Fluence Field Modulated X-ray CT using Multiple Aperture Devices. Acknowledgements
|
|
- Suzanna Campbell
- 6 years ago
- Views:
Transcription
1 Fluence Field Modulated X-ray CT using Multiple Aperture Devices Joseph W. Stayman 1, Aswin J. Mathews 1, Wojciech Zbijewski 1 Grace Gang 1, Jeffrey H. Siewerdsen 1 Satomi Kawamoto 2 Ira Blevis 3, Reuven Levinson 3 1 Department of Biomedical Engineering, Johns Hopkins University 2 Department of Radiology, Johns Hopkins Hospital 3 Philips Medical Systems Johns Hopkins University Schools of Medicine and Engineering Faculty and Scientists Tharindu De Silva Grace Gang Matthew Jacobson Alejandro Sisniega Wojciech Zbijewski Acknowledgements Clinicians John Carey Gary Gallia Kelvin Hong A Jay Khanna Doug Reh Marc Sussman Clifford Weiss AIAI Laboratory Advanced Imaging Algorithms and Instrumentation Lab aiai.jhu.edu web.stayman@jhu.edu I-STAR Laboratory Imaging for Surgery, Therapy, and Radiology istar.jhu.edu jeff.siewerdsen@jhu.edu Students Qian Cao Sarah Ouadah Sureerat Reaungamornrat Steven Tilley II Ali Uneri Jennifer Xu Esme Zhang Funding NIH U01EB (Stayman) Presented at SPIE-Medical Imaging
2 Beam Shaping in Computed Tomography Bowtie Filters Shape via variable thickness Reduce dynamic range of data Help to homogenize noise X-ray Source Limitations Static beam shape Patients are not circular/symmetric Patient positioning (not centered) Patient Detector Split bow-tie designs Fluence Field Modulation Strategies Piecewise linear design Fluid-filled designs (Hsieh, Pelc, Medical Physics, 2013) (Szczykutowicz, Hermus, SPIE, 2015) Presented at SPIE-Medical Imaging
3 Challenges for Fluence Field Modulation in CT Practical implementations require Compact design Limited space for actuators (limited size/number) Limited thickness of attenuators Fast actuation Modern CT spins ~3-5 Hz Min-to-max modulation <100 ms Multiple Aperture Device (MAD) Concept X-ray Source X-rays MAD Filter Patient Detector MAD Features: Binary filter (0% or 100% pass) Small actuation distances (1 period) Thin design (e.g. a few mm of tungsten) Spectrally neutral Presented at SPIE-Medical Imaging
4 Multiple Aperture Device (MAD) Concept X-ray Source X-rays MAD Filter Patient Detector MAD Features: Binary filter (0% or 100% pass) Small actuation distances (1 period) Thin design (e.g. a few mm of tungsten) Spectrally neutral Design Goals Desire smooth fluence patterns Avoid seeing fine MAD structure Principles Local fluence determined by amount of blockage Blockers are too small to resolve (determined by pitch) Individual MAD Design SDD SMD MAD pitch FS X-ray Focal Spot MAD Plane MAD Pitch = FS 1 SMD SDD Detector Plane Presented at SPIE-Medical Imaging
5 Illustration of Single MAD Design - Simulation Constant Optimal Pitch Design Linearly Decreasing Bar Widths Position at MAD (mm) Presampling Fluence at Detector Plane Position at Detector (mm) Sampled Fluence at Detector Detector Element Dual MAD Design MAD Pitch (P) MAD Pitch (P) b 1 (0) b 1 (1) b 1 (2) MAD #1 d(0) d(1) d(2) b 2 (0) b 2 (1) b 2 (2) MAD #2 Dx Presented at SPIE-Medical Imaging
6 MAD Transmission MAD Transmission MAD Transmission Transmission with Relative Motion (Dx) and d-offsets Offset to Right Offset to Left Offset to Center b 1 (0) d = 0 b 1 (0) d = (b 1 -b 2 ) b 1 (0) d = (b 1 -b 2 )/2 b 2 (0) Dx b 2 (0) Dx b 2 (0) Dx Dx Dx Dx Position at MAD (mm) Position at MAD (mm) Position at MAD (mm) Illustration of Fluence Patterns for a Specific Goal Desired Fluence Achieved Fluence Presented at SPIE-Medical Imaging
7 Detector Physical Experiments and Fabrication Object Stage MAD Stages X-ray Source Test-bench Features Stable controllable motion Step-and-shoot, pulsed x-rays kvp, 80 kw, w/ pulse width modulation Flat-panel detector 43x43 cm 2 MAD stages, 2.5 mm steps, 1 mm encoders a = mm 320 mm dia SDD (1040 mm) SAD (803 mm) MADs SMD (341 mm) Source fs ~ 1.19 mm Detector Practical Considerations for Fabrication Focused MAD filters a Dz MAD1 MAD2 Focal Spot Measurements Focal Spot Pinhole Image Variable Line Pairs Projection Image Presented at SPIE-Medical Imaging
8 Slot Thickness (mm) Stayman et al. (Johns Hopkins University) MAD Construction Tungsten Laser Sintering Tungsten Powder 3D CAD File Laser Powder Bed Fusion EDM Wire Cutting and Mounting Single MAD Design to Flatten Fluence for 32 cm Phantom 0.8 Designed Slot Thickness: 1-b 1 (x) Aperture Position (mm) Normalized Detector Signal: Blank Scan 3D Printed Tungsten MAD Minimum Feature Size: ~100 mm Focused at 341 mm Thickness: 1.5 mm Normalized Detector Signal: 32 cm CTDI Phantom Presented at SPIE-Medical Imaging
9 Sample Reconstruction with MAD Filtering Sample Projection CatPhan Resolution Phantom FBP Reconstruction of Test-Bench Data 360 Acquisition, 360 angles, 80 kvp, 10 ma, 50 ms pulses, 180 mas total Sample Projection Sample Reconstruction with MAD Filtering CatPhan Resolution Phantom FBP Reconstruction of Test-Bench Data 360 Acquisition, 360 angles, 80 kvp, 10 ma, 50 ms pulses, 180 mas total Presented at SPIE-Medical Imaging
10 Conclusion Presented a novel FFM design Based on MAD filtration Compact design Large scale shaping with small actuation levels Ongoing Work Dual-MAD Experiments Integration in a commercial CT gantry Conclusion / Ongoing Work Presented at SPIE-Medical Imaging
HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationThermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source
Energy Level (ev) Multi-beam x-ray source array based on carbon nanotube field emission O. Zhou, JP Lu, X. Calderon-Colon, X. Qian, G. Yang, G. Cao, E. Gidcumb, A. Tucker, J. Shan University of North Carolina
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More informationDistributed source x-ray tube technology for tomosynthesis imaging
Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.
More informationPitfalls and Remedies of MDCT Scanners as Quantitative Instruments
intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature
More informationRaySafe X2. Effortless measurements of X-ray
RaySafe X2 Effortless measurements of X-ray At your fingertips We ve grown accustomed to intuitive interactions with our devices. After all, it s not the device that s most important, but what you can
More informationAn EM Reconstruction with Improved Signal-to. to-noise Ratio for Coded Aperture Imaging
An EM Reconstruction with Improved Signal-to to-noise Ratio for Coded Aperture Imaging Cynthia Tozian, PhD UMass Lowell Bristol-Myers Squibb Medical Imaging Young Investigators Symposium April 12, 2006
More informationResearch Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.
Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National
More informationBreast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationdiagnostic examination
RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011
More informationImage Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose
Image Quality and Dose Issues in MSCT Image Quality and Dose Image quality Image noise Spatial resolution Contrast Artefacts Speckle and sharpness S. Edyvean St. George s Hospital London SW17 0QT Radiation
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationHigh-Fidelity Modeling of Shift-Variant Focal-Spot Blur for High-Resolution CT
High-Fidelity Modeling of Shift-Variant Focal-Spot lur for High-Resolution CT Steven Tilley II, Wojciech Zbijewski, J. Webster Stayman bstract Dedicated application-specific CT systems are popular solutions
More informationComputed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology
The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray
More informationLarge-Area CdTe Photon-Counting Pixel Detectors
Large-Area CdTe Photon-Counting Pixel Detectors Tilman Donath, Application Scientist 22.6.2015, DIR2015, Ghent DECTRIS Ltd. 5400 Baden Switzerland www.dectris.com Agenda 1. Introduction Hybrid Photon Counting
More informationA comparison of two methods for the determination of freein-air geometric efficiency in MDCT
A comparison of two methods for the determination of freein-air geometric efficiency in MDCT Theocharis Berris *1, Kostas Perisinakis 1,, Antonios E. Papadakis and John Damilakis 1, 1 Department of Medical
More informationQC by the MPE in Belgium
Acceptance testing of state-of-the-art CT scanners using a new national protocol: first experience on a large number of scanners of different make and model the working group Radiology of the Belgian Hospital
More information2010 Philips BrightView XCT SPECT/CT
2010 Philips BrightView XCT SPECT/CT Unit was purchased from Philips training center in 2015. Installed but never been used by the current facility. (Scroll for pictures) BrightView XCT Camera with PinPoint
More informationFocal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution
Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationDetector technology in simultaneous spectral imaging
Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become
More informationClinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11
Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) Hiroki Kadoya Yukiko Kitagawa MEDIX VOL. 61 P.8 P.11 Clinical Experience Using the Open Bore Multislice CT System Supria
More informationImprovement of CT image quality with iterative reconstruction idose4
Improvement of CT image quality with iterative reconstruction idose4 Poster No.: C-0387 Congress: ECR 2014 Type: Scientific Exhibit Authors: M.-L. Olsson, K. Norrgren, M. Söderberg; Malmö/SE Keywords:
More informationNuclear Associates , , CT Head and Body Dose Phantom
Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More informationPhilip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany.
A new generation of x-ray computed tomography devices for quality inspection and metrology inspection in the field of additive manufacturing and other sciences Philip Sperling Sales Science and New Materials,
More informationT h e P h a n t o m L a b o r a t o r y
T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order
More informationHIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE
HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei
More information1. Patient size AEC. Large Patient High ma. Small Patient Low ma
Comparison of the function and performance of CT AEC systems CTUG meeting by Emily Field Trainee clinical scientist 14 th th Breakdown CT Automatic Exposure Control (AEC) Background Project Description
More informationNew spectral benefi ts, proven low dose
New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers
More information12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
Joseph V. Fritz, PhD Nandor Pintor, MD Dent Neurologic Institute ASN 2017 Friday, January 20, 2017 Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees
More informationInitial setup and subsequent temporal position monitoring using implanted RF transponders
Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements
More informationTOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT
More informationAutomated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London
Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationTest Equipment for Radiology and CT Quality Control Contents
Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3
More informationOptical coherence tomography
Optical coherence tomography Peter E. Andersen Optics and Plasma Research Department Risø National Laboratory E-mail peter.andersen@risoe.dk Outline Part I: Introduction to optical coherence tomography
More informationA Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017
A Fast Monolithic System for Proton Imaging Fritz DeJongh ProtonVDA Inc October 2017 Disclosures I am a cofounder and co-owner of ProtonVDA Inc We hold intellectual property rights on our proton imaging
More informationDALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI
DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI D. Pacella ENEA - Frascati LIMS, Frascati 14-15 ottobre 2015 Come per la fotografia:
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationMTF and NPS of single-shot dual-energy sandwich detectors
MTF and NPS of single-shot dual-energy sandwich detectors Junwoo Kim, a Dong Woon Kim, a Hanbean Youn, b,c Ho Kyung Kim a,c a School of Mechanical Engineering, Pusan National University, Busan 609-735,
More informationJEFFERSON COLLEGE COURSE SYLLABUS BET220 DIAGNOSTIC IMAGING. 3 Credit Hours. Prepared by: Scott Sebaugh Date: 2/20/2012
JEFFERSON COLLEGE COURSE SYLLABUS BET220 DIAGNOSTIC IMAGING 3 Credit Hours Prepared by: Scott Sebaugh Date: 2/20/2012 Mary Beth Ottinger, Division Chair Elizabeth Check, Dean, Career & Technical Education
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More information2012 :15th SESSION of ESMP
2012 :15th SESSION of ESMP Lecture presented in Archamps (Salève Building) by : Elly CASTELLANO (London) Patient dosimetry in x-ray imaging and CT Elly Castellano Objectives measurable dose quantities
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More informationPHYS 1112L - Introductory Physics Laboratory II
PHYS 1112L - Introductory Physics Laboratory II Laboratory Advanced Sheet Thin Lenses 1. Objectives. The objectives of this laboratory are a. to be able to measure the focal length of a converging lens.
More informationOptimized CT metal artifact reduction using the Metal Deletion Technique (MDT)
Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT) F Edward Boas, Roland Bammer, and Dominik Fleischmann Extended abstract for RSNA 2012 Purpose CT metal streak artifacts are
More informationA Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images
A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of
More informationAuthors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2
Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2 Medical Consult, SA; Establish a method to correlate image
More informationThe influence of bowtie filtration on cone-beam CT image quality
The influence of bowtie filtration on cone-beam CT image quality N. Mail Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario M5G 2M9, Canada and Ontario Cancer Institute, University
More informationMC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE
MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for
More informationDevelopment of a Low-order Adaptive Optics System at Udaipur Solar Observatory
J. Astrophys. Astr. (2008) 29, 353 357 Development of a Low-order Adaptive Optics System at Udaipur Solar Observatory A. R. Bayanna, B. Kumar, R. E. Louis, P. Venkatakrishnan & S. K. Mathew Udaipur Solar
More informationK-edge subtraction X-ray imaging with a pixellated spectroscopic detector
K-edge subtraction X-ray imaging with a pixellated spectroscopic detector Silvia Pani Department of Physics, University of Surrey Summary Hyperspectral imaging K-edge subtraction X-ray imaging for mammography
More informationmake it easy, with Ray
Lower dose - Quick scan times - Pulsed X-ray technology - Multiple scan modes 3 Dedicated detectors - Reliable performance - No damage - Long life span Easy upgrade - Ready to upgrade CBCT & Cephalometric
More informationNeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results
NeuViz 16 Computed Tomography Elevating routine imaging for exceptional results Essence NeuViz 16 Raising the bar on clinical utility in routine imaging. Get more. More clinical information for patients.
More informationKit for building your own THz Time-Domain Spectrometer
Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6
More informationSlide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?
Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationPortable Digital X-ray
Portable Digital X-ray Portable Digital X-ray Ideal choice for a complete range of radiographic procedure and eliminates the need for costly room modifications This is a complete outdoor X-ray system with
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationPET/CT Instrumentation Basics
/ Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire
More informationTHE PANTAK THERAPAX SXT 150
THE PANTAK THERAPAX SXT 150 combines the safety and ease-of-use of photon radiation therapy with state-of-the-art technology, including microprocessor control and solid-state electronics. Using Therapax,
More informationabc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.
abc September 2004 MHRA 04099 Philips Mx8000 IDT CT scanner technical evaluation Best choice best practice www.mhra.gov.uk nww.medical-devices.nhs.uk About MHRA evaluation reports. What you can expect.
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationMammography: Physics of Imaging
Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness
More informationFour-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development
ORIGINAL ARTICLE ORIGINAL ARTICLE Radiation Medicine: Vol. 21 No. 1, 17 22 p.p., 2003 Four-dimensional Computed Tomography (4D CT) Concepts and Preliminary Development Masahiro Endo,* Takanori Tsunoo,*
More informationDIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook
DIAGNOSTIC ACCREDITATION PROGRAM Radiology and CT Quality Control Procedures Workbook Quality Control Procedures Radiography/CR/DR Safety Code 35 Summary For more detail about each quality control (QC)
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationPractix. Mobile Radiography. Changing its point of view comes naturally. Practix 300
Practix 300 Specifications Practix Mobile Radiography Changing its point of view comes naturally Practix 300 With its swiveling telescopic tube arm and rotatable column, the Practix 300 adds a new and
More informationQuantitation of clinical feedback on image quality differences between two CT scanner models
Received: 4 August 2016 Revised: 4 November 2016 Accepted: 12 December 2016 DOI: 10.1002/acm2.12050 MEDICAL IMAGING Quantitation of clinical feedback on image quality differences between two CT scanner
More informationUltra-sensitive, room-temperature THz detector using nonlinear parametric upconversion
15 th Coherent Laser Radar Conference Ultra-sensitive, room-temperature THz detector using nonlinear parametric upconversion M. Jalal Khan Jerry C. Chen Z-L Liau Sumanth Kaushik Ph: 781-981-4169 Ph: 781-981-3728
More informationInvestigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras
Egyptian J. Nucl. Med., Vol 2, No. 2, Dec. 2009 82 PHYSICS, Original Artical Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras Abdelsattar, M.B. Ph.D.; BuHumaid,
More informationANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward
More informationDedicated Veterinary Imaging Solutions Digital, CR and Analog Imaging Solutions for any size patient and any size budget.
by Dedicated Veterinary Imaging Solutions Digital, CR and Analog Imaging Solutions for any size patient and any size budget. Serving the Veterinary Profession for Over 75 Years. ... We See Things Differently
More informationModel Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise
Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise Jonas Rydberg, M.D. Professor of Radiology Indiana University School of Medicine Indianapolis, Indiana Medical
More informationDigital Radiography : Flat Panel
Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non
More informationNuclear Associates
Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are
More informationCOMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED
More informationv tome x m microfocus CT
GE Inspection Technologies v tome x m microfocus CT Uniting premium 3D metrology and inspection with quality and speed. gemeasurement.com/ct x plore precision CT line Inspect with precision, power, and
More informationPET: New Technologies & Applications, Including Oncology
PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare
More informationHigh Energy Digital Radiography & 3D-CT for Industrial Systems
DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France High Energy Digital Radiography & 3D-CT for Industrial Systems Non-Destructive
More informationFar field intensity distributions of an OMEGA laser beam were measured with
Experimental Investigation of the Far Field on OMEGA with an Annular Apertured Near Field Uyen Tran Advisor: Sean P. Regan Laboratory for Laser Energetics Summer High School Research Program 200 1 Abstract
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationSuperfast phase-shifting method for 3-D shape measurement
Superfast phase-shifting method for 3-D shape measurement Song Zhang 1,, Daniel Van Der Weide 2, and James Oliver 1 1 Department of Mechanical Engineering, Iowa State University, Ames, IA 50011, USA 2
More informationCOCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION
COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION Revision: 1 Date: June 2015 Approved: June 2015 TABLE OF CONTENT 1. INTRODUCTION...
More information8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements
The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston
More informationWater-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source
2015 International Workshop on EUV and Soft X-Ray Sources Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source T. Parkman 1, M. F. Nawaz 2, M. Nevrkla 2, M. Vrbova 1, A. Jancarek
More information3.003 Lab 3 Part A. Measurement of Speed of Light
3.003 Lab 3 Part A. Measurement of Speed of Light Objective: To measure the speed of light in free space Experimental Apparatus: Feb. 18, 2010 Due Mar. 2, 2010 Components: 1 Laser, 4 mirrors, 1 beam splitter
More informationOverview of Safety Code 35
Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)
More informationOptimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation
Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT
More informationMigrating from traditional to Digital Radiography in Aerospace
Migrating from traditional to Digital Radiography in Aerospace More info about this article: http://www.ndt.net/?id=22663 Abstract Lennart Schulenburg VisiConsult X-ray System & Solutions GmbH Brandenbrooker
More informationSoftware and Hardware in CCTA. Elly Castellano PhD
Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements
More informationDevelopment of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)
Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Liang-Chia Chen 1#, Chao-Nan Chen 1 and Yi-Wei Chang 1 1. Institute of Automation Technology,
More informationSYLLABUS. 1. Identification of Subject:
SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject
More information3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development
Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory
More informationEvaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Measurement of Optical Tissue Phantoms and Human Body Tissue
Journal of the Korean Physical Society, Vol. 51, No. 5, November 2007, pp. 1663 1667 Evaluation of a Chip LED Sensor Module at 770 nm for Fat Thickness Measurement of Optical Tissue Phantoms and Human
More informationRAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II
RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated
More informationWhite Paper: Modifying Laser Beams No Way Around It, So Here s How
White Paper: Modifying Laser Beams No Way Around It, So Here s How By John McCauley, Product Specialist, Ophir Photonics There are many applications for lasers in the world today with even more on the
More informationI. Introduction.
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Accuracy of measuring half- and quarter-value layers and appropriate aperture width of a convenient method using a lead-covered case
More information