Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters

Size: px
Start display at page:

Download "Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters"

Transcription

1 Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters Data Sheet Basic + Good = Elegant Simplicity Features Up to 120,000 counts resolution Up to 0.012% basic DCV accuracy 11 basic measurements and up to six built-in math functions Dual display on bright VFD Selectable resolutions for variable measurement speeds* Kensington lock slot security * U3402A only ** U3401A only Simply right for your basic needs, without compromising quality The U3400 Series digital multimeters (DMMs) come with functions vital for your everyday measurements: DC, AC and AC+ DC voltage and current, 2- and 4-wire resistance, frequency, continuity and diode tests. Math functions dbm, Min/Max, Relative, Compare, Hold and Percentage** are also available to ease your measurement analysis. Not only are the U3400 Series built robust for lasting reliability over the long haul, these DMMs also ensure dependable measurements with up to 0.012% basic DCV accuracy. Efficient testing with dual display and selectable resolutions* Dual display on the U3400 Series lets you view two parameters simultaneously as you measure, and further enhances your troubleshooting tasks. For example, theability to see both AC voltage and frequency simultaneously would help you to measure the frequency response of amplifier circuits more efficiently and effectively. Refer to Page 3 for more typical dual display combinations and applications. Up to three measurement speeds are available on the U3402A: Slow, Medium and Fast. This means greater flexibility in catering to different testing needs: a faster measurement speed at lower resolution or a higher resolution at slower measurement speed. Physical security Your instruments may be at risk of theft or misplacement whenever you leave them unattended on the bench. With the U3400 Series rear Kensington lock slot, you can secure your DMM and be assured that it is where you expect it to be for your continued testing the next day.

2 Take a closer look Primary display Secondary display Input terminals and current fuse Measurement function keypads Autoranging, manual range and comparator operation Math operation keypads Shift Power on/off switch Figure 1 Front panel of the U3402A. The front panel of the U3401A is similar with slight differences in the position of certain functions. Please refer to the U3401A s User s and Service Guide U for details. AC line voltage selector Line voltage fuse selection table AC power connector AC line fuse Chassis ground lug RS232 interface connector (For calibration use only) Figure 2 Rear panel of the U3401A/U3402A. 2

3 Typical dual display combinations and applications Primary display Secondary display Application Available in DCV ACV Testing of DC-to-AC or AC-to-DC converter circuit U3402A, U3401A ACV + DCV DCV Power supply DC level and AC ripple measurement U3402A, U3401A DCV DCI Testing of power supply load regulation U3402A DCV ACI Loop current and voltage drop level checking U3402A ACI + DCI DCV Testing of line and load regulation U3402A ACV DCI Testing of AC-to-DC or DC-to-AC converter U3402A ACI + DCI ACV Power supply DC level and AC ripple measurement U3402A ACV ACI Transformer testing U3402A ACV Hz Amplifier circuit s AC frequency response measurement U3402A, U3401A ACI Hz Adjustment of AC motor control U3402A, U3401A DCI ACI Power supply AC ripple and DC current measurement U3402A, U3401A ACI + DCI DCI Current dissipation measurement U3402A, U3401A dbm Reference W U3401A dbm DCV RF frequency measurement U3402A, U3401A dbm ACV U3402A, U3401A dbm Hz Frequency response checking U3402A 3

4 U3400 Series specifications DC voltage DCV resolution, full-scale reading and accuracy [± (% of reading + count)] Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) Typical input impedance [1] mv 10 µv % MW V 100 µv % MW V 1 mv % MW V 10 mv % MW V 100 mv [2] 0.02% MW [1] Input impedance is in parallel with capacitance <100 pf. [2] In VDC 1000 V range, 1200 V is readable with audio warning. Rate Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) Slow mv 1 µv 0.012% + 8 [2] Medium Fast V V V V mv V V V V mv V V V 1000 V 10 µv 100 µv 1 mv 10 mv 10 µv 100 µv 1 mv 10 mv 100 mv 100 µv 1 mv 10 mv 100 mv 1 V [3] [3] [3] 0.012% % % % % % % % % % % % % % + 2 Typical input impedance [1] 10.0 MW 10.0 MW 11.1 MW 10.1 MW 10.0 MW 10.0 MW 11.1 MW 10.1 MW 10.0 MW 10.0 MW 10.0 MW 11.1 MW 10.1 MW 10.0 MW 10.0 MW [1] Input impedance is in parallel with capacitance <120 pf. [2] Relative (REL) operation is used. [3] In VDC 1000 V range, 1050 V is readable. 4

5 AC voltage (true RMS, AC coupling mode) ACV resolution, full-scale reading and accuracy [± (% of reading + count)] Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) [1] 30 Hz to 50 Hz 50 Hz to 10 khz 10 khz to 30 khz 30 khz to 100 khz mv 10 µv % % % % V 100 µv % % % % V 1 mv % % % % V 10 mv % % + 20 [2] 3% + 50 [2] V 100 mv % + 15 [3] 1% + 20 [2] [1] Accuracy specified at input >5% of full scale. [2] Input voltage <200 V RMS. [3] For 5 khz to 10 khz, accuracy is 0.7% [4] In VAC 750 V range, V RMS is readable with audio warning. U3402A 5.5-digit DMM Rate Range Resolution Maximum reading Slow Medium Fast mv V V V V mv V V V V mv V V V 750 V 1 µv 10 µv 100 µv 1 mv 10 mv 10 µv 100 µv 1 mv 10 mv 100 mv 100 µv 1 mv 10 mv 100 mv 1 V [4] Accuracy (1 year; 23 ºC ± 5 ºC) [1] 20 Hz to 45 Hz 45 Hz to 10 khz 10 khz to 30 khz 30 khz to 100 khz 1% % % % [2] 1% % % % [2] 1% % % % [2] 1% % % % [2] 1% [2] 0.2% % % [3] 1% % % % [2] 1% % % % + 80 [2] 1% % % % + 80 [2] 1% + 40 [2] 0.2% % % + 80 [2] 1% + 40 [2] 0.2% % % + 80 [3] 1% % % % + 15 [2] 1% % + 5 1% + 5 3% + 10 [2] 1% % + 5 1% + 5 3% + 10 [2] 1% + 5 [2] 0.2% + 5 1% + 5 3% + 10 [2] 1% + 5 [2] 0.2% + 5 1% + 5 3% + 10 [3] [1] Accuracy specified at input >5% of full scale. [2] Input voltage <200 V RMS. [3] Input voltage <500 V RMS. [4] In VAC 750 V range, V RMS is readable. 5

6 AC voltage (true RMS, AC+DC coupling mode) AC+DC voltage resolution, full-scale reading and accuracy [± (% of reading + count)] Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) [1] 50 Hz to 10 khz 10 khz to 30 khz 30 khz to 100 khz mv 10 µv % % % V 100 µv % % % V 1 mv % % % V 10 mv % % + 30 [2] 3% + 60 [2] V 100 mv [3] 0.5% + 25 [4] 1% + 30 [2] [1] Accuracy specified at input >5% of full scale. [2] Input voltage <200 V RMS. [3] In VAC 750 V range, V RMS is readable with audio warning. [4] For 5 khz to 10 khz, accuracy is 0.7% U3402A 5.5-digit DMM Rate Range Resolution Maximum reading Slow Medium Fast mv V V V V mv V V V V mv V V V 750 V 1 µv 10 µv 100 µv 1 mv 10 mv 10 µv 100 µv 1 mv 10 mv 100 mv 100 µv 1 mv 10 mv 100 mv 1 V [2] Accuracy (1 year; 23 ºC ± 5 ºC) [1] 45 Hz to 10 khz 10 khz to 30 khz 30 khz to 100 khz 0.2% % % % % % % % % % % % % % % [3] 0.2% % % % % % % % % % % % % % % + 83 [3] 0.2% % % % + 7 1% + 7 3% % + 7 1% + 7 3% % + 7 1% + 7 3% % + 7 1% + 7 3% + 12 [3] [1] Accuracy specified at input >5% of full scale. [2] In VAC 750 V range, V RMS is readable. [3] Input voltage <500 V RMS. 6

7 DC current DCI resolution, full-scale reading and accuracy [± (% of reading + count)] Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) Burden voltage [1] and shunt resistor µa 10 na % + 5 <0.06 V/100 W ma 100 na % + 4 <0.6 V/100 W ma 1 µa % + 4 <0.08 V/1 W ma % + 4 < V/1 W A % + 5 <0.3 V/0.01 W A 1 ma [2] 0.25% + 5 <0.6 V/0.01 W [1] Typical at full-scale reading and voltage across the input terminals. [2] In 10 A range, >10 to 20 ADC is readable for 20 seconds maximum with audio warning. U3402A 5.5-digit DMM Rate Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) Slow ma 0.1 µa 0.05% + 15 [2] Medium Fast ma ma A ma ma ma A ma ma 1200 ma A 1 µa 1 µa 1 ma 1 ma 10 ma % % % % % % % % % % % + 2 Burden voltage [1] and shunt resistor <0.15 V/10 W <1.5 V/10 W <0.3 V/0.1 W <0.6 V/0.01 W <0.5 V/10 W <1.5 V/10 W <0.3 V/0.1 W <0.6 V/0.01 W <0.5 V/10 W <1.5 V/10 W <0.3 V/0.1 W <0.6 V/0.01 W [1] Typical at full-scale reading and voltage across the input terminals. [2] Relative (REL) operation is used. 7

8 AC current (true RMS, AC coupling mode) ACI resolution, full-scale reading and accuracy [± (% of reading + count)] Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) [1] Burden voltage [2] 30 Hz to 50 Hz 50 Hz to 2 khz 2 khz to 5 khz 5 khz to 20 khz and shunt resistor µa 10 na % % % % + 75 [3] <0.06 V/100 W ma 100 na % % % % + 60 <0.6 V/100 W ma 1 µa % % % % + 60 <0.08 V/1 W ma % % % % + 60 < V/1 W A % + 40 [4] 0.5% + 20 <0.3 V/0.01 W A 1 ma [5] 2% + 40 [4] 0.5% + 30 (<1 khz) <0.6 V/0.01 W [1] Accuracy specified at input >5% of full scale and >1 A for 10 A range unless otherwise stated. [2] Typical at full-scale reading and voltage across the input terminals. [3] Input current >35 µa RMS. [4] Input current <3 A RMS. [5] In 10 A range, >10 to 20 ADC is readable for 20 seconds maximum with audio warning. U3402A 5.5-digit DMM Rate Range Resolution Maximum reading Slow Medium Fast ma ma ma A ma ma ma A ma ma 1200 ma A 0.1 µa 1 µa 1 µa 1 ma 1 ma 10 ma Accuracy (1 year; 23 ºC ± 5 ºC) [1] Burden voltage [2] 20 Hz to 45 Hz 45 Hz to 2 khz 2 khz to 10 khz and shunt resistor 1.5% % % <0.15 V/10 W 1.5% % % <1.5 V/10 W 1.5% % % <0.3 V/0.1 W 2% (<1.2 A) 1% <0.6 V/0.01 W 1.5% % % + 80 <0.5 V/10 W 1.5% % % + 30 <1.5 V/10 W 1.5% % % + 30 <0.3 V/0.1 W 1.5% + 12 (<1.2 A) 1% + 12 <0.6 V/0.01 W 1.5% % + 5 2% + 10 <0.5 V/10 W 1.5% % % + 5 <1.5 V/10 W 1.5% % % + 5 <0.3 V/0.1 W 2% + 2 (<1.2 A) 1% +2 <0.6 V/0.01 W [1] Accuracy specified at input >5% of full scale. [2] Typical at full-scale reading and voltage across the input terminals. 8

9 AC current (true RMS, AC+DC coupling mode) AC+DC current resolution, full-scale reading and accuracy [± (% of reading + count)] Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) [1] Burden voltage [2] 50 Hz to 2 khz 2 khz to 5 khz 5 khz to 20 khz and shunt resistor µa 10 na % % % + 85 [3] <0.06 V/100 W ma 100 na % % % + 70 <0.6 V/100 W ma 1 µa % % % + 70 <0.08 V/1 W ma % % % + 70 < V/1 W A % + 30 <0.3 V/0.01 W A 1 ma [4] 0.5% + 40 (<1 khz) <0.6 V/0.01 W [1] Accuracy specified at input >5% of full scale and >1 A for 10 A range unless otherwise stated. [2] Typical at full-scale reading and voltage across the input terminals. [3] Input current >35 µa RMS. [4] In 10 A range, >10 to 20 ADC is readable for 20 seconds maximum with audio warning. U3402A 5.5-digit DMM Rate Range Resolution Maximum reading Slow Medium Fast ma ma ma A ma ma ma A ma ma 1200 ma A 0.1 µa 1 µa 1 µa 1 ma 1 ma 10 ma Accuracy (1 year; 23 ºC ± 5 ºC) [1] Burden voltage [2] 45 Hz to 2 khz 2 khz to 10 khz and shunt resistor 0.5% % <0.15 V/10 W 0.5% % <1.5 V/10 W 0.5% % <0.3 V/0.1 W 1% <0.6 V/0.01 W 0.5% % + 80 <0.5 V/10 W 0.5% % + 30 <1.5 V/10 W 0.5% % + 30 <0.3 V/0.1 W 1% + 15 <0.6 V/0.01 W 0.5% + 7 2% + 12 <0.5 V/10 W 0.5% + 4 2% + 7 <1.5 V/10 W 0.5% + 4 2% + 7 <0.3 V/0.1 W 1% + 4 <0.6 V/0.01 W [1] Accuracy specified at input >5% of full scale. [2] Typical at full-scale reading and voltage across the input terminals. 9

10 Resistance Resistance resolution, full scale reading, and accuracy [± (% of reading + count)] Range [1] Resolution Maximum reading Test current Accuracy (1 year; 23 ºC ± 5 ºC) W 10 mw ma 0.1% + 5 [2] kw 100 mw ma 0.1% + 3 [2] kw 1 W µa 0.1% kw 10 W µa 0.1% MW 100 W na 0.1% MW 1 kw na 0.3% + 3 [1] To reduce noise interference that may be induced by the test leads, we recommend using a shielded test cable for measuring resistances above 500 kw. [2] Relative (REL) operation is used. U3402A 5.5-digit DMM Rate Range [1] Resolution Maximum reading Slow Medium Fast W kw kw kw MW MW MW W kw kw kw MW MW MW W kw kw kw MW MW MW 1 mw 10 mw 100 mw 1 W 10 W 100 W 1 kw 10 mw 100 mw 1 W 10 W 100 W 1 kw 10 kw 100 mw 1 W 10 W 100 W 1 kw 10 kw 100 kw Test current 0.5 ma 0.5 ma 1 µa 100 na 10 na 0.5 ma 50 µa 5 µa 500 na 50 na 10 na 0.5 ma 50 µa 5 µa 500 na 50 na 10 na Accuracy (1 year; 23 ºC ± 5 ºC) 2-wire 4-wire 0.1% + 8 [2] 0.05% + 8 [2] 0.08% + 5 [2] 0.05% + 5 [2] 0.06% + 5 [2] 0.05% % % % % % % + 5 3% + 8 3% % + 5 [2] 0.05% + 5 [2] 0.08% + 3 [2] 0.05% % % % % % % % % % % % + 2 [2] 0.05% % % % % % % % % % % % % + 2 [1] To reduce noise interference that may be induced by the test leads, we recommend using a shielded test cable for measuring resistances above 100 kw. [2] Relative (REL) operation is used. 10

11 Diode test/continuity Range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) V 100 µv V 0.05% + 5 U3402A 5.5-digit DMM Rate Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) Slow V µv 0.012% + 5 Medium V µv 0.012% + 5 Fast V mv 0.012% + 2 Frequency Frequency resolution, full-scale reading and accuracy [± (% of reading + count)] Range Measurement range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) Hz 5 Hz to 500 Hz 0.01 Hz % khz 500 Hz to 5 khz 0.1 Hz % khz 5 khz to 50 khz 1 Hz % khz 50 khz to 500 khz 10 Hz % + 3 Range Input sensitivity for voltage measurement (sine-wave) 5 Hz to 100 khz 100 khz to 500 khz 500 mv 35 mv RMS 200 mv RMS 5 V 0.25 V RMS 0.5 V RMS 50 V 2.5 V RMS 5 V RMS 500 V 25 V RMS 750 V 50 V RMS U3402A 5.5-digit DMM Range Measurement range Resolution Maximum reading Accuracy (1 year; 23 ºC ± 5 ºC) [1] [1] Specified accuracy at input >5% of full scale. Input sensitivity (sine wave) Hz 5 Hz to 1200 Hz 10 mhz % mv RMS khz 10 Hz to 12 khz 100 mhz 0.005% mv RMS khz 100 Hz to 120 khz 1 Hz 0.005% mv RMS MHz 1 khz to 1 MHz 10 Hz % V RMS 11

12 Decibel (db) calculation Range and accuracy (±db) Voltage range [1][2] Input voltage dbm [3] range at 600 W ref Accuracy (1 year; 23 ºC ± 5 ºC) 30 Hz to 50 Hz 50 Hz to 10 khz 10 khz to 100 khz mv 20 mv to 500 mv to V 5000 mv to 5 V 3.80 to V 5 V to 50 V to V 50 V to 500 V to [4] [4] VDC 500 V to 1000 V to [5] VAC 500 V to 750 V to [5] [1] Auto ranging is used when dbm operation is selected. [2] In VDC 1000 V range, 1200 V is readable. In VAC 750 V range, 1000 V is readable. [3] Reading is displayed in db when relative (REL) operation is used. [4] Input voltage <200 V RMS. [5] For input voltage within the frequency range of 50 Hz to 1 khz. U3402A 5.5-digit DMM Rate Voltage range [1][2] Input voltage dbm [3] range at 600 W ref Slow Medium Fast mv V V V VDC V mv V V V VDC V mv V V V 1000 VDC 750 V 6 mv to 120 mv 120 mv to 1.2 V 1.2 V to 12 V 12 V to 120 V 120 V to 1000 V 120 V to 750 V 20 mv to 400 mv 400 mv to 4 V 4 V to 40 V 40 V to 400 V 400 V to 1000 V 400 V to 750 V 20 mv to 400 mv 400 mv to 4 V 4 V to 40 V 40 V to 400 V 400 V to 1000 V 400 V to 750 V to to to to to to to to to to to to to to to to to to Accuracy (1 year; 23 ºC ± 5 ºC) 20 Hz to 45 Hz 45 Hz to 10 khz 10 khz to 100 khz [4] 1.0 [4] [4] 1.0 [4] [4] 1.0 [4] [1] Auto ranging is used when dbm operation is selected. [2] In VAC 750 V range, 5% over-range is readable. [3] Reading is displayed in db when relative (REL) operation is used. [4] For input voltage within the frequency range of 45 Hz to 1 khz. 12

13 Supplemental specifications Full-scale display counts Model Reading rate Display counts U3401A N/A 51,000 U3402A Slow 120,000 Medium 40,000 Fast 4,000 Supplemental measurement specifications Measurement Specification DC voltage Measurement method Sigma Delta A-to-D converter Input resistance Maximum input voltage Input protection 10 MW ± 2% range (typical) 1000 VDC or PEAK AC on all ranges 1000 V on all ranges Response time Approximately 1.0 second when the displayed reading reaches 99.9% DC value of the tested input signal at the same range DC current Shunt resistance U3401A 0.01 W to 100 W for 500 µa to 10 A ranges Maximum input and overload protection (U3402A only) Response time U3402A 0.1 W to 10 W for 12 ma to 1.2 A ranges 0.01 W for 12 A range ma input terminal: 1200 madc or AC RMS. Protected with 1.25 A/500 V, IEC-127 sheet, FB fuse 12 A input terminal: 10 ADC or AC RMS continuous; or 12 ADC or AC RMS for 30 seconds maximum. Protected with 15 A/600 V, breaking capacity 10,000 A FB fuse Approximately 1.0 second when the displayed reading reaches 99.9% DC value of the tested input signal at the same range AC voltage Measurement method AC-coupled true RMS: measures the AC component with up to 400 VDC bias on any range Crest factor Maximum 3:0 at full scale Input impedance U3401A: 1 MW in parallel with <100 pf U3402A: 1 MW ± 2% in parallel with <120 pf Maximum input voltage U3401A: 1000 V RMS/1400 V PEAK U3402A: 750 V RMS/1200 V PEAK 2x10 7 V-Hz product on any range, normal mode input 1x10 6 V-Hz product on any range, common mode input Overload ranging Selects higher range if peak input overload is detected during auto range. Overload is reported in manual ranging Input protection U3401A: 1000 V RMS on all ranges U3402A: 750 V RMS on all ranges Response time Approximately 1.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested input signal at the same range. 13

14 Supplemental measurement specifications Measurement AC+DC voltage Measurement method Specification AC+DC coupled true RMS: measures the AC component with up to 400 VDC bias on any range Crest factor Maximum 3:0 at full scale Input impedance U3401A: 1 MW in parallel with <100 pf U3402A: 1 MW ± 2% in parallel with <120 pf Maximum input voltage U3401A: 1000 V RMS/1400 V PEAK U3402A: 750 V RMS/1100 V PEAK 2x10 7 V-Hz product on any range, normal mode input 1x10 6 V-Hz product on any range, common mode input Overload ranging Selects higher range if peak input overload is detected during auto range. Overload is reported in manual ranging Input protection U3401A: 1000 V RMS on all ranges U3402A: 750 V RMS on all ranges Response time Approximately 1.5 seconds (2.5 seconds for U3402A) when the displayed reading reaches 99.9% AC RMS value of the tested input signal at the same range AC current Measurement method DC coupled to the fuse and current shunt, AC coupled true rms measurement (measures the AC component only) Crest factor Maximum 3:0 at full scale Shunt resistance U3401A 0.01 W to 100 W for 500 µa to 10 A ranges AC+DC current Input protection Response time Measurement method Crest factor Measurement range Response time U3402A 0.1 W to 10 W for 10 ma to 1.2 A ranges 0.01 W for 12 A range U3401A Front panel fuse 630 ma, 500 V; internal 25 A, 440 V U3402A ma input terminal: 1200 madc or AC RMS. Protected with 1.25 A/500 V, IEC-127 sheet, FH fuse 12 A input terminal: 10 ADC or AC RMS continuous; or 12 ADC or AC RMS for 30 seconds maximum. Protected with 15 A/600 V, breaking capacity 10,000 A FH fuse Approximately 1.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested input signal at the same range AC+DC coupled to the fuse and current shunt, AC+DC coupled true rms measurement (measures the AC component only) Maximum 3:0 at full scale Vdc and Vac are automatically set at the same range Approximately 1.5 seconds when the displayed reading reaches 99.9% AC RMS value of the tested input signal at the same range Resistance Measurement method U3401A: 2-wire ohms U3402A: 2-wire ohms or 4-wire ohms Open-circuit voltage U3401A: Limited to <+6 VDC U3402A: Limited to <+5 VDC Zeroing error 0.05 W or less (excluding test lead resistance) in each range when Relative operation is used 14

15 Supplemental measurement specifications Measurement Resistance Input protection 500 V Diode/ Continuity Response time Measurement method Specification U3401A Approximately 1.5 seconds for 5 MW and ranges below 5 MW U3402A Approximately 1.5 seconds for 12 MW and ranges below 12 MW Approximately 5 seconds for 40 MW Approximately 10 seconds for 120 MW Approximately 23 seconds for 300 MW U3401A: 0.5mA +/- 0.2% constant current source, open circuit voltage limited to <6V. U3402A: 0.5mA +/- 0.2% constant current source, open circuit voltage limited to <5V. Test current Approximately 0.5 madc Open-circuit voltage U3401A: Limited to <+6 VDC U3402A: Limited to <+5 VDC Continuity threshold 10 W fixed Continuity level Approximately <+50 mvdc Audible tone Continuous beep for continuity and single tone for normal forward-biased diode or semiconductor junction Input protection 500 V Resistance/ Measurement method 2-wire ohms Continuity Test current Approximately 0.5 madc (U3402A only) Open-circuit voltage Limited to <+5 VDC Audible tone Continuous beep for continuity and single tone for normal forward-biased diode or semiconductor junction Zeroing error 0.05 W or less (excluding test lead resistances) in each range when Relative operation is used Input protection 500 V Frequency Measurement method Reciprocal counting technique. AC coupled input using AC voltage function Crest factor Maximum 3:0 at full scale Signal level 10% of range to full-scale input on all ranges; auto or manual range selection Gate time 0.1 second or 1 period of the input signal, whichever is longer Input impedance U3401A: 1 MW in parallel with <100 pf U3402A: 1 MW ± 2% in parallel with <120 pf Noise rejection Maximum input voltage Input protection Response time Common mode rejection ratio (CMRR) for 1 kw unbalanced LO lead Normal mode rejection ratio (NMRR) U3401A: 1000 V RMS/1400 V PEAK U3402A: 750 V RMS/1100 V PEAK 2x10 7 V-Hz product on any range, normal mode input 1x10 6 V-Hz product on any range, common mode input 750 V RMS on all ranges Approximately 1.5 seconds when the displayed reading reaches 99.9% of frequency value 50/60 Hz ± 0.1%: DC >90 db 50/60 Hz ± 0.1%: >50 db 15

16 Supplemental measurement specifications Measurement dbm operation Specification 0 dbm 1 mw at 600 W reference impedance Resolution U3401A 0.01 db for all ranges U3402A Slow: 0.01 db for all ranges Medium: 0.01 db for all ranges Fast: 0.1 db for all ranges Reference impedance [1] 2 W [2], 4 W [2], 8 W [2], 16 W [2], 50 W, 75 W, 93 W, 110 W, 124 W, 125 W, 135 W, 150 W, 250 W, 300 W, 500 W, 600 W, 800 W, 900 W, 1000 W, 1200 W, 8000 W Math operation U3401A: dbm, Relative, Min/Max, Compare, Hold, Percentage U3402A: dbm, Relative, Min/Max, Compare, Hold I/O interface RS-232 (for calibration use only) [1] Reference impedance is displayed on the secondary display. [2] Reading is displayed in watts (audio power). 16

17 General characteristics Power supply 100 V/120 V/220 V/240 V ± 10% AC line frequency 50 Hz to 60 Hz Power consumption 16 VA maximum Input power option Manual-ranging (100 VAC to 240 VAC ± 10%) Fuse U3401A Terminal: 25 A, 440 V FB fuse; 0.63 A, 500 V FB fuse Power line: 0.25 A, 250 V SB fuse, or A, 250 V SB fuse U3402A Terminal: 15 A, 600 V FB fuse; 1.25 A, 500 V FB fuse Power line: 0.25 A, 250 V SB fuse, or A, 250 V SB fuse Display Highly visible vacuum fluorescent display (VFD) Operating environment Operating temperature from 0 C to +50 C Relative humidity up to 80% at 28 C RH (non-condensing) Altitude up to 2000 meters Pollution degree 2 For indoor use only Storage compliance 20 C to 60 C Relative humidity at 5% to 90% RH (non-condensing) Safety compliance IEC :2001/EN :2001 (2nd Edition) Canada: CAN/CSA-C22.2 No USA: ANSI/UL :2004 EMC compliance IEC :2005/EN :2006 Canada: ICES/NMB-001:2004 Australia/New Zealand: AS/NZS CISPR11:2004 Shock and vibration Tested to IEC/EN I/O connector Output connectors I/O interface RS-232 (for calibration use only) Dimensions (W H D) 255 mm 105 mm 305 mm (with bumpers) 215 mm x 87 mm x 282 mm (without bumpers) Weight 3.44 kg (with bumpers) Warranty Three years for U3401A/U3402A Three months for standard shipped accessories Calibration cycle One year Warm-up time At least 30 minutes 105 mm 255 mm 305 mm 17

18 Ordering Information Each U3400 Series includes these standard shipped accessories: Quick Start Guide Product Reference CD Certificate of Calibration 34138A Test Lead Kit AC Power Cord Accessories 34138A Test Lead Kit U1161A Extended Test Lead Kit 34330A Current Shunt (30 A) 34133A Precision Electronic Test Leads 11059A Kelvin Probe Set U3400A-1CM Rack Mount Kit 18

19 myagilent myagilent A personalized view into the information most relevant to you. AdvancedTCA Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Agilent is a founding member of the AXIe consortium. LAN extensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consortium. PCI extensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system. Three-Year Warranty Beyond product specification, changing the ownership experience. Agilent is the only test and measurement company that offers three-year warranty on all instruments, worldwide. Agilent Assurance Plans Five years of protection and no budgetary surprises to ensure your instruments are operating to specifications and you can continually rely on accurate measurements. Agilent Electronic Measurement Group DEKRA Certified ISO 9001:2008 Quality Management System For more information on Agilent Technologies products, applications or services, please contact your local Agilent office. The complete list is available at: Americas Canada (877) Brazil (11) Mexico United States (800) Asia Pacific Australia China Hong Kong India Japan 0120 (421) 345 Korea Malaysia Singapore Taiwan Other AP Countries (65) Europe & Middle East Belgium 32 (0) Denmark Finland 358 (0) France * *0.125 /minute Germany 49 (0) Ireland Israel /544 Italy Netherlands 31 (0) Spain 34 (91) Sweden United Kingdom 44 (0) Basic + Good = Elegant Simplicity Agilent Channel Partners Get the best of both worlds: Agilent s measurement expertise and product breadth, combined with channel partner convenience. Some of the best equations express large ideas with elegant simplicity. The same is true with the Agilent U3401A and U3402A dual-display DMMs. When you need basic capabilities with good performance at 4½ or 5½ digits, respectively, these affordable benchtop units make the grade in educational, electronics and communications applications. For extra credit, they also include a PC-grade lock slot for physical security. The U3400 Series: basic, good capability in an elegantly simple DMM. For other unlisted countries: (BP ) Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2010, 2013 Published in USA, December 21, EN

Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters

Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters Data Sheet Basic + Good = Elegant Simplicity Features Up to 120,000 counts resolution Up to 0.012% basic DCV accuracy 11 basic measurements and

More information

Keysight Technologies U3400 Series 4½ and 5½ Digit Digital Multimeters

Keysight Technologies U3400 Series 4½ and 5½ Digit Digital Multimeters Keysight Technologies U3400 Series 4½ and 5½ Digit Digital Multimeters Product discontinuance notice: The U3401A/U3402A DMM and all associated options will be discontinued December 1, 2017. The last date

More information

Agilent Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price. Data Sheet

Agilent Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price. Data Sheet Agilent Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price Data Sheet Features 120000 counts resolution 16 built-in measurement functions including temperature

More information

Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size

Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size Data Sheet Features 1-Slot, C-size, message-based DCV, ACV, DCI, ACI, 2/4-wire Ω, frequency, period NULL, MIN/MAX, LIMIT, db, dbm 1000 reading/s

More information

Keysight Technologies 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM. Data Sheet

Keysight Technologies 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM. Data Sheet Keysight Technologies 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet 02 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Features Fast reading speed of up to

More information

Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price

Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price Data Sheet Features 120000 counts resolution 16 built-in measurement functions including temperature and

More information

Agilent U2741A USB Modular 5.5 Digits Digital Multimeter. Data Sheet

Agilent U2741A USB Modular 5.5 Digits Digital Multimeter. Data Sheet Agilent U2741A USB Modular 5.5 Digits Digital Multimeter Data Sheet Features Makes fast measurements with up to 100 readings per second Measures up to 300 VDC with 5.5 digits resolution Introduction The

More information

Agilent U1230 Series Handheld Digital Multimeter (DMM)

Agilent U1230 Series Handheld Digital Multimeter (DMM) Agilent U1230 Series Handheld Digital Multimeter (DMM) Data Sheet Whether it is dark, noisy or even dangerous, the U1230 Series handheld digital multimeter keeps you equipped with features that anticipate

More information

N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes

N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes Data Sheet Oscilloscope users often need to make floating measurements where neither point of the measurement is at earth

More information

Agilent Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price. Data Sheet

Agilent Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price. Data Sheet Agilent Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price Data Sheet DC CHARACTERISTICS [1] ACCURACY ± (% of reading + % of range) Temperature Cefficient

More information

Agilent N432A Thermistor Power Meter

Agilent N432A Thermistor Power Meter Agilent N432A Thermistor Power Meter Data Sheet Key features Frequency range: 100 khz to 18 GHz (sensor dependent) Power range: 30 dbm (1 μw) to +10 dbm (10 mw) High accuracy ± (0.1% of reading + 0.5 μw),

More information

Agilent 34405A Multimeter

Agilent 34405A Multimeter 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price Data Sheet Features 120000

More information

Keysight Technologies 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM. Data Sheet

Keysight Technologies 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM. Data Sheet Keysight Technologies 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet 02 Keysight 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Data Sheet Features Fast reading speed of up to

More information

DATA SHEET A Multimeter 5.5 Digit Dual Display, Benchtop DMM

DATA SHEET A Multimeter 5.5 Digit Dual Display, Benchtop DMM DATA SHEET 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM Features Fast reading speed of up to 190 readings/sec 0.015% DCV accuracy Multiple connectivity options USB 2.0, Serial Interface (RS-232)

More information

N2820A/21A High-Sensitivity, High Dynamic Range Current Probes

N2820A/21A High-Sensitivity, High Dynamic Range Current Probes N2820A/21A High-Sensitivity, High Dynamic Range Current Probes Data Sheet See the details without losing sight of the big picture Key features and specifications Measure currents as low as 50 µa Measure

More information

Agilent U1730C Series Handheld LCR Meters

Agilent U1730C Series Handheld LCR Meters Agilent U1730C Series Handheld LCR Meters Take your expectations higher with the latest LCR meters Data Sheet Agilent s U1730C Series handheld LCR meters allow you to measure at frequencies as high as

More information

Agilent U1240 Series Handheld Digital Multimeters

Agilent U1240 Series Handheld Digital Multimeters Agilent U1240 Series Handheld Digital Multimeters Data Sheet Helping You Check and Fix More Installation and Maintenance Bugs Key Features Check more, fix more ACI, diode, continuity tests Capacitance,

More information

Keysight Technologies U1210 Series Handheld Clamp Meters

Keysight Technologies U1210 Series Handheld Clamp Meters Keysight Technologies U1210 Series Handheld Clamp Meters Handle big currents more safely Data Sheet Introduction Measurements of electrical distribution cables can be challenging and risky. For cables

More information

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes Data Sheet Fast, automatic and reliable characterization of switching mode power devices Today s power supply

More information

Keysight Technologies U1730C Series Handheld LCR Meters Take your expectations higher with the latest LCR meters. Data Sheet

Keysight Technologies U1730C Series Handheld LCR Meters Take your expectations higher with the latest LCR meters. Data Sheet Keysight Technologies U1730C Series Handheld LCR Meters Take your expectations higher with the latest LCR meters Data Sheet Introduction The Keysight Technologies, Inc. U1730C Series handheld LCR meters

More information

N2750A/51A/52A InfiniiMode Differential Active Probes

N2750A/51A/52A InfiniiMode Differential Active Probes N2750A/51A/52A InfiniiMode Differential Active Probes Data Sheet Key Features Measurement Versatility 1.5 GHz, 3.5 GHz, and 6 GHz probe bandwidth models Dual attenuation ratio (2:1/10:1) High input resistance

More information

The Keysight Technologies, Inc. U1730C Series handheld LCR meters allow you to measure at frequencies as high as 100 khz a capability typically found

The Keysight Technologies, Inc. U1730C Series handheld LCR meters allow you to measure at frequencies as high as 100 khz a capability typically found The Keysight Technologies, Inc. U1730C Series handheld LCR meters allow you to measure at frequencies as high as 100 khz a capability typically found only in benchtop meters. Get measurements done faster

More information

7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes.

7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes. 7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes. Achieving maximized measurement dynamic range 1) Use averaging to increase measurement resolution 2) Use high-resolution

More information

Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer. Application Note

Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer. Application Note Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer Application Note Introduction This note details the input impedance of the U8903B Audio Analyzer, and shows that this needs to

More information

Keysight U3606B Multimeter I DC Power Supply

Keysight U3606B Multimeter I DC Power Supply Keysight U3606B Multimeter I DC Power Supply Convenient and Full Featured. One-box source-and-measure device. Data Sheet Introduction Features Combination of a 5.5 digit digital multimeter and 30-W power

More information

MXG X-Series Signal Generator N5183B Microwave Analog

MXG X-Series Signal Generator N5183B Microwave Analog MXG X-Series Signal Generator N5183B Microwave Analog Configuration Guide This configuration guide will help you determine which performance, software applications, accessories, and services to include

More information

Keysight Technologies 1 mw 50 MHz Power Reference Measurement with the N432A Thermistor Power Meter. Application Note

Keysight Technologies 1 mw 50 MHz Power Reference Measurement with the N432A Thermistor Power Meter. Application Note Keysight Technologies 1 mw 50 MHz Power Reference Measurement with the N432A Thermistor Power Meter Application Note Introduction This application note explains the application procedure for using the

More information

Keysight Technologies Making Current-Voltage Measurement Using SMU

Keysight Technologies Making Current-Voltage Measurement Using SMU Keysight Technologies Making Current-Voltage Measurement Using SMU Keysight B2901A/02A/11A/12A Precision Source/Measure Unit Demonstration Guide Introduction The Keysight Technologies, Inc. B2901A/02A/11A/12A

More information

Keysight Technologies N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes. Data Sheet

Keysight Technologies N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes. Data Sheet Keysight Technologies N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes Data Sheet 02 Keysight N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes

More information

Agilent U1250 Series Handheld Digital Multimeters

Agilent U1250 Series Handheld Digital Multimeters Agilent U1250 Series Handheld Digital Multimeters Data Sheet Test more more safely The Agilent U1250 Series handheld digital multimeters (DMMs) exceed your expectations by delivering powerful features

More information

Agilent NFA Noise Figure Analyzer

Agilent NFA Noise Figure Analyzer Agilent NFA Noise Figure Analyzer Configuration Guide Dedicated Noise Figure Analyzer Hard specifications to 26.5 GHz Works with N4000A SNS or 346 Series noise sources Noise figure measurements to 110

More information

Agilent E5061B Network Analyzer. 100 khz to 1.5 GHz/3 GHz 5 Hz to 3 GHz

Agilent E5061B Network Analyzer. 100 khz to 1.5 GHz/3 GHz 5 Hz to 3 GHz Agilent E5061B Network Analyzer 100 khz to 1.5 GHz/3 GHz 5 Hz to 3 GHz E5061B responds to various measurement needs, - from LF to RF The Agilent E5061B is a member of the industry standard ENA Series network

More information

Keysight Technologies N432A Thermistor Power Meter. Data Sheet

Keysight Technologies N432A Thermistor Power Meter. Data Sheet Keysight Technologies N432A Thermistor Power Meter Data Sheet 02 Keysight N432A Thermistor Power Meter Data Sheet Why Keysight s Power Meters and Sensors? Keysight s only power meter that supports thermistor

More information

Keysight Technologies Simultaneous Measurements with a Digital Multimeter

Keysight Technologies Simultaneous Measurements with a Digital Multimeter Keysight Technologies Simultaneous Measurements with a Digital Multimeter Application Brief Test Challenges: Making more confident measurements Making dual measurements in less time 02 Keysight Simultaneous

More information

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options Data Sheet For InfiniiVision 3000, 4000 and 6000 X-Series Oscilloscopes Achieve cost-effective analysis of your switching mode

More information

Agilent N6780 Series Source/Measure Units (SMUs) for the N6700 Modular Power System

Agilent N6780 Series Source/Measure Units (SMUs) for the N6700 Modular Power System Agilent N6780 Series Source/Measure Units (SMUs) for the N6700 Modular Power System Data Sheet N6781A 2-Quadrant Source/Measure Unit for Battery Drain Analysis N6782A 2-Quadrant Source/Measure Unit for

More information

Keysight 8762F Coaxial Switch 75 ohm

Keysight 8762F Coaxial Switch 75 ohm Keysight 8762F Coaxial Switch 75 ohm Technical Overview DC to 4 GHz Exceptional repeatability over 1 million cycle life Excellent isolation The 8762F brings a new standard of performance to 75 ohm coaxial

More information

Introduction. Part 1. Introduction...2

Introduction. Part 1. Introduction...2 Keysight Technologies Simple Scalar Network Analysis of Frequency Converter Devices using the U2000 USB Power Sensor Series with the ENA Network Analyzer Application Note Introduction This application

More information

Evaluating Oscilloscopes for Low-Power Measurements

Evaluating Oscilloscopes for Low-Power Measurements Evaluating Oscilloscopes for Low-Power Measurements Application Note Increasing market demand for products that are portable, mobile, green, and that can stay powered for long periods of time is driving

More information

Keysight Technologies 7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes. Application Note

Keysight Technologies 7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes. Application Note Keysight Technologies 7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes Application Note Seven Hints for Making Power Measurements with Oscilloscopes Achieving maximized

More information

Keysight Technologies Accurate Evaluation of MEMS Piezoelectric Sensors and Actuators Using the E4990A Impedance Analyzer.

Keysight Technologies Accurate Evaluation of MEMS Piezoelectric Sensors and Actuators Using the E4990A Impedance Analyzer. Keysight Technologies Accurate Evaluation of MEMS Piezoelectric Sensors and Actuators Using the E4990A Impedance Analyzer Application Note Introduction Excellent impedance measurement accuracy and repeatability

More information

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview Keysight Technologies N9398C/F/G and N9399C/F DC Block Technical Overview Introduction Key Features Maximize your operating range - 26.5, 50 or 67 GHz Improve calibration accuracy with exceptional return

More information

Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes

Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes Data Sheet Oscilloscope training tools created specifically for electrical engineering and physics undergraduate students

More information

N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes

N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes Data Sheet Oscilloscope users often need to make floating measurements where neither point of the measurement is at earth

More information

Agilent U1700 Series Handheld LCR Meters

Agilent U1700 Series Handheld LCR Meters Agilent U1700 Series Handheld LCR Meters Data Sheet Test passive components conveniently, affordably and reliably with the Agilent U1700 Series LCR meters extending the tradition of industryleading benchtop

More information

Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes

Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes Data Sheet Oscilloscope training tools created specifically for electrical engineering and physics undergraduate students

More information

Agilent N9310A RF Signal Generator. All the capability and reliability of an Agilent instrument you need at a price you ve always wanted

Agilent N9310A RF Signal Generator. All the capability and reliability of an Agilent instrument you need at a price you ve always wanted Agilent N9310A RF Signal Generator All the capability and reliability of an Agilent instrument you need at a price you ve always wanted Reliable Performance. Essential Test Capability The N9310A RF signal

More information

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview Keysight Technologies N9398C/F/G and N9399C/F DC Block Technical Overview Introduction Key Features Maximize your operating range - 26.5, 50 or 67 GHz Improve calibration accuracy with exceptional return

More information

Agilent U9391C/F/G Comb Generators

Agilent U9391C/F/G Comb Generators Agilent U9391C/F/G Comb Generators U9391C (10 MHz to 26.5 GHz) U9391F (10 MHz to 50 GHz) U9391G (10 MHz to 67 GHz) Technical Overview Key Features Excellent amplitude and phase flatness enable it to be

More information

Agilent 2-Port and 4-Port PNA-X Network Analyzer

Agilent 2-Port and 4-Port PNA-X Network Analyzer Agilent 2-Port and 4-Port PNA-X Network Analyzer N5244A - MHz to 43.5 GHz N5245A - MHz to 5. GHz with Option H29 Data Sheet and Technical Specifications Documentation Warranty THE MATERIAL CONTAINED IN

More information

HP 34401A Specifications 8

HP 34401A Specifications 8 8 HP 34401A Specifications 8 DC Characteristics DC Characteristics Accuracy Specifications ± ( % of reading + % of range ) [ 1 ] Function Range [ 3 ] Test Current or Burden Voltage 24 Hour [ 2 ] 23 C ±

More information

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes Data Sheet Fast, automatic and reliable characterization of switching mode power devices Today s power supply

More information

Keysight Technologies U9391C/F/G Comb Generators. U9391C (10 MHz to 26.5 GHz) U9391F (10 MHz to 50 GHz) U9391G (10 MHz to 67 GHz) Technical Overview

Keysight Technologies U9391C/F/G Comb Generators. U9391C (10 MHz to 26.5 GHz) U9391F (10 MHz to 50 GHz) U9391G (10 MHz to 67 GHz) Technical Overview Keysight Technologies U9391C/F/G Comb Generators U9391C (10 MHz to 26.5 GHz) U9391F (10 MHz to 50 GHz) U9391G (10 MHz to 67 GHz) Technical Overview Key Features Excellent amplitude and phase flatness enable

More information

Keysight Technologies Enhance EMC Testing with Digital IF. Application Note

Keysight Technologies Enhance EMC Testing with Digital IF. Application Note Keysight Technologies Enhance EMC Testing with Digital IF Application Note Introduction With today s accelerating business environment and development cycles, EMC measurement facilities that offer rapid

More information

True RMS Bench Multimeters 2831E and 5491B

True RMS Bench Multimeters 2831E and 5491B Data Sheet True RMS Bench Multimeters and True RMS Bench Multimeters with Dual Display The B&K Precision models and are versatile and dependable bench multimeters suitable for applications in education,

More information

Data Sheet. Agilent M9185A PXI Isolated D/A Converter. DISCOVER the Alternatives... Agilent MODULAR Products. 8/16-Channel 16-bit, ±16 V

Data Sheet. Agilent M9185A PXI Isolated D/A Converter. DISCOVER the Alternatives... Agilent MODULAR Products. 8/16-Channel 16-bit, ±16 V Agilent M9185A PXI Isolated D/A Converter Data Sheet 8/16-Channel 16-bit, ±16 V DISCOVER the Alternatives...... Agilent MODULAR Products Overview Introduction The Agilent M9185A is a digital/analog converter

More information

Agilent U1401A/U1401B Handheld Multi-function Calibrator/Meter

Agilent U1401A/U1401B Handheld Multi-function Calibrator/Meter Agilent U1401A/U1401B Handheld Multi-function Calibrator/Meter Data Sheet Calibrate while you measure with just one tool Agilent handheld multi-function calibrator/meter has all you need for quick validation,

More information

InfiniiMax III probing system

InfiniiMax III probing system InfiniiMax III probing system Data Sheet World s highest speed and highest performing probe system Full 30 GHz bandwidth to the probe tip Industry s lowest probe and scope system noise Industry s highest

More information

Keysight Technologies How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer.

Keysight Technologies How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer. Keysight Technologies How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer Application Note Introduction RF IN RF OUT Waveform Generator Pulse Power Amplifier

More information

Keysight Technologies 87405C 100 MHz to 18 GHz Preamplifier. Technical Overview

Keysight Technologies 87405C 100 MHz to 18 GHz Preamplifier. Technical Overview Keysight Technologies 8745C 1 MHz to 18 GHz Preamplifier Technical Overview 2 Keysight 8745C 1 MHz to 18 GHz Preamplifier Technical Overview Introduction The Keysight Technologies, Inc. 8745C preamplifier

More information

Keysight Technologies Making Simpler DC Power Measurements with a Digital Multimeter

Keysight Technologies Making Simpler DC Power Measurements with a Digital Multimeter Keysight Technologies Making Simpler DC Power Measurements with a Digital Multimeter Application Brief Test Challenges: Measuring DC voltage and current with a single digital multimeter Measuring watts

More information

Agilent 34410A and 34411A Multimeters Setting the Standard for Next Generation Benchtop and System Testing

Agilent 34410A and 34411A Multimeters Setting the Standard for Next Generation Benchtop and System Testing Agilent 34410A and 34411A Multimeters Setting the Standard for Next Generation Benchtop and System Testing Product Overview Agilent 34410A 6 1 / 2 -Digit High-Performance DMM 10,000 readings/s @ 5 1 /

More information

Agilent N4916B De-emphasis Signal Converter

Agilent N4916B De-emphasis Signal Converter Agilent N4916B De-emphasis Signal Converter Data Sheet, Version 1.1 NEW! Extended bit rate to 14.2 Gb/s Accurately characterize your multi-gigabit serial interfaces with the 4-tap de-emphasis signal converter

More information

Keysight U1882B Measurement Application for Infiniium Oscilloscopes. Data Sheet

Keysight U1882B Measurement Application for Infiniium Oscilloscopes. Data Sheet Keysight U1882B Measurement Application for Infiniium Oscilloscopes Data Sheet 02 Keysight U1882B Measurement Application for Infiniium Oscilloscopes - Data Sheet Fast, Automatic and Reliable Characterization

More information

Keysight 34401A Multimeter

Keysight 34401A Multimeter Keysight 34401A Multimeter Data Sheet See Keysight's Truevolt Series of DMMs You also get both GPIB and RS-232 interfaces as standard features. Voltmeter Complete and External Trigger signals are provided

More information

Agilent First-generation MXG Signal Generators

Agilent First-generation MXG Signal Generators Agilent First-generation MXG Signal Generators N5181A MXG Analog N5182A MXG Vector Configuration Guide You Can Upgrade! Options can be added after initial purchase. This guide is to assist in the ordering

More information

Keysight Technologies U1230 Series Handheld Digital Multimeters (DMMs) Data Sheet

Keysight Technologies U1230 Series Handheld Digital Multimeters (DMMs) Data Sheet Keysight Technologies U1230 Series Handheld Digital Multimeters (DMMs) Data Sheet Introduction Retool your expectations with the new Keysight Technologies, Inc. U1230 Series Handheld DMMs the first to

More information

Keysight Technologies N432A Thermistor Power Meter. Data Sheet

Keysight Technologies N432A Thermistor Power Meter. Data Sheet Keysight Technologies N432A Thermistor Power Meter Data Sheet 02 Keysight N432A Thermistor Power Meter Data Sheet Why Keysight s Power Meters and Sensors? Keysight s only power meter that supports thermistor

More information

Keysight N9310A RF Signal Generator

Keysight N9310A RF Signal Generator Keysight N9310A RF Signal Generator 9 khz to 3.0 GHz Data Sheet 02 Keysight N9310A RF Signal Generator - Data Sheet Definitions and Conditions Specifications describe the performance of parameters that

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Agilent N1911A/N1912A P-Series Power Meters and N1921A/N1922A Wideband Power Sensors Data Sheet LXI Class-C-Compliant

More information

Agilent U1250 Series Handheld Digital Multimeters

Agilent U1250 Series Handheld Digital Multimeters Agilent U1250 Series Handheld Digital Multimeters Data Sheet Test more more safely The Agilent U1250 Series handheld digital multimeters (DMMs) exceed your expectations by delivering powerful features

More information

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options Data Sheet For InfiniiVision 3000, 4000 and 6000 X-Series Oscilloscopes 02 Keysight DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement

More information

Agilent N9320B RF Spectrum Analyzer

Agilent N9320B RF Spectrum Analyzer Agilent N9320B RF Spectrum Analyzer 9 khz to 3.0 GHz Data Sheet Definitions and Conditions The spectrum analyzer will meet its specifications when: It is within its calibration cycle It has been turned

More information

True RMS Bench Multimeters 2831E and 5491B

True RMS Bench Multimeters 2831E and 5491B Data Sheet True RMS Bench Multimeters and TRUE RMS True RMS Bench Multimeters with Dual Display The B&K Precision models and are versatile and dependable bench multimeters suitable for applications in

More information

Keysight Technologies Achieving Accurate E-band Power Measurements with E8486A Waveguide Power Sensors. Application Note

Keysight Technologies Achieving Accurate E-band Power Measurements with E8486A Waveguide Power Sensors. Application Note Keysight Technologies Achieving Accurate E-band Power Measurements with Waveguide Power Sensors Application Note Introduction The 60 to 90 GHz spectrum, or E-band, has been gaining more millimeter wave

More information

Keysight Technologies 87405C 100 MHz to 18 GHz Preamplifier. Technical Overview

Keysight Technologies 87405C 100 MHz to 18 GHz Preamplifier. Technical Overview Keysight Technologies 8745C 1 MHz to 18 GHz Preamplifier Technical Overview 2 Keysight 8745C 1 MHz to 18 GHz Preamplifier Technical Overview Introduction The Keysight Technologies, Inc. 8745C preamplifier

More information

SDM3045X Digital Multimeter. DataSheet

SDM3045X Digital Multimeter. DataSheet SDM3045X Digital Multimeter DataSheet-2016.06 User-friendly Design Product Overview SDM3045X is a 4½ digit digital (60000 count) multimeter incorporating a dual-display and is especially well suited for

More information

Keysight Technologies 423B, 8470B, 8472B, 8473B/C Low Barrier Schottky Diode Detectors

Keysight Technologies 423B, 8470B, 8472B, 8473B/C Low Barrier Schottky Diode Detectors Keysight Technologies 423B, 8470B, 8472B, 8473B/C Low Barrier Schottky Diode Detectors Keysight 423B Data Sheet Keysight 8470B Keysight 8472B Keysight 8473B Keysight 8473C Introduction Excellent broadband

More information

Agilent U1250 Series Handheld Digital Multimeters

Agilent U1250 Series Handheld Digital Multimeters Agilent U1250 Series Handheld Digital Multimeters Data Sheet 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Test more more safely Visit us at www.testequipmentdepot.com

More information

Keysight N9320B RF Spectrum Analyzer

Keysight N9320B RF Spectrum Analyzer Keysight N9320B RF Spectrum Analyzer 9 khz to 3.0 GHz Data Sheet 02 Keysight N9320B RF Spectrum Analyzer - Data Sheet Definitions and Conditions Specifications describe the performance of parameters covered

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

Essential Capabilities of EMI Receivers. Application Note

Essential Capabilities of EMI Receivers. Application Note Essential Capabilities of EMI Receivers Application Note Contents Introduction... 3 CISPR 16-1-1 Compliance... 3 MIL-STD-461 Compliance... 4 Important features not required by CISPR 16-1-1 or MIL-STD-461...

More information

Agilent 970-Series Handheld Multimeters Data Sheet

Agilent 970-Series Handheld Multimeters Data Sheet Agilent 970-Series Handheld Multimeters Data Sheet Benchtop features and performance with handheld convenience and price 3 1 /2and 4 1 /2 digits with dcv accuracy to 0.05% 1 khz to 100 khz frequency response

More information

Agilent U1270 Series Handheld Digital Multimeters

Agilent U1270 Series Handheld Digital Multimeters Agilent U1270 Series Handheld Digital Multimeters Data Sheet NEW Features OLED display for high brightness level and wider viewing angles (up to 160 degrees) 1 Intelligent features: Z LOW 1, 2, Smart Ω

More information

Keysight Technologies 8490G Coaxial Attenuators. Technical Overview

Keysight Technologies 8490G Coaxial Attenuators. Technical Overview Keysight Technologies 8490G Coaxial Attenuators Technical Overview Introduction Key Specifications Maximize your operating frequency range for DC to 67 GHz application Minimize your measurement uncertainty

More information

Keysight E5063A ENA Series Network Analyzer

Keysight E5063A ENA Series Network Analyzer Keysight E5063A ENA Series Network Analyzer 100 khz to 500 M/1.5 G/3 G/4.5 G/6.5 G/8.5 G/14 G/18 GHz Configuration Guide 02 Keysight E5063A ENA Series Network Analyzer - Configuration Guide Ordering Guide

More information

Keysight Technologies Measuring Low Current Consumption with a Digital Multimeter

Keysight Technologies Measuring Low Current Consumption with a Digital Multimeter Keysight Technologies Measuring Low Current Consumption with a Digital Multimeter Application Brief Test Challenges: Characterizing the power consumption of a battery powered device Testing the current

More information

INPUT RESISTANCE OR CLAMP VOLTAGE. 2 Years 23 C ±5 Voltage mv 3 10 nv > 10 G

INPUT RESISTANCE OR CLAMP VOLTAGE. 2 Years 23 C ±5 Voltage mv 3 10 nv > 10 G Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 www.keithley.com Model 2010 Low-Noise Multimeter Specifications DC CHARACTERISTICS CONDITIONS: MED (1 PLC) 1 or SLOW (5 PLC)

More information

Keysight Technologies Secondary Radar Transponder Testing Using the 8990B Peak Power Analyzer. Application Note

Keysight Technologies Secondary Radar Transponder Testing Using the 8990B Peak Power Analyzer. Application Note Keysight Technologies Secondary Radar Transponder Testing Using the 8990B Peak Power Analyzer Application Note Introduction After a brief review of radar systems and the role of transponders, this application

More information

Keysight Technologies 87222C/D/E Coaxial Transfer Switches

Keysight Technologies 87222C/D/E Coaxial Transfer Switches Keysight Technologies 87C/D/E Coaxial Transfer Switches dc to 6.5, 0, 50 GHz Technical Overview High Performance Transfer Switches for Micro wave and RF Instrumentation and Systems Exceptional repeatability

More information

True RMS Bench Multimeter 5492B

True RMS Bench Multimeter 5492B Data Sheet True RMS Bench Multimeter 5492B True RMS Bench Multimeter The B&K Precision model 5492B is a versatile 5½-digit, 120,000-count bench multimeter suitable for applications in education, service

More information

Keysight Technologies 85072A 10-GHz Split Cylinder Resonator. Technical Overview

Keysight Technologies 85072A 10-GHz Split Cylinder Resonator. Technical Overview Keysight Technologies 85072A 10-GHz Split Cylinder Resonator Technical Overview 02 Keysight 85072A 10-GHz Split Cylinder Resonator - Technical Overview Part of the complete turn-key solution for the IPC

More information

Keysight Technologies N4916B De-emphasis Signal Converter

Keysight Technologies N4916B De-emphasis Signal Converter Keysight Technologies N4916B De-emphasis Signal Converter Data Sheet Version 1.11 Accurately characterize your multi-gigabit serial interfaces with the 4-tap de-emphasis signal converter N4916B with optional

More information

Agilent N9342C Handheld Spectrum Analyzer (HSA)

Agilent N9342C Handheld Spectrum Analyzer (HSA) Agilent N9342C Handheld Spectrum Analyzer (HSA) Data Sheet Field testing just got easier The Agilent N9342C handheld spectrum analyzer (HSA) is more than easy-to-use its measurement performance gives you

More information

Agilent 8761A/B Microwave Switches

Agilent 8761A/B Microwave Switches Agilent 8761A/B Microwave Switches Technical Overview Product Description The Agilent Technologies 8761A and 8761B are single-pole, double-throw coaxial switches with excellent electrical and mechanical

More information

Keysight Technologies N2792A/N2818A 200 MHz and N2793A/N2819A 800 MHz Differential Probes. Data Sheet

Keysight Technologies N2792A/N2818A 200 MHz and N2793A/N2819A 800 MHz Differential Probes. Data Sheet Keysight Technologies N2792A/N2818A 200 MHz and N2793A/N2819A 800 MHz Differential Probes Data Sheet Introduction The Keysight Technologies, Inc. N2792A/93A and N2818A/19A differential probes provide the

More information

Agilent Spectrum Visualizer (ASV) Software. Data Sheet

Agilent Spectrum Visualizer (ASV) Software. Data Sheet Agilent Spectrum Visualizer (ASV) Software Data Sheet Technical Overview The Agilent spectrum visualizer (ASV) software provides advanced FFT frequency domain analysis for the InfiniiVision and Infiniium

More information

Keysight Technologies P9400A/C Solid State PIN Diode Transfer Switches

Keysight Technologies P9400A/C Solid State PIN Diode Transfer Switches Keysight Technologies P9400A/C Solid State PIN Diode Transfer Switches P9400A 100 MHz to 8 GHz PIN transfer switch P9400C 100 MHz to 18 GHz PIN transfer switch Technical Overview Key Features Minimize

More information

Keysight Technologies, Inc. Overcome PCB Loss and Deliver a Clean Eye to Your DUT Using Multi-tap De-emphasis

Keysight Technologies, Inc. Overcome PCB Loss and Deliver a Clean Eye to Your DUT Using Multi-tap De-emphasis Keysight Technologies, Inc. Overcome PCB Loss and Deliver a Clean Eye to Your DUT Using Multi-tap De-emphasis Application Brief Introduction Keysight Technologies, Inc. announces a new 32 Gb/s pattern

More information

Agilent 34410A and 34411A Multimeters Setting the Standard for Next Generation Benchtop and System Testing

Agilent 34410A and 34411A Multimeters Setting the Standard for Next Generation Benchtop and System Testing Agilent 34410A and 34411A Multimeters Setting the Standard for Next Generation Benchtop and System Testing Product Overview Agilent 34410A 6 1 / 2 -Digit High-Performance DMM 10,000 readings/s @ 5 1 /

More information