1 Operating Manual Wedge Angle Scanner
|
|
- Sharyl Lindsay Robinson
- 6 years ago
- Views:
Transcription
1 1 Operating Manual Wedge Angle Scanner Operating Manual Wedge Angle- Scanner (Wedge Angle Scanner Flatscan HUD) OEG GmbH Wildbahn 8i Frankfurt (Oder) Tel.: Fax: Internet: Contents 1. Proper use of the device 2. Safety warnings 3. Connection requirements, space requirements 4. Measurement principle 5. Construction of the device, maintenance instructions 6. Operating the device 7. Completing the measurement in USER Mode (automatically measuring) 8. Service mode 8.1 Measurement template 8.2 Settings 8.3 Single measurements 8.4 Calibrating 8.5 Motor settings 8.6 Camera settings 9. Installation warnings 10. Service contact data OEG GmbH
2 2 Operating Manual Wedge Angle Scanner 1. Proper Use of the Device The device wedge angle scanner serves only to measure the course of a wedge error on glass plates, especially automobile front windshields with the maximum dimensions: - Length 1600mm - Height 1000mm - Thickness 10mm. - The mass of the sample may not exceed 25kg. 2. Safety Warnings Emergency stop: The emergency stop switch must always remain accessible and may not be removed. The emergency stop switch cuts off the electricity supply for the stepper motor control. This function may not be altered. Motor power/travel speeds: The permissible phase current for the step motors is 2.5 A. The permissible travel speed of the portal is 30mm/s. Linear guides/portal: Heed instructional decals. When the device is engaged, do not reach between the rubber lips. (crushing danger) Laser: A semi-conductor laser with a maximum output of 1mW is used (class 2). Avoid looking directly into the laser beam when the device is operationally ready. During maintenance work, disconnect the laser from the power supply (pull power cable on the sensor head out). Program start: Remove all objects from the sample table before starting the measurement program. 3. Connection Requirements / Space Requirement Power supply: 220V..240V, 10A Space requirement: 2.5mx2.5m (device and space for operator) Weight: ca. 300kg Working temperature: 18 C to 25 C, Temperature constancy: During a measurement +/-1K Attention! Avoid draughts during measurement.
3 3 Operating Manual Wedge Angle Scanner 4. Measurement Principle electronic autocollimator Reference mirror Laser half mirror rack Figure 1: Functional Principle The measurement principle is based on highly precise measurements of the beam deflection. It is based on the fact that a collimated light bundle is deflected when it passes through a glass plate whose surfaces are not parallel. The deflection (δ) is proportional to the so-called wedge error (ε). In the case of double light transmission: δ 2k ε (1) with k as calibration factor. Equation (1) applies only to small wedge angles (<1 ) and not to larger angles for the test sample with respect to the measurement system. In the concrete case, the laser beam passes through the test sample more or less deflected to a reference mirrors, will be reflected back by it, passes through the test sample a second time and a then finally comes into a collimating tube through a 90 with a CCD camera. In the video image, an image of the laser spot will appear, whose position is directly associated with the wedge angle of the test sample. Before any measurement, a reference measure of the position of the laser spot will be taken without the test sample (front windshield). This defines the zero point (ε=0).
4 4 Operating Manual Wedge Angle Scanner Attention! The calibration value, that is the proportionality k from the equation (1) for the wedge angle measurement depends on the refractive index of the glass and/or layer system of an automobile windshield. If precise absolute values of the wedge are demanded, theoretically it needs to be re-calibrated for every type of glass. If you are only interested in small fluctuations in wedge angles, you can calibrate using a standard glass since in that case small differences in the refractive index of the various types of glass hardly play any role. Important! The decisive factor is the fixed (unchangeable) angle position of the reference mirror to the AKF. For that reason a reference measurement is taken automatically prior to every wedge angle measurement. One can assume that no change in the relative position between the reference mirror and sensor will occur during the brief time of the measurements. Over long time periods however this cannot be excluded due to expansion caused by temperature fluctuations. For this reason the reference measurement is highly important for measurement precision. Sensor measurement range (laser auto-collimation system) /ε/ <+- 20mRad The measurement range derives from the focal length of the collimator used. Measurement range portal system: X axis: Y axis: Position uncertainty: 1650mm 700 mm 0.1mm
5 5 Operating Manual Wedge Angle Scanner 5. Structure of the Device Figure 2: Structure of the front windshield scanner with the linear advancers, sample holder and automobile windshield Profile frame construction One can see the linear guides in Figure 2. This has the advantage that the device can be completely disassembled for transport and reassembled with relative ease and remains very stable mechanically.
6 6 Operating Manual Wedge Angle Scanner Specimen holder Figure 3: Specimen holder consisting of two plastic blocks with a nut, a stop and two outer supports on which it is leaned (only visible in Figure 1) The two plastic blocks and the stop will determine the position of the automobile windshield with respect to the portal system. All components of the specimen holder can be moved and fixed. Suggestion: The resting blocks for all types of automobile windshields should remain in the same position. The adjustment to the size of the windshield is then affected by the position of the stop. In this way the middle of the windshield should remain in the same position, at least in the x- direction. Linear guides The portal consists of two vertical standing guides that each contain a ball screw. The lift equals around 1600mm. The guides are driven with one step motor using the so-called main shaft. The y-axis (lift 700mm) also contains a ball screw. Each of the housings of the linear guides has a stepper motor integrated into it (max. 3.5A). Maintenance The linear guides should be annually lubricated. (See appendix, Isel components ). No other maintenance should be necessary.
7 7 Operating Manual Wedge Angle Scanner Laser Module Strahlteiler POL Filter Lasermodul Figure 4: Laser Assembly Group The intensity of the laser can be regulated (potentiometer on the back). Since the regulator range is not adequate, the beam will be further weakened by a polarisation filter during measurement. The filter can however be moved out of the beam path for purposes of adjustment so that the laser spot can be seen on a screen (paper). With the polarisation filter the laser spot is practically invisible. Beam adjustment The beam splitter plate will be mounted on a platform with an adjustable angle. Without a test specimen ( measurement in the air ), the laser spot should always appear in the middle of the screen image. The laser itself can only be adjusted by trained personnel. Step motor control The step motors are driven by a high-grade step motor control LSTEP. The LSTEP is for its part linked to the PC with a serial interface. The step motor control is configured with special software ( Win Commander ). The measurement program essentially uses the settings that were saved in the file wincom4.ini. You can find more detailed information in the manufacturer s operating manual (CD Win Commander ). Attention: The switch on the front plate of the LSTEP must be set to A.
8 8 Operating Manual Wedge Angle Scanner Camera This is a CCD camera with a USB interface. Connection lines - The LSTEP control device is connected to the PC with the serial interface COM1. - The power supply for the LSTEP goes through the emergency stop switch - One line from each step motor goes to the control device. They are labelled. - The camera is supplied with power through the USB interface. - The laser is supplied with 12V DC (separate plug-in power supply) 6. Putting the Device into Operation Remove all objects from the measurement device that can restrict the portal s range of motion. - Activate the PC. - Activate the step motor controller (LSTEP). The switch is located on the back. Attention: The switch on the front plate must be set to Auto, otherwise the measurement program will not work. - Start the program Flatscan HUD When using a master- slave- table distributor, the LSTEP will be automatically activated when the PC is activated. Directly after the program starts, the reference movements will automatically be carried out. Wait until the guides come to a stop. Switching it off is done by reversing the sequence. 7. Completing a Measurement in USER Mode In User Mode, only the function measure can be called up after the program starts and is initialised (Figure 5). The video image of the laser spot will appear. Attention: The laser spot should not be overshot (no filter). See Figure 6 and Figure 7.
9 9 Operating Manual Wedge Angle Scanner Figure 5: Program start in user mode Figure 6: Incorrect! Laser spot overshot. Diffraction effects are clearly visible.
10 10 Operating Manual Wedge Angle Scanner Figure 7: Correct! Diffraction effects are hardly visible. In spite of this the laser spot is visible as a bright spot. The operator will be asked to select a suitable measurement program for the windshield from a list. After the automobile windshield has been correctly positioned and the HUD and/or PVB number is typed in, the measurement procedure can be initiated with the F4 Start. The measurement begins with a reference measurement. Then the section of the automobile windshield selected in the background will be scanned. The measurement program can include several scanning ranges and/or measurement jobs. After a measurement job is completed, the operator will be asked to save the results and continue the measurement program.
11 11 Operating Manual Wedge Angle Scanner Figure 8: Selecting the measurement program. Once a measurement job is completed, a report window will appear (Figure 9). The report will include the measurement results along with the prescribed value and an evaluation Good (Green) Bad (red).
12 12 Operating Manual Wedge Angle Scanner Figure 9: Protocol window (report).
13 13 Operating Manual Wedge Angle Scanner 8. Service Mode The necessary program settings are made in the so-called Sevice Mode, and the so-called measurement templates (measurement programs) are generated. While the program settings may need to changed rarely or not at all, the generation or change of measurement programs may occur more frequently. Service Mode is password protected. 8.1 Measurement Template Figure 10: Selecting the function Edit Measurement Templates
14 14 Operating Manual Wedge Angle Scanner Figure 11: Generating measurement templates (measurement programs) A measurement program is divided into two parts: a) Gravity Center: Here the geometric data for the scanned field, the measurement point spacing and the measurement direction are set (measurement template). The position of the measurement field is determined by its mid-point ( Gravity Center ). The mid-point can be approached using a joystick and the position learned then. If a new name is assigned, the measurement template just created can be added to the list. All others can be edited or deleted. b) Programs: Multiple measuremetn templates can be combined into a measurement program. The name of the program is saved in a list. Existing measurement programs can be edited or deleted. Note: Assign as short and concise a name as possible that the operator can quickly comprehend in USER Mode and cannot confuse with another. 8.2 Settings The window settings contains essentially the parameters that influence the measurement process. These should not be changed or only in consultation with the OEG. Only the directory structure can be changed from time to time.
15 15 Operating Manual Wedge Angle Scanner Figure 12: First tab in the settings. The terms are actually self-explanatory. Of particular importance is the input field Measurement stroke for on-the-fly measurements. The time entered is the measurement rate during continuous movement of the measurement system. If the values are too large and the movements of the measurement sled are too fast, too few measurement values would be recorded. Cycle times that are too small unnecessarily burden the computer. The cycle time should be the same or somewhat greater than the time that the camera system needs to transmit a video image (transmission rate). Figure 13: Second tab for settings These parameters should only be changed by specially trained personnel. The values are responsible for recognising the position of the laser spot. They are determined experimentally or are based on experience. The optimisation of the parameters for signal recognition makes the right system robust against disruptions in the video image (light reflections, noise, etc.). Figure 14: Third tab for settings Here you select the directories for the measurement data.
16 16 Operating Manual Wedge Angle Scanner 8.3 Single Measurements For repairs, maintenance and trouble-shooting, it is necessary to examine the AKF system without measurement. Use the window single measurements. Figure 15: Single Measurements Using Set, set the AKF display to zero. Using Start measure initiates a continuous display of measured values. This is the best way to test whether the system has any noise, drifts or if there are relevant environmental problems. Depending on the AKF system, the measurements will fluctuate between and +-1. In order to get any <measurements, the <laser spot must be within the socalled measurement window (Figure 16). They are not automatically generated here.
17 17 Operating Manual Wedge Angle Scanner Figure 16: The laser spot must lie inside the so-called Meas. window position. The measurement window can be generated with the scroll bars. 8.4 Calibration The calibration data establish the relationship between the measurement value in the video image in pixels and the actual wedge angle in mrad or arc seconds (proportionality factor). Calibration data can be uploaded, stored, edited or generated from scratch. A second calibration is done in two steps. A first measurement without the wedge (that is, in the air) and then a second measurement with a known wedge. The pixel difference calculated will then be assigned to the known wedge angle. Figure 17: Calibration window. The window is under Video control in the main menu, Attention: Change or initial generation of calibration data may only be carried out by trained service personnel exercising due care. One needs a reference glass wedge, whose refractive index corresponds to the glass of the test specimens. 8.5 Motor Settings The option Measurement table control is in the main menu. Using the first tab, one can complete any of the movements of the portal and determining standard position angles.
18 18 Operating Manual Wedge Angle Scanner Figure 18: First tab for motor settings. Set and then Go means: Travel to the saved standard position. Using Teach in the positions involved will be overwritten with the current positions. Please note that in Flatscan HUD version of the program Load position and Zero position are identical. Figure 19: Motor settings Set speeds and acceleration here. Changes of the parameters should only be completed by trained personnel. During the warranty period these must only be carried out by OEG.
19 19 Operating Manual Wedge Angle Scanner 8.6 Camera Settings Flatscan-HUD can work with different camera systems. Currently we are using USB 2.0 cameras from the manufacturer IDS (Imaging Source). The camera settings are broken down into several Tabs with two of them being particularly important. Figure 20: Setting for light sensitivity using the exposure time. Keep in mind that exposure time, pixel frequency and transfer rate are related. Attention: There is no reason to change the settings that are entered when the device is was installed. Changes can only be beneficial and necessary if the laser for the AKF system needs to be changed. Figure 21: Camera settings that affect the linearity of the video signal transfer. Attention: Never change
20 20 Operating Manual Wedge Angle Scanner 9. Installations Tips First install the camera drivers. Then install the program. Attention: Do not install the program under C:\Program files\... Just install it under C:\oeg 10. Service Contact Information If you have problems please contact us at:
Top spin Nr /
Top spin Nr. 1840 0000 / 1840 1000 Bedienungsanleitung 21-6680 28052014 / A Made in Germany Ideas for dental technology Top spin Nr. 1840 0000 / 1840 1000 Contents 1. Introduction...2 1.1 Symbols...2 2.
More information3B SCIENTIFIC PHYSICS
3B SCIENTIFIC PHYSICS Equipment Set for Wave Optics with Laser 1003053 Instruction sheet 06/18 Alf 1. Safety instructions The laser emits visible radiation at a wavelength of 635 nm with a maximum power
More information3B SCIENTIFIC PHYSICS
3B SCIENTIFIC PHYSICS Equipment Set for Wave Optics with Laser U17303 Instruction sheet 10/08 Alf 1. Safety instructions The laser emits visible radiation at a wavelength of 635 nm with a maximum power
More informationUsing the USB2.0 camera and guider interface
Using the USB2.0 camera and guider interface The USB2.0 interface is an updated replacement for the original Starlight Xpress USB1.1 unit, released in 2001. Its main function is to provide a USB2 compatible
More informationLASER. Analog Laser Displacement Transducer. LAM Series. Key-Features: Content:
LASER Analog Laser Displacement Transducer LAM Series Key-Features: Content: Overview, Measuring Principle...2 Installation Instructions...3 Technical Data...4 Technical Drawings.7 Electrical Connection...9
More informationScanning Electron Microscope FEI INSPECT F50. Step by step operation manual
Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check
More informationPICO MASTER 200. UV direct laser writer for maskless lithography
PICO MASTER 200 UV direct laser writer for maskless lithography 4PICO B.V. Jan Tinbergenstraat 4b 5491 DC Sint-Oedenrode The Netherlands Tel: +31 413 490708 WWW.4PICO.NL 1. Introduction The PicoMaster
More informationOperating Manual. itech IMPAXX Digital Monument Engraver
Operating Manual itech IMPAXX Digital Monument Engraver Copyright 2013 Allen Datagraph Systems - All Rights Reserved Manual Date August 2013 Congratulations on your new purchase! Thank You for Selecting
More informationS E L E C T I O N. Arm Curl. User manual
S E L E C T I O N T H E S T R E N G T H E V O L U T I O N User manual The identification plate of the and manufacturer, affixed behind the seat, gives the following details: A Name and address of the manufacturer
More informationInstruction sheet VideoCom Retroreflecting Foil Falling Body for VideoCom. 1 Safety notes
Physics Chemistry Biology Technics LEYBOLD DIDACTIC GMBH 8/97-Hund- Instruction sheet 337 47 337 471 337 472 VideoCom Retroreflecting Foil Falling Body for VideoCom Fig. 1 VideoCom (337 47) is a camera
More informationWE BRING QUALITY TO LIGHT DTS 500. Positioner Systems AUTOMATED DISPLAY AND LIGHT MEASUREMENT
WE BRING QUALITY TO LIGHT DTS 500 Positioner Systems AUTOMATED DISPLAY AND LIGHT MEASUREMENT Standalone XYZ positioners (260 to 560 mm max. travel range) Standalone 2-axis goniometers (up to 70 cm diagonal
More informationGeneral Physics Laboratory Experiment Report 2nd Semester, Year 2018
PAGE 1/13 Exp. #2-7 : Measurement of the Characteristics of the Light Interference by Using Double Slits and a Computer Interface Measurement of the Light Wavelength and the Index of Refraction of the
More informationKit for building your own THz Time-Domain Spectrometer
Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6
More informationHandbook for the Starlight Xpress AO unit Issue 1 21/8/2005 Handbook for the Starlight Xpress SXV-AO unit
Handbook for the Starlight Xpress SXV-AO unit Thank you for purchasing an SXV-AO active optics unit. This device should give you much improved guiding accuracy with almost any telescope and mount. Please
More informationFEI Tecnai G 2 F20 Operating Procedures
FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account
More informationAn Optical Characteristic Testing System for the Infrared Fiber in a Transmission Bandwidth 9-11μm
An Optical Characteristic Testing System for the Infrared Fiber in a Transmission Bandwidth 9-11μm Ma Yangwu *, Liang Di ** Center for Optical and Electromagnetic Research, State Key Lab of Modern Optical
More informationPyroMiniUSB Series Operators Guide
PyroMiniUSB Series Operators Guide The PyroMiniUSB is a simple, compact infrared temperature sensor with USB communications. It measures the surface temperature of a variety of materials without contact.
More informationSlimScan BT Users Manual
SlimScan BT Users Manual ASP Microcomputers 456 North Road, Ormond, Victoria, 3204 Australia Telephone: (03) 9578-7600 FAX: (03) 9578-7727 email: solutions@asp.com.au World Wide Web: http://www.asp.com.au
More informationAgilEye Manual Version 2.0 February 28, 2007
AgilEye Manual Version 2.0 February 28, 2007 1717 Louisiana NE Suite 202 Albuquerque, NM 87110 (505) 268-4742 support@agiloptics.com 2 (505) 268-4742 v. 2.0 February 07, 2007 3 Introduction AgilEye Wavefront
More informationS E L E C T I O N. Upper Back. User manual
and S E L E C T I O N T H E S T R E N G T H E V O L U T I O N User manual and and The identification plate of the and manufacturer, affixed to the frame on the side opposite the padded rest, gives the
More informationInstallation Guide. English. English
Installation Guide Safety Instructions For your safety, read all the instructions in this guide before using the setting plate. Incorrect handling that ignores instructions in this guide could damage the
More informationKoPa Scanner. User's Manual A99. Ver 1.0. SHENZHEN OSTEC OPTO-ELECTRONIC TECHNOLOGY CO.,LTD.
KoPa Scanner A99 User's Manual Ver 1.0 SHENZHEN OSTEC OPTO-ELECTRONIC TECHNOLOGY CO.,LTD. http://www.ostec.com.cn Content Chapter 1 Start... 1 1.1 Safety Warnings and Precautions... 1 1.2 Installation
More information1.6 Beam Wander vs. Image Jitter
8 Chapter 1 1.6 Beam Wander vs. Image Jitter It is common at this point to look at beam wander and image jitter and ask what differentiates them. Consider a cooperative optical communication system that
More informationSCREW JACKS Operating Instructions
INS114EN.5 SCREW JACKS Operating Instructions Sweden Int +46 372 265 00 info@swedrive.se Vat.no SE556145631901 138-4551 1 Table of Contents 1. Health and Safety... 3 1.1. Notice about safety... 3 1.2.
More informationDSSIU-6-1U. Dedicated 6-channel system interface unit for ultra-stable, high precision fluxgate technology DS series current transducers.
DSSIU-6-1U Dedicated 6-channel system interface unit for ultra-stable, high precision fluxgate technology DS series current transducers. Powers up to 6 x DS50 to DS2000 at the same time. Supports calibration
More informationOPERATING INSTRUCTIONS
OPERATING INSTRUCTIONS Rotary Microtome CUT 4062 / CUT 5062 / CUT 6062 CUT 6062 illustrated above INS1000GB 2012-01-06 Instructions CUT4062 / CUT 5062 / CUT 6062 2 CONTENTS 1. INTENDED USE... 4 2. SYMBOLS...
More informationMeasuring Procedure the Principle. The laser beam is scanned by means of a specialized measuring tip within a 3D measurement cylinder.
PRIMES FocusMonitor FM For different wavelengths pyroelectric detectors or photodiodes are used. The divergence of the focused laser beam of lasers is rather small. The relationship between the focal length
More informationOpternus GmbH optische Spleiss- und Messtechnik
Spec. No. B-11M3001A Date of Issue : March 3, 2011 FSM-100M+ and FSM-100P+ ARC FUSION SPLICER 1. GENERAL These specifications cover Sophisticated Specialty fiber fusion splicer FSM-100M+ and Sophisticated
More informationOPERATING MANUAL. ACOUSTO OPTIC MODULATOR MODEL NUMBER: X-LTD X= 1, 2, or 3 mm DOCUMENT NUMBER: 51A00620D
OPERATING MANUAL ACOUSTO OPTIC MODULATOR MODEL NUMBER: 23080-1 23080-X-LTD X= 1, 2, or 3 mm DOCUMENT NUMBER: 51A00620D Document approved for release: W Seale Date: 6/20/06 US OFFICE:. 4005 Opportunity
More informationInstallation Guide. English. English
Installation Guide Safety Instructions For your safety, read all the instructions in this guide before using the setting plate. Incorrect handling that ignores instructions in this guide could damage the
More informationCHAINSAW SHARPENER MODEL: ECSS-1
CHAINSAW SHARPENER MODEL: ECSS-1 Part No: 3402075 ASSEMBLY & INSTRUCTION MANUAL LS0409 INTRODUCTION Thank you for purchasing this CLARKE product Before attempting to use the product, it is essential that
More informationUser Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com
User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...
More informationAcceptance test for the linear motion actuator for the scanning slit of the HIE ISOLDE short diagnostic boxes
EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH CERN ACC NOTE 2014 0099 HIE ISOLDE PROJECT Note 0036 Acceptance test for the linear motion actuator for the scanning slit of the HIE ISOLDE short diagnostic boxes
More informationLOS 1 LASER OPTICS SET
LOS 1 LASER OPTICS SET Contents 1 Introduction 3 2 Light interference 5 2.1 Light interference on a thin glass plate 6 2.2 Michelson s interferometer 7 3 Light diffraction 13 3.1 Light diffraction on a
More informationEnglish User's Guide
User's Guide Imacon Flextight 343 2 2003 Imacon A/S. All rights reserved. Imacon Flextight 343 User's Guide, Part No 70030009, revision B. The information in this manual is furnished for informational
More informationUSER MANUAL FOR VISIBLE SPECTROPHOTOMETER
USER MANUAL FOR VISIBLE SPECTROPHOTOMETER 1 Table of Contents 1. MAIN USAGES...3 2. WORKING ENVIRONMENT...3 3. MAIN TECHNICAL DATA AND SPECIFICATIONS...4 4. WORKING PRINCIPLE...5 5. OPTICAL PRINCIPLE...6
More informationTH450A-T TH550A-T THP550-T/TS3000
0 TH450A-T TH550A-T THP550-T/TS3000 INSTRUCTION MANUAL CEILING TYPE (OVERHEAD TRAVELING TYPE) INDUSTRIAL ROBOT SPECIFICATIONS Notice 1. Make sure that this instruction manual is delivered to the final
More informationMinolta Scanner Plugin
Minolta Scanner Plugin For a list of Minolta digitizers and Geomagic software products with which this plugin is compatible, see Release Notes for Geomagic Minolta Plugin 7.6.0.3. Copyright 2005, Raindrop
More informationPolymer Electric. Product information. Ultrasonic Industrial Sensor USi
Polymer Electric Product information Ultrasonic Industrial Sensor USi Master Slave USi-PP 1005632 * 1005633 * USi-IP 1005899 * 1005900 * USi-UP 1005901 * 1005902 * * plus Sensor 1005264 GmbH & Co. KG Polymer
More informationUragan- for LabView. Installation
for LabView Installation Follow these steps to install the LabView driver: 1. Download UraganLabView.zip 2. Unzip UraganLabView.zip in the user.lib directory of your LabView installation 3. Run LabView
More informationClearClick Virtuoso. 22MP Film & Slide Scanner. Quick Start Guide & User s Manual
ClearClick Virtuoso 22MP Film & Slide Scanner Quick Start Guide & User s Manual This document 2016 ClearClick Software LLC. All rights reserved. Do not reproduce in any form without permission. Register
More information16. Sensors 217. eye hand control. br-er16-01e.cdr
16. Sensors 16. Sensors 217 The welding process is exposed to disturbances like misalignment of workpiece, inaccurate preparation, machine and device tolerances, and proess disturbances, Figure 16.1. sensor
More informationWinDIAS 3. Quick Start Guide version 3.2. Delta-T Devices Ltd. Image Analysis System
WinDIAS 3 Image Analysis System Quick Start Guide version 3.2 Delta-T Devices Ltd Overview Note: Not all TWAIN devices will work Page 2 Video Camera Operations 1 Connect camera to PC 2 PC detects camera,
More informationLeuze electronic. 3 Series. Product description
3 Series Product description Contents: 3 Series - overview and advantages Special features of Series 3 Sensor selection table Data sheets All rights reserved, especially the rights of copying and translation.
More informationExperimental Question 2: An Optical Black Box
Experimental Question 2: An Optical Black Box TV and computer screens have advanced significantly in recent years. Today, most displays consist of a color LCD filter matrix and a uniform white backlight
More informationWARNING. BX Ford Explorer With Adaptive Cruise Control & Eco Boost Installation Instructions
Please read BOTH these and the General Instructions before attempting to install or operate this equipment. 1. Blue Ox towing products and accessories are intended to be installed by Blue Ox Dealers who
More informationPlease read BOTH these Installation Instructions and the General Instructions prior to installing or operating this equipment.
Attachment Tab Height: 16-1/2 Serial Number Attachment Tab Width: 24 Please read BOTH these and the General Instructions prior to installing or operating this equipment. 1. Blue Ox towing products and
More informationS E L E C T I O N. Vertical Traction. User manual
and S E L E C T I O N T H E S T R E N G T H E V O L U T I O N User manual and and The identification plate of the and manufacturer, affixed behind the backrest, gives the following details: A Name and
More informationproduct range SMASCANNERS.COM SMA Electronic Document GmbH Södeler Weg Wölfersheim Germany Phone: Fax:
product range SMA Electronic Document GmbH Södeler Weg 2 61200 Wölfersheim Germany Phone: +49 6036 9893010 Fax: +49 6036 9893020 SMASCANNERS.COM book Scanners 4 SCAN MASTER 2 M If you want the high resolution
More informationFig.2: Scanner VistaScan for image plates
RADIOGRAPHIC INSPECTION OF WELDINGS BY DIGITAL SENSORS H. Thiele, H.-J. Friemel RADIS GmbH, Johanniskirchen, Germany Abstract: The newly available digital sensors for radiographic inspection are suitable
More informationMANUAL PLASTIC STRAPPING TOOL MODEL P404
OPERATION MANUAL / SPARE PARTS LIST MANUAL PLASTIC STRAPPING TOOL MODEL P404 43.0404.02 43040402.en/MAS/ 12.05 INDEX PAGE 1 SAFETY INSTRUCTIONS 2 2 TECHNICAL DATA 3 3 OPERATION ELEMENTS 4 4 ADJUSTMENT
More informationHartmann Sensor Manual
Hartmann Sensor Manual 2021 Girard Blvd. Suite 150 Albuquerque, NM 87106 (505) 245-9970 x184 www.aos-llc.com 1 Table of Contents 1 Introduction... 3 1.1 Device Operation... 3 1.2 Limitations of Hartmann
More informationOPERATING MANUAL. ACOUSTO OPTIC MODULATOR MODEL NUMBER: X-1.06-LTD X= 1, 2, or 3 mm DOCUMENT NUMBER: 51A14950A
OPERATING MANUAL ACOUSTO OPTIC MODULATOR MODEL NUMBER: X= 1, 2, or 3 mm DOCUMENT NUMBER: 51A14950A Document approved for release: W Seale Date: 6/06/06 US OFFICE: NEOS Technologies, Inc. 4005 Opportunity
More informationhurryscan, hurryscan II
hurryscan, hurryscan II more Information at: universal and compatible These compact scan heads from SCANLAB provide optimal solutions for nearly all challenges found in industrial laser materials processing.
More informationintelliweld smart welding
intellield more Information at: smart welding Designed for robot-assisted welding applications, this 3D-scan system is capable of swiftly positioning the laser beam along 3D contours. hile a robot guides
More informationClearClick Photo2Digital 10 MP Photo, Slide, & Negative Scanner
ClearClick Photo2Digital 10 MP Photo, Slide, & Negative Scanner Quick Start Guide & User s Manual Full 1-Year Warranty & Free USA Tech Support This product comes with ClearClick s full 1-year warranty
More informationCS276D. CASSESE SA Zone Industrielle - F VERNEUIL L'ETANG. Telecopie : (1)
CS276D S.A. CASSESE Z.I. - 77390 VERNEUIL L'ETANG SIREN 308 305 333 00038 S.A. au capital de 5 000 000 F Téléphone : 01 64 42 49 50 Fax 01 64 42 58 90 CASSESE SA Zone Industrielle - F 77390 VERNEUIL L'ETANG.
More informationINSTRUCTION MANUAL. IBRit - rf1 - usb PC - Station for wireless Data transmission. M e s s t e c h n i k. Messtechnik GmbH & Co.
M e s s t e c h n i k INSTRUCTION MANUAL PC - Station for wireless Data transmission Document No. : D1F604 001 Version : April 2006 Copyright : IBR Messtechnik GmbH & Co. KG Contents 1. Introduction 1.1
More informationStandard Operating Procedure
Standard Operating Procedure Nanosurf Atomic Force Microscopy Operation Facility NCCRD Nanotechnology Center for Collaborative Research and Development Department of Chemistry and Engineering Physics The
More informationUser Manual. Copyright 2010 Lumos. All rights reserved
User Manual The contents of this document may not be copied nor duplicated in any form, in whole or in part, without prior written consent from Lumos. Lumos makes no warranties as to the accuracy of the
More informationPractical work no. 3: Confocal Live Cell Microscopy
Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside
More informationBringing Answers to the Surface
3D Bringing Answers to the Surface 1 Expanding the Boundaries of Laser Microscopy Measurements and images you can count on. Every time. LEXT OLS4100 Widely used in quality control, research, and development
More informationA Revolution in Profile Measurement
2D Laser Displacement Sensor LS Series * FASTUS is a product brand of Optex FA. Linearity ±0.1% of F.S. Sampling period 0.5 ms (max. speed) Superbly affordable 2D measurement A Revolution in Profile Measurement
More informationThermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect tool. for routine measurements
Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as
More informationPart/Reorder Number: Version 1.0
Part/Reorder Number: 870000 Version 1.0 Cross Match Technologies L SCAN 100/100R Operator s Manual Version 1.0 First Edition (August 2006) No portion of this guide may be reproduced in any form or by any
More informationSRV02-Series Rotary Experiment # 3. Ball & Beam. Student Handout
SRV02-Series Rotary Experiment # 3 Ball & Beam Student Handout SRV02-Series Rotary Experiment # 3 Ball & Beam Student Handout 1. Objectives The objective in this experiment is to design a controller for
More informationHuvitz Digital Microscope HDS-5800
Huvitz Digital Microscope HDS-5800 Dimensions unit : mm Huvitz Digital Microscope HDS-5800 HDS-MC HDS-SS50 The world s first, convert the magnification from 50x to 5,800x with a zoom lens HDS-TS50 Huvitz
More informationMANUAL. Textron Motors Diagnostic Tool. This manual is valid for the following Textron Motors Diagnostic Tool:
MANUAL Textron Motors Diagnostic Tool This manual is valid for the following Textron Motors Diagnostic Tool: 0507 TD0507_HB Rev F 6..05 en_english Read the manual before performing the task on the engine.
More informationAG Optical Systems. Newtonian Astrograph Manual Version AG Optical Systems
AG Optical Systems Newtonian Astrograph Manual Version 2 2012 Table of Contents 1. Introduction 2. System Specifications 3. Initial Assembly 4. Collimation 5. Care and Cleaning 6. Cooling Fan Operation
More informationLASER. Analog Laser Displacement Transducer. LAM Series. Key-Features: Content:
LASER Analog Laser Displacement Transducer LAM Series Key-Features: Content: Measuring Principle...2 Installation Instructions...3 Technical Data LAM-S...4 Technical Data LAM-F...5 Technical Drawing...6
More informationBruker Dimension Icon AFM Quick User s Guide
Bruker Dimension Icon AFM Quick User s Guide August 8 2014 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com
More informationSintec Optronics Technology Pte Ltd 10 Bukit Batok Crescent #07-02 The Spire Singapore Tel: Fax:
Sintec Optronics Technology Pte Ltd 10 Bukit Batok Crescent #07-02 The Spire Singapore 658079 Tel: +65 63167112 Fax: +65 63167113 High-power Nd:YAG Self-floating Laser Cutting Head We supply the laser
More informationRPMSP Series Installation Guide
RPMSP Series Installation Guide Contents 1. Overview... page 1 2. Unpacking the Projector...2 3. Projector Configuration...2 4. Projector Throw Distance and Mounting...9 5. Projection Lens Focus...9 6.
More information3. Product Instruction. 3.1 External Components
INDEX INDEX... 0 1. Preface... 1 2. Precautions... 2 3. Product Instruction... 3 3.1 External Components... 3 3.2 Display And Keypad... 4 4. Specifications Check And Adjustment... 5 4.1 Plat Vial... 5
More informationMagnettech GmbH Louis-Bleriot-Straße 5 D Berlin Germany
MiniScope Control 6.51 IInstructiion Manuall Magnettech GmbH Louis-Bleriot-Straße 5 e-mail:info@magnettech.de MiniScope : ESR Spectrometer 2 1. TECHNICAL DATA...3 2. EXTENT OF DELIVERY...3 3. INSTALLATION...4
More informationApplication Tooling Specification Sheet
Tool Kit Order No. 63823-6470 FEATURES Application Tooling Specification Sheet Hand Crimp Tool Order No. 63823-6400 % A full cycle ratcheting hand tool ensures complete crimps % Ergonomically designed
More informationHandbook for the Starlight Xpress AOLF unit Issue 3 17/8/2015. Handbook for the Starlight Xpress SXV-AOLF unit (version 2)
Handbook for the Starlight Xpress SXV-AOLF unit (version 2) Thank you for purchasing an SXV-AOLF2 active optics unit. This device should give you much improved guiding accuracy with almost any telescope
More informationzforce AIR Touch Sensor Specifications
zforce AIR Touch Sensor 2017-12-21 Legal Notice Neonode may make changes to specifications and product descriptions at any time, without notice. Do not finalize a design with this information. Neonode
More informationFilm2USB Converter. Quick Start Guide & User s Manual. Model Number F2USB-05
Film2USB Converter Quick Start Guide & User s Manual Model Number F2USB-05 www.film2usb.com www.clearclicksoftware.com Table of Contents Safety Precautions... 3 Package Contents... 4 Description of Parts...
More informationMAKEBLOCK MUSIC ROBOT KIT V2.0
MAKEBLOCK MUSIC ROBOT KIT V2.0 Catalog Music Robot Kit V2.0 Introduction... 1 1 What is Music Robot Kit V2.0?... 1 1.1 Mechanical part... 1 1.2 Electronic part... 1 1.3 Software part... 1 2 Music Robot
More informationUser s Manual. For correct usage please read these instructions carefully and keeps in a safe place for future reference!
User s Manual For correct usage please read these instructions carefully and keeps in a safe place for future reference! Contents 1: Safety precautions... 2 2: Package contents... 3 3: Description of parts...
More informationKTS- 440RC SERIES TOTAL STATION SERVICE MANUAL
2010 KTS- 440RC SERIES TOTAL STATION SERVICE MANUAL GUANGDONG KOLIDA INSTRUMENT CO., LTD. KOLIDA 2010-10-20 1 KTS 440RC SERIES TOTAL STATION SERVICE MANUAL Tools Oscillograph Multimeter Collimator Screw
More informationChlorophyll Fluorescence Imaging System
Quick Start Guide Chlorophyll Fluorescence Imaging System Quick Start Guide for Technologica FluorImager software for use with Technlogica CFImager hardware Copyright 2006 2015 TECHNOLOGICA LIMITED. All
More informationMASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science
Student Name Date MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science 6.161 Modern Optics Project Laboratory Laboratory Exercise No. 3 Fall 2005 Diffraction
More informationMachine Vision Lyte-MV 2
Machine Vision Lyte-MV 2 The Lyte-MV 2 Range The Lyte-MV 2 provides a reliable industrial light source for a wide range of machine vision applications including triangulation, 3D inspection and alignment.
More informationA510S Operation Manual
A510S Operation Manual REV 1.1 1 Table of Contents 1 General Information 1-1 Description 1-2 Potential Operational Hazards 1-3 Technical Specifications 1-4 Instrument Overview 1-5 Function Summary 2 How
More informationBruker Dimension Icon AFM Quick User s Guide
Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com
More informationIRISYS ISI Series Imager Report Writing Software
IRISYS ISI 4604 4000 Series Imager Report Writing Software User Manual InfraRed Integrated Systems Ltd Park Circle, Tithe Barn Way Swan Valley Northampton NN4 9BG Tel: (0) 1604 594200 Fax: (0) 1604 594210
More informationMICROWAVE MOISTURE SENSOR WITH WIRELESS DATA TRANSMISSION FL-WAPP
- 1 MICROWAVE MOISTURE SENSOR WITH WIRELESS DATA TRANSMISSION FL-WAPP INDEX General Function Description P. 2 Technical Data P. 2 1. Software Installation P. 3 2. System Integration P. 3 2.1 Sensor as
More informationSTEM Spectrum Imaging Tutorial
STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3
More informationIntroduction. Lighting
&855(17 )8785(75(1'6,10$&+,1(9,6,21 5HVHDUFK6FLHQWLVW0DWV&DUOLQ 2SWLFDO0HDVXUHPHQW6\VWHPVDQG'DWD$QDO\VLV 6,17()(OHFWURQLFV &\EHUQHWLFV %R[%OLQGHUQ2VOR125:$< (PDLO0DWV&DUOLQ#HF\VLQWHIQR http://www.sintef.no/ecy/7210/
More informationKnife grinding machine K3 H/K
Knife grinding machine K3 H/K 01.11.2010 GS-Schleiftechnik, Leyher Str. 61 a, D-90431 Nürnberg Phone 0049/9193/4404, Fax 0049/9193/4391 e-mail: info@gs-de.eu 1 Table of Contents A. SAFETY... 3 A.1 THE
More informationLaser contrast scanner. Dimensioned drawing
Laser contrast scanner Dimensioned drawing en 06-2016/08 50110624-03 60mm 10-30 V DC Red light laser transmitter Various teach variants Small light spot adjustment via EasyTune Level adaptation for glossy
More informationQUANTUM Qflash T2 / X2 OPERATING INSTRUCTIONS
QUANTUM Qflash T2 / X2 OPERATING INSTRUCTIONS 1.0 DESIGNATIONS T2 AND X2 1. Removable Reflector, two positions Normal and Wide angle. 2. Flash-tube 2A. Modeling Lamp (for Model X2 only) 3. Bounce Head,
More informationSTPDRV-1 Stepper Motor Driver Data Sheet (R1.0) BFF Design Ltd
STPDRV-1 Stepper Motor Driver Data Sheet (R1.0) BFF Design Ltd 1. Introduction The BFF STPDRV-1 card is a bi-polar stepper motor driver. It is designed to drive the BFF Motorised Trim Wheel or other user-designed
More informationDigital Portable Overhead Document Camera LV-1010
Digital Portable Overhead Document Camera LV-1010 Instruction Manual 1 Content I Product Introduction 1.1 Product appearance..3 1.2 Main functions and features of the product.3 1.3 Production specifications.4
More informationW-Pressing Assembly. the top W roller thus creating a uniform and constant pressure between and throughout both top and lower W rollers.
PRODUCT: NUR Fresco HiQ 3200 ISSUE DATE: 10-Sep-02 SOFTWARE AFFECTED: YES NO TYPE OF CHANGE: MANDATORY COST PRODUCTION MAINTENANCE PROJECT MANAGER: SIGNATURE/DATE PART/ASSEMBLY: FCO KIT P/N: 10-0803 W-Pressing
More informationRadial Polarization Converter With LC Driver USER MANUAL
ARCoptix Radial Polarization Converter With LC Driver USER MANUAL Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Principle of the radial polarization
More informationGT-1050A 2 GHz to 50 GHz Microwave Power Amplifier
Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Giga-tronics GT-1050A Microwave Power Amplifier GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier Operation Manual
More informationTechnische Universität Chemnitz Physikalische Chemie. Laser Safety
Technische Universität Chemnitz Physikalische Chemie Laser Safety. Lasers and laser systems are classified by their ability to cause biological damage to the eye or skin during use. Class I: Lasers or
More information