX / IECEx NEC 505 NEC 506 NEC 500 Class I Class I Class II Class III Zone Zone Division Ex i
Temperature Transmitter Series 9182 www.stahl.de > One unit for nearly all temperature sensors indivdually configurable > Intrinsically safe input [Ex ia] IIC > Signal duplication possible > Galvanic isolation
More informationIsolators A3/1.
> Compact limit value switch with 2 configurable limiting values and 4... 20 ma output > Suitable for 2-, 3-wire transmitters, 2-wire HART transmitters and ma sources > Intrinsically safe input [Ex ia]
More informationGlass-fibre-reinforced polycarbonate. Dead zone Adaptive or fixed from % 20 ms. 60 ms. 60 ms. Approx. 1.3 kg. Approx. 0.9 kg. Approx. 5.
General data Mounting On linear actuators On quarter turn actuators ARCA-integrated or to VDI/VDE 3847-1 or IEC 534-6 (NAMUR) Range of stroke: 3... 130 mm Integrated to VDI/VDE 3847-2 or VDI/VDE 3845 Angle
More informationTechnical Data. General specifications. Rated operating distance s n 3 mm
0102 Model Number Features Direct mounting on standard actuators Compact and stable housing with terminal compartment connection Fixed setting EC-Type Examination Certificate TÜV99 ATEX 1479X Usable up
More informationTechnical Information Barrier RB223
TI00132R/09/EN/13.14 71275790 Products Solutions Services Technical Information Barrier RB223 One or two-channel barrier Loop-powered barrier for the safe separation of 4 to 20 ma standard signal circuits
More informationDigital Input Output Module for Zone 2 Series 9470/33
www.stahl.de > 16 can be adjusted in pairs as digital input or output > Intrinsically safe inputs/outputs Ex ia > For contacts, NAMUR proximity switches and low-power solenoid valves > Up to 8 can be used
More informationTVA 080 / 100 / 120 / 180 / 200 / 220 Isolation amplifier active
Technical manual BA 0318 Signal converter Isolation amplifier active for galvanic isolation, conversion and adjustment of electrical standard signals Evaluable input signals Direct voltage 0...10V / 0...1V
More informationDigital Temperature Transmitters Model T12.10, Universally Programmable, Head Mounting Model T12.30, Universally Programmable, Rail Mounting
Electrical Temperature Measurement Digital Temperature Transmitters Model T12.10, Universally Programmable, Head Mounting Model T12.30, Universally Programmable, Rail Mounting WIKA Data Sheet TE 12.03
More information3 Ex i Isolators. Switching Repeater with Power Relay (250 V / 4 A) Type 9170/ Type 9170/
3 Ex i Isolators Switching Repeater with Power Relay Intrinsically safe input [Ex ia] IIC Galvanic isolation between input, output and power supply Open-circuit / short-circuit monitoring and messaging
More informationIsolating Repeater Input Series 9163
> For 4-wire HART transmitters and voltage sources > Intrinsically safe input [Ex ia] IIC > Galvanic isolation between input, output and power supply > For use up to SIL 2 (IEC 61508) > High accuracy www.stahl.de
More informationINSTALLATION AND OPERATION MANUAL
Version 2.2 MS1016 MTF SERIES VIBRATING PROBES FOR LIQUIDS AND HYC-PKK-312 SWITCHING ISOLATOR INSTALLATION AND OPERATION MANUAL Hycontrol Limited, Larchwood House, Orchard Street, Redditch, Worcestershire,
More informationSafety Instructions Nivotester FTC625, FTC325
XA01351F-A/00/EN/01.14 71273077 Products Solutions Services Safety Instructions FTC625, FTC325 [Ex ia Ga] IIC/IIB TÜV 13.0903 X Segurança TÜV INMETRO OCP 0004 Document: XA01351F-A Safety instructions for
More informationBinary Output without Power Supply Series 9176
Binary Output without Power Supply Series 9176 > For intrinsically safe operation of Ex i solenoid valves, indicators and horns > Power supply by control circuit, loop powered > Intrinsically safe output
More informationIsolators A3/1. Transmitter Supply Unit Field Circuit Ex i Series 9160
Series 9160 > Intrinsically safe input [Ex ia] IIC > Galvanic isolation between input, output and power supply > For use up to SIL 2, special version up to SIL 3 (IEC 61508) > High accuracy www.stahl.de
More informationDigital Input Output Module for Zone 1 Series 9470/32
www.stahl.de > 16 can be adjusted in pairs as digital input or output > Intrinsically safe inputs/outputs Ex ia > For contacts, NAMUR proximity switches and low-power solenoid valves > Up to 8 can be used
More informationIsolators A3/1. Transmitter Supply Unit Field Circuit Ex i Series 9160 Rev. F
Series 9160 Rev. F > Intrinsically safe input [Ex ia] IIC > Galvanic isolation between input, output and power supply > For use up to SIL 2, special version up to SIL 3 (IEC 61508) > High accuracy www.stahl.de
More informationâ itemp Pt100 TMT127 Technical information Pt100 transmitter for DIN rail mounting for high accuracy temperature monitoring
Technical information itemp Pt100 TMT127 Pt100 transmitter for DIN rail mounting for high accuracy temperature monitoring Application Temperature transmitter with fixed measuring range for converting a
More informationPreset counters, electronic
The Codix 716 / 717 can be used universally. These preset pulse, tachometers or preset timers with up to 2 presets can solve a wide variety of control and monitoring tasks in every application. With their
More informationRelative and absolute pressure transmitter type 528
Relative and absolute pressure transmitter type 528 The compact type 528 pressure transmitter is based upon the Huba Control developed and since over 20 years well proven ceramic technology. These transmitters
More informationAnalogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting
Electrical Temperature Measurement Analogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting WIKA Data Sheet TE 24.01 Applications T Machinery, plant construction T Process industry
More informationSBEx-2S BINARY SEPARATOR
SBEx2S BNARY SEPARATOR accompanying device of group of category (M1), group i category (1), two channels in one rail housing (TS35, width 22.5mm), level ia intrinsically safe input circuits compliance
More informationUser manual. Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD
User manual Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD Contents Precautions Intended use General 1 Specification 3
More informationSingle-Channel Safety Barriers with Electronic Current Limitation Series 9004
Single-Channel Safety Barriers with Electronic Current Limitation > Broad product range for all standard applications in the world of automation > Flexible and space saving > Time saving installation due
More informationDigital Input Module 24 V Ex n Inputs, 16 Channels Series 9471
> 16 channels for active 0 / 24 V signals > Zone 2 / Division 2 version for connection of circuits acc. to Ex nl, Ex na, Nonincendive and Non-Ex > Galvanic separation between inputs and system > Two channels
More informationElectronic timer CT-TGD.12 Pulse generator with 1 c/o (SPDT) contact
Data sheet Electronic timer CT-TGD.12 Pulse generator with 1 c/o (SPDT) contact The CT-TGD.12 is an electronic time relay with the function pulse generator. It is from the CT-D range. With their MDRC profile
More informationprosonic T FMU230, FMU231
Technical Information prosonic T FMU230, FMU231 Ultrasonic Level Measurement Compact transmitters for non-contact level measurement of fluids and coarse bulk materials Application The compact transmitters
More informationSubmersible pressure sensor For oils and fuels Model LF-1
Level measurement Submersible pressure sensor For oils and fuels Model LF-1 WIKA data sheet LM 40.04 Applications Level measurement in vessel and storage systems for oils and fuels Overfilling and dry-run
More informationData Sheet KFD2-UFC-EX1.D. Supplied by. .com. Call us on +44 (0)
Data Sheet KFD-UFC-EX1.D Supplied by.com Call us on +44 (0)118 916 940 Email info@47able.com Frequency Converter with Trip Values Features Assembly 1-channel isolated barrier 4 V DC supply (Power Rail)
More informationSystem Components Liquiphant M (NAMUR)/PROFIBUS-PA interface FXA 164
Technical Information TI 343F/00/en Operating Instructions 52004947 System Components Liquiphant M (NAMUR)/PROFIBUS-PA interface FXA 164 With intrinsically safe signal circuits for the connection of up
More informationProcess display RIA 261
Technical Information TI 083R/24/ae Process display RIA 261 Digital loop powered field display for 4 to 20 ma current loops Features and Benefits Loop powered display, no additional power supply cable
More informationIntrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments Model IS-20-H
Replacement product: Model IS-3 Electronic Pressure Measurement Intrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments WIKA Data Sheet PE 81.51 Applications
More informationTemperature transmitter itemp RTD DIN rail TMT 127
Technical information TI 095R/24/ae Nr. 510 06165 Temperature transmitter itemp RTD DIN rail TMT 127 Pt100 transmitter for an economical, high accuracy temperature monitoring. Fixed range measurement,
More informationGESTRA Steam Systems LRS English. Installation & Operating Instructions Conductivity Switch LRS 1-50
GESTRA Steam Systems LRS 1-50 EN English Installation & Operating Instructions 819223-02 Conductivity Switch LRS 1-50 Contents Important notes Page Usage for the intended purpose...4 Function...4 Safety
More informationQUINT-PS/ 3AC/24DC/10
Primary-switched power supply with SFB technology, 3 AC, output current 10 A INTERFACE Data sheet 103131_en_01 1 Description PHOENIX CONTACT - 09/2009 Features QUINT POWER power supply units Maximum system
More informationUltrasonic level meters ULM 70
průmyslová elektronika Ultrasonic level meters ULM 70 For continuous level measurement of liquids (even if polluted), mash and paste materials in open or closed vessels, sumps, open channels, drains, etc.
More informationSINEAX V 624 Programmable Temperature Transmitter for RTD and TC inputs
SINEAX V 6 for rail mounting in housing P/7 or P/7 St 00 II () GD Application SINEAX V 6 (Figs. and ) is designed for measuring temperature in combination with thermocouples or resistance thermometers.
More informationElectronic relays. Monitoring. Monitoring
Electronic relays Neither overload nor underload are particularly good for machines and installations in industrial applications. An unsteady power supply may lead to machine failure and, thus, expensive
More informationTransducer transmitter BILT 4 II 2(1) G. Technical Manual
GB Transducer transmitter BILT 4 II 2(1) G Technical Manual Transducer transmitter BILT 4 Contents Introduction General... 3 Ex.safety description... 4 Technical data... 5 Installation General... 6 Mechanical
More informationJUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: 49 661 60030 Fax: 49 661 6003607 Email: mail@jumo.net Internet: www.jumo.net JUMO
More informationConnection. Input EEx ia IIC. Input I. Input II R 1 R 1 A B A 2+ B LB SC. yellow. red. Composition
solated switch amplifier Output: relay Connection Control circuit EEx ia C 24 V DC nominal supply voltage Reversible mode of operation Lead monitoring (short-circuit LK and interruption LB) with LED indicator
More informationOPTIWAVE 7300 C Supplementary instructions
OPTIWAVE 7300 C Supplementary instructions Radar (FMCW) Level Transmitter for agitated liquids in process applications Supplementary Instructions for ATEX applications KROHNE CONTENTS OPTIWAVE 7300 C General
More informationRemote I/O. 74 Remote I/O. Remote I/O System
System Overview of the System Components IS1+ 76 General Information IS1+ 77 CPU & Power Modules CPU & Power Modules for Zone 1 / Div. 1 IS1+ 9440/22 78 Sockets for CPU & Power Module IS1+ 9490 78 CPU
More informationSubmersible pressure sensor For water and wastewater Model LW-1
Replacement product: Model LF-1 Level measurement Submersible pressure sensor For water and wastewater Model LW-1 WIKA data sheet LM 40.03 Applications Level measurement in rivers and lakes Deep well and
More informationZelio Control Measurement Relays
Selection Guide Application Voltage control Current control Functions 3-phase Single-phase and d.c. Integrated current transformer - Overvoltage and between phases - Overvoltage and between phases and
More informationSINEAX B 811 Power Pack with additional Functions
SINEA B 8 for intelligent and conventional s, in housing S7 for rail and wall mounting 002 II () G Application The power supply unit SINEA B 8 (Figure ) provides the DC power supply for s and transfers
More informationPressure transmitters MBS 4201, MBS 4251, MBS 4701 and MBS 4751
060R9345 Instructions Pressure transmitters MBS 4201, MBS 4251, MBS 4701 and MBS 4751 Content Page 1 Description/application 2 Identification 3 Specifications 4 Safety instructions 5 Installation/dimensions
More informationProcess transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm
Electronic pressure measurement Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm WIKA data sheet PE 86.05 Applications Process technology Machine building and plant
More informationResistance thermometer Model TR30, compact design
Electrical temperature measurement Resistance thermometer Model TR30, compact design WIKA data sheet TE 60.30 Applications Machine building, plant and vessel construction Propulsion technology, hydraulics
More information02/11/2015
Modem communication plug and play solutions GSM Part number 88970119 For remote control of your application Automatic notification of alarms via SMS (GSM Modem) / email or on a PC with M3 ALARM software.
More informationSITRANS F flowmeters SITRANS F US. SONOFLO SONO 3000 transmitter. 4/152 Siemens FI Overview
Overview Design The transmitter is available in different enclosure to fit any demands: IP67 enclosure for compact or wall mounting IP00 as a 19 insert IP65 for wall mounting IP65 enclosure in stainless
More informationField indicator for current loops with HART communication Models DIH50, DIH52
Accessories Field indicator for current loops with HART communication Models DIH50, DIH52 WIKA data sheet AC 80.10 for further approvals see page 6 Applications Process engineering Plant construction General
More informationTechnical Information itemp TMT180
TI00088R/09/EN/14.16 71315611 Technical Information itemp TMT180 Temperature head transmitter for resistance thermometers Pt100 Configurable using a PC, for installation in a terminal head, form B Application
More informationMonitoring Relays True RMS 3-Phase, 3-Phase+N, Multi-function Types DPB01, PPB01
Monitoring Relays True RMS 3-Phase, 3-Phase+N, Multi-function Types, TRMS 3-phase over and under voltage, phase sequence and phase loss monitoring relays Detect when all 3 phases are present and have the
More informationDATASHEET - ETR4-70-A. Delivery program. Timing relay, 2W, 0.05s-100h, multi-function, VAC/DC, potentiometer connection
DATASHEET - ETR4-70-A Delivery program Timing relay, 2W, 0.05s-100h, multi-function, 24-240VAC/DC, potentiometer connection Part no. ETR4-70-A Catalog No. 031888 Eaton Catalog No. XTTR6A100H70B EL-Nummer
More informationIntrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F, IS-21-F
Replacement product: Model IS-3 Electronic Pressure Measurement Intrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F,
More informationICARUS ELECTRONIC GAS VOLUME CONVERTER
ICARUS ELECTRONIC GAS VOLUME CONVERTER 1. INTRODUCTION 2. TECHNICAL DETAILS ICARUS is an electronic volume converter which, combined with a gas meter, provides the gas measurement in Sm3, calculated at
More informationInstallation and Operational Instructions for ROBA -multiswitch Type 019._00.2
Guidelines on the Declaration of Conformity A conformity evaluation has been carried out for the product in terms of the EU Low Voltage Directive 2014/35/ EU and the Electromagnetic Compatibility (EMC)
More informationInstallation guide 971 SmartRadar LTi
Installation guide 971 SmartRadar LTi March 2009 Part no. 4416.715 Revision 3 Enraf B.V. P.O. Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100 Fax : +31 15 2701 111 E-mail : enraf-nl@honeywellenraf.nl
More information