1 PhO 2 Smartphone based Blood Oxygen Level Measurement using Near-IR and RED Wave-guided Light Nam Bui, Anh Nguyen, Phuc Nguyen, Hoang Truong, Ashwin Ashok, Thang Dinh, Robin Deterding, Tam Vu 1/30

2 Chronic Obstructive Pulmonary Disease (COPD) Leading cause of death in the US, Million Americans have COPD Early detection of COPD is key to successful treatment OFTEN FOUND IN LATE STAGE $34,000 treatment in ICU 2/30

3 Vital signs screening in COPD Blood pressure Heart rate Oxygen Saturation needs to be measured frequently Oxygen saturation (SpO2) Breathing rate Wheezing sound 3/30

4 Oxygen Saturation (SpO 2 ) Oxygenated hemoglobin Deoxygenated hemoglobin O 2 O 2 O 2 O 2 Sp02 = No. oxygenated hemoglobin Total No. hemoglobin 100% 4/30

5 Non-invasive Measurement of Oxygen Saturation IR emitter Red emitter Intensity Intensity t AC IR AC R AC t Sp02 = f AC R AC IR Photon Detector R IR 5/30

6 Red and IR Wavelength Selection Harmful Low penetration Absorbed by red pigmentation At Red= 660 nm and IR = 960 nm, the difference of light absorption by oxygenated hemoglobin and deoxygenated hemoglobin is largest. 6/30

7 Dedicated Hardware Solutions Smartphone-based blood oxygen measurement Pros Cons High accuracy Inconvenient Costly

8 A solution to capture the oxygen level and is High accuracy. Low-cost. Easy to make by patients 8/30

9 PhO 2 Smartphone based Blood Oxygen Level Measurement 9/30

10 Hardware Challenges PULSE OXIMETER PHONE BASED PULSE OXIMETER IR emitter Red emitter Light Source 10/30

11 Challenge. Phone s camera IR filter IR filter Sensor 11/30

12 Existing Software-based Solutions DISADVANTAGES Inaccurate: Green channel is mostly absorbed by red pigmentation. Coarse-grained: one record per 10 seconds. RGB Color Image Red Light IR Light DigiDoc ihealthcare

13 Solution. Red and NIR for Oxygen Saturation Measurement NIR NIR NIR has similar characteristics of hemoglobin absorption with the IR lights. is available in most of phone s flashlight and is not filtered by the phone s camera. 14/30

14 PhO 2 NIR Extraction Technique Absorbed by red pigmentation Absorbe d Green filter Red and Green filter cuttoff NIR 15/30

15 And 5 More Challenges RGB Intensity Smartphone s camera picks up most of the lights Conversion between RGB and light intensity Pressure Micro movements Finger movement and Contact pressure Flashlight heat can cause skin burn 16/30

16 3D printed Optical Hardware Add-on 17/30

17 3D printed Optical Hardware Add-on RGB Intensity Smartphone s camera picks up most of the lights Conversion between RGB and light intensity Pressure Micro movements Finger movement and contact pressure Flashlight can cause skin burn 18/30

18 Wavelength selection α = arctan(a/b) ε Hb r ε Hb r R Sp02 = ε Hb ir ε HbO2 r + ε HbO2 ir ε Hb ir R b α a X:678 Y: 724 Z:1.002 λ R = 678 nm λ NIR = 724 nm 20/30

19 Normalized Intensity Normalized Intensity Normalized Intensity Pressure Detection and Recommendation PPG signal in a 7 seconds window Sample PPG signal in a 7 seconds window Sample PPG signal in a 7 seconds window Sample 21/30

20 Heat Reduction with Distributed Lighting Source H E p = hf Modify 2.55 mm 0.6 mm Bandpass filte E = To maintain the high light intensity 22/30

21 PhO 2 Algorithms Detecting pressure Collecting raw data 95 % PPG Calculate peak-topeak ratio Identify the peaks and troughs ROI detection and create the PPG signals 23/30

22 App overview PPG display Image frame 24/30

23 25/30

24 In-lab experiments 3 males, 3 females 5 times performs the hyperventilation (a breath holding technique to reduce oxygen). Totally 560 samples. 26/30

25 SpO2 (%) SpO2 (%) SpO2 (%) DigiDoc PhO2 vs. Existing apps ihealthcare Measurement Indices PhO2 Measurement Indices Time (s) 27/30

26 Experimental results 3.5% 28/30

27 In hospital experiment with patient. In clinic experiment 29/30

28 Fine-grained evaluation of the patient Low Oxygen Level Coughing 30/30

29 Heat Reduction Evaluation Hand multimeter Thermalcouple Probe 31/30

30 Temperature (Celsius) Output temperature of different add-on designs Without Banpass Filter With Bandpass Filter Large Tunnel Small Tunnel Multiple Small Tunnels 32/30

31 Conclusion Accurately measuring oxygen level using low-cost optical film filters. Detect the appropriate contact pressure between finger and phone camera. 3D-printed add-on to handle the problem of light scattering and finger movement. 34/30

32 Application for optical light based sensing Urea Lactate Cholesterol Optical light based sensing Water Alcohol Glucose 35/30

33 PhO 2 Smartphone based Blood Oxygen Level Measurement using Near-IR and RED Wave-guided Light Nam Bui, Anh Nguyen, Phuc Nguyen, Hoang Truong, Ashwin Ashok, Thang Dinh, Robin Deterding, Tam Vu 37/30

D5.1 Report on the design of a fibre sensor based on NIRS

D5.1 Report on the design of a fibre sensor based on NIRS Optical Fibre Sensors Embedded into technical Textile for Healthcare Contract no.: FP6-027 869 Quality control Version : 2.0 Security: PU Nature: Prototype + Report (P, R) Workpackage: WP5 Start date of

More information

An Advanced Architecture & Instrumentation for Developing the System of Monitoring a Vital Sign (Oxygen Saturation) of a Patient.

An Advanced Architecture & Instrumentation for Developing the System of Monitoring a Vital Sign (Oxygen Saturation) of a Patient. An Advanced Architecture & Instrumentation for Developing the System of Monitoring a Vital Sign (Oxygen Saturation) of a Patient. 1 Md.Mokarrom Hossain, 2 A.S.M.Mohsin*, 3 Md.Nasimul Islam Maruf, 4 Md.

More information

Principle of Pulse Oximeter. SpO2 = HbO2/ (HbO2+ Hb)*100% (1)

Principle of Pulse Oximeter. SpO2 = HbO2/ (HbO2+ Hb)*100% (1) Design of Pulse Oximeter Simulator Calibration Equipment Pu Zhang, Jing Chen, Yuandi Yang National Institute of Metrology, East of North Third Ring Road, Beijing, China,100013 Abstract -Saturation of peripheral

More information

WRIST BAND PULSE OXIMETER

WRIST BAND PULSE OXIMETER WRIST BAND PULSE OXIMETER Vinay Kadam 1, Shahrukh Shaikh 2 1,2- Department of Biomedical Engineering, D.Y. Patil School of Biotechnology and Bioinformatics, C.B.D Belapur, Navi Mumbai (India) ABSTRACT

More information

T TH-Typing on Your TeetH: Tongue-Teeth

T TH-Typing on Your TeetH: Tongue-Teeth T TH-Typing on Your TeetH: Tongue-Teeth Localization for Human-Computer Interface Phuc Nguyen, Nam Bui, Anh Nguyen, Hoang Truong, Abhijit Suresh, Matthew Whitlock, Duy Pham, Thang Dinh, and Tam Vu Mobile

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( Pulse Oximetry: The story so far

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (  Pulse Oximetry: The story so far Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) Oxygen is carried in the blood by haemoglobin which has two forms: Hb and HbO 2. These

More information

Pulse Oximetry. Principles of oximetry

Pulse Oximetry. Principles of oximetry Pulse Oximetry The principal advantage of optical sensors for medical applications is their intrinsic safety since there is no electrical contact between the patient and the equipment. (An added bonus

More information

INTEGRATION OF LOW COST SpO2 SENSOR IN A WEARABLE MONITOR

INTEGRATION OF LOW COST SpO2 SENSOR IN A WEARABLE MONITOR INTEGRATION OF LOW COST SpO2 SENSOR IN A WEARABLE MONITOR Ajith K. G. 1, Bony George 1, Aravind B. 2 and Martin K. M. 1 1 NIELIT, Calicut, Kerala, India 2 Mobilexion Technologies, India E-Mail: ajithkallidukkil@gmail.com

More information

Design of Wearable Pulse Oximeter Sensor Module for Capturing PPG Signals

Design of Wearable Pulse Oximeter Sensor Module for Capturing PPG Signals Design of Wearable Pulse Oximeter Sensor Module for Capturing PPG Signals Mr. Vishwas Nagekar 1, Mrs Veena S Murthy 2 and Mr Vishweshwara Mundkur 3 1 Department of ECE, BNMIT, Bangalore 2 Assoc. Professor,

More information

E-health Project Examination: Introduction of an Applicable Pulse Oximeter

E-health Project Examination: Introduction of an Applicable Pulse Oximeter E-health Project Examination: Introduction of an Applicable Pulse Oximeter Mona asseri & Seyedeh Fatemeh Khatami Firoozabadi Electrical Department, Central Tehran Branch, Islamic Azad University, Tehran,

More information

Wireless Sensor Networks. EP2980

Wireless Sensor Networks. EP2980 Wireless Sensor Networks EP2980 Jonas.Wahslen@sth.kth.se Sensors What to sense? How to sense/measure? Available sensors Technology Medical ECG Pulsoximeter Applications Smart Grid Industrial Automation

More information

SFH Photoplethysmography Sensor

SFH Photoplethysmography Sensor SFH 7050 - Photoplethysmography Sensor Application Note draft version - subject to change without notice 1 Introduction This application note describes the use of the SFH 7050 (see Fig. 1) as the sensor

More information

Design Considerations for Wrist- Wearable Heart Rate Monitors

Design Considerations for Wrist- Wearable Heart Rate Monitors Design Considerations for Wrist- Wearable Heart Rate Monitors Wrist-wearable fitness bands and smart watches are moving from basic accelerometer-based smart pedometers to include biometric sensing such

More information

Masimo Corporation 40 Parker Irvine, California Tel Fax

Masimo Corporation 40 Parker Irvine, California Tel Fax Instruments and sensors containing Masimo SET technology are identified with the Masimo SET logo. Look for the Masimo SET designation on both the sensors and monitors to ensure accurate pulse oximetry

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Rybynok, V., May, J.M., Budidha, K. and Kyriacou, P. A. (2013). Design and Development of a novel Multi-channel Photoplethysmographic

More information

OMEGAMONITOR BOM-L1 TR W

OMEGAMONITOR BOM-L1 TR W Laser Tissue Blood Oxygenation Monitor OMEGAMONITOR BOM-L1TR W BA4D9013-3 OMEGAMONITOR BOM-L1 TR W USER'SMANUAL CONTENTS Page 1. Summary 2 2. Part names and Function 3 3. Connection to Recorder and Operation

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Zaman, T., Kyriacou, P. A. & Pal, S. (2013). Free flap pulse oximetry utilizing reflectance photoplethysmography. 35th

More information

DESIGN AND PROTOTYPING OF A MINIATURIZED SENSOR

DESIGN AND PROTOTYPING OF A MINIATURIZED SENSOR DESIGN AND PROTOTYPING OF A MINIATURIZED SENSOR FOR NON-INVASIVE MONITORING OF OXYGEN SATURATION IN BLOOD Roberto Marani, Gennaro Gelao and Anna Gina Perri Electrical and Electronic Department, Polytechnic

More information

Sensors. CSE 666 Lecture Slides SUNY at Buffalo

Sensors. CSE 666 Lecture Slides SUNY at Buffalo Sensors CSE 666 Lecture Slides SUNY at Buffalo Overview Optical Fingerprint Imaging Ultrasound Fingerprint Imaging Multispectral Fingerprint Imaging Palm Vein Sensors References Fingerprint Sensors Various

More information

PHYSIOLOGICAL SIGNALS AND VEHICLE PARAMETERS MONITORING SYSTEM FOR EMERGENCY PATIENT TRANSPORTATION

PHYSIOLOGICAL SIGNALS AND VEHICLE PARAMETERS MONITORING SYSTEM FOR EMERGENCY PATIENT TRANSPORTATION PHYSIOLOGICAL SIGNALS AND VEHICLE PARAMETERS MONITORING SYSTEM FOR EMERGENCY PATIENT TRANSPORTATION Dhiraj Sunehra 1, Thirupathi Samudrala 2, K. Satyanarayana 3, M. Malini 4 1 JNTUH College of Engineering,

More information

BENG 186B Principles of Bioinstrumentation. Week 7 Review. Solutions

BENG 186B Principles of Bioinstrumentation. Week 7 Review. Solutions BENG 186B Principles of Bioinstrumentation Week 7 Review Solutions Selections from: 2015 Homework 5 2015 Homework 6 C d = 0.001 1 2 1.5 Normalized Voltage 1 0.5 0-0.5-1 -1.5-2 Time A B C b C b BENG 186B

More information

DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139

DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139 DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139 Spring 2017 V2 6.101 Introductory Analog Electronics Laboratory Laboratory

More information

The OXY100C outputs four signals simultaneously, as shown in this graph: O 2 Saturation (beat-by-beat, CH 1) Pulse Waveform (beat-by-beat, CH 5)

The OXY100C outputs four signals simultaneously, as shown in this graph: O 2 Saturation (beat-by-beat, CH 1) Pulse Waveform (beat-by-beat, CH 5) Chapter 6 Specialty Modules NIBP100C OXY100C Pulse Oximeter Module The OXY100C Pulse Oximeter Module is primarily used to measure the blood oxygen saturation level in a non-invasive fashion. Via LEDs,

More information

Health and Fitness Analog solution. Wenbin Zhu Medical BDM June, 2015

Health and Fitness Analog solution. Wenbin Zhu Medical BDM June, 2015 Health and Fitness Analog solution Wenbin Zhu Medical BDM June, 2015 1 A Broad Market TI in Medical Devices Today TI HealthTech Engineering components for life. TI Solutions for Wearable Optical Bio-Sensing

More information

Testing Properties of E-health System Based on Arduino

Testing Properties of E-health System Based on Arduino Journal of Automation and Control, 2015, Vol. 3, No. 3, 122-126 Available online at http://pubs.sciepub.com/automation/3/3/17 Science and Education Publishing DOI:10.12691/automation-3-3-17 Testing Properties

More information

Pulse Oximetry Dave Hoff Roy Zhang Tad Stalter Mike Carlson

Pulse Oximetry Dave Hoff Roy Zhang Tad Stalter Mike Carlson Pulse Oximetry Dave Hoff Roy Zhang Tad Stalter Mike Carlson INTRODUCTION TO PULSE OXIMETRY The oxygenation and deoxygenation of blood is a process rarely considered, but occurs with every breath. When

More information

REMOTE HEALTH MONITORING SYSTEM USING PIC MICROCONTROLLER

REMOTE HEALTH MONITORING SYSTEM USING PIC MICROCONTROLLER REMOTE HEALTH MONITORING SYSTEM USING PIC MICROCONTROLLER S.Sakuntala #1 and R.Ramya Dharshini *2 # B.E, ECE, Mepco Schlenk Engineering College, Sivakasi,India * B.E, ECE, Mepco Schlenk Engineering College,

More information

A Technology Overview of the Nellcor OxiMax Pulse Oximetry System

A Technology Overview of the Nellcor OxiMax Pulse Oximetry System A Technology Overview of the Nellcor OxiMax Pulse Oximetry System Nellcor Technical Staff Why Nellcor Developed the OxiMax Pulse Oximetry System The introduction of the OxiMax Pulse Oximetry System brought

More information

(51) Int Cl.: A61B 5/00 ( ) G06F 17/00 ( )

(51) Int Cl.: A61B 5/00 ( ) G06F 17/00 ( ) (19) (11) EP 1 424 934 B1 (12) EUROPEAN PATENT SPECIFICATION (4) Date of publication and mention of the grant of the patent: 06.08.08 Bulletin 08/32 (21) Application number: 01981641.2 (22) Date of filing:

More information

Realizing Human-Centricity: Data-Driven Services

Realizing Human-Centricity: Data-Driven Services Realizing Human-Centricity: Data-Driven Services Ajay Chander R&D Lead, Data Driven Life Innovations Fujitsu Laboratories of America January 22, 2014 INTERNAL USE ONLY Copyright 2014 FUJITSU LIMITED Context:

More information

ABSTRACT. Robust acquisition of Photoplethysmograms using a Camera. Mayank Kumar

ABSTRACT. Robust acquisition of Photoplethysmograms using a Camera. Mayank Kumar ABSTRACT Robust acquisition of Photoplethysmograms using a Camera by Mayank Kumar Non-contact monitoring of vital signs, such as pulse rate, using a camera is gaining popularity because of its potential

More information

MASIMO RADICAL 7 Signal Extraction Pulse CO-Oximeter

MASIMO RADICAL 7 Signal Extraction Pulse CO-Oximeter MASIMO RADICAL 7 Signal Extraction Pulse CO-Oximeter Women s Health Manual MCH Only Policy Group: Cardiovascular Approved by: Heather Crosland Director, Women s Health, Covenant Health, GNH/MCH Site Lead

More information

Distant pulse oximetry based on skin region extraction and multi-spectral measurement

Distant pulse oximetry based on skin region extraction and multi-spectral measurement Distant pulse oximetry based on skin region extraction and multi-spectral measurement Christian Herrmann a,b, Jürgen Metzler a, Dieter Willersinn a, and Jürgen Beyerer a,b a Fraunhofer IOSB, Karlsruhe,

More information

(12) United States Patent

(12) United States Patent (12) United States Patent USOO7313426B2 (10) Patent No.: US 7,313.426 B2 Takeda et al. (45) Date of Patent: Dec. 25, 2007 (54) APPARATUS FOR DETERMINING 4,759,369 A * 7/1988 Taylor... 600,323 CONCENTRATIONS

More information

6.111 Final Project Proposal HeartAware

6.111 Final Project Proposal HeartAware 6.111 Final Project Proposal HeartAware Michael Holachek and Nalini Singh Massachusetts Institute of Technology 1 Introduction Pulse oximetry is a popular non-invasive method for monitoring a person s

More information

DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR

DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR Srijan Banerjee 1, Subhajit Roy 2 1 Department of Electrical Engineering, Siliguri Institute of Technology, 2 Department of Electrical Engineering,

More information

rainbow ReSposable Pulse CO-Oximeter Sensor System Training Please see product s Directions for Use for more information 2012 Masimo Corporation

rainbow ReSposable Pulse CO-Oximeter Sensor System Training Please see product s Directions for Use for more information 2012 Masimo Corporation rainbow ReSposable Pulse CO-Oximeter Sensor System Training Learning Objectives At the end of this lesson, the learner will be able to: > Identify key application, inspection and precautionary guidelines

More information

ELR 4202C Project: Finger Pulse Display Module

ELR 4202C Project: Finger Pulse Display Module EEE 4202 Project: Finger Pulse Display Module Page 1 ELR 4202C Project: Finger Pulse Display Module Overview: The project will use an LED light source and a phototransistor light receiver to create an

More information

PRODUCT SHEET. fnir 2000C Imager fnir2000s Imager fnir2000p Imager

PRODUCT SHEET. fnir 2000C Imager fnir2000s Imager fnir2000p Imager fnir FUNCTIONAL NEAR INFRARED OPTICAL BRAIN IMAGING SYSTEMS fnir 2000C fnir2000s fnir2000p fnir functional near infrared optical imaging systems measure oxygen level changes in the prefrontal cortex of

More information

HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS

HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS Mr. Sunil L. Rahane Department of E & TC Amrutvahini College of Engineering Sangmaner, India Prof. Ramesh S. Pawase Department of E & TC Amrutvahini

More information

Noninvasive PoC Anemia Detection Device

Noninvasive PoC Anemia Detection Device Noninvasive PoC Anemia Detection Device Team 11 - Project Proposal ECE 445 Spring 2018 Jeremy Dejournett Mythri Anumula TA: Yamuna Phal 1 Table of Contents Introduction 3 Objective 3 Background 3 High-level

More information

Interference metal/dielectric filters integrated on CMOS image sensors SEMICON Europa, 7-8 October 2014

Interference metal/dielectric filters integrated on CMOS image sensors SEMICON Europa, 7-8 October 2014 Interference metal/dielectric filters integrated on CMOS image sensors SEMICON Europa, 7-8 October 2014 laurent.frey@cea.fr Outline Spectral filtering applications Consumer Multispectral Prior art Organic

More information

Design and Development of a Wireless Pulse Oximeter System

Design and Development of a Wireless Pulse Oximeter System Design and Development of a Wireless Pulse Oximeter System Abhishek Ekhare Department of Instrumentation and Control, College of Engineering, Pune. Maharashtra, India. abhishek.ekhare@gmail.com Uttam Chaskar

More information

APPLICATION OF HEART PHOTOPLETHYSMOGRAPHY

APPLICATION OF HEART PHOTOPLETHYSMOGRAPHY APPLICATION OF HEART PHOTOPLETHYSMOGRAPHY 1 VICKY KUMAR SINGH, 2 SUMIT KUMAR THAKUR, 3 VINOD KUMAR 1,2,3 Department of Electronics Engineering, Bharati Vidyapeeth College of Engineering Pune E-mail: vickysingh229@gmail.com,

More information

Keywords: Electronic Patch, Wireless Reflectance Pulse Oximetry, SpO2, Heart Rate, Body Temperature.

Keywords: Electronic Patch, Wireless Reflectance Pulse Oximetry, SpO2, Heart Rate, Body Temperature. IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY Electronic Patch Wireless Reflectance Pulse Oximetry for Remote Health Monitoring S.Venkatesh Department of ECE, Anna University,Chennai,

More information

Innovative Experimental Low Cost Electronics Operated Instrumentation for Wearable Health Systems with High Resolution Physiological Measurements

Innovative Experimental Low Cost Electronics Operated Instrumentation for Wearable Health Systems with High Resolution Physiological Measurements Innovative Experimental Low Cost Electronics Operated Instrumentation for Wearable Health Systems with High Resolution Physiological Measurements D.F. Cruz, E.M.G. Rodrigues, R. Godina, C.M.P. Cabrita

More information

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging Nonlinear microscopy I: Two-photon fluorescence microscopy Multiphoton Microscopy What is multiphoton imaging? Applications Different imaging modes Advantages/disadvantages Scattering of light in thick

More information

PULSE OXIMETRY MODULE TO IMPLEMENT IN TEAM MONITOR OF VITAL SIGNS

PULSE OXIMETRY MODULE TO IMPLEMENT IN TEAM MONITOR OF VITAL SIGNS PULSE OXIMETRY MODULE TO IMPLEMENT IN TEAM MONITOR OF VITAL SIGNS A. Soto Otalora 1, L. A. Guzman Trujilloy 2 and A. DiazDiaz 3 1 Industrial Control Engineering, Universidad Surcolombiana Neiva, Avenida

More information

Intelligent Pillow for Heart Rate Monitor

Intelligent Pillow for Heart Rate Monitor International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 8, Issue 4 (August 2013), PP.47-52 Intelligent Pillow for Heart Rate Monitor Everlyn

More information

LOW POWER WIRELESS PULSE OXIMETER TERMINAL

LOW POWER WIRELESS PULSE OXIMETER TERMINAL Technical Sciences and Applied Mathematics LOW POWER WIRELESS PULSE OXIMETER TERMINAL Dan LOZNEANU*, Paul BORZA*, Gheorghe PANĂ*, Horaţiu MOGA* * Transilvania University, Brasov, Romania Abstract: The

More information

Optical Receivers Theory and Operation

Optical Receivers Theory and Operation Optical Receivers Theory and Operation Photo Detectors Optical receivers convert optical signal (light) to electrical signal (current/voltage) Hence referred O/E Converter Photodetector is the fundamental

More information

73 ASSignee: Neller tria Bennet Incorporated, Clinical Investigation-Evaluation of oxygen Saturation

73 ASSignee: Neller tria Bennet Incorporated, Clinical Investigation-Evaluation of oxygen Saturation USOO5842982A United States Patent (19) 11 Patent Number: Mannheimer (45) Date of Patent: Dec. 1, 1998 54 INFANT NEONATAL PULSE OXIMETER FOREIGN PATENT DOCUMENTS SENSOR 4429 845 C 10/1995 Germany. 75 Inventor:

More information

lllllllllllllllllllllll llllllllllllllllllllllllllllllllllllllllllllllll

lllllllllllllllllllllll llllllllllllllllllllllllllllllllllllllllllllllll United States Patent [191 lllllllllllllllllllllll llllllllllllllllllllllllllllllllllllllllllllllll US005437275A [111 Amundsen et a1. [45] Patent Number: Date of Patent Aug. 1, 1995 [54] PULSE OXIMETRY

More information

Instructions to User

Instructions to User Instructions to User Dear Users, Thank you very much for purchasing our product. Please read the manual very carefully before using this device. Please follow these instructions to operate. No part of

More information

Wearables for novel healthcare paradigms Nick Van Helleputte

Wearables for novel healthcare paradigms Nick Van Helleputte Wearables for novel healthcare paradigms Nick Van Helleputte R&D manager biomedical circuits & systems - imec Chronic disease management Chronic disease example: United states 117 million americans suffer

More information

Signal Extraction Technology

Signal Extraction Technology Signal Extraction Technology Technical bulletin Introduction Masimo SET pulse oximetry is a new and fundamentally distinct method of acquiring, processing and reporting arterial oxygen saturation and pulse

More information

Near-infrared image formation and processing for the extraction of hand veins

Near-infrared image formation and processing for the extraction of hand veins Journal of Modern Optics Vol. 57, No. 18, 20 October 2010, 1731 1737 Near-infrared image formation and processing for the extraction of hand veins Nabila Bouzida*, Abdel Hakim Bendada and Xavier P. Maldague

More information

Get your daily health check in the car

Get your daily health check in the car Edition September 2017 Smart Health, Image sensors and vision systems, Sensor solutions for IoT, CSR Get your daily health check in the car Imec researches capacitive, optical and radar technology to integrate

More information

Design of Low Power Pulse Oximeter for Early Detection of Hypoxemia

Design of Low Power Pulse Oximeter for Early Detection of Hypoxemia 2016 International Conference on Micro-Electronics and Telecommunication Engineering, Design of Low Power Pulse Oximeter for Early Detection of Hypoxemia Komal Kashish, Matangi Priya, Piyush Yadav Department

More information

Third Century Initiative: Global Challenges

Third Century Initiative: Global Challenges Third Century Initiative: Global Challenges The Third Century Initiative has been established to inspire innovative programs that enhance the student learning experience and develop creative approaches

More information

arxiv: v1 [cs.cv] 25 May 2015

arxiv: v1 [cs.cv] 25 May 2015 OAGM Workshop 2015 (arxiv:1505.01065) 1 VeinPLUS: A Transillumination and Reflection-based Hand Vein Database Alexander Gruschina Department of Computer Sciences, University of Salzburg, Austria arxiv:1505.06769v1

More information

Reprint (R37) DLP Products DMD-Based Hyperspectral Imager Makes Surgery Easier

Reprint (R37) DLP Products DMD-Based Hyperspectral Imager Makes Surgery Easier Reprint (R37) DLP Products DMD-Based Hyperspectral Imager Makes Surgery Easier Reprinted with permission by Dr. Karel J. Zuzak University of Texas/Arlington October 2008 Gooch & Housego 4632 36 th Street,

More information

A Novel Ring Shaped Photodiode for Reflectance Pulse Oximetry in Wireless Applications

A Novel Ring Shaped Photodiode for Reflectance Pulse Oximetry in Wireless Applications Downloaded from orbit.dtu.dk on: Sep 15, 2018 A Novel Ring Shaped Photodiode for Reflectance Pulse Oximetry in Wireless Applications Duun, Sune Bro; Haahr, Rasmus Grønbek; Birkelund, Karen; Raahauge, P.;

More information

paper title : Analyzing the Components of Dark Circle by Nonlinear Estimation of Chromophore Concentrations and Shading

paper title : Analyzing the Components of Dark Circle by Nonlinear Estimation of Chromophore Concentrations and Shading (1)First page classification of paper : Original Paper paper title : Analyzing the Components of Dark Circle by Nonlinear Estimation of Chromophore Concentrations and Shading author names : Rina Akaho,

More information

NSERC Summer Project 1 Helping Improve Digital Camera Sensors With Prof. Glenn Chapman (ENSC)

NSERC Summer Project 1 Helping Improve Digital Camera Sensors With Prof. Glenn Chapman (ENSC) NSERC Summer 2016 Digital Camera Sensors & Micro-optic Fabrication ASB 8831, phone 778-782-319 or 778-782-3814, Fax 778-782-4951, email glennc@cs.sfu.ca http://www.ensc.sfu.ca/people/faculty/chapman/ Interested

More information

FOR 353: Air Photo Interpretation and Photogrammetry. Lecture 2. Electromagnetic Energy/Camera and Film characteristics

FOR 353: Air Photo Interpretation and Photogrammetry. Lecture 2. Electromagnetic Energy/Camera and Film characteristics FOR 353: Air Photo Interpretation and Photogrammetry Lecture 2 Electromagnetic Energy/Camera and Film characteristics Lecture Outline Electromagnetic Radiation Theory Digital vs. Analog (i.e. film ) Systems

More information

Challenges & Chances

Challenges & Chances What is wrong with AAL? Challenges & Chances Dirk Elias, Director Center for Assistive i Information and Communication i Solutions Fraunhofer Portugal Research, FhP AICOS Back to Index Content Overview

More information

Using blood volume pulse vector to extract rppg signal in infrared spectrum

Using blood volume pulse vector to extract rppg signal in infrared spectrum MASTER Using blood volume pulse vector to extract rppg signal in infrared spectrum Lin, X. Award date: 2014 Link to publication Disclaimer This document contains a student thesis (bachelor's or master's),

More information

Fingerprint Spoof Detection using Multispectral Imaging Robert K. Rowe, Ph.D. Chief Technology Officer

Fingerprint Spoof Detection using Multispectral Imaging Robert K. Rowe, Ph.D. Chief Technology Officer Fingerprint Spoof Detection using Multispectral Imaging Robert K. Rowe, Ph.D. Chief Technology Officer RKRowe@Lumidigm.com September 21, 4 Program Goal and Topics Goal Develop multispectral imaging as

More information

Noncontact simultaneous dual wavelength photoplethysmography: A further step toward noncontact pulse oximetry

Noncontact simultaneous dual wavelength photoplethysmography: A further step toward noncontact pulse oximetry REVIEW OF SCIENTIFIC INSTRUMENTS 78, 044304 2007 Noncontact simultaneous dual wavelength photoplethysmography: A further step toward noncontact pulse oximetry Kenneth Humphreys and Tomas Ward Department

More information

common type of cardiac diseases and may indicate an increased risk of stroke or sudden cardiac death. ECG is the most

common type of cardiac diseases and may indicate an increased risk of stroke or sudden cardiac death. ECG is the most ISSN: 0975-766X CODEN: IJPTFI Available Online through Research Article www.ijptonline.com DESIGNING OF ELECTRONIC CARDIAC EVENTS RECORDER *Dr. R. Jagannathan, K.Venkatraman, R. Vasuki and Sundaresan Department

More information

FAQ. What is the Qualcomm Tricorder XPRIZE competition?

FAQ. What is the Qualcomm Tricorder XPRIZE competition? FAQ What is the Qualcomm Tricorder XPRIZE competition? The Qualcomm Tricorder XPRIZE is a 3.5-year global competition that will award $10 million to teams that develop a consumer-friendly device capable

More information

fnirs Sensor Data Sheet

fnirs Sensor Data Sheet FNIRS25102017 SPECIFICATIONS > Infrared emitter* >Peak emission: 860nm >Half intensity beam angle: ±13 deg >Spectral bandwitdth: 30nm >Radiant intensity: 750mW/sr > Red emitter* >Peak emission: 660nm >Half

More information

Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based On CMOS

Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based On CMOS IOS Journal of Electrical and Electronics Engineering (IOS-JEEE) e-issn: 2278-1676,p-ISSN: 232-3331, Volume 12, Issue 1 Ver. IV (Jan. Feb. 217), PP 64-7 www.iosrjournals.org Design of the Pulse Oximetry

More information

Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects

Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects Matti Huotari 1, Antti Vehkaoja 2, Kari Määttä 1, Juha

More information

Comparison of color sensor characteristics for mobile applications. Kevin Jensen European Imaging Academy 2014

Comparison of color sensor characteristics for mobile applications. Kevin Jensen European Imaging Academy 2014 Comparison of color sensor characteristics for mobile applications Kevin Jensen European Imaging Academy 2014 MAZeT GmbH Kevin Jensen 11.12.2014 (Nr.:1) European Imaging Academy - SpectroNet Collaboration

More information

Laser processing of materials. Laser safety

Laser processing of materials. Laser safety Laser processing of materials Laser safety Prof. Dr. Frank Mücklich Dr. Andrés Lasagni Lehrstuhl für Funktionswerkstoffe Sommersemester 2007 Contents: LASER Safety Laser-tissue interaction Type of interaction

More information

THE PULSE OXIMETRY DATA TRANSFER SYSTEM FOR WIRELESS

THE PULSE OXIMETRY DATA TRANSFER SYSTEM FOR WIRELESS THE PULSE OXIMETRY DATA TRANSFER SYSTEM FOR WIRELESS Baskar.s 1, Suresh kumar.a 2 Assistant Professor, Electronics and Communication Engineering, Kongunadu College of Engineering and Technology Tamil Nadu,

More information

Validation of the Happify Breather Biofeedback Exercise to Track Heart Rate Variability Using an Optical Sensor

Validation of the Happify Breather Biofeedback Exercise to Track Heart Rate Variability Using an Optical Sensor Phyllis K. Stein, PhD Associate Professor of Medicine, Director, Heart Rate Variability Laboratory Department of Medicine Cardiovascular Division Validation of the Happify Breather Biofeedback Exercise

More information

Crop Scouting with Drones Identifying Crop Variability with UAVs

Crop Scouting with Drones Identifying Crop Variability with UAVs DroneDeploy Crop Scouting with Drones Identifying Crop Variability with UAVs A Guide to Evaluating Plant Health and Detecting Crop Stress with Drone Data Table of Contents 01 Introduction Crop Scouting

More information

P.O. Pro WIRELESS REFLECTANCE PULSE OXIMETER Design 2. December 1, 2004 Team # 3 James Hart Sofia Iddir Rob Mahar Naomi Thonakkaraparayil

P.O. Pro WIRELESS REFLECTANCE PULSE OXIMETER Design 2. December 1, 2004 Team # 3 James Hart Sofia Iddir Rob Mahar Naomi Thonakkaraparayil P.O. Pro WIRELESS REFLECTANCE PULSE OXIMETER Design 2 December 1, 2004 Team # 3 James Hart Sofia Iddir Rob Mahar Naomi Thonakkaraparayil Table of Contents Introduction 1) Wireless technology 2) Design

More information

DESIGN AND IMPLEMENTATION OF PULSE OXIMETER USING FUZZY LOGIC

DESIGN AND IMPLEMENTATION OF PULSE OXIMETER USING FUZZY LOGIC DESIGN AND IMPLEMENTATION OF PULSE OXIMETER USING FUZZY LOGIC MAJOR PROJECT REPORT SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE AWARD OF DEGREE OF BACHELOR OF TECHNOLOGY In Electronics

More information

Design of the Diffuse Optical Tomography Device

Design of the Diffuse Optical Tomography Device Design of the Diffuse Optical Tomography Device A thesis submitted in partial fulfillment of the requirements for the degree of Bachelor of Science degree in Physics from the College of William and Mary

More information

PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION

PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION Tatyana Dimitrova Neycheva, Dobromir Petkov Dobrev Centre of Biomedical Engineering Ivan Daskalov Bulgarian Academy of Sciences, Bl. 105

More information

Instructions to User

Instructions to User Instructions to User Dear Users, thank you very much for purchasing our product. This Manual is written and compiled in accordance with the council directive MDD93/42/EEC for medical devices and harmonized

More information

Simultaneous Second Harmonic Generation of Multiple Wavelength Laser Outputs for Medical Sensing

Simultaneous Second Harmonic Generation of Multiple Wavelength Laser Outputs for Medical Sensing Sensors 2011, 11, 6125-6130; doi:10.3390/s110606125 OPEN ACCESS sensors ISSN 1424-8220 www.mdpi.com/journal/sensors Article Simultaneous Second Harmonic Generation of Multiple Wavelength Laser Outputs

More information

Quad Rat Vitals Monitor. Client: Dr. Alex Converse. Advisor: Dr. Tom Yen. Group Members: Jack Ho Team Leader. Nathan Werbeckes Communicator

Quad Rat Vitals Monitor. Client: Dr. Alex Converse. Advisor: Dr. Tom Yen. Group Members: Jack Ho Team Leader. Nathan Werbeckes Communicator Quad Rat Vitals Monitor Client: Dr. Alex Converse Advisor: Dr. Tom Yen Group Members: Jack Ho Team Leader Nathan Werbeckes Communicator Joseph Yuen BSAC, BWIG Mid Semester Paper for BME 402, Spring 2009

More information

An Engineer s Perspective on of the Retina. Steve Collins Department of Engineering Science University of Oxford

An Engineer s Perspective on of the Retina. Steve Collins Department of Engineering Science University of Oxford An Engineer s Perspective on of the Retina Steve Collins Department of Engineering Science University of Oxford Aims of the Talk To highlight that research can be: multi-disciplinary stimulated by user

More information

Real -time multi-spectral image processing for mapping pigmentation in human skin

Real -time multi-spectral image processing for mapping pigmentation in human skin Real -time multi-spectral image processing for mapping pigmentation in human skin Daisuke Nakao, Norimichi Tsumura, Yoichi Miyake Department of Information and Image Sciences, Chiba University, Japan Abstract

More information

Oxygen Saturation measurements from Green and Orange Illuminations of Multi-Wavelength Optoelectronic Patch Sensor

Oxygen Saturation measurements from Green and Orange Illuminations of Multi-Wavelength Optoelectronic Patch Sensor Article Oxygen Saturation measurements from Green and Orange Illuminations of Multi-Wavelength Optoelectronic Patch Sensor Samah Alharbi 1, Sijung Hu 1 *, David Mulvaney 1, Laura Barrett 2 Liangwen Yan

More information

You won t be able to measure the incident power precisely. The readout of the power would be lower than the real incident power.

You won t be able to measure the incident power precisely. The readout of the power would be lower than the real incident power. 1. a) Given the transfer function of a detector (below), label and describe these terms: i. dynamic range ii. linear dynamic range iii. sensitivity iv. responsivity b) Imagine you are using an optical

More information

COLOUR INSPECTION, INFRARED AND UV

COLOUR INSPECTION, INFRARED AND UV COLOUR INSPECTION, INFRARED AND UV TIPS, SPECIAL FEATURES, REQUIREMENTS LARS FERMUM, CHIEF INSTRUCTOR, STEMMER IMAGING THE PROPERTIES OF LIGHT Light is characterized by specifying the wavelength, amplitude

More information

Ti/Au TESs as photon number resolving detectors

Ti/Au TESs as photon number resolving detectors Ti/Au TESs as photon number resolving detectors LAPO LOLLI, E. MONTICONE, C. PORTESI, M. RAJTERI, E. TARALLI SIF XCVI National Congress, Bologna 20 24 September 2010 1 Introduction: What are TES? TESs

More information

Next Generation Biometric Sensing in Wearable Devices

Next Generation Biometric Sensing in Wearable Devices Next Generation Biometric Sensing in Wearable Devices C O L I N T O M P K I N S D I R E C T O R O F A P P L I C AT I O N S E N G I N E E R I N G S I L I C O N L A B S C O L I N.T O M P K I N S @ S I L

More information

Noninvasive PoC Anemia Detection Device

Noninvasive PoC Anemia Detection Device Noninvasive PoC Anemia Detection Device Team 11 - Design Document ECE 445 Spring 2018 Jeremy DeJournett Mythri Anumula TA: Yamuna Phal 1 Table of Contents Introduction 3 Objective 3 Background 3 High-level

More information

An Intelligent Wearable e-belt for Continuous Monitoring of Sinus Rhythm

An Intelligent Wearable e-belt for Continuous Monitoring of Sinus Rhythm 1 An Intelligent Wearable e-belt for Continuous Monitoring of Sinus Rhythm Arun.P 1, Amala M.C 2, Anjaly M 3, Githin T.S 4, Jomin J 5 1 Assistant Professor, 2,3,4,5 UG Scholar, Department of Electronics

More information

VivoSense. User Manual - Equivital Import Module. Vivonoetics, Inc. San Diego, CA, USA Tel. (858) , Fax. (248)

VivoSense. User Manual - Equivital Import Module. Vivonoetics, Inc. San Diego, CA, USA Tel. (858) , Fax. (248) VivoSense User Manual - VivoSense Version 3.0 Vivonoetics, Inc. San Diego, CA, USA Tel. (858) 876-8486, Fax. (248) 692-0980 Email: info@vivonoetics.com; Web: www.vivonoetics.com Cautions and disclaimer

More information

Superconducting Transition-Edge Sensors and Superconducting Tunnel Junctions for Optical/UV Time-Energy Resolved Single-Photon Counters

Superconducting Transition-Edge Sensors and Superconducting Tunnel Junctions for Optical/UV Time-Energy Resolved Single-Photon Counters Superconducting Transition-Edge Sensors and Superconducting Tunnel Junctions for Optical/UV Time-Energy Resolved Single-Photon Counters NHST Meeting STScI - Baltimore 10 April 2003 TES & STJ Detector Summary

More information

Self-contained, passive, non-contact, photoplethysmography: real-time extraction of heart rates from live view within a Canon PowerShot

Self-contained, passive, non-contact, photoplethysmography: real-time extraction of heart rates from live view within a Canon PowerShot Self-contained, passive, non-contact, photoplethysmography: real-time extraction of heart rates from live view within a Canon PowerShot Henry Dietz, Chadwick Parrish, and Kevin D. Donohue; Department of

More information

HIGH RESOLUTION HADAMARD TRANSFORM OXIMETRY. D. Barschdorff, E. Starke

HIGH RESOLUTION HADAMARD TRANSFORM OXIMETRY. D. Barschdorff, E. Starke HGH REOLUTON HADAMARD TRANFORM OXMETRY D. Barschdorff, E. tarke nstitute of Electrical Measurement University of Paderborn, D-33098 Paderborn, Germany Abstract: The principle of pulse oximetry is based

More information