Fallstricke präziser DC- Messungen

Size: px
Start display at page:

Download "Fallstricke präziser DC- Messungen"

Transcription

1 Fallstricke präziser DC- Messungen Sascha Egger, Applications Engineer Group Leader National Instruments Switzerland GmbH

2 Agenda Overview of Precision Test Systems Techniques for: Low-voltage measurements Low-resistance measurements Low-current measurements High-impedance measurements Sweeps Conclusion: Measurement Considerations 2

3 What are Precision Measurements? Precision explicitly means: Reproducibility/repeatability Yield same results each time measurement is taken To man people, it also implies: Accuracy the overall uncertainty of a measurement Resolution number of bits used by the ADC (or DAC) to represent a signal Sensitivity the smallest signal detectable by a particular instrument on its most sensitive range 3

4 What are Precision Measurements? With modern instrumentation, these parameters usually come into question for the following: Voltage under 10 µv Current under 10 µa Resistance under 10 mω A precision measurement system should offer good accuracy, resolution, sensitivity, and precision for measurements under/of? these values. 4

5 Precision Measurement Examples Dielectric absorption Electrochemical Leakage Low-current semiconductor Light measurements Connector testing Surface insulation resistance Resistivity Charge measurements Temperature Contact resistance Superconductor resistance NI CONFIDENTIAL 5

6 Challenges in Precision Measurements Cost/Specialty of measurement equipment Environmental challenges Test times Cabling Channel count This presentation will provide a case study for building a precision DC test system in PXI and point out some pitfalls. 6

7 Case Study Precision Device Under Test (DUT) Reed relay SRC devices form a relay Tests to be performed: Contact resistance Coil resistance Thermal offset Coil pickup voltage, current Coil dropout voltage, current Open contact resistance Settling time 7

8 Relay Specifications 150 mω 10 GΩ-1 TΩ 500 us Ω 3.75 V, 0.8 V 8

9 Equipment PXI-1042 chassis and PXI-8119 controller: Measurements: o PXI ½-digit DMM o PXI wire guard and current amplifier Source: o PXI-4110 programmable DC power supply Switch: o PXI x16 reed relay matrix o PXI channel SPST relays External 100V DC-DC converter or SMU PXI

10 System Architecture 10

11 Low-Voltage Measurement Devices Data acquisition modules: X Series DAQ: 16 bits, sensitivity as low as 15 µv Absolute accuracy as low as 291 ppm (1 V range) Digital multimeters (DMM): PXI-407x FlexDMMs: bits, sensitivity as low as 10 nv Absolute accuracy as low as 28 ppm (wide temperature? range, 2-year calibration, 100 mv range) 11

12 Calibration Instruments must be calibrated at regular intervals. Benefits of calibration: Reduced measurement errors Consistency between measurements Assurance you are making accurate measurements Calibration procedures: Self-calibration Manual calibration procedures NI calibration services 12

13 NI-DMM Driver API, examples and help will be installed with NI-DMM driver. Supported platforms: LabVIEW & LabVIEW Real-Time LabWindows CVI & LabWindows CVI Real-Time Microsoft Visual C++ Microsoft Visual Basic

14 Soft Front Panels for Modular Instruments DMM Soft Front Panel will be installed with NI-DMM Driver. 14

15 Low-Voltage Measurements Major concern #1: Offset voltages Thermal offsets develop from junction points in the system (similar to a thermocouple). Voltmeter or DMM input offsets 15

16 Low-Voltage Measurements Major concern #2: Noise Johnson noise Magnetic fields Ground loops Johnson noise e nr = 1.28 x10-10 * Rf Twisted-pair wires to reduce noise 16

17 Thermal Offset 17

18 Low-Resistance and In-Circuit Measurement Devices Digital multimeters: PXI-407x FlexDMMs: bits, sensitivity as low as 10 µω Absolute accuracy as low as 60 ppm (wide temperature? range, 2-year calibration, 100 Ω range) 2-wire and 4-wire capability In-circuit guard amplifier accessory: PXI wire guard and current amplifier: enables 6-wire resistance measurements ±200 µv guard accuracy, typical 18

19 Low-Resistance Measurements Major concern #1: Lead resistance 2-wire measurement 4-wire measurement 19

20 Low-Resistance Measurements Major concern #2: Thermal EMFs Current reversal method Offset-compensated ohms V OCO = V M1 - V M2 = (I S R X + V THERMAL ) - V THERMAL = I S R X therefore: R X = V OCO /I S 20

21 Contact Resistance 21

22 Low-Level Resistance Measurements with DMM (4-Wire Mode) Internal current source (1mA) External current source (100mA) 22

23 Low-Level Resistance Measurements with a DMM and Additional Current Source External current source 23

24 Low-Current Measurement Devices Digital multimeters (shunt ammeters): PXI-407x FlexDMMs: bits, sensitivity as low as 1 pa Absolute accuracy as low as 390 ppm (2-year calibration, 1 µa range) Max burden voltage: <50 mv on 1 µa range, <700 mv on 3 A range PXI-4022 as a feedback ammeter: PXI-4022 current amplifier: I-V converter used with voltmeter, sensitivity as low as 0.5 pa Absolute accuracy as low as 7,000 ppm Max burden voltage: <20 µv on 100 na range 24

25 Low-Current Measurements Major concern #1: Generated currents Offset currents due to measurement equipment Triboelectric effects cable movements induce charge Piezoelectric effects mechanical strain induces charge Contamination/humidity reduce resistance 25

26 Low-Current Measurements Major concern #2: Leakage currents and guarding i test - i leakage HI Sense DUT HI DMM i test Guard + - i leakage Test Fixture LO Sense LO 26

27 Low-Current Measurements Major concern #3: Burden voltage I measured = (1.5 V V)/(5 Ω) I measured = 0.2 A I actual = 1.5 V/5 Ω I actual = 0.3 A 27

28 High-Resistance Measurements Low-current measurement + high-voltage source PXI-4022 feedback ammeter with sensitivity to 0.5 pa 100 V power supply 100 V / 0.5 pa = 100 TΩ Am 100 V R measured 28

29 Insulation Resistance 29

30 Low Level Current Measurement with a DMM and an Additional Preamplifier Connection diagram for using PXI-4022 with PXI-4071 as picoammeter 30

31 Voltage/Current Sweeps Necessary for device characterization: Example: sweep voltage to measure diode operation Key parameters: Voltage/current programming resolution Voltage/current readback resolution Programming speed (up and down) For our relay DUT, we measure the voltage or current required to operate and release the contacts. 31

32 Operate-Release Voltage-Current 32

33 Settling Time Settling time is the amount of time required for the contacts to close (including bouncing). Use PXI-4071 DMM with digitizer mode to measure microseconds. Apply voltage across contacts. Begin digitizer mode acquisition on contacts. Apply voltage across coil. Voltage across contacts will be shorted, so power supply will current limit. Amount of time to settle at limited voltage is settling time. 33

34 Settling Time 34

35 Testing Scalability Switching/cabling considerations dominate precision measurement scalability. Avoid testing the test equipment instead of the DUT. Perform system calibrations. Software plays a key role in productivity. TestStand, Switch Executive, and LabVIEW enable test system deployment to multiple DUTs. Maximize code reuse and performance with PXI hardware. 35

36 Conclusion: Measurement Considerations When taking measurements with a DMM, consider the following: Normal/Common/Effective -mode rejection ratio (NMRR/CMRR/ECMRR) Input resistance Burden voltage Thermal voltages Settling time Resistor self-heating Dielectric absorption Residual impedance Stray admittance 36

37 For more information Digital multimeter measurement tutorial Modular instruments product demonstrations Understanding instrument specifications Calibration FAQs 37

38 Questions? 38

National Instruments Switches

National Instruments Switches ni.com National Instruments Switches Raviteja Chivukula Webinar Overview A. Switch Basics A. Recap B. Advanced Switch Topics A. High Channel Switches B. Fault Insertion Units C. Resistor Modules D. RF

More information

APPLICATION NOTE. Wide Range of Resistance Measurement Solutions from μω to PΩ

APPLICATION NOTE. Wide Range of Resistance Measurement Solutions from μω to PΩ APPLICATION NOTE Wide Range of Resistance Measurement Solutions from μω to PΩ Introduction Resistance measurement is one of the fundamental characterizations of materials, electronic devices, and circuits.

More information

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU) CALIBRATION PROCEDURE NI PXIe-4139 Single-Channel Precision Source-Measure Unit (SMU) This document contains the verification and adjustment procedures for the PXIe-4139. Refer to ni.com/calibration for

More information

Contents. Software Requirements

Contents. Software Requirements CALIBRATION PROCEDURE NI PXIe-4154 This document contains information for calibrating the NI PXIe-4154 Battery Simulator. For more information about calibration, visit ni.com/calibration. Contents Software

More information

6 1 2-Digit FlexDMM and 1.8 MS/s Isolated Digitizer

6 1 2-Digit FlexDMM and 1.8 MS/s Isolated Digitizer NI PXI-4070 Superior accuracy and measurement rates 00 range or function changes per second 6 ppm basic 4-hour VDC accuracy Built-in self-calibration guarantees accuracy Tight integration with NI switches

More information

6 1 2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter

6 1 2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter NI PXI-4072 FlexDMM Superior accuracy and measurement rates Multifunction device 6 1 2-digit digital multimeter 1.8 MS/s isolated digitizer LCR meter (inductance, capacitance, and resistance) 20 built-in

More information

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter CALIBRATION PROCEDURE PXIe-4081 PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter This document contains the verification and adjustment procedures for the PXIe-4081.

More information

5 th. Low Level Measurements. Handbook. Precision DC Current, Voltage and Resistance Measurements. Edition A GREATER MEASURE OF CONFIDENCE

5 th. Low Level Measurements. Handbook. Precision DC Current, Voltage and Resistance Measurements. Edition A GREATER MEASURE OF CONFIDENCE A GREATER MEASURE OF CONFIDENCE Low Level Measurements Handbook Precision DC Current, Voltage and Resistance Measurements 5 th Edition www.keithley.com LOW LEVEL MEASUREMENTS Precision DC Current,Voltage,

More information

What Is An SMU? SEP 2016

What Is An SMU? SEP 2016 What Is An SMU? SEP 2016 Agenda SMU Introduction Theory of Operation (Constant Current/Voltage Sourcing + Measure) Cabling : Triax vs Coax Advantages in Resistance Applications (vs. DMMs) Advantages in

More information

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply CALIBRATION PROCEDURE NI PXIe-4113 Dual-Output Programmable DC Power Supply This document contains the verification and adjustment procedures for the NI PXIe-4113 (NI 4113). Refer to ni.com/calibration

More information

Precision in Practice Achieving the best results with precision Digital Multimeter measurements

Precision in Practice Achieving the best results with precision Digital Multimeter measurements Precision in Practice Achieving the best results with precision Digital Multimeter measurements Paul Roberts Fluke Precision Measurement Ltd. Abstract Digital multimeters are one of the most common measurement

More information

NI PXIe Contents CALIBRATION PROCEDURE. Four Channel Precision Source-Measure Unit (SMU)

NI PXIe Contents CALIBRATION PROCEDURE. Four Channel Precision Source-Measure Unit (SMU) CALIBRATION PROCEDURE NI PXIe-4141 Four Channel Precision Source-Measure Unit (SMU) This document contains the verification and adjustment procedures for the NI PXIe-4141 (NI 4141). Refer to ni.com/calibration

More information

Racal Instruments 4152A. 6.5 Digit Digital Multimeter. Product Information

Racal Instruments 4152A. 6.5 Digit Digital Multimeter. Product Information Racal Instruments 415A 6.5 Digit Digital Multimeter The 6.5 digit Digital Multimeter (DMM) combines DC and AC current and voltage measurements as well as /4 wire Ω, frequency and DC ratio measurements

More information

Agilent 34970A Data Acquisition / Switch Unit

Agilent 34970A Data Acquisition / Switch Unit Note: Unless otherwise indicated, this manual applies to all serial numbers. The Agilent Technologies 34970A combines precision measurement capability with flexible signal connections for your production

More information

Optimizing System Throughput with the NI PXI ½-Digit FlexDMM

Optimizing System Throughput with the NI PXI ½-Digit FlexDMM Optimizing System Throughput with the NI PXI-4070 6 ½-Digit FlexDMM Introduction How do I maximize my system throughput? is a common question posed by many engineers and scientists. For years, engineers

More information

Model 6517B Electrometer / High Resistance Meter Specifications

Model 6517B Electrometer / High Resistance Meter Specifications VOLTS Accuracy (1 Year) 1 / C 2V 10µV 0.025+4 0.003+2 20V 100µV 0.025+3 0.002+1 200V 1mV 0.06+3 0.002+1 NMRR: 2V and 20V range > 60dB, 200V range > 55dB. 50Hz or 60Hz 2 CMRR: >120dB at DC, 50Hz or 60Hz.

More information

Application Note Series. Solutions for Production Testing of Connectors

Application Note Series. Solutions for Production Testing of Connectors Number 2208 Application Note Series Solutions for Production Testing of Connectors Introduction As electronics have become increasingly pervasive, the importance of electrical connectors has increased

More information

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5160 This document contains the verification and adjustment procedures for the PXIe-5160. Refer to ni.com/calibration for more information about calibration solutions. Contents

More information

Contents. CALIBRATION PROCEDURE NI PXIe-6555/6556. ni.com/manuals

Contents. CALIBRATION PROCEDURE NI PXIe-6555/6556. ni.com/manuals CALIBRATION PROCEDURE NI PXIe-6555/6556 Français Deutsch ni.com/manuals This document contains the verification and adjustment procedures for the NI PXIe-6555 (NI 6555) and NI PXIe-6556 (NI 6556) 200 MHz

More information

NI PXI ½ Digit FlexDMM Calibration Procedure

NI PXI ½ Digit FlexDMM Calibration Procedure NI PXI-4070 6½ Digit FlexDMM Calibration Procedure Contents This document contains step-by-step instructions for writing an external calibration procedure for the NI PXI-4070 6½ digit FlexDMM and 1.8 MS/s

More information

System Cabling Errors and DC Voltage Measurement Errors in Digital Multimeters

System Cabling Errors and DC Voltage Measurement Errors in Digital Multimeters Digital Multimeter Measurement Errors Series System Cabling Errors and DC Voltage Measurement Errors in Digital Multimeters Application Note AN 1389-1 Introduction When making measurements with a digital

More information

Welcome! Device Characterization with the Keithley Model 4200-SCS Characterization System.

Welcome! Device Characterization with the Keithley Model 4200-SCS Characterization System. Welcome! Device Characterization with the Keithley Model 4200-SCS Characterization System Low Current and High Resistance Measurement Techniques 1 Low Current and High Resistance Measurements Sources of

More information

Model Multimeter Specifications ABSOLUTE ACCURACY TYPICAL ACCURACIES 2002 SPECIFIED CALIBRATION INTERVALS

Model Multimeter Specifications ABSOLUTE ACCURACY TYPICAL ACCURACIES 2002 SPECIFIED CALIBRATION INTERVALS Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 www.tek.com/keithley Model 2002 Multimeter Specifications This document contains the complete specifications for the 2002. Every

More information

Keysight Technologies 6½ Digit PXI Digital Multimeters (M9181A, M9182A, M9183A)

Keysight Technologies 6½ Digit PXI Digital Multimeters (M9181A, M9182A, M9183A) Keysight Technologies 6½ Digit PXI Digital Multimeters (M9181A, M9182A, M9183A) Data Sheet 6½ digit resolution High throughput: up to 15,000 rdgs/sec, 66 µs single reading interval DCV basic one year accuracy:

More information

CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni. CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module Français Deutsch ni.com/manuals This document contains the verification

More information

Model 6600A Dual Source High Resistance Bridge

Model 6600A Dual Source High Resistance Bridge Dual Source High Resistance Bridge Based on proven NMI Design Range: 100 kω to 10 PΩ Voltages: 1 V to 1000 V (5000 V Optional) Automatic and Manual Operation Not affected by Temperature change 10 and 20

More information

Keysight Technologies PXI Digital Multimeters 6½ Digit, High Performance (M9182A, M9183A)

Keysight Technologies PXI Digital Multimeters 6½ Digit, High Performance (M9182A, M9183A) Keysight Technologies PXI Digital Multimeters 6½ Digit, High Performance (M9182A, M9183A) Data Sheet 6½ digit resolution High throughput: up to 15,000 rdgs/sec, 66 µs single reading interval DCV basic

More information

Analog front-end electronics

Analog front-end electronics FYS3240 PC-based instrumentation and microcontrollers Analog front-end electronics Spring 2017 Lecture #6 Bekkeng, 30.1.2017 Considerations for analog signals Signal source - grounded or floating Source

More information

SPECIFICATIONS FOR THE NI PXI/PCI-4060

SPECIFICATIONS FOR THE NI PXI/PCI-4060 SPECIFICATIONS FOR THE NI PXI/PCI-4060 This document lists the specifications of the NI PXI/PCI-4060. These specifications are guaranteed between 15 and 35 C unless otherwise specified. DC Voltage (% of

More information

Maximizing your reference multimeter, minimizing measurement uncertainties

Maximizing your reference multimeter, minimizing measurement uncertainties Maximizing your reference multimeter, minimizing measurement uncertainties Introduction Modern precision digital multimeters are sophisticated measuring instruments offering more than just the ability

More information

2.1 Make Accurate Low-Level Measurements with High-Resolution Instrumentation. Incorporating Best Practices for Ensuring Product Quality

2.1 Make Accurate Low-Level Measurements with High-Resolution Instrumentation. Incorporating Best Practices for Ensuring Product Quality 2.1 Make Accurate Low-Level Measurements with High-Resolution Instrumentation Incorporating Best Practices for Ensuring Product Quality 77 Types of Low-Level Measurements DC measurements measuring very

More information

Data Sheet. Agilent M9185A PXI Isolated D/A Converter. DISCOVER the Alternatives... Agilent MODULAR Products. 8/16-Channel 16-bit, ±16 V

Data Sheet. Agilent M9185A PXI Isolated D/A Converter. DISCOVER the Alternatives... Agilent MODULAR Products. 8/16-Channel 16-bit, ±16 V Agilent M9185A PXI Isolated D/A Converter Data Sheet 8/16-Channel 16-bit, ±16 V DISCOVER the Alternatives...... Agilent MODULAR Products Overview Introduction The Agilent M9185A is a digital/analog converter

More information

Discover. Blue Box. the. Difference. High Resistance Metrology Products Guide

Discover. Blue Box. the. Difference. High Resistance Metrology Products Guide Discover the Blue Box Difference High Resistance Metrology Products Guide Metrology is our Science, Accuracy is Our Business Measurements International (MI) is the world s premier metrology company. MI

More information

NI 4070/4072 6½-Digit FlexDMM

NI 4070/4072 6½-Digit FlexDMM CALIBRATION PROCEDURE NI 4070/4072 6½-Digit FlexDMM Contents This document contains step-by-step instructions for writing an external calibration procedure for the National Instruments PXI/PCI-4070 and

More information

4 Things to Consider When Using a DAQ as a Data Logger

4 Things to Consider When Using a DAQ as a Data Logger 4 Things to Consider When Using a DAQ as a Data Logger Introduction There is a growing trend across all industries to design feature-rich products. The more features added to a product, the more complex

More information

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer This document contains the verification procedures for the National Instruments PXIe-5668R (NI 5668R) vector signal analyzer (VSA)

More information

DC Current Source AC and DC Current Source

DC Current Source AC and DC Current Source AC and 6220 and Source and sink (programmable load) 100fA to 100mA 10 14 Ω output impedance ensures stable current sourcing into variable loads 65000-point source memory allows executing comprehensive

More information

Switching in Multipoint Testing

Switching in Multipoint Testing Number 1138 Application Note Series Switching in Multipoint Testing Switching in Multipoint Testing Selection of suitable switching is an extremely important but sometimes under-emphasized function in

More information

SCXI 8-Channel Isolated Analog Input Modules

SCXI 8-Channel Isolated Analog Input Modules SCXI 8-Channel Isolated Analog Input NI, NI SCXI-1120, NI SCXI-1120D 8 channels 333 ks/s maximum sampling rate Gain and lowpass filter settings per channel Up to 300 V rms working isolation per channel

More information

HP 34401A Specifications 8

HP 34401A Specifications 8 8 HP 34401A Specifications 8 DC Characteristics DC Characteristics Accuracy Specifications ± ( % of reading + % of range ) [ 1 ] Function Range [ 3 ] Test Current or Burden Voltage 24 Hour [ 2 ] 23 C ±

More information

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland Calibration Laboratory Registration number: 256C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in

More information

Isolation (ISO), Special Overvoltage functions protection (OP) Voltage measurement PAD-V8-P 8 Voltage, up to ±50 V BW: 6 Hz ISO: 350 V DC

Isolation (ISO), Special Overvoltage functions protection (OP) Voltage measurement PAD-V8-P 8 Voltage, up to ±50 V BW: 6 Hz ISO: 350 V DC Selection Guide PAD Series Modules Multi channel Low bandwidth - for static signals Isolation Digital output (RS-485) Exchangeable For DEWETRON systems with built-in DAQ rack Module # CH Input type Ranges

More information

white paper A primer A utomated Signal Switching reliable data first time every time

white paper A primer A utomated Signal Switching reliable data first time every time white paper A utomated Signal Switching 1 Overview Routing electrical and even optical signals from one source to another requires an overall understanding of the application, signals being switched and

More information

Models 2634B, 2635B and 2636B

Models 2634B, 2635B and 2636B Models 2634B, 2635B and 2636B Keithley Instruments, Inc. System SourceMeter 28775 Aurora Road Instrument Specifications Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com SPECIFICATION CONDITIONS

More information

2601 System SourceMeter 2602 Multi-Channel I-V Test Solutions

2601 System SourceMeter 2602 Multi-Channel I-V Test Solutions 601 System SourceMeter 60 Multi-Channel I-V Test Solutions SPECIFICATION CONDITIONS This document contains specifications and supplemental information for the Models 601 and 60. Specifications are the

More information

Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size

Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size Data Sheet Features 1-Slot, C-size, message-based DCV, ACV, DCI, ACI, 2/4-wire Ω, frequency, period NULL, MIN/MAX, LIMIT, db, dbm 1000 reading/s

More information

Bipolar Output Working Standard

Bipolar Output Working Standard 6166 DC Voltage/Current Source 6½-digit working standard suitable for calibration High-accuracy and highly stable with bipolar output l Wide dynamic range and high resolution Voltage source: ±10nV to ±1200V

More information

National Instruments Switches

National Instruments Switches ni.com National Instruments Switches Raviteja Chivukula Why the Need for Switches? Nearly every system can benefit from switching Increases channel count Adds measurement flexibility Simplifies test fixture

More information

2750 Integra Mainframes

2750 Integra Mainframes Selector Guide Plug-In Modules for 700, 70, 70 Integra Mainframes Module Selector Guide This selector guide may prove helpful in identifying the best module for a specific application. Install up to five

More information

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C

Butler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland Calibration Laboratory Registration number: 256C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in

More information

Models 2601B, 2602B and 2604B

Models 2601B, 2602B and 2604B Models 2601B, 2602B and 2604B Keithley Instruments, Inc. System SourceMeter 28775 Aurora Road Instrument Specifications Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com SPECIFICATION CONDITIONS

More information

Signal Conditioning Systems

Signal Conditioning Systems Note-13 1 Signal Conditioning Systems 2 Generalized Measurement System: The output signal from a sensor has generally to be processed or conditioned to make it suitable for the next stage Signal conditioning

More information

USB-TEMP and TC Series USB-Based Temperature Measurement Devices

USB-TEMP and TC Series USB-Based Temperature Measurement Devices USB-Based Temperature Measurement Devices Features Temperature and voltage measurement USB devices Thermocouple, RTD, thermistor, or semiconductor sensor measurements Eight analog inputs Up to ±10 V inputs*

More information

ni.com Sensor Measurement Fundamentals Series

ni.com Sensor Measurement Fundamentals Series Sensor Measurement Fundamentals Series Introduction to Data Acquisition Basics and Terminology Litkei Márton District Sales Manager National Instruments What Is Data Acquisition (DAQ)? 3 Why Measure? Engineers

More information

Fluke 8845A/8846A Digital Multimeters Extended Specifications

Fluke 8845A/8846A Digital Multimeters Extended Specifications Fluke 8845A/8846A Digital Multimeters Extended Specifications The Fluke 8845A and 8846A 6.5 digit precision multimeters have the precision and versatility to handle your most demanding measurements, on

More information

VXI-TB CHANNEL ISOTHERMAL TERMINAL BLOCK

VXI-TB CHANNEL ISOTHERMAL TERMINAL BLOCK VXI-TB-1303 32-CHANNEL ISOTHERMAL TERMINAL BLOCK Introduction This guide describes how to install and use the VXI-TB-1303 terminal block with a VXI-SC submodule. The VXI-TB-1303 terminal block is a shielded

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 1 Wilson Street Contact: Mr Ibrahim Ibrahim Thornaby Tel: +44 (0) 1642 626148 Stockton-On-Tees Fax: +44 (0) 870 143 1869 TS17 7AR E-Mail: ibrahim@calibrate.co.uk United Kingdom Website: www.calibrate.co.uk

More information

MODEL INFORMATION MODEL 6010D. Automated Primary Resistance/ Thermometry Bridge

MODEL INFORMATION MODEL 6010D. Automated Primary Resistance/ Thermometry Bridge MODEL 6010D Automated Primary Resistance/ Thermometry Bridge Resistance & Temperature Applications Range 0.001 Ω to 100 KΩ Accuracy < 40 x 10-9 Linearity < 5 x 10-9 Featuring true ratio self calibration

More information

Model 2651A Specifications

Model 2651A Specifications Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-55 http://www.tek.com/keithley Model 2651A Specifications High Power System SourceMeter Instrument Specifications SPECIFICATION CONDITIONS

More information

CX1100 Series Current and Differential Sensors

CX1100 Series Current and Differential Sensors DATA SHEET CX00 Series Current and Differential Sensors CX0A Current Sensor, Single Channel CX0A Current Sensor, Dual Channel CX03A Current Sensor, Low Side CX04A Current Sensor, Selectable Resistive Sensor

More information

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer CALIBRATION PROCEDURE NI PXIe-5665 3.6 GHz and 14 GHz RF Vector Signal Analyzer This document contains the verification procedures for the National Instruments PXIe-5665 (NI 5665) RF vector signal analyzer

More information

Keysight 34401A Multimeter

Keysight 34401A Multimeter Keysight 34401A Multimeter Data Sheet See Keysight's Truevolt Series of DMMs You also get both GPIB and RS-232 interfaces as standard features. Voltmeter Complete and External Trigger signals are provided

More information

Basic Analog Circuits

Basic Analog Circuits Basic Analog Circuits Overview This tutorial is part of the National Instruments Measurement Fundamentals series. Each tutorial in this series, will teach you a specific topic of common measurement applications,

More information

6½-Digit, ±300 V, 1.8 MS/s Isolated Digitizer Included, PCI Digital Multimeter Device

6½-Digit, ±300 V, 1.8 MS/s Isolated Digitizer Included, PCI Digital Multimeter Device SPECIFICATIONS PCI-4070 6½-Digit, ±300 V, 1.8 MS/s Isolated Digitizer Included, PCI Digital Multimeter Device These specifications apply to the PCI-4070. Contents Definitions...2 Conditions... 2 DC Specifications...

More information

Technical Services Department, BHEL, Piplani, Bhopal, Madhya Pradesh. Discipline Electro Technical Calibration Issue Date

Technical Services Department, BHEL, Piplani, Bhopal, Madhya Pradesh. Discipline Electro Technical Calibration Issue Date Last Amended on 29.02.2016 Page 1 of 6 SOURCE 1. DC VOLTAGE $ 100mV to 220 mv 220 mv to 220 V 220 V to 1000 V 2. AC VOLTAGE $ 50 Hz to 1 khz 3 mv to 220 mv 220 mv to 220 V 220 V to 1000 V 3. DC CURRENT

More information

2.5 GHz 75 Ω Multiplexer and SPDT Relay Switches

2.5 GHz 75 Ω Multiplexer and SPDT Relay Switches 2.5 GHz Multiplexer and SPDT Relay Switches NI PXI-255x NEW! 2.5 GHz bandwidth characteristic impedance 30 V max switching voltage 0.5 A max switching current 10 W max switching power Mini SMB direct connectivity

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Metrology Division Contact: Mr K Pallett Calibration House Tel: +44 (0)1332 867 700 Castings Road E-Mail: sales@comech.co.uk Derby Website:

More information

Ultra-Fast I-V Module for the Model 4200-SCS

Ultra-Fast I-V Module for the Model 4200-SCS Provides voltage outputs with programmable timing from 60ns to DC in 10ns steps Measure I and V simultaneously, at acquisition rates of up to 200 megasamples/second (MS/s) Choose from two voltage source

More information

NI Contents CALIBRATION PROCEDURE

NI Contents CALIBRATION PROCEDURE CALIBRATION PROCEDURE NI 5450 Contents This document describes processes to calibrate the National Instruments PXIe-5450 (NI 5450) differential I/Q signal generator. This document provides performance

More information

DUAL ULTRA MICROPOWER RAIL-TO-RAIL CMOS OPERATIONAL AMPLIFIER

DUAL ULTRA MICROPOWER RAIL-TO-RAIL CMOS OPERATIONAL AMPLIFIER ADVANCED LINEAR DEVICES, INC. ALD276A/ALD276B ALD276 DUAL ULTRA MICROPOWER RAILTORAIL CMOS OPERATIONAL AMPLIFIER GENERAL DESCRIPTION The ALD276 is a dual monolithic CMOS micropower high slewrate operational

More information

DMM ½-Digit Graphical Sampling Multimeter

DMM ½-Digit Graphical Sampling Multimeter DMM 7510 7½-Digit Graphical Sampling Multimeter Datasheet The DMM7510 combines all the advantages of a precision digital multimeter, a graphical touchscreen display, and a high speed, high resolution digitizer

More information

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents. Required Software CALIBRATION PROCEDURE CALIBRATION PROCEDURE PXIe-5113 This document contains the verification and adjustment procedures for the PXIe-5113. Refer to ni.com/calibration for more information about calibration solutions. Contents

More information

Lecture Notes Unit-III

Lecture Notes Unit-III Lecture Notes Unit-III FAQs Q1: An operational amplifier has a differential gain of 103 and CMRR of 100, input voltages are 120µV and 80µV, determine output voltage. 2 MARKS

More information

Model 1140A Thermocouple Simulator-Calibrator

Model 1140A Thermocouple Simulator-Calibrator BULLETIN 2031 Model 1140A Thermocouple Simulator-Calibrator The Model 1140A represents the latest innovation in thermocouple simulator-calibrators from Ectron, the originator of the Thermocouple Simulator

More information

Contents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator

Contents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator CALIBRATION PROCEDURE NI 5421/5441 100 MS/s Arbitrary Waveform Generator This document contains the verification and adjustment procedures for the NI 5421/5441 arbitrary waveform generator. This calibration

More information

5320A. Multifunction Electrical Tester Calibrator. Extended Specifications

5320A. Multifunction Electrical Tester Calibrator. Extended Specifications 5320A Multifunction Electrical Tester Calibrator Extended Specifications General Specifications Warm-Up Time...30 minutes Specifications Confidence Level...99 % Specifications Interval...1 year Temperature

More information

DC Resistance Meters 2840 Series

DC Resistance Meters 2840 Series Data Sheet DC Meters Touch screen to zoom, select, and enter values The DC resistance meters feature high accuracy and resolution measurements in the milliohm range. Both meters are ideally suited for

More information

High Voltage Component Production Testing with Two Model 2410 SourceMeter Units. Application Note Series. Introduction. Test System Configuration

High Voltage Component Production Testing with Two Model 2410 SourceMeter Units. Application Note Series. Introduction. Test System Configuration A Tektronix Company Application Note Series Number 2058 igh Voltage Component Production Testing with Two Model 2410 SourceMeter Units Introduction Various production test applications require the use

More information

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS DC POWER SUPPLIES We will discuss these instruments one at a time, starting with the DC power supply. The simplest DC power supplies are batteries which

More information

6.5-Digit Multimeter, High- Accuracy, C-Size HP E1410A Technical Specifications

6.5-Digit Multimeter, High- Accuracy, C-Size HP E1410A Technical Specifications PDFINFO H 5 5 6 1-0 1 6.5-Digit Multimeter, High- Accuracy, C-Size Technical Specifications 1-Slot, C-size, message based Vdc/ac, 2- & 4-wire Ω Noise rejection with long integration times/ guarding Quality

More information

Analog Signal Conditioning Accessories

Analog Signal Conditioning Accessories NI 64-channel multiplexer mv, V, current, and thermocouple inputs NI 8-channel simultaneous sample-and-hold mv, V inputs NI SC-2042-RTD 8-channel RTD/thermistor RTD, thermistor, mv, V inputs NI 8-channel

More information

Contents CALIBRATION PROCEDURE NI 5412

Contents CALIBRATION PROCEDURE NI 5412 CALIBRATION PROCEDURE NI 5412 Contents Introduction... 2 Software... 2 Documentation... 3 Password... 4 Calibration Interval... 4 Test Equipment... 4 Test Conditions...5 Self-Calibration Procedures...

More information

SOURCE-MEASURE UNITS INCREASE PRODUCTIVITY AND ACCURACY IN AUTOMATED TESTING. Lee Stauffer. Keithley Instruments, Inc.

SOURCE-MEASURE UNITS INCREASE PRODUCTIVITY AND ACCURACY IN AUTOMATED TESTING. Lee Stauffer. Keithley Instruments, Inc. SOURCE-MEASURE UNITS INCREASE PRODUCTIVITY AND ACCURACY IN AUTOMATED TESTING Lee Stauffer Keithley Instruments, Inc. Introduction Source-Measure Units (SMUs) are more than the next generation of power

More information

DC MHZ PXI Differential Instrumentation Amplifier

DC MHZ PXI Differential Instrumentation Amplifier DC - 100 MHZ PXI Differential Instrumentation Amplifier Differential 100 V Common Mode Input DC - 100 MHz Bandwidth AC/DC Coupling Programmable Attenuation/Gain/ Offset 9 nv/ Input Noise 50 Ω Output Impedance

More information

350MHz, Ultra-Low-Noise Op Amps

350MHz, Ultra-Low-Noise Op Amps 9-442; Rev ; /95 EVALUATION KIT AVAILABLE 35MHz, Ultra-Low-Noise Op Amps General Description The / op amps combine high-speed performance with ultra-low-noise performance. The is compensated for closed-loop

More information

Signal Conditioning Fundamentals for PC-Based Data Acquisition Systems

Signal Conditioning Fundamentals for PC-Based Data Acquisition Systems Application Note 048 Signal Conditioning Fundamentals for PC-Based Data Acquisition Systems Introduction PC-based data acquisition (DAQ) systems and plugin boards are used in a very wide range of applications

More information

Chapter 6 Specifications

Chapter 6 Specifications Chapter 6 Specifications DC Characteristics Function Range [2] Test Current or Burden Voltage DC Voltage DC Current Resistance [6] 24 Hour [3] T CAL ±1 Accuracy Specifications: ± (% of reading + % of range)

More information

Measurement & Control of energy systems. Teppo Myllys National Instruments

Measurement & Control of energy systems. Teppo Myllys National Instruments Measurement & Control of energy systems Teppo Myllys National Instruments National Instruments Direct operations in over 50 Countries More than 1,000 products, 7000+ employees, and 700 Alliance Program

More information

INPUT RESISTANCE OR CLAMP VOLTAGE. 2 Years 23 C ±5 Voltage mv 3 10 nv > 10 G

INPUT RESISTANCE OR CLAMP VOLTAGE. 2 Years 23 C ±5 Voltage mv 3 10 nv > 10 G Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 www.keithley.com Model 2010 Low-Noise Multimeter Specifications DC CHARACTERISTICS CONDITIONS: MED (1 PLC) 1 or SLOW (5 PLC)

More information

Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL

Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL TECHNICAL INFORMATION When comparing two products features and specification using sales information it is often difficult to determine which product

More information

PXI Contents SPECIFICATIONS. 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer, L and C Measurement Support, PXI Digital Multimeter

PXI Contents SPECIFICATIONS. 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer, L and C Measurement Support, PXI Digital Multimeter SPECIFICATIONS PXI-4072 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer, L and C Measurement Support, PXI Digital Multimeter These specifications apply to the PXI-4072. Contents Definitions...2 Conditions...

More information

Data acquisition and instrumentation. Data acquisition

Data acquisition and instrumentation. Data acquisition Data acquisition and instrumentation START Lecture Sam Sadeghi Data acquisition 1 Humanistic Intelligence Body as a transducer,, data acquisition and signal processing machine Analysis of physiological

More information

Solutions for Production Testing of Connectors. Application Note Series. Introduction. Test Description. Number 2208

Solutions for Production Testing of Connectors. Application Note Series. Introduction. Test Description. Number 2208 Number 2208 Application Note Series Solutions for Production Testing of Connectors Introduction As electronics have become increasingly pervasive, the importance of electrical connectors has increased

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, Unit 8 The Centre Holywell Business Park Northfield Road Southam Warwickshire CV47 0FP Contact: Mr N Morgan Tel: +44 (0)1926 812066 Fax: +44

More information

PCI Express, 6½-Digit, ±300 V Digital Multimeter Device

PCI Express, 6½-Digit, ±300 V Digital Multimeter Device SPECIFICATIONS PCIe-4065 PCI Express, 6½-Digit, ±300 V Digital Multimeter Device These specifications apply to the PCIe-4065. Contents Definitions...1 Conditions... 2 DC Specifications... 2 DC System Speed

More information

Unequalled combination of Accuracy, Stability and Versatility

Unequalled combination of Accuracy, Stability and Versatility üthe Source for Calibration Professionals The new microk family of precision thermometry bridges Unequalled combination of Accuracy, Stability and Versatility www.isotech.co.uk Introduction The microk

More information

Integrating a PXI Automated Test System

Integrating a PXI Automated Test System Integrating a PXI Automated Test System Top Considerations to Integrating a PXI Automated Test System Agilent Technologies Software and Modular Solutions Alan Lesko Application Engineer April 24 th, 2012

More information

High Power Monolithic OPERATIONAL AMPLIFIER

High Power Monolithic OPERATIONAL AMPLIFIER High Power Monolithic OPERATIONAL AMPLIFIER FEATURES POWER SUPPLIES TO ±0V OUTPUT CURRENT TO 0A PEAK PROGRAMMABLE CURRENT LIMIT INDUSTRY-STANDARD PIN OUT FET INPUT TO- AND LOW-COST POWER PLASTIC PACKAGES

More information

2790 SourceMeter Airbag Test System

2790 SourceMeter Airbag Test System 2790 Meter Airbag Test System Single-instrument solution for continuity and hi-pot type leakage resistance measurements Programmable constant V source (50 500V) supports high speed, high resistance measurements

More information

Calsytech, # 38 North Mada Street Nandambakkam, Chennai, Tamil Nadu. Discipline Electro-Technical Calibration Issue Date

Calsytech, # 38 North Mada Street Nandambakkam, Chennai, Tamil Nadu. Discipline Electro-Technical Calibration Issue Date Last Amended on 23.11.2016 Page 1 of 7 SOURCE 1. DC VOLTAGE $ 1mV to 10 mv 10 mv to100 mv 100 mv to1000 V 0.43% to 0.05% 0.05% to 0.01% 0.01% to 0.004% Calibrator, Calibrator by DC VOLTAGE 1mV to 20 mv

More information