1 Fallstricke präziser DC- Messungen Sascha Egger, Applications Engineer Group Leader National Instruments Switzerland GmbH
2 Agenda Overview of Precision Test Systems Techniques for: Low-voltage measurements Low-resistance measurements Low-current measurements High-impedance measurements Sweeps Conclusion: Measurement Considerations 2
3 What are Precision Measurements? Precision explicitly means: Reproducibility/repeatability Yield same results each time measurement is taken To man people, it also implies: Accuracy the overall uncertainty of a measurement Resolution number of bits used by the ADC (or DAC) to represent a signal Sensitivity the smallest signal detectable by a particular instrument on its most sensitive range 3
4 What are Precision Measurements? With modern instrumentation, these parameters usually come into question for the following: Voltage under 10 µv Current under 10 µa Resistance under 10 mω A precision measurement system should offer good accuracy, resolution, sensitivity, and precision for measurements under/of? these values. 4
5 Precision Measurement Examples Dielectric absorption Electrochemical Leakage Low-current semiconductor Light measurements Connector testing Surface insulation resistance Resistivity Charge measurements Temperature Contact resistance Superconductor resistance NI CONFIDENTIAL 5
6 Challenges in Precision Measurements Cost/Specialty of measurement equipment Environmental challenges Test times Cabling Channel count This presentation will provide a case study for building a precision DC test system in PXI and point out some pitfalls. 6
7 Case Study Precision Device Under Test (DUT) Reed relay SRC devices form a relay Tests to be performed: Contact resistance Coil resistance Thermal offset Coil pickup voltage, current Coil dropout voltage, current Open contact resistance Settling time 7
8 Relay Specifications 150 mω 10 GΩ-1 TΩ 500 us Ω 3.75 V, 0.8 V 8
9 Equipment PXI-1042 chassis and PXI-8119 controller: Measurements: o PXI ½-digit DMM o PXI wire guard and current amplifier Source: o PXI-4110 programmable DC power supply Switch: o PXI x16 reed relay matrix o PXI channel SPST relays External 100V DC-DC converter or SMU PXI
10 System Architecture 10
11 Low-Voltage Measurement Devices Data acquisition modules: X Series DAQ: 16 bits, sensitivity as low as 15 µv Absolute accuracy as low as 291 ppm (1 V range) Digital multimeters (DMM): PXI-407x FlexDMMs: bits, sensitivity as low as 10 nv Absolute accuracy as low as 28 ppm (wide temperature? range, 2-year calibration, 100 mv range) 11
12 Calibration Instruments must be calibrated at regular intervals. Benefits of calibration: Reduced measurement errors Consistency between measurements Assurance you are making accurate measurements Calibration procedures: Self-calibration Manual calibration procedures NI calibration services 12
13 NI-DMM Driver API, examples and help will be installed with NI-DMM driver. Supported platforms: LabVIEW & LabVIEW Real-Time LabWindows CVI & LabWindows CVI Real-Time Microsoft Visual C++ Microsoft Visual Basic
14 Soft Front Panels for Modular Instruments DMM Soft Front Panel will be installed with NI-DMM Driver. 14
15 Low-Voltage Measurements Major concern #1: Offset voltages Thermal offsets develop from junction points in the system (similar to a thermocouple). Voltmeter or DMM input offsets 15
16 Low-Voltage Measurements Major concern #2: Noise Johnson noise Magnetic fields Ground loops Johnson noise e nr = 1.28 x10-10 * Rf Twisted-pair wires to reduce noise 16
17 Thermal Offset 17
18 Low-Resistance and In-Circuit Measurement Devices Digital multimeters: PXI-407x FlexDMMs: bits, sensitivity as low as 10 µω Absolute accuracy as low as 60 ppm (wide temperature? range, 2-year calibration, 100 Ω range) 2-wire and 4-wire capability In-circuit guard amplifier accessory: PXI wire guard and current amplifier: enables 6-wire resistance measurements ±200 µv guard accuracy, typical 18
19 Low-Resistance Measurements Major concern #1: Lead resistance 2-wire measurement 4-wire measurement 19
20 Low-Resistance Measurements Major concern #2: Thermal EMFs Current reversal method Offset-compensated ohms V OCO = V M1 - V M2 = (I S R X + V THERMAL ) - V THERMAL = I S R X therefore: R X = V OCO /I S 20
21 Contact Resistance 21
22 Low-Level Resistance Measurements with DMM (4-Wire Mode) Internal current source (1mA) External current source (100mA) 22
23 Low-Level Resistance Measurements with a DMM and Additional Current Source External current source 23
24 Low-Current Measurement Devices Digital multimeters (shunt ammeters): PXI-407x FlexDMMs: bits, sensitivity as low as 1 pa Absolute accuracy as low as 390 ppm (2-year calibration, 1 µa range) Max burden voltage: <50 mv on 1 µa range, <700 mv on 3 A range PXI-4022 as a feedback ammeter: PXI-4022 current amplifier: I-V converter used with voltmeter, sensitivity as low as 0.5 pa Absolute accuracy as low as 7,000 ppm Max burden voltage: <20 µv on 100 na range 24
25 Low-Current Measurements Major concern #1: Generated currents Offset currents due to measurement equipment Triboelectric effects cable movements induce charge Piezoelectric effects mechanical strain induces charge Contamination/humidity reduce resistance 25
26 Low-Current Measurements Major concern #2: Leakage currents and guarding i test - i leakage HI Sense DUT HI DMM i test Guard + - i leakage Test Fixture LO Sense LO 26
27 Low-Current Measurements Major concern #3: Burden voltage I measured = (1.5 V V)/(5 Ω) I measured = 0.2 A I actual = 1.5 V/5 Ω I actual = 0.3 A 27
28 High-Resistance Measurements Low-current measurement + high-voltage source PXI-4022 feedback ammeter with sensitivity to 0.5 pa 100 V power supply 100 V / 0.5 pa = 100 TΩ Am 100 V R measured 28
29 Insulation Resistance 29
30 Low Level Current Measurement with a DMM and an Additional Preamplifier Connection diagram for using PXI-4022 with PXI-4071 as picoammeter 30
31 Voltage/Current Sweeps Necessary for device characterization: Example: sweep voltage to measure diode operation Key parameters: Voltage/current programming resolution Voltage/current readback resolution Programming speed (up and down) For our relay DUT, we measure the voltage or current required to operate and release the contacts. 31
32 Operate-Release Voltage-Current 32
33 Settling Time Settling time is the amount of time required for the contacts to close (including bouncing). Use PXI-4071 DMM with digitizer mode to measure microseconds. Apply voltage across contacts. Begin digitizer mode acquisition on contacts. Apply voltage across coil. Voltage across contacts will be shorted, so power supply will current limit. Amount of time to settle at limited voltage is settling time. 33
34 Settling Time 34
35 Testing Scalability Switching/cabling considerations dominate precision measurement scalability. Avoid testing the test equipment instead of the DUT. Perform system calibrations. Software plays a key role in productivity. TestStand, Switch Executive, and LabVIEW enable test system deployment to multiple DUTs. Maximize code reuse and performance with PXI hardware. 35
36 Conclusion: Measurement Considerations When taking measurements with a DMM, consider the following: Normal/Common/Effective -mode rejection ratio (NMRR/CMRR/ECMRR) Input resistance Burden voltage Thermal voltages Settling time Resistor self-heating Dielectric absorption Residual impedance Stray admittance 36
37 For more information Digital multimeter measurement tutorial Modular instruments product demonstrations Understanding instrument specifications Calibration FAQs 37
38 Questions? 38
National Instruments Switches
ni.com National Instruments Switches Raviteja Chivukula Webinar Overview A. Switch Basics A. Recap B. Advanced Switch Topics A. High Channel Switches B. Fault Insertion Units C. Resistor Modules D. RF
More informationAPPLICATION NOTE. Wide Range of Resistance Measurement Solutions from μω to PΩ
APPLICATION NOTE Wide Range of Resistance Measurement Solutions from μω to PΩ Introduction Resistance measurement is one of the fundamental characterizations of materials, electronic devices, and circuits.
More informationNI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)
CALIBRATION PROCEDURE NI PXIe-4139 Single-Channel Precision Source-Measure Unit (SMU) This document contains the verification and adjustment procedures for the PXIe-4139. Refer to ni.com/calibration for
More informationContents. Software Requirements
CALIBRATION PROCEDURE NI PXIe-4154 This document contains information for calibrating the NI PXIe-4154 Battery Simulator. For more information about calibration, visit ni.com/calibration. Contents Software
More information6 1 2-Digit FlexDMM and 1.8 MS/s Isolated Digitizer
NI PXI-4070 Superior accuracy and measurement rates 00 range or function changes per second 6 ppm basic 4-hour VDC accuracy Built-in self-calibration guarantees accuracy Tight integration with NI switches
More information6 1 2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter
NI PXI-4072 FlexDMM Superior accuracy and measurement rates Multifunction device 6 1 2-digit digital multimeter 1.8 MS/s isolated digitizer LCR meter (inductance, capacitance, and resistance) 20 built-in
More informationPXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter
CALIBRATION PROCEDURE PXIe-4081 PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter This document contains the verification and adjustment procedures for the PXIe-4081.
More information5 th. Low Level Measurements. Handbook. Precision DC Current, Voltage and Resistance Measurements. Edition A GREATER MEASURE OF CONFIDENCE
A GREATER MEASURE OF CONFIDENCE Low Level Measurements Handbook Precision DC Current, Voltage and Resistance Measurements 5 th Edition www.keithley.com LOW LEVEL MEASUREMENTS Precision DC Current,Voltage,
More informationWhat Is An SMU? SEP 2016
What Is An SMU? SEP 2016 Agenda SMU Introduction Theory of Operation (Constant Current/Voltage Sourcing + Measure) Cabling : Triax vs Coax Advantages in Resistance Applications (vs. DMMs) Advantages in
More informationNI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply
CALIBRATION PROCEDURE NI PXIe-4113 Dual-Output Programmable DC Power Supply This document contains the verification and adjustment procedures for the NI PXIe-4113 (NI 4113). Refer to ni.com/calibration
More informationPrecision in Practice Achieving the best results with precision Digital Multimeter measurements
Precision in Practice Achieving the best results with precision Digital Multimeter measurements Paul Roberts Fluke Precision Measurement Ltd. Abstract Digital multimeters are one of the most common measurement
More informationNI PXIe Contents CALIBRATION PROCEDURE. Four Channel Precision Source-Measure Unit (SMU)
CALIBRATION PROCEDURE NI PXIe-4141 Four Channel Precision Source-Measure Unit (SMU) This document contains the verification and adjustment procedures for the NI PXIe-4141 (NI 4141). Refer to ni.com/calibration
More informationRacal Instruments 4152A. 6.5 Digit Digital Multimeter. Product Information
Racal Instruments 415A 6.5 Digit Digital Multimeter The 6.5 digit Digital Multimeter (DMM) combines DC and AC current and voltage measurements as well as /4 wire Ω, frequency and DC ratio measurements
More informationAgilent 34970A Data Acquisition / Switch Unit
Note: Unless otherwise indicated, this manual applies to all serial numbers. The Agilent Technologies 34970A combines precision measurement capability with flexible signal connections for your production
More informationOptimizing System Throughput with the NI PXI ½-Digit FlexDMM
Optimizing System Throughput with the NI PXI-4070 6 ½-Digit FlexDMM Introduction How do I maximize my system throughput? is a common question posed by many engineers and scientists. For years, engineers
More informationModel 6517B Electrometer / High Resistance Meter Specifications
VOLTS Accuracy (1 Year) 1 / C 2V 10µV 0.025+4 0.003+2 20V 100µV 0.025+3 0.002+1 200V 1mV 0.06+3 0.002+1 NMRR: 2V and 20V range > 60dB, 200V range > 55dB. 50Hz or 60Hz 2 CMRR: >120dB at DC, 50Hz or 60Hz.
More informationApplication Note Series. Solutions for Production Testing of Connectors
Number 2208 Application Note Series Solutions for Production Testing of Connectors Introduction As electronics have become increasingly pervasive, the importance of electrical connectors has increased
More informationPXIe Contents. Required Software CALIBRATION PROCEDURE
CALIBRATION PROCEDURE PXIe-5160 This document contains the verification and adjustment procedures for the PXIe-5160. Refer to ni.com/calibration for more information about calibration solutions. Contents
More informationContents. CALIBRATION PROCEDURE NI PXIe-6555/6556. ni.com/manuals
CALIBRATION PROCEDURE NI PXIe-6555/6556 Français Deutsch ni.com/manuals This document contains the verification and adjustment procedures for the NI PXIe-6555 (NI 6555) and NI PXIe-6556 (NI 6556) 200 MHz
More informationNI PXI ½ Digit FlexDMM Calibration Procedure
NI PXI-4070 6½ Digit FlexDMM Calibration Procedure Contents This document contains step-by-step instructions for writing an external calibration procedure for the NI PXI-4070 6½ digit FlexDMM and 1.8 MS/s
More informationSystem Cabling Errors and DC Voltage Measurement Errors in Digital Multimeters
Digital Multimeter Measurement Errors Series System Cabling Errors and DC Voltage Measurement Errors in Digital Multimeters Application Note AN 1389-1 Introduction When making measurements with a digital
More informationWelcome! Device Characterization with the Keithley Model 4200-SCS Characterization System.
Welcome! Device Characterization with the Keithley Model 4200-SCS Characterization System Low Current and High Resistance Measurement Techniques 1 Low Current and High Resistance Measurements Sources of
More informationModel Multimeter Specifications ABSOLUTE ACCURACY TYPICAL ACCURACIES 2002 SPECIFIED CALIBRATION INTERVALS
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 www.tek.com/keithley Model 2002 Multimeter Specifications This document contains the complete specifications for the 2002. Every
More informationKeysight Technologies 6½ Digit PXI Digital Multimeters (M9181A, M9182A, M9183A)
Keysight Technologies 6½ Digit PXI Digital Multimeters (M9181A, M9182A, M9183A) Data Sheet 6½ digit resolution High throughput: up to 15,000 rdgs/sec, 66 µs single reading interval DCV basic one year accuracy:
More informationCALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.
CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module Français Deutsch ni.com/manuals This document contains the verification
More informationModel 6600A Dual Source High Resistance Bridge
Dual Source High Resistance Bridge Based on proven NMI Design Range: 100 kω to 10 PΩ Voltages: 1 V to 1000 V (5000 V Optional) Automatic and Manual Operation Not affected by Temperature change 10 and 20
More informationKeysight Technologies PXI Digital Multimeters 6½ Digit, High Performance (M9182A, M9183A)
Keysight Technologies PXI Digital Multimeters 6½ Digit, High Performance (M9182A, M9183A) Data Sheet 6½ digit resolution High throughput: up to 15,000 rdgs/sec, 66 µs single reading interval DCV basic
More informationAnalog front-end electronics
FYS3240 PC-based instrumentation and microcontrollers Analog front-end electronics Spring 2017 Lecture #6 Bekkeng, 30.1.2017 Considerations for analog signals Signal source - grounded or floating Source
More informationSPECIFICATIONS FOR THE NI PXI/PCI-4060
SPECIFICATIONS FOR THE NI PXI/PCI-4060 This document lists the specifications of the NI PXI/PCI-4060. These specifications are guaranteed between 15 and 35 C unless otherwise specified. DC Voltage (% of
More informationMaximizing your reference multimeter, minimizing measurement uncertainties
Maximizing your reference multimeter, minimizing measurement uncertainties Introduction Modern precision digital multimeters are sophisticated measuring instruments offering more than just the ability
More information2.1 Make Accurate Low-Level Measurements with High-Resolution Instrumentation. Incorporating Best Practices for Ensuring Product Quality
2.1 Make Accurate Low-Level Measurements with High-Resolution Instrumentation Incorporating Best Practices for Ensuring Product Quality 77 Types of Low-Level Measurements DC measurements measuring very
More informationData Sheet. Agilent M9185A PXI Isolated D/A Converter. DISCOVER the Alternatives... Agilent MODULAR Products. 8/16-Channel 16-bit, ±16 V
Agilent M9185A PXI Isolated D/A Converter Data Sheet 8/16-Channel 16-bit, ±16 V DISCOVER the Alternatives...... Agilent MODULAR Products Overview Introduction The Agilent M9185A is a digital/analog converter
More informationDiscover. Blue Box. the. Difference. High Resistance Metrology Products Guide
Discover the Blue Box Difference High Resistance Metrology Products Guide Metrology is our Science, Accuracy is Our Business Measurements International (MI) is the world s premier metrology company. MI
More informationNI 4070/4072 6½-Digit FlexDMM
CALIBRATION PROCEDURE NI 4070/4072 6½-Digit FlexDMM Contents This document contains step-by-step instructions for writing an external calibration procedure for the National Instruments PXI/PCI-4070 and
More information4 Things to Consider When Using a DAQ as a Data Logger
4 Things to Consider When Using a DAQ as a Data Logger Introduction There is a growing trend across all industries to design feature-rich products. The more features added to a product, the more complex
More informationContents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer
CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer This document contains the verification procedures for the National Instruments PXIe-5668R (NI 5668R) vector signal analyzer (VSA)
More informationDC Current Source AC and DC Current Source
AC and 6220 and Source and sink (programmable load) 100fA to 100mA 10 14 Ω output impedance ensures stable current sourcing into variable loads 65000-point source memory allows executing comprehensive
More informationSwitching in Multipoint Testing
Number 1138 Application Note Series Switching in Multipoint Testing Switching in Multipoint Testing Selection of suitable switching is an extremely important but sometimes under-emphasized function in
More informationSCXI 8-Channel Isolated Analog Input Modules
SCXI 8-Channel Isolated Analog Input NI, NI SCXI-1120, NI SCXI-1120D 8 channels 333 ks/s maximum sampling rate Gain and lowpass filter settings per channel Up to 300 V rms working isolation per channel
More informationHP 34401A Specifications 8
8 HP 34401A Specifications 8 DC Characteristics DC Characteristics Accuracy Specifications ± ( % of reading + % of range ) [ 1 ] Function Range [ 3 ] Test Current or Burden Voltage 24 Hour [ 2 ] 23 C ±
More informationButler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C
Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland Calibration Laboratory Registration number: 256C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in
More informationIsolation (ISO), Special Overvoltage functions protection (OP) Voltage measurement PAD-V8-P 8 Voltage, up to ±50 V BW: 6 Hz ISO: 350 V DC
Selection Guide PAD Series Modules Multi channel Low bandwidth - for static signals Isolation Digital output (RS-485) Exchangeable For DEWETRON systems with built-in DAQ rack Module # CH Input type Ranges
More informationwhite paper A primer A utomated Signal Switching reliable data first time every time
white paper A utomated Signal Switching 1 Overview Routing electrical and even optical signals from one source to another requires an overall understanding of the application, signals being switched and
More informationModels 2634B, 2635B and 2636B
Models 2634B, 2635B and 2636B Keithley Instruments, Inc. System SourceMeter 28775 Aurora Road Instrument Specifications Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com SPECIFICATION CONDITIONS
More information2601 System SourceMeter 2602 Multi-Channel I-V Test Solutions
601 System SourceMeter 60 Multi-Channel I-V Test Solutions SPECIFICATION CONDITIONS This document contains specifications and supplemental information for the Models 601 and 60. Specifications are the
More informationAgilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size
Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size Data Sheet Features 1-Slot, C-size, message-based DCV, ACV, DCI, ACI, 2/4-wire Ω, frequency, period NULL, MIN/MAX, LIMIT, db, dbm 1000 reading/s
More informationBipolar Output Working Standard
6166 DC Voltage/Current Source 6½-digit working standard suitable for calibration High-accuracy and highly stable with bipolar output l Wide dynamic range and high resolution Voltage source: ±10nV to ±1200V
More informationNational Instruments Switches
ni.com National Instruments Switches Raviteja Chivukula Why the Need for Switches? Nearly every system can benefit from switching Increases channel count Adds measurement flexibility Simplifies test fixture
More information2750 Integra Mainframes
Selector Guide Plug-In Modules for 700, 70, 70 Integra Mainframes Module Selector Guide This selector guide may prove helpful in identifying the best module for a specific application. Install up to five
More informationButler Technologies. Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland. Calibration Laboratory Registration number: 256C
Unit 14, Block G, Maynooth Business Campus, Co. Kildare, Ireland Calibration Laboratory Registration number: 256C is accredited by the Irish National Board (INAB) to undertake calibration as detailed in
More informationModels 2601B, 2602B and 2604B
Models 2601B, 2602B and 2604B Keithley Instruments, Inc. System SourceMeter 28775 Aurora Road Instrument Specifications Cleveland, Ohio 44139 1-888-KEITHLEY http://www.keithley.com SPECIFICATION CONDITIONS
More informationSignal Conditioning Systems
Note-13 1 Signal Conditioning Systems 2 Generalized Measurement System: The output signal from a sensor has generally to be processed or conditioned to make it suitable for the next stage Signal conditioning
More informationUSB-TEMP and TC Series USB-Based Temperature Measurement Devices
USB-Based Temperature Measurement Devices Features Temperature and voltage measurement USB devices Thermocouple, RTD, thermistor, or semiconductor sensor measurements Eight analog inputs Up to ±10 V inputs*
More informationni.com Sensor Measurement Fundamentals Series
Sensor Measurement Fundamentals Series Introduction to Data Acquisition Basics and Terminology Litkei Márton District Sales Manager National Instruments What Is Data Acquisition (DAQ)? 3 Why Measure? Engineers
More informationFluke 8845A/8846A Digital Multimeters Extended Specifications
Fluke 8845A/8846A Digital Multimeters Extended Specifications The Fluke 8845A and 8846A 6.5 digit precision multimeters have the precision and versatility to handle your most demanding measurements, on
More informationVXI-TB CHANNEL ISOTHERMAL TERMINAL BLOCK
VXI-TB-1303 32-CHANNEL ISOTHERMAL TERMINAL BLOCK Introduction This guide describes how to install and use the VXI-TB-1303 terminal block with a VXI-SC submodule. The VXI-TB-1303 terminal block is a shielded
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
1 Wilson Street Contact: Mr Ibrahim Ibrahim Thornaby Tel: +44 (0) 1642 626148 Stockton-On-Tees Fax: +44 (0) 870 143 1869 TS17 7AR E-Mail: ibrahim@calibrate.co.uk United Kingdom Website: www.calibrate.co.uk
More informationMODEL INFORMATION MODEL 6010D. Automated Primary Resistance/ Thermometry Bridge
MODEL 6010D Automated Primary Resistance/ Thermometry Bridge Resistance & Temperature Applications Range 0.001 Ω to 100 KΩ Accuracy < 40 x 10-9 Linearity < 5 x 10-9 Featuring true ratio self calibration
More informationModel 2651A Specifications
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-55 http://www.tek.com/keithley Model 2651A Specifications High Power System SourceMeter Instrument Specifications SPECIFICATION CONDITIONS
More informationCX1100 Series Current and Differential Sensors
DATA SHEET CX00 Series Current and Differential Sensors CX0A Current Sensor, Single Channel CX0A Current Sensor, Dual Channel CX03A Current Sensor, Low Side CX04A Current Sensor, Selectable Resistive Sensor
More informationContents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer
CALIBRATION PROCEDURE NI PXIe-5665 3.6 GHz and 14 GHz RF Vector Signal Analyzer This document contains the verification procedures for the National Instruments PXIe-5665 (NI 5665) RF vector signal analyzer
More informationKeysight 34401A Multimeter
Keysight 34401A Multimeter Data Sheet See Keysight's Truevolt Series of DMMs You also get both GPIB and RS-232 interfaces as standard features. Voltmeter Complete and External Trigger signals are provided
More informationBasic Analog Circuits
Basic Analog Circuits Overview This tutorial is part of the National Instruments Measurement Fundamentals series. Each tutorial in this series, will teach you a specific topic of common measurement applications,
More information6½-Digit, ±300 V, 1.8 MS/s Isolated Digitizer Included, PCI Digital Multimeter Device
SPECIFICATIONS PCI-4070 6½-Digit, ±300 V, 1.8 MS/s Isolated Digitizer Included, PCI Digital Multimeter Device These specifications apply to the PCI-4070. Contents Definitions...2 Conditions... 2 DC Specifications...
More informationTechnical Services Department, BHEL, Piplani, Bhopal, Madhya Pradesh. Discipline Electro Technical Calibration Issue Date
Last Amended on 29.02.2016 Page 1 of 6 SOURCE 1. DC VOLTAGE $ 100mV to 220 mv 220 mv to 220 V 220 V to 1000 V 2. AC VOLTAGE $ 50 Hz to 1 khz 3 mv to 220 mv 220 mv to 220 V 220 V to 1000 V 3. DC CURRENT
More information2.5 GHz 75 Ω Multiplexer and SPDT Relay Switches
2.5 GHz Multiplexer and SPDT Relay Switches NI PXI-255x NEW! 2.5 GHz bandwidth characteristic impedance 30 V max switching voltage 0.5 A max switching current 10 W max switching power Mini SMB direct connectivity
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Metrology Division Contact: Mr K Pallett Calibration House Tel: +44 (0)1332 867 700 Castings Road E-Mail: sales@comech.co.uk Derby Website:
More informationUltra-Fast I-V Module for the Model 4200-SCS
Provides voltage outputs with programmable timing from 60ns to DC in 10ns steps Measure I and V simultaneously, at acquisition rates of up to 200 megasamples/second (MS/s) Choose from two voltage source
More informationNI Contents CALIBRATION PROCEDURE
CALIBRATION PROCEDURE NI 5450 Contents This document describes processes to calibrate the National Instruments PXIe-5450 (NI 5450) differential I/Q signal generator. This document provides performance
More informationDUAL ULTRA MICROPOWER RAIL-TO-RAIL CMOS OPERATIONAL AMPLIFIER
ADVANCED LINEAR DEVICES, INC. ALD276A/ALD276B ALD276 DUAL ULTRA MICROPOWER RAILTORAIL CMOS OPERATIONAL AMPLIFIER GENERAL DESCRIPTION The ALD276 is a dual monolithic CMOS micropower high slewrate operational
More informationDMM ½-Digit Graphical Sampling Multimeter
DMM 7510 7½-Digit Graphical Sampling Multimeter Datasheet The DMM7510 combines all the advantages of a precision digital multimeter, a graphical touchscreen display, and a high speed, high resolution digitizer
More informationPXIe Contents. Required Software CALIBRATION PROCEDURE
CALIBRATION PROCEDURE PXIe-5113 This document contains the verification and adjustment procedures for the PXIe-5113. Refer to ni.com/calibration for more information about calibration solutions. Contents
More informationLecture Notes Unit-III
Lecture Notes Unit-III FAQs Q1: An operational amplifier has a differential gain of 103 and CMRR of 100, input voltages are 120µV and 80µV, determine output voltage. 2 MARKS
More informationModel 1140A Thermocouple Simulator-Calibrator
BULLETIN 2031 Model 1140A Thermocouple Simulator-Calibrator The Model 1140A represents the latest innovation in thermocouple simulator-calibrators from Ectron, the originator of the Thermocouple Simulator
More informationContents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator
CALIBRATION PROCEDURE NI 5421/5441 100 MS/s Arbitrary Waveform Generator This document contains the verification and adjustment procedures for the NI 5421/5441 arbitrary waveform generator. This calibration
More information5320A. Multifunction Electrical Tester Calibrator. Extended Specifications
5320A Multifunction Electrical Tester Calibrator Extended Specifications General Specifications Warm-Up Time...30 minutes Specifications Confidence Level...99 % Specifications Interval...1 year Temperature
More informationDC Resistance Meters 2840 Series
Data Sheet DC Meters Touch screen to zoom, select, and enter values The DC resistance meters feature high accuracy and resolution measurements in the milliohm range. Both meters are ideally suited for
More informationHigh Voltage Component Production Testing with Two Model 2410 SourceMeter Units. Application Note Series. Introduction. Test System Configuration
A Tektronix Company Application Note Series Number 2058 igh Voltage Component Production Testing with Two Model 2410 SourceMeter Units Introduction Various production test applications require the use
More informationAPPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS
APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS DC POWER SUPPLIES We will discuss these instruments one at a time, starting with the DC power supply. The simplest DC power supplies are batteries which
More information6.5-Digit Multimeter, High- Accuracy, C-Size HP E1410A Technical Specifications
PDFINFO H 5 5 6 1-0 1 6.5-Digit Multimeter, High- Accuracy, C-Size Technical Specifications 1-Slot, C-size, message based Vdc/ac, 2- & 4-wire Ω Noise rejection with long integration times/ guarding Quality
More informationAnalog Signal Conditioning Accessories
NI 64-channel multiplexer mv, V, current, and thermocouple inputs NI 8-channel simultaneous sample-and-hold mv, V inputs NI SC-2042-RTD 8-channel RTD/thermistor RTD, thermistor, mv, V inputs NI 8-channel
More informationContents CALIBRATION PROCEDURE NI 5412
CALIBRATION PROCEDURE NI 5412 Contents Introduction... 2 Software... 2 Documentation... 3 Password... 4 Calibration Interval... 4 Test Equipment... 4 Test Conditions...5 Self-Calibration Procedures...
More informationSOURCE-MEASURE UNITS INCREASE PRODUCTIVITY AND ACCURACY IN AUTOMATED TESTING. Lee Stauffer. Keithley Instruments, Inc.
SOURCE-MEASURE UNITS INCREASE PRODUCTIVITY AND ACCURACY IN AUTOMATED TESTING Lee Stauffer Keithley Instruments, Inc. Introduction Source-Measure Units (SMUs) are more than the next generation of power
More informationDC MHZ PXI Differential Instrumentation Amplifier
DC - 100 MHZ PXI Differential Instrumentation Amplifier Differential 100 V Common Mode Input DC - 100 MHz Bandwidth AC/DC Coupling Programmable Attenuation/Gain/ Offset 9 nv/ Input Noise 50 Ω Output Impedance
More information350MHz, Ultra-Low-Noise Op Amps
9-442; Rev ; /95 EVALUATION KIT AVAILABLE 35MHz, Ultra-Low-Noise Op Amps General Description The / op amps combine high-speed performance with ultra-low-noise performance. The is compensated for closed-loop
More informationSignal Conditioning Fundamentals for PC-Based Data Acquisition Systems
Application Note 048 Signal Conditioning Fundamentals for PC-Based Data Acquisition Systems Introduction PC-based data acquisition (DAQ) systems and plugin boards are used in a very wide range of applications
More informationChapter 6 Specifications
Chapter 6 Specifications DC Characteristics Function Range [2] Test Current or Burden Voltage DC Voltage DC Current Resistance [6] 24 Hour [3] T CAL ±1 Accuracy Specifications: ± (% of reading + % of range)
More informationMeasurement & Control of energy systems. Teppo Myllys National Instruments
Measurement & Control of energy systems Teppo Myllys National Instruments National Instruments Direct operations in over 50 Countries More than 1,000 products, 7000+ employees, and 700 Alliance Program
More informationINPUT RESISTANCE OR CLAMP VOLTAGE. 2 Years 23 C ±5 Voltage mv 3 10 nv > 10 G
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 www.keithley.com Model 2010 Low-Noise Multimeter Specifications DC CHARACTERISTICS CONDITIONS: MED (1 PLC) 1 or SLOW (5 PLC)
More informationMultimeter Selection Guide Fluke 8508A & Agilent 3458/HFL
Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL TECHNICAL INFORMATION When comparing two products features and specification using sales information it is often difficult to determine which product
More informationPXI Contents SPECIFICATIONS. 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer, L and C Measurement Support, PXI Digital Multimeter
SPECIFICATIONS PXI-4072 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer, L and C Measurement Support, PXI Digital Multimeter These specifications apply to the PXI-4072. Contents Definitions...2 Conditions...
More informationData acquisition and instrumentation. Data acquisition
Data acquisition and instrumentation START Lecture Sam Sadeghi Data acquisition 1 Humanistic Intelligence Body as a transducer,, data acquisition and signal processing machine Analysis of physiological
More informationSolutions for Production Testing of Connectors. Application Note Series. Introduction. Test Description. Number 2208
Number 2208 Application Note Series Solutions for Production Testing of Connectors Introduction As electronics have become increasingly pervasive, the importance of electrical connectors has increased
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, Unit 8 The Centre Holywell Business Park Northfield Road Southam Warwickshire CV47 0FP Contact: Mr N Morgan Tel: +44 (0)1926 812066 Fax: +44
More informationPCI Express, 6½-Digit, ±300 V Digital Multimeter Device
SPECIFICATIONS PCIe-4065 PCI Express, 6½-Digit, ±300 V Digital Multimeter Device These specifications apply to the PCIe-4065. Contents Definitions...1 Conditions... 2 DC Specifications... 2 DC System Speed
More informationUnequalled combination of Accuracy, Stability and Versatility
üthe Source for Calibration Professionals The new microk family of precision thermometry bridges Unequalled combination of Accuracy, Stability and Versatility www.isotech.co.uk Introduction The microk
More informationIntegrating a PXI Automated Test System
Integrating a PXI Automated Test System Top Considerations to Integrating a PXI Automated Test System Agilent Technologies Software and Modular Solutions Alan Lesko Application Engineer April 24 th, 2012
More informationHigh Power Monolithic OPERATIONAL AMPLIFIER
High Power Monolithic OPERATIONAL AMPLIFIER FEATURES POWER SUPPLIES TO ±0V OUTPUT CURRENT TO 0A PEAK PROGRAMMABLE CURRENT LIMIT INDUSTRY-STANDARD PIN OUT FET INPUT TO- AND LOW-COST POWER PLASTIC PACKAGES
More information2790 SourceMeter Airbag Test System
2790 Meter Airbag Test System Single-instrument solution for continuity and hi-pot type leakage resistance measurements Programmable constant V source (50 500V) supports high speed, high resistance measurements
More informationCalsytech, # 38 North Mada Street Nandambakkam, Chennai, Tamil Nadu. Discipline Electro-Technical Calibration Issue Date
Last Amended on 23.11.2016 Page 1 of 7 SOURCE 1. DC VOLTAGE $ 1mV to 10 mv 10 mv to100 mv 100 mv to1000 V 0.43% to 0.05% 0.05% to 0.01% 0.01% to 0.004% Calibrator, Calibrator by DC VOLTAGE 1mV to 20 mv
More information