Development of a Vibration Measurement Method for Cryocoolers

Size: px
Start display at page:

Download "Development of a Vibration Measurement Method for Cryocoolers"

Transcription

1 REVTEX 3.1 Released September 2 Development of a Vibration Measurement Method for Cryocoolers Takayuki Tomaru, Toshikazu Suzuki, Tomiyoshi Haruyama, Takakazu Shintomi, Akira Yamamoto High Energy Accelerator Research Organization (KEK), 1-1 Oho, Tsukuba, Ibaraki, 35-81, Japan Takashi Uchiyama, Shinji Miyoki Institute for Cosmic Ray Research, University of Tokyo, Kashiwanoha, Kashiwa, Chiba, , Japan Tsutomu Shimonosono, Yasumi Ohtani, Toru Kuriyama Toshiba Corporation, 2-4, Suehiro, Tsurumi, Yokohama, 23-45, Japan Tomohiro Koyama, Rui Li Sumitomo Heavy Industries Ltd., Yato, Nishitokyo, Tokyo , Japan A vibration measurement method for cryocoolers has been developed. The method can measure the vibration of the cold stage at the thermally steady state after cool-down as well as the measurement at room temperature. It can also separately monitor not only the motion of the cold stage, but also the acceleration of the whole cold head caused by inertial forces of the cryocooler. From the result of vibration measurements for a pulse tube cryocooler, we found that the vibration of the cryocooler has different profiles, at the steady state operation after the cool-down from that at the beginning of the cool-down at room temperature. This result showed an advantage of the method. I. INTRODUCTION A well-developed Gifford-McMahon (GM) cryocooler is widely used in various fields of science and industries because of its convenient handling. However, since a GM cryocooler has a large vibration due to the displacer motion, there is a demand for vibration-free cryocoolers in many fields, such as a cryogenic interferometric gravitational wave detector [1], a superconducting mixer for submillimeter waves, a magnetic resonant imaging device and an electron microscope. A pulse tube (PT) cryocooler is expected to be quieter than a conventional GM cryocooler, because a PT cryocooler has no mechanical displacer in the cold head. However, there are a few reports of vibration evaluations for cryocoolers [2,3]. Most of these evaluations were performed in an early stage of the cool-down process of cryocoolers, far from a steady state. Therefore, we developed a vibration measurement method in which it is possible to measure the vibrations of the cold stage and the whole cold head separately at the thermally steady state condition of the cryocooler. This method can evaluate the vibration coming from different vibration sources in a more realistic situation during its steady state operation. We report on the vibration measurement method for the cryocoolers in this paper. Corresponding author. tomaru@post.kek.jp 7-1 c 23 American Institute of Physics

2 REVTEX 3.1 Released September 2 II. MEASUREMENT METHOD Figs. 1 and 2 show the developed vibration measurement method for cryocoolers. A cold head of the cryocooler was set on a top flange of a vacuum chamber. The size of the vacuum chamber was about φ45 mm in diameter and 67 mm in length, its weight was 1 kg, and there was no radiation shield in the chamber for simplicity. It was set on small rubber sheets to isolate the vibration coming through the ground. The rotary valve unit and the cold head were connected by a rigid copper-tube of 4 m, and the tube was mechanically anchored onto a block of 24 kg at about 1 m apart from the cold head to reduce the transmitted vibration from the valve unit and the compressor directly. The compressor was set in the next room, and a steel door was partitioned between the measurement apparatus and the compressor to reduce the effect of sounds. The vibration of the cold stage and the whole cold head should be measured separately, because their vibration source is different. As shown in Fig. 1 (a), the vibration of the cold stage was measured by an optical displacementsensor and the vibration of the whole cold head (vacuum chamber) was measured by an accelerometer. To measure the vibration of the cold stage in a high-frequency range, the resonant frequency of the table on which the optical sensor was set had to be sufficiently high. This table consisted of four stainless-steel rods in 38 mm diameter and in 5 mm length, and an aluminium plate in 3 mm diameter and in 2 mm thickness. The resonant frequency of the table was 21 Hz for the vibration measurement as given in section III. Fig. 1 (b) shows a picture of the PT cold head and the measurement apparatus. Fig. 3 (a) shows a schematic diagram of the optical sensor. The light emitted from a LED is reflected by a reflector attached under the cold stage, and detected by photo-detectors. The vibration of the cold stage is detected as the change of the light power corresponding to the change of the distance between the sensor and the reflector. The LED is a high-power GaAlAs diode with a peak wavelength of light of 89 nm, and the photo-detector is two InGaAs PIN photo-diodes. This optical sensor can work even at liquid-helium temperature. 1 The reason that two photo-detectors were used was to reduce the effect of the tilt between the sensor and the reflector. The cut-off frequency of the pre-amplifier of the sensor was 1.6 khz. The reflector was made of a polished aluminum block (75 g). The noises of the sensor mainly came from the line noise, the intensity noise of the LED and the noise of the pre-amplifier for the 1 Finally, the sensor was hardly cooled in this measurement, since no radiation shield was used. 7-2

3 REVTEX 3.1 Released September 2 photo-detectors. Since the thermal contraction and tilt due to the thermal stress of the cold stage occurred at cryogenic temperature, and also the reflectivity of the reflector changes by absorption of residual gas, the sensor had to be calibrated under this situation. Therefore, we set the optical sensor on a motor-driven X-stage with an encoder, and calibrated the sensor just before each measurement. Fig. 3 (b) show the relationship between the displacement and the output signal of the optical sensor. The sensing position of the sensor was chosen to be at about 5 mm from the reflector, since this position has an almost linear response. The conversion coefficients from the sensor-output to the displacement were calculated from the gradient of the graph. We adopted the accelerometer to measure the vibration of the vacuum chamber, since it needs no reference of position for the measurement. A commercial laser accelerometer 2 was set on the top flange of the vacuum chamber. The resonant frequency of the accelerometer was about 8 Hz. Since the accelerometer was set on outside of the vacuum chamber, the effect of sound transmitted through the air could not be eliminated. Although a principle sensitivity limitation for the acceleration measurement was ground-motion, the pre-amplifier noise of the accelerometer limited the sensitivity in the measurement as given in section III, because a wide dynamic range of the measurement was used. III. EXPERIMENT We measured the vibrations of a 4 K PT cryocooler produced by Sumitomo Heavy Industries Ltd. [5] using this measurement method. In this measurement, we compared the vibration amplitude at the thermally steady state at the beginning of the cool-down. The typical cooling capacity of the cryocooler was about 2 W at 4 K for the first stage and.5 W at 4.2 K for the second stage. The weight of the cold head was 12.4 kg and its length was 625 mm. An air-cooling type compressor of 7 kw was used in this experiment. The filling pressure was 1.7 MPa, and the operating pressure was measured at the outlet of the rotary valve unit. The operating frequency was 1 Hz. The vacuum pumps (a rotary and a turbo pump) were stopped during the measurements. Typical pressure in the chamber was 1 2 Pa. The measurement was performed for a vertical and two horizontal directions. 2 The accelerometer was a LA5 produced by RION Co. Ltd. The system of typical laser accelerometer is described in reference [4]. 7-3

4 REVTEX 3.1 Released September 2 Fig. 4 is the experimental result for the vertical vibration of the cryocooler. The output of the optical sensor was low-pass filtered at 1 Hz to eliminate the large line noise of 5 Hz. The optical sensor was calibrated just before each measurement. The temperature of the cold stage at the thermally steady state was 32 K, which is higher than the nominal temperature, since no radiation shield was used. We found that the vibrations of the cold stage, the chamber and the pressure at the thermally steady state were smaller than those at the beginning of the cool-down. Also, we found that the vibration of the cold stage at the beginning of the cool-down was very complicated, as shown typically in Fig. 5. Since the pressure oscillation had a different profile from the cold stage vibration, the complicated motion of the cold stage can come from the thermal stress of the stage. On the contrary, the vibrations in the thermally steady state were stable. The vibration amplitude of the cold stage was 12 µm in peak to peak, and the acceleration of the chamber was below.2m/s 2. Although there was a vertical-horizontal coupling, the horizontal vibration of the cold stage was smaller than that in the vertical direction. The vibrations of the vacuum chamber were almost same for all directions. IV. CONCLUSION We developed a vibration measurement method for the cryocoolers. This method can measure the vibrations of the cold stage and the whole cold head separately at the thermally steady state of the cryocooler with calibration of the optical sensor. By using this method, we found that the vibration of the cryocooler has different profiles, at the steady state operation after the cool-down from that at the beginning of the cool-down at room temperature. ACKNOWLEDGMENTS This study was supported by the Joint Research and Development Program of KEK and by a Grant-in-Aid prepared by Ministry of Education, Science, Sports and Culture. We express our appreciation to Dr. R. Takahashi at National Astronomical Observatory of Japan for useful advice during our measurement. [1] K. Kuroda et al., Int. J. Mod. Phys. D 8 (1999) 557. [2] C. Wang et al., Adv. Cryo. Eng. 47 (22) 641. [3] R. Lienerth et al., Proceedings of ICEC 18, (2) 555. [4] A. Araya et al., Rev. Sci. Instrum. 64 (5), (1993) [5] M.Y.Xu et al., Cryocoolers 12, Proceedings of the 12th International Cryocooler Conference, (23)

5 REVTEX 3.1 Released September 2 FIG. 1. Vibration measurement apparatus for the cryocoolers. An optical displacement sensor monitors the motion of the cold stage and an accelerometer measures the acceleration of the whole cold head (vacuum chamber), separately. The optical sensor is calibrated by moving with a motor-driven X-stage. FIG. 2. Whole setup of the vibration measurement apparatus. The rotary valve unit and the cold head were connected by a rigid copper-tube. The tube was mechanically anchored onto a block of 24 kg. The compressor was set in the next room, and a steel door was partitioned between the measurement apparatus and the compressor to reduce sound effect. 7-5

6 REVTEX 3.1 Released September 2 FIG. 3. Optical displacement-sensor (reflective type). (a): schematic diagram of the sensor, (b): output signal of the sensor versus displacement. (a) 2 Displacement [m] x1-6 (b) Acceleration [m/s 2 ] (c) Pressure [MPa] FIG. 4. Experimental results of the vibration of the 4 K PT cryocooler for the vertical direction. (a): vibration of the cold stage (displacement), (b): vibration of the chamber (acceleration) and (c): vibration of the pressure. The bright gray lines show noise of the sensor when the cryocooler stopped, the dark gray lines show the vibration at the beginning of the cool-down and the black lines show the vibration at the thermally steady state (32 K). The graph of (c) has no bright gray line of sensor noise. The vibration data of the cold stage at the beginning of the cool-down is a typical value, since its amplitude was not constant as shown in Fig

7 REVTEX 3.1 Released September 2 Displacement [m] (a) x1-6 (b) Acceleration [m/s 2 ] (c) Pressure [MPa] FIG. 5. Long-term data of the vibration of the 4 K PT cryocooler for the vertical direction at the beginning of the cool-down. (a): vibration of the cold stage (displacement), (b): vibration of the chamber (acceleration) and (c): vibration of the pressure

Vibration-Free Pulse Tube Cryocooler System for Gravitational Wave Detectors II - Cooling Performance and Vibration -

Vibration-Free Pulse Tube Cryocooler System for Gravitational Wave Detectors II - Cooling Performance and Vibration - 1 Vibration-Free Pulse Tube Cryocooler System for Gravitational Wave Detectors II - Cooling Performance and Vibration - R. Li A, Y. Ikushima A, T. Koyama A, T. Tomaru B, T. Suzuki B, T. Haruyama B, T.

More information

Vibration measurement in the cryogenic interferometric gravitational wave detector (CLIO interferometer)

Vibration measurement in the cryogenic interferometric gravitational wave detector (CLIO interferometer) Vibration measurement in the cryogenic interferometric gravitational wave detector (CLIO interferometer) ICRR Univ. of Tokyo, Dept. of geophysics Kyoto University A, KEK B, Dept. of advanced materials

More information

Development of the accelerometer for cryogenic experiments II

Development of the accelerometer for cryogenic experiments II Development of the accelerometer for cryogenic experiments II ICRR Univ. of Tokyo, KEK A, Dept. of advanced materials science Univ. of Tokyo B K. Yamamoto, H. Hayakawa, T. Uchiyama, S. Miyoki, H. Ishitsuka,

More information

Cold-Head Vibrations of a Coaxial Pulse Tube Refrigerator

Cold-Head Vibrations of a Coaxial Pulse Tube Refrigerator Cold-Head Vibrations of a Coaxial Pulse Tube Refrigerator T. Koettig 1, F. Richter 2, C. Schwartz 2, R. Nawrodt 2, M. Thürk 2 and P. Seidel 2 1 CERN, AT-CRG-CL, CH-1211 Geneva 23, Switzerland 2 Friedrich-Schiller-Universität

More information

Low Vibration, Low Thermal Fluctuation System for Pulse Tube and Gifford- McMahon Cryocoolers

Low Vibration, Low Thermal Fluctuation System for Pulse Tube and Gifford- McMahon Cryocoolers Low Vibration, Low Thermal Fluctuation System for Pulse Tube and Gifford- McMahon Cryocoolers L. Mauritsen, D. Snow, A. Woidtke, M. Chase, and I. Henslee S2 Corporation Bozeman, MT ABSTRACT A compact,

More information

HIGH POWER INPUT COUPLERS FOR THE STF BASELINE CAVITY SYSTEM AT KEK

HIGH POWER INPUT COUPLERS FOR THE STF BASELINE CAVITY SYSTEM AT KEK HIGH POWER INPUT COUPLERS FOR THE STF BASELINE CAVITY SYSTEM AT KEK E. Kako #, H. Hayano, S. Noguchi, T. Shishido, K. Watanabe and Y. Yamamoto KEK, Tsukuba, Ibaraki, 305-0801, Japan Abstract An input coupler,

More information

A SUBMILLIMETER SIS RECEIVER COOLED BY A COMPACT STIRLING-YT REFRIGERATOR

A SUBMILLIMETER SIS RECEIVER COOLED BY A COMPACT STIRLING-YT REFRIGERATOR Eighth International Symposium on Space Terahertz Technology. Harvard Universit y. March 1997 A SUBMILLIMETER SIS RECEIVER COOLED BY A COMPACT STIRLING-YT REFRIGERATOR J.Inatani, T.Noguchi, S.C.Shi, and

More information

MULTIPACTING IN THE CRAB CAVITY

MULTIPACTING IN THE CRAB CAVITY MULTIPACTING IN TH CRAB CAVITY Y. Morita, K. Hara, K. Hosoyama, A. Kabe, Y. Kojima, H. Nakai, KK, 1-1, Oho, Tsukuba, Ibaraki 3-81, JAPAN Md. M. Rahman, K. Nakanishi, Graduate University for Advanced Studies,

More information

Energy Efficient Operation of 4 K Pulse Tube Cryocoolers

Energy Efficient Operation of 4 K Pulse Tube Cryocoolers 187 Energy Efficient Operation of 4 K Pulse Tube Cryocoolers C. Wang, A. Beyer, J. Cosco, B. Lichtenwalter and E. Brown Cryomech, Inc., Syracuse, NY 13211, USA ABSTRACT An inverter compressor has been

More information

Micro-manipulated Cryogenic & Vacuum Probe Systems

Micro-manipulated Cryogenic & Vacuum Probe Systems Janis micro-manipulated probe stations are designed for non-destructive electrical testing using DC, RF, and fiber-optic probes. They are useful in a variety of fields including semiconductors, MEMS, superconductivity,

More information

Development of Shock Acceleration Calibration Machine in NMIJ

Development of Shock Acceleration Calibration Machine in NMIJ IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. Development of Shock Acceleration Calibration Machine

More information

DEVELOPMENT OF A BETA 0.12, 88 MHZ, QUARTER WAVE RESONATOR AND ITS CRYOMODULE FOR THE SPIRAL2 PROJECT

DEVELOPMENT OF A BETA 0.12, 88 MHZ, QUARTER WAVE RESONATOR AND ITS CRYOMODULE FOR THE SPIRAL2 PROJECT DEVELOPMENT OF A BETA 0.12, 88 MHZ, QUARTER WAVE RESONATOR AND ITS CRYOMODULE FOR THE SPIRAL2 PROJECT G. Olry, J-L. Biarrotte, S. Blivet, S. Bousson, C. Commeaux, C. Joly, T. Junquera, J. Lesrel, E. Roy,

More information

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G APPLICATION NOTE M06 attosnom I: Topography and Force Images Scanning near-field optical microscopy is the outstanding technique to simultaneously measure the topography and the optical contrast of a sample.

More information

7 Telsa SQUID Magnetometer

7 Telsa SQUID Magnetometer 7 Telsa SQUID Magnetometer Cryogen Free / Liquid Helium Cooled www.cryogenic.co.uk Introduction S700X - For better magnetic measurements Cryogen free or Liquid Helium based system High homogeneity 7 Tesla

More information

THE CRYOGENIC SYSTEM OF TESLA

THE CRYOGENIC SYSTEM OF TESLA THE CRYOGENIC SYSTEM OF TESLA S. Wolff, DESY, Notkestr. 85, 22607 Hamburg, Germany for the TESLA collaboration Abstract TESLA, a 33 km long 500 GeV centre-of-mass energy superconducting linear collider

More information

KEK ERL CRYOMODULE DEVELOPMENT

KEK ERL CRYOMODULE DEVELOPMENT KEK ERL CRYOMODULE DEVELOPMENT H. Sakai*, T. Furuya, E. Kako, S. Noguchi, M. Sato, S. Sakanaka, T. Shishido, T. Takahashi, K. Umemori, K. Watanabe and Y. Yamamoto KEK, 1-1, Oho, Tsukuba, Ibaraki, 305-0801,

More information

High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE*

High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE* High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE* Y. Owadano, E. Takahashi, I. Okuda, I. Matsushima, Y. Matsumoto, S. Kato, E. Miura and H.Yashiro 1), K. Kuwahara 2)

More information

Tests of the Spoke Cavity RF Source and Cryomodules in Uppsala

Tests of the Spoke Cavity RF Source and Cryomodules in Uppsala FREIA Report 2012/03 October 2012 DEPARTMENT OF PHYSICS AND ASTRONOMY UPPSALA UNIVERSITY Tests of the Spoke Cavity RF Source and Cryomodules in Uppsala ESS TDR Contribution R. Ruber, T. Ekelöf, R.A. Yogi.

More information

ngvla Advanced Cryocoolers For ngvla NATIONAL RADIO ASTRONOMY OBSERVATORY Larry D Addario, Caltech ngvlaworkshop, Socorro, 2017 June 26

ngvla Advanced Cryocoolers For ngvla NATIONAL RADIO ASTRONOMY OBSERVATORY Larry D Addario, Caltech ngvlaworkshop, Socorro, 2017 June 26 NATIONAL RADIO ASTRONOMY OBSERVATORY Advanced Cryocoolers For ngvla Larry D Addario, Caltech ngvlaworkshop, Socorro, 2017 June 26 ngvla Outline How cold do we need to get? Tutorial on cryocoolers (just

More information

S.No Description/Specifications Qty 01. Post office box Trainer.

S.No Description/Specifications Qty 01. Post office box Trainer. Specification of Equipments for Physics lab S.No Description/Specifications Qty 01. Post office box Trainer. 06 The trainer should have: On Board DC Power Supply : 5V Galvanometer ; Deflection : 30 0 30

More information

PRELIMINARY SPECIFICATIONS MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED CRYO-COOLED MAGNET SYSTEM

PRELIMINARY SPECIFICATIONS MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED CRYO-COOLED MAGNET SYSTEM PRELIMINARY SPECIFICATIONS MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED CRYO-COOLED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton,

More information

Observation of Remanent Vortices Attached to Rough Boundaries in Superfluid 4 He

Observation of Remanent Vortices Attached to Rough Boundaries in Superfluid 4 He Journal of Low Temperature Physics - QFS9 manuscript No. (will be inserted by the editor) Observation of Remanent Vortices Attached to Rough Boundaries in Superfluid 4 He Y. Nago T. Ogawa A. Mori Y. Miura

More information

queensgate a brand of Elektron Technology

queensgate a brand of Elektron Technology NanoSensors NX/NZ NanoSensor The NanoSensor is a non-contact position measuring system based on the principle of capacitance micrometry. Two sensor plates, a Target and a Probe, form a parallel plate capacitor.

More information

PI piezo Life Time Test Report. A. Bosotti, R. Paparella, F. Puricelli

PI piezo Life Time Test Report. A. Bosotti, R. Paparella, F. Puricelli PI piezo Life Time Test Report A. Bosotti, R. Paparella, F. Puricelli 1. Introduction...3 1.1. Vacuum...4 1.2. Temperature...4 1.3. Preload...4 1.4. Driving signal...4 2. General features and conceptual

More information

Enabling cutting of Reflective Materials with Back Reflection protection

Enabling cutting of Reflective Materials with Back Reflection protection Enabling cutting of Reflective Materials with Back Reflection protection March 2018 Marking Cutting Welding Micro Machining Additive Manufacturing Slide: 1 02/10/2017 CM-F00003 Rev 6 Laser Cutting of Bright

More information

Uncertainty in measurements of micro-patterned thin film thickness using Nanometrological AFM - Reliability of parameters for base straight line -

Uncertainty in measurements of micro-patterned thin film thickness using Nanometrological AFM - Reliability of parameters for base straight line - Uncertainty in measurements of micro-patterned thin film thickness using Nanometrological AFM - Reliability of parameters for base straight line - Ichiko Misumi,, Satoshi Gonda, Tomizo Kurosawa, Yasushi

More information

QUARTER WAVE COAXIAL LINE CAVITY FOR NEW DELHI LINAC BOOSTER*

QUARTER WAVE COAXIAL LINE CAVITY FOR NEW DELHI LINAC BOOSTER* QUARTER WAVE COAXIAL LINE CAVITY FOR NEW DELHI LINAC BOOSTER* P.N. Prakash and A.Roy Nuclear Science Centre, P.O.Box 10502, New Delhi 110 067, INDIA and K.W.Shepard Physics Division, Argonne National Laboratory,

More information

RF STATUS OF SUPERCONDUCTING MODULE DEVELOPMENT SUITABLE FOR CW OPERATION: ELBE CRYOSTATS

RF STATUS OF SUPERCONDUCTING MODULE DEVELOPMENT SUITABLE FOR CW OPERATION: ELBE CRYOSTATS RF STATUS OF SUPERCONDUCTING MODULE DEVELOPMENT SUITABLE FOR CW OPERATION: ELBE CRYOSTATS J. Teichert, A. Büchner, H. Büttig, F. Gabriel, P. Michel, K. Möller, U. Lehnert, Ch. Schneider, J. Stephan, A.

More information

4. Superconducting sector magnets for the SRC 4.1 Introduction

4. Superconducting sector magnets for the SRC 4.1 Introduction 4. Superconducting sector magnets for the SRC 4.1 Introduction The key components for the realization for the SRC are: the superconducting sector magnet and the superconducting bending magnet (SBM) for

More information

ALMA Interferometer and Band 7 Cartridge

ALMA Interferometer and Band 7 Cartridge ALMA Interferometer and Band 7 Cartridge B7 Cartridge designed, assembled and tested by: S. Mahieu, D. Maier (mixer team lead), B. Lazareff (now at IPAG) G. Celestin, J. Chalain, D. Geoffroy, F. Laslaz,

More information

ACCURACY IMPROVEMENT ON NON-INVASIVE ULTRASONIC-DOPPLER FLOW MEASUREMENT BY UTILZING SHEAR WAVES IN METAL PIPE

ACCURACY IMPROVEMENT ON NON-INVASIVE ULTRASONIC-DOPPLER FLOW MEASUREMENT BY UTILZING SHEAR WAVES IN METAL PIPE 4th International Symposium on Ultrasonic Doppler Method for Fluid Mechanics and Fluid Engineering Sapporo, 6.-8. September, 24 ACCURACY IMPROVEMENT ON NON-INVASIVE ULTRASONIC-DOPPLER FLOW MEASUREMENT

More information

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 S.V. Roth, R. Döhrmann, M. Dommach, I. Kröger, T. Schubert, R. Gehrke Definition of the upgrade The wiggler beamline BW4 is dedicated to

More information

Seal Mechanism of Tip Seal in Scroll Compressor

Seal Mechanism of Tip Seal in Scroll Compressor Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 214 Seal Mechanism of Tip Seal in Scroll Compressor Mitsuhiro Fukuta Shizuoka University,

More information

Vibration studies of a superconducting accelerating

Vibration studies of a superconducting accelerating Vibration studies of a superconducting accelerating module at room temperature and at 4.5 K Ramila Amirikas, Alessandro Bertolini, Wilhelm Bialowons Vibration studies on a Type III cryomodule at room temperature

More information

SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM

SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU,

More information

Mechanical study of the «Saclay piezo tuner» PTS (Piezo Tuning System) P. Bosland, Bo Wu DAPNIA - CEA Saclay. Abstract

Mechanical study of the «Saclay piezo tuner» PTS (Piezo Tuning System) P. Bosland, Bo Wu DAPNIA - CEA Saclay. Abstract SRF Mechanical study of the «Saclay piezo tuner» PTS (Piezo Tuning System) P. Bosland, Bo Wu DAPNIA - CEA Saclay Abstract This report presents the piezo tuner developed at Saclay in the framework of CARE/SRF.

More information

Multi-spectral acoustical imaging

Multi-spectral acoustical imaging Multi-spectral acoustical imaging Kentaro NAKAMURA 1 ; Xinhua GUO 2 1 Tokyo Institute of Technology, Japan 2 University of Technology, China ABSTRACT Visualization of object through acoustic waves is generally

More information

SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC

SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC SUPERCONDUCTING GANTRY AND OTHER DEVELOPMENTS AT HIMAC Y. Iwata *, K. Noda, T. Shirai, T. Murakami, T. Fujita, T. Furukawa, K. Mizushima, Y. Hara, S. Suzuki, S. Sato, and K. Shouda, NIRS, 4-9-1 Anagawa,

More information

XYZ Stage. Surface Profile Image. Generator. Servo System. Driving Signal. Scanning Data. Contact Signal. Probe. Workpiece.

XYZ Stage. Surface Profile Image. Generator. Servo System. Driving Signal. Scanning Data. Contact Signal. Probe. Workpiece. Jpn. J. Appl. Phys. Vol. 40 (2001) pp. 3646 3651 Part 1, No. 5B, May 2001 c 2001 The Japan Society of Applied Physics Estimation of Resolution and Contact Force of a Longitudinally Vibrating Touch Probe

More information

XFEL Cryo System. Project X Collaboration Meeting, FNAL September 8-9, 2010 Bernd Petersen DESY MKS (XFEL WP10 & WP13) 1 st stage. Possible extension

XFEL Cryo System. Project X Collaboration Meeting, FNAL September 8-9, 2010 Bernd Petersen DESY MKS (XFEL WP10 & WP13) 1 st stage. Possible extension XFEL Cryo System Possible extension 1 st stage Project X Collaboration Meeting, FNAL September 8-9, 2010 (XFEL WP10 & WP13) Outline 2 XFEL accelerator structure TESLA technology Basic cryogenic parameters

More information

Diseño del Criostato del Receptor de Banda Ancha

Diseño del Criostato del Receptor de Banda Ancha Diseño del Criostato del Receptor de Banda Ancha José Manuel Serna, Beatriz Vaquero, Félix Tercero, Samuel López Informe Técnico IT-CDT 2015-18 [Los desarrollos descritos en este informe técnico han sido

More information

The VIRGO injection system

The VIRGO injection system INSTITUTE OF PHYSICSPUBLISHING Class. Quantum Grav. 19 (2002) 1829 1833 CLASSICAL ANDQUANTUM GRAVITY PII: S0264-9381(02)29349-1 The VIRGO injection system F Bondu, A Brillet, F Cleva, H Heitmann, M Loupias,

More information

MPI TS300-SE 300 mm Manual Probe System with ShielDEnvironment TM For accurate and reliable DC/CV, RF and mmw measurements

MPI TS300-SE 300 mm Manual Probe System with ShielDEnvironment TM For accurate and reliable DC/CV, RF and mmw measurements MPI TS300-SE 300 mm Manual Probe System with ShielDEnvironment TM For accurate and reliable DC/CV, RF and mmw measurements FEATURES / BENEFITS Universal Use Designed for wide variety of applications such

More information

SONO 1500 CT Ultrasonic flow sensor

SONO 1500 CT Ultrasonic flow sensor Data sheet SONO 1500 CT Ultrasonic flow sensor Description/Application MID examination certificate no.: DE-09-MI004-PTB011 The SONO 1500 CT is an ultrasonic flow sensor especially designed for heating,

More information

DEVELOPMENT OF MOS-FET BASED MARX GENERATOR WITH SELF-PROVED GATE POWER

DEVELOPMENT OF MOS-FET BASED MARX GENERATOR WITH SELF-PROVED GATE POWER DEVELOPMENT OF MOS-FET BASED MARX GENERATOR WITH SELF-PROVED GATE POWER A. Tokuchi 1,2,3, W. Jiang 2, K. Takayama 3, T. Arai 3, T. Kawakubo 3 and T. Adachi 3 1 Pulsed Power Japan Laboratory Ltd., Kusatsu,

More information

Cryocoolers for Space Applications #4

Cryocoolers for Space Applications #4 2015 CEC Cryocooler Short Course Cryocoolers for Space Applications #4 R.G. Ross, Jr. Jet Propulsion Laboratory California Institute of Technology Topics Space Cryocooler Historical Overview and Applications

More information

Description of options, upgrades and accessories for the laser beam stabilization system Compact

Description of options, upgrades and accessories for the laser beam stabilization system Compact Description of options, upgrades and accessories for the laser beam stabilization system Compact The basic configuration of the Compact laser beam stabilization system is fully equipped for stabilization

More information

TECHNICAL SPECIFICATIONS. FOR AN MRBR 7.0 TESLA / 160mm ACTIVELY SHIELDED ROOM TEMPERATURE BORE MAGNET SYSTEM

TECHNICAL SPECIFICATIONS. FOR AN MRBR 7.0 TESLA / 160mm ACTIVELY SHIELDED ROOM TEMPERATURE BORE MAGNET SYSTEM TECHNICAL SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 160mm ACTIVELY SHIELDED ROOM TEMPERATURE BORE MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial

More information

Design of a mechanical resonator to be coupled to a thermoacoustic stirling-engine M.E.H. Tijani Srinivas Vanapalli S. Spoelstra

Design of a mechanical resonator to be coupled to a thermoacoustic stirling-engine M.E.H. Tijani Srinivas Vanapalli S. Spoelstra Design of a mechanical resonator to be coupled to a thermoacoustic stirling-engine M.E.H. Tijani Srinivas Vanapalli S. Spoelstra Proceedings of ASME 2010 3 rd Joint US-European Fluids Engineering Summer

More information

NON-TRADITIONAL MACHINING PROCESSES ULTRASONIC, ELECTRO-DISCHARGE MACHINING (EDM), ELECTRO-CHEMICAL MACHINING (ECM)

NON-TRADITIONAL MACHINING PROCESSES ULTRASONIC, ELECTRO-DISCHARGE MACHINING (EDM), ELECTRO-CHEMICAL MACHINING (ECM) NON-TRADITIONAL MACHINING PROCESSES ULTRASONIC, ELECTRO-DISCHARGE MACHINING (EDM), ELECTRO-CHEMICAL MACHINING (ECM) A machining process is called non-traditional if its material removal mechanism is basically

More information

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components Thomas J. Dunn, Robert Michaels, Simon Lee, Mark Tronolone, and Andrew Kulawiec; Corning Tropel

More information

Stability of a Fiber-Fed Heterodyne Interferometer

Stability of a Fiber-Fed Heterodyne Interferometer Stability of a Fiber-Fed Heterodyne Interferometer Christoph Weichert, Jens Flügge, Paul Köchert, Rainer Köning, Physikalisch Technische Bundesanstalt, Braunschweig, Germany; Rainer Tutsch, Technische

More information

School of Instrument Science and Opto-electronics Engineering, Hefei University of Technology, Hefei, China 2

School of Instrument Science and Opto-electronics Engineering, Hefei University of Technology, Hefei, China 2 59 th ILMENAU SCIENTIFIC COLLOQUIUM Technische Universität Ilmenau, 11 15 September 2017 URN: urn:nbn:de:gbv:ilm1-2017iwk-009:9 Low-Frequency Micro/Nano-vibration Generator Using a Piezoelectric Actuator

More information

AVN Training HartRAO 2016

AVN Training HartRAO 2016 AVN Training HartRAO 2016 Microwave 1 Overview Introduction to basic components used in microwave receivers. Performance characteristics of these components. Assembly of components into a complete microwave

More information

1272. Phase-controlled vibrational laser percussion drilling

1272. Phase-controlled vibrational laser percussion drilling 1272. Phase-controlled vibrational laser percussion drilling Chao-Ching Ho 1, Chih-Mu Chiu 2, Yuan-Jen Chang 3, Jin-Chen Hsu 4, Chia-Lung Kuo 5 National Yunlin University of Science and Technology, Douliou,

More information

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS S-94,316 PATENTS-US-A96698 BEAM POSITION MONITOR RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS CONTRACTUAL ORIGIN OF THE INVENTION The United States Government has rights in this invention pursuant

More information

SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM

SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU,

More information

attocube systems Probe Stations for Extreme Environments CRYOGENIC PROBE STATION fundamentals principles of cryogenic probe stations

attocube systems Probe Stations for Extreme Environments CRYOGENIC PROBE STATION fundamentals principles of cryogenic probe stations PAGE 88 & 2008 2007 PRODUCT CATALOG CRYOGENIC PROBE STATION fundamentals...................... 90 principles of cryogenic probe stations attocps I.......................... 92 ultra stable cryogenic probe

More information

SY-SNSPD-001 Superconducting Nanowire Single Photon Detector System

SY-SNSPD-001 Superconducting Nanowire Single Photon Detector System SY-SNSPD-001 Superconducting Nanowire Single Photon Detector System www.ali-us.com Overview Advanced Lab Instruments SY-SNSPD-001 single-photon detectors system is integrated one or more units Advanced

More information

New Long Stroke Vibration Shaker Design using Linear Motor Technology

New Long Stroke Vibration Shaker Design using Linear Motor Technology New Long Stroke Vibration Shaker Design using Linear Motor Technology The Modal Shop, Inc. A PCB Group Company Patrick Timmons Calibration Systems Engineer Mark Schiefer Senior Scientist Long Stroke Shaker

More information

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Partial Replication of Storms/Scanlan Glow Discharge Radiation Partial Replication of Storms/Scanlan Glow Discharge Radiation Rick Cantwell and Matt McConnell Coolescence, LLC March 2008 Introduction The Storms/Scanlan paper 1 presented at the 8 th international workshop

More information

Cryogenic Systems and Receiver Maintenance

Cryogenic Systems and Receiver Maintenance Cryogenic Systems and Receiver Maintenance Christian Plötz Email: christian.ploetz@bkg.bund.de Federal Agency for Cartography and Geodesy Geodetic Observatory Wettzell Germany Objective Provide basic knowledge

More information

Results from the Stanford 10 m Sagnac interferometer

Results from the Stanford 10 m Sagnac interferometer INSTITUTE OF PHYSICSPUBLISHING Class. Quantum Grav. 19 (2002) 1585 1589 CLASSICAL ANDQUANTUM GRAVITY PII: S0264-9381(02)30157-6 Results from the Stanford 10 m Sagnac interferometer Peter T Beyersdorf,

More information

ECRH on the Levitated Dipole Experiment

ECRH on the Levitated Dipole Experiment ECRH on the Levitated Dipole Experiment S. Mahar, J. Kesner, A.C. Boxer, J.E. Ellsworth, I. Karim, A. Roach MIT PSFC A.K. Hansen, D.T. Garnier, M.E. Mauel, E.E.Ortiz Columbia University Presented at the

More information

Introduction... 3 Slits for AIR Operation... 4 Slits in Vacuum Vessels... 5 Slits for High Vacuum Operation... 6 Custom Slits... 7 Steel Slits...

Introduction... 3 Slits for AIR Operation... 4 Slits in Vacuum Vessels... 5 Slits for High Vacuum Operation... 6 Custom Slits... 7 Steel Slits... Introduction... 3 Slits for AIR Operation... 4 Slits in Vacuum Vessels... 5 Slits for High Vacuum Operation... 6 Custom Slits... 7 Steel Slits... 10 Non-magnetic Options for Slits... 12 Slits with Passive

More information

Measuring the speed of light

Measuring the speed of light 1 Purpose and comments Determine the speed of light by sending a laser beam through various mediums. Unless you want to see like Helen Keller, do not place your eyes in the beam path. Also, Switch the

More information

DEVELOPMENT OF CAPACITIVE LINEAR-CUT BEAM POSITION MONITOR FOR HEAVY-ION SYNCHROTRON OF KHIMA PROJECT

DEVELOPMENT OF CAPACITIVE LINEAR-CUT BEAM POSITION MONITOR FOR HEAVY-ION SYNCHROTRON OF KHIMA PROJECT DEVELOPMENT OF CAPACITIVE LINEAR-CUT BEAM POSITION MONITOR FOR HEAVY-ION SYNCHROTRON OF KHIMA PROJECT Ji-Gwang Hwang, Tae-Keun Yang, Seon Yeong Noh Korea Institute of Radiological and Medical Sciences,

More information

1. INTRODUCTION 2. LASER ABSTRACT

1. INTRODUCTION 2. LASER ABSTRACT Compact solid-state laser to generate 5 mj at 532 nm Bhabana Pati*, James Burgess, Michael Rayno and Kenneth Stebbins Q-Peak, Inc., 135 South Road, Bedford, Massachusetts 01730 ABSTRACT A compact and simple

More information

Generation Laser Scanning Method for Visualizing Ultrasonic Waves Propagating on a 3-D Object

Generation Laser Scanning Method for Visualizing Ultrasonic Waves Propagating on a 3-D Object 1st International Symposium on Laser Ultrasonics: Science, Technology and Applications July 16-18 2008, Montreal, Canada Generation Laser Scanning Method for Visualizing Ultrasonic Waves Propagating on

More information

Active Vibration Isolation of an Unbalanced Machine Tool Spindle

Active Vibration Isolation of an Unbalanced Machine Tool Spindle Active Vibration Isolation of an Unbalanced Machine Tool Spindle David. J. Hopkins, Paul Geraghty Lawrence Livermore National Laboratory 7000 East Ave, MS/L-792, Livermore, CA. 94550 Abstract Proper configurations

More information

HIGH MAGNETIC FIELD SUPERCONDUCTING MAGNETS FABRICATED IN BUDKER INP FOR SR GENERATION

HIGH MAGNETIC FIELD SUPERCONDUCTING MAGNETS FABRICATED IN BUDKER INP FOR SR GENERATION HIGH MAGNETIC FIELD SUPERCONDUCTING MAGNETS FABRICATED IN BUDKER INP FOR SR GENERATION K.V. Zolotarev *, A.M. Batrakov, S.V. Khruschev, G.N. Kulipanov, V.H. Lev, N.A. Mezentsev, E.G. Miginsky, V.A. Shkaruba,

More information

Interferometer signal detection system for the VIRGO experiment. VIRGO collaboration

Interferometer signal detection system for the VIRGO experiment. VIRGO collaboration Interferometer signal detection system for the VIRGO experiment VIRGO collaboration presented by Raffaele Flaminio L.A.P.P., Chemin de Bellevue, Annecy-le-Vieux F-74941, France Abstract VIRGO is a laser

More information

Improving seismic isolation in Advanced LIGO using a ground rotation sensor

Improving seismic isolation in Advanced LIGO using a ground rotation sensor Improving seismic isolation in Advanced LIGO using a ground rotation sensor 04/16/2016 Krishna Venkateswara for UW- Michael Ross, Charlie Hagedorn, and Jens Gundlach aligo SEI team LIGO-G1600083 1 Contents

More information

two pairs of dipole steering windings that t inside the quadrupole yoke an RF beam position monitor (BPM) consisting of a pill box RF cavity,

two pairs of dipole steering windings that t inside the quadrupole yoke an RF beam position monitor (BPM) consisting of a pill box RF cavity, Chapter 6 Quadrupole Package The quadrupole package is shown in Fig. 6.1. It consists of a superferric quadrupole doublet powered in series enclosed in a stainless steel vessel and cooled by 4 K LHe; two

More information

CAVITY DIAGNOSTIC SYSTEM FOR THE VERTICAL TEST OF THE BASELINE SC CAVITY IN KEK-STF

CAVITY DIAGNOSTIC SYSTEM FOR THE VERTICAL TEST OF THE BASELINE SC CAVITY IN KEK-STF CAVITY DIAGNOSTIC SYSTEM FOR THE VERTICAL TEST OF THE BASELINE SC CAVITY IN KEK-STF Y. Yamamoto #, H. Hayano, E. Kako, S. Noguchi, T. Shishido, K. Umemori, K. Watanabe, KEK, Tsukuba, 305-0801, Japan, H.

More information

NEW SITUATION IN COLD MILLS SINCE VTLG THICKNESS MEASUREMENT WITH LASER IS POSSIBLE

NEW SITUATION IN COLD MILLS SINCE VTLG THICKNESS MEASUREMENT WITH LASER IS POSSIBLE IWCC TECHNICAL SEMINAR 2018, CHICAGO NEW SITUATION IN COLD MILLS SINCE VTLG THICKNESS MEASUREMENT WITH LASER IS POSSIBLE IWCC Legal Disclaimer: The purpose of this presentation is to guide programs benefiting

More information

Engineering Model of a High Power Low Temperature Pulse Tube Cryocooler for Space Application

Engineering Model of a High Power Low Temperature Pulse Tube Cryocooler for Space Application 1 Engineering Model of a High Power Low Temperature Pulse Tube Cryocooler for Space Application Y. Pennec 1, J. Butterworth 1, G. Coleiro 1, P. Barbier 1, S. Martin 1,2, P. Crespi 1, I. Charles 2, J-M

More information

Chapter 3. Experimental set up. 3.1 General

Chapter 3. Experimental set up. 3.1 General Chapter 3 Experimental set up 3.1 General Experimental set up and various swirl flow generators such as full length twisted tapes, increasing and decreasing order of twist ratio sets and full length screw

More information

Initial Results from the C-Mod Prototype Polarimeter/Interferometer

Initial Results from the C-Mod Prototype Polarimeter/Interferometer Initial Results from the C-Mod Prototype Polarimeter/Interferometer K. R. Smith, J. Irby, R. Leccacorvi, E. Marmar, R. Murray, R. Vieira October 24-28, 2005 APS-DPP Conference 1 Abstract An FIR interferometer-polarimeter

More information

GA MICROWAVE WINDOW DEVELOPMENT

GA MICROWAVE WINDOW DEVELOPMENT P GA421874 e a MILESTONE NO. 1 TASK ID NOS. T243 (U.S. task 3.2) and T242 (JA Task 2.1) GA MICROWAVE WINDOW DEVELOPMENT by C.P. MOELLER, General Atomics A. KASUGAI, K. SAKAMOTO, and K. TAKAHASHI, Japan

More information

RF power tests of LEP2 main couplers on a single cell superconducting cavity

RF power tests of LEP2 main couplers on a single cell superconducting cavity RF power tests of LEP2 main couplers on a single cell superconducting cavity H.P. Kindermann, M. Stirbet* CERN, CH-1211 Geneva 23, Switzerland Abstract To determine the power capability of the input couplers

More information

Squeezed light and radiation pressure effects in suspended interferometers. Thomas Corbitt

Squeezed light and radiation pressure effects in suspended interferometers. Thomas Corbitt Squeezed light and radiation pressure effects in suspended interferometers Thomas Corbitt MIT Sarah Ackley, Tim Bodiya, Keisuke Goda, David Ottaway, Eugeniy Mihkailov, Daniel Sigg, Nicolas, Smith, Chris

More information

DEVELOPMENT OF STABILIZED AND HIGH SENSITIVE OPTICAL FI- BER ACOUSTIC EMISSION SYSTEM AND ITS APPLICATION

DEVELOPMENT OF STABILIZED AND HIGH SENSITIVE OPTICAL FI- BER ACOUSTIC EMISSION SYSTEM AND ITS APPLICATION DEVELOPMENT OF STABILIZED AND HIGH SENSITIVE OPTICAL FI- BER ACOUSTIC EMISSION SYSTEM AND ITS APPLICATION HIDEO CHO, RYOUHEI ARAI and MIKIO TAKEMOTO Faculty of Mechanical Engineering, Aoyama Gakuin University,

More information

SodiumStar 20/2 High Power cw Tunable Guide Star Laser

SodiumStar 20/2 High Power cw Tunable Guide Star Laser SodiumStar 20/2 High Power cw Tunable Guide Star Laser Laser Guide Star Adaptive Optics Facilities LIDAR Atmospheric Monitoring Laser Cooling SodiumStar 20/2 High Power cw Tunable Guide Star Laser Existing

More information

INTERPLANT STANDARD - STEEL INDUSTRY

INTERPLANT STANDARD - STEEL INDUSTRY INTERPLANT STANDARD - STEEL INDUSTRY IPSS SPECIFICATION OF SENSOR MEASUREMENTS OF LENGTH OF ROLLED MATERIALS IPSS: 2-07-037-13 (First Revision) Corresponding Indian Standard does not exist Formerly-: IPSS:

More information

Optimization of the LCLS Single Pulse Shutter

Optimization of the LCLS Single Pulse Shutter SLAC-TN-10-002 Optimization of the LCLS Single Pulse Shutter Solomon Adera Office of Science, Science Undergraduate Laboratory Internship (SULI) Program Georgia Institute of Technology, Atlanta Stanford

More information

LISA and SMART2 Optical Work in Europe

LISA and SMART2 Optical Work in Europe LISA and SMART2 Optical Work in Europe David Robertson University of Glasgow Outline Overview of current optical system work Title Funded by Main focus Prime Phase Measuring System LISA SMART2 SEA (Bristol)

More information

Cavity BPMs for the NLC

Cavity BPMs for the NLC SLAC-PUB-9211 May 2002 Cavity BPMs for the NLC Ronald Johnson, Zenghai Li, Takashi Naito, Jeffrey Rifkin, Stephen Smith, and Vernon Smith Stanford Linear Accelerator Center, 2575 Sand Hill Road, Menlo

More information

Design, Development and Testing of RF Window for C band 250 kw CW Power Klystron

Design, Development and Testing of RF Window for C band 250 kw CW Power Klystron Available online www.ejaet.com European Journal of Advances in Engineering and Technology, 2016, 3(6): 26-30 Research Article ISSN: 2394-658X Design, Development and Testing of RF Window for C band 250

More information

19 th INTERNATIONAL CONGRESS ON ACOUSTICS MADRID, 2-7 SEPTEMBER 2007

19 th INTERNATIONAL CONGRESS ON ACOUSTICS MADRID, 2-7 SEPTEMBER 2007 19 th INTERNATIONAL CONGRESS ON ACOUSTICS MADRID, 2-7 SEPTEMBER 2007 LARGE CAPACITY ULTRASONIC COMPLEX VIBRATION SOURCES USING MULTIPLE TRANSDUCERS INTEGRATED USING A CIRCULAR VIBRATION DISK PACS: 43.55.Cs

More information

Laser Telemetric System (Metrology)

Laser Telemetric System (Metrology) Laser Telemetric System (Metrology) Laser telemetric system is a non-contact gauge that measures with a collimated laser beam (Refer Fig. 10.26). It measure at the rate of 150 scans per second. It basically

More information

High Power and Energy Femtosecond Lasers

High Power and Energy Femtosecond Lasers High Power and Energy Femtosecond Lasers PHAROS is a single-unit integrated femtosecond laser system combining millijoule pulse energies and high average powers. PHAROS features a mechanical and optical

More information

Development of a Low Cost 3x3 Coupler. Mach-Zehnder Interferometric Optical Fibre Vibration. Sensor

Development of a Low Cost 3x3 Coupler. Mach-Zehnder Interferometric Optical Fibre Vibration. Sensor Development of a Low Cost 3x3 Coupler Mach-Zehnder Interferometric Optical Fibre Vibration Sensor Kai Tai Wan Department of Mechanical, Aerospace and Civil Engineering, Brunel University London, UB8 3PH,

More information

III III 0 IIOI DID IIO 1101 I II 0II II 100 III IID II DI II

III III 0 IIOI DID IIO 1101 I II 0II II 100 III IID II DI II (19) United States III III 0 IIOI DID IIO 1101 I0 1101 0II 0II II 100 III IID II DI II US 200902 19549A1 (12) Patent Application Publication (10) Pub. No.: US 2009/0219549 Al Nishizaka et al. (43) Pub.

More information

PROGRESS IN IFMIF HALF WAVE RESONATORS MANUFACTURING AND TEST PREPARATION

PROGRESS IN IFMIF HALF WAVE RESONATORS MANUFACTURING AND TEST PREPARATION PROGRESS IN IFMIF HALF WAVE RESONATORS MANUFACTURING AND TEST PREPARATION G. Devanz, N. Bazin, G. Disset, H. Dzitko, P. Hardy, H. Jenhani, J. Neyret, O. Piquet, J. Plouin, N. Selami, CEA-Saclay, France

More information

Reliability Studies of the Nozzle/Piezo Units for the WASA-at-COSY Pellet Target

Reliability Studies of the Nozzle/Piezo Units for the WASA-at-COSY Pellet Target Reliability Studies of the Nozzle/Piezo Units for the WASA-at-COSY Pellet Target Florian Bergmann DPG Spring Meeting March 2012 WASA Wide Angle Shower Apparatus Constructed for production and decay studies

More information

INFRARED MEASUREMENTS OF THE SYNTHETIC DIAMOND WINDOW OF A 110 GHz HIGH POWER GYROTRON

INFRARED MEASUREMENTS OF THE SYNTHETIC DIAMOND WINDOW OF A 110 GHz HIGH POWER GYROTRON GA A23723 INFRARED MEASUREMENTS OF THE SYNTHETIC DIAMOND WINDOW by I.A. GORELOV, J. LOHR, R.W. CALLIS, W.P. CARY, D. PONCE, and M.B. CONDON JULY 2001 This report was prepared as an account of work sponsored

More information

Vibration Fundamentals Training System

Vibration Fundamentals Training System Vibration Fundamentals Training System Hands-On Turnkey System for Teaching Vibration Fundamentals An Ideal Tool for Optimizing Your Vibration Class Curriculum The Vibration Fundamentals Training System

More information

SECOND HARMONIC GENERATION AND Q-SWITCHING

SECOND HARMONIC GENERATION AND Q-SWITCHING SECOND HARMONIC GENERATION AND Q-SWITCHING INTRODUCTION In this experiment, the following learning subjects will be worked out: 1) Characteristics of a semiconductor diode laser. 2) Optical pumping on

More information

PRECISION CUTTING MICRACUT 202

PRECISION CUTTING MICRACUT 202 PRECISION CUTTING MICRACUT 202 MICRACUT 202 MICRACUT 202 precision cutters are used for precise and deformation-free cutting of Metals, Ceramics, Electronic Components, Crystals, Composites, Biomaterials,

More information