LDR Prostate Medical Event Definition. John Wochos, MS Gundersen Lutheran La Crosse, Wisconsin April 22, 2011

Size: px
Start display at page:

Download "LDR Prostate Medical Event Definition. John Wochos, MS Gundersen Lutheran La Crosse, Wisconsin April 22, 2011"

Transcription

1 LDR Prostate Medical Event Definition John Wochos, MS Gundersen Lutheran La Crosse, Wisconsin April 22, 2011

2 Conflicts of Interest NONE Three LDR Medical Events at GLMC

3 Outline Definition of Medical Event Recent Prostate Medical Events Significance / Purpose of Medical Event Radiation Oncology Safety Issues Wisconsin LDR Prostate Medical Events NRC, ASTRO position Proposed NCCAAPM Recommendation

4 Philosophy I am NOT a Tea Party Member I do believe there is a strong role for regulators in the use of radiation and radioactive material I do think Radiation Oncology needs significant work in reporting and compiling incidents I do think each institution needs to periodically review procedures from a patient endpoint perspective.

5 Oct 24, 2010

6 Definition of Medical Event

7 Definition cont.

8 Definition Details & Prostate LDR Improper Administrations implies the sources were not administered according to the prescription What is the Prescribed Dosage? Many have taken this to be D90; the dose received by 90% of the target (PTV or prostate)

9 RTOG Protocol 0232 Study designed to compare External Beam with LDR prostate brachytherapy (no where are regulatory implications mentioned). CTV = Prostate PTV = CTV + (2-5mm) all directions except Posterior CT 3-5 weeks post procedure per protocol D90 = 90%-130% Acceptable D90 = 80%-90% OR D90>130% Unacceptable D90 < 80%

10 Philadelphia VA 2008 LDR Prosate Medical Events Analysis by Hagan, et.al., (Brachytherapy, in press) of 62 of the reported Medical Events at the VA Post implant Day 1 CT images were used to evaluate the implants. Edema resulted in a 56% increase in volume (SD 56% also) over the intraoperative ultrasound. Correcting for this edema reduces the number of MEs to 20 using dose criteria.

11 Cont. Philadelphia VA MEs Re-evaluation based on total activity (strength) 11 cases had >20% of the seeds beyond the treatment site (included designed margin). 3 of these had >40% outside treatment site. 10 of these 11 cases also fail the D90 >80% criteria and the one that did not had questionable volume

12 NRC Reported Prostate LDR Med Events (Jan 2010 March 2011) State Level Comments Colorado 32% underdose mistake Air Kerma and Activity Kentucky D90 5.2Gy vs 133 planned Not acknowledge by Institution 27% mistake Air Kerma and Activity Illinois Missed gland by 4-5 cm Indiana D90 72% D90 64% "Dose Received" 44% "Dose Received" 44% North Carolina Peripheral implant; seeds outside gland 59% discovered day of implant Ohio 67% of Presc Dose 75% of Presc Dose Dose <20% of prescribed Cold Cold Cold 72% 68% 71% D90 63% aborted implant 13 seeds in bladder Bones too narrow

13 Cont. MEs State Level Comments Louisianna D90 = 9% Missed by 3 cm Nevada 79% of Prescribed Connecticut D90 58% Gave external beam boost D90 = 59% felt to be adequate coverage Mississippi VA 10 cold events D90 < 80% 1 cold event D90 < 50% Missouri 11% of prescribed

14 Cont. MEs State Level Comments Wisconsin D90 differ 20% D90 = 75% D90 = 75% D90 = 77% D90 = 78% D90 = 121% D90 = 121% D90 = 55% D90 = 78% D90 = 74% D90 = 64% D90 = 75% Overdose Overdose Overdose Overdose Overdose Overdose Underdose Underdose Underdose Underdose D90 = 118% D90 = 69% D90 = 77% D90 = 79% D90 = 70%

15 NRC Reported MEs for LDR Prostate Brachytherapy Nine state with two or fewer events One state (Indiana) with 4 events One state (Mississippi) with 11 events, all VA One state (Ohio) with 11 events One state (Wisconsin) with 28 events

16 Possible Conclusions From Report Medical Events Three states do most of the LDR prostate seed implants? Three states perform significantly worse prostate seed implants? Three states (really two), in particular Wisconsin, have radiation regulators that are investigating differently?

17 Report Non-Dosimetry MEs *W Butler ABS Gross target misses or wrong target treated 9 wrong strength used 11 leaking sources (mostly MICK jams) 8 reports of lost seeds 1 prescription error (monotherapy vs. boost) 1 wrong patient (and I-125 vs Pd-103) 3 shipping errors

18 Nov 21, 2010

19 August 10, 2010 NRC Proposed Amendment voted down Unanimous vote by 5 commissioners NRC staff was recommended to work with their Advisory Committee and the Medical Community to develop a new approach.

20 What happened at Gundersen 3 on-site visits 2 conference calls July 2010 initial on-site inspection indicated we had established criteria, we were thanked and told there would like be no recommendations Follow-up phone call indicated they did not like our criteria. We needed to establish our own criteria and implement them. Subsequent follow up visit September 2010

21 September 2010 visit Long discussions on criteria. We did suggest they follow the NRC recommendation and back off on the grey zone cases, zeroing in on clear events. Again, we were told we could establish our criteria and then we were required to stand by it (sounds like phone call recommendation)

22 Gundersen Initial Criteria and Rationale (Pd-103 exclusively) Since D90 of 80% is acceptable by RTOG standards, a level 20% below that should be an event. Alternatively, if D90 of 90% is a good implant, a D90 of 70% could be an event Upper limit for Pd-103 less well defined, from Brachytherapy 8, D90>140% as upper event. Urethra, no clear limit in literature Bladder, no clear limit in literature Rectum: ESTRO 2cc < 100%

23 Subsequent Conference Call We could pick our own criteria and live by it as long as it was the State s criteria. We were told our interest was clinical results, their interest was a regulation. They felt the +/-20% dose in the regulations was all that was needed. The range of dose recommended in the regulation was intended for I-131 thyroid treatment and dose not apply here. Thus, a D90 of 100% is the implied prescription.

24 What we agreed on D90 < 80% is event D90 > 160% is event (gave us higher value for Pd-103) Subsequent review over 2+ years indicated 3 events out of 187 implants. D90 = 78% (PSA to 0.54) D90 = 71% (Follow-up April 22, 2011) D90 = 77% (PSA to 0.06)

25 Example of Med Event Contouring uncertainty

26 D90 of 80% within uncertainty

27 Definition of Medical Event Improper Administration of radioactive material should we be looking at the RESULT of the procedure to determine if the Administration was Improper? Probably, but.. Total Dosage differs by more than 20% from intended. -->should we be looking at dose? Or, administration of the radioactive material?

28 ASTRO Working Group Problems in Using Dose Dose based definition has many problems Timing of post implant imaging Imaging modality differences Observer variability Planning margins used vary Planning system uncertainties Seed migration (administration of material vs. result)

29 ASTRO Working Group Conclusion If a dose based definition for permanent implants is applied strictly many medically acceptable implants would be inappropriately classified as a Medical Event and rather than achieving the intended goal of patient safety, there will be a detrimental effect on brachytherapy.

30 ASTRO Recommendation Source strength based criterion (>20% of source strength implanted outside PTV) is more appropriate for defining ME in permanent brachytherapy.

31 Thomadsen Proposal Proposed definition For permanent prostate implants, a medical event should be defined when either: A. For the target, D 90 <70% of the clinical target volume (CTV) AND Less than 5% of sources occupy any octant of the PTV, except by design (e.g. for preservation of normal tissue or escalated treatment to a particular region), as specified in the written directive. B. For the normal tissues, 1. For the bladder and rectum, the D 5cc on post-implant dosimetry exceeds 150% of the prescription dose OR 2. For the urethra, the D 5cc on post-implant dosimetry exceeds 150% of its value on the planned, approved dose distribution.

32 Wochos Request NCCAAPM submit a letter to Wisconsin Radiation Safety Section requesting they hold off on the enforcement of a Medical Event using D90 criteria of <80% or >140%. Wait for NRC guidance. There is no radiation safety benefit being seen by the patients (of the 28 MEs found there are NO adverse reactions). Why enforce an ambiguous regulation when there is no benefit to anyone? This is a waste of resources for both the Medical Institutions and the Radiation Section Staff

33 Until Further Guidance Received Current regulation is Dose based There is still a need to call a bad implant a Medical Event D90 was not invented as a regulatory tool, but it is being used as a quantitative measure of implant quality Higher D90 s have better outcomes. But D90 s of 80% still have satisfactory outcomes compared to other modalities Suggest a D90 < 60% for the time being

34 Radiation Oncology & Safety Need to reduce number of incidents (not limited to Medical Events). Even though percent of incidents is low, the results can be horrific (Cure, palliation, side effects, death). Need a national data base on incidents and near misses. Both at institution and national level we need to determine methods to reduce errors. Outside professional review (peer review, e.g. ACR survey (ASTRO)

35 Acknowledgements Bruce Thomadsen Mary Ellen Jafari Kimberly Schmidt Kari Adank (Compliance)

36 Feb 6, 2011

A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate cancer *

A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate cancer * IOP PUBLISHING Phys. Med. Biol. 52 (7) 2147 2156 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/52/8/7 A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate

More information

Aperture Based Inverse Planning AAPM Summer School 2003

Aperture Based Inverse Planning AAPM Summer School 2003 Aperture Based Inverse Planning AAPM Summer School 003 D.M. Shepard, M.A. Earl, Y. Xiao, C.X. Yu Acknowledgements Ziping Jiang Allen Li Shahid Naqvi James Galvin Di Yan Prowess, Inc. University of Maryland

More information

Computers and Medicine

Computers and Medicine Illinois Institute of Technology Computers and Medicine Alexander M. Nicoara CS485: History of Computers Professor Charles Bauer April 10th, 2016 What is the background of the topic? Computers play an

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

Patient-Assisted Compression Impact on Image Quality and Workflow

Patient-Assisted Compression Impact on Image Quality and Workflow Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled

More information

The Safety Case and the Risk-Informed Performance-Based Approach for Management of US Commercial Low-Level Waste (Paper #190)

The Safety Case and the Risk-Informed Performance-Based Approach for Management of US Commercial Low-Level Waste (Paper #190) The Safety Case and the Risk-Informed Performance-Based Approach for Management of US Commercial Low-Level Waste (Paper #190) Rateb (Boby) Abu-Eid, David Esh, and Christopher Grossman Division of Decommissioning,

More information

This is a repository copy of Real-time in vivo dosimetry in high dose rate prostate brachytherapy.

This is a repository copy of Real-time in vivo dosimetry in high dose rate prostate brachytherapy. This is a repository copy of Real-time in vivo dosimetry in high dose rate prostate brachytherapy. White Rose Research Online URL for this paper: http://eprints.whiterose.ac.uk/100200/ Version: Accepted

More information

Initial Certification

Initial Certification Initial Certification Nuclear Medical Physics (NMP) Study Guide Part 2 Content Guide and Sample Questions The content of all ABR exams is determined by a panel of experts who select the items based on

More information

SHTG primary submission process

SHTG primary submission process Meeting date: 24 April 2014 Agenda item: 8 Paper number: SHTG 14-16 Title: Purpose: SHTG primary submission process FOR INFORMATION Background The purpose of this paper is to update SHTG members on developments

More information

Consultation for the Formulation of the NMS IR Programme Dr Dave Rayner Formulator Business Leader, Radiation Dosimetry

Consultation for the Formulation of the NMS IR Programme Dr Dave Rayner Formulator Business Leader, Radiation Dosimetry Consultation for the Formulation of the NMS IR Programme 2007-10 Dr Dave Rayner Formulator Business Leader, Radiation Dosimetry 31 May 2006 The Timetable 24 May Orientation with MAC WG 5 Jun NPL staff

More information

The Finding Respect and Ending Stigma around HIV (FRESH) Study Intervention Post-Workshop Survey Community Participants

The Finding Respect and Ending Stigma around HIV (FRESH) Study Intervention Post-Workshop Survey Community Participants The Finding Respect and Ending Stigma around HIV (FRESH) Study Intervention Post-Workshop Survey Community Participants Date: / / Study ID Number: Thank you for participating in this study. Please fill

More information

The Finding Respect and Ending Stigma around HIV (FRESH) Study Intervention Workshop Survey Community Participants

The Finding Respect and Ending Stigma around HIV (FRESH) Study Intervention Workshop Survey Community Participants The Finding Respect and Ending Stigma around HIV (FRESH) Study Intervention Workshop Survey Community Participants Date: / / Study ID Number: Thank you for participating in this study. Please fill out

More information

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance

More information

Quality and GLP for Histology and Pathology of Drug Safety Studies

Quality and GLP for Histology and Pathology of Drug Safety Studies Quality and GLP for Histology and Pathology of Drug Safety Studies Roger Alison BVSc MRCVS DiplECVP Consultant Toxicological Pathologist What is Quality Histology? It depends upon the purpose - Answer

More information

A Brief Introduction to the Regulatory Environment of Medical Device Supervision. CFDA Department of Legal Affairs Liu Pei

A Brief Introduction to the Regulatory Environment of Medical Device Supervision. CFDA Department of Legal Affairs Liu Pei A Brief Introduction to the Regulatory Environment of Medical Device Supervision CFDA Department of Legal Affairs Liu Pei Development Trend of Medical Device Industry Development Opportunities of Medical

More information

Electronic Brachytherapy Sources. Thomas W. Rusch

Electronic Brachytherapy Sources. Thomas W. Rusch Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

Aim. Images for this section: Page 2 of 13

Aim. Images for this section: Page 2 of 13 Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094

More information

Progress Report from the HDR Brachytherapy Working Party. Margaret Bidmead

Progress Report from the HDR Brachytherapy Working Party. Margaret Bidmead Progress Report from the HDR Brachytherapy Working Party Margaret Bidmead HDR Ir-192 calibration service provided by NPL since May 2004 Primary standard irradiated with a Nucletron classic HDR IR-192 source

More information

8.2.1 Therac-25 Radiation Overdoses

8.2.1 Therac-25 Radiation Overdoses Reuse of software: the Ariane 5 rocket and No Fly lists 8.2 Case Study: The Therac-25 377 Less than 40 seconds after the first launch of France s Ariane 5 rocket, the rocket veered off course and was destroyed

More information

Andalusian Agency for Health Technology Assessment (AETSA)

Andalusian Agency for Health Technology Assessment (AETSA) Andalusian Agency for Health Technology Assessment (AETSA) Seville, 22 nd of July, 2016 Comments on the concept paper Facilitating the translation of advanced therapies to patients in Europe 1 Introduction

More information

The Friends and Family Test score is calculated as outlined in the NHS England guidance issued in Oct-14. The calculation is as follows:

The Friends and Family Test score is calculated as outlined in the NHS England guidance issued in Oct-14. The calculation is as follows: 10 Children s outpatients department - NDDH - Friends and Family Test - Jun-15 to Jul-15 Friends and Family Test question: If your friends and family needed similar care or treatment to you, do you think

More information

Medical Images Analysis and Processing

Medical Images Analysis and Processing Medical Images Analysis and Processing - 25642 Emad Course Introduction Course Information: Type: Graduated Credits: 3 Prerequisites: Digital Image Processing Course Introduction Reference(s): Insight

More information

Quality assurance in the supply chain for pharmaceuticals from the WHO perspective

Quality assurance in the supply chain for pharmaceuticals from the WHO perspective 1 Quality assurance in the supply chain for pharmaceuticals from the WHO perspective Dr Sabine Kopp Quality Assurance and Safety: Medicines Medicines Policy and Standards World Health Organization Presentation

More information

Primary IVF Conditions for Registration For Assisted Reproductive Treatment Providers under the Assisted Reproductive Treatment Act 2008

Primary IVF Conditions for Registration For Assisted Reproductive Treatment Providers under the Assisted Reproductive Treatment Act 2008 Primary IVF Conditions for Registration For Assisted Reproductive Treatment Providers under the Assisted Reproductive Treatment Act 2008 Effective: 1 June 2018 Contents SECTION 1: Background... 3 SECTION

More information

LOREM IPSUM XXX MEDICAL DEVICES NEWS OCTOBER 2012 SPECIAL POINTS OF INTEREST: XXX EDITORIAL

LOREM IPSUM XXX MEDICAL DEVICES NEWS OCTOBER 2012 SPECIAL POINTS OF INTEREST: XXX EDITORIAL LOREM IPSUM MEDICAL DEVICES NEWS OCTOBER 2012 XXX XXX MEDICAL DEVICES NEWS N X XXXXXXXX OCTOBER 2008 2012 EDITORIAL Dear Reader, Welcome to the latest issue of our Medical Devices Newsletter. Within this

More information

2012 ACCE Industry Advisory Board Best Practices Positioning Your Firm After the Great Recession

2012 ACCE Industry Advisory Board Best Practices Positioning Your Firm After the Great Recession 2012 ACCE Industry Advisory Board Best Practices Positioning Your Firm After the Great Recession 2012 FMI Corporation 0 The Great Recession 2012 FMI Corporation 1 FMI Corporation 2010 1 Market Truths You

More information

TGA Discussion Paper 3D Printing Technology in the Medical Device Field Australian Regulatory Considerations

TGA Discussion Paper 3D Printing Technology in the Medical Device Field Australian Regulatory Considerations TGA Discussion Paper 3D Printing Technology in the Medical Device Field Australian Regulatory Considerations MTAA Response - October 2017 October 2017 Australian Regulatory Considerations Page 1 of 7 Level

More information

WANT TO PARTICIPATE IN RESEARCH? THERE S AN APP FOR THAT!

WANT TO PARTICIPATE IN RESEARCH? THERE S AN APP FOR THAT! WANT TO PARTICIPATE IN RESEARCH? THERE S AN APP FOR THAT! SECURITY AND PRIVACY ISSUES WITH THE INCREASED USE OF CONNECTED DEVICES, APPS, AND SOCIAL MEDIA IN RESEARCH. PRESENTED BY: DAVID MATA, SENIOR ASSOCIATE,

More information

How to be a Proactive Patient

How to be a Proactive Patient How to be a Proactive Patient Part I: A Urologist s Perspective September 11, 2017 Presented by: is currently an associate professor with the department of urology in the Levine Cancer Institute at the

More information

3D Brachytherapy with Afterloading Machines

3D Brachytherapy with Afterloading Machines 3D Brachytherapy with Afterloading Machines 3D Brachytherapy/MS Page 1 Introduction 3D-Brachytherapy refers to the case when the planning is performed based on a set of CT, MR or UltraSound (US) images.

More information

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???

More information

CEOCFO Magazine. Pat Patterson, CPT President and Founder. Agilis Consulting Group, LLC

CEOCFO Magazine. Pat Patterson, CPT President and Founder. Agilis Consulting Group, LLC CEOCFO Magazine ceocfointerviews.com All rights reserved! Issue: July 10, 2017 Human Factors Firm helping Medical Device and Pharmaceutical Companies Ensure Usability, Safety, Instructions and Training

More information

Initial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA. Katja Langen Mariana Guerrero Shifeng Chen Shh..

Initial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA. Katja Langen Mariana Guerrero Shifeng Chen Shh.. Initial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA Katja Langen Mariana Guerrero Shifeng Chen Shh..Mobius3D ACKOWLEDGEMENTS Dr. Katja Langen Dr. Shifeng Chen Dr.

More information

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,

More information

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017 A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

Software as a Medical Device (SaMD)

Software as a Medical Device (SaMD) Software as a Medical Device () Working Group Status Application of Clinical Evaluation Working Group Chair: Bakul Patel Center for Devices and Radiological Health US Food and Drug Administration NWIE

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

E5 Implementation Working Group Questions & Answers (R1) Current version dated June 2, 2006

E5 Implementation Working Group Questions & Answers (R1) Current version dated June 2, 2006 INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE E5 Implementation Working Group & (R1) Current version dated June 2, 2006 ICH Secretariat,

More information

Transmission Availability Data System Phase II Final Report

Transmission Availability Data System Phase II Final Report Transmission Availability Data System Phase II Final Report Prepared by the Transmission Availability Data System Task Force for the NERC Planning Committee Approved by the Planning Committee on: Table

More information

What is a coverage determination?

What is a coverage determination? Coverage Determinations Contact Information CALL 1-800-645-3965 TTY/TDD: 1-888-857-4816 WRITE: Medicare Member Appeals Unit, P.O. Box 41820, Philadelphia, PA 19101-1820 VISIT: 1901 Market Street, 1st Floor,

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

IQM Detector Characteristics: Signal reproducibility

IQM Detector Characteristics: Signal reproducibility The Integral Quality Monitor (IQM) System is a real-time beam verification system that monitors the accuracy of radiation delivery throughout each patient treatment without any user interaction. IQM continuously

More information

SAUDI ARABIAN STANDARDS ORGANIZATION (SASO) TECHNICAL DIRECTIVE PART ONE: STANDARDIZATION AND RELATED ACTIVITIES GENERAL VOCABULARY

SAUDI ARABIAN STANDARDS ORGANIZATION (SASO) TECHNICAL DIRECTIVE PART ONE: STANDARDIZATION AND RELATED ACTIVITIES GENERAL VOCABULARY SAUDI ARABIAN STANDARDS ORGANIZATION (SASO) TECHNICAL DIRECTIVE PART ONE: STANDARDIZATION AND RELATED ACTIVITIES GENERAL VOCABULARY D8-19 7-2005 FOREWORD This Part of SASO s Technical Directives is Adopted

More information

June Phase 3 Executive Summary Pre-Project Design Review of Candu Energy Inc. Enhanced CANDU 6 Design

June Phase 3 Executive Summary Pre-Project Design Review of Candu Energy Inc. Enhanced CANDU 6 Design June 2013 Phase 3 Executive Summary Pre-Project Design Review of Candu Energy Inc. Enhanced CANDU 6 Design Executive Summary A vendor pre-project design review of a new nuclear power plant provides an

More information

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality

More information

Ai Group Submission. in response to the REVIEW OF ELECTRICITY (CONSUMER SAFETY) ACT 2004 ISSUES PAPER

Ai Group Submission. in response to the REVIEW OF ELECTRICITY (CONSUMER SAFETY) ACT 2004 ISSUES PAPER Ai Group Submission in response to the REVIEW OF ELECTRICITY (CONSUMER SAFETY) ACT 2004 ISSUES PAPER APRIL 2010 EXECUTIVE SUMMARY The Australian Industry Group (Ai Group) welcomes the opportunity to comment

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Evolution of Research Consortia Increasing Clinical Focus

Evolution of Research Consortia Increasing Clinical Focus 1 July 11, 2015 Evolution of Research Consortia Increasing Clinical Focus Elekta Synergy Research Partnerships Lead to development of Synergy Systems Served as model for other technology research groups

More information

NEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode

NEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode NEMA XR 27-2012 X-ray Equipment for Interventional Procedures User Quality Control Mode Published by: National Electrical Manufacturers Association 1300 North 17th Street, Suite 1752 Rosslyn, Virginia

More information

Mobius3D. Software based IMRT QA

Mobius3D. Software based IMRT QA Mobius3D Software based IMRT QA What is Mobius Medical Systems? Clinical Expertise Software Expertise Nathan Childress, Ph.D., Founder Eli Stevens, Chief Technical Officer Support Expertise Physicists

More information

Tool for reducing uncertainties in the evidence generation for specialised treatments for rare diseases

Tool for reducing uncertainties in the evidence generation for specialised treatments for rare diseases Tool for reducing uncertainties in the evidence generation for specialised treatments for rare diseases In 2017 the multi-stakeholder initiative was set up to: Facilitate a shared understanding of the

More information

LICENSING THE PALLAS-REACTOR USING THE CONCEPTUAL SAFETY DOCUMENT

LICENSING THE PALLAS-REACTOR USING THE CONCEPTUAL SAFETY DOCUMENT LICENSING THE PALLAS-REACTOR USING THE CONCEPTUAL SAFETY DOCUMENT M. VISSER, N.D. VAN DER LINDEN Licensing and compliance department, PALLAS Comeniusstraat 8, 1018 MS Alkmaar, The Netherlands 1. Abstract

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-6 Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for Fluoroscopy Basics Renato Padovani EFOMP Joint ICTP-IAEA Advanced

More information

by Artnarong Thansandote, David W. Lecuyer and Gregory B. Gajda Key Words: Microwave radiation leakage, microwave ovens

by Artnarong Thansandote, David W. Lecuyer and Gregory B. Gajda Key Words: Microwave radiation leakage, microwave ovens Radiation Leakage Of Before-Sale and Used Microwave Ovens by Artnarong Thansandote, David W. Lecuyer and Gregory B. Gajda This paper presents data from surveys of both before-sale and used microwave ovens

More information

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities

More information

Prescription Drug-Use-Related Software; Establishment of a Public Docket; Request for Comments (Docket No. FDA-2018-N-3017)

Prescription Drug-Use-Related Software; Establishment of a Public Docket; Request for Comments (Docket No. FDA-2018-N-3017) January 22, 2019 Via Electronic Submission Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Rm. 1061 Rockville, MD 20852 Re: Prescription Drug-Use-Related Software;

More information

YOUR RIGHTS. In Local Authority Services. Texas Department of Aging and Disability Services. Published by

YOUR RIGHTS. In Local Authority Services. Texas Department of Aging and Disability Services. Published by YOUR RIGHTS In Local Authority Services Published by Texas Department of Aging and Disability Services YOUR RIGHTS This book belongs to: Your Rights in Local Authority Programs Table of contents A note

More information

Is the scanned image stored as a color, grayscale, or black and white image? If applicable, what resolution is used?

Is the scanned image stored as a color, grayscale, or black and white image? If applicable, what resolution is used? Topic: Trademarks, Trade names, Service marks Question by: Mandy Harlan/LA s IT staff Jurisdiction: Louisiana Date: 25 August 2010 Jurisdiction Question(s) Manitoba Corporations Canada Alabama Alaska Arizona

More information

Canada s Ballast Water Requirements. September 2016

Canada s Ballast Water Requirements. September 2016 Canada s Ballast Water Requirements September 2016 Applicability of Canada s Regulations Ballast Water Control and Management Regulations require vessels from outside Canada s Exclusive Economic Zone (EEZ)

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

3D Bio printing. Nazrawit Mekonnen IT /29/15. "By placing this statement on my webpage, I certify that I have read and understand the GMU

3D Bio printing. Nazrawit Mekonnen IT /29/15. By placing this statement on my webpage, I certify that I have read and understand the GMU 3-D Bio printing 1 3D Bio printing Nazrawit Mekonnen IT-104-006 Professor Jayasree Jayaram 9/29/15 "By placing this statement on my webpage, I certify that I have read and understand the GMU Honor Code

More information

Re: Examination Guideline: Patentability of Inventions involving Computer Programs

Re: Examination Guideline: Patentability of Inventions involving Computer Programs Lumley House 3-11 Hunter Street PO Box 1925 Wellington 6001 New Zealand Tel: 04 496-6555 Fax: 04 496-6550 www.businessnz.org.nz 14 March 2011 Computer Program Examination Guidelines Ministry of Economic

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

JIS G3465 Seamless Steel Tubes for Drilling

JIS G3465 Seamless Steel Tubes for Drilling JIS G3465 Seamless Steel Tubes for Drilling 1. Scope This Japanese Industrial Standard specifies the seamless steel tubes, hereinafter referred to as the "tubes", used for casing tube, core tube and drill

More information

IEEE ICES Exposure Limits Above 6 GHz

IEEE ICES Exposure Limits Above 6 GHz Mobile Manufacturers Forum Workshop EMF Exposure Limits and Compliance Assessment of Future Wireless Devices Above 6 GHz Exposure Limits Above 6 GHz Dr. C-K. Chou* TC95 Chairman International Committee

More information

Phase 2 Executive Summary: Pre-Project Review of AECL s Advanced CANDU Reactor ACR

Phase 2 Executive Summary: Pre-Project Review of AECL s Advanced CANDU Reactor ACR August 31, 2009 Phase 2 Executive Summary: Pre-Project Review of AECL s Advanced CANDU Reactor ACR-1000-1 Executive Summary A vendor pre-project design review of a new nuclear power plant provides an opportunity

More information

Ethics in Materials Engineering

Ethics in Materials Engineering Ethics in Materials Engineering Dr. Parviz Yavari Dr. Ehsan Barjasteh Picture : https://www.linkedin.com/topic/ethical-reasoning Contents 1.Ethics/ Morality/Laws 2.Ethics in Engineering 3.Ethics in material

More information

HOW TO GET SPECIALTY CARE AND REFERRALS

HOW TO GET SPECIALTY CARE AND REFERRALS Insert for HARP Member Handbooks THE BELOW SECTIONS OF YOUR MEMBER HANDBOOK HAVE BEEN REVISED TO READ AS FOLLOWS HOW TO GET SPECIALTY CARE AND REFERRALS If you need care that your PCP cannot give, he or

More information

Reliability Guideline: Generating Unit Operations During Complete Loss of Communications

Reliability Guideline: Generating Unit Operations During Complete Loss of Communications 1 1 1 1 1 1 1 1 0 1 0 1 0 1 Reliability Guideline: Generating Unit Operations During Complete Loss of Communications Preamble: It is in the public interest for the North American Electric Reliability Corporation

More information

2005 HSC Notes from the Marking Centre Food Technology

2005 HSC Notes from the Marking Centre Food Technology 2005 HSC Notes from the Marking Centre Food Technology 2006 Copyright Board of Studies NSW for and on behalf of the Crown in right of the State of New South Wales. This document contains Material prepared

More information

American Nuclear Society

American Nuclear Society American Nuclear Society 1 Unraveling the Mystery of Consensus Standards Presented by: The American Nuclear Society Standards Committee January 31, 2017 Copyright 2017 by American Nuclear Society Purpose

More information

Reliability Guideline: Generating Unit Operations During Complete Loss of Communications

Reliability Guideline: Generating Unit Operations During Complete Loss of Communications 1 1 1 1 1 1 1 1 0 1 0 1 0 1 Reliability Guideline: Generating Unit Operations During Complete Loss of Communications Preamble It is in the public interest for the North American Electric Reliability Corporation

More information

A BALANCING ACT: STREAMLINING METHOD TRANSFERS WITHOUT COMPROMISING COMPLIANCE OR SCIENCE JULIE FROST SENIOR SPECIALIST, QUALITY CONTROL

A BALANCING ACT: STREAMLINING METHOD TRANSFERS WITHOUT COMPROMISING COMPLIANCE OR SCIENCE JULIE FROST SENIOR SPECIALIST, QUALITY CONTROL A BALANCING ACT: STREAMLINING METHOD TRANSFERS WITHOUT COMPROMISING COMPLIANCE OR SCIENCE JULIE FROST SENIOR SPECIALIST, QUALITY CONTROL PRESENTATION OUTLINE Introduction The Improved Comparative Study

More information

Half value layer and AEC receptor dose compliance survey in Estonia

Half value layer and AEC receptor dose compliance survey in Estonia Half value layer and AEC receptor dose compliance survey in Estonia K. Kepler, A. Vladimirov Training Centre of Medical Physics, University of Tartu Testing Centre of the University of Tartu, Estonia E-mail:

More information

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA

More information

State of the Art Film Dosimetry

State of the Art Film Dosimetry State of the Art Film Dosimetry Micke A., Lewis D. Advanced Materials Ashland proprietary technology, patents pending Film Dosimetry Radiochromic Film EBT2/EBT3 One-Scan Protocol Multi-channel Film Dosimetry

More information

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S. Practical Medical Physics Session: TG-151 Dose Monitoring August 5, 2013 Katie Hulme, M.S. Digital Imaging and Dose Creep Images courtesy of Agfa Healthcare Under-Exposed Over-Exposed Freedman et al.,

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Stage 2: eligibility screening. Stage 3: prioritisation. Stage 4: selection

Stage 2: eligibility screening. Stage 3: prioritisation. Stage 4: selection Digital therapy technology endorsement for IAPT project eligibility and prioritisation criteria NICE has been commissioned by NHS England to assess selected, digitally enabled therapies for depression

More information

NATIONAL POLICY ON OILED BIRDS AND OILED SPECIES AT RISK

NATIONAL POLICY ON OILED BIRDS AND OILED SPECIES AT RISK NATIONAL POLICY ON OILED BIRDS AND OILED SPECIES AT RISK January 2000 Environment Canada Canadian Wildlife Service Environnement Canada Service canadien de la faune Canada National Policy on Oiled Birds

More information

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes

More information

Who Benefited From Transportation Improvements?

Who Benefited From Transportation Improvements? Who Benefited From Transportation Improvements? We ve seen that many of the transportation improvements led to major reductions in shipping costs but didn t necessarily lead to big profits for investors

More information

The Cremation (Scotland) Regulations Consultation. Introduction. The regulations

The Cremation (Scotland) Regulations Consultation. Introduction. The regulations The Cremation (Scotland) Regulations 2017 Consultation Introduction 1. The Burial and Cremation (Scotland) Act 2016 1 provides a modern, comprehensive legislative framework for burial and cremation. The

More information

Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp , Jul Presented by Dror Feitelson

Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp , Jul Presented by Dror Feitelson Nancy G. Leveson and Clark S. Turner, An Investigation of the Therac-25 Accidents. Computer 26(7), pp. 18-41, Jul 1993. Presented by Dror Feitelson The Big Picture The Therac-25 was a computerized radiation

More information

PMDA perspective on Quality by Design for pharmaceutical products

PMDA perspective on Quality by Design for pharmaceutical products PMDA perspective on Quality by Design for pharmaceutical products Junichi Fukuchi, Ph.D. Office of Cellular and Tissue-based Products Pharmaceuticals and Medical Devices Agency (PMDA) Annual conference

More information

Frequently Asked Questions

Frequently Asked Questions Table of Contents Who should an Investigator contact to submit an ERP proposal? What type of information is needed for Medtronic to review? Is a protocol required? Why is so much information necessary?

More information

Consistent with Labeling Final Guidance: Implications for Drug Products

Consistent with Labeling Final Guidance: Implications for Drug Products Consistent with Labeling Final Guidance: Implications for Drug Products Timothy Candy, Principal Consultant, Regulatory Affairs, Opus Regulatory, Inc. Sue Gregory, Managing Counsel, Merck & Co., Inc. Kate

More information

CENTER FOR DEVICES AND RADIOLOGICAL HEALTH. Notice to Industry Letters

CENTER FOR DEVICES AND RADIOLOGICAL HEALTH. Notice to Industry Letters CENTER FOR DEVICES AND RADIOLOGICAL HEALTH Standard Operating Procedure for Notice to Industry Letters PURPOSE This document describes the Center for Devices and Radiological Health s (CDRH s, or Center

More information

TECHNOLOGY QUALIFICATION MANAGEMENT

TECHNOLOGY QUALIFICATION MANAGEMENT OFFSHORE SERVICE SPECIFICATION DNV-OSS-401 TECHNOLOGY QUALIFICATION MANAGEMENT OCTOBER 2010 FOREWORD (DNV) is an autonomous and independent foundation with the objectives of safeguarding life, property

More information

Drug Safety. Electronic Supplementary Material. Krska J and Morecroft CW

Drug Safety. Electronic Supplementary Material. Krska J and Morecroft CW Drug Safety Patients Use of Information About Medicine Side Effects in Relation to Experiences of Suspected Adverse Drug Reactions: A Cross- Section Survey in Medical In-Patients Krska J and Morecroft

More information

Identifying and Managing Joint Inventions

Identifying and Managing Joint Inventions Page 1, is a licensing manager at the Wisconsin Alumni Research Foundation in Madison, Wisconsin. Introduction Joint inventorship is defined by patent law and occurs when the outcome of a collaborative

More information

Registration of Innovative Medical Devices in China

Registration of Innovative Medical Devices in China Registration of Innovative Medical Devices in China July 2017 Ⅰ Policies on Registration of Innovative Medical Devices Ⅱ Registration of Innovative Medical Devices Ⅲ Registration of Medical Devices with

More information

FDA Oversight Of Drug Safety: What Works, What Doesn t. Geoffrey Levitt Chief Counsel, Regulatory and Research Wyeth Pharmaceuticals August 24, 2006

FDA Oversight Of Drug Safety: What Works, What Doesn t. Geoffrey Levitt Chief Counsel, Regulatory and Research Wyeth Pharmaceuticals August 24, 2006 FDA Oversight Of Drug Safety: What Works, What Doesn t Geoffrey Levitt Chief Counsel, Regulatory and Research Wyeth Pharmaceuticals August 24, 2006 FDA Oversight of Drug Safety: The Stakes 100,000 deaths/year

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Medical Devices cyber risks and threats

Medical Devices cyber risks and threats Medical Devices cyber risks and threats David Grainger Senior Medical Device Specialist MHRA The challenges of software medical device regulation. david.grainger@mhra.gov.uk Current framework 1998 In Vitro

More information

Applying Advanced Technologies to Improve NPP Productivity

Applying Advanced Technologies to Improve NPP Productivity Applying Advanced Technologies to Improve NPP Productivity Lew Hanes, Consultant Joe Naser, EPRI Plant Productivity Improvement Through Advanced Technology - Group Kickoff Meeting June 29-30, 2010 Presentation

More information

Questioning $1 Million Fee in a Needle Deal

Questioning $1 Million Fee in a Needle Deal July 19, 2002 Questioning $1 Million Fee in a Needle Deal By BARRY MEIER with MARY WILLIAMS WALSH group that buys medical supplies for a third of the nation's hospitals received two highly unusual payments

More information