Ratio Imaging. Dividing one image by another to detect changing conditions. Images collected at different times, wavelengths, polarities, etc
|
|
- Susanna Cain
- 6 years ago
- Views:
Transcription
1 Ratio Imaging Dividing one image by another to detect changing conditions Images collected at different times, wavelengths, polarities, etc Most common use of ratio imaging is to provide a quick spectral fingerprint
2 Ratio Imaging 458 excitation (ph-insensitive) stomach 514 excitation (ph-sensitive) Ratio image (calculation) ph4 perfusate + dye ph3 70 µm
3 Ratio Imaging What are the best wavelengths to choose for ratioing? Dynamic range of ratio values A: 9-fold change B (isosbestic point): no change C: 2-fold change A B C Reality of your noisy detector and specimen may mean that too extreme a change gives poor data
4 Ratio Imaging Ratios help compensate for Loss of dye (photobleach/leak) Variability in local dye concentration Variability in cell thickness (non-confocal) Fluctuations in excitation light intensity Changes in detector sensitivity/gain Ratios will NOT help compensate for Saturation in detector or fluorophore response Incorrectly adjusted gains/settings on imaging device Contribution of background autofluorescence Nothing gets around the garbage in garbage out rule
5 Probes Ion selective probes K+ PBFO, PBFI Na+ SBFO, SBFI Ca++ Fura2, Indo1, calcium cameleon FP. Fluo3, Fura Green, FuraRed, calcein ph BCECF, SNARFs, SNAFL, all FPs Mg++ Mag-Fura2 Cl- SPQ, MQAE, YFP HCO 3 -? Membrane potential 8-ANEPPS, oxynols, styrl dyes, DiOC 7 Volume Calcein Sensors must change response over a relevant physiological range Sensors must be selective for the event of interest
6 Probes Fura2-acetoxymethylester (AM) 1-10 µm AM fluorescence Toxic cleavage products Partially cleaved forms Buffering Fura2-AM esterases µm Fura2 Cell homeostasis perturbation Cell survival mm SPQ Hypotonic shock µm SPQ
7 Collect images with minimal time separation Camera detector: entire image1, change settings, entire image2 PMT based confocal: emission ratio multiple images simultaneously essential for processes on msec scale excitation ratio frame switching (same as camera) line switching (1-10 ms betw. colors) point switching? Speed of your imaging device (and your perfusion system) must be compatible with the timing of the biological process you want to capture
8 Quantitative Imaging Set detector background correctly: Background image pixels all greater than zero #pixels #pixels Gray level
9 Quantitative Imaging Minimal noise is that your values will vary ±1 gray level ~ 10% noise ~1% noise 10 ± 1 10 ± 1 ~ 20% noise 100 ± ± 1 ~ 2% noise Ratio value and images are noisier than raw data High gray values (more significant digits) for best precision
10 Quantitative Imaging Test linearity of response: Doubling light to detector should give double the signal Fluorescence microscope: neutral density filter Laser scanning microscope: reflectance, AOTF control No saturated signals: Detector Digital image dynamic range Fluorophore saturation
11 Quantitative Imaging 150 Cl-NERF 514nm/458nm intensity ratio x ph Even with optimized, linear responses of excitation and emission, you must always work within range of dye response
12 Quantitative Imaging Calculating Ratios Threshold: eliminate low values from image that would give noisy ratio values. ONLY if most data is clearly separated from noise floor Subtract background (autofluorescence + detector) Extract average values from regions of interest in fluorescence images, then manually ratio for highest precision Calculate ratio image for presentation
13 Quantitative Imaging Daily calibration Never assume the system is the same as yesterday Calibrate at the end of every experiment Suggestion: single point ratio calibration (dye in known solution) so you can scale response to known calibration curve Gold standard Ratio 5 Today ph 4 ratio = /10 Ratio Scaled calibration curve ph ph
14 Thick specimen imaging Light scatter Light absorption More trouble to get light in, get light out and still believe you know what your numbers mean. Increasing depth of focus means checking the fidelity of your light path even before you get a biological specimen on the stage!
15 Picking an objective for an aqueous environment Zeiss 63X NA 1.4 oil 640nm 580nm Normalized fluorescence Leitz 50X NA 1.0 water Normalized fluorescence Depth of focus into SNARF-1 ocean (µm)
16 Picking an objective for an aqueous environment SNARF-1 emission ratio µm µm µm ph 8.0 Step 1 All AOK in solution Step 2 How check if it will work in thick specimens like native tissue samples?
17 An inert control: Lucifer Yellow Broad excitation Broad emission Boring EX EM
18 An inert control: Lucifer Yellow ph and ion insensitive Cheap Small: 450 g/mole Dextran conjugates Use to check fidelity of optical path Ideal condition is to match excitation and emission colors with your real probe 640nm/580nm ratio SNARF-1 Lucifer yellow ph
19 The live tissue nemesis: Light Scatter LUCIFER YELLOW fluorescence (normalized) 640nm 580nm Depth of focus into tissue (µm)
20 Does ratio imaging correct for light scatter artifacts in tissue? Lucifer Yellow fluorescence ratio (640nm/580nm) Crypt lumen Adjacent tissue Depth of focus into tissue (µm)
21 Where am I? Confocal reflectance (rat gastric surface) pits lumen mucus epithelium 50 µm XZ plane XY plane
ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006
ADVANCED METHODS FOR CONFOCAL MICROSCOPY II Jean-Yves Chatton Sept. 2006 Workshop outline Confocal microscopy of living cells and tissues X-Z scanning Time series Bleach: FRAP, photoactivation Emission
More informationLSM 510 META in Chang Gung University
Content LSM 510 META in Chang ung University LSM 510 META 路 理 The features and applications of LSM 510 META 01-09 Introduction of the hardware 10-12 Fluorescence observation in conventional microscope
More informationInvitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner
Invitation for a walk through microscopy Sebastian Schuchmann Jörg Rösner joerg.roesner@charite.de Techniques in microscopy Conventional (light) microscopy bright & dark field, phase & interference contrast
More informationMultifluorescence The Crosstalk Problem and Its Solution
Multifluorescence The Crosstalk Problem and Its Solution If a specimen is labeled with more than one fluorochrome, each image channel should only show the emission signal of one of them. If, in a specimen
More informationWhy and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005
Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the
More informationFluorescent Indicators. Martin Thomas, Cairn Research Ltd
Fluorescent Indicators Martin Thomas, Cairn Research Ltd ptical Measurements Are Sensitive! Electric current 1A = 6.25 x 10 18 electrons/sec Squid axon voltage clamp 1mA 10 16 charges/sec Microelectrode
More informationFundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy. Integrated Microscopy Course
Fundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy Integrated Microscopy Course Review Lecture 1: Microscopy Basics Light train Kohler illumination*
More informationBasics of confocal imaging (part I)
Basics of confocal imaging (part I) Swiss Institute of Technology (EPFL) Faculty of Life Sciences Head of BIOIMAGING AND OPTICS BIOP arne.seitz@epfl.ch Lateral resolution BioImaging &Optics Platform Light
More informationBi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging
Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this
More informationModes of light microscopy Choosing the appropriate system
Modes of light microscopy Choosing the appropriate system Wide-field microscopy Confocal microscopy Multi-photon microscopy Wide-field, inverted fluorescence microscope Nikon MicroscopyU Endosome migration
More informationSupplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each
Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses
More informationBoulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement
Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement CONFOCAL MICROSCOPY BioVis Uppsala, 2017 Jeremy Adler Matyas Molnar Dirk Pacholsky Widefield & Confocal Microscopy
More information1.The Problem LIGHT-LEVEL LEVEL IMAGING. light-level level Cameras. 3. Solutions. 2. Low-light LOW-LIGHT
LOW-LIGHT LIGHT-LEVEL LEVEL IMAGING 1.The Problem 2. Low-light light-level level Cameras 3. Solutions How Much Light? I. Illumination system: 75 W Xenon Arc (~1mW/nm in visible) 490/10 nm exciter filter
More informationConfocal Microscopy. Kristin Jensen
Confocal Microscopy Kristin Jensen 17.11.05 References Cell Biological Applications of Confocal Microscopy, Brian Matsumoto, chapter 1 Studying protein dynamics in living cells,, Jennifer Lippincott-Schwartz
More informationFLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky
FLUORESCENCE MICROSCOPY Matyas Molnar and Dirk Pacholsky 1 The human eye perceives app. 400-700 nm; best at around 500 nm (green) Has a general resolution down to150-300 μm (human hair: 40-250 μm) We need
More informationLSM 710 Confocal Microscope Standard Operation Protocol
LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.
More informationConfocal Microscopy. (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7. November 2017
Confocal Microscopy (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7 November 2017 3 Flavours of Microscope Confocal Laser Scanning Problem: Out of Focus Light Spinning disc 2-Photon
More informationMicroscopy from Carl Zeiss
Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path
More information3. are adherent cells (ie. cells in suspension are too far away from the coverslip)
Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive
More informationProducts - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software
Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software Arrayit offers the world s only next generation microarray scanning technology, with proprietary rotary motion control,
More informationImaging Retreat - UMASS Customized real-time confocal and 2-photon imaging
Imaging Retreat - UMASS 2012 Customized real-time confocal and 2-photon imaging Mike Sanderson Department of Microbiology and Physiological Systems University of Massachusetts Medical School Thanks for
More informationImaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal
Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Todays Goal: Introduce some additional functionalities of the Leica SP8 confocal HyD vs. PMT detectors Dye Assistant Scanning By
More informationShreyash Tandon M.S. III Year
Shreyash Tandon M.S. III Year 20091015 Confocal microscopy is a powerful tool for generating high-resolution images and 3-D reconstructions of a specimen by using point illumination and a spatial pinhole
More informationOperation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009
Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton
More informationNature Methods: doi: /nmeth Supplementary Figure 1. Comparison of HySP and linear unmixing under different signal-to-noise ratios (SNRs).
Supplementary Figure 1 Comparison of HySP and linear unmixing under different signal-to-noise ratios (SNRs). (a) TrueColor images of 32 channel datasets of zebrafish labeled with H2B-Cerulean, kdrl:egfp,
More informationMaria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging
Nonlinear microscopy I: Two-photon fluorescence microscopy Multiphoton Microscopy What is multiphoton imaging? Applications Different imaging modes Advantages/disadvantages Scattering of light in thick
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power
More informationExamination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,
KTH Applied Physics Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, 2009-06-05, 8-13, FB51 Allowed aids: Compendium Imaging Physics (handed out) Compendium Light Microscopy
More informationSupplemental Method Information Zeiss LSM710
Supplemental Method Information Zeiss LSM710 1 Under the Light Path window set up the confocal for imaging a green dye (Alexa488-EGFP). For example, set up the light path as shown here using the 488 nm
More information3D light microscopy techniques
3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 1 1 2! NA = 0.5! NA 2D imaging
More informationZeiss 780 Training Notes
Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and
More informationScanArray Overview. Principle of Operation. Instrument Components
ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional
More informationPractical work no. 3: Confocal Live Cell Microscopy
Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside
More informationBi Imaging. Multicolor Imaging: The Important Question of Co-Localization. Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK
Multicolor Imaging: The Important Question of Co-Localization Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK The use of specific fluorescent probes, combined with confocal or multiphoton microscopy
More informationConfocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:
Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn
More information1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the
1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the light path as shown here using the 488 nm LASER (Laser Unit 1) reflecting off of the 405/488 nm Dichroic mirror
More information3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.
1 Set up the confocal light path for imaging a green dye (e.g. Alexa488-EGFP). For example, under the Configuration Control window the light path could be set up as shown here using the 488 nm LASER (found
More informationConfocal Imaging Through Scattering Media with a Volume Holographic Filter
Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,
More informationZeiss 880 Training Notes Zen 2.3
Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start
More informationQuick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s
LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1
More information3D light microscopy techniques
3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 2D imaging 3D imaging Resolution
More informationWorking Simultaneously. The Next Level of TIRF Microscopy. cell^tirf Illuminator Motorized Total Internal Reflection Fluorescence
cell^tirf Illuminator Motorized Total Internal Reflection Fluorescence Four individually aligned illumination beams for simultaneous multi-color TIRF imaging Working Simultaneously The Next Level of TIRF
More informationMULTIPHOTON MICROSCOPY. Matyas Molnar Dirk Pacholsky
MULTIPHOTON MICROSCOPY Matyas Molnar Dirk Pacholsky Information Information given here about 2 Photon microscopy were mainly taken from these sources: Background information on 2-Photon microscopy: http://micro.magnet.fsu.edu/primer/techniques/fluorescence/multiphoton/
More information長庚大學共軛焦顯微鏡課程 長庚大學共軛焦顯微鏡課程. Spot light 長庚大學
長庚大學共軛焦顯微鏡課程 Spot light 長庚大學共軛焦顯微鏡課程 20071030 長庚大學 Basic principle of Laser Scanning Confocal Microscopy The application of LSM 510 META detector Multiphoton microscopy basic principle and introduction
More informationConfocal Application Letter No. 13. Sequential Scan for Leica TCS NT/SP systems
Confocal Application Letter No. 13 Sequential Scan for Leica TCS NT/SP systems Leica Microsystems Heidelberg GmbH Im Neuenheimer Feld 518 D-69120 Heidelberg Telephone +49 6221 4148 0 Fax +49 6221 414833
More informationLSM 780 Confocal Microscope Standard Operation Protocol
LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch
More informationDigital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal
Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Yashvinder Sabharwal, 1 James Joubert 2 and Deepak Sharma 2 1. Solexis Advisors LLC, Austin, TX, USA 2. Photometrics
More informationFinal Exam, 150 points PMB 185: Techniques in Light Microscopy
Final Exam, 150 points Name PMB 185: Techniques in Light Microscopy Point value is in parentheses at the end of each question. Note: GFP = green fluorescent protein ; CFP = cyan fluorescent protein ; YFP
More informationChemical Imaging. Whiskbroom Imaging. Staring Imaging. Pushbroom Imaging. Whiskbroom. Staring. Pushbroom
Chemical Imaging Whiskbroom Chemical Imaging (CI) combines different technologies like optical microscopy, digital imaging and molecular spectroscopy in combination with multivariate data analysis methods.
More informationConfocal Laser Scanning Microscopy
Name of the Core Facility: Confocal Laser Scanning Microscopy CORE Forschungszentrum Immunologie Mainz Welcome to the CSLM Core Facility: The CLSM Core Facility enables working groups to incorporate high
More informationZeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling
Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front
More informationLife Science Instrumentation. New Generation. Light Sheet Fluorescence Microscope. Alph
Life Science Instrumentation Light Sheet Fluorescence Microscope New Generation Alph Modular Light Sheet Microscope Alpha 3 is a new generation of light sheet fluorescence microscope addressing the needs
More informationDynamic Confocal Imaging of Living Brain. Advantages and risks of multiphoton microscopy in physiology
Dynamic Confocal Imaging of Living Brain Advantages and risks of multiphoton microscopy in physiology Confocal laser scanning microscopy In conventional optical microscopy focused and out-offocus light
More informationWe attempted to separate the two dyes by acquiring images using a single excitation wavelength and just two emission wavelengths.
TN437: Spectral Separation of monochrome images using Volocity 4.0 Introduction Spectral Separation is a technique that allows the user to separate images containing data from more than one fluorochrome
More informationQuick Start Guide. Leica SP5 X
Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons
More informationGuide to Confocal 5. Starting session
Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope
More informationNature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup.
Supplementary Figure 1 Schematic of 2P-ISIM AO optical setup. Excitation from a femtosecond laser is passed through intensity control and shuttering optics (1/2 λ wave plate, polarizing beam splitting
More informationIntroduction to light microscopy
Center for Microscopy and Image Anaylsis Introduction to light microscopy (an overview) Microscopy with light Components of a light microscope 1. Light source 2. Objective 3. Sample or specimen holder
More informationVery short introduction to light microscopy and digital imaging
Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and
More informationLeica Sp5 II Confocal User Guide
Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on
More informationNon-Descanned FLIM Detection in Multiphoton Microscopes
Non-Descanned FLIM Detection in Multiphoton Microscopes Abstract. Multiphoton microscopes use a femtosecond NIR laser to excite fluorescence in the sample. Excitation is performed via a multi-photon absorption
More informationPoint Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy
Bi177 Lecture 5 Adding the Third Dimension Wide-field Imaging Point Spread Function Deconvolution Confocal Laser Scanning Microscopy Confocal Aperture Optical aberrations Alternative Scanning Microscopy
More informationBASICS OF CONFOCAL IMAGING (PART I)
BASICS OF CONFOCAL IMAGING (PART I) INTERNAL COURSE 2012 LIGHT MICROSCOPY Lateral resolution Transmission Fluorescence d min 1.22 NA obj NA cond 0 0 rairy 0.61 NAobj Ernst Abbe Lord Rayleigh Depth of field
More informationRapid Adaptive Optical Recovery of Optimal Resolution over Large Volumes
SUPPLEMENTARY MATERIAL Rapid Adaptive Optical Recovery of Optimal Resolution over Large Volumes Kai Wang, Dan Milkie, Ankur Saxena, Peter Engerer, Thomas Misgeld, Marianne E. Bronner, Jeff Mumm, and Eric
More informationTechnology Note ZEISS LSM 880 with Airyscan
Technology Note ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode Author: Dr. Annette Bergter Carl Zeiss Microscopy GmbH,
More informationTRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO
TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped
More informationObserving Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros
PHARMACEUTICAL MICROBIOLOGY JIGAR SHAH INSTITUTE OF PHARMACY NIRMA UNIVERSITY Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens.
More informationNikon Instruments Europe
Nikon Instruments Europe Recommendations for N-SIM sample preparation and image reconstruction Dear customer, We hope you find the following guidelines useful in order to get the best performance out of
More informationLMT F14. Cut in Three Dimensions. The Rowiak Laser Microtome: 3-D Cutting and Imaging
LMT F14 Cut in Three Dimensions The Rowiak Laser Microtome: 3-D Cutting and Imaging The Next Generation of Microtomes LMT F14 - Non-contact laser microtomy The Rowiak laser microtome LMT F14 is a multi-purpose
More informationImaging Introduction. September 24, 2010
Imaging Introduction September 24, 2010 What is a microscope? Merriam-Webster: an optical instrument consisting of a lens or combination of lenses for making enlarged images of minute objects; especially:
More information1 Co Localization and Working flow with the lsm700
1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with
More informationThe Next Level of TIRF Microscopy. cell^tirf Illuminator Motorized Total Internal Reflection Fluorescence
cell^tirf Illuminator Motorized Total Internal Reflection Fluorescence Four individually aligned illumination beams for simultaneous multi-color TIRF imaging The Next Level of TIRF Microscopy Mario Faretta,
More informationIC 2 S High Performance Objectives
M i c r o s c o p y f r o m C a r l Z e i s s IC 2 S igh Performance Objectives for Biomedical Applications with Laser Based Imaging Systems LSM,, ConfoCor, TIRF and ELYRA Carl Zeiss offers a large range
More informationMicroscopic Structures
Microscopic Structures Image Analysis Metal, 3D Image (Red-Green) The microscopic methods range from dark field / bright field microscopy through polarisation- and inverse microscopy to techniques like
More informationMIF ZEISS LSM510 CONFOCAL USER PROTOCOL
MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and
More informationInformation & Instructions
KEY FEATURES 1. USB 3.0 For the Fastest Transfer Rates Up to 10X faster than regular USB 2.0 connections (also USB 2.0 compatible) 2. High Resolution 4.2 MegaPixels resolution gives accurate profile measurements
More informationNIH Public Access Author Manuscript Opt Lett. Author manuscript; available in PMC 2010 August 9.
NIH Public Access Author Manuscript Published in final edited form as: Opt Lett. 2010 January 1; 35(1): 67 69. Autoconfocal transmission microscopy based on two-photon induced photocurrent of Si photodiodes
More informationa) How big will that physical image of the cells be your camera sensor?
1. Consider a regular wide-field microscope set up with a 60x, NA = 1.4 objective and a monochromatic digital camera with 8 um pixels, properly positioned in the primary image plane. This microscope is
More informationMAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION
Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION
More informationDevelopment of a Next-Generation Laser-Scanner System for Life Science Research
Development of a Next-Generation Laser-Scanner System for Life Science Research Masaki TAKAMATSU* Yasutake TANAKA* Takashi KOBAYASHI* Hiromi ISHIKAWA* and Akira YAMAGUCHI* Abstract We developed a next-generation
More informationSupplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots
Supplementary Information Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots Bin Dong 1,, Xiaochen Yang 2,, Shaobin Zhu 1, Diane C.
More informationTraining Guide for Leica SP8 Confocal/Multiphoton Microscope
Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence
More informationAn 8-Channel Parallel Multispectral TCSPC FLIM System
An 8-Channel Parallel Multispectral TCSPC FLIM System Abstract. We describe a TCSPC FLIM system that uses 8 parallel TCSPC channels to record FLIM data at a peak count rate on the order of 50 10 6 s -1.
More informationMegapixel FLIM with bh TCSPC Modules
Megapixel FLIM with bh TCSPC Modules The New SPCM 64-bit Software Abstract: Becker & Hickl have recently introduced version 9.60 of their SPCM TCSPC data acquisition software. SPCM version 9.60 not only
More informationOPERATING INSTRUCTIONS
Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms
More informationTraining Guide for Carl Zeiss LSM 510 META Confocal Microscope
Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches
More informationNikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON
N-SIM guide NIKON IMAGING CENTRE @ KING S COLLEGE LONDON Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least
More informationNikon C1si Spectral Laser Scanning Confocal Microscope. User Guide
Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for
More informationOpterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity
Opterra II Multipoint Scanning Confocal Microscope Enabling 4D Live-Cell Fluorescence Imaging through Speed, Sensitivity, Viability and Simplicity Innovation with Integrity Fluorescence Microscopy The
More informationReflecting optical system to increase signal intensity. in confocal microscopy
Reflecting optical system to increase signal intensity in confocal microscopy DongKyun Kang *, JungWoo Seo, DaeGab Gweon Nano Opto Mechatronics Laboratory, Dept. of Mechanical Engineering, Korea Advanced
More informationOperating Instructions for Zeiss LSM 510
Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users
More informationLSM 800 Confocal Microscope Standard Operation Protocol
LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power
More informationCell Biology and Bioimaging Core
Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account
More informationThe DCS-120 Confocal Scanning FLIM System
he DCS-120 Confocal Scanning FLIM System he bh DCS-120 confocal scanning FLIM system converts a conventional microscope into a high-performance fluorescence lifetime imaging system. he system is based
More informationZEISS LSM510META confocal manual
ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including
More informationLSM 510 Meta Training Notes
LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.
More informationSHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014
CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1
More informationDCS-120. Confocal Scanning FLIM Systems. Based on bh s Multidimensional Megapixel FLIM Technology
Based on bh s Multidimensional Megapixel FLIM Technology Complete Laser Scanning FLIM Microscopes FLIM Upgrades for Existing Conventional Microscopes Multidimensional TCSPC technique High throughput dual-channel
More informationZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL
ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE
More information