D2.1 Operating 2D STED Microscope

Size: px
Start display at page:

Download "D2.1 Operating 2D STED Microscope"

Transcription

1 D2.1 Operating 2D STED Microscope Nature: Report Dissemination Level: Public Lead Beneficiary: UNIVDUN Author(s): Piotr Zdankowski Work Package: WP2 Task: ESR5 Version: 0.02 Last modified: 24/04/2017 Status: Complete Approved by: Kees Weijer Date: 24/04/2017 <delete the text below if the Dissemination Level is not Public> Disclaimer: The material contained in this document is provided for information purposes only. No warranty is given in relation to use that may be made of it and neither the copyright owners or the European Commission accept any liability for loss or damage to a third party arising from such use. Copyright Notice: Copyright PHOQUS Consortium All rights reserved. Page 1 of 16 Version: 0.02 Status: Released

2 Table of Contents 1 Introduction STED Microscopy STED Microscope design Microscope optical design Depletion laser beam path Excitation laser beam path: Pinhole size calculations Microscope control software Results Evaluation of the microscope performance Gold beads measurements Laser pulse measurements Fluorescent beads measurements Conclusions Page 2 of 16 Version: 0.02 Status: Released

3 1 Introduction Fluorescence microscopy has been a well-established and extremely useful technique to image biological structures. However, its resolution has always been restricted by the diffraction limit. The introduction of super-resolution microscopy techniques [1] that emerged at the turn of the last century allowed us to see into nanostructures and step into the world of nanoscopy. Among the various super-resolution techniques, Stimulated Emission Depletion (STED) Microscopy [2] has been especially interesting as it relies solely on optical phenomena and does not require any additional computational processing. STED has proven to be able to resolve structures as small as 20nm [3]. To date, there are a few companies that offer a commercial STED microscopes, but their price tag is extremely high and they do not offer the flexibility of a home-built system. What is more, imaging through thick tissues is a very challenging task as every biological sample introduces aberrations and scatters propagating light. This problem becomes notably more significant in the STED microscopy as even small amount of aberrations can degrade the depletion beam shape substantially [4]. Lastly, STED microscopy relies on very high depletion beam powers which can cause significant photodamaging of samples and it is important to minimise the exposition of the sample to the STED beam [5]. This can be achieved using fast resonant scanning over the sample. Our approach was to develop a custom-built system with adaptive optics (AO) correction of the depletion beam to overcome the issues with commercial systems and which can be tailored to the end applications. 1.1 STED Microscopy STED microscope is a type of confocal scanning fluorescence microscope which allows the acquisition of images with a resolution well beyond the Abbe s resolution limit [6], which is dependent on the wavelength of the imaging light and the numerical aperture (NA) of the imaging optics. The principle of operation of the fluorescent microscope is that first, we have to label the chosen structures of the sample with fluorescent markers. Then, by illuminating them with an appropriate wavelength, we cause the molecules of illuminated structure to change their energetic state from ground to the excited state (increasing their energy). Since the molecules prefer to stay in the ground state, they will start to release their energy in the form of fluorescent photons. This phenomenon is called spontaneous emission. Those photons are later separated from the excitation photons and collected with the photodetector. In order to build an image, we need to scan the sample point by point, collecting fluorescent photons from each point of the structure. The problem appears, when we want to image structures smaller than the resolution limit. Due to the diffraction and limited size of imaging optics, the microscope objective has a confined NA. The light that passes through the objective can never be focused into an infinitely small focal spot meaning that we will collect photons from all the structures that are in the area of the illuminating focal spot. This has the consequence that we will not be able to distinguish objects smaller than the size of that spot. STED microscopy is one solution to this problem. If we shine the light onto the excited molecules we can turn them off (deplete them) by making them release the photons in the form of stimulated emission. Now, if we shape the depleting beam in the form of the donut and superimpose it with the excitation beam, we will effectively be exciting the structures that are in the area of the donut hole, which can be significantly smaller than the area of the focal spot. Scanning the sample with those two superimposed beams enables the acquisition of images with the resolution much higher than classical diffraction limited microscopes [7][8]. Page 3 of 16 Version: 0.02 Status: Released

4 2 STED Microscope design 2.1 Microscope optical design Fig. 1. Schematic of the STED microscope. M1 to M6 correspond to broadband mirrors, SLM stands for Spatial Light modulator, f1 to f11 are lenses, GLP is a Glan Laser Polariser, DM1 and DM2 are dichroic mirrors, GM and RM are galvanometric and resonant scanning mirrors respectively, MO is a microscope objective, EF is a fluorescence emission filter, MPPD and SPAD are photon counting detectors Depletion laser beam path The 766nm depletion laser is a 500ps pulsed diode laser (VisIR STED PicoQuant), specially designed for the purpose of STED microscopy. It has a beam diameter of 2.2mm. Then it passes through the half-wave plate /2 to rotate its polarisation. Lenses f1 and f2 are responsible for beam expansion. After expansion beam size is 16mm, so that the beam diameter overfills the Spatial Light Modulator (SLM) aperture. After the expansion, the depletion beam passes through the half-wave plate which rotates the polarisation in order to maximise the phase modulation of the SLM. The SLM displays an appropriate phase mask to shape the depletion beam. This phase mask is a circular aperture with a blazed grating, which is necessary, since it separates the diffraction orders after the beam reflects from the SLM. The modulated phase mask is represented in the 1 st diffraction order of the blazed grating. The size of the aperture is set to 557 pixels which equating to a 11mm beam size diameter Page 4 of 16 Version: 0.02 Status: Released

5 (SLM pixel pitch is equal to 20 m). After the SLM, the depletion beam is decreased to 1.83mm by a 4f system and passes through the Glan-Laser polariser (GLP), that only transmits polarisation corresponding to the SLM polarisation. The fast axis of the GLP is set so that the 1 st diffraction order is maximised. DM1 is a fluorescence dichroic mirror (Semrock FF720- SDi01), which reflects the STED beam and transmits fluorescence. Further on, the STED beam passes through the dichroic mirror DM2 (Chroma ZT647rdc-UF3), where it is combined with the excitation beam. There are three 4f systems in the beam path: 4f1: f6 (f=400mm) and f7 (f=40mm), imaging the SLM plane containing the hologram onto the plane in-between two galvanometric mirrors GM 4f2: f8(f=80mm) and f9(f=100mm) imaging the RM plane onto resonant mirror RM plane 4f3: f10(f=50mm) and f11(400mm) imaging RM plane onto the back focal plane of the microscope objective MO (100x Nikon Plan Apo, NA=1.45) Excitation laser beam path: The excitation beam laser is a diode pulsed laser with the wavelength of 637nm and is coupled into polarisation maintaining fibre. At the output of the fibre it passes through the half-wave plate /2, which rotates the polarisation to match the STED beam polarisation and then is collimated by the aspheric lens collimator f3 (f=10.99mm). After f3 the beam size is 2.1mm. Lenses f4 (f=60mm) and f5 (f=200mm) expand the excitation beam to the size of 7mm. After the beam expander the excitation beam is reflected from DM2, where it is combined with the STED beam. After DM2 both the STED and excitation beams propagate coaxially. Both beams then pass through two 4f systems (4f1 and 4f2) before passing through the quarter-wave plate /4, which converts the beam polarisation to circular. Then the beams go through the MO and are focused on the sample. The excitation beam that goes into the MO should overfill the pupil diameter, however size of the phase mask image of the depletion beam should be exactly matching the pupil diameter of MO. The pupil diameter can be calculated from the following formula: D = 2NA f For the applied Nikon microscope objective, that has NA = 1.45 and f = 2mm, the pupil diameter is equal D = 5.8mm. Depending on the sample type that is imaged, there are two different possible imaging paths: a reflective path and a fluorescent path. The imaging beam propagates back through the same beam path as the excitation/sted beam path, until it reaches: a) a pellicle beamsplitter P1 which reflects the imaging beam onto the imaging lens FiR (f=80mm), which focuses it through the pinhole PHR and falls onto the photon counting detector MPPC; b) a dichroic mirror DM1, which transmits the fluorescent beam, then focusses it through the imaging lens FiF (f=80mm), passing through the set of 3 filters: EF1 (Semrock FF01-750/SP), EF2 (Semrock FF01-676/37) and EF3 (Semrock FF650/100). After the filters the beam passes through the pinhole PHF and falls on the photon counting detector SPAD Pinhole size calculations. An Airy disc has a radius according to the Abbe s formula of: Page 5 of 16 Version: 0.02 Status: Released

6 r = 0.61λ nm = = 268nm NA 1.45 This means that for the excitation beam wavelength 637nm, the Airy disc radius size is equal to r=268nm. Taking into account the magnification of the whole setup, the Airy disc radius size in the fluorescent pinhole plane PHF is equal to: r F = f11 f MO f9 f10 f7 f8 FiF f6 = 10.72μm Similarly, the Airy disc size for imaging reflective objects PHR is equal to: r R = f11 f MO f9 f10 f7 f8 FiR f6 = 10.72μm The chosen pinhole size for fluorescent imaging is 1A.U. (Airy units), which means that the actual pinhole size is 20 m. For reflective objects, the chosen pinhole size is approx. 20A.U, for which the actual pinhole size is 400 m, which allows the system to have quasi-widefield imaging. 2.2 Microscope control software The software used for the STED microscope control has been written in Labview. The acquisition and scanning is carried out using a National Instruments FPGA card. The STED microscope uses a 16kHz resonant mirror for x-axis scanning and a slower, galvanometric mirror for y-axis scanning. 16kHz scanning means that the mirror does a return scan in 62.5 s meaning that one line is scanned in half of the period, s. The FPGA can acquire digital input data with 80MHz clock frequency (1 clock tick is equal to 12.5ns), meaning there are 2500 ticks available for the photon counting. The number of pixels per line (ppl) determines the number of photon counts per pixel that can be acquired by the detector in the resonant mirror scan time (e.g. for 250ppl, the maximum number of counts is 10, for 500ppl the maximum number of counts is 5). The acquisition software has the option of multiple line scans, so that the count number in the image can be increased. Page 6 of 16 Version: 0.02 Status: Released

7 3 Results 3.1 Evaluation of the microscope performance Gold beads measurements. In order to measure the performance of both excitation beam quality and STED beam quality it is best to image the beams and calculate their point spread functions (PSF). It is not possible to measure both beams using fluorescent detection, since there are no such beads that would be excited at both wavelengths (637nm and 766nm). Hence the measurements of the PSF for both beams are carried out by imaging gold beads and collecting their reflected signal. Firstly, the PSF of the microscope has been simulated using the following parameters: Pixel size: x = 26nm, y = 26nm, z = 30nm. Number of z stacks: 100 NA = 1.45 Tables 1-2 contain the calculated full width at half maximum (FWHM) of the simulated beads of different pixel sizes and for different wavelengths. a) b) d) e) f) c) Fig x256 pixels simulation of the PSF for 637nm using Born&Wolf algorithm [9]. a) presents xy plane of the PSF, b) yz profile, c) xz profile; d) x profile and Gaussian fitting, e) y profile and Gaussian fitting, f) z profile and Gaussian fitting. Table 1. Calculated FWHM for the simulated PSF shown in fig. 2. Number of Pixels Pixel size FWHM x nm μm y nm μm z nm μm Page 7 of 16 Version: 0.02 Status: Released

8 a) b) d) e) f) c) Fig x256 pixels simulation of the PSF for 766nm using Born&Wolf algorithm. a) presents xy plane of the PSF, b) yz profile, c) xz profile; d) x profile and Gaussian fitting, e) y profile and Gaussian fitting, f) z profile and Gaussian fitting. Table 2. Calculated FWHM for the simulated PSF shown in fig. 3. Number of Pixels Pixel size FWHM x nm μm y nm μm z nm μm Having the simulated PSF for the experimental microscope configuration the next step is experimental verification carried out by imaging gold beads. 150nm gold nanoparticles from BBI solutions were diluted in ethanol and spread onto the coverslip, which was then mounted with a 97% solution of Thiodiethanol on the microscope slide. Fig. 4-5 shows the obtained PSF for both excitation beam and STED beam. Tables 3 and 4 shown the FWHM values calculated for both beams. Gold beads were recorded for two different pixel sizes and total number of pixels. The field of view was 5 m x 5 m. a) b) d) e) f) c) Fig x256 pixels PSF of 150nm gold bead image recorded with excitation laser. a) presents xy plane of the PSF, b) yz profile, c) xz profile; d) x profile and Gaussian fitting, e) y profile and Gaussian fitting, f) z profile and Gaussian fitting. Page 8 of 16 Version: 0.02 Status: Released

9 Table 3. Calculated FWHM for the recorded PSF of 256x256 pixels 150nm gold bead recorded with excitation laser. Number of pixels/stacks Pixel size FWHM x nm μm y nm μm z nm μm a) b) d) e) f) c) Fig x256pixels PSF of 150nm gold bead image recorded with STED laser. a) presents xy plane of the PSF, b) yz profile, c) xz profile; d) x profile and Gaussian fitting, e) y profile and Gaussian fitting, f) z profile and Gaussian fitting. Table 4. Calculated FWHM for the recorded PSF of 256x256 pixels 150nm gold bead recorded with STED laser. Number of pixels/stacks Pixel size FWHM x nm y nm z nm Imaging gold beads also helps to image and evaluate the quality of the phase mask applied on the SLM as well as spatial alignment of both STED and excitation beams. Imaging the phase mask is very important, as aberrations can significantly distort the Laguerre-Gaussian beam. Corrected and aligned beams with different phase masks are shown on fig Page 9 of 16 Version: 0.02 Status: Released

10 a) b) Fig. 6. STED beam phase mask imaged with the 150nm gold beads. a) vortex phase mask creating Laguerre-Gaussian beam creating 2D STED beam (donut beam); b) π-step phase mask creating a 3D STED beam a) b) Fig. 7. STED beam phase mask imaged with the 150nm gold beads merged with the gold bead recorded with the excitation beam. This shows spatial alignment of the excitation and STED beams. a) vortex phase mask creating Laguerre-Gaussian beam creating 2D STED beam (donut beam); b) π-step phase mask creating a 3D STED beam Laser pulse measurements With good spatial alignment of the STED and excitation beams, the next step is to check the temporal alignment of the pulses of both lasers. In order to do that, pulses were measured with a 2GHz photodiode (DET025A/M) and a fast 2GHz oscilloscope. Pulses can be finely tuned using the electronic pulse delay unit (MPD Picosecond Delayer). Results of the pulse measurements are shown in fig. 8. Page 10 of 16 Version: 0.02 Status: Released

11 Fig. 8. Pulse measurements for excitation beam and STED beam. Pulses had to be measured separately for each laser, since the time resolution of the diode was not good enough to be able to distinguish both pulses simultaneously. The STED pulse should arrive approx. 150ps after the excitation pulse. Measurements shown in fig. 8 show both pulses measured for the delay set to 2020ps, which makes the STED beam arrive ~160ps after the excitation pulse. The pulses are set to 40MHz repetition rate Fluorescent beads measurements Fig. 9 shows images of fluorescent beads acquired in the confocal mode and for both 2D and 3D STED phase mask. There is a significant increase in the resolution of the STED images well below the diffraction limit. The beads are 23nm GattaBeads filled with Atto647n dye. FWHM was calculated and is shown in tables 5-7 for all 3 cases: no STED (table 5) phase mask, 2D phase mask (table 6) and 3D phase mask (table 7). Fig. 9. Comparison of the images with a) no STED beam, b) 2D STED phase mask, c) 3D STED phase mask. 512x512 images, pixel size 19nmx19nm, 40MHz laser repetition rate Page 11 of 16 Version: 0.02 Status: Released

12 250nm 620nm 55nm 600nm 100nm 165nm Fig. 10. Comparison of the confocal (a), 2D STED (b) and 3D STED (c) images and their orthogonal views. Images on the left show acquired bead images, images in the right show zoomed bead shown in the blue frame and its orthogonal views YZ (right) and XZ (bottom). Page 12 of 16 Version: 0.02 Status: Released

13 Table 5. Full width at half maximum of the confocal image obtained for the beads shown in fig. 9. Bead Number Average FWHM X [nm] FWHM Y [nm] FWHM XZ [nm] FWHM YZ [nm] Table 6. Full width at half maximum of the 2D STED image obtained for the beads shown in fig. 9. Bead Number Average FWHM X [nm] FWHM Y [nm] Table 7. Full width at half maximum of the 3D STED image obtained for the beads shown in fig. 9. Bead Number Average FWHM X [nm] FWHM Y [nm] FWHM XZ [nm] FWHM YZ [nm] Experiments to demonstrate the STED microscope usability on the biological samples were carried out. To show the increased resolution achieved by STED microscopy microtubule networks in He-La and RPE-1 cells were investigated. In both samples, α-tubulin was stained with the Atto647n anti-mouse secondary antibody. Resulting images are shown in fig. 11 and 12 (microtubules) and fig. 13 (mitotic cell). Fig. 11. He-La cell microtubules stained with ATTO647N dye. Left confocal image, right 2D STED image. Page 13 of 16 Version: 0.02 Status: Released

14 Fig. 12. RPE-1 cell microtubules stained with ATTO647N dye. Left confocal image, right 2D STED image. Fig. 13. RPE-1 mitotic cell stained with ATTO647N dye. Left confocal image, right 2D STED image. Page 14 of 16 Version: 0.02 Status: Released

15 4 Conclusions. In the previous sections, we presented clear evidence that we have designed and built a fully functional custom built STED microscope. The performance of the microscope in terms of 2 and 3D resolution is comparable to that of available commercial systems, but it has the huge advantage of the customisation possibilities. The microscope can be used with widely available fluorescent dyes. By selecting an appropriate phase mask on the SLM it can acquire 2D and 3D super-resolution images. The achievable lateral resolution is 50nm and the axial resolution is 165nm. The microscope has been used to acquire super-resolution images of microtubule networks in fixed interphase and mitotic cells. Page 15 of 16 Version: 0.02 Status: Released

16 5 References [1] E. Betzig, S. W. Hell, and W. E. Moerner, How the optical microscope became a nanoscope, pp. 1 7, [2] S. W. Hell and J. Wichmann, Breaking the diffraction resolution limit by stimulated emission: stimulated-emission-depletion fluorescence microscopy., Opt. Lett., vol. 19, no. 11, pp , Jun [3] F. Göttfert, C. a Wurm, V. Mueller, S. Berning, V. C. Cordes, A. Honigmann, and S. W. Hell, Coaligned dual-channel STED nanoscopy and molecular diffusion analysis at 20 nm resolution., Biophys. J., vol. 105, no. 1, pp. L01-3, Jul [4] B. R. Patton, D. Burke, R. Vrees, and M. J. Booth, Is phase-mask alignment aberrating your STED microscope?, Methods Appl. Fluoresc., vol. 3, no. 2, p , [5] Y. Wu, Reducing photobleaching in STED microscopy with higher scanning speed, arxiv Prepr. arxiv , [6] E. Abbe, Beiträge zur Theorie des Mikroskops und der mikroskopischen Wahrnehmung, Arch. für Mikroskopische Anat., vol. 9, no. 1, pp , Dec [7] J. B. Pawley, Handbook Of Biological Confocal Microscopy. Boston, MA: Springer US, [8] a Cornea and P. Conn, Fluorescence Microscopy: Super-resolution and Other Novel Techniques. Academic Press, [9] B. Richards and E. Wolf, Electromagnetic Diffraction in Optical Systems. II. Structure of the Image Field in an Aplanatic System, Proc. R. Soc. London. Ser. A. Math. Phys. Sci., vol. 253, no. 1274, p. 358 LP-379, Dec Page 16 of 16 Version: 0.02 Status: Released

Development of a High-speed Super-resolution Confocal Scanner

Development of a High-speed Super-resolution Confocal Scanner Development of a High-speed Super-resolution Confocal Scanner Takuya Azuma *1 Takayuki Kei *1 Super-resolution microscopy techniques that overcome the spatial resolution limit of conventional light microscopy

More information

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy Bi177 Lecture 5 Adding the Third Dimension Wide-field Imaging Point Spread Function Deconvolution Confocal Laser Scanning Microscopy Confocal Aperture Optical aberrations Alternative Scanning Microscopy

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

Introduction to light microscopy

Introduction to light microscopy Center for Microscopy and Image Anaylsis Introduction to light microscopy Basic concepts of imaging with light Urs Ziegler ziegler@zmb.uzh.ch Light interacting with matter Absorbtion Refraction Diffraction

More information

TCSPC at Wavelengths from 900 nm to 1700 nm

TCSPC at Wavelengths from 900 nm to 1700 nm TCSPC at Wavelengths from 900 nm to 1700 nm We describe picosecond time-resolved optical signal recording in the spectral range from 900 nm to 1700 nm. The system consists of an id Quantique id220 InGaAs

More information

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup.

Nature Methods: doi: /nmeth Supplementary Figure 1. Schematic of 2P-ISIM AO optical setup. Supplementary Figure 1 Schematic of 2P-ISIM AO optical setup. Excitation from a femtosecond laser is passed through intensity control and shuttering optics (1/2 λ wave plate, polarizing beam splitting

More information

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots

Supplementary Information. Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots Supplementary Information Stochastic Optical Reconstruction Microscopy Imaging of Microtubule Arrays in Intact Arabidopsis thaliana Seedling Roots Bin Dong 1,, Xiaochen Yang 2,, Shaobin Zhu 1, Diane C.

More information

Shaping light in microscopy:

Shaping light in microscopy: Shaping light in microscopy: Adaptive optical methods and nonconventional beam shapes for enhanced imaging Martí Duocastella planet detector detector sample sample Aberrated wavefront Beamsplitter Adaptive

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis Center for Microscopy and Image Analysis Bio 407 Applied Introduction into light José María Mateos Fundamentals of light Compound microscope Microscope composed of an objective and an additional lens (eyepiece,

More information

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Qiyuan Song (M2) and Aoi Nakamura (B4) Abstracts: We theoretically and experimentally

More information

EUV microscopy - a user s perspective Dimitri Scholz EUV,

EUV microscopy - a user s perspective Dimitri Scholz EUV, EUV microscopy - a user s perspective Dimitri Scholz EUV, 09.11.2011 Imaging technologies: available at UCD now and in the next future Begin ab ovo - Simple approaches direct to the goal - Standard methods

More information

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging.

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging. Supplementary Figure 1 Optimized Bessel foci for in vivo volume imaging. (a) Images taken by scanning Bessel foci of various NAs, lateral and axial FWHMs: (Left panels) in vivo volume images of YFP + neurites

More information

5/4/2015 INTRODUCTION TO LIGHT MICROSCOPY. Urs Ziegler MICROSCOPY WITH LIGHT. Image formation in a nutshell. Overview of techniques

5/4/2015 INTRODUCTION TO LIGHT MICROSCOPY. Urs Ziegler MICROSCOPY WITH LIGHT. Image formation in a nutshell. Overview of techniques INTRODUCTION TO LIGHT MICROSCOPY Urs Ziegler ziegler@zmb.uzh.ch MICROSCOPY WITH LIGHT INTRODUCTION TO LIGHT MICROSCOPY Image formation in a nutshell Overview of techniques Widefield microscopy Resolution

More information

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging Nonlinear microscopy I: Two-photon fluorescence microscopy Multiphoton Microscopy What is multiphoton imaging? Applications Different imaging modes Advantages/disadvantages Scattering of light in thick

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 1 1 2! NA = 0.5! NA 2D imaging

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN Detector

Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN Detector Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN 392-1000 Detector Abstract: We present a wide-field TCSPC FLIM system consisting of a position-sensitive MCP PMT of the delay-line type,

More information

Introduction to light microscopy

Introduction to light microscopy Center for Microscopy and Image Anaylsis Introduction to light Basic concepts of imaging with light Urs Ziegler ziegler@zmb.uzh.ch Microscopy with light 1 Light interacting with matter Absorbtion Refraction

More information

Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets

Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets SUPPLEMENTARY MATERIAL Multicolor 4D Fluorescence Microscopy using Ultrathin Bessel Light sheets Teng Zhao, Sze Cheung Lau, Ying Wang, Yumian Su, Hao Wang, Aifang Cheng, Karl Herrup, Nancy Y. Ip, Shengwang

More information

The DCS-120 Confocal Scanning FLIM System

The DCS-120 Confocal Scanning FLIM System he DCS-120 Confocal Scanning FLIM System he bh DCS-120 confocal scanning FLIM system converts a conventional microscope into a high-performance fluorescence lifetime imaging system. he system is based

More information

Single-photon excitation of morphology dependent resonance

Single-photon excitation of morphology dependent resonance Single-photon excitation of morphology dependent resonance 3.1 Introduction The examination of morphology dependent resonance (MDR) has been of considerable importance to many fields in optical science.

More information

Heisenberg) relation applied to space and transverse wavevector

Heisenberg) relation applied to space and transverse wavevector 2. Optical Microscopy 2.1 Principles A microscope is in principle nothing else than a simple lens system for magnifying small objects. The first lens, called the objective, has a short focal length (a

More information

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM INTRODUCTION TO MICROSCOPY Urs Ziegler ziegler@zmb.uzh.ch THE PROBLEM 1 ORGANISMS ARE LARGE LIGHT AND ELECTRONS: ELECTROMAGNETIC WAVES v = Wavelength ( ) Speed (v) Frequency ( ) Amplitude (A) Propagation

More information

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D Microscopy Outline 1. Resolution and Simple Optical Microscope 2. Contrast enhancement: Dark field, Fluorescence (Chelsea & Peter), Phase Contrast, DIC 3. Newer Methods: Scanning Tunneling microscopy (STM),

More information

Dynamic beam shaping with programmable diffractive optics

Dynamic beam shaping with programmable diffractive optics Dynamic beam shaping with programmable diffractive optics Bosanta R. Boruah Dept. of Physics, GU Page 1 Outline of the talk Introduction Holography Programmable diffractive optics Laser scanning confocal

More information

Supplementary Figure S1: Schematic view of the confocal laser scanning STED microscope used for STED-RICS. For a detailed description of our

Supplementary Figure S1: Schematic view of the confocal laser scanning STED microscope used for STED-RICS. For a detailed description of our Supplementary Figure S1: Schematic view of the confocal laser scanning STED microscope used for STED-RICS. For a detailed description of our home-built STED microscope used for the STED-RICS experiments,

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

Flatness of Dichroic Beamsplitters Affects Focus and Image Quality

Flatness of Dichroic Beamsplitters Affects Focus and Image Quality Flatness of Dichroic Beamsplitters Affects Focus and Image Quality Flatness of Dichroic Beamsplitters Affects Focus and Image Quality 1. Introduction Even though fluorescence microscopy has become a routine

More information

Confocal Microscopy and Related Techniques

Confocal Microscopy and Related Techniques Confocal Microscopy and Related Techniques Chau-Hwang Lee Associate Research Fellow Research Center for Applied Sciences, Academia Sinica 128 Sec. 2, Academia Rd., Nankang, Taipei 11529, Taiwan E-mail:

More information

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember Günter Toesko - Laserseminar BLZ im Dezember 2009 1 Aberrations An optical aberration is a distortion in the image formed by an optical system compared to the original. It can arise for a number of reasons

More information

BASICS OF CONFOCAL IMAGING (PART I)

BASICS OF CONFOCAL IMAGING (PART I) BASICS OF CONFOCAL IMAGING (PART I) INTERNAL COURSE 2012 LIGHT MICROSCOPY Lateral resolution Transmission Fluorescence d min 1.22 NA obj NA cond 0 0 rairy 0.61 NAobj Ernst Abbe Lord Rayleigh Depth of field

More information

Confocal Microscopy. Kristin Jensen

Confocal Microscopy. Kristin Jensen Confocal Microscopy Kristin Jensen 17.11.05 References Cell Biological Applications of Confocal Microscopy, Brian Matsumoto, chapter 1 Studying protein dynamics in living cells,, Jennifer Lippincott-Schwartz

More information

Aberrations and adaptive optics for biomedical microscopes

Aberrations and adaptive optics for biomedical microscopes Aberrations and adaptive optics for biomedical microscopes Martin Booth Department of Engineering Science And Centre for Neural Circuits and Behaviour University of Oxford Outline Rays, wave fronts and

More information

Administrative details:

Administrative details: Administrative details: Anything from your side? www.photonics.ethz.ch 1 What are we actually doing here? Optical imaging: Focusing by a lens Angular spectrum Paraxial approximation Gaussian beams Method

More information

More fancy SPIM, Even fancier SPIM

More fancy SPIM, Even fancier SPIM More fancy SPIM, Even fancier SPIM Last class Light sheet microscopy Fancy SPIM (ispim, dspim, etc ) This class Multi camera SPIM SIM SPIM Bessels d x,y = λ em 2 NA d z = 2 NA λ ex + n(1 cosθ λ em 1 IsoView

More information

IR Antibunching Measurements with id201 InGaAs Gated SPAD Detectors

IR Antibunching Measurements with id201 InGaAs Gated SPAD Detectors IR Antibunching Measurements with id201 GaAs Gated SPAD Detectors Abstract. Antibunching measurements with GaAs SPAD detectors are faced with the problems of high background count rate, afterpulsing, and

More information

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Finite-difference time-domain calculations of the optical transmittance through

More information

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON N-SIM guide NIKON IMAGING CENTRE @ KING S COLLEGE LONDON Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least

More information

Reflecting optical system to increase signal intensity. in confocal microscopy

Reflecting optical system to increase signal intensity. in confocal microscopy Reflecting optical system to increase signal intensity in confocal microscopy DongKyun Kang *, JungWoo Seo, DaeGab Gweon Nano Opto Mechatronics Laboratory, Dept. of Mechanical Engineering, Korea Advanced

More information

Femtosecond laser microfabrication in. Prof. Dr. Cleber R. Mendonca

Femtosecond laser microfabrication in. Prof. Dr. Cleber R. Mendonca Femtosecond laser microfabrication in polymers Prof. Dr. Cleber R. Mendonca laser microfabrication focus laser beam on material s surface laser microfabrication laser microfabrication laser microfabrication

More information

Optics and Lasers. Matt Young. Including Fibers and Optical Waveguides

Optics and Lasers. Matt Young. Including Fibers and Optical Waveguides Matt Young Optics and Lasers Including Fibers and Optical Waveguides Fourth Revised Edition With 188 Figures Springer-Verlag Berlin Heidelberg New York London Paris Tokyo Hong Kong Barcelona Budapest Contents

More information

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement CONFOCAL MICROSCOPY BioVis Uppsala, 2017 Jeremy Adler Matyas Molnar Dirk Pacholsky Widefield & Confocal Microscopy

More information

Dynamic closed-loop system for focus tracking using a spatial light modulator and a deformable membrane mirror

Dynamic closed-loop system for focus tracking using a spatial light modulator and a deformable membrane mirror Dynamic closed-loop system for focus tracking using a spatial light modulator and a deformable membrane mirror Amanda J. Wright, Brett A. Patterson, Simon P. Poland, John M. Girkin Institute of Photonics,

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 2D imaging 3D imaging Resolution

More information

Technology Note ZEISS LSM 880 with Airyscan

Technology Note ZEISS LSM 880 with Airyscan Technology Note ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode ZEISS LSM 880 with Airyscan Introducing the Fast Acquisition Mode Author: Dr. Annette Bergter Carl Zeiss Microscopy GmbH,

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Optically reconfigurable metasurfaces and photonic devices based on phase change materials S1: Schematic diagram of the experimental setup. A Ti-Sapphire femtosecond laser (Coherent Chameleon Vision S)

More information

Practical Flatness Tech Note

Practical Flatness Tech Note Practical Flatness Tech Note Understanding Laser Dichroic Performance BrightLine laser dichroic beamsplitters set a new standard for super-resolution microscopy with λ/10 flatness per inch, P-V. We ll

More information

Rapid Non linear Image Scanning Microscopy, Supplementary Notes

Rapid Non linear Image Scanning Microscopy, Supplementary Notes Rapid Non linear Image Scanning Microscopy, Supplementary Notes Calculation of theoretical PSFs We calculated the electrical field distribution using the wave optical theory developed by Wolf 1, and Richards

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope

Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope Summary of the PhD thesis Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope Author: Dudás, László Supervisors: Prof. Dr. Szabó, Gábor and Dr.

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

LOPUT Laser: A novel concept to realize single longitudinal mode laser

LOPUT Laser: A novel concept to realize single longitudinal mode laser PRAMANA c Indian Academy of Sciences Vol. 82, No. 2 journal of February 2014 physics pp. 185 190 LOPUT Laser: A novel concept to realize single longitudinal mode laser JGEORGE, KSBINDRAand SMOAK Solid

More information

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design Computer Aided Design Several CAD tools use Ray Tracing (see

More information

PHY 431 Homework Set #5 Due Nov. 20 at the start of class

PHY 431 Homework Set #5 Due Nov. 20 at the start of class PHY 431 Homework Set #5 Due Nov. 0 at the start of class 1) Newton s rings (10%) The radius of curvature of the convex surface of a plano-convex lens is 30 cm. The lens is placed with its convex side down

More information

Invitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner

Invitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner Invitation for a walk through microscopy Sebastian Schuchmann Jörg Rösner joerg.roesner@charite.de Techniques in microscopy Conventional (light) microscopy bright & dark field, phase & interference contrast

More information

Travel to New Dimensions- LSM 880. The Resolution of a Microscope is limited. The Resolution of a Microscope is limited. Image. Image. Object.

Travel to New Dimensions- LSM 880. The Resolution of a Microscope is limited. The Resolution of a Microscope is limited. Image. Image. Object. Travel to New Dimensions- LSM 880 LSM 880: The Power of Sensitivity Our Latest Member of the LSM 880 with GaAsP Detectors Sensitivity, and Ease of Use Innovative High-End Laser Scanning Microscopes from

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Introduction to light microscopy

Introduction to light microscopy Center for Microscopy and Image Anaylsis Introduction to light Imaging with light / Overview of techniques Urs Ziegler ziegler@zmb.uzh.ch Light interacting with matter Absorbtion Refraction Diffraction

More information

Micro- and Nano-Technology... for Optics

Micro- and Nano-Technology... for Optics Micro- and Nano-Technology...... for Optics 3.2 Lithography U.D. Zeitner Fraunhofer Institut für Angewandte Optik und Feinmechanik Jena Printing on Stones Map of Munich Stone Print Contact Printing light

More information

Design Description Document

Design Description Document UNIVERSITY OF ROCHESTER Design Description Document Flat Output Backlit Strobe Dare Bodington, Changchen Chen, Nick Cirucci Customer: Engineers: Advisor committee: Sydor Instruments Dare Bodington, Changchen

More information

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack

Application Note. The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a Z-stack The New 2D Superresolution Mode for ZEISS Airyscan 120 nm Lateral Resolution without Acquiring a

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

Basics of confocal imaging (part I)

Basics of confocal imaging (part I) Basics of confocal imaging (part I) Swiss Institute of Technology (EPFL) Faculty of Life Sciences Head of BIOIMAGING AND OPTICS BIOP arne.seitz@epfl.ch Lateral resolution BioImaging &Optics Platform Light

More information

Shreyash Tandon M.S. III Year

Shreyash Tandon M.S. III Year Shreyash Tandon M.S. III Year 20091015 Confocal microscopy is a powerful tool for generating high-resolution images and 3-D reconstructions of a specimen by using point illumination and a spatial pinhole

More information

Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells

Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells F e a t u r e A r t i c l e Feature Article Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells Yasuhiro Awatsuji The author invented and developed a technique capable

More information

DCS-120. Confocal Scanning FLIM Systems. Based on bh s Multidimensional Megapixel FLIM Technology

DCS-120. Confocal Scanning FLIM Systems. Based on bh s Multidimensional Megapixel FLIM Technology Based on bh s Multidimensional Megapixel FLIM Technology Complete Laser Scanning FLIM Microscopes FLIM Upgrades for Existing Conventional Microscopes Multidimensional TCSPC technique High throughput dual-channel

More information

R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad.

R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad. R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad. DEPARTMENT OF PHYSICS QUESTION BANK FOR SEMESTER III PAPER III OPTICS UNIT I: 1. MATRIX METHODS IN PARAXIAL OPTICS 2. ABERATIONS UNIT II

More information

Rates of excitation, emission, ISC

Rates of excitation, emission, ISC Bi177 Lecture 4 Fluorescence Microscopy Phenomenon of Fluorescence Energy Diagram Rates of excitation, emission, ISC Practical Issues Lighting, Filters More on diffraction Point Spread Functions Thus Far,

More information

Nature Methods: doi: /nmeth Supplementary Figure 1

Nature Methods: doi: /nmeth Supplementary Figure 1 . Supplementary Figure 1 Schematics and characterization of our AO two-photon fluorescence microscope. (a) Essential components of our AO two-photon fluorescence microscope: Ti:Sapphire laser; optional

More information

SETTING UP OF A TOTAL INTERNAL REFLECTION FLUORESCENT MICROSCOPE (TIRFM) SYSTEM: A DETAILED OVERVIEW

SETTING UP OF A TOTAL INTERNAL REFLECTION FLUORESCENT MICROSCOPE (TIRFM) SYSTEM: A DETAILED OVERVIEW PK ISSN 0022-2941; CODEN JNSMAC Vol. 51, (2011) PP 31-45 SETTING UP OF A TOTAL INTERNAL REFLECTION FLUORESCENT MICROSCOPE (TIRFM) SYSTEM: A DETAILED OVERVIEW A. R. KHAN 1 *, S. AKHLAQ 1, M. N. B. ABID

More information

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal

Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Digital Camera Technologies for Scientific Bio-Imaging. Part 2: Sampling and Signal Yashvinder Sabharwal, 1 James Joubert 2 and Deepak Sharma 2 1. Solexis Advisors LLC, Austin, TX, USA 2. Photometrics

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Parallel Mode Confocal System for Wafer Bump Inspection

Parallel Mode Confocal System for Wafer Bump Inspection Parallel Mode Confocal System for Wafer Bump Inspection ECEN5616 Class Project 1 Gao Wenliang wen-liang_gao@agilent.com 1. Introduction In this paper, A parallel-mode High-speed Line-scanning confocal

More information

6/3/15. The Anatomy of a Digital Image. Representative Intensities. Specimen: (molecular distribution)

6/3/15. The Anatomy of a Digital Image. Representative Intensities. Specimen: (molecular distribution) 2015 LMIC Imaging Workshop Sidney L. Shaw Technical Director An introduction of concepts for Super-Resolution Light Microscopy The Anatomy of a Digital Image Representative Intensities Specimen: (molecular

More information

BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP LASER SCANNING CONFOCAL MICROSCOPY PRACTICAL CONSIDERATIONS

BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP LASER SCANNING CONFOCAL MICROSCOPY PRACTICAL CONSIDERATIONS LASER SCANNING CONFOCAL MICROSCOPY PRACTICAL CONSIDERATIONS IMPORTANT PARAMETERS Pixel dwell time Zoom and pixel number PIXEL DWELL TIME How much time signal is collected at every pixel Very small values,

More information

Nikon Instruments Europe

Nikon Instruments Europe Nikon Instruments Europe Recommendations for N-SIM sample preparation and image reconstruction Dear customer, We hope you find the following guidelines useful in order to get the best performance out of

More information

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES Shortly after the experimental confirmation of the wave properties of the electron, it was suggested that the electron could be used to examine objects

More information

Introduction to light microscopy

Introduction to light microscopy Center for Microscopy and Image Anaylsis Introduction to light microscopy (an overview) Microscopy with light Components of a light microscope 1. Light source 2. Objective 3. Sample or specimen holder

More information

Multiphoton Microscopy

Multiphoton Microscopy Multiphoton Microscopy A. Neumann, Y. Kuznetsova Introduction Multi-Photon Fluorescence Microscopy is a relatively novel imaging technique in cell biology. It relies on the quasi-simultaneous absorption

More information

An opening a = λ would put the first minima at θ = 90

An opening a = λ would put the first minima at θ = 90 Microscopy Outline Resolution & definitions Fluorescence microscopy Some other optical microscopy techniques Electron microscopes X-ray microscopy Scanning tunneling microscopes 2 Microscopy history First

More information

Advanced Optical Microscopy lecture. 03. December 2012 Kai Wicker

Advanced Optical Microscopy lecture. 03. December 2012 Kai Wicker Advanced Optical Microscopy lecture 03. December 2012 Kai Wicker Today: Optical transfer functions (OTF) and point spread functions (PSF) in incoherent imaging. 1. Quick revision: the incoherent wide-field

More information

Katarina Logg, Kristofer Bodvard, Mikael Käll. Dept. of Applied Physics. 12 September Optical Microscopy. Supervisor s signature:...

Katarina Logg, Kristofer Bodvard, Mikael Käll. Dept. of Applied Physics. 12 September Optical Microscopy. Supervisor s signature:... Katarina Logg, Kristofer Bodvard, Mikael Käll Dept. of Applied Physics 12 September 2007 O1 Optical Microscopy Name:.. Date:... Supervisor s signature:... Introduction Over the past decades, the number

More information

Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By

Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By imaging FITC-containing solution, the thickness of the light sheet generated by the P3A-DSLM

More information

MOM#3: LIGHT SHEET MICROSCOPY (LSM) Stanley Cohen, MD

MOM#3: LIGHT SHEET MICROSCOPY (LSM) Stanley Cohen, MD MOM#3: LIGHT SHEET MICROSCOPY (LSM) Stanley Cohen, MD Introduction. Although the technical details of light sheet imaging and its various permutations appear at first glance to be complex and require some

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

plasmonic nanoblock pair

plasmonic nanoblock pair Nanostructured potential of optical trapping using a plasmonic nanoblock pair Yoshito Tanaka, Shogo Kaneda and Keiji Sasaki* Research Institute for Electronic Science, Hokkaido University, Sapporo 1-2,

More information

Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces. Ali Mahmoudi

Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces. Ali Mahmoudi 1 Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces Ali Mahmoudi a.mahmoudi@qom.ac.ir & amahmodi@yahoo.com Laboratory of Optical Microscopy,

More information

Precision-tracking of individual particles By Fluorescence Photo activation Localization Microscopy(FPALM) Presented by Aung K.

Precision-tracking of individual particles By Fluorescence Photo activation Localization Microscopy(FPALM) Presented by Aung K. Precision-tracking of individual particles By Fluorescence Photo activation Localization Microscopy(FPALM) Presented by Aung K. Soe This FPALM research was done by Assistant Professor Sam Hess, physics

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006

ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006 ADVANCED METHODS FOR CONFOCAL MICROSCOPY II Jean-Yves Chatton Sept. 2006 Workshop outline Confocal microscopy of living cells and tissues X-Z scanning Time series Bleach: FRAP, photoactivation Emission

More information

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, KTH Applied Physics Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy, 2009-06-05, 8-13, FB51 Allowed aids: Compendium Imaging Physics (handed out) Compendium Light Microscopy

More information

Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope. Advanced Lab Course University of Basel

Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope. Advanced Lab Course University of Basel Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope Advanced Lab Course University of Basel October 6, 2015 Contents 1 Introduction 2 2 The Confocal Microscope

More information

Supporting Information

Supporting Information Copyright WILEY-VCH Verlag GmbH & Co. KGaA, 69469 Weinheim, Germany, 2012. Supporting Information for Adv. Mater., DOI: 10.1002/adma.201203033 Solid Immersion Facilitates Fluorescence Microscopy with Nanometer

More information

Εισαγωγική στην Οπτική Απεικόνιση

Εισαγωγική στην Οπτική Απεικόνιση Εισαγωγική στην Οπτική Απεικόνιση Δημήτριος Τζεράνης, Ph.D. Εμβιομηχανική και Βιοϊατρική Τεχνολογία Τμήμα Μηχανολόγων Μηχανικών Ε.Μ.Π. Χειμερινό Εξάμηνο 2015 Light: A type of EM Radiation EM radiation:

More information

3.0 Alignment Equipment and Diagnostic Tools:

3.0 Alignment Equipment and Diagnostic Tools: 3.0 Alignment Equipment and Diagnostic Tools: Alignment equipment The alignment telescope and its use The laser autostigmatic cube (LACI) interferometer A pin -- and how to find the center of curvature

More information

arxiv: v1 [physics.optics] 7 Sep 2007

arxiv: v1 [physics.optics] 7 Sep 2007 Measurement of focusing properties for high numerical aperture optics using an automated submicron beamprofiler arxiv:0709.1004v1 [physics.optics] 7 Sep 2007 J. J. Chapman, B. G. Norton, E. W. Streed and

More information