CDAX 605 High Precision Capacitance & Dissipation Factor Test Set
|
|
- Dustin Allison
- 6 years ago
- Views:
Transcription
1 CDAX 605 High Precision Capacitance & Dissipation Factor Test Set 1
2 Tan delta and capacitance measurements Hi V A Lo Ground C HL Measure with AC test signal, use Ohms law to calculate: Dissipation factor (tan delta) Power factor Capacitance C L C H 2
3 Megger test sets for capacitance/tan delta testing Application/Measurement CB100 CDAX605 IDAX/VAX DELTA4000 Power frequency tan delta/power factor Yes Yes Yes Yes Capacitance measurements Yes Yes Yes Yes External voltage source No Yes No No High voltage tan delta/power factor No Yes Yes Yes High power tan delta/power factor No Yes No Yes Laboratory precision measurements No Yes No No Dielectric frequency response No No 0.1 mhz-10 khz Hz Moisture assessment No No Yes No Individual temperature correction No No Yes Yes Temperature dependence analysis No No Yes No Tip-up testing No Yes Yes Yes 3
4 CDAX 605 General High-precision impedance, capacitance and dissipation factor measurement instrument Used together with a high-voltage AC voltage source and a calibrated standard capacitor to form a complete setup for insulation testing IDAX605 can be used for high precision ratio measurements, e.g. calibration of voltage transformers Testing can be performed at any voltage level depending on the rating of the equipment, the power source and the reference capacitor IDAX605 performs measurements in all four quadrants i.e. the instrument handles measurement of capacitive, resistive and inductive components (also in combination) 4
5 CDAX 605 Key features Direct readings of capacitance and dissipation factor, no balancing or calculation required Direct reading of measured ratio Very high accuracy Capacitance 0.02% Dissipation factor 0.001% Wide measurement range Capacitance up to 100 µf Dissipation factor phase measurements. Reference and test objects can be capacitive, resistive or inductive Works with any reference value without any recalculations Measurements synchronized to power frequency or generator/reference signal (5-400 Hz) Measures UST-R, UST-B, UST-RB, GST-GND, GSTg-R, GSTg-B and GSTg-RB configurations using 3 inputs and two separate current measurement channels IDAX605 SW + LabView and C# interfaces 5
6 CDAX 605 Application Capacitance & tan delta 6
7 CDAX 605 Application Ratio measurements - + STANDARD CAPACITOR 2 GENERATOR STANDARD CAPACITOR 7
8 CDAX 605 Application Ratio measurements Test voltage is measured using a standard capacitor connected to the voltage source Another standard capacitor (or resistor) is connected to the secondary output of the VT The capacitance value is measured Voltage ratio = Capacitance value/measured capacitance value Direct reading of measured ratio Example: Test voltage 100 kv VT ratio 230 kv/115 V = 2000 Capacitor value pf Measured capacitance = pf Measured voltage ratio / = Secondary voltage V 8
9 CDAX Software 9
10 CDAX605 test modes UST: Ungrounded Specimen Testing Test mode Measure Ground Guard UST-R Red Blue UST-B Blue Red Guard = Ground UST-RB Red and Blue --- GST: Grounded Specimen Testing Test mode Measure Ground Guard GST-GND Ground Red and Blue --- GSTg-R Ground Blue Red GSTg-B Ground Red Blue GSTg-RB Ground --- Red and Blue 10
11 C HL : UST-R A C HL C L C H 11
12 C HL +C H : GST-GND A C H L C L C H 12
13 C H : GSTg-RB 2 A C HL C L C H 13
14 CDAX 605 Standard capacitors, kv 14
15 CDAX 605 Test generator examples, kv 15
16 High Precision Ratio Measurements of Capacitive Coupled Voltage Transformers 16
17 CVT General CVT/CCVT Capacitive Coupled Voltage Transformer Typically used for voltages 69 kv 765 kv (1200 kv in India?) Provides voltage potential for revenue meters and sometimes protection circuits Capacitive divider down to 5-12 kv followed by a standard inductive step-down voltage transformer to typically 115 V Series reactor connected from capacitive divider to step-down transformer. Reactor impedance is designed so its impedance cancels the impedance of the capacitor Full intermediate voltage to the step-down transformer Typical ratio range: 500 to 5000 Capacitive: Electromagnetic: Accuracy down to 0.1% (revenue metering) 17
18 Typical CVT design (ABB) 1. Primary terminal 2. Expansion tank 3. Porcelain 4. Expansion tank for multi-section capacitors 5. Low-voltage terminal switch 6. Compensating reactor 7. Main primary windings 8. Secondary windings 9. Trimming windings 10. Damping device 11. Ground terminal C1 High voltage capacitor C2 Intermediate voltage capacitor 18
19 Typical CVT schematic diagram (Trench) 1. Series reactors 2. Intermediate voltage transformers 3. Harmonic suppression filter 4. Sealed protective gap 5. Secondary terminal board 6. Faraday shield 7. Potential ground switch 8. Choke coil and gap assembly 9. Drain coil, gap and carrier ground switch assembly 19
20 CVT testing Factory Dielectric (insulation) tests designed to verify the CVT s insulation integrity Accuracy tests designed to verify CVT s compliance to a specified accuracy class Leak tests (unit is pressurized and checked for leaks) 20
21 CVT ratio testing LV measurements (100 V TTR) High ratio very low secondary voltage. May be difficult to measure accurately, especially in the field Test voltage from TTR instrument is stepped down times in the capacitive divider Very low test voltage to step-down transformer Impedance error risk Inaccuracy is generally > 1% HV measurements (2-5 kv TTR) Higher test voltage reduces accuracy issues Inaccuracies < 0.5% is achievable HV measurements ( kv) capacitance method Very high test voltage (up to rated voltage) removes accuracy issues Final accuracy is pending accuracy of standard capacitors (as well as test equipment) Inaccuracies < 0.1% is achievable 21
22 CVT ratio measurement factory setup (bridge) 22
23 CVT ratio testing with traditional bridge High accuracy by using bridge technique Manual instrument Needs balancing Manual data collection, interpretation and reporting Difficult to create an automated test No computer interface Can only measure capacitance Reference capacitor on LV circuit Higher capacitance (>1000 pf) is desired to get higher test current Expensive and limited number of suppliers 23
24 CVT ratio measurement factory setup (direct) Can be exchanged for a reference resistor 24
25 CVT ratio testing using direct measurements Processor based instrument No balancing Direct readings Automated data collection and interpretation Computer interface to standard lab SW Can measure capacitance, resistance and inductance in combination Possibility to use a reference resistor on LV circuit Less expensive and can easily be designed for higher measurement current Accuracy in line with traditional bridge technique 25
26 Summary CVT ratio measurements CVT ratio needs to be verified in factory tests (and sometimes also in the field ) CVT s with 0.1% accuracy class needs to be tested with a test setup that provides an inaccuracy well below 0.1%. Only HV methods are possible to use Traditional tan delta and capacitance measurement bridges provides very good accuracy but Needs balancing Have manual data collection, interpretation and reporting Are difficult to integrate in an automated test setup Reference components are limited to capacitors Direct voltage-current measurement instruments can measure capacitance, resistance and inductance in combination and provides No balancing, direct readings, automated data collection Computer interface to standard lab SW e.g. LabView Possibility to use a reference resistor on LV circuit Accuracy in line with traditional bridge technique 26
27 CDAX605 Application example Ratio measurements on revenue class CCVT:s Ratio and insulation measurements on revenue class CCVTs (0.1% accuracy) Direct replacement of CB-605 (old Multi-Amp measuring bridge) Direct reading of measured ratio Possibility to use a resistor reference on LV Calibration of CCVTs up to highest accuracy grade, with support for full automation using LabVIEW and/or C# - 27
28 CDAX605 Demo box Easy to demonstrate Both capacitance and ratio Adjustable output voltage 28
29 CRD605 HV resistor 20 MΩ Max 2 kv (100 µa) To be used for ratio measurements Or in any other application requesting a stable high-voltage resistor 29
30 30
Capacitive Voltage Transformer (CVT) Commissioning Test
Capacitive Voltage Transformer (CVT) Commissioning Test Fahad Rashid, Engineer, Revenue Metering, IESO June 12, 2018 Outline IT Checks Market Rule Reference Commissioning Market Manual Requirements Capacitive
More information4. CAPACITIVE VOLTAGE TRANSFORMERS AND COUPLING CAPACITORS Oil-paper insulation
4. CPCITIVE VOLTGE TRNSFORMERS ND COUPLING CPCITORS 420 kv Capacitive voltage transformers. Fingrid, Visulahti (Finland). 34 Instrument transformers High voltage INTRODUCTION Capacitive voltage transformers
More informationIDAX 300 Insulation Diagnostic Analyzer. Dielectric Frequency Response Also known as: Frequency Domain Spectroscopy
IDAX 300 Insulation Diagnostic Analyzer Dielectric Frequency Response Also known as: Frequency Domain Spectroscopy 1 Frequency Domain Spectroscopy Hi V A Lo Ground C HL Measure at several frequencies Use
More informationTesting & Calibration Lab, 204, Diamond Industrial Estate No. 2, Ketki Pada Road, (Near Dahisar Toll Naka), Dahisar (East), Mumbai, Maharashtra
Last Amended on - Page 1 of 5 SOURCE 1. DC VOLTAGE 1 mv to 32 mv 0.43 % to 0.02 % 32 mv to 32 V 0.02 % to 0.008 % 32 V to 1000 V 0.008 % to 0.02 % 2. DC CURRENT 10 µa to 320 µa 0.15 % to 0.03 % 32 µa to
More informationMatz Ohlen Director Transformer Test Systems. Megger Sweden
Matz Ohlen Director Transformer Test Systems Megger Sweden Frequency response analysis of power transformers Measuring and analyzing data as function of frequency, variable frequency diagnostics Impedance
More informationCapacitive voltage transformers
Capacitive voltage transformers Outdoor operation Oil-paper insulated ECF (72 550) kv General description Capacitive voltage transformers of type ECF are used in high-voltage switchgears from 72 to 550
More informationLecture 4 Power System Instrumentation. Course map
Lecture 4 Power System Instrumentation 1 Course map 2 1 Outline of the Lecture Instrument Transformers Voltage Transformer Current Transformers Measurement Setups Instrumentation 3 The Current Transformer
More informationPower Factor Insulation Diagnosis: Demystifying Standard Practices
Power Factor Insulation Diagnosis: Demystifying Standard Practices Dinesh Chhajer, PE 4271 Bronze Way, Dallas Tx Phone: (214) 330 3238 Email: dinesh.chhajer@megger.com ABSTRACT Power Factor (PF) testing
More informationTechnical Services Department, BHEL, Piplani, Bhopal, Madhya Pradesh. Discipline Electro Technical Calibration Issue Date
Last Amended on 29.02.2016 Page 1 of 6 SOURCE 1. DC VOLTAGE $ 100mV to 220 mv 220 mv to 220 V 220 V to 1000 V 2. AC VOLTAGE $ 50 Hz to 1 khz 3 mv to 220 mv 220 mv to 220 V 220 V to 1000 V 3. DC CURRENT
More informationMEDIUM & HIGH VOLTAGE
MEDIUM & HIGH VOLTAGE TESTING EQUIPMENT VOLTAGE WITHSTAND SGM Series Resonant Systems The SGM series are used for generating high AC voltages at a fixed frequency (mainly 50 or 60 Hz) by means of an excited
More informationUnderstanding the Value of Electrical Testing for Power Transformers. Charles Sweetser, OMICRON electronics Corp. USA
Understanding the Value of Electrical Testing for Power Transformers Charles Sweetser, OMICRON electronics Corp. USA Understanding the Value of Electrical Testing for Power Transformers Charles Sweetser,
More informationELTEL CALIBRATION LABORATORY
NABL ACCREDITED ELTEL CALIBRATION LABORATORY Our laboratory is NABL ACCREDITED in Electro-Technical Discipline as per the international standard IS/ISO/IEC-17025:2005 for both In-House & Site Calibrations.
More informationTRAX Transformer and Substation Test System
TRAX Replaces need for multiple test sets Saves time by eliminating need for multiple instruments learning User-friendly interface reduces training and testing time Portable and compact system components
More informationCalsytech, # 38 North Mada Street Nandambakkam, Chennai, Tamil Nadu. Discipline Electro-Technical Calibration Issue Date
Last Amended on 23.11.2016 Page 1 of 7 SOURCE 1. DC VOLTAGE $ 1mV to 10 mv 10 mv to100 mv 100 mv to1000 V 0.43% to 0.05% 0.05% to 0.01% 0.01% to 0.004% Calibrator, Calibrator by DC VOLTAGE 1mV to 20 mv
More informationTRAX Transformer and Substation Test System
Replaces need for multiple test sets Saves time by eliminating need for multiple instruments learning User-friendly interface reduces training and testing time Portable and compact system components for
More informationKNOW MORE ABOUT THE TRANSFORMERS. Glossary Transformers
KNOW MORE ABOUT THE TRANSFORMERS Glossary Transformers Ambient temperature The existing temperature of the atmosphere surrounding a transformer installation. Ampere The practical unit of electric current.
More informationSubstation Preventive Maintenance
Substation Preventive Maintenance PROVINCIAL ELECTRICITY AUTHORITY 1 Presentation Contents 1) A kind of substation 2) Electrical equipment details of AIS substation 3) Electrical equipment details of GIS
More informationROEVER ENGINEERING COLLEGE ELAMBALUR, PERAMBALUR DEPARTMENT OF ELECTRICAL & ELECTRONICS ENGINEERING
ROEVER ENGINEERING COLLEGE ELAMBALUR, PERAMBALUR 621 212 DEPARTMENT OF ELECTRICAL & ELECTRONICS ENGINEERING EE1003 HIGH VOLTAGE ENGINEERING QUESTION BANK UNIT-I OVER VOLTAGES IN ELECTRICAL POWER SYSTEM
More informationTest description for dry-type transformers chapter for special tests
Test description for dry-type transformers chapter for special tests 1. SCOPE 4 2. STANDARDS 5 3. LIGHTNING IMPULSE TEST 6 4. SOUND LEVEL MEASUREMENT 7 4.1. STANDARD 7 4.2. AIM 7 4.3. THEORETICAL PRINCIPAL
More informationMitigating Murphy s Law While Test. Frédéric Dollinger
Mitigating Murphy s Law While Test Frédéric Dollinger Curriculum Vitae Frédéric Dollinger HAEFELY HIPOTRONICS factory Basel Switzerland Area Sales & Marketing Manager Dipl. Ing. / M.Sc. Mechatronic Language:
More informationApplication Note. Application for Precision Impedance Meters in a Standards Laboratory. Required Capabilities for Precision Measurements
Application for Precision Impedance Meters in a Standards Laboratory The IET Labs 1689 Precision RLC Digibridge, which measures resistance, capacitance and inductance, has found wide acceptance in production
More informationCapacitive voltage transformers
Capacitive voltage transformers Outdoor operation Oil-paper insulated ECF (72 550) kv General description Capacitive voltage transformers of type ECF are used in high-voltage switchgears from 72 to 550
More informationTRAX Transformer and Substation Test System
TRAX Replaces need for multiple test sets Saves time by eliminating need for multiple instruments learning User-friendly interface reduces training and testing time Portable and compact system components
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Standards & Calibration Laboratory Contact: Mr Martin Matthews Archcliffe Road Tel: +44 (0)1304 502121 Dover Fax: +44 (0)1304 502268 Kent
More informationPower Engineering II. High Voltage Testing
High Voltage Testing HV Test Laboratories Voltage levels of transmission systems increase with the rise of transmitted power. Long-distance transmissions are often arranged by HVDC systems. However, a
More informationHigh Voltage Engineering
High Voltage Engineering Course Code: EE 2316 Prof. Dr. Magdi M. El-Saadawi www.saadawi1.net E-mail : saadawi1@gmail.com www.facebook.com/magdi.saadawi 1 Contents Chapter 1 Introduction to High Voltage
More informationDEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING QUESTION BANK SUBJECT CODE & NAME : EE 1402 HIGH VOLTAGE ENGINEERING UNIT I
DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING QUESTION BANK SUBJECT CODE & NAME : EE 1402 HIGH VOLTAGE ENGINEERING YEAR / SEM : IV / VII UNIT I OVER VOLTAGES IN ELECTRICAL POWER SYSTEMS 1. What
More informationModule 2 : Current and Voltage Transformers. Lecture 8 : Introduction to VT. Objectives. 8.1 Voltage Transformers 8.1.1Role of Tuning Reactor
Module 2 : Current and Voltage Transformers Lecture 8 : Introduction to VT Objectives In this lecture we will learn the following: Derive the equivalent circuit of a CCVT. Application of CCVT in power
More informationElectronics Test and Development Centre, STQC Directorate, Agriculture College Campus, Shivajinagar, Pune, Maharashtra
Last Amended on 23.02.2015 Page 1 of 8 SOURCE 1. DC VOLTAGE # 100 V to 1 mv 0.5 % to 0.06 % Using Multi-Function Calibrator 1 mv to 10 mv 0.06 % to 0.007 % by 10 mv to 100 mv 0.007 % to 0.0014 % 100 mv
More informationUniversity of Pennsylvania Department of Electrical and Systems Engineering ESE319
University of Pennsylvania Department of Electrical and Systems Engineering ESE39 Laboratory Experiment Parasitic Capacitance and Oscilloscope Loading This lab is designed to familiarize you with some
More information10. DISTURBANCE VOLTAGE WITHSTAND CAPABILITY
9. INTRODUCTION Control Cabling The protection and control equipment in power plants and substations is influenced by various of environmental conditions. One of the most significant environmental factor
More informationA New Method for the Calibration of the mv Ranges of an AC Measurement Standard
A New Method for the Calibration of the mv Ranges of an AC Measurement Standard Speaker/Author Neil Faulkner Fluke Corporation PO Box 9090, Everett, WA 98206 Phone: (425) 446-5538 FAX: (425) 446-5649 E-mail:
More informationChapter 30 Inductance, Electromagnetic. Copyright 2009 Pearson Education, Inc.
Chapter 30 Inductance, Electromagnetic Oscillations, and AC Circuits 30-7 AC Circuits with AC Source Resistors, capacitors, and inductors have different phase relationships between current and voltage
More informationPrimary Metering. What is Primary Metering?
NWEMS Primary Metering August 22, 2018 Bill Unbehaun, Tacoma Power Exchanging Expertise Since 1893 What is Primary Metering? Metering energy flow past a point at high voltage above 600v Both PTs and CTs
More informationUnderstanding the Value of Electrical Testing for Power Transformers. Charles Sweetser - OMICRON
Understanding the Value of Electrical Testing for Power Transformers Charles Sweetser - OMICRON Transformers Diagnostic Testing - OVERALL DGA Oil Screen Power Factor / Capacitance Exciting Current Transformer
More informationTMS 580. Transformer Loss Measuring System FEATURES BENEFITS APPLICATIONS
TMS 580 Transformer Loss Measuring System The measurement of the losses in power transformers is an indispensable quality-verification process. Due to the fact that normally the transformer user puts a
More informationHIGH VOLTAGE ENGINEERING(FEEE6402) LECTURER-24
LECTURER-24 GENERATION OF HIGH ALTERNATING VOLTAGES When test voltage requirements are less than about 300kV, a single transformer can be used for test purposes. The impedance of the transformer should
More informationAdvanced Test Equipment Rentals ATEC (2832) CP RC. Resonance circuit for GIS testing
Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) CP RC Resonance circuit for GIS testing A new approach to testing gas-insulated switchgear Testing gas-insulated switchgear
More information700 Series AC Dielectric Test Sets
700 Series AC Dielectric Test Sets AC Test Systems HAEFELY HIPOTRONICS standard line of AC Test Systems is designed to perform high voltage AC tests on electrical apparatus in accordance with IEC60, IEEE
More informationPower Measurements and Basic Electrical Diagnostic Tests
Power Measurements and Basic Electrical Diagnostic Tests Instrument Basics Burden VA Sources V and I Meters V and I KVL and KCL Kelvin Connection KVL and KCL Kelvin Connection 4-Wire Technique Exclude
More informationConnection Impact Assessment Application
Connection Impact Assessment Application This form is for generators applying for Connection Impact Assessment (CIA) and for generators with a project size >10 kw. Please return the completed form by email,
More informationUsing DFR to determine dissipation factor temperature dependence
Using DFR to determine dissipation factor temperature dependence by Matz Ohlen and Peter Werelius, Megger With an aging power component population, today s electrical utility industry faces a tough challenge
More informationDELTA Reference Manual Applications Guide. 12 kv Insulation Diagnostic System ZM-AH02E
DELTA 4000 12 kv Insulation Diagnostic System Reference Manual Applications Guide WWW.MEGGER.COM ZM-AH02E DELTA 4000 12 kv Insulation Diagnostic System Reference Manual Applications Guide NOTICE OF COPYRIGHT
More informationFerroresonance in MV Voltage Transformers: Pragmatic experimental approach towards investigation of risk and mitigating strategy
Ferroresonance in MV Voltage Transformers: Pragmatic experimental approach towards investigation of risk and mitigating strategy W. Piasecki, M. Stosur, T. Kuczek, M. Kuniewski, R. Javora Abstract-- Evaluation
More informationTransformer Factory Testing
Transformer Factory Testing John J. Foschia Test Engineer John.Foschia@spx.com September 2018 Reasons for Testing Compliance with user specifications Assessment of quality and reliability Verification
More informationFGJTCFWP"KPUVKVWVG"QH"VGEJPQNQI[" FGRCTVOGPV"QH"GNGEVTKECN"GPIKPGGTKPI" VGG"246"JKIJ"XQNVCIG"GPIKPGGTKPI
FGJTFWP"KPUKWG"QH"GEJPQNQI[" FGRTOGP"QH"GNGETKEN"GPIKPGGTKPI" GG"46"JKIJ"XQNIG"GPIKPGGTKPI Resonant Transformers: The fig. (b) shows the equivalent circuit of a high voltage testing transformer (shown
More informationRLCG measurements (imitance)
RLCG measurements (imitance) Ján Šaliga KEMT FEI TU Košice 2017 Real electronic components Real components are NT ideal (perfect) It they were ideal the would have only one physical property: resistivity,
More informationMODEL INFORMATION. AccuLoss MODELS ALMS 2100, ALMS 4100, ALMS 4200, ALMS 6200 & ALMS 4300
AccuLoss MODELS ALMS 2100, ALMS 4100, ALMS 4200, ALMS 6200 & ALMS 4300 The measurement of electric power and energy at high voltages and currents at low power factors is becoming increasingly important
More informationUnderstanding and Extracting Valuable Information from Basic and Advanced Power Transformer Testing Techniques
Understanding and Extracting Valuable Information from Basic and Advanced Power Transformer Testing Techniques Charles Sweetser, Services Manager, PRIM Engineering, Waltham, Mass. Topics of Discussion
More informationVoltage Probe Manual and Data North Star High Voltage, Inc. Rev January 2016
561 Rose Loop NE Bainbridge Island, WA, USA 9811 (52)78-93; (26)219-425 FAX http://www.highvoltageprobes.com probes@highvoltageprobes.com Voltage Probe Manual and Data North Star High Voltage, Inc. Rev
More informationON-SITE TESTING OF SPECIAL TRANSFORMERS
ON-SITE TESTING OF SPECIAL TRANSFORMERS Simanand GANDHI JEYARAJ, Megger Ltd, UK, Simanand.gandhi@megger.com, Robert MILNE, UK Power Networks, UK, robert.milne83@googlemail.com Grant MITCHELL, Transmag
More informationTab 2 Voltage Stresses Switching Transients
Tab 2 Voltage Stresses Switching Transients Distribution System Engineering Course Unit 10 2017 Industry, Inc. All rights reserved. Transient Overvoltages Decay with time, usually within one or two cycles
More informationPRE COMMISSIONING TESTS ON EQUIPMENT AT 33/11 KV SUB STATIONS. IR Values are to be read on the megger by meggering the Power transformer
PRE COMMISSIONING TESTS ON EQUIPMENT AT 33/11 KV SUB STATIONS TESTS ON TRANSFORMERS 1. IR Values This is measured to measure the Insulation Resistance of the whole transformer. a) For 33/11 KV Power Transformer
More informationError vs. Uncertainty Historical Perspective
1 Error vs. Uncertainty Historical Perspective Jim McBride Chairman PSIM Committee Vice-Chairman HVTT Subcommittee IEEE PES SPDC Fall 2017 Clearwater, FL Discussions on Uncertainty PSIM - HVTT Subcommittee
More informationChapter 7 Conclusion 7.1 General
Chapter 7 7.1 General The mechanical integrity of a transformer winding is challenged by several mechanisms. Many dielectric failures in transformers are direct results of reduced mechanical strength due
More informationTR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:
CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL LF METROLOGY Approved By: Senior Manager: Mpho Phaloane Revised By: Specialist Technical Committee Date of Approval: 2015-08-26 Date of
More informationDC current interruption tests with HV mechanical DC circuit breaker
http: //www.cigre.org CIGRÉ A3/B4-124 CIGRÉ Winnipeg 2017 Colloquium Study Committees A3, B4 & D1 Winnipeg, Canada September 30 October 6, 2017 DC current interruption tests with HV mechanical DC circuit
More informationHands-On Transformer Testing and Maintenance
Hands-On Course Description This Hands-On course will teach you how to prioritize your transformer maintenance strategy, stretch your maintenance budget and at the same time maximize the life and condition
More informationMethods of secondary short circuit current control in single phase transformers
2015; 1(8): 412-417 ISSN Print: 2394-7500 ISSN Online: 2394-5869 Impact Factor: 5.2 IJAR 2015; 1(8): 412-417 www.allresearchjournal.com Received: 17-05-2015 Accepted: 20-06-2015 Parantap Nandi A/2, Building
More informationECE2019 Sensors, Signals, and Systems A Lab #6: Electromagnetic Field Sensing
ECE2019 Sensors, Signals, and Systems A 2012 Lab #6: Electromagnetic Field Sensing Introduction This lab involves construction of circuits which demonstrate electromagnetic properties of cables used in
More informationLook over Chapter 31 sections 1-4, 6, 8, 9, 10, 11 Examples 1-8. Look over Chapter 21 sections Examples PHYS 2212 PHYS 1112
PHYS 2212 Look over Chapter 31 sections 1-4, 6, 8, 9, 10, 11 Examples 1-8 PHYS 1112 Look over Chapter 21 sections 11-14 Examples 16-18 Good Things To Know 1) How AC generators work. 2) How to find the
More informationHigh Votage Module AC/DC/Impulse Test System
TSGADI Series High Votage Module AC/DC/Impulse Test System A digital control and measuring system is used to be control the difference output AC/DC/Impulse and related protection device such as over voltage
More informationPRODUCT PORTFOLIO TEST AND MEASURING SYSTEMS
PRODUCT PORTFOLIO TEST AND MEASURING SYSTEMS n Quality made in Germany n Standardized and customized solutions n Turn key projects, all from one hand 0.1-1/1 Product Portfolio Test and Measuring Systems
More informationISSN: [IDSTM-18] Impact Factor: 5.164
IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY A REVIEW OF ROUTINE TESTING ON DISTRIBUTION TRANSFORMER Sukhbir Singh 1, Parul Jangra 2, Anoop Bhagat 3, Vipin Saini 4 1 Assistant
More informationRAP Ripple Control Coupling for parallel injection of RC signals in Medium and High Voltage Networks
RAP Ripple Control Coupling for parallel injection of RC signals in Medium and High Voltage Networks We reserve all rights to this document, and the subject-matter it deals with. Duplication, dissemination
More information3. COMBINED TRANSFORMERS Oil-paper insulation
3. COMBINED TRANSFORMERS Oil-paper insulation 123 kv Combined transformers. 26 Instrument transformers High voltage INTRODUCTION Combined instrument transformers contain a current transformer and an inductive
More informationHigh-Voltage Test Techniques
High-Voltage Test Techniques Dieter Kind Kurt Feser 2nd Revised and Enlarged Edition With 211 Figures and 12 Laboratory Experiments Translated from the German by Y. Narayana Rao Professor of Electrical
More informationDifferential-Mode Emissions
Differential-Mode Emissions In Fig. 13-5, the primary purpose of the capacitor C F, however, is to filter the full-wave rectified ac line voltage. The filter capacitor is therefore a large-value, high-voltage
More informationOutdoor Instrument Transformers Buyer s Guide
Outdoor Instrument Transformers Buyer s Guide Table of content Page Products Introduction 3 Explanations 4 Silicone Rubber (SIR) Insulators 7 Design features and Advantages Currrent transformers IMB 8
More informationOil insulated Outdoor Instrument Transformers Buyer s Guide
Oil insulated Outdoor Instrument Transformers Buyer s Guide Table of content Page Products Introduction 3 Explanations 4 Silicone Rubber (SIR) Insulators 7 Design features and Advantages Current transformers
More informationTransformer condition assessment with an integrated test van
Transformer condition assessment with an integrated test van 1 2012 SebaKMT Measuring and locating techniques MADE in GERMANY 2 Testing and Standards for Power Transformers CIGRE CIGRE Brochure 342 (SFRA-FRAX)
More informationAC Circuit. What is alternating current? What is an AC circuit?
Chapter 21 Alternating Current Circuits and Electromagnetic Waves 1. Alternating Current 2. Resistor in an AC circuit 3. Capacitor in an AC circuit 4. Inductor in an AC circuit 5. RLC series circuit 6.
More information(2) New Standard IEEE P (3) Core : (4) Windings :
(d) Electrical characteristics (such as short-circuit withstand, commutating reactance, more number of windings, etc); (e) Longer life expectancy; (f) Energy efficiency; (g) more demanding environment.
More informationVariation in SFRA plot due to design and external parameter
Chapter 6 Variation in SFRA plot due to design and external parameter 6.1 Introduction As the experience grows with Sweep Frequency Response Analysis in world, it is useful to discuss the measurements
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION In terms of section 22(2) (b) of the Accreditation for Conformity Assessment, Calibration and Good Laboratory Practice Act, 2006 (Act 19 of 2006), read with sections 23(1),
More informationTHE ELECTROM itig II MOTOR TESTER AND WINDING ANALYZER
THE ELECTROM itig II MOTOR TESTER AND WINDING ANALYZER AUTOMATED, SAFE, AND EASY TO USE TESTER. HIGH AND LOW VOLTAGE TESTS IN A LIGHTWEIGHT PACKAGE. The state of the art Electrom itig II provides a wide
More information700 Series AC Dielectric Test Sets
700 Series AC Dielectric Test Sets AC Test Systems HAEFELY HIPOTRONICS standard line of AC Dielectric Test Systems is designed to perform high voltage AC tests on electrical apparatus in accordance with
More information9. How is an electric field is measured?
UNIT IV - MEASUREMENT OF HIGH VOLTAGES AND HIGH CURRENTS PART-A 1. Mention the techniques used in impulse current measurements. Hall generators, Faraday generators and current transformers. 2.Mention the
More informationINSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad
Course Name Course Code Class Branch INSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad -500 043 AERONAUTICAL ENGINEERING TUTORIAL QUESTION BANK : ELECTRICAL AND ELECTRONICS ENGINEERING : A40203
More information1433 SERIES. High-Accuracy Decade Resistor User and Service Manual
1433 SERIES High-Accuracy Decade Resistor User and Service Manual Contents Chapter 1 Introduction...1 1.1 Product Overview... 1 Chapter 2 Specifications...2 Decade Specifications... 2 Chapter 3 Operation...5
More informationElectromagnetic Oscillations and Currents. March 23, 2014 Chapter 30 1
Electromagnetic Oscillations and Currents March 23, 2014 Chapter 30 1 Driven LC Circuit! The voltage V can be thought of as the projection of the vertical axis of the phasor V m representing the time-varying
More informationAC/DC Resistance Standard DRR-112
AC/DC Resistance Standard DRR-112 Operator's Manual DTM Instruments, LLC www.dtminstruments.com Version 2.0 November 2013 Introduction The AC/DC resistance standard DRR-112 is used to calibrate resistance
More informationAccurate Current Measurement Transducer for Relaying Purpose
Accurate Current Measurement Transducer for Relaying Purpose Ashish S. Paramane 1, Dr.P.K.Katti 2 Department of Electrical Engineering Dr. Babasaheb Ambedkar Technological University, Lonere, Maharashtra
More informationExercise 1: Series RLC Circuits
RLC Circuits AC 2 Fundamentals Exercise 1: Series RLC Circuits EXERCISE OBJECTIVE When you have completed this exercise, you will be able to analyze series RLC circuits by using calculations and measurements.
More informationMODEL INFORMATION. MODEL AccuLoss ALMS 1058, 2058 and Capabilities. Measurements
MODEL AccuLoss ALMS 1058, 2058 and 4058 The measurement of electric power and energy at high voltages and currents at low power factors is becoming increasingly important economically as a way to reduce
More informationMSR Series. Modular Series Resonant Systems kV; 500kVA...60,000kVA FEATURES BENEFITS APPLICATIONS 1/6
MSR Series Modular Series Resonant Systems - 250...2200kV; 500kVA...60,000kVA The MSR Series is designed to provide power for tests on cables, HV and EHV transformers, gasinsulated switchgear, bushings,
More informationVARIABLE FREQUENCY RESONANT TEST SYSTEM
VARIABLE FREQUENCY RESONANT TEST SYSTEM 400kV,1200kVA Tank Type Reactor for Outdoor Cable Test Field 650kV, 5850kVA, Cylinder Type Reactors for Onsite Testing of GIS and Cable PHENIX Variable Frequency
More informationTesting Potential Transformers
Testing Potential Transformers By Les Warner (PCA Valence) and Chris Werstiuk ( LLC) Potential transformers (PTs) are necessary to a power system for metering and protective relaying to convert higher
More informationKIT 4.0 HIGH VOLTAGE CONSTRUCTION KIT. reliable. precision. HIGH V O L T A G E T E S T S OLUTION S
KIT 4.0 HIGH VOLTAGE CONSTRUCTION KIT reliable. precision. HIGH V O L T A G E T E S T S OLUTION S 2 General Description GENERAL DESCRIPTION APPLICATION The KIT 4.0 is a high voltage system made of easily
More informationfor the test of CT, VT and PT
SPECIFICATIONS TDX 5000 for the test of CT, VT and PT DOCUMENT No. SIE65175, Rev. 0, January 2016 TABLE OF CONTENTS SPECIFICATIONS... 1 1 GENERAL... 7 2 MEASURE ACCURACY...14 2.1 APPLICABLE TEST SET...
More informationNote: 1. All the students must strictly follow all the safety precautions. 2. In case of any question or concern, please contact LAB INSTRUCTOR or TA.
UNIVERSITY OF WATERLOO ELECTRICAL & COMPUTER ENGINEERING DEPARTMENT FALL 2006 E&CE 261: Energy Systems and Components EXPERIMENT 1: THREE-PHASE SYSTEMS Contents covered in this laboratory exercise: 1.
More informationTECHNICAL BULLETIN 004a Ferroresonance
May 29, 2002 TECHNICAL BULLETIN 004a Ferroresonance Abstract - This paper describes the phenomenon of ferroresonance, the conditions under which it may appear in electric power systems, and some techniques
More informationPower Voltage Transformers for Air Insulated Substations. THE PROVEN POWER.
Power Voltage Transformers for Air Insulated Substations THE PROVEN POWER. Introduction Trench Power Voltage Transformers (Power VTs) combine the attributes of an inductive voltage transformer with the
More informationAligarh College of Engineering & Technology (College Code: 109) Affiliated to UPTU, Approved by AICTE Electrical Engg.
Aligarh College of Engineering & Technology (College Code: 19) Electrical Engg. (EE-11/21) Unit-I DC Network Theory 1. Distinguish the following terms: (a) Active and passive elements (b) Linearity and
More informationAC Resonant Test Systems Tank Type Up to 500 kv
R AC Resonant Test Systems Tank Type Up to 500 kv PHENIX TECHNOLOGIES Testing Applications High power AC Resonant Test Systems are used where the load is largely capacitive with very low real power losses
More informationDescription, operating instructions and hardware specification of the HL Transverse balance measurement module (TBMM).
, operating instructions and hardware specification of the HL 8240 - Transverse balance measurement module (TBMM). The increasing use of telecom equipment in ultrafast applications has enhanced the need
More information5080A Multi-Product Calibrator
5080A Multi-Product Calibrator Calibration solutions for your analog and digital workload 5080A Multi-Product Calibrator: expanded workload coverage Calibrate analog and digital meters, and much more The
More informationPARAMETER CONDITION VALUE Minimum Input Voltage. Maximum Input Voltage. Maximum Output Current for 3.3Voutput. Maximum Output Current for 2.
ECRIPTION QUICK TART UIE FOR EMONTRATION CIRCUIT 541A LTC3723-1, LTC3901, LT3710 and LT1431 emonstration circuit 541A is an isolated synchronous push-pull converter featuring the LTC3723-1, LTC3901, LT1431
More informationInnovative Test Techniques and Diagnostic Measurements to Improve the Performance and Reliability of Power System Transformers
Innovative Test Techniques and Diagnostic Measurements to Improve the Performance and Reliability of Power System Transformers Dr. Michael Krüger, Alexander Kraetge, OMICRON electronics GmbH, Austria Alexander
More informationTransformers handling and transport
Special tests (Credit: http://www.breakbulk.com/wp-content/uploads/2015/02/20141117160247x.jpg) Transformers handling and transport Damages that may arise and how to find them Table of contents summary
More information