CP CU1. Reference Manual COUPLING UNIT FOR LINE IMPEDANCE AND K FACTOR MEASUREMENTS, COUPLING MEASUREMENTS ON POWER LINES AND SIGNAL CABLES AND

Size: px
Start display at page:

Download "CP CU1. Reference Manual COUPLING UNIT FOR LINE IMPEDANCE AND K FACTOR MEASUREMENTS, COUPLING MEASUREMENTS ON POWER LINES AND SIGNAL CABLES AND"

Transcription

1 CP CU1 Reference Manual COUPLING UNIT FOR LINE IMPEDANCE AND K FACTOR MEASUREMENTS, COUPLING MEASUREMENTS ON POWER LINES AND SIGNAL CABLES AND GROUND IMPEDANCE MEASUREMENTS OF LARGE SUBSTATIONS V1.4

2 CP CU1 Reference Manual V 1.4 Article Number VESD Manual Version: CPCU1.AE.4 OMICRON electronics All rights reserved. This Reference Manual is a publication of OMICRON electronics GmbH. All rights including translation reserved. Reproduction of any kind, e.g., photocopying, microfilming or storage in electronic data processing systems, requires the explicit consent of OMICRON electronics. Reprinting, wholly or in part, is not permitted. This Reference Manual represents the technical status at the time of printing. The product information, specifications, and all technical data contained within this Reference Manual are not contractually binding. OMICRON electronics reserves the right to make changes at any time to the technology and/or configuration without announcement. OMICRON electronics is not to be held liable for statements and declarations given in this Reference Manual. The user is responsible for every application described in this Reference Manual and its results. OMICRON electronics explicitly exonerates itself from all liability for mistakes in this manual. 2

3 Contents Contents Using This Manual Operator Qualifications and Safety Standards Conventions and Symbols Used Related Documents Safety Rules General Operating the Measurement Setup Orderly Measures Disclaimer Hardware Information Overview Circuit Diagram of CP CU Operating Controls of CP CU CP CU1 Accessories CP GB1 Grounding Box Description Shorting the Phases Changing the Surge Arrestors Clamp-on Ammeter Operation Measurement Setup Operating Principle Configuring CPC Setting CP CU Applications Template Usage Safety Instructions for Connecting CP CU1 to Power Lines Before Starting Recommended Current Range Settings Estimating the Open-Line Voltage Connecting the Measurement Setup to Power Lines k Factor Measurement Why k Factor Measurement?

4 CP CU1 Reference Manual V Performing Measurements Interpretation of Measurement Results Ground Impedance and Step Voltage Measurement Introduction Performing Measurements Interpretation of Measurement Results Measurement of Coupling into Signal Cables Introduction Performing Measurements Technical Data CP CU1 Output Ranges CP CU1 Measuring Transformers CP CU1 Inputs CP GB1 Specifications Output Power Accuracy Environmental Conditions Mechanical Data Clamp-on Ammeter (Accessory) Specifications Contact Information / Technical Support Index

5 Using This Manual Using This Manual This Reference Manual provides detailed information on how to use the CP CU1 coupling unit safely, properly and efficiently. The CP CU1 Reference Manual contains important safety instructions for working with CP CU1, gets you familiar with operating CP CU1, and provides typical application examples. Following the instructions in this Reference Manual will help you to prevent danger, repair costs and possible down time due to incorrect operation. The CP CU1 Reference Manual always has to be available at the site where CP CU1 is used. It must be read and observed by all users of CP CU1. Reading the CP CU1 Reference Manual alone does not release you from the duty of complying with all national and international safety regulations relevant to working with CPC 100 and CP CU1. The regulation EN "The Erection and Operation of Electrical Test Equipment" as well as all the applicable regulations for accident prevention in the country and at the site of operation has to be fulfilled. Operator Qualifications and Safety Standards Working on overhead lines is extremely dangerous. Testing with CP CU1 must be carried out only by qualified, skilled and authorized personnel. Before starting to work, clearly establish the responsibilities. Personnel receiving training, instructions, directions, or education on CP CU1 must be under constant supervision of an experienced operator while working with the equipment. Testing with CP CU1 must comply with the relevant national and international safety standards listed below: EN (VDE 0104) "Erection and Operation of Electrical Equipment" EN (VDE 0105 Part 100) "Operation of Electrical Installations" IEEE 510 "IEEE Recommended Practices for Safety in High-Voltage and High-Power Testing" LAPG NASA "Electrical Safety" Moreover, additional relevant laws and internal safety standards have to be followed. 5

6 CP CU1 Reference Manual V 1.4 Conventions and Symbols Used In this manual, the following symbols indicate paragraphs with special safety relevant meaning: Symbol Description Equipment damage or loss of data possible. Personal injury or severe damage to objects possible. Related Documents The following documents complete the information covered in the CP CU1 Reference Manual: Title CPC 100 User Manual CPC 100 Reference Manual Description Contains information on how to use the CPC 100 test system and relevant safety instructions. Contains detailed hardware and software information on CPC 100 including relevant safety instructions. 6

7 Safety Rules Safety Rules Before operating the CP CU1 coupling unit, read the following safety rules carefully. If you do not understand some safety rules, contact OMICRON electronics before proceeding. CP CU1 is designated for use with the CPC 100 test system. Therefore observe the safety rules both in this Reference Manual and in the CPC 100 User/Reference Manual when working with CP CU1. Depending on the application and the device under test, specific safety instructions must be observed. Very often the danger coming from the device under test is even higher that the danger from CP CU1 itself. For application specific safety instructions, see 3.2 "Safety Instructions for Connecting CP CU1 to Power Lines" on page 26. General Always observe the five safety rules: Disconnect completely Secure against re-connection Verify that the installation is dead Carry out grounding and short-circuiting Provide protection against adjacent live parts Do not touch any terminals without a visible connection to ground. Before handling CP CU1 or CPC 100 in any way, connect them with a solid connection of at least 6 mm 2 cross-section to ground. Ground CP CU1 as close as possible to CPC 100. Use the CP GB1 grounding box to connect CP CU1 to overhead lines and power cables. For detailed information, see the application specific 3.2 "Safety Instructions for Connecting CP CU1 to Power Lines" on page 26. When using CP GB1, ground it near the place where the connection to the test object is made. Make sure that the grounding stud is in good condition, clean and free of oxidation. Make sure that all studs and cables of CP GB1 are screwed tight. Make sure that the test object s terminals to be connected to CP CU1 do not carry any voltage potential. During a test, the only power source for a test object may be CP CU1 (powered by CPC 100). The only exception are measurements on overhead lines as described in 3 "Applications" on page 25. Do not open the CP CU1 s or CP GB1 s housing. 7

8 CP CU1 Reference Manual V 1.4 Do not repair, modify, extend, or adapt CP CU1, CP GB1 or any accessories. Use only original accessories available from OMICRON electronics. Use CP CU1, CP GB1 and their accessories only in a technically sound condition and when its use is in accordance with the regulations. In particular, avoid disruptions that could in turn affect safety. Do not use CP CU1 if you have a cardiac pacemaker. Before operating CP CU1 make sure that there is no person with a cardiac pacemaker in the immediate vicinity of the measurement setup. Operating the Measurement Setup Before operating CP CU1, CPC 100 and, when used, CP GB1, ground them as described in "General" on page 7. When using CP GB1, ground it near the place where the connection to the test object is made. Make sure that the grounding stud is in good condition, clean and free of oxidation. Life threatening voltages up to 600 V can appear on all CP GB1 s contacts and on all clamps and cables connected to CP CU1 during the test. Keep safe distance from them. Before handling CP CU1 or CP GB1 in any way (even before setting the current range switch), make sure that the device under test (e.g. overhead lines or power cables) are well grounded (e.g. by closing the grounding switch) near the measurement setup. Power CP CU1 only from the CPC 100 s EXT. BOOSTER output. Use only booster cables supplied by OMICRON electronics. Ensure that the short-circuit bar is always plugged in the CP CU1 s I AC output whenever the output is not connected to the I AC input of CPC 100. Connect the CP CU1 s I AC output exclusively to the I AC input of CPC 100. Before connecting CP CU1 with CPC 100, turn off CPC 100 either by the POWER ON/OFF switch or the Emergency Stop button. Set the current range switch on the CP CU1 s front panel only when CPC 100 is turned off and the test object is grounded. In addition to the above safety rules follow the application specific 3.2 "Safety Instructions for Connecting CP CU1 to Power Lines" on page 26. 8

9 Safety Rules Orderly Measures The CP CU1 Reference Manual or alternatively the e-book in PDF format has always to be available on the site where CP CU1 is being used. It must be read and observed by all users of CP CU1. CP CU1 may be used only as described in 3 "Applications" on page 25. Any other use is not in accordance with the regulations. The manufacturer and/or distributor is not liable for damage resulting from improper usage. The user alone assumes all responsibility and risk. Following the instructions provided in this Reference Manual is also considered part of being in accordance with the regulations. Disclaimer If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. 9

10 10 CP CU1 Reference Manual V 1.4

11 Hardware Information 1 Hardware Information 1.1 Overview CP CU1 is a coupling unit designated for measurements with the CPC 100 test system mainly on overhead lines and power cables. Typical applications include line and cable impedance measurements, measurements of k factors, mutual coupling of power lines, measurements of coupling between power lines and signal cables, and ground impedance measurements. The CP GB1 grounding box (see 1.5 "CP GB1 Grounding Box" on page 14) available as an accessory from OMICRON electronics is a surge arrestor unit protecting the operating staff and equipment from high-voltage hazards during measurements on overhead lines and power cables in case of unexpected events on the power line. 1.2 Circuit Diagram of CP CU1 Figure 1-1: "Circuit Diagram of CP CU1" below shows the principal circuit diagram of the coupling unit. Figure 1-1: Circuit Diagram of CP CU1 Fuse 30 A Power transformer Current range switch IAC (0 2.5A) Surge arrestor BOOSTER CT I OUT (0 100 A) Voltmeter V1 AC (0 30 V) V SENSE (0 600 V) Surge arrestor VT 11

12 CP CU1 Reference Manual V Operating Controls of CP CU1 The front panel of CP CU1 provides the following functional elements: BOOSTER input for connecting with the CPC 100 s EXT. BOOSTER output Current range switch for setting the current range of CP CU1 Voltmeter for measuring the voltage at the test object s terminals I OUT current output I AC output for measuring the output current using a CT (current transformer) with the 100 A : 2.5 A transformation ratio The output is to be connected with the I AC input of CPC 100. V SENSE input for measuring the voltage at the test object s terminals V1 AC output for measuring the voltage at the test object s terminals using a VT (voltage transformer) with the 600 V : 30 V transformation ratio The output is to be connected with the V1 AC input of CPC 100. Short-circuit bar for shorting the I AC output whenever the output is not connected to the I AC input of CPC 100 Equipotential ground terminal for grounding CP CU1 close to the position of the operating staff Figure 1-2: "Front Panel" below shows the CP CU1 s functional elements. Figure 1-2: Front Panel Equipotential ground terminal Location to I AC output V1 AC output V SENSE input store the shortcircuit bar Fuse 30 A BOOSTER input Current range switch Voltmeter I OUT output 12

13 Hardware Information 1.4 CP CU1 Accessories The following accessories are delivered with the CP CU1 coupling unit: Table 1-1: CP CU1 Accessories Accessories Booster cable Description Power connection from the CPC 100 s EXT. BOOSTER output to the CP CU1 s BOOSTER input V1 AC coax. cable Connection from the CPC 100 s V1 AC input to the CP CU1 s V1 AC output 4 Banana cable Connection from the CPC 100 s I AC input to the CP CU1 s I AC output and connection from the CP CU1 s V SENSE input to the Kelvin clamps voltage sense outputs 2 Kelvin cable Connection from the CP CU1 s I OUT output to the current feed-in point (usually at CP GB1) Grounding cable Connection from the CP CU1 s equipotential ground terminal to the substation ground Short-circuit bar A bar for shorting the CP CU1 s I AC output when the output is not connected to the I AC input of CPC

14 CP CU1 Reference Manual V CP GB1 Grounding Box Description The CP GB1 grounding box (see Figure 1-3: "CP GB1 Grounding Box" below) is a surge arrestor unit for connecting CP CU1 to the test object. If high voltage appears for a short time on the test object s terminals, an arc discharges the voltage and distinguishes without destroying the grounding box. If the arc persists for a longer time period, the surge arrestor insulator melts and the terminals are short-circuited to ground, thereby protecting the operating staff, CP CU1 and CPC 100. Figure 1-3: CP GB1 Grounding Box Other stud standards available optionally L1 line stud A/L1/red Equipotential ground stud L2 line stud B/L2/yellow L3 line stud C/L3/blue Caution: The CP GB1 grounding box must be used for measurements on overhead lines or power cables. The CP GB1 grounding box is available for three different ground connection types: cylindrical grounding studs of 16 mm diameter or ball studs of 20 mm and 25 mm (1 inch) diameter. The grounding socket clamp is needed for secure ground connection of CP GB1 to the substation ground. The grounding socket clamps compatible with the grounding studs in the substation are given in Table 14

15 Hardware Information 1-2: "Grounding Studs and Socket Clamps" below. For ordering information, contact OMICRON electronics sales office. When ordering CP GB1, choose one connection set; additional connection sets are available optionally. Table 1-2: Grounding Studs and Socket Clamps Grounding Stud in the Substation 16 mm cylindrical grounding stud Proper Grounding Socket Clamp mm grounding socket clamp (shipped with the 16 mm cylindrical and 20 mm ball CP GB1 s studs) 12 mm 20 mm ball grounding stud 25 mm ball grounding stud 25 mm grounding socket clamp (shipped with the 25 mm CP GB1 s studs) 16 mm 15

16 CP CU1 Reference Manual V 1.4 Figure 1-4: Screwing the CP GB1 s Studs Caution: Depending on the type of grounding studs in the substation, the appropriate connection set and socket clamp have to be used. Connecting socket clamps of one type to a grounding point of another system is highly dangerous on both the connection of the grounding set to CP GB1 and the connection of CP GB1 to the grounding point in the substation. The 16 to 20 mm socket clamps are designed and tested for fault currents up to 26.5 ka, the 25 mm (1 inch) socket clamp for fault currents up to 30 ka, both for a maximum duration of 100 ms. On locations where higher fault currents are possible, CP CU1 and CP GB1 must not be used. For transportation, the CP GB1 s studs are usually removed. If this is the case, mount them onto CP GB1 using the delivered wrench and screw them tight (see Figure 1-4: "Screwing the CP GB1 s Studs" on page 16). 16

17 Hardware Information Figure 1-5: Three-Lead Cable Shorting the Phases A three-lead cable is delivered with CP GB1 for shorting all phases for L1 L2 L3-E measurements (see Figure 3-5: "Zero-Sequence Impedance Measurement" on page 34, Figure 3-8: "Ground Impedance and Step Voltage Measurement" on page 41, Figure 3-13: "Measurement with the Loop Between Parallel Lines and Ground" on page 48 and Figure 3-14: "Calibration with the Loop Between Parallel Lines and Ground" on page 49). Figure 1-6: Shorting the Phases To short the phases, connect the line studs of CP GB1 as shown in Figure 1-6: "Shorting the Phases" below. 17

18 CP CU1 Reference Manual V 1.4 Figure 1-7: Surge Arrestors Changing the Surge Arrestors The surge arrestors of CP GB1 can permanently short-circuit the CP GB1 s terminals to ground if overvoltage appears on the terminals. Even short transients can cause a discharge and, if the energy is too high, possibly damage the surge arrestor. Defective surge arrestors can result in erroneous measurement results. If the measurement results obtained using CP GB1 differ considerably from the expected values, check the surge arrestors using CPC 100 as follows. Apply a voltage of 500 V for at least 10 seconds using the VWithstand test card from the resistance test cards. Set a test current of 0.01 A. If the current is exceeded, an error message is displayed. In this case the surge arrestor under test is defective and you have to replace it. If no message is displayed, the surge arrestor is intact. For detailed information on this test, see the CPC 100 Reference Manual. Repeat the test for all three studs A/L1, B/L2 and C/L3. Replace defective surge arrestors only by spare parts from OMICRON electronics (see Figure 1-7: "Surge Arrestors" below). For ordering information, contact OMICRON electronics sales office. Note: Before changing the surge arrestors, check whether there is a fault that caused the problem and remove it. 18

19 Hardware Information Figure 1-8: Opening the Surge Arrestor Chamber To replace a surge arrestor: 1. Disconnect CP GB1 completely and observe the five safety rules in "Safety Rules" on page Open the surge arrestor chamber using a 22 mm wrench by removing the contact screw (see Figure 1-8: "Opening the Surge Arrestor Chamber" on page 19). Contact screw 3. Turn CP GB1 upright and move the stud over the surge arrestor chamber until the surge arrestor falls out. 4. Replace the defective surge arrestor by the spare one. 5. Screw the contact screw very tight (torsional moment of Nm). 19

20 CP CU1 Reference Manual V 1.4 Figure 1-9: Clamp-on Ammeter 1.6 Clamp-on Ammeter A clamp-on ammeter for AC 50/60 Hz (see Figure 1-9: "Clamp-on Ammeter" below) is available from OMICRON electronics as an accessory. For ordering information, contact OMICRON electronics sales office. The clamp-on ammeter provides the following features: Hold function Battery check Auto power off Bar display Voltmeter Ohmmeter Contact check with beeper 20

21 Operation 2 Operation 2.1 Measurement Setup The measurement setup consists of the CPC 100 test system, of the CP CU1 coupling unit and, in case of measurements on overhead lines and power cables, of the CP GB1 grounding box. Figure 2-1: "Measurement Setup" below shows the functional block diagram. Caution: CP CU1 must be connected to the test object through the CP GB1 grounding box for measurements on overhead lines or power cables. For these applications, connect the measurement setup by following 3.2 "Safety Instructions for Connecting CP CU1 to Power Lines" on page 26. Figure 2-1: Measurement Setup Dangerous zone V1 AC IAC IAC V1 AC CP GB1 IOUT (optional) CPC 100 CP CU1 Device under test EXT. BOOSTER BOOSTER V SENSE 2.2 Operating Principle CP CU1 is a coupling unit controlled by the CPC 100 test system via the BOOSTER interface. CP CU1 provides programmable current signals at the I OUT output and facilitates measuring of the output current and the voltage at the test object s terminals. The current range of 10 A, 20 A, 50 A or 100 A is set by the CPC 100 software (see 2.3 "Configuring CPC 100" on page 22) and the current range switch on the front panel of CP CU1 is set manually by the user (see 2.4 "Setting CP CU1" on page 23). The output current and the voltage at the test object s terminals connected to the CP CU1 s V SENSE input are processed by the coupling unit for measuring with CPC 100. The output current is transformed by a current transformer with the transmission ratio 100 A : 2.5 A and the secondary transformer winding wired to the I AC output. The transformed current at the I AC output is measured via the 21

22 CP CU1 Reference Manual V 1.4 I AC input of CPC 100. The test object s voltage is transformed by a voltage transformer with the transmission ratio 600 V : 30 V and the secondary transformer winding wired to the V1 AC output. The transformed voltage at the V1 AC output is measured via the V1 AC input of CPC 100. The measurements are performed frequency selective, i.e. only signal components at the generated frequency different from the mains frequency are analyzed. Due to the high-power disturbances at the mains frequency, the spectral components around the mains frequency and its harmonics are filtered out. For detailed information on the frequency selective measurement, see "Getting Started with Quick: The frequency selective measurement" in the CPC 100 Reference Manual. 2.3 Configuring CPC 100 Figure 2-2: Options Window Note: The minimum CPC 100 software version required is V 1.4. If you have an earlier version installed, upgrade the software from the CD-ROM delivered with your CP CU1. CPC 100 must be configured for CP CU1. To configure CPC 100: 1. Press the Options view selector button. The Options window with the Device Setup tab selected (see Figure 2-2: "Options Window" below) opens. 2. Select CU1 from the External booster combo box. The CT and VT settings are set according to the built-in current and voltage transformers automatically. 3. Press the Test Card View view selector button and insert the Quick (default), Sequencer or Ramping test card. Figure 2-3: "Quick Test Card Window" below shows the setting using the Quick test card as example. 22

23 Operation Figure 2-3: Quick Test Card Window Output Range combo box Measured Quantities combo box 4. Open the item list from the Output Range combo box. The item list displays the CU1:10A, CU1:20A, CU1:50A and CU1:100A CP CU1 relevant current ranges. 5. Select VT sel and/or CT sel from the Measured Quantities combo boxes when measuring with the CP CU1 s built-in voltage and current transformers to account for their transformation ratio. 6. Select the current range you want to use. 2.4 Setting CP CU1 Set the current range of CP CU1 using the current range switch (see 1.3 "Operating Controls of CP CU1" on page 12) to the value configured by the CPC 100 software. Caution: Set the current range switch on the CP CU1 s front panel only when CPC 100 is turned off and the test object is connected to ground with closed grounding switch near the measurement setup. Note: Current range settings on the test card and on the CP CU1 s front panel must be the same. 23

24 24 CP CU1 Reference Manual V 1.4

25 Applications 3 Applications 3.1 Template Usage The test procedures running on the measurement setup are controlled by templates available on the CPC Explorer CD-ROM shipped with your CP CU1 or in the customer area of the OMICRON electronics home page The templates are pairs of XML documents and Microsoft Excel templates designed by OMICRON electronics for designated applications. The XML templates are predefined test procedures, often with comments, that run on CPC 100 and guide the user through the test. Once completed, the XML file is saved, downloaded to the PC using CPC Explorer and then loaded into the corresponding Microsoft Excel template. There the results are post processed and a final test report is generated. The template pairs facilitate and speed testing with CP CU1 and the evaluation of results. Note: Some template pairs allow version control. If an error message appears after loading the XML template, use a template pair of the same version. To run a test procedure according a template: 1. Using CPC Explorer, upload the XML template for the intended application from the PC to CPC Open the template on CPC Run the test procedure according to the template. 4. After completing the test procedure, save the test in a new file. 5. Using CPC Explorer, download the test results from CPC 100 to your preferred working directory on the PC. 6. Open the corresponding Microsoft Excel template by double-clicking the *.xlt file in the folder labeled with the test procedure name. A Microsoft Excel workbook appears. 7. Click the Load XML-File button and open the *.xml file saved in your preferred working directory before to load the test results. 8. After all worksheets are filled with data, the test results are calculated. 25

26 CP CU1 Reference Manual V Safety Instructions for Connecting CP CU1 to Power Lines Before Starting Caution: A lightning discharge to the line under test can cause injury or possibly death of the operating staff. Do not connect the measurement setup to overhead lines if there is a possibility of a thunderstorm over any part of the lines to be measured. Caution: Connecting the measurement setup to overhead lines with a life parallel system bring about high-voltage hazards. It is strongly recommended to take all parallel lines out of service before proceeding. Before connecting CP CU1 to overhead lines or power cables (further on referred to as power lines), you must estimate the open-line voltage as follows. Follow the instructions below exactly and sequentially to protect yourself from high-voltage hazards. In addition to the following safety instructions, observe "Safety Rules" on page Recommended Current Range Settings The highest current range allowed by the open-line voltage (see "Connecting the Measurement Setup to Power Lines" on page 28) provides the best measurement accuracy. However, depending on the length of the power line under test, this setting may result in CPC 100 overload due to low driving voltage. As a rule of thumb, the current range required for the power line length is given in Table 3-1: "Recommended Current Range Settings" below. Set the current range switch of CP CU1 to the value according to the table. Table 3-1: Recommended Current Range Settings Line Impedance Line Length Current Range Compliance Voltage Ω 0 2 km/0 1.5 miles 100 A 50 V Ω 1 10 km/0.5 5 miles 50 A 100 V Ω 5 50 km/3 30 miles 20 A 250 V > 16 Ω > 20 km/15 miles 10 A 500 V 26

27 Applications Estimating the Open-Line Voltage To estimate the open-line voltage: 1. Switch off, short-circuit and ground the power line on both sides using an installed grounding switch or, if no grounding switch is available on site, using grounding cables (further on, the grounding switch or these extra grounding cables are referred to as grounding switch). 2. Make sure that the connection to ground at the far end of the power line is not removed during the complete test procedure. 3. In addition to the grounding switch, ground the line at the near end using a grounding set consisting of three cables rated for the maximum short-circuit current possible on the line. This connection is called working ground further on. 4. Open the grounding switch at the near end of the power line and measure the current through the working ground using a clamp-on ammeter on all three phases. 5. Close the grounding switch. 6. Calculate the estimated open-line voltage after removal of the grounding cables as follows: V est [V] = I meas [A] 0.4 [Ω/km] 2 l line [km] (Eq. 3-1) or V est [V] = I meas [A] 0.64 [Ω/mile] 2 l line [miles] (Eq. 3-2) where V est [V] is the estimated open-loop voltage in volts, I meas is the highest measured current in ampers, 0.4 [Ω/km] = 0.64 [Ω/mile] is the constant of a typical overhead line per wire and l line [km] and l line [miles] is the length of the line in km and miles respectively. Caution: If the estimated open-line voltage is > 500 V, stop. The measurement is not possible due to high-voltage hazard. Try to take parallel lines out of service V, the measurement is possible only in the 10 A range V, the measurement is possible in the 10 A or 20 A range V, the measurement is possible in the 10 A, 20 A or 50 A range. < 50 V, the measurement is possible in all current ranges. 7. If the current range allowed by the estimated open-line-voltage is lower as the current range set according to Table 3-1: "Recommended Current Range Settings" on page 26, set the current range switch of CP CU1 to the value allowed by the open-line voltage. 27

28 CP CU1 Reference Manual V 1.4 Caution: During the grounding switch at the near end of the power line is open, the area around CP GB1 in the range of 5 m/15 ft and around CP CU1 in the range of 2 m/5 ft is a dangerous zone due to high-voltage and mechanical hazards. Do not enter the dangerous zone. Keep the grounding switch open for a time as short as possible. Caution: If you see or hear anything uncommon in the test equipment, e.g. noise of electrical discharge or lightening of surge arrestors, close the grounding switch before touching the measurement setup Connecting the Measurement Setup to Power Lines If the estimated open-line voltage (see "Estimating the Open-Line Voltage" on page 27) allows measurement in the current range you want to use, connect the measurement setup to the power line as follows: 1. Make sure that the grounding switch is closed. 2. Connect CP GB1 to ground using the delivered cable near the place where the connection to the line is made. Make sure that the grounding stud is in good condition, clean and free of oxidation. Caution: Depending on the type of grounding points in the substation, the appropriate connection set and socket clamp have to be used. Connecting socket clamps of one type to a grounding point of another system is highly dangerous on both the connection of the grounding set to CP GB1 and the connection of CP GB1 to the grounding point in the substation. The 16 to 20 mm socket clamps are designed and tested for fault currents up to 26.5 ka, the 25 mm (1 inch) socket clamp for fault currents up to 30 ka, both for a maximum duration of 100 ms. On locations where higher fault currents are possible, CP CU1 and CP GB1 must not be used. 3. Disconnect the grounding cables from the ground (the grounding switch is closed!) and connect them to the CP GB1 s line studs. 4. Position CP CU1 at a minimum distance of 5 m/15 ft from CP GB1. 5. Position CPC 100 at a minimum distance of 5 m/15 ft from CP CU1 and 10 m/30 ft from CP GB1. 6. Ground CP CU1 using a cable of at least 6 mm 2 cross-section close to CPC 100 and the position of the operator. 7. Connect CP CU1 with CP GB1 as shown in Figure 3-1: "Wiring the Measurement Setup" on page

29 Applications Figure 3-1: Wiring the Measurement Setup Connection using grounding sets on site L3/C L2/B L1/A 8. Ground CPC 100 using a cable of at least 6 mm 2 cross-section close to the position of the operator. 9. Connect CP CU1 with CPC 100 as shown in Figure 3-1: "Wiring the Measurement Setup" above. 10.Mark the area around CP GB1 in the range of at least 5 m/15 ft and around CP CU1 in the range of at least 2 m/5 ft as dangerous zone. 11.Open the grounding switch and read the voltmeter on the CP CU1 s front panel from outside of the dangerous zone. Caution: If the voltmeter s reading is > 500 V, stop. The measurement is not possible due to high-voltage hazard V, the measurement is possible only in the 10 A range V, the measurement is possible in the 10 A or 20 A range V, the measurement is possible in the 10 A, 20 A or 50 A range. < 50 V, the measurement is possible in all current ranges. If the open-line voltage allows measurement, proceed as described in the respective "Performing Measurements" section of the following applications. Caution: Make sure that the grounding switch is always closed when no measurement is performed and especially when the wiring is modified or the current range switch of CP CU1 is set. 29

30 CP CU1 Reference Manual V k Factor Measurement Why k Factor Measurement? On most modern secondary distance protection relays, the value of the positivesequence (line) and zero-sequence (line-to-ground) impedance or the ground impedance matching factor (k factor) is required to make the relay settings. The line impedance can be readily calculated but the chosen values for the ground impedance often do not match the actual conditions. The accuracy of these settings is crucial to the operation of the relay because they directly affect the reach of the different protection zones e.g. in case of a line-to-ground fault. Measurements show that in significant number of cases the k factor of the measured lines is set more than 20% from its actual value. This can result in zone under- or overreach and consequently, the selectivity is lost. This situation is particularly relevant to underground power cables. The k factors are line parameters independent of the fault location describing the ratio of the line and ground impedances. The following k factor definitions are commonly used: The complex ratio of the ground impedance Z E and the line impedance Z L k L = Z E /Z L = (Z 0 /Z 1 1)/3, (Eq. 3-3) the complex ratio of the zero-sequence impedance Z 0 and the positivesequence impedance Z 1 (see Figure 3-2: "Zero-Sequence Impedance Definition" below) k 0 = Z 0 /Z 1 (Eq. 3-4) and a couple of real values R E /R L (Eq. 3-5) X E /X L (Eq. 3-6) where R E and X E are the real and imaginary parts respectively of the ground impedance and R L and X L are the real and imaginary parts respectively of the line impedance. 30

31 Applications Figure 3-2: Zero-Sequence Impedance Definition The k factor is an important setting of distance protection relays. The precision of this setting affects the accuracy of distance protection relays dramatically. The k factor can be calculated, but the calculation results give only a rough estimate of the actual value. As a wrong k factor setting can cause worse power quality, higher risk to lose the system stability and loss of power supply, k factor measurements are essential for fast, selective and reliable distance protection. Because there are usually strong disturbances by other lines in service, measurement at the mains frequency is not feasible. All measurements running according to the templates are done below and above the mains frequency and the results are interpolated. From these results the positive- and zero-sequence impedances as well as the k factor in various formats are calculated Performing Measurements Connect the measurement setup to the overhead lines or power cables under test following 3.2 "Safety Instructions for Connecting CP CU1 to Power Lines" on page 26. Note: For line length below 5 km/3 miles it is recommended to connect the V SENSE input of CP CU1 as close as possible to the VT of the line to reduce the additional impedance of the current feed in the path. For longer lines, you can connect the V SENSE input with the Kelvin clamps directly on CP GB1. 31

32 CP CU1 Reference Manual V 1.4 In the course of the k factor test procedure, the following measurements are performed: Line-to-line impedance measurements: L1-L2, L1-L3, L2-L3 (Figure 3-3: "Line-to-Line Impedance Measurements" below shows the L1-L2 measurement as example.) Figure 3-3: Line-to-Line Impedance Measurements Far end Near end Overhead line V1 AC IAC CPC 100 EXT. BOOSTER IAC V1 AC IOUT CP CU1 BOOSTER V SENSE CP GB1 32

33 Applications Line-to-ground impedance measurements: L1-E, L2-E, L3-E (Figure 3-4: "Line-to-Ground Impedance Measurements" below shows the L1-E measurement as example.) Figure 3-4: Line-to-Ground Impedance Measurements Far end Near end Overhead line V1 AC IAC CPC 100 EXT. BOOSTER IAC V1 AC IOUT CP CU1 BOOSTER V SENSE CP GB1 33

34 CP CU1 Reference Manual V 1.4 Zero-sequence impedance measurements: L1 L2 L3-E (see Figure 3-5: "Zero-Sequence Impedance Measurement" below) Short the three phases with the delivered three-lead cable as shown in Figure 1-6: "Shorting the Phases" on page 17. Figure 3-5: Zero-Sequence Impedance Measurement Far end Near end Overhead line V1 AC IAC CPC 100 EXT. BOOSTER IAC V1 AC IOUT CP CU1 BOOSTER V SENSE CP GB1 The test procedure is controlled by templates available on the CPC Explorer CD-ROM shipped with your CP CU1 or in the customer area of the OMICRON electronics home page. For detailed information on the templates and instructions how to use them, see 3.1 "Template Usage" on page 25. It is recommended to use the same test current for all measurements. To find out the highest test current possible, start the test procedure with the measurement featuring the highest impedance, i.e., the L1-L3 measurement on power cables and the L1-E measurement on overhead lines. 34

35 Applications After wiring the measurement setup to the line under test proceed as follows: 1. Configure CPC 100 as described in 2.3 "Configuring CPC 100" on page 22 for the CP CU1 s current range set by the current range switch. Caution: The configured current range must not exceed the limit by the open-line voltage. 2. Choose the XML template for the mains frequency (e.g. "Line Imp CU1 60Hz.xmt" for the 60 Hz mains frequency) and open the template. Caution: Open the grounding switch at the near end before making the test and keep it open only during the measurement. Close the grounding switch after the test and before reconnecting the measurement setup. 3. Run the test procedure. The following measurements are performed: Line-to-line measurements: L1-L2, L1-L3, L2-L3 For each measurement, connect the I OUT and V SENSE inputs of CP CU1 to the corresponding CP GB1 s line studs. Line-to-ground-measurements: L1-E, L2-E, L3-E For each measurement, connect the I OUT and V SENSE inputs of CP CU1 to the corresponding CP GB1 s line studs. Zero-sequence impedance measurement: L1 L2 L3-E 4. If an overload of CPC 100 occurs, reduce the test current or set a lower current range and run the test procedure once again. Lower test currents at the two highest frequencies are recommended. 5. Save the test procedure as a file on CPC Download the test file from CPC 100 to the PC using CPC Explorer. 7. Load the test file into the Microsoft Excel template. The measurement results are displayed. The measurement of mutual coupling of two power lines is performed in a similar way. All three phases of both systems under test are connected together and grounded at the far end of the line. The current I I is fed into all the phases of the system I and the voltage U II is measured on the near end of the system II which is not grounded here. The mutual coupling factor for the relay II is then given by k M0 = (U II /I I )/Z LII (Eq. 3-7) 35

36 CP CU1 Reference Manual V Interpretation of Measurement Results To interpret the results of line impedance measurements correctly, you have to know details about the overhead line or power cable under test. You will find below some useful notes about how to interpret the measurement results. Usually, the resistive part of the line impedance is relatively constant over the L1-L2, L1-L3 and L2-L3 as well as L1-E, L2-E and L3-E measurements. If the measurement results differ considerably, typically contact problems are the reason. In some cases, the grounding switches at the far end of the line are not as good as necessary for the measurement. Additional grounding cables could help to avoid the contact problems. For the lines under test shorter than 5 km/3 miles, do not connect the V SENSE input of CP CU1 with the Kelvin clamps, but rather use additional clamps directly on the wires of the power line. The inductive part of the line impedance increases with the distance between the lines. This is documented by the measurement results stored in an example file delivered with the line impedance templates (see the marked results in Figure 3-6: "Measurement Results" on page 37). The measured overhead line with the shortest distance between the lines L1 and L3 is shown in Figure 3-7: "Measured Overhead Line" on page 38. Note: For each line of measurement results there is a separate overload indication top right on the CPC 100 s screen (or in the report) explained below. No overload indication means no overload during that step of the measurement sequence. Dotted overload indication means that there was an overload during that step of the measurement sequence but not all the time. Solid overload indication means a permanent overload during that step of the measurement sequence. 36

37 Applications Figure 3-6: Measurement Results 37

38 CP CU1 Reference Manual V 1.4 Figure 3-7: Measured Overhead Line L2 L1 L3 The L2-E measurement features the lowest X component because the line is very close to the ground wire. The X component of the L3-E measurement is decreased by a parallel system taking course close to L3 on the other side of the tower. Short-circuiting of the parallel system during the measurement would have increased the effect and would have lead to erroneous results because this is not the normal operating condition. The interpretation of the result of zerosequence impedance measurement is more difficult. Some methods for verifying it are shown below. Several intermediate results are hidden in the Microsoft Excel template due to their minor importance. However, by clicking on the "+" symbol on the left (see the lines 32 and 51 in Figure 3-6: "Measurement Results" on page 37) you can view the individual k factor results for each line. If the individual measurements are very close to each other and close to the overall k factor, the measurement is most likely correct. Broadly spread results with the average value close to the overall k factor value indicate a very asymmetrical line under test but the results are most likely correct. If the individual k factor measurements differ considerably from each other, the relay should be set to a value smaller than the average k factor to avoid zone overreaches. 38

39 Applications Another interesting effect can be observed when measuring power cables. If the screen or shield is very close to the conductors but the conductors are relatively wide from each other, the inductive part of the line-to-line measurements is higher than the inductive part of the line-to-ground measurements, resulting in a negative X component of the calculated impedance Z E. This seemingly strange result is explained as follows. Recalling (see "Why k Factor Measurement?" on page 30) that the zero-sequence impedance is given by Z 0 = Z 1 + 3Z E (Eq. 3-8) and hence Z E = (Z 0 Z 1 )/3 (Eq. 3-9) where Z 1 is the positive-sequence impedance and Z E is defined as a difference between the line-to-ground loop measurement and a half of the line-to-line loop measurement, the X component of Z E can become negative. 3.4 Ground Impedance and Step Voltage Measurement Introduction A good substation grounding system is crucial to prevent people injury and damage of equipment. International standards such as DIN VDE 0101/CENELEC HD637S1, IEEE Std or IEEE Std give guidelines how to measure such impedances. This application describes the measurement of large substations. Smaller grounding systems could be tested without connection to an existing overhead line or power cable and therefore the use of CP CU1 and CP GB1 would not be necessary. However, the procedure is the same when injecting the current into a long wire directly from the 6 A AC output. Anyway, when injecting the current into a test probe, special safety measures are required to avoid hazards when people approach the test probe. The current-voltage method as called in CENELEC HD637S1 or fall of potential method as called in IEEE standards is a good solution to measure the ground impedance of a substation. Before starting the test procedure, take one overhead line or power cable leaving the substation under test out of service and ground it at the far end. Feed a current via this power line into a remote ground. For larger substations, a distance of the remote ground of at least 5 km/3 miles is recommended, the minimum distance is 10 times the size of the grounding system. Then measure the voltages with a test probe at various distances 39

40 CP CU1 Reference Manual V 1.4 around the substation. If possible, choose the measurement points in a 90º angle (bird s-eye view) relative to the current path. In any case, avoid measuring close (<60º) to the current path. For an accurate estimate of the step voltage, set the measurement points as close as 1 m/3 ft to the substation and to each other. The step voltage is calculated for a certain fault current. Enter the highest possible fault current for the substation under test in the relevant field of the Microsoft Excel template. Measurement data at a large distance (typically three times the length of the substation) from the substation allow the calculation of the overall substation ground impedance Z ground as defined in VDE Performing Measurements To measure ground impedance and step voltage: 1. Connect the measurement setup to an overhead line or a power cable leading from the substation under test following 3.2 "Safety Instructions for Connecting CP CU1 to Power Lines" on page Short the three phases with the delivered three-lead cable as shown in Figure 1-6: "Shorting the Phases" on page Connect one pin of the CPC 100 s V1 AC input to the ground substation ground, the other pin to a test probe as shown in Figure 3-8: "Ground Impedance and Step Voltage Measurement" on page 41. The V SENSE input and the V1 AC output of CP CU1 are not used in this application. The voltage is measured directly using the V1 AC input of CPC

41 Applications Figure 3-8: Ground Impedance and Step Voltage Measurement 90º V1 AC IAC CPC 100 EXT. BOOSTER IAC IOUT CP CU1 BOOSTER CP GB1 The test procedure is controlled by templates available on the CPC Explorer CD-ROM shipped with your CP CU1 or in the customer area of the OMICRON electronics home page. For detailed information on the templates and instructions how to use them, see 3.1 "Template Usage" on page 25. Using the CPC 100 s Sequencer test card, the test procedure runs without user interaction, performing the recommended six measurements per test probe position. Caution: Do not touch the test probe without insulating gloves outside of the substation area. In case of a high-current ground fault within the substation during the test, considerably high voltages could arise in any wire connected to the substation and leading away from it. 41

42 CP CU1 Reference Manual V 1.4 After wiring the measurement setup to the line proceed as follows: 1. Configure CPC 100 as described in 2.3 "Configuring CPC 100" on page 22 for the CP CU1 s current range set by the current range switch. Caution: The configured current range must not exceed the limit by the open-line voltage. 2. Choose the XML template for the mains frequency (e.g. "Ground Imp CU1 60Hz.xmt" for the 60 Hz mains frequency) and open the template. 3. Select the "Enter Distance Here" card from the template. 4. Select "Save as Default" to reuse this card later on. 5. Stick the test probe into the ground at the specified distance from the substation and proceed from short to long distances. Recommended distances for the step voltage measurement are 1 to 15 m in 1 m steps (3 to 45 ft in 3 ft steps) in the first file and 15 to 30 m in 1 m steps (45 to 90 ft) in the second file. For the ground impedance measurement, 1, 2, 5, 10, 20, 50, 100, 150, 200 m and continuing in 100 m steps (5, 10, 20, 50, 100, 200, 500, 750, 1000 ft and continuing in 250 ft steps) seem to be a good choice. You can measure the distance for this application conveniently with a commercial GPS device. Note: Make sure that the measurement points take course in a 90º angle (bird s-eye view) relative to the current path and, if possible, avoid additional overhead lines or power cables as well as current paths. 6. Start the test card for the current test point. 7. Label the test card with the distance in units m or ft without blanks, e.g. "10m" or "30ft". 8. Add one Sequencer test card for every test point you want to measure. 9. Proceed with step 6 as long as you want to measure at another distance. 10.Save the test procedure as a file on CPC 100. Note: It is recommended to save at most 15 test cards in one file, but having more files is possible. 11.Download the test file(s) from CPC 100 to the PC using CPC Explorer. 12.Load the test file(s) into the Microsoft Excel template. The ground impedance and the step voltage are displayed as a function of the distance from the substation. Note: If there are more files, load one after another. 42

43 Applications Interpretation of Measurement Results Ground Impedance Measurement Due to the inductance of the feed-in line, a considerable part of the current I out injected into the ground does not flow back through the ground but through the ground wire or the line shield. This current I shield has to be subtracted from I out and, consequently, the ground impedance is given by Z ground = V meas /(I out I shield ) (Eq. 3-10) This effect is compensated by the current reduction factor r as defined in the CENELEC HD637 S :1999. A field in the XML template allows setting the current reduction factor between 0.01 and 1.00 (1.00 means no current compensation). For 110 kv overhead lines, the standard gives typical values of r = 0.98 for steel ground wires and down to 0.60 for steel/aluminum ground wires. For current feeding via power cables, the r factor can be as low as The effect of the current I shield can be eliminated by disconnecting the line shield or the ground wire of the feed-in line. If the disconnection is not possible, it is recommended to measure the current I shield with a clamp-on ammeter and to calculate the current reduction factor as r = 1 I shield /I out (Eq. 3-11) The example file delivered with the templates includes measurement results for a terrain approaching the optimum. Figure 3-9: "Ground Impedance vs. Distance for a Terrain Approaching Optimum" on page 44 shows a graph displayed after loading the example file into the Microsoft Excel template. The voltage between the grounding system under test and the test probe (and hence the ground impedance) increases slowly until it reaches an approximately constant value. This limit value corresponds to the ground impedance Z ground, the impedance of the substation against the "rest of the world". In the example considered, Z ground is approximately 60 mω. 43

44 CP CU1 Reference Manual V 1.4 Figure 3-9: Ground Impedance vs. Distance for a Terrain Approaching Optimum Figure 3-10: Ground Impedance vs. Distance for a Difficult Terrain In some cases, measurement results shows peaks and drops until an area free of buildings and buried conductors or pipes is reached. Until then, erroneous results can be obtained. Figure 3-10: "Ground Impedance vs. Distance for a Difficult Terrain" below shows typical measurement results obtained under such conditions. The graph shows a peak due to a street lamp close to the test probe. The street lamp was connected to a protective ground wire approaching the location of the remote grounding system. Voltage drops can be observed when the measurement points are set close to objects (i.e. towers of power lines leaving the substation under test) connected to the grounding system under test. In a difficult terrain, the measurement results may show no noticeable trend until values converging to a constant are obtained. Consequently, the terrain analysis on site is of highest importance. Setting the measurement points close to the line injecting the current results in erroneously high measured values. An area within the angle of at least 60º relative to the line should be avoided. 44

CP CU1. Reference Manual

CP CU1. Reference Manual CP CU1 Reference Manual CP CU1 Reference Manual Article Number: VESD0671 Manual Version: CPCU1.AE.5 OMICRON electronics 2010. All rights reserved. This manual is a publication of OMICRON electronics GmbH.

More information

CP Cu1. Advanced Test Equipment Rentals ATEC (2832) Multi-purpose coupling unit for CPC 100. Measurement System for

CP Cu1. Advanced Test Equipment Rentals ATEC (2832) Multi-purpose coupling unit for CPC 100. Measurement System for Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) CP Cu1 Multi-purpose coupling unit for CPC 100 Measurement System for Line Impedances and k-factors Mutual Coupling

More information

CP CU1. Coupling unit for line and ground testing

CP CU1. Coupling unit for line and ground testing CP CU1 Coupling unit for line and ground testing Line and ground test system CPC 100 The CPC 100 is a multifunctional test set for primary assets. When combined with the CP CU1 it covers the following

More information

RMO500 M I C R O O H M M E T E R

RMO500 M I C R O O H M M E T E R M I C R O O H M M E T E R Reference Manual IBEKO POWER AB Sweden 2 IBEKO POWER AB Manual Version:. MV.01 IBEKO POWER AB 2003 This Reference Manual is a publication of IBEKO POWER AB, 18 125 Lidingö, Sweden.

More information

Cable Identifying Generator LCI TX / LCI TX-440

Cable Identifying Generator LCI TX / LCI TX-440 Operation manual Cable Identifying Generator LCI TX / LCI TX-440 Mess- und Ortungstechnik Measuring and Locating Technologies Elektrizitätsnetze Power Networks Kommunikationsnetze Communication Networks

More information

External brake resistor AX2090-BW5x

External brake resistor AX2090-BW5x Documentation External brake resistor AX2090-BW5x Accessories for Beckhoff servo drive AX5000 Version: Date: 1.3 2018-03-14 Table of content Table of content 1 Foreword... 5 1.1 Notes on the documentation...

More information

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com # # AAC Clamp Meter. Instruction Manual

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com # # AAC Clamp Meter. Instruction Manual 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com #61-732 #61-736 400 AAC Clamp Meter Instruction Manual AC HOLD APO DC KMΩ mva WARNING Read First: Safety Information Understand

More information

200Amp AC Clamp Meter + NCV Model MA250

200Amp AC Clamp Meter + NCV Model MA250 User's Guide 200Amp AC Clamp Meter + NCV Model MA250 Introduction Congratulations on your purchase of this Extech MA250 Clamp Meter. This meter measures AC Current, AC/DC Voltage, Resistance, Capacitance,

More information

2011 / Circuit Tracer

2011 / Circuit Tracer INSTRUCTION MANUAL 2011 / 00521 Circuit Tracer Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. 52044992 2008 Greenlee Textron

More information

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL Table of Contents TITLE PAGE 1. GENERAL INSTRUCTIONS 1 1.1 Precaution safety measures 1 1.1.1 Preliminary 1 1.1.2 During use 2 1.1.3 Symbols

More information

Cisco Aironet 13.5-dBi Yagi Mast Mount Antenna (AIR-ANT1949)

Cisco Aironet 13.5-dBi Yagi Mast Mount Antenna (AIR-ANT1949) Cisco Aironet 13.5-dBi Yagi Mast Mount Antenna (AIR-ANT1949) Overview This document describes the 13.5-dBi Yagi mast mount antenna and provides instructions for mounting it. The antenna operates in the

More information

Electrical Measurement Safety. Sponsored By:

Electrical Measurement Safety. Sponsored By: Electrical Measurement Safety Sponsored By: About the Viewer Panel Slides: Go to the Links tab at the top and click on the link to download the PDF of the slides If you re watching the archive version,

More information

returned to your distributor for attention. 14. Do not operate the function switch while the instrument is connected to a circuit.

returned to your distributor for attention. 14. Do not operate the function switch while the instrument is connected to a circuit. 1. This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions. KYORITSU will not accept liability for any damage or injury caused by misuse

More information

Model 4210-MMPC-W. Multi-measurement Prober Cable Kit. Overview

Model 4210-MMPC-W. Multi-measurement Prober Cable Kit. Overview Model 4210-MMPC-W Keithley Instruments, Inc. Multi-measurement Prober Cable Kit 28775 urora Road Quick Start Guide Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com Overview The Keithley Instruments

More information

A New Technology. for Measuring Overhead Line and Cable Impedance Values and the Ground Impedance of Large Substations. Feature

A New Technology. for Measuring Overhead Line and Cable Impedance Values and the Ground Impedance of Large Substations. Feature Feature A New Technology for Measuring Overhead Line and Cable Impedance Values and the Ground Impedance of Large Substations Introduction The knowledge of overhead line and cable impedances is very important

More information

Agilent U1253B True RMS OLED Multimeter. Quick Start Guide

Agilent U1253B True RMS OLED Multimeter. Quick Start Guide Agilent U1253B True RMS OLED Multimeter Quick Start Guide The following items are included with your multimeter: Silicone test leads 4 mm probes Alligator clips Printed Quick Start Guide Rechargeable 8.4

More information

312, 316, 318. Clamp Meter. Users Manual

312, 316, 318. Clamp Meter. Users Manual 312, 316, 318 Clamp Meter Users Manual PN 1989445 July 2002 Rev.2, 2/06 2002, 2006 Fluke Corporation. All rights reserved. Printed in China. All product names are trademarks of their respective companies.

More information

Model 4210-MMPC-L. Multi-measurement Prober Cable Kit. Overview. Quick start guide topics. Related documents

Model 4210-MMPC-L. Multi-measurement Prober Cable Kit. Overview. Quick start guide topics. Related documents Model 0-MMPC-L Keithley Instruments, Inc. Multi-measurement Prober Cable Kit 877 Aurora Road Quick Start Guide Cleveland, Ohio 9-888-KEITHLEY http://www.keithley.com Overview The Keithley Instruments Model

More information

Keysight U1231A/ U1232A/U1233A Handheld Multimeter. Quick Start Guide

Keysight U1231A/ U1232A/U1233A Handheld Multimeter. Quick Start Guide Keysight UA/ UA/UA Handheld Multimeter Quick Start Guide Contacting Keysight www.keysight.com/find/assist (worldwide contact information for repair and service) Safety and EMC Information This meter is

More information

CPC 100. Multi-functional Primary Test System for Substation Commissioning and Maintenance

CPC 100. Multi-functional Primary Test System for Substation Commissioning and Maintenance CPC 100 Multi-functional Primary Test System for Substation Commissioning and Maintenance CPC 100 - The Revolutionary All-in-One Test System The patented test system replaces numerous individual testing

More information

User s Guide. 400A AC/DC Clamp Meter. Model MA220

User s Guide. 400A AC/DC Clamp Meter. Model MA220 User s Guide 400A AC/DC Clamp Meter Model MA220 Introduction Thank you for selecting the Extech MA200 AC/DC Clamp Meter. This meter measures AC/DC Current, AC/DC Voltage, Resistance, Capacitance, Frequency,

More information

Installation and Operational Instructions for ROBA -switch Type 017._00.2

Installation and Operational Instructions for ROBA -switch Type 017._00.2 OBA -switch Type 017._00.2 Guidelines on the Declaration of Conformity A conformity evaluation has been carried out for the product in terms of the EC Low Voltage Directive 2014/35/ EC and the EMC Directive

More information

SA 125 mixer amplifier. user manual

SA 125 mixer amplifier. user manual SA 125 mixer amplifier user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 27.10.2016,

More information

MODEL 3810/2 Line Impedance Stabilization Network

MODEL 3810/2 Line Impedance Stabilization Network EMC TEST SYSTEMS FEBRUARY 1996 REV C PN 399197 MODEL 3810/2 Line Impedance Stabilization Network OPERATION MANUAL USA P.O. Box 80589 Austin, Texas 78708-0589 2205 Kramer Lane, Austin, Texas 78758-4047

More information

Current Probes. User Manual

Current Probes. User Manual Current Probes User Manual ETS-Lindgren Inc. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall

More information

Technical manual. Microstep driver SMC11. NANOTEC ELECTRONIC GmbH & Co. KG Gewerbestraße 11 D Landsham near Munich, Germany

Technical manual. Microstep driver SMC11. NANOTEC ELECTRONIC GmbH & Co. KG Gewerbestraße 11 D Landsham near Munich, Germany Technical manual Microstep driver NANOTEC ELECTRONIC GmbH & Co. KG Gewerbestraße 11 D-85652 Landsham near Munich, Germany Tel. +49 (0)89-900 686-0 Fax +49 (0)89-900 686-50 info@nanotec.de Editorial Editorial

More information

Agilent U1231A/U1232A/U1233A Handheld Multimeter. Quick Start Guide

Agilent U1231A/U1232A/U1233A Handheld Multimeter. Quick Start Guide Agilent U3A/U3A/U33A Handheld Multimeter Quick Start Guide Verify that you received the following items in the shipment of your multimeter: One pair of red and black test leads Four.5 V AAA alkaline battery

More information

400Amp True RMS AC Clamp Meter + NCV

400Amp True RMS AC Clamp Meter + NCV User Guide 400Amp True RMS AC Clamp Meter + NCV Model MA410T Introduction Thank you for selecting the Extech MA410 Clamp Meter. This meter measures AC Current, AC/DC Voltage, Resistance, Capacitance, Frequency,

More information

USER MANUAL 600A AC Clamp Meter + NCV Model MA610

USER MANUAL 600A AC Clamp Meter + NCV Model MA610 USER MANUAL 600A AC Clamp Meter + NCV Model MA610 Additional User Manual Translations available at www.extech.com Introduction Thank you for selecting the Extech MA610 Clamp Meter. This meter measures

More information

MM V 10A ENGLISH. INSTRUCTION MANUAL Auto-Ranging DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE

MM V 10A ENGLISH. INSTRUCTION MANUAL Auto-Ranging DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE INSTRUCTION MANUAL Auto-Ranging Digital Multimeter MM400 DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE 600V 10A 40MΩ 2 GENERAL SPECIFICATIONS Klein Tools MM400 is an auto-ranging

More information

DM-45 Digital Multimeter

DM-45 Digital Multimeter INSTRUCTION MANUAL DM-45 Digital Multimeter Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Description The Greenlee DM-45 Digital

More information

Education & Training

Education & Training Distribution System Operator Certificate This program provides you with a proficient working knowledge in modern electric power distribution systems. These four classes are designed to walk students through

More information

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH AC/DC LOW CURRENT CLAMP-ON METER CM605 ENGLISH User Manual Statement of Compliance Chauvin Arnoux, Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated using standards and instruments

More information

Safety. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.

Safety. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present. 9305 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information. This symbol, adjacent to a terminal, indicates

More information

T/3000 T/3000. Substation Maintenance and Commissioning Test Equipment

T/3000 T/3000. Substation Maintenance and Commissioning Test Equipment T/3000 Substation Maintenance and Commissioning Test Equipment MULTI FUNCTION SYSTEM FOR TESTING SUBSTATION EQUIPMENT SUCH AS: CURRENT, VOLTAGE AND POWER TRANSFORMERS, ALL TYPE OF PROTECTION RELAYS, ENERGY

More information

User s Manual. Ground Resistance Clamp On Tester MODEL

User s Manual. Ground Resistance Clamp On Tester MODEL User s Manual Ground Resistance Clamp On Tester MODEL 382357 Warranty EXTECH INSTRUMENTS CORPORATION warrants the basic instrument to be free of defects in parts and workmanship for one year from date

More information

CPC 100. Multi-functional primary test system for substation commissioning and maintenance

CPC 100. Multi-functional primary test system for substation commissioning and maintenance CPC 100 Multi-functional primary test system for substation commissioning and maintenance CPC 100 The revolutionary all-in-one test system The patented test system replaces numerous individual testing

More information

INSTRUCTION MANUAL. IBRit - rf1 - usb PC - Station for wireless Data transmission. M e s s t e c h n i k. Messtechnik GmbH & Co.

INSTRUCTION MANUAL. IBRit - rf1 - usb PC - Station for wireless Data transmission. M e s s t e c h n i k. Messtechnik GmbH & Co. M e s s t e c h n i k INSTRUCTION MANUAL PC - Station for wireless Data transmission Document No. : D1F604 001 Version : April 2006 Copyright : IBR Messtechnik GmbH & Co. KG Contents 1. Introduction 1.1

More information

CL900. True RMS 1000V 2000A 60MΩ ENGLISH. INSTRUCTION MANUAL 2000A Digital Clamp Meter. Measurement Technology

CL900. True RMS 1000V 2000A 60MΩ ENGLISH. INSTRUCTION MANUAL 2000A Digital Clamp Meter. Measurement Technology ENGLISH INSTRUCTION MANUAL 2000A Digital Clamp Meter True RMS Measurement Technology NON-CONTACT VOLTAGE TESTING INRUSH CURRENT LOW IMPEDANCE DATA HOLD RANGE HOLD AUDIBLE CONTINUITY DIODE TEST CAPACITANCE

More information

PES Eddy Current Proximity Sensor User s Manual P/N M8A-100

PES Eddy Current Proximity Sensor User s Manual P/N M8A-100 PES -110 Eddy Current Proximity Sensor User s Manual P/N 9431-09M8A-100 Safety Information The following manual contains information and warnings. They must be followed in order to keep the instrument

More information

Digital Function Generator

Digital Function Generator Digital Function Generator 13654-99 PHYWE Systeme GmbH & Co. KG Robert-Bosch-Breite 10 37079 Göttingen Germany Tel. +49 (0) 551 604-0 Fax +49 (0) 551 604-107 E-mail info@phywe.de Operating Instructions

More information

POWER TRACE TM TASCO, INC.

POWER TRACE TM TASCO, INC. POWER TRACE TM Instruction Manual PTL610 TASCO, INC. HOW YOUR POWER TRACE TM OPERATES The Power Trace TM is composed of two primary components: the Receiver and the Transmitter. When the Transmitter is

More information

CONTENTS MS2033A. 1.Introduction Safety Information...01

CONTENTS MS2033A. 1.Introduction Safety Information...01 MS2033A AC Digital Clamp Meter User s Manual CAT III 600 V CONTENTS 1.Introduction...01 2.Safety Information...01 2.1 Precautions...02 2.2 Safety Symbols...03 3. Description...04 3.1 Front Panel...04 3.2

More information

MS8250A/B OPERATION MANUAL MS8250A. Hz% FUNC REL RANGE REL HOLD OFF 10A. Hz% A NCV. Hz% COM. A ma 10A FUSED 600V CAT IV.

MS8250A/B OPERATION MANUAL MS8250A. Hz% FUNC REL RANGE REL HOLD OFF 10A. Hz% A NCV. Hz% COM. A ma 10A FUSED 600V CAT IV. MS8250A/B DIGITAL MULTIMETER OPERATION MANUAL AUTO DC AC REL hfe PCLINK % C F kmωkz nµmfav MS8250A DIGITAL MULTIMETER Auto Power Off RANGE REL HOLD FUNC NCV A ma OFF 10A A ma 10A FUSED 600V CAT IV COM

More information

COOPER POWER SERIES. Upgrade instructions for CL-6 series control firmware. Voltage Regulators MN225071EN

COOPER POWER SERIES. Upgrade instructions for CL-6 series control firmware. Voltage Regulators MN225071EN Voltage Regulators MN225071EN Effective March 2017 Supersedes September 2012 (S225-11-4) COOPER POWER Upgrade instructions for CL-6 series control firmware SERIES DISCLAIMER OF WARRANTIES AND LIMITATION

More information

AT Advanced Wire Tracer. Users Manual

AT Advanced Wire Tracer. Users Manual AT-1000 Advanced Wire Tracer Users Manual AT-1000 Advanced Wire Tracer English Users Manual AT1000_Rev001 2008 Amprobe Test Tools. All rights reserved. Limited Warranty and Limitation of Liability Your

More information

DIGITAL MULTIMETER CONTENTS DIGITAL MULTIMETER CONTENTS

DIGITAL MULTIMETER CONTENTS DIGITAL MULTIMETER CONTENTS CONTENTS CONTENTS CONTENTS 1. SAFETY INFORMATION...1 1.1 Preliminary...1 1.2 Dos and don ts...2 1.3 Symbols...3 1.4 Precautions...4 2. DESCRIPTION...5 2.1 Names of parts...6 2.2 Switches, buttons and input

More information

SIMEAS-T. Operating Instructions Transducer without auxiliary power. 7KG6111 and 7KG6101. Operating Instructions

SIMEAS-T. Operating Instructions Transducer without auxiliary power. 7KG6111 and 7KG6101. Operating Instructions Operating Instructions SIMEAS-T s Operating Instructions Transducer without auxiliary power for alternating current for alternating voltage for alternating voltage with expanded end range 7KG6111 and 7KG6101

More information

Heritage MedCall. Sentry E-Call Model HM-527 Resident Host Panel

Heritage MedCall. Sentry E-Call Model HM-527 Resident Host Panel Heritage MedCall Sentry E-Call Model HM-527 Resident Host Panel 430-527B 0305 Heritage MedCall, Inc. Issue 1, March 2005 Heritage Medcall Sentry Emergency Call System Model 527 Host Panel Installation

More information

Agilent U1251B and U1252B Handheld Digital Multimeter. Quick Start Guide

Agilent U1251B and U1252B Handheld Digital Multimeter. Quick Start Guide Agilent U1251B and U1252B Handheld Digital Multimeter Quick Start Guide The following items are included with your multimeter: Silicone test leads 4 mm probes Alligator clips Printed Quick Start Guide

More information

ProfiScale MULTI Multimeter

ProfiScale MULTI Multimeter 1,5 V 9V 200 mv 600 V 200 ma 1/10 A ProfiScale MULTI Multimeter en Operating instructions BURG-WÄCHTER KG Altenhofer Weg 15 58300 Wetter Germany Introduction Want the reassurance of knowing whether current

More information

Remote Maintenance Box, RMB

Remote Maintenance Box, RMB OPERATOR S MANUAL Remote Maintenance Box, RMB Warnings and legal information Operating procedure Voltage measurement Document no.: 4189340730B SW version AGC 3.40.0 or later and AGC 4.00.0 or later Table

More information

Earthing of Electrical Devices and Safety

Earthing of Electrical Devices and Safety Earthing of Electrical Devices and Safety JOŽE PIHLER Faculty of Electrical Engineering and Computer Sciences University of Maribor Smetanova 17, 2000 Maribor SLOVENIA joze.pihler@um.si Abstract: - This

More information

MS2109A AC/DC Clamp Meter. User Manual. Contents

MS2109A AC/DC Clamp Meter. User Manual. Contents MS2109A AC/DC Clamp Meter User Manual Contents 1. Safety information 1 1.1 Preparation 1 1.2 Usage 1 1.3 Signs and Labels 2 1.4 Maintenance 2 2. Description 2 2.1 Part name 3 2.2 Switch and button description

More information

No. 139 Jinghai Third Road, BDA, Beijing, China, Office TEL. +86 (10) Website

No. 139 Jinghai Third Road, BDA, Beijing, China, Office TEL. +86 (10) Website PONOVO POWER CO., LTD No. 139 Jinghai Third Road, BDA, Beijing, China, 100176 Office TEL. +86 (10) 5908 9666 E-Mail Info@relaytest.com Website www.relaytest.com T1000 Series Universal Tester VERSION: DATE:

More information

User's Guide. 800 Amp AC/DC True RMS Clamp Meter. Model EX Washington Street Melrose, MA Phone Toll Free

User's Guide. 800 Amp AC/DC True RMS Clamp Meter. Model EX Washington Street Melrose, MA Phone Toll Free User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com 800 Amp AC/DC True RMS Clamp Meter Model EX730 Introduction Congratulations

More information

SPES Underwater Eddy Current Proximity Sensor User s Manual P/N M2A-107

SPES Underwater Eddy Current Proximity Sensor User s Manual P/N M2A-107 SPES -108 Underwater Eddy Current Proximity Sensor User s Manual P/N 9431-09M2A-107 Safety Information The following manual contains information and warnings. They must be followed in order to keep the

More information

DM-46 Instruction Manual

DM-46 Instruction Manual Auto Meter Products Inc. Test Equipment DM-46 Instruction Manual Automotive Multimeter and Inductive Amp Probe The DM-46 is the auto industry s answer to pocket portability in a 20 2650-1552-00 3/8/11

More information

E-800 power amplifier. user manual

E-800 power amplifier. user manual E-800 power amplifier user manual Musikhaus Thomann Thomann GmbH Hans-Thomann-Straße 1 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 20.05.2016,

More information

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No Portable Appliance Testers OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver. 1.2. Code No. 20 750 684 Distributor: Producer: METREL d.d. Ljubljanska 77 SI-1354 Horjul E-mail: metrel@metrel.si http://www.metrel.si

More information

MM700. True RMS ENGLISH. INSTRUCTION MANUAL Auto-Ranging. Measurement Technology

MM700. True RMS ENGLISH. INSTRUCTION MANUAL Auto-Ranging. Measurement Technology INSTRUCTION MANUAL Auto-Ranging Digital Multimeter er True RMS Measurement Technology MM700 DATA & RANGE HOLD LOW IMPEDANCE AUDIBLE CONTINUITY MIN / MAX / RELATIVE TEMPERATURE DIODE TEST CAPACITANCE &

More information

2015 RIGOL TECHNOLOGIES, INC.

2015 RIGOL TECHNOLOGIES, INC. Service Guide DG000 Series Dual-channel Function/Arbitrary Waveform Generator Oct. 205 TECHNOLOGIES, INC. Guaranty and Declaration Copyright 203 TECHNOLOGIES, INC. All Rights Reserved. Trademark Information

More information

DVM1190 DIGITAL MULTIMETER

DVM1190 DIGITAL MULTIMETER DIGITAL MULTIMETER 1. Introduction Thank you for buying the. This digital multimeter has a large LCD, a data-hold function and a backlight. The device uses a very practical safety mechanism that keeps

More information

E-800 power amplifier. user manual

E-800 power amplifier. user manual E-800 power amplifier user manual Musikhaus Thomann e.k. Treppendorf 30 96138 Burgebrach Germany Telephone: +49 (0) 9546 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 03.06.2013 Table of contents

More information

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation User s Manual 701932 Current Probe Yokogawa Electric Corporation 1st Edition Foreword Revisions Thank you for purchasing the 701932 Current Probe. This user's manual contains useful information about the

More information

GFL-1000 User Manual Ground Fault Locator

GFL-1000 User Manual Ground Fault Locator GFL-Series User Manual V1.1 GFL-1000 User Manual Ground Fault Locator Contents Contents... 1 1 Declaration of Conformity... 3 2 Introduction... 3 3 Equipment Information... 3 3.1 Safety Precautions...

More information

Type CP-S, CP-C & CP-A Switch mode

Type CP-S, CP-C & CP-A Switch mode Switch mode power CP-S, CP-C & CP-A Switch mode Characteristics CP-S and CP-C range Output current 5 A, 10 A and 20 A Integrated power reserve of up to 50 % 5 A and 10 A devices with pluggable connecting

More information

Independent Technology Service Inc Independence Ave. Chatsworth, California Toll Free:

Independent Technology Service Inc Independence Ave. Chatsworth, California Toll Free: Independent Technology Service Inc. 9182 Independence Ave. Chatsworth, California 91311 www.itscnc.com Toll Free: 1.800.342.3475 NEW Brush Amplifiers For Fadal Machines AMP-0006N-ITS AMP-0021N-ITS NEW

More information

TETRIS 1000 High Impedance Active Probe. Instruction Manual

TETRIS 1000 High Impedance Active Probe. Instruction Manual TETRIS 1000 High Impedance Active Probe Instruction Manual Copyright 2015 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications

More information

Operating Instructions PROFITEST H+E TECH. Diagnostics Unit for Electric Charging Stations (Type 2 Connector Socket and Plug) /3.

Operating Instructions PROFITEST H+E TECH. Diagnostics Unit for Electric Charging Stations (Type 2 Connector Socket and Plug) /3. Diagnostics Unit for Electric Charging Stations (Type 2 Connector Socket and Plug) 3-349-878-03 1/3.16 Opening the Instrument / Repairs The instrument may only be opened by authorized, trained personnel

More information

RAGU 81D DIGITAL MULTIMETER OPERATION MANUAL

RAGU 81D DIGITAL MULTIMETER OPERATION MANUAL RAGU 81D DIGITAL MULTIMETER OPERATION MANUAL Contents I. General...- 1 - Ⅱ. Open-package Inspection...- 2 - III. Safety Considerations... - 3 - IV.Instrument Panel & Button Function Description...- 9 -

More information

Compact Autoranging Clamp Meters. Models (400 A AC), (400 A AC/DC)

Compact Autoranging Clamp Meters. Models (400 A AC), (400 A AC/DC) User Manual Compact Autoranging Clamp Meters with NIST-Traceable Calibration Models 20250-55 (400 A AC), 20250-56 (400 A AC/DC) THE STANDARD IN PRECISION MEASUREMENT 1065DGMAN_20250-55,-56 DS Clamp Meter

More information

Keysight U1461A Insulation Multimeter/ U1453A Insulation Tester

Keysight U1461A Insulation Multimeter/ U1453A Insulation Tester Keysight U46A Insulation Multimeter/ U453A Insulation Tester Quick Start Guide Standard Accessories Included In Your Purchase Standard Accessories Included In Your Purchase The following accessories are

More information

C&G Level 3 Award in the Periodic Inspection, Testing and Certification of Electrical Installations. Earth Fault Loop Impedance Tests

C&G Level 3 Award in the Periodic Inspection, Testing and Certification of Electrical Installations. Earth Fault Loop Impedance Tests C&G 2395-01 Level 3 Award in the Periodic Inspection, Testing and Certification of Electrical Installations Earth Fault Loop Impedance Tests 1 Revision Inspections are made to verify that the installed

More information

VT1586A Rack Mount Terminal Panel Installation and User s Manual

VT1586A Rack Mount Terminal Panel Installation and User s Manual VT1586A Rack Mount Terminal Panel Installation and User s Manual Manual Part Number: 82-0095-000 Rev. June 16, 2003 Printed in U.S.A. Certification VXI Technology, Inc. certifies that this product met

More information

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español Owner's Manual AC/DC Clamp Meter Model No. 82369 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance Español

More information

PM-311x Quick Start Ver PM-311x introduction 1.1. Caution & Warning 1.2. Product Warranty & Customer Support

PM-311x Quick Start Ver PM-311x introduction 1.1. Caution & Warning 1.2. Product Warranty & Customer Support PM-311x Quick Start Ver. 1.1 1. PM-311x introduction ICP DAS brings the most powerful, cost-effective, advanced Smart Power Meters PM-3000 series that gives you access to real-time electric usage for single-phase

More information

MVCT. Megger VT & CT Analyzer. MVCT Megger VT & CT Analyzer

MVCT. Megger VT & CT Analyzer. MVCT Megger VT & CT Analyzer MVCT Ability to test both VT and CT Easy to use one-button automated test plans Industry leading test duration using patented simultaneous tap measurements Smallest and lightest unit on the market CT Kneepoints

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-920 The Service Information provides the following information: Precautions and safety information Specifications Performance test procedure Calibration

More information

2 Grounding of power supply system neutral

2 Grounding of power supply system neutral 2 Grounding of power supply system neutral 2.1 Introduction As we had seen in the previous chapter, grounding of supply system neutral fulfills two important functions. 1. It provides a reference for the

More information

PCT200 SERIES CT/PT TEST SYSTEM USER MANUAL

PCT200 SERIES CT/PT TEST SYSTEM USER MANUAL PONOVO POWER CO., LTD 2F, 4Cell, Tower C, In.Do Mansion No.48A Zhichun Road, Haidian District Beijing, China (Post Code 100098) TEL. +86 (10) 82755151 ext. 8887 Office FAX +86 (10) 82755257 E-Mail Info@ponovo.com.cn

More information

Central Inverter SUNNY CENTRAL 200 / 250 / 250HE / 350

Central Inverter SUNNY CENTRAL 200 / 250 / 250HE / 350 Central Inverter SUNNY CENTRAL 200 / 250 / 250HE / 350 Installation Guide SC20_25_35-IEN094521 98-4004821 Version 2.1 EN SMA Solar Technology AG Table of Contents Table of Contents 1 Notes on this Manual..............................

More information

Instruction Manual CT-6 High Frequency AC Current Probe

Instruction Manual CT-6 High Frequency AC Current Probe Instruction Manual CT-6 High Frequency AC Current Probe 071-0453-00 Revision A Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending.

More information

Bode 100. User Manual

Bode 100. User Manual Bode 100 User Manual Bode 100 User Manual Article Number VESD0661 - Manual Version: Bode100.AE.3 OMICRON Lab 2008. All rights reserved. This User Manual is a publication of OMICRON electronics GmbH. This

More information

TAG5000 WIRELESS PHASER. Instruction Manual HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS U.S.A.

TAG5000 WIRELESS PHASER. Instruction Manual HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS U.S.A. TAG5000 WIRELESS PHASER Instruction Manual TM HD ELECTRIC COMPANY 1475 LAKESIDE DRIVE WAUKEGAN, ILLINOIS 60085 U.S.A. PHONE 847.473.4980 FAX 847.473.4981 website: www.hdelectriccompany.com DESCRIPTION

More information

Keysight U1241C/ U1242C Handheld Digital Multimeter. Quick Start Guide

Keysight U1241C/ U1242C Handheld Digital Multimeter. Quick Start Guide Keysight U4C/ U4C Handheld Digital Multimeter Quick Start Guide Safety Notices CUTION CUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly

More information

Bulk Current Injection Probe Test Procedure

Bulk Current Injection Probe Test Procedure Bulk Current Injection Probe Test Procedure 1 TABLE OF CONTENTS INTRODUCTION 3 GENERAL INFORMATION 4 TEST METHODS 6 SAFETY 8 FIGURES 9 FORMULAS 12 MAINTENANCE 13 WARRANTY 14 2 INTRODUCTION CURRENT PROBE

More information

MS2030 CAT III 600 V A V AUTO RS232

MS2030 CAT III 600 V A V AUTO RS232 MS2030 AC Digital Clamp Meter User s Manual CAT III 600 V AUTO RS232 A V CONTENTS 1.Introduction...1 2.Safety Information...1 2.1 Precautions...1 2.2 Safety Symbols...3 3. Description...4 3.1 Front Panel...4

More information

Dissipation / Power Factor and Capacitance Measurement on Rotating Machines using the CPC100/TD1 and CR500

Dissipation / Power Factor and Capacitance Measurement on Rotating Machines using the CPC100/TD1 and CR500 Application Note Dissipation / Power Factor and Capacitance Measurement on Rotating Machines using the CPC100/TD1 and CR500 Author s Fabian Öttl Stefan Böhler Martin Anglhuber Fabian.oettl@omicronenergy.com

More information

TA MHz oscilloscope probe TA MHz oscilloscope probe

TA MHz oscilloscope probe TA MHz oscilloscope probe TA375 100 MHz oscilloscope probe TA386 200 MHz oscilloscope probe User's Guide X1 X10 TA386 X1/X10 Max. 600 Vp Introduction This passive high-impedance oscilloscope probe is suitable for most oscilloscopes

More information

Overvoltage Protection

Overvoltage Protection Overvoltage Protection S T U D E N T M A N U A L March 31, 2005 2 STUDENT TRAINING MANUAL Prerequisites: Single-Phase Transformer Load Checks Objectives: From memory, you will be able to describe the electrical

More information

BC145 SIGNAL ISOLATOR BOARD

BC145 SIGNAL ISOLATOR BOARD BC145 SIGNAL ISOLATOR BOARD 4/17 Installation & Operating Manual MN1373 Any trademarks used in this manual are the property of their respective owners. Important: Be sure to check www.baldor.com to download

More information

Installation Manual DOC R1 Date: 05/27/99. Plug-in Meter Installation Manual

Installation Manual DOC R1 Date: 05/27/99. Plug-in Meter Installation Manual Installation Manual DOC-4004-5015-R1 Date: 05/27/99 Plug-in Meter Installation Manual Document Number: Doc-4004-5015-R1 Document Release Date: 5/99 Copyright 1999 TeCom Inc. All rights reserved. No part

More information

Advanced Test Equipment Rentals ATEC (2832) CP RC. Resonance circuit for GIS testing

Advanced Test Equipment Rentals ATEC (2832) CP RC. Resonance circuit for GIS testing Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) CP RC Resonance circuit for GIS testing A new approach to testing gas-insulated switchgear Testing gas-insulated switchgear

More information

V1.3. TBLC08 50mH AC-LISN TBLC08

V1.3. TBLC08 50mH AC-LISN TBLC08 V1.3 TBLC08 The TBLC08 is a Line Impedance Stabilization Network for the measurement of line-conducted interference within the range of 9kHz to 30MHz, according to the CISPR16 standard. The device is designed

More information

MT /6 Smart Digital Clamp Meter. User s Manual 1 st Edition, 2016 Copyright by Prokit s Industries Co., Ltd.

MT /6 Smart Digital Clamp Meter. User s Manual 1 st Edition, 2016 Copyright by Prokit s Industries Co., Ltd. MT-3110 3 5/6 Smart Digital Clamp Meter User s Manual 1 st Edition, 2016 Copyright by Prokit s Industries Co., Ltd. 1.Safety Information Warnings Special attention shall be paid when using the meter, improper

More information

Power supply CP-E 24/2.5

Power supply CP-E 24/2.5 2CDC 271 015 F0t06 a OUTPUT L+, L : terminals output b DC OK: terminal signalling output c INPUT L, N, PE: terminals input d OUTPUT OK: green LED output voltage OK e OUTPUT Adjust: potentiometer adjustment

More information

Tube Facing Tool.

Tube Facing Tool. www.swagelok.com Tube Facing Tool This manual contains important information for the safe and effective operation of the Swagelok TF72 series tube facing tool. Users should read and understand its contents

More information

E-400 power amplifier. user manual

E-400 power amplifier. user manual E-400 power amplifier user manual Musikhaus Thomann e.k. Treppendorf 30 96138 Burgebrach Germany Telephone: (09546) 9223-0 E-mail: info@thomann.de Internet: www.thomann.de 07.03.2014, ID: 173888 Table

More information

INSTRUCTION MANUAL DIGITAL PSC-LOOP TESTER MODEL 4118A KYORITSU ELECTRICAL INSTRUMENTS WORKS,LTD.

INSTRUCTION MANUAL DIGITAL PSC-LOOP TESTER MODEL 4118A KYORITSU ELECTRICAL INSTRUMENTS WORKS,LTD. INSTRUCTION MANUAL DIGITAL PSC-LOOP TESTER MODEL 4118A KYORITSU ELECTRICAL INSTRUMENTS WORKS,LTD. CONTENTS 1. SAFE TESTING... 1 2. PROCEDURE OF REMOVING COVER... 4 3. FEATURES... 5 3.1 Instrument Layout...

More information