1 Pressure measurement Process pressure/hydrostatic VEGAWELL 5 Product Information

2 Contents Contents Description of the measuring principle Type overview Mounting instructions Electrical connection 4. General requirements Power supply Connection cable Cable screening and grounding Wiring plan VEGAWELL ma Wiring plan VEGAWELL ma/hart - Pt Operation 5. Overview Adjustment with PACTware Technical data Dimensions Product code Process pressure/hydrostatic Pressure measurement

3 Contents Take note of safety instructions for Ex applications Please note the Ex specific safety information which you can find on our homepage and which comes with every instrument. In hazardous areas you should take note of the appropriate regulations, conformity and type approval certificates of the sensors and power supply units. The sensors must only be operated on intrinsically safe circuits. The permissible electrical values are stated in the certificate. Process pressure/hydrostatic Pressure measurement

4 Description of the measuring principle Description of the measuring principle Measuring principle VEGAWELL 5 pressure transmitters work according to the hydrostatic measuring principle, which functions independently of the dielectric properties of the product and is not influenced by foam generation. The sensor element of VEGAWELL 5 is the dry ceramic-capacitive CERTEC measuring cell in two sizes. Base element and diaphragm consist of high purity sapphire-ceramic. The hydrostatic pressure of the product causes via the diaphragm a capacitance change in the measuring cell. This capacitance change is converted into an appropriate output signal. 6 µm Fig. : Configuration of the CERTEC measuring cell with VEGAWELL 5 Diaphragm Soldered glass bond Base element The advantages of the CERTEC measuring cell are: l Very high overload resistance l No hysteresis l Excellent long-term stability l Completely front flush installation l Good corrosion resistance l Very high abrasion resistance Wide application range VEGAWELL 5 is suitable for level measurement in deep wells and ballast tanks as well as for gauge measurement in open flumes. Typical media are drinking water and waste water as well as water containing abrasive substances. All signal outputs are available in 4 0 ma and 4 0 ma/hart - Pt 00. In the 4 0 ma/hart - Pt 00 version, a temperature sensor Pt 00 in four-wire technology is integrated in the transducer. Power supply or processing are carried out via an external temperature transducer. 4 Process pressure/hydrostatic Pressure measurement

5 Type overview Type overview VEGAWELL 5 Measuring cell: CERTEC Media: Process fitting: drinking water and waste water Material process fitting: 6L Material, suspension cable: PE, PUR, FEP Material transmitter: Diameter transmitter: Measuring range: Process temperature: Straining clamp, screw connection, thread 6L,.446 (Duplex), each also with PE coating, PVDF, Titanium depending on material and version at least mm 0 0. bar up to 0 5 bar C ( F) Deviation: < 0. %, < 0. % Signal output: Operation: 4 0 ma, 4 0 ma/hart depending on the version via PACTware/PC Process pressure/hydrostatic Pressure measurement 5

6 Mounting instructions Mounting instructions Mounting position The following illustration shows a mounting example for VEGA- WELL 5. The VEGA price list contains suitable mounting brackets under the section Accessories. With these parts, standard mounting arrangements can be realised quickly and reliably. Mounting with straining clamp Fig. : Straining clamp Suspension cable Suspension opening Clamping jaws Mounting with screw connection Fig. : VEGAWELL 5 in a pump shaft with VEGABOX 0 VEGAWELL 5 must be mounted in a calm area or in a suitable protective tube. This avoids lateral movements of the transmitter and the resulting corruption of measurement data. Note: As an alternative to fixing the transmitter, the use of a measuring instrument holder from VEGA's line of mounting accessories is recommended. Beside the connection and suspension cables, the suspension cable also contains a capillary for atmospheric pressure compensation. All versions can be shortened on site. With VEGAWELL 5, the electronics is completely integrated in the transmitter. The cable end can be lead directly to a dry connection compartment. Pressure compensation is then carried out via the filter element of the capillaries. Note: The pressure compensation housing VEGABOX 0 is recommended for connecting VEGAWELL 5. It contains a high-quality ventilation filter and terminals. A protective cover is optionally available for use outdoors. Fig. 4: Screw connection Suspension cable Seal screw Cone bushing 4 Seal cone 5 Screw connection 6 Seal Mounting versions The following illustrations show the different mounting versions depending on the instrument type. 6 Process pressure/hydrostatic Pressure measurement

7 Mounting instructions Mounting with housing and thread Fig. 5: Housing with thread G½ A Housing Seal Thread Process pressure/hydrostatic Pressure measurement 7

8 Electrical connection 4 Electrical connection 4. General requirements The supply voltage range can differ depending on the instrument version. You can find exact specifications in chapter "Technical data". The national installation standards as well as the valid safety regulations and accident prevention rules must be observed. 4.5 Wiring plan VEGAWELL ma Direct connection 4 In hazardous areas you should take note of the appropriate regulations, conformity and type approval certificates of the sensors and power supply units. Fig. 6: Wire assignment, suspension cable 4. Power supply Supply voltage and current signal are carried on the same twowire cable. The requirements on the power supply are specified in chapter "Technical data". blue (-): to power supply or to the processing system brown (+): to power supply or to the processing system Shielding 4 Breather capillaries with filter element Connection via VEGABOX 0 The VEGA power supply units VEGATRENN 49AEx, VEGAS- TAB 690, VEGADIS 7 as well as VEGAMET signal conditioning instruments are suitable for power supply. When one of these instruments is used, a reliable separation of the supply circuits from the mains circuits according to DIN VDE 006 part 0 is ensured Connection cable In general An outer diameterof 5 9 mm ensures the seal effect of the cable entry. If electromagnetic interference is expected, screened cable should be used for the signal lines. Fig. 7: Terminal assignment VEGABOX 0 To power supply or the processing system Shielding ) The sensors are connected with standard two-wire cable without screen. In Ex applications, the corresponding installation regulations must be noted for the connection cable. 4.4 Cable screening and grounding If screened cable is necessary, the cable screen must be connected on both ends to ground potential. If potential equalisation currents are expected, the connection on the evaluation side must be made via a ceramic capacitor (e.g. nf, 500 V). ) Connect screen to ground terminal. Connect ground terminal on the outside of the housing as prescribed. The two terminals are galvanically connected. 8 Process pressure/hydrostatic Pressure measurement

9 Electrical connection Connection via housing Connection via VEGABOX Fig. 0: Terminal assignment VEGABOX 0 To power supply or the processing system (signal pressure transmitter) To power supply or the processing system (connection cables resistance thermometer Pt 00) Shielding ) Connection via VEGABOX 0 with integrated temperature sensor Fig. 8: Terminal assignment of the housing To power supply or the processing system Shielding ) 4.6 Wiring plan VEGAWELL ma/ HART - Pt 00 Direct connection ma Fig. : Terminal assignment VEGABOX 0 To power supply or the processing system (signal pressure transmitter) For voltage supply or to processing system (resistance thermometer Pt 00) Shielding 4) Fig. 9: Wire assignment, connection cable Blue (-): to power supply or to the processing system Brown (+): to power supply or to the processing system White: for processing of the integrated Pt 00 (power supply) 4 Yellow: for processing of the integrated Pt 00 (measurement) 5 Red: for processing of the integrated Pt 00 (measurement) 6 Black: for processing of the integrated Pt 00 (power supply) 7 Shielding 8 Breather capillaries with filter element ) ) 4) Connect screen to ground terminal. Connect ground terminal on the outside of the housing as prescribed. The two terminals are galvanically connected. Connect screen to ground terminal. Connect ground terminal on the outside of the housing as prescribed. The two terminals are galvanically connected. Connect screen to ground terminal. Connect ground terminal on the outside of the housing as prescribed. The two terminals are galvanically connected. Process pressure/hydrostatic Pressure measurement 9

10 6 Electrical connection Connection via housing 4 5 Fig. : Terminal assignment of the housing To power supply or the processing system (signal pressure transmitter) For voltage supply or to processing system (resistance thermometer Pt 00) Shielding 5) 5) Connect screen to ground terminal. Connect ground terminal on the outside of the housing as prescribed. The two terminals are galvanically connected. 0 Process pressure/hydrostatic Pressure measurement

11 Operation 5 Operation 5. Overview VEGAWELL ma VEGAWELL ma has no adjustment options. VEGAWELL ma/hart - Pt 00 l Adjustment software according to FDT/DTM standard, e.g. PACTware and PC l HART handheld 5. Adjustment with PACTware Connecting the PC to the signal cable 4 5 Fig. : Connection of the PC to VEGABOX 0 or communication resistor PC with PACTware RS interface (with VEGACONNECT ), USB interface (with VEGA- CONNECT 4) VEGACONNECT or 4 4 Communication resistor 50 Ω 5 Power supply unit Necessary components: l VEGAWELL 5 l PC with PACTware and suitable VEGA DTM l VEGACONNECT with HART adapter cable l HART resistor approx. 50 Ω l Power supply unit Note: With power supply units with integrated HART resistance (internal resistance approx. 50 Ω), an additional external resistance is not necessary (e. g. VEGATRENN 49A, VEGAMET 8/64/65, VEGASCAN 69). In such cases, VEGACONNECT can be connected parallel to the 4 0 ma cable. Process pressure/hydrostatic Pressure measurement

12 Technical data 6 Technical data Materials and weights Materials, wetted parts - Transmitter 6L, 6L with PE coating,.446 (Duplex),.446 with PE coating, PVDF, Titanium - Diaphragm sapphire ceramic (99.9 % oxide ceramic) - Measuring cell seal FKM (VP/A) - FDA and KTW approved, FFKM (Perlast G75S), EPDM (A+P 75.5/KW75F) - Suspension cable PE (FDA and KTW-approved), FEP, PUR - Cable gland on the transmitter 6L - Process fitting 6L - Straining clamp.40 - Unassembled screw connection 6L, PVDF - Threaded connection on the housing 6L Materials, non-wetted parts - Housing plastic PBT (Polyester), 6L Weight approx. - Basic weight 0.8 kg (.764 lbs) - Suspension cable 0. kg/m (0.07 lbs/ft) - Straining clamp 0. kg (0.44 lbs) - Screw connection 0.4 kg (0.88 lbs) - Plastic housing 0.8 kg (.764 lbs) - Stainless steel housing.6 kg (.58 lbs) Input variable Measured value Level Measuring range see product code Recommended max. turn down 0 : Output variable 4 0 ma Output signal 4 0 ma Signal resolution µa Failure signal <.6 ma Max. output current ma Run-up time s Step response time 00 ms (ti: 0 s, 0 6 %) Fulfilled NAMUR recommendations NE ma/hart - Pt 00 Output signal 4 0 ma/hart Signal resolution µa Failure signal <.6 ma; 0.5 ma; ma; unchanged (adjustable via PACTware) Max. output current ma Run-up time 5 s Step response time 00 ms (ti: 0 s, 0 6 %) Fulfilled NAMUR recommendations NE 4 Additional output parameter - temperature integrated resistance thermometer Pt 00 according to DIN EN 6075 Range C (-58 + F) Resolution K Deviation for 4 0 ma version 6) 6) Specifications refer to the set span. Turn down (TD) = nominal measuring range/set span. Determined according to the limit point method according to IEC 60770, incl. non-linearity, hysteresis and non-repeatability. Process pressure/hydrostatic Pressure measurement

13 Technical data Deviation with version < 0. % - Turn down : up to 5 : < 0. % - Turn down > 0 : < 0.04 % x TD Deviation with version < 0. % - Turn down : up to 5 : < 0. % - Turn down > 0 : < 0.0 % x TD Deviation for version 4 0 ma/hart - Pt 00 7) Applies to digital HART interface as well as to analogue current output 4 0 ma. Specifications refer to the set span. Turn down (TD) is the relation nominal measuring range/set span. Deviation with version < 0. % - Turn down : up to 5 : < 0. % - Turn down > 0 : < 0.04 % x TD Deviation with version < 0. % - Turn down : up to 5 : < 0. % - Turn down > 0 : < 0.0 % x TD Influence of the product or ambient temperature Applies to digital HART interface as well as to analogue current output 4 0 ma. Specifications refer to the set span. Turn down (TD) is the relation nominal measuring range/set span. Average temperature coefficient of the zero signal In the compensated temperature range of C (+ +76 F), reference temperature 0 C (68 F). Average temperature coefficient of the zero signal - Turn down : < 0.05 %/0 K - Turn down : up to 5 : < 0. %/0 K - Turn down > 0 : < 0.5 %/0 K Outside the compensated temperature range Average temperature coefficient of the zero signal - Turn down : typ. < 0.05 %/0 K Long-term stability (similar to DIN 6086, DINV 959- and IEC ) Applies to digital HART interface as well as to analogue current output 4 0 ma. Specifications refer to the set span. Turn down (TD) is the relation nominal measuring range/set span. Long-term drift of the zero signal < (0. % x TD)/year Ambient conditions Ambient temperature - Connection cable PE C ( F) - Connection cable PUR, FEP C ( F) Storage and transport temperature C ( F) 7) Determined according to the limit point method according to IEC 60770, incl. non-linearity, hysteresis and non-repeatability. Process pressure/hydrostatic Pressure measurement

14 Technical data Process conditions Process pressure Max. process pressure, transmitter 8) - Measuring range 0. bar (.45 psig) 5 bar (8 psig) - Measuring range 0. bar (.9 psig) 0 bar (90 psig) - Measuring range 0.4 bar (5.8 psig) 5 bar (6 psig)pressure stage, process fitting - Unassembled screw connection 6L: PN, PVDF: unpressurized - Thread on the housing PN Product temperature, depending on the version Transmitter All All PE coating All PE coating Product temperature C ( F) C ( F) C ( F) C ( F) C ( F) Vibration resistance mechanical vibrations with 4 g and 5 00 Hz 9) Electromechanical data Suspension cable - Configuration six wires, one suspension cable, one breather capillary, screen braiding, foil, mantle - Tensile strength 00 N (70 pound force) - Max. length 000 m (80 ft) - Min. bending radius 5 mm (with 5 C/77 F) - Diameter approx. 8 mm (0.5 in) - colour (non-ex/ex) - PE black/blue - colour (non-ex/ex) - PUR, FEP blue/blue Cable entry housing or VEGABOX 0 x cable gland M0 x.5 (cable: ø 5 9 mm), x blind stopper M0 x.5 Screw terminals for wire cross section.5 mm² (AWG 6), screen up to 4 mm² (AWG ) Supply voltage ma Operating voltage 8 6 V DC Permissible residual ripple - < 00 Hz U ss < V - 00 Hz 0 khz U ss < 0 mv Load see diagram 8) 9) Limited by the overpressure resistance of the measuring cell. Tested according to the regulations of German Lloyd, GL directive. 4 Process pressure/hydrostatic Pressure measurement

15 Technical data Ω V Fig. 4: Voltage diagram Voltage limit Operating voltage Supply voltage ma/hart - Pt 00 Operating voltage V DC Permissible residual ripple - < 00 Hz U ss < V - 00 Hz 0 khz U ss < 0 mv Load see diagram Ω , V Fig. 5: Voltage diagram HART load Voltage limit Operating voltage Electrical protective measures Protection - Transmitter IP 68 (0 bar) - Housing IP 66/IP 67 - VEGABOX 0 IP 65 Overvoltage category III Protection class III Process pressure/hydrostatic Pressure measurement 5

16 Technical data Existing approvals or approvals applied for Gas explosion protection Fire-damp protection Overfill protection Ship approval e.g. according to ATEX and IEC e.g. according to ATEX e.g. according to WHG e.g. according to GL, LRS, ABS, RINA The available approvals can be selected via the configurator on Depending on the version, instruments with approvals can have different technical data. For these instruments, please note the corresponding approval documents. They can be downloaded in the download section on CE conformity EMC (004/08/EG) EN 66-: 006 LVD (006/95/EG) EN 600-: 00 Environmental instructions VEGA environment management system certified according to DIN EN ISO 400 You can find detailed information under 6 Process pressure/hydrostatic Pressure measurement

17 L L L Technical data 7 Dimensions VEGAWELL 5 - suspension cable VEGAWELL 5 - suspension cable ~ 69 mm (.7") ø 77 mm (.0") 48 mm (.89") mm (0.87") 75mm (6.89") 5mm (.05") 48mm (.89") mm (0.87") 8,5mm 4mm (.5") (0.55") G ½A/ NPT ½ 67mm (.64") 0mm (0.79") mm (4.4") mm (0.87") SW46 G ½A/ NPT ½ M0x,5/ ½ NPT SW46 (.8") G ½A/ NPT ½ 6 L 75 mm (6.89") 5 mm (.05") ø 8 mm (0.") 8,5 mm 4 mm (.5") (0.55") SW0 (.8") G ½A/ NPT ½ ø 8mm (0.") 0 mm (9.06") ø mm (0.87") 0 mm (9.06") ø mm (0.87") 06mm (8.") ø mm (.6") 7mm (8.54") ø mm (.6") 0mm (9.06") ø 6mm (.4") 8mm (8.58") ø 44mm (.7") 0mm (8.7") 50mm (9.84") (*) ø mm (.0") 4 5 Fig. 6: VEGAWELL 5 - suspension cable Transmitter Duplex, with straining clamp TransmitterDuplex for deep wells, with unassembledscrew connectiong½ A (½ NPT) and closing cap Transmitter Duplex, with PE coating 4 Transmitter with screwed connection of PVDF 5 TransmitterTitanium/Titaniumwith glass leadthrough, with threadg A ( NPT) and plastic housing Fig. 7: VEGAWELL 5 - suspension cable Transmitter 6L, with straining clamp Transmitter Titanium, with unassembled screw connection G A ( NPT) VEGAWELL 5 - threaded fitting ø 8 mm (.89") ø 8 mm (.89") 4 mm (9.49") ø mm (.6") ø 40 mm (.58") 0 mm (9.06") ø mm (.6") ø 40 mm (.58") 74 mm (.9") 0 mm (0.79") mm (0.") SW 7 mm (.06") G½ 45 mm (.77") SW 4 mm (.6") G G¼ ø 7,5 mm (0.69") 8 mm (0.7") Fig. 8: VEGAWELL 5 - thread Threaded fitting G½ inner G¼ Threaded fitting G Process pressure/hydrostatic Pressure measurement 7

18 Product code 8 Product code VEGAWELL 5 Approval XX without XM Ship approval AX ATEX II G EEx ia IIC T6 AM ATEX II G EEx ia IIC T6 + Ship approval AI IEC Ex ia IIC T6 Fastening / Material X4 without A4 Straining clamp /.40(04) GA Threaded fitting, unassembled G½A PN / 6L NP Threaded fitting, unassembled G½A PN0, / PVDF GC Threaded fitting, unassembled GA PN / 6L GK Thread G½A PN / 6L with plastic housing GV Thread G½A PN / 6L w.hous. StSt (precision casting) Version / Process temperature A Suspension cable PE / C D Suspension cable PUR / C B Suspension cable FEP / C Length K 6 m suspension cable PE L m suspension cable PE M 7 m suspension cable PE T individually selectable length (PE/PUR/FEP) Transmitter material / Diameter D Duplex.446 / mm V 6L / mm K Duplex.446 with PE coating / 5mm P PVDF / 44 mm Seal measuring cell FKM (VP/A) EPDM (A+P 75.5/KW75F) P FFKM (Perlast G75S) Measuring range A rel. / bar (0...0 kpa) B rel. / bar (0...0 kpa) C rel. / bar ( kpa) D rel. / 0... bar ( kpa) E rel. / bar ( kpa) F rel. / bar ( kpa) G rel. / bar ( kpa) abs bar (0...50kPa) abs bar ( kPa) Electronics C 4...0mA D 4...0mA/HART + PT00 4-wire Deviation in characteristic Transmitter options X without V for deep wells WL5. 8 Process pressure/hydrostatic Pressure measurement

19 Process pressure/hydrostatic Pressure measurement 9

20 Ohmart/VEGA 470 Rosslyn Drive Cincinnati, Ohio 4509 USA Telephone Fax You can find at downloads of the following l operating instructions manuals l menu schematics l software l certificates l approvals and much, much more Subject to change without prior notice

Pressure measurement. Process pressure/hydrostatic VEGAWELL 52. Product Information

Pressure measurement. Process pressure/hydrostatic VEGAWELL 52. Product Information Pressure measurement Process pressure/hydrostatic VEGAWELL 5 Product Information Contents Contents Description of the measuring principle........................................................... Type

More information

Pressure measurement. Process pressure/hydrostatic VEGAWELL 72. Product Information

Pressure measurement. Process pressure/hydrostatic VEGAWELL 72. Product Information Pressure measurement Process pressure/hydrostatic VEGAWELL 7 Product Information Content Content Description of the measuring principle........................................................... Type overview..............................................................................

More information

OEM radar transmitter, for aggressive media level measurement

OEM radar transmitter, for aggressive media level measurement OEM radar transmitter, for aggressive media level measurement C ompact for level measurement up to 20 m 4... 20 ma/hart - 2 wires Adjustable with P C ATEX approvals Type 8136 can be combined with... Type

More information

Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm

Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm Electronic pressure measurement Process transmitter Model UPT-20, with pressure port Model UPT-21, with flush diaphragm WIKA data sheet PE 86.05 Applications Process technology Machine building and plant

More information

Submersible pressure sensor For oils and fuels Model LF-1

Submersible pressure sensor For oils and fuels Model LF-1 Level measurement Submersible pressure sensor For oils and fuels Model LF-1 WIKA data sheet LM 40.04 Applications Level measurement in vessel and storage systems for oils and fuels Overfilling and dry-run

More information

vega-line price list 2011/12

vega-line price list 2011/12 vega-line price list 0/ Measurement technology for level, switching, pressure vega-line Standard solutions Quickly and price-favourable VEGA world leader in measurement technology VEGA is a world leader

More information

Submersible pressure sensor For superior applications Model LF-1

Submersible pressure sensor For superior applications Model LF-1 Level Submersible pressure sensor For superior applications Model LF-1 WIKA data sheet LM 40.04 Applications Level measurement in vessel and storage systems Overfilling and no-load operation monitoring

More information

Submersible pressure sensor For water and wastewater Model LW-1

Submersible pressure sensor For water and wastewater Model LW-1 Replacement product: Model LF-1 Level measurement Submersible pressure sensor For water and wastewater Model LW-1 WIKA data sheet LM 40.03 Applications Level measurement in rivers and lakes Deep well and

More information

Assembly 249 (9.80) 259 (10.2) Example: level probe with Pt100 and HUT temperature head transmitter ( ma) ma. Core colors: RD BK YE BU

Assembly 249 (9.80) 259 (10.2) Example: level probe with Pt100 and HUT temperature head transmitter ( ma) ma. Core colors: RD BK YE BU Level probe Features Assembly Device for hydrostatic pressure measurement in liquids High resistance to overload High-precision, robust ceramic measuring cell with longterm stability Simultaneous level

More information

Ultrasonic level measuring device, non-contact

Ultrasonic level measuring device, non-contact Ultrasonic level measuring device, non-contact For level measurement up to 8 m 4... 20 ma/hart - 2 wires Suitable for solids ATEX approvals Type 8177 can be combined with... Type 8611 Type 8793 Type 8802-GB

More information

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. appmeas.co.uk +44 (0)

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. appmeas.co.uk +44 (0) DMP 4 Industrial Pressure Transmitter Without Media Isolation accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 0 mbar up to 0... 000 mbar Product characteristics excellent linearity

More information

Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted

Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted Electrical temperature measurement Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted WIKA data sheet TE 60.34 Applications Machine building, plant and vessel construction

More information

Process pressure transmitter Model IPT-10, standard version Model IPT-11, flush diaphragm

Process pressure transmitter Model IPT-10, standard version Model IPT-11, flush diaphragm Electronic pressure measurement Process pressure transmitter Model IPT-10, standard version Model IPT-11, flush diaphragm WIKA data sheet PE 86.11 Applications Process engineering Pharmaceutical industry

More information

Rod probe With coaxial tube reference probe

Rod probe With coaxial tube reference probe Capacitive level transmitters GENERAL DESCRIPTION NIVOCAP 2-wire capacitive level transmitters provide an ideal solution for level measurement of conductive or non-conductive liquids. The probe of the

More information

SALES SERVICE +358 (0)

SALES SERVICE +358 (0) SALES SERVICE +358 (0)17 262 3555 info@econosto.fi ADCATROL ELECTRO-PNEUMATIC POSITIONERS PE 986 (ATEX) DESCRIPTION The ADCATROL PE986 positioner requires an input signal of 4 20 ma for proportional control

More information

DS 400 DS 400. Intelligent Electronic Pressure Switch Completely in Stainless Steel. piezoresistive stainless steel sensor

DS 400 DS 400. Intelligent Electronic Pressure Switch Completely in Stainless Steel. piezoresistive stainless steel sensor GmbH Postal address: House address Fon (+49) 0 22 42-870-0 Postbox 26 Löhestr. 7 Fax (+49) 0 22 42-870-20 http: // www.tematec.de 5759 Hennef 577 Hennef e-mail: team@tematec.de Data Sheet Page /5 Intelligent

More information

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure

DMP 343. Industrial Pressure Transmitter. Without Media Isolation. accuracy according to IEC 60770: 0.35 % FSO. Nominal pressure DMP 4 Industrial Pressure Transmitter Without Media Isolation accuracy according to IEC 60770: 0.5 % FSO Nominal pressure from 0... 0 mbar up to 0... 000 mbar Product characteristics excellent linearity

More information

FOR CLEAN WATER AND SEWAGE WATER HYDROSTATIC LEVEL TRANSMITTERS

FOR CLEAN WATER AND SEWAGE WATER HYDROSTATIC LEVEL TRANSMITTERS FOR CLEAN WATER AND SEWAGE WATER HYDROSTATIC LEVEL TRANSMITTERS O U R P R O F E S S I O N NIVOPRESS N HYDROSTATIC LEVEL TRANSMITTERS FOR CLEAN WATER AND SEWAGE APPLICATIONS FEATURES Measuring range up

More information

ED701 General Industry Pressure Transmitter

ED701 General Industry Pressure Transmitter ED701 General Industry Pressure Transmitter Standard industrial process connections Complete range of electrical connections 4... 20 ma and Voltage outputs Accuracy: 0.1%, 0.2% and 0.4% FS Quick response

More information

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F Replacement product: Model IS-3 Electronic pressure measurement Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F WIKA data sheet PE

More information

MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400

MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400 Data Sheet MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400 Submersible Hydrostatic s Rugged and versatile We reserve the right to continuously improve our products and make any change in

More information

Level detection in liquid. Conductive. VEGAKON 61 VEGAKON 66 Probes EL 1, 3, 4, 6, 8. Product Information

Level detection in liquid. Conductive. VEGAKON 61 VEGAKON 66 Probes EL 1, 3, 4, 6, 8. Product Information Level detection in liquid Conductive VEGAKON 6 VEGAKON 66 Probes EL,, 4, 6, 8 Product Information Content Content Description of the measuring principle...........................................................

More information

Intrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F, IS-21-F

Intrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F, IS-21-F Replacement product: Model IS-3 Electronic Pressure Measurement Intrinsically Safe Pressure Transmitter for applications in hazardous environments and shipbuilding industry Model IS-20-S, IS-21-S, IS-20-F,

More information

Analogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting

Analogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting Electrical Temperature Measurement Analogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting WIKA Data Sheet TE 24.01 Applications T Machinery, plant construction T Process industry

More information

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F

Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F Replacement product: Model IS-3 Electronic pressure measurement Intrinsically safe pressure transmitter For applications in hazardous areas Models IS-20-S, IS-21-S, IS-20-F, IS-21-F WIKA data sheet PE

More information

Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version

Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version Electrical temperature measurement Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version WIKA data sheet TE 15.01 Applications Process industry Machine building

More information

Features. Picture. Specifications

Features. Picture. Specifications Type: for level measurement Order No.: see Table 1 / Page 5 Features Picture Two wire system Piezoresistive measuring element Output signal 4-20 ma, adjustable within 1 : 4 of the nominal pressure Conformity

More information

Digital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version

Digital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version Temperature Digital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version WIKA data sheet TE 15.01 for further approvals see page 10 Applications

More information

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted Electrical temperature measurement OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted WIKA data sheet TE 60.31 further approvals see page 10 Applications Machine building, plant

More information

For clean water and sewage water NIVOPRESS N HYDROSTATIC LEVEL TRANSMITTERS LEVEL TRANSMITTERS

For clean water and sewage water NIVOPRESS N HYDROSTATIC LEVEL TRANSMITTERS LEVEL TRANSMITTERS For clean water and sewage water NIVOPRESS N HYDROSTATIC LEVEL TRANSMITTERS O U R P R O F E S S I O N I S Y O U R L E V E L LEVEL TRANSMITTERS O U R P R O F E S S I O N NIVOPRESS N HYDROSTATIC LEVEL TRANSMITTERS

More information

3DLevelScanner. Technical Data Sheet. Level Sensor. Non-Contact, Multiple-Point Measurement

3DLevelScanner. Technical Data Sheet. Level Sensor. Non-Contact, Multiple-Point Measurement Technical Data Sheet Non-Contact, Multiple-Point Measurement The 3DLevelScanner is a noncontact, dust-penetrating bin measurement device that detects irregular surfaces in powder or solid materials and

More information

Resistance thermometer Model TR30, compact design

Resistance thermometer Model TR30, compact design Electrical temperature measurement Resistance thermometer Model TR30, compact design WIKA data sheet TE 60.30 Applications Machine building, plant and vessel construction Propulsion technology, hydraulics

More information

OPTIFLUX 5000 Technical Datasheet

OPTIFLUX 5000 Technical Datasheet OPTIFLUX 5000 Technical Datasheet Electromagnetic flowmeter in flanged version Exceptional long-term stability and accuracy For highly aggressive and abrasive fluids Fully vacuum-resistant with high-tech

More information

3DLevelScanner. Technical Data Sheet. Level Sensor. Non-Contact, Multiple-Point Measurement

3DLevelScanner. Technical Data Sheet. Level Sensor. Non-Contact, Multiple-Point Measurement Technical Data Sheet Distribué par : Contact : hvssystem@hvssystem.com Tél : 0326824929 Fax : 0326851908 Non-Contact, Multiple-Point Measurement The 3DLevelScanner is a noncontact, dust-penetrating bin

More information

SITRANS T measuring instruments for temperature

SITRANS T measuring instruments for temperature Overview Our field devices for heavy industrial use HART, Universal 4 to 20 ma, universal Field indicator for 4 to 20 ma signals The temperature transmitter SITRANS TF works where others feel uncomfortable.

More information

Digital Temperature Transmitters Model T12.10, Universally Programmable, Head Mounting Model T12.30, Universally Programmable, Rail Mounting

Digital Temperature Transmitters Model T12.10, Universally Programmable, Head Mounting Model T12.30, Universally Programmable, Rail Mounting Electrical Temperature Measurement Digital Temperature Transmitters Model T12.10, Universally Programmable, Head Mounting Model T12.30, Universally Programmable, Rail Mounting WIKA Data Sheet TE 12.03

More information

Level and Pressure. Product Information. Vibrating level switches

Level and Pressure. Product Information. Vibrating level switches Level and Pressure Product Information Vibrating level switches Contents Contents 1 Product description 1.1 VEGAVIB... 3 1.2 VEGASWING... 3 2 Function and application 2.1 Functional principle... 5 2.2

More information

Operating Instructions

Operating Instructions Level and Pressure Operating Instructions VEGATOR 620, 621, 622 max. 0 10 min. 0 10 on VEGATOR 622! 6 7 8 9 10 11 12 13 14 in out Contents Contents Safety information... 2 Note Ex area... 2 1 Product description

More information

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751 Data sheet Intrinsically safe pressure transmitter MBS 420, MBS 425, MBS 470 and MBS 475 The intrinsically safe pressure transmitter program is designed for use in hazardous environments and offers a reliable

More information

Absolute pressure gauge with output signal For the process industry Models APGT and APGT43.160, NS 100 and 160

Absolute pressure gauge with output signal For the process industry Models APGT and APGT43.160, NS 100 and 160 Mechatronic pressure measurement Absolute pressure gauge with output signal For the process industry Models APGT43.100 and APGT43.160, NS 100 and 160 k Applications for further approvals see page 4 WIKA

More information

Level measurement in liquids. Radar VEGAPULS 61 VEGAPULS 62 VEGAPULS 63 VEGAPULS 65 VEGAPULS 66. Product Information

Level measurement in liquids. Radar VEGAPULS 61 VEGAPULS 62 VEGAPULS 63 VEGAPULS 65 VEGAPULS 66. Product Information Level measurement in liquids Radar VEGAPULS 6 VEGAPULS 62 VEGAPULS 63 VEGAPULS 65 VEGAPULS 66 Product Information Content Content Application, function, configuration..............................................................

More information

PRODUCT CATALOGUE HYDROSTATIC PROBES SCREW-IN TRANSMITTERS

PRODUCT CATALOGUE HYDROSTATIC PROBES SCREW-IN TRANSMITTERS PRODUCT CATALOGUE HYDROSTATIC PROBES SCREW-IN TRANSMITTERS PRESSURE AT THE HIGHEST LEVEL Successful medium-sized companies are not successful because they are active in many areas, but rather because they

More information

Product information. Guided Wave Radar. Level and interface measurement in liquids VEGAFLEX 81 VEGAFLEX 83 VEGAFLEX 86. Document ID: 46597

Product information. Guided Wave Radar. Level and interface measurement in liquids VEGAFLEX 81 VEGAFLEX 83 VEGAFLEX 86. Document ID: 46597 Product information Guided Wave Radar Level and interface measurement in liquids VEGAFLEX 8 VEGAFLEX 8 VEGAFLEX 86 Document ID: 697 Contents Contents Measuring principle... Type overview... Instrument

More information

Positioner with HART Communication Type 3780

Positioner with HART Communication Type 3780 Positioner with HART Communication Type 780 Application Single-acting or double-acting positioner for attachment to pneumatic control valves. Supplied with an electric input signal from to 20 ma Travels

More information

Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters

Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters ll 2G EEx ia llc T6 As our products are continuously improved, we reserve the right to make any change in the stated specifications

More information

0 2 1 â Ó. itemp HART TMT182. Technical Information

0 2 1 â Ó. itemp HART TMT182. Technical Information Technical Information itemp HART TMT182 Temperature head transmitter Head transmitter for resistance thermometers (RTD), thermocouples (TC), resistance and voltage transmitters (mv), HART -protocol, for

More information

Temperature transmitter itemp PCP DIN rail TMT 121

Temperature transmitter itemp PCP DIN rail TMT 121 Technical information TI 087R/09/en 510 03671 Temperature transmitter itemp PCP DIN rail TMT 121 Universal temperature transmitter for resistance thermometers (RTD), thermocouples, resistance and voltage

More information

ATM.ECO - Analog Pressure Transmitter with Temperature Compensation

ATM.ECO - Analog Pressure Transmitter with Temperature Compensation Pressure Transmitter ATM.ECO - Analog Pressure Transmitter with Temperature Compensation CUSTOMER BENEFITS Entry level series into precision pressure measurement High measurement accuracy is ensured by

More information

Connection. Input EEx ia IIC. Sense. Fine tuning adjustment. Fine tuning amplifying. Amplifying. Adjustment. Output. Composition

Connection. Input EEx ia IIC. Sense. Fine tuning adjustment. Fine tuning amplifying. Amplifying. Adjustment. Output. Composition Converter Connection 1-channel EEx ia IIC for 4-wire and 6-wire bridges Analogue output 0/4 ma... 20 ma or 4mA...-12mA Full bridge load cells and strain gauges circuit for resistance bridges up to 17 Ω

More information

Temperature Transmitter itemp RTD DIN rail TMT 127

Temperature Transmitter itemp RTD DIN rail TMT 127 Technical Information TI 095R/09/en No. 51004790 Temperature Transmitter itemp RTD DIN rail TMT 127 Temperature Transmitter for Pt100 Resistance Thermometer Application Temperature transmitter with fixed

More information

Design and applications

Design and applications Design and applications The measuring device operates largely independent of viscosity and is suitable for indicating the flow rate of water, acids, alkaline solutions and gases. Every device is calibrated

More information

RMA42. Technical Information

RMA42. Technical Information Technical Information RMA42 Process transmitter with control unit Digital process transmitter for mounting on a top-hat rail for monitoring and visualizing analog measured values Application Plant and

More information

OPTIWAVE 1010 Technical Datasheet

OPTIWAVE 1010 Technical Datasheet OPTIWAVE 1010 Technical Datasheet Radar (FMCW) Level Transmitter for bypass chambers and magnetic level indicators (BM 26 Advanced) Device welded to a bypass chamber with an optional IP68 level indicator

More information

Glass-fibre-reinforced polycarbonate. Dead zone Adaptive or fixed from % 1.5 s adjustable. 20 ms. 60 ms. 60 ms. Approx. 1.

Glass-fibre-reinforced polycarbonate. Dead zone Adaptive or fixed from % 1.5 s adjustable. 20 ms. 60 ms. 60 ms. Approx. 1. General data Mounting On linear actuators On quarter turn actuators ARCA-integrated or to VDI/VDE 3847-1 or IEC 534-6 (NAMUR) Range of stroke: 3... 130 mm Integrated to VDI/VDE 3847-2 or VDI/VDE 3845 Angle

More information

Miniature resistance thermometer Model TR33, thread-mounted

Miniature resistance thermometer Model TR33, thread-mounted Electrical temperature measurement Miniature resistance thermometer Model TR33, thread-mounted WIKA data sheet TE 60.33 Applications Machine building, plant and vessel construction Propulsion technology,

More information

Temperature transmitter itemp RTD DIN rail TMT 127

Temperature transmitter itemp RTD DIN rail TMT 127 Technical information TI 095R/24/ae Nr. 510 06165 Temperature transmitter itemp RTD DIN rail TMT 127 Pt100 transmitter for an economical, high accuracy temperature monitoring. Fixed range measurement,

More information

0 2 1 â Ó. itemp HART TMT182. Technical Information

0 2 1 â Ó. itemp HART TMT182. Technical Information Technical Information itemp HART TMT182 Temperature head transmitter Head transmitter for resistance thermometers (RTD), thermocouples (TC), resistance and voltage transmitters (mv), HART -protocol, for

More information

Operating Instruction

Operating Instruction Level and Pressure Operating Instruction Capacitive electrodes EK 4 20 ma - compact Safety information Safety information The described module must only be installed and operated as described in this operating

More information

Pressure transmitter for general industrial applications Model A-10

Pressure transmitter for general industrial applications Model A-10 Electronic pressure measurement Pressure transmitter for general industrial applications Model A-10 WIKA data sheet PE 81.60 Applications Machine building Machine tools Measurement and control technology

More information

Float Switches with Permanent Magnet For Vertical Installation Model RSM

Float Switches with Permanent Magnet For Vertical Installation Model RSM Level Measurement Float Switches with Permanent Magnet For Vertical Installation Model RSM WIKA Data Sheet LM 30.01 Applications Level measurement for almost all liquid media Pump/level control and monitoring

More information

OPTIFLUX 5000 Technical Datasheet

OPTIFLUX 5000 Technical Datasheet OPTIFLUX 5000 Technical Datasheet Electromagnetic flowmeter in flanged version Exceptional long-term stability and accuracy For highly aggressive and abrasive fluids Fully vacuum-resistant with high-tech

More information

Resistance thermometer Model TR30, compact design

Resistance thermometer Model TR30, compact design Electrical temperature measurement Resistance thermometer Model TR30, compact design WIKA data sheet TE 60.30 Applications Machine building, plant and vessel construction Propulsion technology, hydraulics

More information

itemp HART DIN rail TMT112

itemp HART DIN rail TMT112 Temperature transmitter itemp HART DIN rail TMT112 Universal temperature transmitter for resistance thermometers (RTD), thermocouples, resistance and voltage transmitters, incorporating HART protocol Application

More information

Pressure transmitter for general industrial applications Model A-10

Pressure transmitter for general industrial applications Model A-10 Electronic pressure measurement Pressure transmitter for general industrial applications Model A-10 WIKA data sheet PE 81.60 Applications Machine building Machine tools Measurement and control technology

More information

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted Temperature OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted WIKA data sheet TE 60.31 further approvals see page 11 Applications Machine building, plant and vessel construction

More information

Pressure transmitter For general industrial applications Model A-10

Pressure transmitter For general industrial applications Model A-10 Electronic pressure measurement Pressure transmitter For general industrial applications Model A-10 WIKA data sheet PE 81.60 Applications Machine building Ship building Measurement and control technology

More information

KFS. Flap-type flow meter. Design and range of application

KFS. Flap-type flow meter. Design and range of application Design and range of application The measuring device operates largely independent of viscosity and is suitable for indicating the flow rate of water, acids, alkaline solutions and gases. Every device is

More information

Intrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments Model IS-20-H

Intrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments Model IS-20-H Replacement product: Model IS-3 Electronic Pressure Measurement Intrinsically Safe Pressure Transmitter for highest pressure applications in hazardous environments WIKA Data Sheet PE 81.51 Applications

More information

prosonic T FMU230, FMU231

prosonic T FMU230, FMU231 Technical Information prosonic T FMU230, FMU231 Ultrasonic Level Measurement Compact transmitters for non-contact level measurement of fluids and coarse bulk materials Application The compact transmitters

More information

Transducer transmitter BILT 4 II 2(1) G. Technical Manual

Transducer transmitter BILT 4 II 2(1) G. Technical Manual GB Transducer transmitter BILT 4 II 2(1) G Technical Manual Transducer transmitter BILT 4 Contents Introduction General... 3 Ex.safety description... 4 Technical data... 5 Installation General... 6 Mechanical

More information

Operating Instruction

Operating Instruction Level and Pressure Operating Instruction VEGASEL 643 6 5 + + 3 5 + + max. min. on VEGASEL 643 5 6 7 8 9 10 11 12 13 14 in out Contents Contents Safety information... 2 1 Product description 1.1 Function

More information

Electrical Temperature Measurement

Electrical Temperature Measurement Electrical Temperature Measurement Resistance Thermometers Model TR10-B, for Additional Thermowell Applications Machinery, plant and tank construction Energy and power plant technology Chemical industry

More information

SATRON VB Pressure Transmitter

SATRON VB Pressure Transmitter SATRON VB pressure transmitter belongs to the series V transmitters. SATRON VB is userfriendly, through the ball valve mounted transmitter which is used for 04 kpa... 03 MPa ranges. The transmitter communicates

More information

Operating Instruction

Operating Instruction Level and Pressure Operating Instruction Capacitive electrodes EK with signal conditioning instrument Safety information Safety information The described module must only be installed and operated as described

More information

MHPS MHPS. Modular pressure transmitter. Technical documentation. Table of content. Characteristics - applications - technical data

MHPS MHPS. Modular pressure transmitter. Technical documentation. Table of content. Characteristics - applications - technical data Technical documentation Table of content Page 2: Page 3: Page 4: Page 5: Page 6: Page 7: Page 8: Characteristics - applications - technical data Technical data - input quantity - output quantity Electrical

More information

Pressure transmitter For general industrial applications Model A-10

Pressure transmitter For general industrial applications Model A-10 Electronic pressure measurement Pressure transmitter For general industrial applications Model A-10 WIKA data sheet PE 81.60 for further approvals see page 9 Applications Machine building Shipbuilding

More information

Digital inductive conductivity transmitter

Digital inductive conductivity transmitter Digital inductive conductivity transmitter Optimal solution for conductivity measurements in difficult fluids (polluted, dirty,...) PEEK/PPA version for CIP applications Large range of process connections

More information

Electromagnetic Flow Transmitter

Electromagnetic Flow Transmitter Electromagnetic Flow Transmitter Sensor in solid state technology Working as a transmitter and/or as an On/Off controller Automatic-calibration of full scale: Teach-In Clean in place (CIP) Type can be

More information

Glass-fibre-reinforced polycarbonate. Dead zone Adaptive or fixed from % 20 ms. 60 ms. 60 ms. Approx. 1.3 kg. Approx. 0.9 kg. Approx. 5.

Glass-fibre-reinforced polycarbonate. Dead zone Adaptive or fixed from % 20 ms. 60 ms. 60 ms. Approx. 1.3 kg. Approx. 0.9 kg. Approx. 5. General data Mounting On linear actuators On quarter turn actuators ARCA-integrated or to VDI/VDE 3847-1 or IEC 534-6 (NAMUR) Range of stroke: 3... 130 mm Integrated to VDI/VDE 3847-2 or VDI/VDE 3845 Angle

More information

HART field temperature transmitter Models TIF50, TIF52, TIF62

HART field temperature transmitter Models TIF50, TIF52, TIF62 Electrical temperature measurement HART field temperature transmitter Models TIF50, TIF52, TIF62 WIKA data sheet TE 62.01 Applications Plant construction Process engineering General industrial applications

More information

Mechatronic pressure measurement. Applications. Special features. Description

Mechatronic pressure measurement. Applications. Special features. Description Mechatronic pressure measurement Diaphragm pressure gauge with electrical output signal For the process industry High overload safety up to the 10-fold full scale value, max. 40 bar Models PGT43.100 and

More information

DTM.OCS.S/N - Modbus Level Transmitter

DTM.OCS.S/N - Modbus Level Transmitter Digital Submersible Level Transmitter with Modbus Interface DTM.OCS.S/N - Modbus Level Transmitter CUSTOMER BENEFITS High precision digital level transmitter with standard Modbus interface Fast customization

More information

Pressure transmitter with flameproof enclosure For applications in explosion-protected areas Models E-10 and E-11

Pressure transmitter with flameproof enclosure For applications in explosion-protected areas Models E-10 and E-11 Electronic pressure measurement Pressure transmitter with flameproof enclosure For applications in explosion-protected areas Models E-10 and E-11 WIKA data sheet PE 81.27 for further approvals see page

More information

Temperature head transmitter itemp PCP TMT 181

Temperature head transmitter itemp PCP TMT 181 Technical information TI 070R/09/en Mat.-Nr.: 510 01110 Temperature head transmitter itemp PCP TMT 181 Universal head transmitter for resistance thermometers (RTD), thermocouples, resistance and voltage

More information

Resistance Thermometers Model Series TR7X0, Sheathed Design

Resistance Thermometers Model Series TR7X0, Sheathed Design Electrical Temperature Measurement Resistance Thermometers Model Series TR7X0, Sheathed Design WIKA Data Sheet TE 60.40 Applications T Suitable for all industrial and laboratory applications Special Features

More information

Picture. Digital output signal (pressure and temperature) RS-485 Modbus RTU, 9600 bps Protocol details see Data Sheet SW

Picture. Digital output signal (pressure and temperature) RS-485 Modbus RTU, 9600 bps Protocol details see Data Sheet SW Type: for level measurement Order No.: see Table 1 / Page 4 Features Picture Four wire system Piezoresistive measuring element Output signal 4-20 m and RS-485 nalogue output adjustable within 1 : 4 of

More information

Technical Information RMA42

Technical Information RMA42 TI00150R/09/EN/05.16 71315973 Products Solutions Services Technical Information RMA42 Process transmitter with control unit Digital process transmitter for monitoring and visualizing analog measured values

More information

TT2780. Temperature head transmitter For resistance thermometers Pt100, settable using a PC, for installation in a sensor head Form B, BUZ, BUZ-H

TT2780. Temperature head transmitter For resistance thermometers Pt100, settable using a PC, for installation in a sensor head Form B, BUZ, BUZ-H Postal address: House address Fon (+49) 0 22 42-8703-0 Postbox 1261 Löhestr. 37 Fax (+49) 0 22 42-8703-20 http: // www.tematec.de 53759 Hennef 53773 Hennef e-mail: team@tematec.de TT2780 Data Sheet TT2780

More information

Temperature head transmitter itemp TC TMT 188

Temperature head transmitter itemp TC TMT 188 Technical information TI 077R/24/ae No.: 510 06145 Temperature head transmitter itemp TC TMT 188 Universal head transmitter for thermocouples for an economical, high accuracy temperature monitoring. Fixed

More information

SITRANS P. Transmitters for level. Introduction 1/42. Application

SITRANS P. Transmitters for level. Introduction 1/42. Application Introduction HK series S series Fig. / transmitters for level, with built-in analog indicator or digital display Application The transmitter with mounting flange measures the level of corrosive and non-corrosive

More information

omnigrad T - TSM 480, TR 480. Technical Information. RTD Temperature Sensor Hygienic design TSM 480 with electronics programmable via PC

omnigrad T - TSM 480, TR 480. Technical Information. RTD Temperature Sensor Hygienic design TSM 480 with electronics programmable via PC Technical Information omnigrad T - TSM 480, TR 480 RTD Temperature Sensor Hygienic design TSM 480 with electronics programmable via PC Application areas The onmigrad T easytemp TSM 480 and TR 480 thermometers

More information

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted Electrical temperature measurement OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted WIKA data sheet TE 60.31 Applications Machine building, plant and vessel construction Propulsion

More information

OEM pressure transmitter for general industrial applications Models O-10 (T), O-10 (5)

OEM pressure transmitter for general industrial applications Models O-10 (T), O-10 (5) Electronic pressure measurement OEM pressure transmitter for general industrial applications Models O-10 (T), O-10 (5) WIKA data sheet PE 81.65 Applications Hydraulics and pneumatics Pumps and compressors

More information

Incremental encoders Blind hollow shaft or cone shaft pulses per revolution

Incremental encoders Blind hollow shaft or cone shaft pulses per revolution Features TTL output driver for cable length up to 550 m Very high resistance to shock and vibrations Hybrid bearing for extended service life Shaft insulation up to.8 kv Large terminal box, turn by 80

More information

Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters

Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters Class l, Division 1 Group A-D IIC T4 As our products are continuously improved, we reserve the right to make any change in the stated

More information

Level measurement Radar

Level measurement Radar Level measurement Radar Overview 6 VEGAPULS series 60 8 Dimensions 20 AL-EN - 5 Level measurement Radar VEGAPULS: Non-contact measurement of liquids and solids under arduous process conditions Measuring

More information

Coriolis Massflowmeter

Coriolis Massflowmeter Coriolis Massflowmeter for liquids and gas measuring monitoring analysing Measuring range: 0-0.8 kg / h 0-65 000 kg / h water Accuracy: ± 0.1 of reading ±zero-point stability (liquids) pmax: PN40 tmax:

More information

Level Measurement. Continuous level measurement - Radar transmitters SITRANS LR400 5/229. Overview

Level Measurement. Continuous level measurement - Radar transmitters SITRANS LR400 5/229. Overview Overview Application It provides excellent results on low dielectric media. is available for standard applications and for applications that require explosion proof protection. features robust enclosure,

More information

â Ó itemp TC TMT128 Technical information DIN rail mounted Temperature Transmitter for Thermocouples (TC)

â Ó itemp TC TMT128 Technical information DIN rail mounted Temperature Transmitter for Thermocouples (TC) Technical information itemp TC TMT128 DIN rail mounted Temperature Transmitter for Thermocouples (TC) Application Temperature transmitter with fixed measuring ranges for converting TC input signals into

More information

Resistance thermometer Model TR30, compact design

Resistance thermometer Model TR30, compact design Temperature Resistance thermometer Model TR30, compact design WIKA data sheet TE 60.30 for further approvals see page 8 Applications Machine building, plant and vessel construction Propulsion technology,

More information