In vitro optical performance of a new aberration-free intraocular lens
|
|
- Della Black
- 1 years ago
- Views:
Transcription
1 (2014) 28, & 2014 Macmillan Publishers Limited All rights reserved X/14 LABORATORY STUDY Optometry Research Group, Optics Department, University of Valencia, Valencia, Spain Correspondence: D Madrid-Costa, Optometry Research Group, Optics Department, University of Valencia, C/ Dr Moliner 50, Burjassot, Valencia, Spain Tel: ; Fax: Received: 21 January 2013 Accepted in revised form: 15 January 2014 Published online: 21 February 2014 In vitro optical performance of a new aberration-free intraocular lens Abstract Purpose To assess the optical performance of the new EnVista intraocular lens (IOL). Materials and methods Four aspheric IOLs were evaluated; the new EnVista is one amoung them. This IOL, similarly to the Z-Flex HB and the Bi-Flex AB ones, has a neutral aspheric design, whereas the fourth IOL under test (AcrySof IQ IOL SN60WF) presents a negative spherical aberration (SA). The IOL s aberration patterns were measured in vitro, by setting them up on an optical bench. From these aberration-pattern data, the modulation transfer function (MTF), the average modulation values, and the points spread function (PSF) were calculated. Furthermore, in order to assess the potential optical quality that these IOLs would yield once they are implanted, an average cornealaberration pattern was juxtaposed to the in-vitro profiles and the same parameters were calculated again. Results For the IOL-only scenario (ie, without including the corneal factor), it was the EnVista IOL, which is aberration-free that showed the higher MTF, PSF values. This was followed by the other two aberrationfree IOL models. However, when the effect of an average corneal pattern was also taken into consideration, the AcrySof IQ IOL SN60WF always outperformed the other neutral-asphericity IOLs. Conclusions The in-vitro optical performance of the EnVista IOL was good, but it decreases substantially in a whole-eye scenario, when the wavefront profile of an average cornea is added. Other designs with different degrees of SA should be considered for this IOL in order to surpass these results. (2014) 28, ; doi: /eye ; published online 21 February 2014 D Madrid-Costa, J Ruiz-Alcocer, T Ferrer-Blasco, S García-Lázaro and R Montés-Micó Introduction In order to improve their product offering, the specialized industry of intraocular lenses (IOLs) has launched periodic innovations into the market. One of the most important areas of innovation is related to wavefront technology. The introduction of this technology has highlighted the role of higherorder aberrations (HOAs), the spherical aberration (SA) being the most important one. Classic IOLs are spherical (consequently, with positive SA), whereas the latest models have aspheric designs. Within this aspheric-lens category, some of the IOLs have had its intrinsic SA wholly corrected (ie, they have no SA, which is why they are usually designated: aberration-free IOLs), whereas other aspheric IOLs have been devised to compensate (totally or partially) the corneal SA (ie, the IOL itself has a negative SA), which is why they are often called aberrationcorrecting IOLs. At the same time, the materials from which the IOL is manufactured represent the other main field of innovation; inthisrespect,new materials are continuously being developed. Nowadays, the most widely implanted IOLs are made of hydrophobic acrylic materials, as these lenses are soft and foldable and, as a result, only a small incision is required for it to be inserted. These small incisions do not usually need sutures, thus allowing faster patient recovery times, minimizing residual astigmatism and providing better visual outcomes. 1 3 The aim of this study is to assess the optical performance of a novel hydrophobic acrylic IOL recently launched into the market.thisnewlensistheenvista (Bausch and Lomb, Rochester, NY, USA). The optical performance of this new IOL was compared with three other commercially available IOLs.
2 Optical quality of a new aberration-free IOL 615 Materials and methods Aberration-free aspheric IOLs Among the four IOLs evaluated in this study, three of them look for attaining zero SA. The EnVista IOL is marketed as an aberration-free IOL. This lens is manufactured in hydrophobic acrylic material that incorporates a UV filter. This hydrophobic acrylic onepiece monofocal IOL has a total diameter of 12.5 mm, with a 6.0 mm optical zone. The lens is available with an optical power that ranges from 0.0 to 34.0 D, in steps of 0.5 D (within the þ 10.0 D to þ 30.0 D interval) or of 1 D (within the 0.0 to þ 9.0 D and the þ 31.0 to þ 34.0 D intervals). The other two aspheric aberration-free lenses under assessment were the Bi-flex AB (Medicontour, Geneva, Switzerland) and the Z-Flex HB 860FAB IOLs (Medicontour). The Bi-Flex AB is a one-piece IOL, manufactured from a combination of a hydrophilic and a hydrophobic copolymer. It includes a UV-absorber filter. The IOL s total diameter is 13.0 mm, with a 6.0 mm optical zone. The lens is available with an optical power that ranges from 0.0 to 45.0 D, in steps of 0.5 D (within the þ 10.0 D to þ 30.0 D interval) or 1 D (within the 0.0 to þ 9.0 D and the þ 31.0 to þ 45.0 D intervals). Finally, as for the Z-flex HB 860FAB, it is a hydrophobic acrylic IOL that incorporates a blue filter. Its total diameter is 13.0 mm, with a 6.00 mm optical zone. This lens is available with an optical power that ranges from 10.0 to 45.0 D, in steps of 0.5 D (within the þ 10.0 D to þ 30.0 D interval) or 1 D (within the 10.0 to þ 9.0 D and the þ 31.0 to þ 45.0 D intervals). Aberration-correcting IOL The AcrySof IQ IOL SN60WF (Alcon Laboratories, Fort Worth, TX, USA) IOL features an aspheric modification to the posterior optic surface. The aim of this design is to provide a negative amount of SA ( 0.20 mm for a 6-mm pupil), thus compensating part of the positive SA that the eye shows after crystalline lens extraction. The lens is a foldable single-piece hydrophobic acrylic IOL that incorporates both a UV and a blue-blocker filter (yellow tint). The optical zone is 6.0 mm in diameter, whereas the lens total width is 13.0 mm. The optical power of AcrySof IQ IOL SN60WF ranges from þ 6.0 to þ 30.0 D, in 0.5 D steps. Image quality assessment To assess the IOLs image quality we used the NIMO instrument (LAMBDA-X, Nivelles, Belgium), which includes an optical bench, together with its software version Figure 1 shows a schematic layout of the Figure 1 Schematic layout of the NIMO TR1504 (LAMBDA-X). This instrument consists of a cold cathode tube backlight sources that emits green light at 546 nm, incorporating an additional diffuser and a ±10-nm bandwidth filter to the source to homogenize and limit the spectral width of the light beam. A liquid crystal display (LCD) placed at the focal length of the lens L1 that is the lens responsible for collimating the light beam. Two lenses, L2 and L3, which form an image on camera through a telecentric arrangement and a CCD camera with a native pixel resolution of pixels corresponding to a spacing of 69 pixel/mm or 1761 pixel/inch. The high resolution of the camera is directly related to the high instrument resolution (36 mm). The intraocular lens under study is placed in the object plane of the instrument, between lens L1 and L2. NIMO TR1504. The working principle of this instrument is based on a phase-shifting schlieren techonology. 4 By combining the principle of schlieren imaging with the phase-shifting method, the NIMO instrument allows the measurement of light beam deviations, which can be used to calculate the wavefront analysis considering the 36 Zernikes coefficients. This technology has been shown to effectively measure in vitro optical quality of intraocular lenses. 5 The apparatus complies with the International Standard Organization (ISO) All IOLs were measured while being immersed in a saline solution whose composition was milliequivalents per milliliter of NaCl (Laboratoires Sterop SA, Anderlecht, Belgium). The cuvettes or wet cells used to hold the IOLs and the saline solution in place during the measurements have been verified by means of an interferometer; and were shown to have a power o0.005 D. This additional cross-check on the wet cells was carried out to rule out potential interferences with the measurement. Moreover, accurate power measurements are only possible if the setup has been thoroughly calibrated, which is why the instrument was calibrated for each measurement.
3 616 Optical quality of a new aberration-free IOL All measurements were recorded for a 5.0 mm aperture, as it generally corresponds to the average pupil size of patients who are over 60 years of age with under mesopic conditions. 7 In this study, Zernikes coefficients values were retained as the average of five measurements. Image quality parameters The IOLs optical quality was quantified by means of the modulation transfer function (MTF) and the point spread function (PSF). The MTF represents the contrast attenuation induced by an optical system as a function of the spatial frequency or target size. This fall in image contrast, which is sharper for higher spatial frequencies, 8 can lead to a decline in the optical system s visual performance. 9 In the current study, the mean one-dimensional MTF was calculated as the average over all orientation of the two-dimensional MTF. Both the MTF and the PSF were computed from the experimental wavefront data using a custom-made MATLAB-based application (Mathworks, Nantick, MA, USA). In order to be able to compare the MTFs yielded by the four IOLs, we took the average modulation value as optical quality metrics. 10,11 The average modulation has been considered to be the modulation averaged across all frequencies ranging from 0.0 to cycles per millimetre; this value has been shown to be proportional to the area under the MTF curve between 0.0 and cycles per millimetre. As it was mentioned before, the image quality of the lenses alone was measured in vitro. Furthermore, to simulate the optical quality that the IOLs could provide after implantation, a whole-eye wavefront aberration map was simulated by adding a corneal aberration profile to the IOLs aberration profiles, and from these data new MTFs and PSFs were computed. Previous studies have shown that corneal HOAs remain rather stable with age From those studies, an average corneal pattern could be obtained. To combine (add) this average corneal pattern with each IOL s pattern we used the abovementioned custom-made MATLAB-based application. The results yielded by the IOLs alone and by the IOL þ cornea systems were compared in order to have an estimate of the optical quality that the implanted IOLs could provide. Results Figure 2 presents the HOAs Zernike coefficients for each of the IOLs alone, measured in vitro, as well as the Zernike coefficients corresponding to the averaged cornea. With this figure it is possible to estimate how the Zernike modes would be modified when the corneal aberration coefficients were added to the HOA profile of Figure 2 Zernike coefficients of the higher order aberrations corresponding to the intraocular lenses (IOL) alone and the intraocular lenses when an average cornea was added (IOL þ cornea system).
4 Optical quality of a new aberration-free IOL 617 each IOL. In the top of the figure it can be observed that, within all the Zernike modes, the most dramatic differences can be founded for the Z(4,0) (SA) coefficient if the AcrySof IQ IOL SN60WF is compared to the other three lenses. The figure shows how the AcrySof IQ IOL SN60WF would neutralize the positive SA while the other three would show a positive increment in the SA coefficient when they were combined with the averaged cornea. How this coupling effect would impact on the optical quality of the patients is showed in Figures 3 and 4. Figures 3 and 4 and show the MTF curves and the PSF for the four IOLs alone and for the four IOL þ cornea systems, respectively. From Figure 3 it is possible to observe that the EnVista IOL that provided the best MTF curve when the IOLs were assessed by themselves, without including the effect of corneal aberrations. This is followed by the other 2 aberration-free IOLs. The MTF for the AcrySof IQ IOL SN60WF alone was lower than for the other three IOLs. This is expected because this lens incorporates a negative amount of SA ( 0.20 mm for a 6-mm pupil) into its design. However, when the cornea was added (IOL þ cornea), the situation reversed completely, since out of the four IOLs under study it was the AcrySof IQ IOL SN60WF that yielded the best results: for all spatial frequencies the MTF values obtained for the AcrySof IQ IOL SN60WF þ cornea system were higher than for the other three systems. It is also interesting to observe that the AcrySof IQ IOL SN60WF was the only lens whose MTF curve improved upon addition of the cornea, which is comparable to the postimplatation status of the IOL inside of the eye. Table 1 presents the average modulation values for the four IOLs under study in both scenarios (IOL by itself or combined with the average corneal profile). The data reveal that the EnVista IOL offered the best results when the lenses were considered without the cornea, whereas when the IOLs were considered as part of an IOL þ cornea system, it was the AcrySof IQ IOL SN60WF that yielded the best results out of the four IOLs. Discussion In the current study we have evaluated a new aberrationfree IOL that it has been launched into the market. This new lens is the EnVista IOL and in the present study it has been experimentally compared by means of in-vitro measurements of their optical quality with three other aspheric monofocal IOLs: AcrySof IQ IOL SN60WF, Bi-Flex AB and Z-Flex HB 860FAB. The EnVista, the Z-Flex and the Bi-Flex IOLs have all neutral aspheric designs that is, they are IOLs with no intrinsic SA (aberration-free), whereas the AcrySof IQ IOL SN60WF is an aberration-correcting aspheric IOL that has been designed to have a negative amount of SA in an attempt to partially compensate for the positive ocular SA that the eye normally has once the lens has been extracted. Optical quality metrics (MTF, PSF) were computed both Figure 3 Modulation transfer function curves of the four intraocular lenses (IOL) measured in vitro, both alone (without a cornea) and when the average corneal pattern was added (IOL þ cornea system).
5 618 Optical quality of a new aberration-free IOL Figure 4 Point spread function representation of the four intraocular lenses (IOL) measured in vitro, both alone (without a cornea) and when the average corneal pattern was added (IOL þ cornea system). Table 1 Average modulation values for the four IOLs under study, both alone (without a cornea) and when the average corneal pattern was added (IOL þ cornea system), for a 5.0 mm aperture AcrySof IQ Bi-flex Z-Flex Envista IOL IOL þ cornea for the IOLs alone as well as for a hypothetical optical system comprising each IOL plus an average cornea, whose aberration data map was taken from the literature This latter scenario was included in order to estimate the performance of the IOL once it is implanted in the aphakic eye. Figures 3 and 4 show the MTF curves and the PSF, respectively. These figures present the results of the IOLs measured in vitro and the IOLs when the average corneal HOAs were added. Both the MTF curves and the PSF suggest that when the IOLs are measured without cornea, the aberration-free IOLs performed better than the aberration-correcting IOL. However, when the HOAs of the cornea were added to the aberration profile of the IOLs, the aberration-correcting IOL was the only IOL that improves its optical quality when compared with the situation without cornea. All these results clearly show the impact of SA upon the optical performance of an IOL: one can see that if the cornea is not taken into account, the aberration-free IOLs present the lowest amounts of SA and higher MTF and PSF values. Conversely, when the cornea was added, the aberration-correcting IOL present the lowest amount of SA and the best results. In terms of visual quality, different studies have reported the importance of SA, showing that an increase in the total amount of the eye s SA results in an increase of the total amount of HOAs, which, in turn, decrease the patients visual function. 15,16 The dramatic changes in the results observed when the corneal profile was added to the IOL profile, remark the importance of including a corneal profile to the IOL profile in order to represent, as near as possible to a real life scenario, the potential optical performance of an IOL once it is implanted in the aphakic eye. In the current study we have quantified the optical performance by means of the MTF. In relation to visual quality, a decrease in the MTF entails variations in image contrast that could reduce an optical system s image quality, thus affecting the patients visual performance. 9 Moreover, Felipe et al 17 performed a study to determinate whether there is a correlation between IOL optics and visual parameters. In order to do that, the authors analysed the optical quality of the IOL by average modulation parameter and the patient s visual quality by the visual acuity. The authors found that the visual acuity varied 0.24 decimal units per 10 units of average modulation variation under mesopic conditions. Table 1 shows the average modulation values of the four IOLs for a 5-mm aperture (it generally corresponds to the average pupil size that patients over 60 years of age have under mesopic conditions. 7 ). In this study, when the cornea was added to the IOLs, it was the AcrySof IQ IOL SN60WF (aberration-correcting) that provided the best results, showing a substantial difference with the other three IOLs (see Table 1). It is important to take into account these results and the differences across the lenses due to the reported impact that they would have on the visual quality of the patients. 17 For example, considering the study of Felipe et al, 17 the average modulation values obtained in the current study could be equivalent to visual acuity values under mesopic conditions about 20/16 for IQ, 20/20 for Z-Flex and 20/25 for Bi-Flex and EnVista. Therefore, the correction of the SA under the mentioned conditions improves the optical quality of the IOL þ cornea system and consequently, it is possible to suggest that the visual quality of the patients would also be improved in regular tasks performed in mesopic
6 Optical quality of a new aberration-free IOL 619 conditions such night-driving. These outcomes have showed the importance of correcting the SA of the IOL þ cornea system in order to attain an optimum visual performance. It should be also noted that we used an average corneal profile that is a theoretical representation of the corneal profile of the population. This allows to estimates the optical performance when the IOLs profile is added to the averaged cornea. Nevertheless, the corneal profile could slightly vary within normal individuals but largely within patients that have undergone corneal refractive surgeries For these cases, due to the HOA profiles of the IOL are also showed and the surgeons could assess the corneal profile of each patient, it would be possible to estimate which is the best IOL to each individual case. At the same time, it is also important to bear in mind that various factors related to the surgical implantation of the IOLs could also have an impact upon the patient s visual quality. One of the most important factors is the required accuracy of the centration of the lens Decentration of an IOL can lead to other common complication: tilt. In fact, tilt and decentration of the implanted IOLs could limit or even cancel out the theoretical advantages of aspheric IOLs In summary, the results of the current study show that the EnVista IOL studied provides a good optical performance, however, the optical quality results for this IOL were worse than for the aberration-correcting IOL. Hence, it seems that it would be interesting to consider other designs for this new IOL with other degrees of SA in order to optimize its results in a wider spectrum of patients. Summary What was known before K In order to improve their product offering, the specialized industry of intraocular lenses (IOLs) has launched periodic innovations into the market. Two of the most important areas of innovation are related to the role of higher-order aberrations (HOAs), being the spherical aberration (SA) the most important one, and the materials from which the IOLs are manufactured. Recently, it has been launched into the market a new IOL manufactured from a novel hydrophobic acrylic material with an aberration-free design; this new lens is the EnVista. The optical quality related to the visual quality of patients with these new IOL had not been studied. What this study adds K The in vitro optical performance of the EnVista IOL was good, but it decreases substantially in a whole-eye scenario, when the wavefront profile of an average cornea is added. Other designs with different degrees of SA should be considered for this new IOL in order to surpass these results. Conflict of interest The authors declare no conflict of interest. Acknowledgements This research was supported in part by a Research Grant (#SAF #) of the Ministerio de Ciencia e Innovación (Ministry of Science and Innovation) awarded to Robert Montés-Micó. References 1 Olson RJ, Crandall AS. Prospective randomized comparison of phacoemulsification cataract surgery with a 3.2-mm vs a 5.5-mm sutureless incision. Am J Ophthalmol 1998; 125: Oshika T, Nagahara K, Yaguchi S, Emi K, Takenaka H, Tsuboi S et al. Three year prospective, randomized evaluation of intraocular lens implantation through 3.2 and 5.5 mm incisions. J Cataract Refract Surg 1998; 24: Zheng L, Merriam JC, Zaider M. Astigmatism and visual recovery after large incision extracapsular cataract surgery and small incisions for phakoemulsification. Trans Am Ophthalmol Soc 1997; 95: Joannes L, Dubois F, Legros JC. Phase-shifting schlieren: highresolution quantitative schlieren that uses the phaseshifting technique principle. Appl Opt 2003; 42(25): Pérez-Vives C, Ferrer-Blasco T, Madrid-Costa D, García-Lázaro S, Montés-Micó R. Optical quality comparison of conventional and hole-visian implantable collamer lens at different degrees of decentering. Am J Ophthalmol 2013; 156(1): International Organization for Standardization. Ophthalmic Implants Intraocular Lenses Part 2: Optical Properties and Test Methods. ISO: Geneva, Switzerland, 1999, ISO Winn B, Whitaker D, Elliott DB, Phillips NJ. Factors affecting light-adapted pupil size in normal human subjects. Invest Ophthalmol Vis Sci 1994; 35: Gatinel D, Pagnoulle C, Houbrechts Y, Gobin L. Design and qualification of a diffractive trifocal optical profile for intraocular lenses. J Cataract Refract Surg 2011; 37: Artigas JM, Menezo JL, Peris C, Felipe A, Díaz-Llopis M. Image quality with multifocal intraocular lenses and the effect of pupil size; comparison of refractive and hybrid refractive-diffractive designs. J Cataract Refract Surg 2007; 33: Artigas JM, Peris C, Felipe A, Menezo JL, Sánchez-Cortina I, López-Gil N. Modulation transfer function: rigid versus foldable phakic intraocular lenses. J Cataract Refract Surg 2009; 35: Marsack JD, Thibos LN, Applegate RA. Metrics of optical quality derived from wave aberration predict visual performance. JVis2004; 4: Oshika T, Klyce SD, Applegate RA, Howland HC. Changes in corneal wavefront aberrations with aging. Invest Ophthalmol Vis Sci 1999; 40: Guirao A, Redondo M, Artal P. Optical aberrations of the human cornea as a function of age. J Opt Soc Am A Opt Image Sci Vis 2000; 17:
7 620 Optical quality of a new aberration-free IOL 14 Wang L, Dai E, Koch DD, Nathoo A. Optical aberrations of the human anterior cornea. J Cataract Refract Surg 2003; 29: Li J, Xiong Y, Wang N, Li S, Dai Y, Xue L et al. Effects of spherical aberration on visual acuity at different contrasts. J Cataract Refract Surg 2009; 35: Barbero S, Marcos S, Jimenez-Alfaro I. Optical aberrations of intraocular lenses measured in vivo and in vitro. J Opt Soc Am A Opt Image Sci Vis 2003; 20: Felipe A, Pastor F, Artigas JM, Diez-Ajenjo A, Gené A, Menezo JL. Correlation between optics quality of multifocal intraocular lenses and visual acuity: tolerance to modulation transfer function decay. J Cataract Refract Surg 2010; 36: Madrid-Costa D, Ruiz-Alcocer J, Pérez-Vives C, Ferrer-Blasco T, López-Gil N, Montés-Micó R. Visual simulation through different intraocular lenses using adaptive optics: effect of tilt and decentration. J Cataract Refract Surg 2012; 38: Madrid-Costa D, Pérez-Vives C, Ruiz-Alcocer J, Albarrán-Diego C, Montés-Micó R. Visual simulation through different intraocular lenses in patients with previous myopic corneal ablation using adaptive optics: effect of tilt and decentration. J Cataract Refract Surg 2012; 38: Ruiz-Alcocer J, Pérez-Vives C, Madrid-Costa D, López-Gil N, Montés-Micó R. Effect of simulated IOL tilt and decentration on spherical aberration after hyperopic LASIK for different intraocular lenses. J Refract Surg 2012; 28: Madrid-Costa D, Ruiz-Alcocer J, García-Lázaro S, Albarrán-Diego C, Montés-Micó R. Visual performance of the Akreos Adapt AO intraocular lens in patients with different corneal profiles measured with an adaptive optics visual simulator. Br J Ophthalmol 2012; 96: Montés-Micó R, López-Gil N, Pérez-Vives C, Bonaque S, Ferrer-Blasco T. In vitro optical performance of nonrotational symmetric and refractive-diffractive aspheric multifocal intraocular lenses: Impact of tilt and decentration. J Cataract Refract Surg 2012; 38: Gimbel HV, Condon GP, Kohnen T, Olson RJ, Halkiadakis I. Late in-the-bag intraocular lens dislocation: incidence, prevention, and management. J Cataract Refract Surg 2005; 31: Carlson AN, Stewart WC, Tso PC. Intraocular lens complications requiring removal or exchange. Surv Ophthalmol 1998; 42: Atchison DA. Design of aspheric intraocular lenses. Ophthalmic Physiol Opt 1991; 11: Altamann GE, Nichamin LD, Lane SS, Pepose JS. Optical performance of 3 intraocular lens designs in the presence of decentration. J Cataract Refract Surg 2005; 31: Montés-Micó R, Ferrer-Blasco T, Cerviño A. Analysis of the possible benefits of aspheric intraocular lenses: review of the literature. J Cataract Refract Surg 2009; 35:
HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design
HOYA aspherical IOL with ABC (Aspheric Balanced Curve) Design Contents Basics of asphericity Visual quality and aspheric IOL Features of HOYA ABC Design 2 What is asphericity? Deviating from the spherical
Visual Simulation: application to monofocal intraocular lens analysis
ARTICLE Visual Simulation: application to monofocal intraocular lens analysis Alberto Domínguez Vicent, OD; Cari Pérez-Vives, MSc; Lurdes Belda-Salmerón, MSc; César Albarrán-Diego, MSc; Santiago García-Lázaro,
The Aberration-Free IOL:
The Aberration-Free IOL: Advanced Optical Performance Independent of Patient Profile Griffith E. Altmann, M.S., M.B.A.; Keith H. Edwards, BSc FCOptom Dip CLP FAAO, Bausch & Lomb Some of these results were
Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes
European Journal of Ophthalmology / Vol. 18 no. 5, 2008 / pp. 728-732 Comparison of higher order aberrations with spherical and aspheric IOLs compared to normal phakic eyes M. RĘKAS, K. KRIX-JACHYM, B.
The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1
The design is distinctive. The outcomes are clear. Defocus tolerance 1 Glistening-free performance 1,2 Predictable outcomes 1 The clear choice for consistent visual excellence. For over 165 years Bausch
Modulation transfer function: Rigid versus foldable phakic intraocular lenses
LABORATORY SCIENCE Modulation transfer function: Rigid versus foldable phakic intraocular lenses José M. Artigas, PhD, Cristina Peris, MD, PhD, Adelina Felipe, PhD, José L. Menezo, MD, PhD, FEBO, Isaias
Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations
Corneal Asphericity and Retinal Image Quality: A Case Study and Simulations Seema Somani PhD, Ashley Tuan OD, PhD, and Dimitri Chernyak PhD VISX Incorporated, 3400 Central Express Way, Santa Clara, CA
Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN
Crystalens AO: Accommodating, Aberration-Free, Aspheric Y. Ralph Chu, MD Chu Vision Institute Bloomington, MN Financial Disclosure Advanced Medical Optics Allergan Bausch & Lomb PowerVision Revision Optics
Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12
Choices and Vision Jeffrey Koziol M.D. How does the eye work? What is myopia? What is hyperopia? What is astigmatism? What is presbyopia? How the eye works Light rays enter the eye through the clear cornea,
Update on Aspheric IOL Technology
Peer-Reviewed Literature: Update on Aspheric IOL Technology Editor: Ming Wang, MD, PhD, Clinical Associate Professor of Ophthalmology at the University of Tennessee and Director of the Wang Vision Institute
*Simulated vision. **Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most
Simulated vision. Individual results may vary and are not guaranteed. Visual Performance When It s Needed Most The aspheric design of the AcrySof IQ IOL results in improved clarity and image quality. The
Raise your expectations. Deliver theirs.
66 EXTENDED RANGE OF VISION MONOFOCAL-LIKE DISTANCE Raise your expectations. Deliver theirs. Now you can give your patients the best of both worlds with the first and only hybrid designed monofocal-multifocal
Reproducibility of contact lens power measurements using the phase shifting schlieren method
Reproducibility of contact lens power measurements using the phase shifting schlieren method Luc Joannes, Tony Hough, Xavier Hutsebaut, Xavier Dubois, Renaud Ligot, Bruno Saoul, Philip Van Donink, Kris
Clinical Update for Presbyopic Lens Options
Clinical Update for Presbyopic Lens Options Gregory D. Searcy, M.D. Erdey Searcy Eye Group Columbus, Ohio The Problem = Spherical Optics Marginal Rays Spherical IOL Light Rays Paraxial Rays Spherical Aberration
Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation
J CATARACT REFRACT SURG - VOL 31, SEPTEMBER 2005 Comparison of contrast sensitivity and color discrimination after clear and yellow intraocular lens implantation Antonio Rodríguez-Galietero, MD, PhD, FEBO,
The Blueprint for Improved Image Quality
Insert to January 2009 The Blueprint for Improved Image Quality The function and design of the AcrySof IQ IOL. Supported by an educational grant from Alcon Laboratories, Inc. The Function and Design of
Treatment of Presbyopia during Crystalline Lens Surgery A Review
Treatment of Presbyopia during Crystalline Lens Surgery A Review Pierre Bouchut Bordeaux Ophthalmic surgeons should treat presbyopia during crystalline lens surgery. Thanks to the quality and advancements
What is Wavefront Aberration? Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World?
Custom Contact Lenses For Vision Improvement Are They Feasible In A Disposable World? Ian Cox, BOptom, PhD, FAAO Distinguished Research Fellow Bausch & Lomb, Rochester, NY Acknowledgements Center for Visual
ROTATIONAL STABILITY MAKES THE DIFFERENCE
The Bi-Flex platform the proven platform of Excellence 01 Proven Stability less than 02 Optimal biomaterials 2 degrees long term rotation 03 Posterior Toric Lens surface with marks indicating the flat
Evolution of Diffractive Multifocal Intraocular Lenses
Evolution of Diffractive Multifocal Intraocular Lenses Wavefront Congress February 24, 2007 Michael J. Simpson, Ph.D. Alcon Research, Ltd., Fort Worth, Texas Presentation Overview Multifocal IOLs two lens
10/25/2017. Financial Disclosures. Do your patients complain of? Are you frustrated by remake after remake? What is wavefront error (WFE)?
Wavefront-Guided Optics in Clinic: Financial Disclosures The New Frontier November 4, 2017 Matthew J. Kauffman, OD, FAAO, FSLS STAPLE Program Soft Toric and Presbyopic Lens Education Gas Permeable Lens
Research Article Intraocular Telescopic System Design: Optical and Visual Simulation in a Human Eye Model
Hindawi Ophthalmology Volume 27, Article ID 63793, 8 pages https://doi.org/.55/27/63793 Research Article Intraocular Telescopic System Design: Optical and Visual Simulation in a Human Eye Model Georgios
Optical aberrations of intraocular lenses measured in vivo and in vitro
Barbero et al. Vol. 20, No. 10/October 2003/J. Opt. Soc. Am. A 1841 Optical aberrations of intraocular lenses measured in vivo and in vitro Sergio Barbero and Susana Marcos Instituto de Óptica, Consejo
PROGRESSIVE VISION WITHIN FULL ACCOMMODATIVE RANGE
PROGRESSIVE VISION WITHIN FULL ACCOMMODATIVE RANGE PROGRESSIVE VISION Progressive vision within full accomodative range 03 Suitable for sub 2 mm MICS 01 PAD Progressive- Apodized-Diffractive 02 Aspheric
Abetter understanding of the distribution of aberrations in
Predicting the Optical Performance of Eyes Implanted with IOLs to Correct Spherical Aberration Juan Tabernero, 1 Patricia Piers, 2 Antonio Benito, 1 Manuel Redondo, 3 and Pablo Artal 1 PURPOSE. To use
CONTENTS. HYDROPHILIC IOLs. Bi-Flex PIL-MA hydrophilic preloaded P / PY 677PT / PTY 677PMY 677PMTY. Bi-Flex HL...
Bi-Flex IOL RANGE CONTENTS HYDROPHILIC IOLs Bi-Flex PIL-MA hydrophilic preloaded... 1 677P / PY 677PT / PTY 677PMY 677PMTY Bi-Flex HL... 2 677AB / ABY Bi-Flex T... 3 677TA / TB 677TAY / TBY Bi-Flex M...
Retinal image quality is determined by the combined optical. In Vivo Chromatic Aberration in Eyes Implanted With Intraocular Lenses
Visual Psychophysics and Physiological Optics In Vivo Chromatic Aberration in Eyes Implanted With Intraocular Lenses Pablo Pérez-Merino, 1 Carlos Dorronsoro, 1 Lourdes Llorente, 1 Sonia Durán, Ignacio
Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens
ARTICLE Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens José F. Alfonso, MD, PhD, Luis Fernández-Vega, MD, PhD, M. Begoña
OptiSpheric IOL. Integrated Optical Testing of Intraocular Lenses
OptiSpheric IOL Integrated Optical Testing of Intraocular Lenses OPTICAL TEST STATION OptiSpheric IOL ISO 11979 Intraocular Lens Testing OptiSpheric IOL PRO with in air tray on optional instrument table
Ocular Scatter. Rayleigh Scattering
Ocular Scatter The are several sources of stray light in the eye including the cornea, transmission through the iris and the crystalline lens. Cornea tends to have Rayleigh Scatter Lens follows inverse
Centre Hospitalier Universitaire et Psychiatrique de Mons-Borinage. B-Flex Multifocal. Dr Emmanuel Van Acker Belgium
Centre Hospitalier Universitaire et Psychiatrique de Mons-Borinage B-Flex Multifocal Dr Emmanuel Van Acker Belgium Comparison of clinical outcomes and patient satisfaction after implantation of two different
Development of a Calibration Standard for Spherical Aberration
Development of a Calibration Standard for David C. Compertore, Filipp V. Ignatovich, Matthew E. Herbrand, Michael A. Marcus, Lumetrics, Inc. 1565 Jefferson Road, Rochester, NY (United States) ABSTRACT
THE ASPHERIC PRELOADED INJECTION
New from STAAR for Safe and Easy Implantation THE ASPHERIC PRELOADED INJECTION S Y S T E M PRELOADED INJECTION SYSTEM FOR SIMPLE AND EASY IOL DELIVERY SAFE The KS3-Ai features an aspheric IOL preloaded
Transferring wavefront measurements to ablation profiles. Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich
Transferring wavefront measurements to ablation profiles Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich corneal ablation Calculation laser spot positions Centration Calculation
The pupil of the eye is a critical limiting factor in the optics
Pupil Location under Mesopic, Photopic, and Pharmacologically Dilated Conditions Yabo Yang, 1,2 Keith Thompson, 3 and Stephen A. Burns 1 PURPOSE. To determine whether there are systematic changes in pupil
Optical Quality of the Eye in Subjects with Normal and Excellent Visual Acuity METHODS. Subjects
Optical Quality of the ye in Subjects with Normal and xcellent Visual Acuity loy A. Villegas, ncarna Alcón, and Pablo Artal From the Laboratorio de Optica, Departamento de Fisica, Universidad de Murcia,
Through-Focus Optical Bench Performance of Extended Depth-of-Focus and Bifocal Intraocular Lenses Compared to a Monofocal Lens
ORIGINAL ARTICLE Through-Focus Optical Bench Performance of Extended Depth-of-Focus and Bifocal Intraocular Lenses Compared to a Monofocal Lens Young-Sik Yoo, MD; Woong-Joo Whang, MD; Yong-Soo Byun, MD,
What s New in Ocular Biomechanics?
What s New in Ocular Biomechanics? The International Congress of Wavefront Sensing & Optimized Refractive Corrections Wavefront Course January 28, 2006 Torrence A. Makley Research Professor Department
ORIGINAL ARTICLES. Image Metrics for Predicting Subjective Image Quality
1040-5488/05/8205-0358/0 VOL. 82, NO. 5, PP. 358 369 OPTOMETRY AND VISION SCIENCE Copyright 2005 American Academy of Optometry ORIGINAL ARTICLES Image Metrics for Predicting Subjective Image Quality LI
NEW. AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients
Premium Trifocal MICS OVDs IOLs NEW AT LISA tri 839MP and AT LISA tri toric 939MP from ZEISS The innovative trifocal IOL concept providing True Living Vision to more patients Trifocal toric IOL The moment
ORIGINAL ARTICLE. Power Profiles of Commercial Multifocal Soft Contact Lenses. Eon Kim*, Ravi C. Bakaraju, and Klaus Ehrmann*
1040-5488/16/0000-0000/0 VOL. 00, NO. 00, PP. 00Y00 OPTOMETRY AND VISION SCIENCE Copyright * 2016 American Academy of Optometry ORIGINAL ARTICLE Power Profiles of Commercial Multifocal Soft Contact Lenses
Glistening-Free Hydrophobic Acrylic IOL. Glistening-Free Hydrophobic Acrylic IOL
Glistening-Free Hydrophobic Acrylic IOL Proprietary Polymer Composition Packaged Dry UD Aspheric Optic Optimized Refractive Index All SquareTM Enhanced Edge Largest Diopter Range -05.00D to 42.00D Modified
Theoretical Considerations
Bedeutung der Linsenposition für die optische Abbildugsqualität- Ein neues Messverfahren mittels Purkinje Reflexbildern SA(-) Acri.Tec 35A Hoya FY60AD Tecnis ZCB00 AcrySof IQ U. Mester, T. Sauer, H.Kaymak
Theoretical Comparison of Aberrationcorrecting Customized and Aspheric Intraocular Lenses
Theoretical Comparison of Aberrationcorrecting Customized and Aspheric Intraocular Lenses Patricia A. Piers; Henk A. Weeber; Pablo Artal, PhD; Sverker Norrby, PhD ABSTRACT PURPOSE: To assess the performance
Trust your eyes. Presbyopic treatment methods on the cornea. PresbyMAX Decision criteria and patient s acceptance
Trust your eyes. Directory Presbyopic treatment methods on the cornea PresbyMAX The Principle PresbyMAX Expectations and Key Factors PresbyMAX Decision criteria and patient s acceptance PresbyMAX Upcoming
UIN: IOL Date of Prep: April Avansee Preload1P. Technical Slide Kit April 2017
Avansee Preload1P Technical Slide Kit April 2017 Agenda 1. Avansee Preload1P: Summary of product characteristics 2. Excellent IOL characteristics 2.1 Glistening-free optic 2.2 360 degree square edge 2.3
Corneal Mapping over the Contact Lens. Challenge: Getting the Most out of Soft Contact Lens Multifocals
Contact Lens Management of the Challenging Patient Disclosures: Alcon Bausch + Lomb SpecialEyes Valley Contax Vistakon Contact Lens Challenges Matthew J. Lampa, OD, FAAO lampa@pacificu.edu Challenge: Getting
Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses
ARTICLE Comparison of Contrast Sensitivity in Eyes with Toric and Aspheric Intraocular Lenses Francisco Pastor-Pascual, PhD 1 ; Aitor Lanzagorta-Aresti, MD 1 ; Pablo Alcocer-Yuste, MD 1 ; Cristina Puchades-Lladró
Image Quality of the Human Eye. Susana Marcos, Ph.D.
Image Quality of the Human Eye Susana Marcos, Ph.D. Factors Contributing to Retinal Image Degradation The eye is an optical instrument that projects scenes of the visual world onto the retina. It has been
Generation of third-order spherical and coma aberrations by use of radially symmetrical fourth-order lenses
López-Gil et al. Vol. 15, No. 9/September 1998/J. Opt. Soc. Am. A 2563 Generation of third-order spherical and coma aberrations by use of radially symmetrical fourth-order lenses N. López-Gil Section of
Forget Most Everything! The Surgical Management of Presbyopia 2/23/2016. Refraction vs. Diffraction. Presbyopic IOL s Patient Expectations
The Surgical Management of Presbyopia Presbyopic IOL s 2011 B I L L T U L L O, O D Patient Expectations What they say is I want to be able to read The Center of a Presbyope s World What they want is Accommodation
Optical Characteristics of Next Generation Dual Optic IOL
Optical Characteristics of Next Generation Dual Optic IOL Scott Evans, MD Sanjeev Kasthurirangan, PhD Val Portney, PhD Financial Disclosures Scott Evans is an employee of Abbott Medical Optics Inc. Sanjeev
ORIGINAL ARTICLE. Metrics of Retinal Image Quality Predict Visual Performance in Eyes With 20/17 or Better Visual Acuity
1040-5488/06/8309-0635/0 VOL. 83, NO. 9, PP. 635 640 OPTOMETRY AND VISION SCIENCE Copyright 2006 American Academy of Optometry ORIGINAL ARTICLE Metrics of Retinal Image Quality Predict Visual Performance
The eye-care profession s advancement from cataract
Cataract surgery and optimal spherical aberration: as simple as you think? Shamrozé Khan,* BSc, OD; Guillermo Rocha, MD, FRCSC ABSTRACT RÉSUMÉ This paper reviews the optics of higher-order and spherical
Comparison of through-focus image quality across five presbyopia-correcting intraocular lenses (an American Ophthalmological Society Thesis)
Washington University School of Medicine Digital Commons@Becker Open Access Publications 2011 Comparison of through-focus image quality across five presbyopia-correcting intraocular lenses (an American
University of Groningen. Young eyes for elderly people van Gaalen, Kim
University of Groningen Young eyes for elderly people van Gaalen, Kim IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the
The complete choice in refractive lens solutions
The complete choice in refractive lens solutions The VeriFlex phakic IOL is part of Advanced Medical Optics range of high-contrast, high-quality refractive lens solutions. In addition to excellent training
Página 1 de 9 TopPage > Eye Care > Diagnostic > Wave-Front Analyzer KR-1W Wave-Front Analyzer KR-1W Perfection for Professionals : KR-1W Topcon, with its wealth of experience in designing and manufacturing
Visian implantable contact lens versus AcrySof Cachet phakic intraocular lenses: comparison of aberrmetric profiles
Clinical Ophthalmology open access to scientific and medical research Clinical Ophthalmology downloaded from https://www.dovepress.com/ by 13.23.136.75 on 19-Sep-216 Open Access Full Text Article Visian
Surgical data reveals that Q-Factor is important for good surgical outcome
Surgical data reveals that Q-Factor is important for good surgical outcome Michael Mrochen, PhD Michael Bueeler, PhD Tobias Koller, MD Theo Seiler, MD, PhD IROC AG Institut für Refraktive und Ophthalmo-Chirurgie
NIH Public Access Author Manuscript J Refract Surg. Author manuscript; available in PMC 2007 January 8.
NIH Public Access Author Manuscript Published in final edited form as: J Refract Surg. 2005 ; 21(5): S547 S551. Influence of Exposure Time and Pupil Size on a Shack-Hartmann Metric of Forward Scatter William
Implantable collamer lens and femtosecond laser for myopia: comparison using an adaptive optics visual simulator
Original Article Implantable collamer lens and femtosecond laser for myopia: comparison using an adaptive optics visual simulator Lente implantável de collamer e laser de femtosegundo para miopia: comparação
Puntino. Shack-Hartmann wavefront sensor for optimizing telescopes. The software people for optics
Puntino Shack-Hartmann wavefront sensor for optimizing telescopes 1 1. Optimize telescope performance with a powerful set of tools A finely tuned telescope is the key to obtaining deep, high-quality astronomical
CLINICAL SCIENCE INTRODUCTION
CLINICS 2009;64(10):953-60 CLINICAL SCIENCE A CONTRALATERAL EYE STUDY COMPARING APODIZED DIFFRATIVE AND FULL DIFFRATIVE LENSES: WAVEFRONT ANALYSIS AND DISTANCE AND NEAR UNCORRECTED VISUAL ACUITY Marcony
PERSPECTIVE THE PRESENCE OF OPTICAL ABERRATIONS THAT BLUR. Making Sense Out of Wavefront Sensing
PERSPECTIVE Making Sense Out of Wavefront Sensing JAY S. PEPOSE, MD, PHD AND RAYMOND A. APPLEGATE, OD, PHD THE PRESENCE OF OPTICAL ABERRATIONS THAT BLUR retinal images were the subject of popular lectures
Measurement of the Modulation Transfer Function (MTF) of a camera lens. Laboratoire d Enseignement Expérimental (LEnsE)
Measurement of the Modulation Transfer Function (MTF) of a camera lens Aline Vernier, Baptiste Perrin, Thierry Avignon, Jean Augereau, Lionel Jacubowiez Institut d Optique Graduate School Laboratoire d
Disclosures. Opportunities for speciality contact lenses in the multifocal market. Principals. Prof James Wolffsohn
Disclosures Opportunities for speciality contact lenses in the multifocal market Prof James Wolffsohn Ophthalmic Research Group, Life and Health Sciences, Aston University, Birmingham, UK Vested Interest
Aberration Interaction In Wavefront Guided Custom Ablation
Aberration Interaction In Wavefront Guided Custom Ablation Scott M. MacRae MD Professor of Ophthalmology Professor of Visual Science University of Rochester Collaborators and Disclosures: Manoj Subbaram
Evaluate Your Patient s Total Visual System With the OPD-Scan III
0811_OMD553-3MH.qxd:Layout 1 8/9/11 3:08 PM Page 1 SEPTEMBER 2011 Evaluate Your Patient s Total Visual System With the OPD-Scan III New features benefit you and your patients Complete data for IOL selection
EXCHANGE. Financial Disclosure. Clinical pearls In advanced anterior segment surgery being able to do a IOL exchange is a must. Why Do an Exchange
Financial Disclosure D. Ayres, MD Cornea Service IOLBrandon Wills Eye Hospital EXCHANGE Alcon Allergan AMO Bausch and Lomb TearScience BioTissue Why Do an Exchange Refractive surprise after cataract surgery
The Impact of New Generation Aspherical Soft Contact Lenses on Quality of Vision: A Comparison with Spherical Contact Lenses and Spectacle Correction
Deniz Oral, Maryo C. Kohen, Melda Yenerel, Ebru Gorgun, Sule Ziylan, Ferda Ciftci Yeditepe University Faculty of Medicine, Department of Ophthalmology, Istanbul Introduction The correction of higher order
Clinical Trials of Exterior Non Implanted Interference-Based Extended Depth of Focus Intra Ocular Lens Design
Photonics 2014, 1, 296-302; doi:10.3390/photonics1040296 Communication OPN ACCSS photonics ISSN 2304-6732 www.mdpi.com/journal/photonics Clinical Trials of xterior Non Implanted Interference-Based xtended
Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation
J CATARACT REFRACT SURG - VOL 33, FEBRUARY 27 Quality of vision after cataract surgery after Tecnis Z9 intraocular lens implantation Effect of contrast sensitivity and wavefront aberration improvements
Multifocal and Accommodative
What is an IOL? An intraocular lens (or IOL) is a tiny, artificial lens for the eye. It replaces the eye's natural lens. Retina Cornea Lens Macula The eye's normally clear lens bends (refracts) light rays
Fitting Manual Use with kerasofttraining.com
Fitting Manual Use with Fitting Manual: Contents This fitting manual is best used in conjunction with KeraSoft IC online training. To register, please visit www. 01 Kerasoft IC Design - Outlines the KeraSoft
Accommodation with higher-order monochromatic aberrations corrected with adaptive optics
Chen et al. Vol. 23, No. 1/ January 2006/ J. Opt. Soc. Am. A 1 Accommodation with higher-order monochromatic aberrations corrected with adaptive optics Li Chen Center for Visual Science, University of
Exercises Advanced Optical Design Part 5 Solutions
2014-12-09 Manuel Tessmer M.Tessmer@uni-jena.dee Minyi Zhong minyi.zhong@uni-jena.de Herbert Gross herbert.gross@uni-jena.de Friedrich Schiller University Jena Institute of Applied Physics Albert-Einstein-Str.
In recent years there has been an explosion of
Line of Sight and Alternative Representations of Aberrations of the Eye Stanley A. Klein, PhD; Daniel D. Garcia, PhD ABSTRACT Several methods for representing pupil plane aberrations based on wavefront
Impact of scatter on double-pass image quality and contrast sensitivity measured with a single instrument
Impact of scatter on double-pass image quality and contrast sensitivity measured with a single instrument Juan M. Bueno, Guillermo Pérez, Antonio Benito and Pablo Artal * Laboratorio de Óptica, Instituto
EDoF IOL. ZEISS AT LARA 829MP Next generation Extended Depth of Focus Intraocular Lens. NEW EDoF IOL from ZEISS
EDoF IOL Next generation Extended Depth of Focus Intraocular Lens NEW EDoF IOL from ZEISS Introducing the next generation EDoF IOL with the widest range of focus.* ZEISS AT LARA The new premium lens from
Optical Profile Following High Hyperopia Correction With a 500-Hz Excimer Laser System
ORIGINAL ARTICLE Optical Profile Following High Hyperopia Correction With a 5-Hz Excimer Laser System Ana B. Plaza-Puche, MSc; Amr El Aswad, MD; Samuel Arba-Mosquera, PhD; Dominika Wróbel-Dudzinska, MD;
Optics of the crystalline lens and accommodative response
Basic Optics Course, Maastricht 2017 Optics of the crystalline lens and accommodative response Rafael Navarro* *No financial interest 1. Optics of the lens Biconvex lens with complex inner structure Simulation
Confocal Imaging Through Scattering Media with a Volume Holographic Filter
Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,
Fitting Manual Use with
Fitting Manual Use with The KeraSoft IC Lens for and Other Irregular Corneas The KeraSoft IC is a front surface asphere or aspheric toric prism ballasted lens with balanced overall thickness and wavefront
Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens
Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens George Curatu a, Brent Binkley a, David Tinch a, and Costin Curatu b a LightPath Technologies, 2603
Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA
Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Abstract: Speckle interferometry (SI) has become a complete technique over the past couple of years and is widely used in many branches of
Pablo Artal. Adaptive Optics visual simulator ( and depth of focus) LABORATORIO DE OPTICA UNIVERSIDAD DE MURCIA, SPAIN
Adaptive Optics visual simulator ( and depth of focus) Pablo Artal LABORATORIO DE OPTICA UNIVERSIDAD DE MURCIA, SPAIN 8th International Wavefront Congress, Santa Fe, USA, February New LO UM building! Diego
The Eye as an Optical Instrument Pablo Artal
285 12 The Eye as an Optical Instrument Pablo Artal 12.1 Introduction 286 12.2 The Anatomy of the Eye 288 12.3 The Quality of the Retinal Image 290 12.4 Peripheral Optics 294 12.5 Conclusions 295 References
Digital Wavefront Sensors Measure Aberrations in Eyes
Contact: Igor Lyuboshenko contact@phaseview.com Internet: www.phaseview.com Digital Measure Aberrations in Eyes 1 in Ophthalmology...2 2 Analogue...3 3 Digital...5 Figures: Figure 1. Major technology nodes
Multifocal IOL Basics
Multifocal IOL Basics Types of Multifocal Designs REFRACTIVE Concentric zones: M-flex (Rayner) Segmented: LENTIS Mplus (Oculentis) DIFFRACTIVE Progressive: 677MY (Medicontur) Bifocal: IQ AcrySof ReSTOR
CATARACT SURGERY AND DEPTH OF FIELD (D.O.F.)
Prof.Paolo Vinciguerra, M.D. 1, 2 Antonio Calossi 4 Riccardo Vinciguerra, M.D. 1-3 1 Humanitas University 1 Humanitas Clinical and Research Center IRCS 2 Columbus, Ohio State University 3 University of
UNIVERSIDAD COMPLUTENSE DE MADRID
UNIVERSIDAD COMPLUTENSE DE MADRID FACULTAD DE OPTICA Y OPTOMETRÍA Departamento de Óptica TESIS DOCTORAL Vision under manipulated aberrations : towards improved multifocal corrections MEMORIA PARA OPTAR
Visual and Optical Performances of Multifocal Intraocular Lenses with Three Different Near Additions: 6-Month Follow-Up
Send Orders for Reprints to reprints@benthamscience.ae The Open Ophthalmology Journal, 2015, 9, 1-7 1 Open Access Visual and Optical Performances of Multifocal Intraocular Lenses with Three Different Near
1.1 Singlet. Solution. a) Starting setup: The two radii and the image distance is chosen as variable.
1 1.1 Singlet Optimize a single lens with the data λ = 546.07 nm, object in the distance 100 mm from the lens on axis only, focal length f = 45 mm and numerical aperture NA = 0.07 in the object space.